Bureau International des Poids et Mesures

Similar documents
APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-energy X-Rays (APMP.RI(I)-K3)

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Production of X-rays and Interactions of X-rays with Matter

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Introduction to Geiger Counters

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Variance reduction techniques used in BEAMnrc

Introduction to the Monte Carlo method

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, Bologna, Italy

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Radiation Detection and Measurement

Radiation Strip Thickness Measurement Systems

GAMMA-RAY SPECTRA REFERENCES

Lectures about XRF (X-Ray Fluorescence)

Coating Thickness and Composition Analysis by Micro-EDXRF

16th International Toki Conference on Advanced Imaging and Plasma Diagnostics

Efficiency calibration of the Ge(Li) detector of the BIPM for SIR-type ampoules

Atomic and Nuclear Physics Laboratory (Physics 4780)

Appendix A. An Overview of Monte Carlo N-Particle Software

Spectral distribution from end window X-ray tubes

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

AAPM protocol for kv x-ray beam dosimetry in radiotherapy and radiobiology

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

Tutorial 4.6 Gamma Spectrum Analysis

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

2 Absorbing Solar Energy

Principles of dosimetry The ionization chamber

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

Gamma and X-Ray Detection

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations

Diagnostic x-ray imaging relies on the attenuation of

X Ray Flourescence (XRF)

Proton tracking for medical imaging and dosimetry

Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Qualitätsmanagement-Handbuch

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

1090 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 50, NO. 4, AUGUST Intelligent Gamma-Ray Spectroscopy Using 3-D Position-Sensitive Detectors

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

EDXRF of Used Automotive Catalytic Converters

Calorimetry in particle physics experiments

SphinX:A Fast Solar Photometer in X-rays. J. Sylwester 1,, S. Kuzin 2, Yu. D. Kotov 3, F. Farnik 4 & F. Reale 5

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

The photoionization detector (PID) utilizes ultraviolet

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

Fiber Optics: Fiber Basics

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge

Physics testing of image detectors

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

This document is the author s post-print version of this article, i.e. the final draft version after review. The final document will be printed in

GRID AND PRISM SPECTROMETERS

Introduction to Powder X-Ray Diffraction History Basic Principles

Experimental study of beam hardening artefacts in photon counting breast computed tomography

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

Quantitative assessment of the Effect of Anode Surface Roughness on Diagnostic X-ray Spectra: A Monte Carlo Simulation Study

A COMPLETE DOSIMETRIC MODEL OF THE GAMMA KNIFE PERFEXION USING PENELOPE MONTE CARLO CODES RYAN C.M. BEST

Conversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

The Calibration Chain: Role of BIPM,

arxiv: v1 [physics.ins-det] 14 Jan 2014

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

Status of the Free Electron Laser

Audit Report and Requirements of Mantid Software from the Molecular Spectroscopy Group (2013)

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Chapter 5 TREATMENT MACHINES FOR EXTERNAL BEAM RADIOTHERAPY

Vacuum Evaporation Recap

BRUKER ADVANCED X-RAY SOLUTIONS GUIDE TO XRF BASICS

Data. microcat +SPECT

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Permanent City Research Online URL:

Information about the T9 beam line and experimental facilities

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July Reprinted from Photonics Spectra, Laurin Publishing

Energy Dispersive Spectroscopy on the SEM: A Primer

Laue lens for Nuclear Medicine

Transcription:

Rapport BIPM-06/08 Bureau International des Poids et Mesures Establishment of simulated mammography radiation qualities at the BIPM C. Kessler BIPM September 2006 Pavillon de Breteuil, F-92312 SEVRES cedex

BIPM Rapport-06/08 Abstract Establishment of simulated mammography radiation qualities at the BIPM C. Kessler New radiation qualities have been established at the BIPM using the tungsten-anode low-energy x-ray tube with molybdenum and rhodium as filters to simulate the radiation beams used in clinical mammography. The evaluation of the correction factors for the primary standard and the measurements of the x-ray spectra are described. 1. Introduction One of the critical parameters in mammographic procedures is the dose to the breast tissue, which should be kept to a minimum while maintaining adequate diagnostic quality in the recorded image. Accurate dose determinations require measurements performed with radiation measuring instruments calibrated in the x-ray beams used in mammography, traceable to standards laboratories. Clinical mammography x-ray tubes use a combination of different anode materials (molybdenum or rhodium) and filters (molybdenum, rhodium or paladium). At present, few national standards laboratories are equipped with these x-ray tubes. However, similar radiation qualities can be produced using x-ray tubes with a tungsten-anode and suitable filters. A set of nine such x-ray qualities has been established at the BIPM using the present low-energy x-ray facility. 2. Determination of the beam quality and the air kerma rate The low-energy x-ray tube has a tungsten-anode with an inherent filtration of 1 mm beryllium. Combinations of the tungsten anode with two filter materials (molybdenum and rhodium) and different tube voltages in the clinical mammography range were used to implement the new radiation qualities, given in Table 1. The determinations of the beam quality, expressed in terms of the half-value layer (HVL), and the air kerma rates were measured with the BIPM low-energy x-ray standard (free-air chamber) [1]. Table 1. Characteristics of the radiation qualities Radiation quality M23 M25 M28 M30 M35 M40 M50 R25 R30 Generating potential / kv Additional filtration 23 25 28 30 35 40 50 25 30 60 µm Mo 50 µm Rh HVL / mm Al 0.332 0.342 0.356 0.364 0.388 0.417 0.489 0.464 0.505 µ air / 10 3 mm 1 0.213 0.208 0.203 0.198 0.190 0.182 0.166 0.156 0.147 * 0.423 mm of Be was added to all the qualities for reasons related to the measurement of air attenuation 2/8

