Optimal acquisition parameter selection for CT simulators in radiation oncology

Similar documents
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Purchasing a cardiac CT scanner: What the radiologist needs to know

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Multi-slice Helical CT Scanning of the Chest

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Assessing Radiation Dose: How to Do It Right

How To Improve Your Ct Image Quality

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

Protocol Management and Review Strategies The MD Anderson Experience

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

American College of Radiology CT Accreditation Program. Testing Instructions

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

Practical exercise: Effective dose estimate in CT

Copyright March 1, 2016 by AAPM. All rights reserved.

Low-dose CT for Pulmonary Embolism

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

Physics testing of image detectors

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

Key words: treatment planning, quality assurance, 3D treatment planning

Prepublication Requirements

SITE IMAGING MANUAL ACRIN 6698

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

WHERE IN THE WORLD JILL LIPOTI?

The Challenge of CT Dose Records

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Quantitative Imaging In Clinical Trials Using PET/CT: Update

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

IAEA HUMAN HEALTH SERIES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Diagnostic Exposure Tracking in the Medical Record

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

Gated Radiotherapy for Lung Cancer

Techniques and Applications of Automatic Tube Current Modulation for CT 1

Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan

Quality Assurance of Radiotherapy Equipment

Proton tracking for medical imaging and dosimetry

State-of-the-Art Technology in Cardiac CT

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

AAPM Medical Physics Practice Guideline 1.a: CT Protocol Management and Review Practice Guideline

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Comparison of Epson scanner quality for radiochromic film evaluation

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

DICOM Correction Item

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

Brilliance CT 64-channel confi guration

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

CBCT for Prone Breast. Todd Jenkins, MS, DABR Nash Cancer Treatment Center

THE DOSIMETRIC EFFECTS OF

GE Healthcare. Great treasures in small places. BrightSpeed Elite

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.

A d v a n c e d 1 2 8

Cardiac CT for Calcium Scoring

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

INTRODUCTION. A. Purpose

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

Intensity-Modulated Radiation Therapy (IMRT)

First Three Years After Project Proton Therapy Facility:

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

HITACHI Proton Beam Therapy System

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS


3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Gestión global de la dosis en TC. Sistema de registro y gestión

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Optimal acquisition parameter selection for CT simulators in radiation oncology Ruijie Rachel Liu, 1,a Karl L Prado, 1 and Dianna Cody 2 Departments of Radiation Physics 1 and Imaging Physics 2,The University of Texas M. D. Anderson Cancer Center, Houston, TX, U.S.A. ruijieliu@hotmail.com Received 2 May 2008; accepted 12 June 2008 The purpose of this study was to identify optimal CT acquisition parameter settings for each make and model of scanners used in a large Radiation Oncology (RO) Department, considering the special requirements of CT simulation. Two CT phantoms were used to evaluate the image quality of the five different multichannel CT scanners using helical scan mode. We compared the effects of various pitch, detector configurations, and rotation time parameters on image artifacts, and on spatial and contrast resolution. We found that helical artifact was closely related to pitch and detector configuration settings. This artifact was scannerspecific and generally more obvious when the channel width or detector collimation was equal to the image thickness. Different acquisition parameter settings produced slight differences in observed high- and low-contrast resolution. Short rotation time degraded image quality for certain scanners, but only slightly, while other rotation times, such as 0.75 sec/rotation and above, had no obvious effect on resolution. An optimized combination of acquisition parameters was determined for each scanner make and model, based on phantom image quality and other considerations for clinical applications. This information may be directly useful for physicists whose CT simulation scanners match one of the five examined in this study. If not, the strategy reported here may be used as a guide to perform a similar evaluation of the scanner. Key words: CT protocol, helical artifact, pitch, detector configuration, multichannel CT PACS numbers: 87.57.Ce, 87.57.qp, 87.55.Gh, 87.57.C- I. INTRODUCTION Multi-channel CT has been widely accepted for diagnostic applications as well as for radiotherapy virtual simulation purposes. The effect of CT acquisition parameters on image quality is important to all clinical applications, though the goals of the applications are quite different. A few studies (1-6) regarding optimizing acquisition parameters for certain applications in diagnostic imaging (DI) using multi-channel CT scanners partially addressed the effects of machine settings on image quality. However, a more thorough study is needed regarding image quality for newer scanners for diagnostic and therapeutic applications. CT technology continues to develop and there is a vast difference in acquisition parameter options between manufacturers. a Corresponding author: Ruijie Rachel Liu, Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, TX, U.S.A.; phone: 713-218-8899; fax: 713-218-8899; email: ruijieliu@hotmail.com 151 151

