10.2 Fourier Transform Infrared Spectroscopy

Similar documents
FTIR Instrumentation

Introduction to Fourier Transform Infrared Spectrometry

Experiment 5. Lasers and laser mode structure

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Synthetic Sensing: Proximity / Distance Sensors

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

where h = J s

ORIEL. FT-IR Spectroscopy SECTION FOUR FEATURES. Glossary of Terms Introduction to FT-IR Spectroscopy

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p

Time out states and transitions

Infrared Spectroscopy: Theory

E190Q Lecture 5 Autonomous Robot Navigation

Today. next two weeks

Raman spectroscopy Lecture

Module 13 : Measurements on Fiber Optic Systems

Blackbody Radiation References INTRODUCTION

Experiment #5: Qualitative Absorption Spectroscopy

Lab 9: The Acousto-Optic Effect

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

PHYS 331: Junior Physics Laboratory I Notes on Noise Reduction

Preview of Period 3: Electromagnetic Waves Radiant Energy II

Data Transmission. Data Communications Model. CSE 3461 / 5461: Computer Networking & Internet Technologies. Presentation B

The Calculation of G rms

APPLICATION NOTE AP050830

Determination of Molecular Structure by MOLECULAR SPECTROSCOPY

A Guide to Acousto-Optic Modulators

Infrared Spectroscopy 紅 外 線 光 譜 儀

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Organic Chemistry Tenth Edition

Acoustics: the study of sound waves

Modern Classical Optics

Signal to Noise Instrumental Excel Assignment

INFRARED PARTS MANUAL

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

Projects. Objective To gain hands-on design and measurement experience with real-world applications. Contents

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

Symmetric Stretch: allows molecule to move through space

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

Raman Scattering Theory David W. Hahn Department of Mechanical and Aerospace Engineering University of Florida

- thus, the total number of atoms per second that absorb a photon is

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)

Optical Communications

Lock - in Amplifier and Applications

Uses of Derivative Spectroscopy

Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range

It has long been a goal to achieve higher spatial resolution in optical imaging and

Waves Sound and Light

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

13C NMR Spectroscopy

Raman Spectroscopy Basics

Robot Perception Continued

Using the Spectrophotometer

AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

5.33 Lecture Notes: Introduction to Spectroscopy

Laser Audio Surveillance Device

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.

SIGNAL PROCESSING FOR EFFECTIVE VIBRATION ANALYSIS

Aliasing, Image Sampling and Reconstruction

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Solar Energy. Outline. Solar radiation. What is light?-- Electromagnetic Radiation. Light - Electromagnetic wave spectrum. Electromagnetic Radiation

INFRARED SPECTROSCOPY (IR)

The front end of the receiver performs the frequency translation, channel selection and amplification of the signal.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

Does Quantum Mechanics Make Sense? Size

Antennas & Propagation. CS 6710 Spring 2010 Rajmohan Rajaraman

Raman Spectroscopy. 1. Introduction. 2. More on Raman Scattering. " scattered. " incident

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

Selecting Receiving Antennas for Radio Tracking

Experiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))

Hydrogen Bonds in Water-Methanol Mixture

A down-under undergraduate optics and photonics laboratory

AS COMPETITION PAPER 2008

An Introduction to the MTG-IRS Mission

Programmable-Gain Transimpedance Amplifiers Maximize Dynamic Range in Spectroscopy Systems

Interference. Physics 102 Workshop #3. General Instructions

ENGINEERING METROLOGY

THE FIBRE-OPTICS TRAINER MANUAL

A More Efficient Way to De-shelve 137 Ba +

Trigonometric functions and sound

Using light scattering method to find The surface tension of water

Application Note AN1

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques

Limiting factors in fiber optic transmissions

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Treasure Hunt. Lecture 2 How does Light Interact with the Environment? EMR Principles and Properties. EMR and Remote Sensing

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

2. Molecular stucture/basic

Polarization of Light

Short-time FFT, Multi-taper analysis & Filtering in SPM12

Copyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley.

