A COMPLETE DOSIMETRIC MODEL OF THE GAMMA KNIFE PERFEXION USING PENELOPE MONTE CARLO CODES RYAN C.M. BEST
|
|
|
- Stella Payne
- 9 years ago
- Views:
Transcription
1 A COMPLETE DOSIMETRIC MODEL OF THE GAMMA KNIFE PERFEXION USING PENELOPE MONTE CARLO CODES BY RYAN C.M. BEST A Dissertation Submitted to the Graduate Faculty of WAKE FOREST UNIVERSITY GRADUATE SCHOOL OF ARTS AND SCIENCES in Partial Fulfillment of the Requirements for the Degree of DOCTOR OF PHILOSOPHY Physics August 01 Winston-Salem, North Carolina Approved By: J. Daniel Bourland, Ph.D., Advisor William H. Hinson, Ph.D., Chair George Holzwarth, Ph.D. Michael T. Munley, Ph.D. Fred R. Salsbury Jr., Ph.D.
2 ii ACKNOWLEDGEMENTS A few people I would like to thank: The administrators of the DEAC Linux cluster, Tim Miller and David Chin, for their technical help and patience when I wrecked the cluster. My advisor, Dr Bourland, and all the physicists and therapists in Radiation Oncology. Wake Forest students and TRADONC postdocs Jacob Gersh, David Wiant, Marcus Bennett, and BJ Sintay for advice, encouragement, and trailblazing. My parents and Korinne Chiu, without whom this would not have been possible.
3 iii TABLE OF CONTENTS Page List of Tables and Figures v List of Abbreviations viii Abstract ix Introduction 1 Small field dosimetry 1 History of Monte Carlo simulations in radiation transport 13 Penelope background and theory 15 Penelope in practice 16 Methods and Materials 1 Gamma Knife Perfexion 1 Output Factors 7 Monte Carlo simulation parameters for the GK Perfexion 31 Source and Collimator Geometry 31 Photon fluence definition 34 Dose calculation grids and volumes 35 Methods to improve efficiency and reduce variance Phase Space Files step simulations 36. Azimuthal Particle Redistribution 4 Collimator output factor calculation and dependencies 43 Computational Resources 43 Radiochromic Film 44 Maximum Allowed Dose Difference Method 48 MADD Methodology 49 Results 57 Energy Spectra 57 Output factors 59 Collimator Output Factors 6 Row Output Factors 63 Comparison of Profile Curves 64
4 iv MC Comparison with Elekta-Calculated Dose Distributions 74 TMR10 Comparison with Film Measurements 81 MC Comparison with Film Measurements 89 Discussion 97 Conclusions 10 References 104 Appendices 110 Single Sector Dose Profiles with Error Range 110 Derivation of Compton Energy Equations 111 Vita 115
5 v LIST OF TABLES AND FIGURES Page Figure 1: Penumbra from finite source size and scatter Figure : Mass attenuation coefficients versus photon energy plots 5 Figure 3: Angular distribution of Compton scattering from Klein Nishina cross section 7 Figure 4: Cobalt-60 spectrum as measured by NaI spectrometer 9 Figure 5: Angular distribution of photoelectric effect scattering from Sauter distribution 11 Figure 6: Simulation of 15 Cobalt-60-energy photons on a slab of water 15 Figure 7: Source and collimator assembly of Perfexion in cross section Figure 8: Source positions for each collimator in Perfexion 3 Figure 9: The arrangement of sources within one sector 4 Figure 10: Eight sectors can vary collimations to produce a variety of dose distributions 6 Table I: MC-calculated output factors for collimators and rows from Elekta 9 Figure 11: Cutaway of collimator geometry from Elekta schematics 3 Figure 1: Co-60 sources move to collimator positions without rotating 33 Figure 13: Seventeen point sources along the cylinder of Cobalt 34 Figure 14: Cutaway of the collimator simulation geometry 38 Figure 15: Dose distributions from one sector of each row of the 16 mm collimator 39 Figure 16: Dose distributions from one sector of each row of the 8 mm collimator 40 Figure 17: Dose distributions from one sector of each row of the 4 mm collimator 41 Figure 18: Irradiation setup for film calibrations 45 Figure 19: Optical density to dose calibration data and curve 47 Figure 0: Films used for calibration and dose maps created from films 48 Figure 1: Dose distribution comparison model for definition of variables 50 Figure : Comparison model for defining dose difference and distance-to-agreement 51 Figure 3: Acceptance regions for several dose comparison methods 53 Figure 4: MADD defined as a shift to tangency with acceptance region 55 Figure 5: Photon spectra delivered to dosimetry sphere 57
6 vi Figure 6: Normalized photon spectra delivered to dosimetry sphere 58 Table II: Collimator output factors varying as a function of measured volume 60 Figure 7: Output factors varying as a function of measurement volume 61 Table III: Collimator Output Factors 6 Table IV: Row Output Factors 64 Figure 8: Dose profile curves from a single sector of the 16 mm collimator 65 Figure 9: Dose profile curves from a single sector of the 8 mm collimator 66 Figure 30: Dose profile curves from a single sector of the 4 mm collimator 67 Figure 31: Dose profile curves from all eight sectors of the 16 mm collimator 69 Figure 3: Dose profile curves from all eight sectors of the 8 mm collimator 70 Figure 33: Dose profile curves from all eight sectors of the 4 mm collimator 71 Figure 34: Dose profile curves from all sectors of all collimators old version 73 Table V: MADD pass rates comparing LGP and MC dose distributions 75 Figure 35: 16 mm Collimator Dose Distributions from TMR10 and MC Calculations 76 Figure 36: 8 mm Collimator Dose Distributions from TMR10 and MC Calculations 77 Figure 37: 4 mm Collimator Dose Distributions from TMR10 and MC Calculations 78 Figure 38: 16 mm Collimator Dose Distributions from TMR10 and MC Calculations magnified 79 Figure 39: 8 mm Collimator Dose Distributions from TMR10 and MC Calculations magnified 80 Figure 40: 4 mm Collimator Dose Distributions from TMR10 and MC Calculations magnified 80 Table VI: MADD pass rates comparing film and LGP dose distributions 81 Figure 41: 16 mm Collimator Dose Distributions from Film Measurements and LGP Calculations - Axial Plane 83 Figure 4: 8 mm Collimator Dose Distributions from Film Measurements and LGP Calculations - Axial Plane 84 Figure 43: 4 mm Collimator Dose Distributions from Film Measurements and LGP Calculations - Axial Plane 85 Figure 44: 16 mm Collimator Dose Distributions from Film Measurements and LGP Calculations - Coronal Plane 86
7 vii Figure 45: 8 mm Collimator Dose Distributions from Film Measurements and LGP Calculations - Coronal Plane 87 Figure 46: 4 mm Collimator Dose Distributions from Film Measurements and LGP Calculations - Coronal Plane 88 Table VII: MADD pass rates comparing film and MC dose distributions 89 Figure 47: 16 mm Collimator Dose Distributions from Film Measurements and MC Calculations - Axial Plane 90 Figure 48: 8 mm Collimator Dose Distributions from Film Measurements and MC Calculations - Axial Plane 91 Figure 49: 4 mm Collimator Dose Distributions from Film Measurements and MC Calculations - Axial Plane 9 Figure 50: 16 mm Collimator Dose Distributions from Film Measurements and MC Calculations - Coronal Plane 93 Figure 51: 8 mm Collimator Dose Distributions from Film Measurements and MC Calculations - Coronal Plane 94 Figure 5: 4 mm Collimator Dose Distributions from Film Measurements and MC Calculations - Coronal Plane 95 Figure 53: Dose profile curves from LGP and MC with error bars for MC data 110 Figure 54: Diagram of Compton scattering event 111
8 viii Model 4C Model B Co-60 CT CPE DCS ev FWHM GK GSR IMRT LGP LINAC MADD MC MeV MR OD OF OSLD PE PENELOPE PSF QA ROI SRS TLD TMR9, TMR10 Model U high-z LIST OF ABBREVIATIONS GK model preceding Perfexion Early (1997) model of GK Cobalt-60 Computed Tomography Charged particle equilibrium Differential cross-section Electron-volts Full width at half-max Gamma Knife Gamma stereotactic radiosurgery Intensity-modulated radiation therapy Leksell GammaPlan Linear Accelerator Maximum Allowed Dose Difference Monte Carlo Mega electron-volts Magnetic resonance Optical density Output factor Optically stimulated luminescence dosimeters Photoelectric Effect PENetration and Energy LOss of Positrons and Electrons in matter Phase space file Quality assurance Region of interest Stereotactic radiosurgery Thermoluminescent detector Tissue Maximum Ratio two LGP dose algorithms Early (1987) model of GK high atomic number
9 ix ABSTRACT Best, Ryan C.M. A COMPLETE DOSIMETRIC MODEL OF THE GAMMA KNIFE PERFEXION USING PENELOPE MONTE CARLO CODES Dissertation under the direction of J. Daniel Bourland, Ph.D., Professor of Radiation Oncology, Biomedical Engineering, and Physics (Adjunct) The Gamma Knife Perfexion is a radiosurgical treatment device used to irradiate targets in the head. It uses 19 high-activity Cobalt-60 sources collimated to a common focus to impart radiological damage to the target structure while minimizing dose to adjacent healthy tissue. Tungsten collimators are organized into eight sectors of 4 sources each; each sector can be set to one of three collimation sizes. The collimators can produce nominal spheres of dose 16 mm, 8 mm, or 4 mm in diameter at the focus. Planning for Gamma Knife stereotactic radiosurgery (GK SRS) is performed using images from computed tomography (CT) and/or magnetic resonance (MR) imaging, as well as manual skull measurements. The manufacturer s software used to plan the delivery of these treatments uses a simplistic set of assumptions that do not consider heterogeneous tissue structures.
10 x Monte Carlo (MC) methods that model radiation interactions and track radiation transport have been demonstrated to calculate dose more accurately in heterogeneous volumes than other dosimetry models. This study used Penelope, an established set of Monte Carlo dosimetry codes, to create a more complete radiological model of a sector of sources in the Perfexion and verified the model both by comparison to other models and to radiochromic film measurements. Specifically, this model was verified by 1) determining the relative dose rate factors, or output factors (OF), for the three collimator sizes (16 mm, 8 mm, 4 mm), relative to the 16 mm collimator; ) comparing one-dimensional dose profile curves generated by MC simulation, the manufacturer s dose calculation engine, and film measurement, 3) using quantitative mathematical tools to compare MC-calculated 3D dose distributions to the manufacturer s model dose distributions; and 4) quantitatively comparing MC-calculated D dose distributions to radiochromic film-measured dose distributions. Excellent agreement, within 1.0%, was found for calculated 4 mm and 8 mm collimator OFs compared to those distributed by the manufacturer. Dose profile curves compared between MC-calculated data and dosimetry calculated by Elekta showed strong agreement. Full 3D dose-distance comparisons between experimentally-acquired radiochromic films, the MC dose distributions and Elekta-calculated dose distributions were strong overall, while showing small differences in higher dose regions. Film measurements indicated the MC model slightly outperformed the Elekta-calculated model.