152 Liu et al.: Optimal CT simulator settings 152 Though there is some information available regarding technical details, such as image thickness for CT radiotherapy simulation, (7) the optimal CT image acquisition parameter settings for radiotherapy simulation have not yet been thoroughly investigated. The purpose of this study was to address this gap, and determine the optimal image acquisition parameter settings for conventional CT simulation protocols, including pitch (defined as couch movement per gantry rotation divided by beam width), detector configuration, and rotation time using several different CT scanners in our radiation oncology (RO) department. Routine radiation dose values were also collected in order to help understand image quality characteristics between scanners. II. MATERIALS AND METHODS A. Rationale for CT Simulation in Radiation Oncology The implementation of CT simulators in radiation oncology (RO) is somewhat different from CT scanners in DI, though image quality is a main concern for both. Unlike the regular practice in DI, breath holding, contrast injection, variable image thickness, and gantry tilt are generally discouraged in RO. Because most patients undergoing CT simulation in RO have diagnosed cancers, and generally only one scan is used for CT simulation, radiation dose is of less concern than it is in DI. Instead, image quality has singular importance, as accuracy in RO treatment planning is particularly vulnerable to artifacts. In order to simulate radiation treatment conditions, free or gated breathing is required during the CT scan. Fast scanning is generally not desirable because a snapshot of one breathing phase does not represent the average anatomy position during treatment delivery. A scan time of 1 to 2 minutes is considered acceptable. For CT simulation, patient skin surface must be included in the image for treatment planning. Image truncation is intolerable; therefore a larger display field of view (DFOV) is used, usually 50 cm or 65 cm. For a head scan, a body size DFOV is generally used, because part of the shoulders must be scanned as part of patient positioning for most procedures. For CT simulation, the choice of image thickness involves balancing the conflicting considerations of the increased resolution of digitally reconstructed radiographs (DRR) obtained from thin images and increased physician effort required to contour multiple images. The standard choice of the physicians and physicists in our institution is 2.5 mm or 3 mm image thickness and 2.5 mm interval, which agrees with general recommendations. (7) B. Phantoms and Scanners A Catphan phantom (8) (The Phantom Laboratory, Salem, New York) was primarily used to evaluate image quality. Three modules were used in this study: 1) CTP528, high-contrast resolution bar patterns; 2) CTP515, 1.0% low-contrast objects in the Supra Slice portion; 3) CTP401, artifact evaluation. The high-contrast resolution section of the GE CT daily QA phantom (9) (GE Medical Systems, Milwaukee, Wisconsin) was used for additional artifact evaluation. The high contrast region was chosen to evaluate artifacts because objects with high-edge gradients are more likely to produce helical artifacts. Both phantoms were positioned by suspending each from an appropriate bracket, following the manufacturers standard setup recommendations. Five different CT scanners were evaluated: General Electric (GE) LightSpeed (LS) RT (4-channel, 80 cm bore) LS RT16 (16-channel, 80 cm bore) LS16 (16-channel, 70 cm bore) Philips Medical Systems AcQsim CT (1-channel, 80 cm bore) MX8000 (16-channel, 70 cm bore).