For a free-air ionization chamber standard with measuring volume V, the air kerma rate is determined by the relation I K = ρ V W air air e 1 1 g air & (1) i k i where ρ air is the density of air under reference conditions, I is the ionization current under the same conditions, W air is the mean energy expended by an electron of charge e to produce an ion pair in air, g air is the fraction of the initial electron energy lost by bremsstrahlung production in air, and Π k i is the product of the correction factors to be applied to the standard, given in Table 2. By a suitable choice of anode current, the air kerma rate was set to 1.00 mgy s 1 for all the radiation qualities. The correction factors for the standard listed in Table 2 were determined by interpolation in terms of HVL from the existing data for the CCRI reference qualities. The factors for the CCRI qualities for electron loss k e, photon scatter k sc and fluorescence k fl were calculated by Burns [2] with Monte Carlo techniques for monoenergetic photons and the results were convoluted using spectral measurements performed with the NMi spectrometer at the BIPM; the correction factor for photon transmission through the aperture edge k l was calculated analytically and the correction for wall transmission k p was measured. For the present work, these correction factors were recalculated with the Monte Carlo code PENELOPE [3] using a more detailed simulation of the BIPM standard. The results for monoenergetic photons were convoluted with the spectra for the new qualities measured with the BIPM Compton spectrometer (described in section 4). The air attenuation factor k a was measured for each quality using the method of reduction of the air pressure in a tube placed between the filter and the free-air chamber (the beryllium windows for this tube, of total thickness 0.423 mm, are used as additional filters at all times for all the qualities for consistency). Table 2. Correction factors for the BIPM standard Radiation quality M23 M25 M28 M30 M35 M40 M50 R25 R30 Scattered radiation k sc 0.9974 0.9974 0.9974 0.9974 0.9974 0.9974 0.9975 0.9975 0.9975 Fluorescence k fl 0.9972 0.9972 0.9972 0.9972 0.9973 0.9973 0.9975 0.9974 0.9975 Electron loss k e 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 Wall transmission k p 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 Field distortion k d 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 Aperture edge transmission k l 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 1.0000 Air attenuation k a 1.0218 1.0213 1.0208 1.020 1.0195 1.0187 1.0169 1.0159 1.0150 Values for 293.15 K and 100.0 kpa; each measurement is corrected using the air density measured at the time. 3. Uncertainties The uncertainty components associated with the determination of the air kerma rate are listed in Table 3. 3/8

Table 3. Estimated relative standard uncertainties in the BIPM determination of air kerma rate for mammography x-ray qualities 100 s i (1) 100 u i (2) Physical constant dry air density (273.15 K, 101 325 Pa) - 0.01 W/e / (J C 1 ) - 0.15 g - 0.01 Correction factors k sc scattered radiation - 0.03 k e electron loss - 0.05 k fl fluorescence - 0.01 k s ion recombination 0.01 0.01 k pol polarity 0.01 - k a air attenuation 0.02 0.01 k d field distortion - 0.07 k l transmission through edges of aperture - 0.01 k p transmission through walls of standard 0.01 - k h humidity - 0.03 Measurement of I/vρ v volume /cm 3 0.03 0.05 I ionization current correction concerning ρ (temperature, pressure, air compressibility) 0.02 0.02 positioning of standard 0.01 0.01 Relative standard uncertainty in K & BIPM quadratic sum 0.05 0.19 combined uncertainty 0.20 (1) s i represents the relative standard Type A uncertainty, estimated by statistical methods; (2) u i represents the relative standard Type B uncertainty, estimated by other means. 4. Determination of spectra 4.1 Measurement of spectra An accurate knowledge of each spectrum is required to evaluate the energy-dependent correction factors involved in the air kerma determination. One of the problems in determining the spectra is the high flux of the x-ray beams. While several techniques exist to solve this problem, the Compton scattering method, which consists of placing a scattering material in the primary beam, measuring the scattered photons at a certain angle and then reconstructing the primary beam, was chosen for the present work. 4/8