11 1 INTRODUCTION Small field dosimetry Radiation dosimetry is the measurement of energy imparted per unit mass by ionizing radiation to matter. 1 External beam radiation therapy uses one or more beams of ionizing radiation that are aimed from outside the patient and shaped to closely conform to the projection of a target volume, such as a tumor. Medical physicists typically use a 10x10 cm field as the reference field for medical linear accelerator-based dosimetry calculations. For dose rate calibrations a linear accelerator (LINAC) beam is collimated to form a square 10 cm on a side and a radiation dosimeter, typically an ion chamber, is placed on the beam s central axis at a prescribed depth in a known material at a reference distance from the radiation source. In clinical use the goal is to cover an entire tumor with a margin of about 1 cm or less. This margin is included due to uncertainty in definition of the target volume as well as to account for setup error and patient motion. As radiation oncology and medical physics treatment techniques have progressed, smaller, more conformal fields are being used that irradiate target tissues with closer margins while minimizing exposure to adjacent healthy tissue. A small field is described as a field size of less than 3x3 cm which needs special attention in dose measurements and in dose calculations. The 3x3 cm guideline applies to clinical beams from LINACs at 6 megavolts (MV) peak, whereas the average energy of photons from Cobalt-60 is about 1.5 megaelectron volts (MeV). No specific guideline exists for these low therapy-level energies. From a physics point of view, a small field is one where lateral electronic equilibrium is not obtained, or field flatness is compromised, as shown in Figure 1. Essentially, the field size is smaller than the average range of the secondary electrons and
12 electronic equilibrium, whereby the charged particles leaving the dosimetric region of interest are compensated for by others that enter the volume, is not obtained. 3 There are three primary difficulties associated with small field size dosimetry: penumbra, lateral electron range, and detector size. Dose measurements under these non-equilibrium conditions are very challenging. Figure 1: The penumbra is caused by the finite size of the source, and radiological scatter from collimating components and the interaction material (water equivalent phantom). For large square fields (a), the penumbra is a relatively small proportion of the dose compared to the central plateau, where the y-axis is relative dose. For small fields (b,c), the entire beam can be within the penumbra. The figure shows how full width at half max (FWHM) becomes a progressively less accurate description of field size, relative to collimator settings, as field size shrinks. (Figure from Das, 008. Used with permission.) The study of small field dosimetry is important for gamma stereotactic radiosurgery (GSR) since the largest field sizes used in GSR are 16 mm or 18 mm in diameter, falling
13 3 below the threshold size. The 3x3 cm rule also assumes the use of a 10x10 cm single (LINAC) reference beam. 4,5 A GSR device cannot produce a beam of that size, nor can it have only one source active. Thus, the standard LINAC-based dosimetry procedure does not apply to the small beams used in GSR because 1) small field dosimetry measurements are challenging due to finite detector sizes, ) only small fields are possible, and 3) it is not possible to use a 10x10 cm reference field. Monte Carlo (MC) dose computation is a promising alternative option when dosimetry measurements are difficult to obtain. Dosimetry for uniquely-sized fields (e.g., irregular, oblong shapes) is typically scaled according to measurements from known, measured field data. 6 The two main components for the size of beam penumbra at the field edge are geometric penumbra, due to finite source size, and radiological penumbra, due to lateral scatter of secondary electrons and photons. For small fields, the size of the source of radiation becomes far more important as the proportion of the field within the edge penumbra, the less-than-full-strength irradiation at the beam fringes, increases, as seen in Figure 1. MC simulations may assume a point source of radiation, or, as relevant for small fields, may simulate a finitesized source to ensure accurate modeling of geometric penumbra. Lateral electron range and charged particle equilibrium are also significant to small field dosimetry. When high-energy photons produce ionizations, most of the charged particles generated proceed in a forward direction relative to the photons, but some electrons travel laterally (e.g. the photoelectrons and Compton electrons). At the beam edges, the laterally-traveling electrons enter areas unirradiated by the primary photon beam fluence. Lateral electron range is highly energy dependent, 7 so a zero-field size beam s penumbra
14 4 size depends on the beam energy and essentially represents a dose kernel for an infinitesimal beam. The two most probable photon interactions for photons below 1.33 MeV are photoelectric absorption and Compton scattering. 8 The two primary Cobalt-60 γ rays are emitted (energies are γ 1 = 1.17 MeV, γ = 1.33 MeV) and interact within a material to scatter through lower energies. A particular material has photon attenuation properties that relate to the material s atomic and nuclear composition. The mass attenuation coefficient (cm /g) is a measure of interaction probability, and is calculated as the attenuation coefficient (cm -1 ) divided by the interaction material s density (g/cm 3 ). Mass attenuation coefficients for photons vary by material and energy. Figure shows the mass attenuation coefficients of photons at the relevant energy levels in water and tungsten.
15 5 Mass attenuation coefficients vs. photon energy for: Water Tungsten (A) (B) Figure : Mass attenuation coefficient versus photon energy in water (A) and tungsten (B) from NIST data. 9 Compton, or Incoherent scattering is shown in dotted blue; photoelectric absorption is shown in pink. The total mass attenuation coefficient, incorporating all interaction mechanisms, is in solid black, showing that other photon interaction mechanisms contribute minimally to the total cross section over these energies. In (B), the electron energy levels contribute to the photoelectric cross section at k-edges. Water (hydrogen and oxygen) lacks the electron shells to generate this step effect. Figure shows how Compton and Photoelectric effects are the principal interaction mechanisms for photons below 1.33 MeV. In both water and tungsten, photoelectric absorption is most probable at lower energies, below about 30 kev. Tungsten s greater prevalence of bound electrons increases the odds of photoelectric absorption at higher energies too, where photoelectric absorption is most likely up to about 700 kev. The mechanisms of Compton scattering and photoelectric absorption are significant for
16 6 understanding their mass attenuation coefficients as shown, and for understanding how to simulate such photon interactions accurately. The Klein-Nishina cross section, shown in Figure 3, describes the Compton photon scattering angle as a function of the incident photon energy. A wide range of incident energies are given in Figure 3, though the energy dependence is less significant for nearly monoenergetic Cobalt-60 (γ 1 = 1.17 MeV, γ = 1.33 MeV). Photon energy losses after scattering off the collimator walls will cause small spectral changes, where the scattering of lower energy photons will need to be understood. Compton interactions are the dominant effect for photons above 100 kev in these simulations. 8 As shown in Figure 3, the odds of lateral scattering increase as photon energy decreases.
17 7 Figure 3: The Klein-Nishina cross section describes the scattering of photons by electrons, yielding charged particles and photons after Compton interactions. Photons are incident from the left and interact with a free electron at the center of the plot. The isocurves show the relative probability of photon scattering through given angles. For any angle, an isocurve s distance from the edge of the plot indicates the probability of scattering through that angle. Colors indicate how incident energy changes the probability of scattering through each angle. Forward bias increases with energy. (Image by the author based on an equation in Salvat 10 ) In addition to the two photopeaks for Cobalt-60, a photon spectrum will also show Compton edges generated by the maximum energy a photon can impart on a free electron. 8 Equation 1 shows the source of Compton edges, where hv is the incident photon energy and mc is the rest energy of the electron (0.511 MeV). T max is the peak kinetic energy that can be imparted to an electron.
18 8 T max = hv + mc / hv (1) Equation 1 and other Compton scattering equations can be derived from only the conservations of momentum energy. A complete derivation of Equation 1 is included in the appendix. Given the photopeak energies of Cobalt-60, we expect to see Compton edges at 1.1 MeV and MeV. These peak energy interactions occur in a direct hit, a 180-degree photon deflection, according to Figure 3. Compton edges and other scattering phenomena can be seen in the Cobalt-60 photon spectrum in Figure 4.
19 9 Figure 4: A Cobalt-60 spectrum collected by a Sodium Iodide (NaI) spectrometer. The vertical axis counts the number of photons with a given energy, and the horizontal axis indicates the energy channel, or bin, in which the photon fell. This spectrometer has 104 channels which can be calibrated to show energy directly through the use of calibration sources which emit photons at known energies, such as Co-60. The two primary photopeaks at 1.17 and 1.33 MeV are clearly visible near channel 700 and 85. The few counts in channels higher than the 1.33 MeV peak are in error. The NaI detector double-counted two photon interactions as one, summing the energies of each. The higher-energy photopeak is shorter because some of the higher-energy photons scatter into the lower-energy peak s bins. Compton peaks are visible near channel 575, with a spectrum of Compton photons falling in lower channels. The low energy bins show high count rates relative to the number of events near the primary Co-60 energies. (Image from Singh 11 through GNU Free Documentation License) The photon energy spectrum shown in Figure 4 was collected with a sodium iodide (NaI) scintillation counter which records a lower resolution spectrum than other spectrometers, and at lower resolution than the MC-generated results presented later. A high-resolution spectrum would show narrow peaks at the characteristic Cobalt-60 photon energies, but
20 10 Figure 4 shows wide peaks. A detector s resolution depends on factors like the efficiency of light collection and charge collection, and on electronic noise. 8 The number of electrons that initiate a pulse in the detector has a certain statistical spread depending on the specific hardware of the device as well. Though Figure 4 gives a general idea of what to expect in a Cobalt-60 spectrum, it is expected that a MC-generated spectrum will be higher resolution. The other dominant photon interaction mechanism, the photoelectric effect (PE), dominates photon interactions at low energies, below 100 kev. 8 Under this effect, photons are absorbed by bound electrons, which leave their shell with energy equal to the photon s kinetic energy minus their binding energy. The MC system used in this work, Penelope 10 by Salvat et al., uses K-shell cross sections calculated by Sauter 1 to simulate energy depositions and angular deflections under photoelectric interactions. 10 The Sauter distribution is shown in Figure 5.