153 Liu et al.: Optimal CT simulator settings 153 C. Evaluation Methods Various pitch, detector configuration, and gantry rotation time (1 sec and less) settings were used to acquire images, while maintaining the following constant settings: 120 kvp, 240 effective mas (defined as tube current (ma) rotation time (sec) / pitch; where pitch = table movement per gantry rotation / total beam collimation), 1s rotation time unless specified otherwise, 25 cm display field of view (DFOV), standard reconstruction filter, 2.5 mm image interval, and 2.5 mm image thickness for GE or 3 mm image thickness for Philips. On multi-channel CT scanners, there are multiple data channels that route the signals collected on the detector surface to the reconstruction computer. These data channels can be utilized in a flexible manner to produce more than one thin image during one rotation of the gantry, or fewer thicker images during one rotation of the gantry. The combination of the number of active channels and the detector width (z-direction) assigned to each channel is called the detector configuration. The product of these two values represents the nominal x-ray beam width (z-direction). In general, a combination of fine detector spacing (also called small channel width) and a wide beam width are the most desirable in CT imaging. Fine detector spacing produces a denser helical dataset from which to interpolate, and a wide beam allows for higher table speeds and faster exams. The phantom images were evaluated from three aspects: 1) high-contrast resolution; 2) lowcontrast resolution; and 3) the presence of helical artifacts. A single observer primarily evaluated all the images. Zoom, display window and level were adjusted to best display the image region of interest for high-contrast and low-contrast resolution evaluation. For low-contrast resolution, since CTP 515 module spans 70 mm, the images with best resolution were selected for evaluation. Three levels were used to score the presence of artifact: Yes, Slight, and No. Yes meant the artifact was clearly visible, No meant no artifact was present, and Slight meant the artifact was present in the phantom images, but it was very subtle and most likely invisible in clinical images. To examine the effect of gantry rotation time, 1 sec and sub-second rotations were evaluated for scanners that had these options. The fastest gantry rotation for GE RT and Philips AcQsim was 1 sec. Therefore they were not included in the evaluation of gantry rotation time effect. The single-channel Philips AcQsim CT is limited by prolonged scan time and the effects of excessive tube heating, so only pitches greater than 1 were evaluated for this single-channel CT. In order to interpret results such as low-contrast resolution, the standard annual radiation dose assessment made for each scanner was also reported on a normalized (per 100 mas) basis. These dose evaluations were obtained by using the standard CT Dose Index (CTDI) approach. (10) III. RESULTS A. Artifact evaluation For all CT scanners, helical artifacts were more severe when the channel width was equal to the image thickness and a relatively high pitch was used. For example, for the GE LS16 (Fig.1), a detector configuration of 8 2.5 mm and a pitch of 1.675 resulted in 2.5 mm thick images with severe artifacts (Fig.1a), while a 16 1.25 mm, 1.75 pitch setting showed slight artifacts (Fig.1b), and a 16 1.25 mm, 0.562 pitch had no artifacts in the images (Fig.1c). For GE scanners, a channel width setting equal to half of the image thickness was sufficient to produce artifact-free images for pitch values close to 1. For GE LS RT, 4 1.25 mm and a pitch of 0.75 (Fig. 2a) offered the best image quality for 2.5 mm thick images, while 4 2.5 mm and 1.5 pitch (Fig. 2d) settings caused severe helical artifacts. There was no substantial difference in the severity of artifacts for the settings 4 1.25 mm and pitch 1.5 (Fig. 2b), and 4 2.5mm and pitch 0.75 (Fig. 2c), though the former produced better low-contrast resolution.

154 Liu et al.: Optimal CT simulator settings 154 (a) (b) (c) FIG. 1. Artifacts in 2.5 mm thick images of the Daily QA phantom for GE LS16: (a) 8 2.5 mm, pitch 1.675; (b) 16 1.25 mm, pitch 1.75; (c) 16 1.25 mm, pitch 0.562. (a) (b) (c) (d) FIG. 2. Artifacts in 2.5 mm thick images of the Catphan phantom for GE LS RT: (a) 4 1.25 mm, pitch 0.75; (b) 4 1.25 mm, pitch 1.5; (c) 4 2.5 mm, pitch 0.75; and (d) 4 2.5 mm, pitch 1.5. The Philips CT scanners behaved quite differently from the GE CT scanners (Fig. 3). For the Philips MX8000, 16 1.5 mm detector configuration for 3 mm image thickness setting (i.e., a channel width equal to half image thickness) caused severe artifacts for all available pitches (Fig. 3d). A channel width of 1/4 of the image thickness with pitch less than 1 was needed to eliminate helical artifacts (Fig. 3a) on this scanner. The other two settings, 16 0.75mm, pitch