A commercial Compton spectrometer was used that consists of a scattering chamber with lead walls, lead collimators and a PMMA rod of circular cross section used as the scatterer (Fig. 1). The scattered photons were detected at 90 with a low-energy pure germanium detector (LEGe) coupled to a multichannel analyser (MCA). The calibration of the MCA was performed using the known energies of the x- and γ-rays emitted by radioactive sources of 125 I and 241 Am. The primary x-ray spectra were reconstructed from the resulting pulse height distribution using commercial software [4]. primary x-ray beam scattering rod Scattered photons Low-energy germanium (LEGe) detector Signal to MCA 4.2 Simulation of the spectra Fig. 1. Schematic diagram of the Compton spectrometer The mammography spectra were also obtained by simulation with Monte Carlo techniques using the PENELOPE code [3]. The x-ray tube configuration (target, filters and collimation) has been simulated with the PENELOPE geometry code PENGEOM, as shown in Fig. 2. electron source Be filter Mo filter W target Be window lead collimators steel and Al holders Fig. 2. Model of the x-ray tube, filters and collimators In the first step, this model was used to create a phase-space file of photons crossing the first lead collimator (plane at 5 cm from the centre of the target), including the energy, angle and position of each particle. The photon and electron cut-off energies were set to 1 kev. The following values were chosen for the electron transport parameters in PENELOPE 1 : C1 = C2 = 0.2, Wcc = Wcr = 1 kev. Interaction forcing for bremsstrahlung emission was 1 C1 and C2 determine the cutoff angle that separates hard from soft elastic interactions; Wcc and Wcr are the cutoff energies for the production of hard inelastic and bremsstrahlung events, respectively. 5/8

applied to the primary electrons. This phase-space file was used as input for the transport of photons through the collimators and filters; the distribution of the photon number with energy was generated at 50 cm from the centre of the target. For the second step, the photon cut-off energy was also set to 1 kev, while the electron transport cut-off was raised to the maximum photon energy value (that is, no electron transport). 5. Results 5.1 Spectral measurements The measured spectra corresponding to 23 kv, 30 kv, 50 kv with the Mo filter and 30 kv with the Rh filter are shown in Fig. 3. The energy bins are 0.2 kev. No correction was made for the production of x-ray fluorescence in the LEGe detector. 1.10 1.00 0.90 0.80 Mammography spectra 23 kv Mo 30 kv Mo 50 kv Mo 30 kv Rh Normalized number of photons 0.70 0.60 0.50 0.40 0.30 0.20 0.10 5.2 Spectral simulations 0.00 6 10 14 18 22 26 30 34 38 42 46 50 Energy / kev Fig. 3. Spectra measured with the Compton spectrometer Figures 4 and 5 show the spectra for the 30 kv qualities with Mo and Rh filters, respectively, simulated with the PENELOPE code. Each spectrum is compared with that measured using the Compton spectrometer. 1.20 1.00 Measured spectrum Simulated spectrum Normalizer number of photons 0.80 0.60 0.40 0.20 0.00 6 8 10 12 14 16 18 20 22 24 26 28 30 32 Energy / kev Fig. 4. Comparison of the simulated and measured spectra for the 30 kv, Mo filter quality 6/8

1.20 Measured spectrum Simulated spectrum 1.00 0.80 Normalised number of photons 0.60 0.40 0.20 0.00 6 8 10 12 14 16 18 20 22 24 26 28 30 32 Energy / kev Fig. 5. Comparison of the simulated and measured spectra for the 30 kv, Rh filter quality 6. Discussion and Conclusions The correction factors used for the primary air kerma determination were derived by interpolation in terms of the HVL from the correction factors calculated for the CCRI qualities. The subsequent recalculation of these factors using the measured Mo and Rh spectra and an improved model of the standard showed the factors to be insensitive at the 3 10 4 level. The measured and calculated spectra for the 30 kv, Mo filter quality show slightly better agreement than those for the Rh filter qualities, although the maximum deviation between the curves in the latter case remains less than 0.35 kev. This deviation might be due to statistical fluctuations, which could be improved with further calculations. The energy broadening observed in the measured spectra at the absorption edge of the two filters may arise from the energy resolution of the Compton spectrometer. However, these differences have no significant effect in the calculation of the standard correction factors. The suitability of these simulated mammography x-ray qualities for the calibration of ionization chambers currently used in mammography is under investigation. For this separate study, the response of two different ionization chamber types calibrated in these beams is compared with that obtained from calibration in various mammography beams [5]. 7. References [1] BOUTILLON M., HENRY W.H. and LAMPERTI P.J., Comparison of exposure standards in the 10-50 kv x-ray region, Metrologia, 1969, 5, 1-11. [2] BURNS, D.T., New correction factors for the BIPM free-air chamber standards. CCRI(I)/03-28. [3] SALVAT, F., FERNANDEZ-VAREA, J.M., ACOSTA, E., SEMPAU, J., PENELOPE A Code System for Monte Carlo Simulation of Electron and Photon Transport, NEA/NSC/DOC(2001)19 (ISBN 92-64-18475-9). 7/8

[4] MATSCHEKO, G., RIBBERFORS, R., A generalised algorithm for spectral reconstruction in Compton spectroscopy with corrections for coherent scattering, Phys. Med. Biol. 34 (1989) 835-841. [5] KESSLER, C., BURNS, D.,T., BÜERMANN, L., DE PREZ, L., Study of the response of ionization chambers to simulated mammography beams Rapport BIPM-2006/xx (in preparation). 8/8