21 11 Figure 5: The Sauter distribution describes the angular deflections of photoelectrons, where a photon incident from the left scatters off a bound electron at the center of the plot. Electrons scattered by the photoelectric effect always travel laterally relative to the incident photon s path, so they are responsible for much of the dose deposited outside the central photon beam. Photons with higher initial energy tend to cause more forward-directed scattering. (Image by the author based on an equation in Salvat 10 ) Another photon interaction mechanism is pair production, which occurs when a photon passes close to an atomic nucleus, creating an electron-positron pair. Twice the electron rest energy, MeV, is required for this interaction to occur, so only the 1.33 MeV Cobalt-60 gamma has enough energy for pair-production. If pair-production is taking place, a recorded energy spectrum should show a peak at the difference between the
22 1 initial photon energy and twice the electron rest energy. In this case the difference is approximately 308 kev. 8 Since the total attenuation in Figure is nearly all accounted for by Compton scattering and photoelectric absorption, we expect pair production interactions to contribute a negligible amount to total dose. Small field dosimetry is also difficult due to the finite size of dosimeters, and the fact that any dosimeter necessarily perturbs the field it measures. The simplest measurement device is an ion chamber, which relies on Bragg-Gray cavity theory, stating that the chamber is only nonperturbing if the range of charged particles is greater than the diameter of the chamber cavity. For a sufficiently small chamber, maintenance of electronic equilibrium 13 across the chamber volume is possible, provided equilibrium existed before the introduction of the negligibly perturbing small volume chamber; this is not always the case with very narrow beams with high dose gradients. Small detectors, such as small and micro ionization chambers (~0.05 cm 3 to <0.01 cm 3 active volume, respectively), diamond detectors, and thermoluminescent detectors (TLDs) are useful for measuring dose at the central plateau of the beam, the beam s output factor 14,15, but cannot provide accurate data off-center due to dose gradient, which prevents charged particle equilibrium and averaging over peaks and valleys of signal if the detector is smaller than a uniform region of dose. In the future, even these micro-detectors may be too large to provide useful data with smaller photon beams, just as larger detectors today (Farmer chamber, PR-05P, etc) are limited in usefulness to large dose volumes. A Monte Carlo approach provides a much-needed alternative to experiment. Existing work has validated the MC approach to high energy (>1 MeV) photon dosimetry. 16,17,18 MC essentially allows a detector size of zero, eliminating spatial effects on dose, and
23 13 eliminating the perturbing effects of dose measurement. Dosimetry characteristics can be found by measuring a plane of data, then using a scaling factor measured experimentally (at a peak, flat-dose area for example) or by direct calculation of dose. These MCgenerated dose distributions are calculable even when electronic equilibrium is absent, 19 an advantage over measurement. History of Monte Carlo simulations in radiation transport Monte Carlo methods first appeared in physics literature in an article by Kahn 0 in 1950 and the topic s publication rate has increased steadily since that time. 1 Monte Carlo simulations use random number generators to pick among a series of probability distributions to simulate step-by-step the path of an ionizing radiation particle through a given geometry, recording dose deposited along the way. The earliest simulations were conducted by hand in a lab book. Later, enormous mainframe computers took months to calculate thousands of photon or particle histories. Now, 10 8 histories can be simulated for simple geometries on a standard laptop computer in under an hour. A simple example of a MC simulation is the calculation of step length, s, within a homogenous media. Step length is the distance a photon or electron takes between interacting with the surrounding media. In Equation, λ is the mean free path for the particle given its energy and the attenuation coefficient of the medium, and ζ is a random number evenly distributed between zero and one.
24 14 s = λ ln( 1 ζ ) () A series of ζ will produce a set of distances s that match the exponential attenuation, e -µx, found in experiment. Given that radiation is emitted isotropically from a point source, it is easy to demonstrate how random numbers can be used to choose the initial direction of travel for a particle or photon, choose its initial energy from a given spectrum, and pick what type of interaction occurs based on additional, known probability distributions. Once the interaction occurs, deposited energy is recorded and the new direction of travel calculated, the process begins again with the less-energetic secondary particle or photon. This type of thorough, microscopic simulation is extremely accurate but computationally expensive, especially for the simulation of electron scattering. The slowing down of an electron in aluminum from 0.5 MeV to 1 kev takes around 10 4 collisions, when modeled without approximations. 3 A comparably energetic photon takes only a few hundred collisions to come to a stop. This difference, approximately a factor of 10, is shown in Figure 6, where the photons show few turning points and the electrons bend their paths frequently. Most modern MC simulations now use a condensed history or macroscopic approach for certain types of computationally-intensive interactions in order to keep simulation times reasonable.
25 15 Figure 6: A simulation of 15 Cobalt-60 energy photons impinging on a slab of water from the left. The orange/black photons travel in long, straight paths before their few interactions. (Black photons are higher energy than orange photons, as shown in the legend at top-left.) The few red electrons generated by photon interactions follow short, tortuous paths. Note that the length of the red electron track is roughly proportional to the angle of deflection by the ionizing photon. Although 15 photons were initially simulated, only 5 interact in this image which spans a depth of 8 cm; the remaining photons pass off to the right side of the image. Penelope background and theory Penelope is a mixed procedure 4 MC simulation system which divides electron and positron collision events into soft and hard events. Photons are simulated in a detailed way, event-by-event, without approximations. In the energy ranges of concern
26 16 for this experiment the positron contribution will be negligible. Hard electron events are distinguished by a significant change in the electron s direction or a significant deposition of charge, characteristics which are positively correlated. Soft events deposit less energy and result in smaller changes in direction. Soft events are approximated and not simulated in a detailed way. The energy thresholds for distinguishing between hard and soft events are configurable in Penelope, so the degree of approximation is adjustable at the cost of longer computing time. 10 The default Penelope energy thresholds of 10 kev for electron simulation were used in this work. For high energy electrons, the differential cross sections (DCS) for interactions drop off quickly for larger scattering angles and increased energy loss. 5 Thus, relatively few interactions will take place on that track and each individual event can be simulated for little computational cost. Even in large numbers, soft events have little effect on the particle s direction. Soft events therefore use a multiple scattering approach, where low energy events are dropped along a straight track at fixed intervals without individual computation. Penelope in practice Penelope has been experimentally verified, demonstrating its effectiveness at accurately predicting the behavior of ionizing radiation. In the literature, Penelope is compared to other Monte Carlo implementations as well as to experiment. 9,3-30 Sempau et al. modeled an electron beam from a LINAC head using Penelope as well as BEAM and DOSEXYZ, codes based on another MC modeling package called EGS4. 6
27 17 Modeling a LINAC is in many ways more difficult than modeling a device with a radioactive source like Cobalt-60. Radiations from radioactive sources have well-defined energies (monoenergetic photons and continuous electron energies), but the exact energies and spectrum of a LINAC can vary from machine to machine, and can be tuned. Sempau modeled the complex internal geometry of a Siemens Mevatron LINAC using the Penelope geometry engine, PENGEOM. PENGEOM uses simple geometric shapes described in quadric geometry to define the area to be modeled. Planes, cylinders, cones, paraboloids, and other shapes are defined and pieced together into homogeneous bodies. Many bodies pieced together can form complex shapes, like the head of a LINAC or simplified patient anatomy. 10 Quadric surfaces are defined by the reduced quadric equation, Equation 3. Plotting where F s =0 shows a surface whose shape is defined by the I n coefficients, which can each be set to -1, 0, or 1. Setting I n =(1,1,1,0,-1) produces a sphere, I n =(1,1,0,0,-1) produces a cylinder about the z-axis, I n =(1,1,-1,0,0) produces a cone, and so on. Paraboloids, hyperboloids, and planes can also be created. F s = + (3) ( x, y, z) I1x + I y + I3z + I 4z I5 PENGEOM defines a volume from a surface by a side pointer which is set to ±1 per surface, indicating which direction is filled. A side pointer of -1 on a spherical surface produces a solid ball of material, while a side pointer of +1 would fill everything outside the sphere. A second, planar surface could divide the ball into a hemisphere, and which side remains is determined by the plane s side pointer. Scaling factors are used to turn spheres into ellipsoids, or turn cones narrower or wider, etc. This robust quadric
28 18 geometry is used by Penelope due to its low computational cost compared to voxelized geometries. 10 Penelope includes a useful tool for separating this complex computational type of problem into several steps: A phase space file (PSF) of particles can be recorded at a certain plane in one step of the simulation, then the PSF can be played back exactly at the next step. For Sempau s simulation a plane was defined at the exit window of the accelerator so every particle leaving the complex geometry of collimators, flattening filter, etc. was recorded into the PSF. For the next step, the previously simulated complex geometry of the LINAC was retained in the PSF, and to save computer time the PSF was played back as a beam of particles impinging on a slab of water. Two types of detectors were used to measure the output in the water and both were compared with the Penelope and BEAM results along the central axis, and several lateral planes. For a 6 MeV beam, Penelope was consistent with experiment, with no more than 3% error in depth dose and 5% error in lateral dose. For a 18 MeV beam, Penelope s depth dose results were within 3.5% and lateral dose results were within 5.5%. Lateral dose errors are larger in part due to the difficulties in accurately measuring the position and dose experimentally at the beam edges, where a penumbra of dose exists. Between BEAM and Penelope, small but notable differences were found in their PSFs. These variations were attributed to differences in the way the two algorithms simulate particle and photon interactions. The authors note that these differences, however, are shown to have a negligible effect on the calculated dose distributions. 6
29 19 Leksell GammaPlan (LGP) is the software tool from Elekta AB used to plan GK treatments and calculate dosimetry. Until recently 7, LGP had a significant limitation: In the classic mode of dose calculation, LGP assumes all material is either air, outside the patient, or water, causing predictable dosimetric inaccuracies when the beams interact with bone and air cavities within the patient, such as skull and sinus. Within this homogenous-medium dose calculation mode, LGP does not use image data for heterogeneous dose computation so the calculation model is unaware of the presence of any materials other than those that are water equivalent. The latest version of LGP (version 10.1) includes a convolution 7 dose algorithm which accounts for heterogeneities by modifying the dose kernel s distribution based on electron densities obtained from a x- ray computed tomography (CT) scan. Convolution-based dose calculation has been shown to handle heterogeneities better than methods assuming homogeneity, but is less accurate than Monte Carlo methods in theory and, where tested, in practice. 8,9 The convolution dose calculation engine is not yet commissioned in our facility so comparisons to convolution-based dosimetry are not a part of this study. All simulations and experiments in this work take place in water-equivalent media, so heterogeneous calculations would be negligibly different from the homogenous media approximation. Monte Carlo is generally well-suited for the heterogeneous material problem, and Penelope in particular due to its high accuracy at interfaces between materials 30 and high- Z materials. 31 Several studies discussed below compared MC results to LGP predicted results for different GK models. Penelope was used by Moskvin et al. to model the Gamma Knife model 4C. 3 PENGEOM was used to design the internal collimating structures of a single source (the
30 0 4C has 01 sources) and a PSF was collected at the exit of the secondary collimator. The single source was modeled at five planes at fixed distances down the beam path in air. The results were compared to experiments using film dosimetry. Using previouslypublished information about source locations 33, Moskvin modeled the collected PSF as originating from each of the 01 source origins. The output of each PSF was recorded on a dose plane within a simulated plastic 16 cm diameter homogeneous dosimetry sphere. Film was placed inside an identical experimental sphere and was compared to the calculated dose distribution plane. The dose distribution plane was also calculated in LGP. Comparison of the 50% isodose curve diameters in Penelope, film, and LGP found small discrepancies but overall strong agreement. The general shapes of the isodose curves were followed closely in each simulation method, with the largest differences at 0.5 mm, and less than 0. mm, which is within Elekta specifications in either case. The strong agreement between LGP and Monte Carlo methods in this case was due to the homogenous density of the dosimetry sphere. When the same team later used the same Penelope-generated GK PSFs on a heterogeneous sphere the results were different. They built a dosimetry sphere with air cavities and simulated bone to more accurately model the human head, then modeled an identical head phantom in PENGEOM. 9 Dosimetry differences in this case were as much as 7% near air/tissue interfaces due to secondary electron disequilibrium and lowered energy deposition near the heterogeneous areas. The authors proposed the inelegant solution of simply plugging all the sources whose beams passed through air cavities. It is surprising that LGP still uses a very simple algorithm for dose calculation (exponential attenuation: e -µx ) when current commercial linear accelerator treatment planning software
31 1 packages like have used CT data to estimate attenuation characteristics of pixel-sized regions for many years. Elekta AB is working on both convolution and MC models that will require a CT image for dose computation, with higher-fidelity results expected for heterogeneous media. METHODS AND MATERIALS Gamma Knife Perfexion The Elekta Gamma Knife (GK) Perfexion is a GSR treatment device that collimates gamma ray radiation from 19 linear sources of Cobalt-60 to a small nominal sphere (4, 8, or 16 mm in diameter) centered about a geometric focus in order to deliver a high radiation dose to a small target positioned at the focus while delivering a <5% dose rate to healthy tissue a few millimeters from the target. The sources are arranged in five rows outside a 30 cm radius tungsten cylinder (nominally 30 cm outer radius, 0 cm inner radius, 10 cm wall thickness) that has precision machined holes to collimate each source to 4, 8, or 16 mm full-width at half maximum (FWHM) in diameter at a focal point in the center of the device, (Figures 7 and 8). 34 The mechanical positioning of the device is advertised as within 0. mm. Nominal source-to-focus distance is 40 cm, however, as is discussed later, each source row has a unique distance value. Sources are arranged in 8 translatable sectors of 4 sources each. Each sector can move independently to one of four longitudinal positions: fully blocked, 4 mm, 8 mm, or 16 mm positions. In this manner, a target located at the focus can be irradiated with a desired grouping of the 8 sectors.