155 Liu et al.: Optimal CT simulator settings 155 1.438 (Fig. 3b) and 16 1.5mm, pitch 0.688 (Fig. 3c), resulted in less severe and different artifacts; the setting for Fig. 3c were better than those for Fig. 3b. For Philips AcQsim CT, helical artifacts increased with pitch (Fig. 4). Helical artifacts appeared to be scanner/model specific, not only different across manufacturers, but also different across models made by a single manufacturer. In Fig.5, we compared the images from GE LS RT16 (Fig. 5a, 5b) with those from GE LS16 (Fig. 5c). For the same pitch (1.375) and detector configuration (16 1.25 mm), the GE LS RT16 produced more severe artifacts than the GE LS16 (Fig. 5b, 5c). The combinations of the largest channel width and largest pitch that produced artifact-free images for all scanners are listed in Table 1. The helical streak artifact appears to depend on the combination of detector configuration and pitch, and that the relationships between these characteristics are not consistent between or even within vendors. The severity of artifacts increased with increasing pitch. The artifact evaluation in Tables 2, 3 and 4 based on the high-contrast section of the Catphan phantom gives additional information for specific scanner settings. (a) (b) (c) (d) FIG. 3. Artifact evaluation of 3 mm thick images of the Catphan phantom for Philips MX8000: (a) 16 0.75 mm, pitch 0.688; (b) 16 0.75 mm, pitch 1.438; (c) 16 1.5 mm, pitch 0.688; (d) 16 1.5 mm, pitch 1.313. Note: Windows and levels were adjusted separately for optimal display.

156 Liu et al.: Optimal CT simulator settings 156 (a) (b) FIG. 4. Artifact evaluation of 3 mm thick images of the Catphan phantom for Philips AcQsim single-channel CT. (a) pitch 1.0; (b) pitch 1.5. (a) (b) (c) FIG. 5. GE 16-channel CT comparison, 2.5 mm thick images, for detector configuration: 16 1.25 mm. GE LS RT16, pitch 0.562 (a), pitch 1.375 (b), and GE LS16, pitch 1.375 (c). TABLE 1. Combination of the largest channel width and largest pitch that produced artifact-free images. Model Image Pitch Detector Head CTDI W / Body CTDI W / thickness configuration 100mAs 100mAs (mm) (mm) (mgy) (mgy) Philips MX8000 3 0.688 16 0.75 = 12 13.2 4.2 Philips AcQsim 3 1 1 3.00 = 3 13.0 3.5 GE LS16 2.5 1.375 16 1.25 = 20 14.4 7.3 GE LS RT16 2.5 0.938 16 1.25 = 20 12.4 6.8 GE LS RT 2.5 0.75 4 1.25 = 5 23.8 10.4 2.5 0.75 4 2.5 = 10* 17.7 6.4 * Not artifact-free for helical scan, but for dose comparison.

157 Liu et al.: Optimal CT simulator settings 157 TABLE 2. Image quality results for GE LS RT16. Rotation Mode / Pitch High Contrast 1% Low Contrast Artifact Present? Time (s) Resolution (lp/cm) Diameter (mm) 0.5 Axial 7 4 No 1.0 Axial 7 4.5 No 0.5 0.562 7 4 No 0.5 0.938 7 4 No 0.8 0.938 8 4 No 0.5 1.375 7 4.5 Yes 0.8 1.375 7 4 Yes Technique: effective mas = 240, image thickness 2.5 mm, detector configuration 16 1.25 mm. Low contrast was defined as the diameter of the smallest circular target detectable. TABLE 3. Image quality results for GE LS16. Rotation Pitch High Contrast 1% Low Contrast Artifact Time (s) Resolution (lp/cm) Diameter (mm) Present? 0.5 0.562 8 3.5 No 0.8 0.562 8 4 No 0.8 0.938 7 4 No 0.8 1.375 7.5 3.5 Slight 1 1.75 7 3 Yes Technique: effective mas = 240, image thickness 2.5 mm, detector configuration 16 1.25 mm. Low contrast was defined as the diameter of the smallest circular target detectable. B. Resolution evaluation Image resolution (low and high contrast) was evaluated using the phantom sections shown in Fig. 6. For these GE scanners (Tables 2 and 3), the image resolution did not appear to be affected by pitch and rotation time, provided that ma was adjusted appropriately. For high-contrast resolution, the variation was within 1 line pair (lp) /cm around the average value of 7.3 lp/cm. For low contrast resolution, the variation was within 0.6 mm over the average diameter of 3.9 mm for 1% detectable low-contrast objects. These variations may be due to quantum and statistical (a) (b) FIG. 6. Sample image sections used for resolution evaluation: (a) Catphan CTP515 mode, the marked area is 1% contrast which was used for low-contrast resolution evaluation; (b) Catphan CTP528 mode, use for high-contrast resolution evaluation. Techniques: GE LightSpeed RT16, pitch 0.938, rotation time 0.8 sec, 16 1.25 mm detector configuration, 2.5 mm image thickness (display not optimized).