Transcription:

10.2 : 1/22 10.2 Fourier Transform Infrared Spectroscopy overview of the infrared including measurement difficulties spectrometer throughput instead of a t f transform use an x ν transform introduction to the Michelson interferometer, and its behavior as an optical autocorrelator block diagram of an FTIR spectrometer, including details of mirror movement and controlling when the ADC takes data example far infrared spectrum of K 2 ReCl 6 optical resolution including an example of gas-phase ammonia apodization to eliminate sinc function oscillations overall signal-to-noise enhancement that can be expected historic types of measurements that drove the development of FTIR

Overview of the Infrared The highest frequency vibration is the H 2 stretching motion which occurs at 4,400 cm -1. Although this transition is not infrared active (it is Raman active), ~4,000 cm -1 is usually taken to be the upper limit of the infrared. The lower limit is determined by instrumental considerations and is often near 50 cm -1. The mid infrared extends from 4,000 to 200 cm -1. This region is usually composed of characteristic bond frequencies or normal modes of molecules. The far infrared extends from 200 to 50 cm -1. This region is usually composed of characteristic metal bond frequencies, or normal modes of large molecules. The end of the range overlaps with high energy rotations. The resolution of spectra depends upon their use. Solution phase spectra are usually taken with bandwidths of 5-10 cm -1, while gas phase spectra of small molecules usually require a bandwidth of 0.1 cm -1. 10.2 : 2/22

Measurement Difficulties infrared photons are difficult to distinguish from thermal noise, with the result that the measurement is detector noise limited P N = 4kT f for f = 1 Hz and T = 298 K P N = 828 cm -1 infrared transitions have very low molar absorptivities, ε ~ 1 L mol -1 cm -1, requiring high concentrations to obtain a reasonable level of absorption high resolution spectrometers have a very low optical throughput, which limits the signal-to-noise ratio of the absorption measurement the experiment would benefit enormously from multiplexed detection 10.2 : 3/22

Spectrometer Optical Throughput The standard measure of optical collection efficiency is the f/# (pronounced f-number). An f/# is the focal length of a lens or mirror, divided by the diameter of the lens or mirror. Collection efficiency is inversely proportional to the square of the f/#. f/# 1 2 5 10 20 percent 6.2 1.5 0.25 0.06 0.02 A grating or prism bends different wavelengths at different angles. To obtain high resolution the wavelengths are separated by moving as far as possible away from the focusing lens or mirror. That is, long focal lengths are required. It is very difficult to construct a grating or prism monochromator with a low f/# because the size of the grating or prism would need to increase in proportion to the focal length. Large gratings and prisms are either impossible to manufacture or are very expensive. A typical f/# for a 0.1 cm -1 resolution grating spectrometer is 12, which is about a 0.05% throughput. 10.2 : 4/22

Impossibility of a t f Transform To obtain an infrared spectrum via a Fourier transform, the signal would be recorded (digitized) as a function of time to obtain F(t). Then the spectrum would be computed using the following expression, i2πft f F t e dt φ = ( ) ( ) This will not work in the infrared because the frequencies are too fast. An infrared spectrum has frequencies from 1.2 10 14 Hz (4,000 cm -1 ) to 6.0 10 12 Hz (200 cm -1 ). The fastest electronic temporal resolution with commercial instrumentation is 7 10-12 s or ~840-42 times too slow. 10.2 : 5/22

Possibility of an x ν Transform Fortunately, an infrared wave can be written in an alternative form using distance and wave number, where ν = 1/λ. x cos 2π = cos(2 πνx) λ The Fourier transform equation takes the same form, simply exchanging ν for f and x for t. It is numerically solved using the same computer program as that used for t to f transforms. i2 ( ) F( xe ) φν = The infrared spectral range from 4,000 cm -1 to 250 cm -1 corresponds to wavelengths of 2.5 µm to 40 µm. These lengths are easy to measure, all one has to do if figure out how to "stop" the wave from moving! πνx dx 10.2 : 6/22

The Michelson Interferometer light from an infrared source is collimated and directed to a beam splitter at the beam splitter, half is reflected toward the fixed mirror and half is transmitted toward the movable mirror when light reaches each mirror it is reflected back to the beam splitter at the beam splitter, half the light from the fixed mirror is reflected back to the source and half is transmitted to the detector source beam splitter fixed mirror detector sample at the beam splitter, half the light from the movable mirror is transmitted back to the source and half is reflected to the detector half the light reaches the detector, 0.25 from each path light of every wavelength is always traveling through the sample to the detector (multiplexed) an interferogram is obtained by moving the mirror and measuring the resultant intensity at the detector movable mirror 10.2 : 7/22