32 Co -60 Sources Source Tray moves sources Tungsten Collimators - color coded by size Focus (approximate) 0 cm Figure 7: The source arrangement and collimator assembly shown in cross section. 35 Cobalt-60 sources are blue and move together on the source tray in red. Three collimator sizes are shown: 4 mm in yellow, 8 mm in green, and 16 mm in red. Sources move to a collimator to irradiate the focus. (Image source: Elekta AB. Used with permission) A significant portion of the work involved in MC modeling of the Perfexion is determining and describing the internal geometry of the device accurately. For this reason the details of the GK design will be described at some length.
33 3 Focus 4mm collimator position 16mm collimator pos n 8mm collimator pos n Figure 8: The source tray moves longitudinally to position the sources in the three treatment positions aligned with the three sets of collimators. 34 The illustration with yellow beams is the 4 mm collimator, 16 mm in green, and 8 mm in red. The collimators have a common focus. The collimator selected determines the volume irradiated at the focus. (Image source: Elekta AB. Used with permission) Previous versions of the GK underwent few significant changes since the first Model U version (US version, 1987), essentially the original GK device, with 01 sources arranged in a hemispheric geometry at the same distance (40 cm) from the focus. In the Model B (US version, 1997) 35, the 01 static sources were arranged in a vertically oriented hemisphere, again with each source the same distance from the focus as for the Model U (40 cm). The modeling of a nearly spherically-symmetric system was comparatively easy once the geometric coordinates of the sources were known. 33 In contrast, for the Perfexion each sector of 4 sources can be moved independently and the source-to-focus distance is different for each of the five rows of sources within a sector, varying from 37 to 43 cm. 34 A radiochromic film recorded on the surface of a cylinder 10 cm from the isocenter, Figure 9 shows the arrangement of the source beams from one sector at the 16 mm setting. Some of the beams in Figure 9 are starting to overlap on their way to convergence at isocenter. The row E beams are more ellipsoidal than the row A beams
34 4 because row E intersected the film at a more oblique angle than row A, which was nearly perpendicular to the film. (A) (B) Figure 9: (A) Positioning of sources within a sector. Sectors are visibly in several different collimator positions in this image since the rows are not lined up between sectors. (Image source: Elekta AB. Used with permission.) (B) A radiochromic film taken 8 cm from the isocenter showing the arrangement of Cobalt-60 beams delivered by one sector of the Perfexion. In the GK models before Perfexion the 01 sources were fixed in location regardless of the collimator size used to change the beam size one collimator was dismounted and another manually mounted in its place. With Perfexion s moving source trays, the positions of the sources vary for each collimator size (Figure 9A). One important aspect of the sources that does not vary between collimator positions is the angular orientation
35 5 of each linear source within the tray. When the sources are in the 4 mm collimator position, the linear source in its bushing is aligned with the collimator along a common axis. The 4 mm collimator is the only collimator size with this well-aligned geometry. For the 16 mm and 8 mm positions, the source axes are not perfectly aligned with each collimator axis. The angular difference is small, but is significant in a device which demonstrates sub-millimeter positional accuracy. 35 This issue is discussed in more detail later and the small angular difference in alignment is shown in Figure 1. Another unique aspect of the Perfexion relative to the previous models is the ability to use multiple collimation sizes at once. Previous GK models used a collimation helmet that attached prior to treatment. The collimation helmets for the GK Models U, B, and 4C came in four collimator sizes (4, 8, 14, and 18 mm), and it was not possible to simultaneously mix the collimators of different sizes. The only dose shaping option was to block individual sources entirely, called plugging. Plugging is a time-consuming process performed one source at a time. In the Perfexion, since the eight sectors (4 sources per sector) can be in one of four positions: fully blocked, 4 mm, 8 mm, or 16 mm, unique dose distributions can be produced from a single shot based on clinical needs. See Figure 10.
36 6 Figure 10: The eight sectors of sources can be sized individually as shown by the small graphics in the bottom right of each image. When all sectors are collimated the same, the isodose curves in the X-Y plane (constant Z) shown appear roughly circular. The unique shapes seen here are produced by different dose contributions from different directions due to different collimators active in each sector. Shown isodose lines are 15%, 5%, 50% (yellow), 60%, 75%, 90%. 35 (Image source: Elekta AB. Used with permission.) LGP handles patient images and uses a simple dosimetry algorithm, exponential attenuation (e -µx ), relative to the dose rate at the focus for an 80 mm radius waterequivalent sphere to calculate dose within the patient s head as a function of radiological depth for each of the 19 sources. The dose computation algorithm back-calculates along each source trajectory from the center of an 80 mm radius sphere, using the path length (depth) for each individual shot. Skull radii measurements are used to provide a simple model of the surface of the head, to determine the distance from each source to the skin surface for purposes of dose computation. As stated previously, the LGP algorithm assumes that tissue located under the skin surface is radiologically water equivalent. Several papers have found significant problems with the LGP dose calculation method, especially in heterogeneous regions like bone and air cavities. 9,30 These differences in dose computations may be significant for research conditions such as small animal irradiations which include small, complex anatomical regions. 36
37 7 The most accurate dose computation algorithms use MC calculations that take into account the many radiological parameters of materials present in the irradiated region. A system that accurately simulates physical phenomena will necessarily be more accurate than the simpler approximations currently used. To date, no complete MC simulation of the GK Perfexion exists in the literature. Output Factors The output factor for the GK Perfexion model is the experimentally determined dose rate delivered by a specified collimator size, or collimator row, relative to the dose rate delivered by a reference collimator or reference row. 37 Relative to a dose profile curve, which looks similar to a Gaussian distribution, the output is measured across the flat dose plateau (peak dose) at the central region of the dose distribution. A collimator output factor (OF) is the dose profile curve s height in the plateau region relative to the height of the 16 mm collimator s profile curve. Since there are five rows and three collimator sizes, fifteen row OFs exist. The dose distributions for the fifteen row OFs are recorded near the focus and compared to the dose over the same volume of the 16 mm-row B, which is normalized to unity. Rows of sources cannot be activated individually with the Perfexion, only full sectors of 4 sources consisting of five rows, so row OFs are not measurable on the Perfexion. Only the collimator OFs are measurable experimentally. Elekta provides the row output factors based on their in-house MC calculations and publishes these data with each patient treatment plan and online. 38 The collimator OFs are also provided by Elekta and are based on their own MC simulations and small detector measurements.
38 8 Collimator output factors in MC simulations are the weighted sum of the row outputs normalized to the 16 mm collimator output. The weighting factor, W, is the number of sources in the given row: 6 sources in row A, 4 in row B, and so on up to the 4 sources in a sector as seen in Figure 9B and equation 4. Weighting is required because the row OFs account for only one source per row. i i= A, B, C, D, E W OF Row i OF coll = (4) 4 W i = 6,4,5,4,5 See Table I for the output factors provided by Elekta. Two different sets of OFs now exist for the older and newer versions of the LGP dose calculation algorithm. The TMR9, or classic, algorithm uses the older output factors; TMR10 is the updated version that uses the newer output factors. 39 Changes and uncertainty in OF values are part of the justification for additional models of the Perfexion like the one developed here. Wake Forest Baptist Medical Center uses the TMR10 algorithm as of January 01. Output factor comparisons in this work address the newer TMR10 values unless otherwise noted.
39 9 Collimator Row Row OF Collimator OF TMR9 TMR10 TMR9 TMR10 A B mm C D E A B mm C D E A B mm C D E TABLE I: Monte Carlo calculated output factors for three collimator sizes and five rows as calculated by Elekta Instrument AB, Stockholm, Sweden. 40 Only the Collimator OFs may be measured experimentally because the rows cannot be activated individually. TMR9 columns indicate OFs used by the older LGP dose calculation algorithm; TMR10 columns show the OFs used by the newer LGP algorithms. 39 (* All row OFs are defined as a fraction of the 16 mm-row B output. Collimator OFs are relative to the 16 mm collimator output.) Output factors measurements have been reported in a many articles over the several versions of the Gamma Knife, covering dozens of devices capable of measuring ionizing radiation. Bednarz et al found the OFs for the GK model U using diamond detectors and miniature ion chambers. 41 The OFs of the GK model B were measured by Mack et al., Tsai et al., and Araki et al. using a liquid ion chamber, pin-point ion chamber, diode detectors, diamond detectors,
40 30 TLDs in microcube and microrod forms, alanine pellets, radiochromic films, silicon diodes, radiographic film, radiophotoluminescent glass rod dosimeters, and p-type silicon detectors Further GK model B OFs were measured using film which also documented the effect of collimator size on end effect times the time during which the patient is moving into place. 4 Measurements on the model 4C Gamma Knife, the version immediately preceding the Perfexion, were performed by Kurjewicz et al. 43 among others using film, optically stimulated luminescence dosimeters (OSLDs) 44, and polymer gels 45. GK Perfexion output factor measurements have been performed by Novotny et al. 40 and others 5,53 using various types of film and found their OFs in good agreement with the vendor recommended Monte Carlo calculated values. Monte Carlo simulations were used to measure the collimator output factors of previous GK versions with results close to Elekta recommended values. 30,35,47 As a result of internal and external investigations of the 4 mm output factor, Elekta recommended changing the Gamma Knife model 4C s 4 mm output factor by close to 10% (0.80 to 0.87). 46 Largely based on improved MC simulations, Elekta modified the 8 mm and 4 mm output factors of the Perfexion with the software update from TMR9 to TMR (8 mm: from 0.94 to 0.900, and 4 mm: from to 0.814)
41 31 Monte Carlo simulation parameters for the GK Perfexion Source and Collimator Geometry The bulk of the work on this project consists of accurately modeling the interior geometry of the GK Perfexion collimators and sources. Initially, significant progress was made mocking up the GK geometry based on back-calculating the positions of sources by using radiochromic film images obtained along the beam paths (e.g., Figure 9B). With a nondisclosure agreement with Elekta, simplified machinist plans 47 were made available, allowing a more accurate physical MC simulation. These specifications to a large degree verified our earlier film-based measurements, locating the source angles no more than one degree off of the radiochromic film-based calculations. The simplified machinist drawings provided by Elekta describe a slab of tungsten, and concentric cylinders of various diameters forming a collimator straight through this slab (see Figure 11). The distances from the focal point to shielding interfaces and the source bushing are shown. Not shown, but included as data are the collimator angles from normal, and the axial rotation for each source s collimator. These various parameters are confidential information and cannot be stated in this document.