158 Liu et al.: Optimal CT simulator settings 158 fluctuations. However, GE LS16 had systematically better high- and low-contrast resolution than GE LS RT16. GE LS16 has a maximum tube current limit of 440 ma, thus a pitch of less than 1 had to be used to achieve 240 effective mas for 0.5s gantry rotation. Of these two GE models, the larger bore CT scanner had lower image quality and was more prone to helical artifacts. A similar approach was used for images obtained using Philips MX8000 (Table 4). A clear trend was observed that short rotation time (0.5 sec or less) caused slight low-contrast resolution degradation, though high-contrast resolution remained unaffected. The image resolution was best for Philips MX8000, followed by GE LS16, and GE RT16. TABLE 4. Image quality results for Philips MX8000. Rotation Pitch High Contrast 1% Low Contrast Artifact Time (s) Resolution (lp/cm) Diameter (mm) Present? 0.4 0.564 8 4 No 0.5 0.563 8 3.5 No 0.75 0.563 8 3 No 1.5 0.563 8 3 No Technique: effective mas = 240, image thickness 3 mm, detector configuration 16 0.75 mm. Low contrast was defined as the diameter of the smallest circular target detectable. C. Radiation Dosimetry The CTDIW was normalized for 100 mas using the standard head and body phantoms (11) and represents the most recent annual dosimetry evaluations. CTDIW / 100 mas varied by nearly a factor of 2 from 12.4 mgy to 23.8 mgy among the five scanners for the head phantom. For the head CTDI phantom, normalized dose (CTDIW / 100 mas) was highest for GE RT, followed by GE LS16, Philips MX8000, and Philips AcQsim, and was lowest for GE LS RT16. For the body CTDI phantom, CTDIW / 100 mas ranged from 3.5-10.4mGy; this value was highest for GE LS RT, followed by GE LS16, GE LS RT16, and Philips MX8000, and was lowest for Philips AcQsim. For this study, the head CTDI dose results would be more relevant than the body CTDI values. The normalized body CTDI results were included for completeness. IV. DISCUSSION We found that for GE scanners, a channel width setting equal to half of the image thickness was sufficient to produce artifact-free and good quality images for pitch values close to 1. Our result with GE RT 4-channel CT agrees with Takahashi, et al. (3) who determined that for this scanner model, the settings that achieved the best IQ were found to be 4 1.25 mm detector configuration and a pitch of 0.75 (though that study was only focused on depicting arterial stenosis, and imaging artifact was not specifically evaluated). For GE RT, the detector setting 4 1.25 mm resulted in a high CTDIW of 2.38 cgy (Table 1), while 4 2.5 mm setting lowered CTDIW to 1.77cGy due to improved dose efficiency. If a faster scan speed is needed, we would suggest using 4 y2.5 mm and pitch 0.75 rather than 4 1.25 mm and pitch 1.5, though they have the same table speed and similar image quality. For Philips MX8000, a channel width of 1/4 of the image thickness with a pitch of less than 1 was needed to eliminate helical artifacts. However, the MX8000 had overall somewhat better high- and low-contrast resolution than GE scanners (Tables 2, 3, 4). Differences between these two manufacturers may be due to the different reconstruction algorithms or different dose values (matching technique parameters does not necessarily result in equivalent radiation output or quality between scanners). We observed a larger than expected difference in normalized dose