Optical Autocorrelator (1) Light traveling toward the detector can be described as the sum of two cosines, 1 1 cos(2 πν x) + cos(2 πν [ x+ ξ]) 2 2 where ξ is the extra distance (nearer or farther) traveled by light along the path involving the movable mirror. moving the mirror a distance ξ/2 farther from the beam splitter delays light by the distance +ξ. the beam splitter divides intensity, not amplitude. Thus, 1/4 of the total photons reach the detector from each path. This corresponds to 1/2 the amplitude from each path. the detector responds to intensity, which is the square of the electric field. 10.2 : 8/22

Optical Autocorrelator (2) 10.2 : 9/22 Although the detector output is averaged as a function of time, determination of a functional form for the signal requires that integration be over distance. In the FTIR instrument the optical signal (square of the amplitude) is averaged for 100 µs or longer. This corresponds to averaging the waveform over a distance of x 0 = 3 10 4 meters. 1 x x 0 0 0 ( 0.5cos(2 π ) + 0.5cos(2 π ( +ξ)) ) vx v x dx Expansion of the square yields three integrals. x x x 2 0 0 0 0.25 0.25 2 0.50 cos(2 x) dx cos( 2 ( x )) dx cos(2 x)cos( 2 ( x )) dx x πν + πν + ξ + πν πν + ξ x x 0 0 0 0 0 0 By letting x 0, the first two integrals each reduce to 0.125. x 0 0.50 0.125 + 0.125 + cos(2 πνx) cos( 2 πν ( x+ ξ) ) dx x 0 0 2

Optical Autocorrelator (3) The third integral on the previous slide is the autocorrelation function, 1,1 ( ) x 0 1 C1,1 ( ξ ) = cos(2 πνx)cos(2 πν ( x+ ξ)) dx x C 0 0 ξ = 0.5cos(2 πνξ) where Fourier transforms were used to evaluate the integral. The detector output is then given by C 1,1 (ξ) plus the dc offset. 0.25 + 0.25cos(2 πνξ) As the mirror is moved, the output of the detector will be an offset cosine. Each individual wavelength of light will have its own offset cosine. These offset cosines are added together in the detector output. As the mirror is moved the output of the detector is digitized with sufficient resolution to reproduce all optical wavelengths. After collecting the data a Fourier transform is computed to extract the spectrum - amplitude versus ν. 10.2 : 10/22

Instrument Block Diagram source/aperture the aperture controls the maximum possible resolution optical filters minimize aliasing of non-infrared radiation resolution controlling aperture source mirror piston with air bearing movable mirror beam splitter sample chamber has a sample and reference so that transmission can be computed the collimated white light source is used to align the sample and reference with the infrared beam optical filter TGS pyroelectric detector reference mirrors flip 90 deg between sample and reference fixed mirror detector output is electronically amplified electronically low pass filtered digitized at specified mirror displacements from zero sample mirror swings out of the way for alignment collimated white alignment light 10.2 : 11/22

Mirror Movement & Data Collection mirror design supported by a near frictionless air bearing driven back and forth by an electromagnetic solenoid the range of motion is determined by two light emitting diode/photodiode pairs when the position paddle interrupts the travel transducers, the polarity of the electromagnet is changed and the mirror travel direction changes zero delay is about where the "position paddle" is shown helium:neon single-frequency laser @ 0.6328 µm 15,802 cm -1 visible beam splitter made from two prisms photodiode detectors moving visible mirror white light fixed visible mirror piston travel electromagnetic solenoid driver "frictionless" air bearing mirror travel start of travel transducer position indicating paddle end of travel transducer moving infrared mirror infrared beam to infrared beam splitter 10.2 : 12/22

Measuring ξ zero delay determination determined by white light and a visible interferometer constructed on the backside of the moving mirror the interferogram of white light is a "spike" when the delay is zero an electronic circuit monitoring the output of the white light photodiode can easily detect this spike the fixed visible mirror is factory adjusted so that zero delay with the white light beam corresponds to zero delay with the infrared beam in the infrared interferometer distance determination distance is measured by single frequency light from a He:Ne laser the wavelength of the laser is 0.6328 µm, meaning that the output of the photodiode is an offset cosine with a maximum every 0.6328 µm the analog-to-digital converter is set so that it only takes data whenever the He:Ne cosine is at a maximum 10.2 : 13/22