42 3 Figure 11: Cutaway of collimator geometry data received from Elekta. Diameters of the collimator at several key points were provided, along with distances to sources and source orientations. The collimator is a series of concentric cylinders drilled out of tungsten shielding. Drawing is not to scale. The Penelope version 003 quadric geometry 10 is ideally suited to modeling a simple system of stacked cylinders set on a tilt. More complex to model using quadric geometry were the individual GK sources and source bushings which are made up of six nested cylinders. 48 Not shown in Figure 11 are the small tilts and shifts of the source bushing, which vary for each collimator row and collimator size. Since the sources move along a fixed channel to each of the collimating positions, the source s central axis can only be coaxial with one source position, which is the 4 mm collimation position. For the 16 mm and 8 mm collimators the source axes sit at a small angle relative to the collimator axes, as shown in
43 33 Figure Precise values vary for each collimator row. These small angular changes in the source-collimator alignment have an effect on output factors since the effective width of a linear source increases when it sits at an angle instead of end-on. Figure 1: A single Cobalt-60 source can move between three collimator positions. In the 4 mm position the source and its bushing are aligned with the collimator. In the 16 mm and 8 mm positions the source and bushing are not on a common axis with the respective collimators, and are slightly oblique. Accurate modeling of the source shape is also important. The current model uses seventeen point sources set 1 mm apart along the 17 mm length of the cobalt source, as seen in Figure 13. Seventeen was selected because the cylinder of Cobalt is nominally 17 mm long and 1 mm in diameter.
44 34 Figure 13: The simulation modeled seventeen point sources of radiation (shown as white dots in the center of the bushing) along the blue Cobalt volume inside the green and purple source bushing. The initial solid angle of emitted photons from each white point was 45 degrees, as shown by the semitransparent red cones. Four red cones are shown; each point emitted an equal number of photons in a 45 degree solid angle toward the focus. The 008 version of Penelope allows definition of linear sources which would eliminate problem of approximating a linear source as discrete points by randomly selecting a source position along a line for each individual photon shower. 49 This most recent, updated version was not able to be implemented and remains a future improvement to the current model. Photon fluence definition Although an isotropic source fluence of 4π steradians would be more physically realistic, it would require about 6 times more initial photons to achieve similar statistical certainty in the final OF results. In our internal tests we found a.5-degree cone, measured from the central axis of the beam to the beam edge (a 45 degree cone, Figure 13), was sufficient. The GK source row closest to the focus is approximately 38.5 cm from
45 35 isocenter, so an uncollimated.5-degree cone would be 9 cm across at isocenter. The extra width provides authentic scattering effects within the bushing and collimator such that the photon spectra reaching the target include far more scattered photons than photons at the two primary Cobalt-60 energies. In the simulation, initial photon energy was randomly selected from the two energies (γ 1 = 1.17 MeV, γ = 1.33 MeV) that Cobalt-60 emits, with each energy emitted 50% of the time. Each of the seventeen points emitting photons along the linear source produced 8.3*10 7 initial photons, for a total of about 1.4*10 9 initial photons per GK source. This number of initial photons is greater than the number used by other simulations of earlier GK models. 3,50 Internal testing against larger PSFs showed smaller files could be played back repeatedly with no difference in dosimetric results, and only modest cost in terms of statistical significance. Dose calculation grids and volumes A three-dimensional grid of dose was recorded over the center of the dosimetry sphere. The dimensions of the grid varied by application, between 5x5x5 cm 3 and.5x.5x.5 cm 3, always centered on the focus. The grid was 100x100x100, producing a bin size of 0.5 to 0.5 mm cubes. The dose deposited in each bin was collected from the grid using a Matlab program. These three-dimensional grids were used by Matlab to compare dose distributions. When dose to a volume within the grid was needed, Matlab was used to integrate over the volume and sum the dose over several bins.
46 36 Also imported was a grid of equivalent size giving the uncertainty in the dose to each element in the dose grid, data which are calculated by Penelope and reported in the raw dose grid file. These uncertainties were used to keep track of propagation of error in reported dose results. Methods to improve efficiency and reduce variance Two principal methods were used to improve simulation efficiency and reduce variance in dosimetric results: -step simulation via PSFs and azimuthal particle redistribution. 1. Phase Space Files- step simulations: Monte Carlo simulations of LINACs and other radiosurgical devices are typically divided into multiple steps. 6 First, particles are generated at a small- or point-source from a given energy spectrum and directed toward the target within an initial solid-angle cone, limiting the starting directions. After interacting with various shielding and beam-modifying components, but before reaching the patient or target volumes, the particles hit a limiting surface which records their final state (position, direction, and energy) into a PSF. After recording the PSF, the second step of the simulation is playing-back the recorded PSF particles. There are several ways that dividing the simulation in pieces improves calculation efficiency. Of the initially-generated particles, sometimes only a small fraction reaches this recording plane. For some 4 mm collimator simulations, less than 1% of the initial photons reach the PSF plane and are recorded. A 4 mm diameter circle inscribed on the surface of a sphere of radius 38.5 cm, the approximate distance from a source to the focus, is approximately 7x10-6 the area of the full sphere. By this calculation, % of
47 37 photons generated by an isotropically emitting source would never reach the 4 mm diameter target. More photons than this end up reaching the target volume since many are scattered back toward the collimator, but it s clear that a small minority of photons generated in the bushing will end up exiting the collimator at the far end. Since a PSF is typically played back several times, the PSF eliminates the need to simulate particles which will never make it near the target. Additionally, it eliminates the need to simultaneously model the entire geometry to be simulated. Dividing the process into steps allows first a collimator to be modeled, then on playback the collimator is removed and the target volume put in place. Minimizing the complexity of geometry to be modeled significantly improves simulation speed. A final advantage of multi-step simulations using PSFs is that the target geometry does not need to be defined during the first step. Once recorded, the PSF can be played back to impart fluence on a phantom, simulated animal anatomy, voxelized patient anatomy, or any other target. MC modeling is the ideal method for investigating these subtle differences since it is possible to determine the dosimetric differences across a single beam, including energy differences within a beam, by analyzing the PSF. Experimental methods such as film or gel dosimetry cannot record such spectral or directional information.
48 38 Figure 14: A cutaway of the collimator simulation geometry. The surface where a phase space file was recorded is visible on the right, flush with the opening of the collimator. Photons entering the PSF collection surface were recorded into a PSF. Simulations generated photons at the position of the Cobalt-60 source and directed them within a.5-degree cone, measured from the beam center, down the series of concentric cylinders that make up a GK collimator (45 degree solid angle cone). Photons which passed out of the tungsten collimator assembly reached the PSF plane, shown in Figure 14, and were recorded. Photons reaching this plane were mostly near their initial energies of 1.33 or 1.17 MeV, although a predictable amount of the spectrum was also filled with lower energy scattered photons, similar to the spectrum in Figure 4. On playback, the photons are sequentially re-emitted from their recorded position in the recorded direction with the recorded energy toward the target of choice. Figures show the dose distribution from each row of one sector s irradiation within the 16 cm diameter Elekta dosimetry sphere from the 16 mm, 8 mm, and 4 mm collimators, respectively. See Figure 9 for the source distribution within each row. Each
49 39 source in a row was collected as a single PSF and then rotated about the z-axis to the proper source position. Different random seed values mean the cloned beams differ after the first interaction. (A) (B) (C) (D) (E) Figure 15: Dose delivered by each row of one sector set to the 16 mm collation size. High dose areas are red; low dose areas are blue, with different linear scales for each image. The images in A-E correspond to each collimator row, A-E. Some beams overlap before reaching the sphere s surface.
50 40 (A) (B) (C) (D) (E) Figure 16: Dose delivered by each row of one sector set to the 8 mm collation size. High dose areas are red; low dose areas are blue, with different linear scales for each image. The images in A-E correspond to each collimator row, A-E. Some beams overlap before reaching the sphere s surface.
51 41 (A) (B) (C) (D) (E) Figure 17: Dose delivered by each row of one sector set to the 4 mm collation size. High dose areas are red; low dose areas are blue, with different linear scales for each image. The planes in A-E correspond to each collimator row, A-E. Figures show how a single recorded PSF can be rotated about the z-axis to model the several sources in a row. The sources in row A for each collimator are uniformly distributed, but one lone source is apparent in rows B-E on the right side of Figures Figure 9 shows the one lone source in rows B-E as well. The screw thread that drives the sector passes through this gap in sources. The number of sources per row was selected by the manufacturer to maximize the number of cylindrical sources in the limited sector space.
52 4. Azimuthal Particle Redistribution: Collection of sufficiently large PSFs requires many computing hours to reach levels where the variance in resulting dose distributions becomes low enough for practical use. Azimuthal particle redistribution (APR) was developed to make this process more efficient by taking advantage of symmetrical components in the simulated volume. 51 For example: A LINAC model is simulating a square field and a phase space file is collected where the photons exit the collimators. If the geometry that created these photons is to some degree symmetrical about the azimuthal axis, the center of the beam, then all the collected photons in this square beam can be rotated in 90-degree increments about the azimuthal axis without loss of simulation accuracy. The collected photons could also be mirrored over the x- or y-axes. By this method, a PSF designed to produce a square field can be increased in size by eight-fold with little computing necessary. Since the GK beams are circular, even more azimuthal redistribution of particles is possible. Elekta made use of APR when generating their convolution dosimetry algorithm. 9 For our model, we decided against using APR with the assumption that the beams were symmetric about the beam s central axis because the finite-sized, tilted source (see Figure 1) breaks the axial symmetry. This geometry is only symmetric when mirrored over the plane produced by the central axis of the tilted source and the beam s central axis. Mirroring produces a relatively modest doubling of PSF size. Although the 4 mm collimator s source geometry is symmetric about the azimuthal axis, we only mirrored
53 43 the 4 mm collimator s PSFs in order to keep parity with the other collimator s simulations. Collimator output factor calculation and dependencies Output factors are calculated by computing the dose deposited to a small volume of water at the center of the dosimetry sphere. We evaluated dose to a spherical volume 1.6 mm in diameter. The energy deposited in this volume, reported in ev, was recorded for each row s output. The absolute value of the dose is typically not critical outside of spectroscopy measurements, so only doses relative to the reference collimator outputs are reported here. In theory an output factor is the dose deposited at the infinitesimal point of maximum dose at the focus, but any useful measurement must take place over a finite volume. We measured the OF over a range of volumes from 0.5 to 4 mm diameter and found the 1.6 mm diameter produced OFs closest to the manufacturer s OF values. Our comparisons with Elekta s OFs should be interpreted only as degrees of agreement between different MC models. Computational Resources All MC-based radiation simulations are examples of embarrassingly parallel problems, a class of problems where pieces of the problem can be computed in parallel without the need for communication between the parallel tasks.