159 Liu et al.: Optimal CT simulator settings 159 (CTDIW / 100 mas) among these scanners; when using the head CTDI phantom, there was a range of a factor of 2 in CTDIW / 100 mas. When using the body CTDI phantom, we observed a difference between vendors by a factor of 3. In general, when spatial resolution was enhanced, artifacts present in the image were also enhanced. In helical CT, data re-binning and interpolation are used to generate planar transverse images from helical attenuation data. This process yields better quality images if helical data is more finely spaced. The choice of detector configuration is generally a much more powerful method of reducing helical artifact than decreasing scan pitch. Sometimes, the selection of the maximum number of active channels and the finest channel width must be balanced against the need for a faster couch speed. The small differences in spatial resolution shown in Tables 2 to 4 are not likely to be detectable in clinical images, because the DFOV of 50 cm or greater over the 512 512 matrix used in CT simulation limits the achievable spatial resolution. Some difference in contrast resolution is expected due to the different bore sizes, detector designs and reconstruction algorithms. Increasing exposure, or effective mas, will increase low-contrast resolution across all scanners. Though phantom studies are appropriate for comparative evaluations, they are not accurate simulations of patient scans. A major difference between a phantom and a patient is that a patient often moves (voluntarily or involuntarily) during a CT acquisition, especially for those of the chest and abdomen. The resulting blur is not present on the stationary phantom images. Faster rotation time reduces motion-blurring artifact (12) and improves patient image quality, though it should not affect the phantom image quality. Because fast scan time (breath-hold) is not desired in CT simulation, a reasonably low pitch, fast rotation and fine detector configuration should be used to improve image quality and also to acquire patient images in a more averaged position rather than a snapshot of one moment. In general, we recommend using fast gantry rotation for chest and abdomen scans, 0.5-0.8 sec for example, if the options are available. Image resolution evaluation can be used as a reference when setting up scan protocols for radiotherapy simulation. Tube current modulation should not affect the appearance of helical artifacts. However, tube current modulation may affect low-contrast resolution, because effective mas will vary in the x/y plane, along the axis, or both. V. CONCLUSIONS In conclusion, we have quantified the effects of changes in various parameters used for RO CT simulation on specific image metrics. Given the differences we found between CT machines in our own study, choosing optimal image acquisition parameter settings for CT simulation purposes may require a phantom evaluation for each scanner model to be used. REFERENCES 1. Gupta AK, Nelson RC, Johnson GA, Paulson EK, Delong DM, Yoshizumi TT. Optimization of eight-element multidetector row helical CT technology for evaluation of the abdomen. Radiology 2003;227(3):739-745. 2. Schindera ST, Nelson RC, Lee ER, DeLong DM, et al. Abdominal Multislice CT for Obese Patients: Effect on Image Quality and Radiation Dose in a Phantom Study. Acad Radiol. 2007;14(4):486-494. 3. Takahashi S, Takamura M, Yamamoto S, et al. Technical optimization of four-channel multidetector-row helical computed tomography for depicting arterial stenosis: A phantom study. Acta Radiol. 2007;48:173-179. 4. Tsili AC, Efremidis SC, Kalef-Ezra J, et al. Multi-detector row CT urography on a 16-row CT scanner in the evaluation of urothelial tumors, Eur Radiol. 2007;17(4):1046-1054. 5. Kulama E. Scanning protocols for multislice CT scanners. Br J Radiol. 2004;77:S2 S9. 6. Power NP, Pryor MD, Martin A, Horrocks J, McLean AM, Reznek RH. Optimization of scanning parameters for CT colonography. Br J Radiol. 2002;75:401-408. 7. Aird EGA, Conway J. CT simulation for radiotherapy treatment planning. Br J of Radiol. 2002;75:937-949. 8. Catphan Manual, The Phantom Laboratory Inc., Salem, New York, 2002. 9. LightSpeed 5.X Technical Reference Manual, GE Medical Systems, Milwaukee, Wisconsin, 2003.

160 Liu et al.: Optimal CT simulator settings 160 10. McNitt-Gray MF. AAPM/RSNA Physics Tutorial for Residents: Topics on CT - Radiation Dose in CT. Radiographics. 2002;22:1541-1553. 11. McCollough C, Cody D, Edyvean S, et al. The measurement, reporting, and management of radiation dose in CT: report of Task Group No. 23 of the Diagnostic Imaging Council CT Committee of the American Association of Physicists in Medicine. College Park (MD): AAPM 2008. 12. Gagné IM, Robinson DM. The impact of tumor motion upon CT image integrity and target delineation. Med Phys. 2004;31(12):3378-3392.