Analog-to-Digital Conversion 10.2 : 14/22 mirror velocity since the velocity of the mirror can be experimentally controlled, the apparent frequency of the cosine can be controlled to match the speed of the detector electronics in this instrument the mirror velocity is adjusted to make the He:Ne laser produce a 5 khz cosine at the detector the highest un-aliased frequency that can be measured by sampling every peak of the cosine is 15,802/2 = 7,901 cm -1 the spectrum then appears at frequencies from 2.5 khz to dc undersampling since the upper end of the infrared spectrum is near 4,000 cm -1, lower digitizing rates can be used - this is called under sampling for the far infrared the data will be under sampled by 16 to produce a highest optical frequency of 493.8 cm -1 corresponding to a 313 Hz for a fixed resolution, under sampling decreases the size of the data set that needs to be numerically processed under sampling permits the use of low pass filters with lower f 3dB frequencies, thus a better SNR

Example Data: Infrared Source The two figures below show the source (reference) interferogram (distance vs. amplitude) and the spectrum (frequency vs. amplitude). 10.2 : 15/22

Example Data: K 2 ReCl 6 The three figures below show the sample interferogram, the sample intensity, and the sample transmission. The very sharp line at ~375 cm -1 is aliasing of an electric interference. The 330 cm -1 band is the Re-Cl stretch and the 180 cm -1 band is a Cl-Re-Cl bend. The peak at 80 cm -1 is an artifact due to low source intensity. 10.2 : 16/22

Optical Resolution Consider a cosine signal. The process of digitizing the interferogram involves starting and stopping, thus the cosine is multiplied by a rectangle with width ξ 0. The result is convolution by a sinc function (with ν 0 = 1/ξ 0 ). 0 ν 8 4 2 1 0.1 ξ 0 (cm) 0.125 0.25 0.5 1 10 mirror (cm) 0.0625 0.125 0.25 0.5 5 The spectral resolution (bandpass) depends only upon how far the mirror is moved! 10.2 : 17/22

10.2 : 18/22 Gas-Phase Ammonia Spectrum (1)

10.2 : 19/22 Gas-Phase Ammonia Spectrum (2)

Apodization effect of abruptly starting and stopping the mirror movement the process of starting and stopping the digitization of the cosine produces a spectrum convolved with a sinc function a sinc function is said to have "feet," and the feet often make spectral interpretation difficult apodization is literally "no feet making," and involves multiplication of the collected data by some function other than a rectangle apodization functions when computers were slow and had little memory, simple functions were used for apodization, i.e. trapezoid and triangle with modern computers any desirable function can be used, e.g. Gaussian or exponential note that apodization will decrease resolution via convolution with the spectral peaks 10.2 : 20/22

Signal-to-Noise Enhancement multiplexing (Fellgett advantage) 0.1 cm -1 resolution: decrease in measurement time: 31,000 = ~ 176 increase in SNR: 5 cm -1 resolution decrease in measurement time: = 5 increase in SNR: 620 = 25 optical throughput (Jacquinot advantage) 0.1 cm -1 resolution scanning f/# = ~12 for ~0.05% throughput interferometer f/# = ~1 for ~6.2% throughput result = 124 more light reaches the detector 5 cm -1 resolution scanning f/# = ~5 for 0.25% throughput result = 25 more light reaches the detector sum of the two advantages 0.1 cm -1 resolution = 21,824 5 cm -1 resolution = 625 3,500 400 0.1 3,500 400 = 31,000 620 10.2 : 21/22

Original Drivers for FTIR low concentrations (large amount of light on the detector, allows precise measurement of transmission) high resolution spectra (large mirror movement) unstable or transient species (fast measurement time) gas chromatography detector (fast measurement time) mixture spectra using Beer's Law (a precise transmission allows a precise absorption) far infrared spectra (interferometer instead of grating - tough to rule appropriate gratings) high sample throughput (fast measurement time) 10.2 : 22/22