54 44 Computations were performed on Wake Forest s DEAC Linux cluster, a centrallymanaged high-performance computing environment consisting of ~140 computer nodes capable of performing about 1300 simultaneous jobs. 5 In this work between fifteen and seventy-five nodes were given different portions of the simulation at once and the results summed or merged. Our code provided each simulation instance with a different set of pseudorandom seeds generated by the Linux command $RANDOM in a Bash shell. Penelope internally generates its own random numbers thereafter. 10 At peak computation times, 75 jobs were performed simultaneously over approximately 15 hours per simulation. Under typical circumstances, depending on geometry, a node computed showers per second, where a shower is complete simulation of a primary particle and all secondaries. Radiochromic Film Radiochromic films are frequently used to measure ionizing radiation fields due to their high spatial resolution, low energy dependence, and near tissue-equivalence. 53,54 For this work, we used GafChromic EBT QD+ film (Ashland Specialty Products, Wayne, NJ) and an HP ScanJet 5550c scanner. Flat-bed document scanners like the one used have been shown capable of providing accurate dosimetry, within 1%. 55 EBT film is rated by the manufacturer as accurate over dose ranges of 1 cgy to 10 Gy measured in the red color channel. A dose versus pixel intensity calibration curve was generated by irradiating seven films at the isocenter of the 16 cm diameter solid water dosimetry sphere, as shown in Figure 18.
55 45 Due to the design of the solid water dosimetry sphere, films can be placed only in the axial (XY) and coronal (XZ) planes, and not in the sagittal (YZ) plane. Due to sector geometric symmetry about the longitudinal (Z) axis, in principle, correctly positioned coronal and sagittal films should yield the same dose distribution, however, differing source activities in each sector could conceivably result in slightly different dose distributions. Support Frame Z Film X Y Solid Water Sphere Figure 18: Irradiation setup for film calibrations with coordinate system for reference. Pieces of the solid water Elekta dosimetry sphere were removed to show the center of the setup. The yellow radiochromic film is held at the isocenter of the sphere, here shown in the coronal (XZ) plane. The film has two holes punched and solid water rods hold the film in place.
56 46 Doses from 0 to 300 cgy were planned in LGP and delivered at regular dose intervals using the 16 mm collimator of the Perfexion to provide a wide central area of uniform dose. Following established protocols, 55 the films were left in the dark for 1 hours then scanned in color at 300 dpi and saved in a lossless format. Using ImageJ 56 software, the red channel of each image was isolated and a central square of each film was cropped-out measuring 545 pixels / 4.61 cm on a side. These square images were re-saved in a lossless format. The square images were imported into Matlab. A central square.54 mm on a side was selected out of each image and the average pixel intensity was calculated over that 900 pixel area. The pixel intensity and dose were correlated and a 4th order polynomial was generated to fit the curve. The points and curve are shown in Figure 19. The dose range of up to 300 cgy was selected because the curve becomes progressively less sensitive at higher doses, as can be seen in Figure 19.
57 47 Figure 19: Pixel intensity value to dose calibration curve (dotted) and film measurement points. Using the calibration curve equation, a function was written in Matlab to convert the pixel intensity numbers in radiochromic film images to dose maps, as shown in Figure 0.
58 48 Figure 0: (Top row) The red channel portion of three radiochromic film images used in creating the calibration curve. The pixel intensity near the center of each image was correlated with the prescribed dose delivered by the GK. Images are 4.6 x 4.6 cm. (Bottom row) The same images with each pixel value converted to dose. Numbers along axes are number of pixels (545x545 pixels each). The colorbar indicates dose in cgy. Using this method, we were able to record dose distributions experimentally at high resolution for comparison with our MC results. Maximum Allowed Dose Difference Method A quantitative tool is needed to measure differences between dose distributions in order to verify the accuracy of dose calculation algorithms. The most straight-forward method of comparing calculated versus measured dose distributions on a pixel-by-pixel basis runs into problems very quickly in regions with steep dose gradients (~ 40%/mm in the
59 49 penumbra). 39 A method that accounts for deviations in both dose and position is required for comprehensive analysis. 57 The Maximum Allowed Dose Difference (MADD) method was developed by Steve Jiang as a general framework for comparing dose distributions in both dose- and spatial-domains. 58 MADD has been evaluated and used by at least 30 scholarly articles A Matlab code example of the method was acquired through personal correspondence with Dr Jiang and recoded into a Matlab function without modification. Proper operation of the code was confirmed using known test distributions included with the code. Like other similar methods, MADD compares a reference and a test dose distribution and assigns a pass or fail to each point in the reference distribution. The fraction of points that pass provides a simple quantitative measure of agreement between the distributions. Examining where failures occurred helps a physicist evaluate the clinical significance of the differences failures in a low-dose region are usually less clinically significant than failures in a high-dose region. MADD Methodology Dose distribution comparison methods start with a reference dose distribution (D r ) and a test dose distribution (D t ). The reference distribution is taken as the ground truth to evaluate the test distribution, 58 and is typically derived directly from experimental data. If the distributions differ in size/shape or resolution, the distributions are cropped to the area common to both distributions, and the test distribution is interpolated to match the reference point-for-point. This can mean direct three-dimensional (3D) to 3D comparison, as takes place between two different algorithms, or if the reference measurement is 3D as
60 50 with polymer gel dosimetry. Alternatively, two-dimensional (D) to 3D comparisons truncates the 3D data not matched to the D reference, which is typically radiochromic film or an array of ion chambers. The reference dose D r (r) is compared at each point r to the test dose D t in the nearby neighborhood, as shown in Figure 1. Figure 1: This illustration helps define some variables. The vertical axis D is dose and the horizontal axis r represents space. The origin (r, D r (r)) defines the comparison point r in the reference and the reference dose D r (r) at that point. The curve D t is the test dose distribution in the neighborhood of the origin. DD(r) is the dose difference at r and DTA(r) is the distance-toagreement along the r axis. (Illustration based on Jiang et al. 004) The dose difference DD(r), calculated in equation 5, is the difference between the test and reference distributions at r. DD( r) = D ( r) D ( r) (5) The distance-to-agreement DTA(r) is the distance along the r axis to a point where the test and reference distributions are equal. Figure 1 shows a one-dimensional (1D) test t r
61 51 dose distribution curve. When comparing D test dose distributions, there would be several test curves representing nearby test points in all directions from the reference point, and DTA(r) would be measured to the test curve nearest the origin. To evaluate what DD(r) and DTA(r) are acceptable, thresholds must be set. The predetermined dose tolerance δd 0 is the dose difference threshold, the maximum acceptable DD(r). The predetermined spatial tolerance δr 0 is the distance threshold defining the neighborhood in r where the test dose D t must match the reference D(r). These terms are defined visually in Figure. Figure : The vertical axis D is dose and the horizontal axis r represents one dimensional space, or distance from the reference point. δd 0 is the predetermined dose difference threshold. If the absolute dose difference DD(r) is less than δd 0, the comparison passes in dose at this point. δr 0 is the predetermined spatial tolerance. If the distance-to-agreement DTA(r) is less than δr 0, the comparison passes in the spatial domain at this point. Based on the curve shown, the point D r (r=0) fails in dose and passes in space. (Illustration based on Jiang et al. 004) Common criteria for evaluating dose differences vary from δd 0 =-5% and δr 0 =-5 mm for DTA. 39 The Comprehensive Cancer Center at Wake Forest University uses a 3% /
62 5 3mm criterion with the gamma index algorithm, discussed below, when evaluating Intensity Modulated Radiation Therapy (IMRT) plans for physics approval with a passing rate of 90% required. Intensity modulated treatments are one of the most complex radiation therapy modalities, so measurement of each planned field is required before treatment. When evaluating their own dose algorithms, Elekta uses the criterion of 3% / 0.5mm for dose and distance, respectively, which reflects the tighter mechanical tolerances of the Perfexion relative to a LINAC. This study used the same 3% / 0.5mm criterion when evaluating dose distributions. Several methods of applying these thresholds in dose and space exist to compare distributions. The simplest, the dose difference test, evaluates the reference and test doses at point r only, where point r passes if DD(r) < δd 0. Composite analysis 65 passes a point r if it passes the dose-difference test or if the DTA is under the threshold: if DD(r) < δd 0 or DTA(r) < δr 0. Composite analysis is shown in Figure 3(a).
63 53 Figure 3: Acceptance regions are shaded in the spatial-dose domain for different dose comparison methods. The vertical axis represents dose and the horizontal axis represents space. (a) Composite analysis passes the reference point if the test curve crosses the vertical axis between 0 and 1, or if the test curve crosses the horizontal axis between 0 and 1; (b) γ index passes if the curve traverses the shaded area; (c) box method passes if the curve traverses the shaded area. (Illustration based on Jiang et al. 004) The γ-index method 66 is a popular method that combines dose and spatial thresholds into a single pass/fail test. This method is defined in equation (6) The coordinate (r,d t (r )) is defined as the point on the test curve closest to the origin at (r,d r (r)). r= r-r and D= D r (r)- D t (r ). As shown in Figure 3 (b), the effect of equation 6 is that any curve crossing the ellipse limited by the axes δr 0 and δd 0 passes the γ-index test for that point. The γ-index method is used at Wake Forest for evaluating IMRT plans using MapCheck software and the ArcCheck measurement device by Sun Nuclear. 67
64 54 The box method, seen in Figure 3(c), simplifies the γ-index method by passing any test point within the box about the origin limited by δr 0 and δd 0 in the dose versus space coordinate system. The MADD algorithm uses a modified version of the box method when passing or failing evaluation points. MADD turns the pass/fail test from one in the spatial and dose domains to a test at a single point in the dose domain only. As shown in Figure 4, the test dose distribution is shifted in the dose direction (vertically) until it is tangent to the ellipse of the γ-index acceptance area. The value MADD(r) is calculated as the shifted test dose D t (r) evaluated at r. The evaluation point passes if the dose difference DD(r) is less than or equal to MADD(r) at r.
65 55 Figure 4: Schematic of MADD method: The vertical axis D is dose and the horizontal axis r represents space. δd 0 and δr 0 are the predetermined dose difference and spatial difference thresholds. The ellipse is the acceptance region defined by the γ-index method. The solid line is the test dose distribution D t and the dashed line is the shifted test dose distribution D t which has been shifted in the dose direction (vertically) until the curve is tangent to the acceptance region ellipse. MADD is the distance along the D axis from the origin (r, D r ) to the shifted test dose D t (r). The test point passes if DD(r) is less than or equal to MADD. (Illustration based on Jiang et al. 004) The degree of similarity between a reference and test dose distribution is given by the pass rate, which the percentage of reference points that passed the analysis method s (gamma, MADD, etc) test. One inherent flaw in any dose distribution comparison method is the inclusion of irrelevant points in the comparison. For example: A tiny radiation field that occupied only 5% of the measured area could, when compared to a zero-dose field, get a 95% pass rate since only 5% of the points differ from the reference, uniform dose of zero. It is important therefore to either manually define a region-of-interest (ROI) that eliminates
66 56 low-dose areas from that pass-rate analysis, or apply a threshold to define the ROI. The industry standard threshold for gamma index analysis and similar methods is 10% of the maximum dose, meaning that the pass/failure of points with doses less than 10% of maximum are ignored when counting up the passing fraction. 68 Some articles 39 that make use of MADD analysis, however, do not clearly apply a threshold ROI, so in this work we report pass rates both with and without a 10% threshold applied. The dose/space pass criteria of 3% / 0.5 mm, and threshold of 10% were selected to allow direct comparison with other users and articles that used the same pass criteria and threshold.
67 57 RESULTS Energy Spectra A simple check on the validity of the simulation is found in the photon energy spectrum delivered to the dosimetry sphere, shown in Figure 5. This energy spectrum matches the expected results for radioactive decay of Cobalt-60, as shown previously in Figure 4. This result indicates that both photon generation and subsequent interactions are being modeled properly. Figure 5: Photon spectra delivered to the dosimetry sphere. The two energy peaks for unscattered photons from Cobalt-60 (γ 1 = 1.17 MeV, γ = 1.33 MeV) are clearly visible, along with other scattering phenomena.
68 58 The two Cobalt-60 photopeaks are clearly visible in Figure 5 and 6. Compton edges are also seen at 1.1 MeV and MeV, as predicted by equation 1, along with decaying probability of interaction to the left of the edges, as predicted by the Klein- Nishina cross section from Figure 3. The total number of interactions from a given collimator s photons scales roughly with the area of the collimator. Figure 6: Normalized energy spectra of photons delivered to the dosimetry sphere. The counts for each collimator are normalized to the bin with the maximum number of counts, at the 1.17 MeV Co-60 peak for all 3 collimators. The 4 mm collimator has the largest relative number of scatter counts at lower energy, indicating a larger portion of the dose delivered by the 4 mm collimator is from scattered photons than the other two collimators. Figure 6, which normalizes the spectra from Figure 5 by setting the highest peak in each spectrum equal, shows the relative number of counts in the low energy scatter area
69 59 is higher for the 4 mm collimator than the other two collimators, and higher in the 8 mm collimator than the 16 mm collimator. The normalizing of the counts exaggerates this effect. Un-normalized, as in Figure 5, the number of photons delivered to the target visibly scales with the cross-sectional area of the collimator. Photons traveling down the narrower collimator have a smaller chance of making it through without interacting along the way than those in a larger collimator, so a larger proportion of photons as low energy scatter makes physical sense. Output Factors Collimator OFs and row OFs were calculated by measuring the dose to a small sphere of water at the center of the 16 cm diameter Elekta dosimetry sphere, simulated as a sphere of water in Penelope. Since there is no clear consensus on what the volume should be to measure the OFs, we varied the size of the measurement volume to find a volume that produced results close to the manufacturer s values. Table II and Figure 7 shows how OF results vary with the volume over which they are measured.
70 60 Meas. Volume Diameter (cm) 4 mm Collimator OF 8 mm Collimator OF ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± TABLE II: Collimator OFs calculated by MC in water spheres of varying diameters located at the radiation focus at the center of the Elekta dosimetry sphere. OFs decrease as the measurement volume increases. Uncertainty in the OF result decreases as the measurement volume increases due to the larger number of interactions incorporated into the sum. These data are shown graphically in Figure 7 with a larger number of volumes measured.
71 61 (A) (B) (C) (D) Figure 7: For all images: the horizontal axis is the diameter of the sphere over which dose was measured, and the vertical axis is the output factor calculated over a spherical dosimeter of the given volume. The horizontal dashed lines are the reference OF values provided by the manufacturer. A MC-calculated OF exactly agrees with the manufacturer s value at the point where the solid MC curve crosses the dashed reference line of the same color. (A) The collimator output factors show the strong dependence of the 4 mm collimator on the volume of measurement. The error bars around each point show how uncertainty in OF results decreases as the volume of measurement increases, due to the larger number of interactions measured in the larger volumes. (B-D) The row output factors for the 16 mm, 8 mm, and 4 mm collimator. The dotted lines at the bottom of each plot are the absolute difference between the MC OF value measured over that diameter and the reference value; a small difference indicates a good choice of measurement diameter.
72 6 Figure 7(D) shows that the difference between our 4 mm row OF calculations and Elekta s values comes to a minimum when we measure over a 1.6 mm diameter volume. The other collimator and row OF results do fairly well at the 1.6 mm measurement volume as well, so we selected the 1.6 mm diameter sphere as our measurement volume for all OF calculations. Collimator Output Factors Collimator OFs are calculated as the dose delivered at the focus by the 4 mm or 8 mm collimators compared to the dose delivered to the same volume by the 16 mm collimator. Our collimator OF results are reported in Table III. Collimator MC Output Factor Elekta Output Factors % Difference 16 mm 1.000* mm 0.904± % 4 mm 0.806± % TABLE III: MC-calculated OFs measured over a 1.6 mm diameter sphere at the center of the 16 mm diameter dosimetry sphere. *The 16 mm collimator is normalized as the reference value. Both MC calculated results differ from the Elekta values by more than the statistical MC uncertainty, but by less than 1%.
73 63 Row Output Factors Row OFs are the dose delivered by a single source in a particular row at the focus divided by the dose delivered by the 16 mm collimator s row B to the same measurement volume. The 16 mm collimator s row B is used as the reference value because the dose it delivers to the focus is the greatest. 16 mm row B s sources are closest to the focus, which due to the inverse square law, means the dose delivered is greater than more distant rows. Table IV shows our row OF results measured over the 1.6 mm diameter dosimeter. All results were within.5% of the Elekta recommended values. The greatest deviation from manufacturer values was seen with the 8 mm collimator row OFs, one of which was.44% higher than expected.
74 64 Collimator 16 mm 8 mm 4 mm Row MC Elekta Output Factor Output Factor % Difference A 0.968± % B 1.000* C 0.987± % D 0.94± % E 0.85± % A 0.944± % B 0.99± % C 0.883± % D 0.796± % E 0.75± % A 0.799± % B 0.81± % C 0.798± % D 0.79± % E 0.667± % TABLE IV: Row OF results for each row of each collimator. *16 mm row B is the highest-output reference row. All row OF results are within.5% of Elekta s recommended values. Comparison of Profile Curves The single sector dose distributions generated by MC and the TMR10 algorithm of LGP were initially compared using their dose profile curves. Figures 8 show the profile curves at 0.5 mm resolution (100 points per curve) along all axes from irradiation with one sector. This comparison is for Sector 1, which is oriented in the vertical plane from anterior to posterior along the Y-axis, (from top to bottom in Figures 7 and 9).
75 65 (A) (B) (C) Figure 8: Dose profile curves along all axes from a single sector of the 16 mm collimator. The blue curves were calculated by MC code; the green curves were generated by LGP s TMR10 algorithm.
76 66 (A) (B) (C) Figure 9: Dose profile curves along all axes from a single sector of the 8 mm collimator. The blue curves were calculated by MC code; the green curves were generated by LGP s TMR10 algorithm.
77 67 (A) (B) (C) Figure 30: Dose profile curves along all axes from a single sector of the 4 mm collimator. The blue curves were calculated by MC code; the green curves were generated by LGP s TMR10 algorithm. The output (height) of each curve matches the output factor for the calculation method that generated the profile. The MC curves have collimator OFs of and 0.806, and the TMR10 curves show collimator OFs of and for 8 mm and 4 mm collimators, respectively. If a correction factor was applied to match the collimator OFs (meaning, use of the same normalization factor), in most cases the profiles would show closer agreement. The profile along the Y-axis is elongated, as expected for the use of
78 68 one sector only. For instance, Figure 10 shows the effect of using blocked sectors or different sector combinations with Figure 10a showing an elongated distribution along the Y-axis when two opposed sectors are used(sectors 1 and 5). The MC profiles have slightly irregular shapes relative to the smooth TMR10 profiles due to the noise inherent to MC simulations, and the fact that the MC curves have not been artificially smoothed. Figures 8-30 show good agreement between the dose distributions generated by MC and TMR10. The dose differences appear to exist more in the spatial dimension than the dose dimension in most cases. For example, in Figure 8(A), the left sides of the curves do not overlap but have very similar shapes, indicating the problem may be due to misalignment between coordinate systems. In some cases the offset between the profiles is smaller than the resolution of the data, as in Figure 8(A). In other cases, such as the right side of Figure 8(B), the offset is significant. The single-sector dose profile curves are reproduced in the Appendix, Figure 54, with the addition of error bars for MC data. Figures show profile curves for irradiation from all eight sectors for each collimator, a full shot of dose, as calculated at the focus in the Elekta dosimetry sphere by MC and LGP s TMR10 algorithm.
79 69 (A) (B) (C) Figure 31: Dose profile curves along the x, y, and z axes from all eight sectors of the 16 mm collimator. The blue curves were calculated by MC code; the green curves were generated by LGP s TMR10 algorithm.
80 70 (A) (B) (C) Figure 3: Dose profile curves along the x, y, and z axes from all eight sectors of the 8 mm collimator. The blue curves were calculated by MC code; the green curves were generated by LGP s TMR10 algorithm.
81 71 (A) (B) (C) Figure 33: Dose profile curves along the x, y, and z axes from all eight sectors of the 4 mm collimator. The blue curves were calculated by MC code; the green curves were generated by LGP s TMR10 algorithm. The full profile curves show strong agreement between the MC and TMR10 algorithms. As in Figures 8-30, the height of each curve matches the collimator OF of the algorithm that generated the curve. In some places gaps between the curves, such as Figure 9(A) and (B), are of a width less than the resolution of the figure. Attempts to improve agreement between the curves in
82 7 these figures by shifting one curve over by 0.5 mm creates a gap of equal size on the opposite side of the curve. This offset difference can be seen by comparing Figure 3 (A) and (B). The x-axis and y-axis profiles should be identical in this geometry, but the offset by a single pixel makes the two figures mirror images of one-another. For reference, Figure 34 shows the y-axis profile curves generated by the MC model created for the author s M.S. thesis. 69 In this geometry the x-axis curve is identical to the y-axis curve.
83 73 (A) (B) (C) Figure 34: Dose profile curves along the y-axis from all eight sectors of the 16, 8, and 4 mm collimator in A, B, and C, respectively, as calculated by the version of the simulation used for the author s M.S. thesis. The blue curves were calculated by MC code; the black dotted curves were generated by LGP s TMR9 algorithm, and the black solid curve is based on a film measurement performed by an Elekta engineer. Collimator OF values were ignored for these curves.
84 74 The dose profile curves previously created for the M.S. thesis were generated by rotating the dose distribution from a single source model with a constant angular increment about the z-axis as opposed to simulating the particles from those different angles within the MC model itself. The LGP curves in Figure 34 were created by the old TMR9 algorithm. The former disagreements in dose at the shoulders of the 16 mm and 8 mm collimator profiles are absent in the new profile curves. MC Comparison with Elekta-Calculated Dose Distributions LGP s TMR10 algorithm was used to create a dose distribution from a single sector for each collimator size delivered to a model of the Elekta dosimetry sphere. As shown in Figure 9, a single sector consists of 4 sources. All single-sector experimental data in this report use sector #1, the sector anterior to the patient. All sectors are mechanically identical and symmetric about the z-axis in 45 degree steps for eight total sectors. The TMR10-calculated dose distribution of a single sector s output was exported in the DICOM-RT format at a resolution of 0.5x0.5x0.5 mm voxel size, the smallest size available. The exported dose distribution was imported into Matlab and compared with a Monte Carlo-generated dose distribution of the same geometry and spatial resolution. The two dose distributions centered about the focus of the Perfexion, representing cubes.5 cm on a side. The dose distributions were shifted relative to one-another up to 1 mm in each direction to optimize for dosimetric agreement. Shifts are sometimes used when comparing dose distributions to account for setup error and differences in positions of the isocenter or focus between distributions. The coordinates of the focus are not directly
85 75 reported in the DICOM-RT files exported by LGP, so shifts are necessary to manually determine the position of the focus within the LGP-generated matrix. The two relative dose distributions were compared using the MADD method with a 3% / 0.5 mm pass criteria. The passing rates for each collimator are shown in Table V. Collimator MADD Pass Rate MADD Pass Rate (no threshold) (10% threshold) 16 mm 99.97% 56.87% 8 mm 99.55% 91.45% 4 mm 99.8% 70.93% TABLE V:.5 cm cube dose distributions at 0.5 mm resolution generated by LGP and our MC model were compared using the MADD algorithm with 3% / 0.5 mm criteria. The MADD pass rate is the percent of passing points in the reference volume (LGP) when compared to the test volume (MC). Contour plots of the single sector dose distributions generated by LGP s TMR10 algorithm and our MC code are shown in Figures Both TMR10 and MC generate three-dimensional dose distributions, so the three principal planes are shown. A single sector of the 16 mm collimator is shown in Figure 35, while the 8 mm and 4 mm collimator single sector distributions are shown in Figures 36 and 37, respectively.
86 76 16 mm Collimator Dose Distributions from TMR10 and MC Calculations Figure 35: Planes of dose distributions from a single sector of the 16 mm collimator are shown on the left side. The right side shows points in the same plane that fail the MADD algorithm test, with no 10% threshold used. The axial (XY) plane, sagittal (ZY), and coronal (XZ) planes are shown. Since the comparison is between 3D distributions, many failing points are not in the selected planes and are not shown.
87 77 8 mm Collimator Dose Distributions from TMR10 and MC Calculations Figure 36: Planes of dose distributions from a single sector of the 8 mm collimator are shown on the left side. The right side shows points in the same plane that fail the MADD algorithm test, with no 10% threshold used. The axial (XY) plane, sagittal (ZY), and coronal (XZ) planes are shown. Since the comparison is between 3D distributions, many failing points are not in the selected planes and are not shown.
88 78 4 mm Collimator Dose Distributions from TMR10 and MC Calculations Figure 37: Planes of dose distributions from a single sector of the 4 mm collimator are shown on the left side. The right side shows points in the same plane that fail the MADD algorithm test, with no 10% threshold used. The axial (XY) plane, sagittal (ZY), and coronal (XZ) planes are shown. Since the comparison is between 3D distributions, many failing points are not in the selected planes and are not shown.
89 79 By looking at the contour plots in Figures side by side with the points of failure, it s possible to see what regions of the models disagree in terms of dose. Another way to view the points of failure is by looking at the distance between the same contour levels between models, here indicated by the same color of solid versus dashed lines. Figures are magnified versions of Figures to allow closer examination of those contour lines by cropping out the low-dose regions. Note that this transformation in some cases makes the axes unequal. 16 mm Collimator Dose Distributions from TMR10 and MC Calculations Figure 38: Planes of dose distributions from a single sector of the 16 mm collimator. The axial (XY) plane, sagittal (ZY), and coronal (XZ) planes are shown. The vertical and horizontal dimensions are not equal in every image.
90 80 8 mm Collimator Dose Distributions from TMR10 and MC Calculations Figure 39: Planes of dose distributions from a single sector of the 8 mm collimator. The axial (XY) plane, sagittal (ZY), and coronal (XZ) planes are shown. The vertical and horizontal dimensions are not equal in every image. 4 mm Collimator Dose Distributions from TMR10 and MC Calculations Figure 40: Planes of dose distributions from a single sector of the 4 mm collimator. The axial (XY) plane, sagittal (ZY), and coronal (XZ) planes are shown. The vertical and horizontal dimensions are not equal in every image.
91 81 TMR10 Comparison with Film Measurements To provide a measure of validation for the film dosimetry measurements, the previous section s TMR10-generated single-sector irradiations were compared to film measurements of the same irradiation setup. The same film measurements are compared to the MC-generated dose distributions in the following section. Film measurements of a single-sector irradiation were taken in the plane of the focus at the center of the solid water dosimetry sphere as shown in Figure 18. Film pixel intensities were converted to dose and compared to MC-calculated dose distributions in the same plane, as shown in Figures 19 and 0. Table VI shows the results of MADD comparisons between film and LGP dose distributions. Collimator 16 mm 8 mm 4 mm Plane MADD Pass Rate MADD Pass Rate (no threshold) (10% threshold) Axial (XY) 8.6% 6.3% Coronal (XZ) 9.0% 5.4% Axial (XY) 86.5% 55.% Coronal (XZ) 96.9% 66.8% Axial (XY) 98.9% 87.1% Coronal (XZ) 97.3% 63.1% TABLE VI: MADD pass rates from comparisons between film and LGP-calculated dose distributions with the 3% / 0.5 mm MADD criteria. The film versus TMR10 results show high pass rates when no threshold was applied, and significantly lower pass rates when a threshold was in place. Figures compare the
92 8 film dose distributions to TMR10 calculations in contour plots and figures showing which points passed and failed the MADD algorithm test, starting with results in the axial plane. Figures show the MADD test results in the coronal plane and axial planes, including magnified images for better detail.
93 83 16 mm Collimator Dose Distributions from Film Measurements and LGP Calculations Axial Plane (A) (B) (C) Figure 41: (A) Normalized film and LGP-calculated dose distributions from one sector of the 16 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
94 84 8 mm Collimator Dose Distributions from Film Measurements and LGP Calculations Axial Plane (A) (B) (C) Figure 4: (A) Normalized film and LGP-calculated dose distributions from one sector of the 8 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
95 85 4 mm Collimator Dose Distributions from Film Measurements and LGP Calculations Axial Plane (A) (B) (C) Figure 43: (A) Normalized film and LGP-calculated dose distributions from one sector of the 4 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
96 86 16 mm Collimator Dose Distributions from Film Measurements and LGP Calculations Coronal Plane (A) (B) (C) Figure 44: (A) Normalized film and LGP-calculated dose distributions from one sector of the 16 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
97 87 8 mm Collimator Dose Distributions from Film Measurements and LGP Calculations Coronal Plane (A) (B) (C) Figure 45: (A) Normalized film and LGP-calculated dose distributions from one sector of the 8 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
98 88 4 mm Collimator Dose Distributions from Film Measurements and LGP Calculations Coronal Plane (A) (B) (C) Figure 46: (A) Normalized film and LGP-calculated dose distributions from one sector of the 4 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
99 89 Though the contour plots in many cases show strong agreement between the film and TMR10 dose distributions, the MADD algorithm with a 10% threshold applied gave pass rates of 50-60% in most cases, indicating significant differences in the dose distributions in high dose rate areas. Another possibility is that the film dosimetry data are inaccurate. Film data are susceptible to thin-film interference artifacts, like those evident in the sawtooth-patterns in Figure 43(C), as well as other problems relating to the accurate measurement of absolute dose. MC Comparison with Film Measurements The films discussed in the previous section were compared to the MC-calculated dose distributions. Table VII shows the results of MADD comparisons between film and MC dose distributions. Collimator 16 mm 8 mm 4 mm Plane MADD Pass Rate MADD Pass Rate (no threshold) (10% threshold) Axial (XY) 87.4% 7.9% Coronal (XZ) 93.0% 59.0% Axial (XY) 97.1% 87.3% Coronal (XZ) 98.4% 73.0% Axial (XY) 97.9% 77.4% Coronal (XZ) 98.3% 83.5% TABLE VII: MADD pass rates from comparisons between film and MC-calculated dose distributions using the 3% / 0.5 mm criteria. Figures 47-5 show contours of the paired dose distributions in the axial (XY) and coronal (XZ) planes, and the points of failure for each distribution.
100 90 16 mm Collimator Dose Distributions from Film Measurements and MC Calculations Axial Plane (A) (B) (C) Figure 47: (A) Normalized film and MC-calculated dose distributions from one sector of the 16 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
101 91 8 mm Collimator Dose Distributions from Film Measurements and MC Calculations Axial Plane (A) (B) (C) Figure 48: (A) Normalized film and MC-calculated dose distributions from one sector of the 8 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
102 9 4 mm Collimator Dose Distributions from Film Measurements and MC Calculations Axial Plane (A) (B) (C) Figure 49: (A) Normalized film and MC-calculated dose distributions from one sector of the 4 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
103 93 16 mm Collimator Dose Distributions from Film Measurements and MC Calculations Coronal Plane (A) (B) (C) Figure 50: (A) Normalized film and MC-calculated dose distributions from one sector of the 16 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
104 94 8 mm Collimator Dose Distributions from Film Measurements and MC Calculations Coronal Plane (A) (B) (C) Figure 51: (A) Normalized film and MC-calculated dose distributions from one sector of the 8 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
105 95 4 mm Collimator Dose Distributions from Film Measurements and MC Calculations Coronal Plane (A) (B) (C) Figure 5: (A) Normalized film and MC-calculated dose distributions from one sector of the 4 mm collimator. (B) Points which failed the MADD algorithm test with 3% / 0.5 mm criteria. (C) Dose distribution contours with low dose areas cropped out, making the axes unequal.
106 96 Although the isodose plots in most cases appear to show strong agreement between dose distributions, the 3% / 0.5 mm criteria for the MADD comparison is very tight in the spatial domain, as seen in the points of failure plots. Radiochromic film dosimetry has challenges for precision that include film positioning errors for irradiation as well as acquisition errors during scanning. The dose comparisons for film versus the TMR10 and MC-calculated dose distributions show similar differences that relate to film accuracy. The more important comparison is for direct comparison of the TMR10 and MCcalculated dose distributions, as shown in Figures
107 97 DISCUSSION The photon energy spectra in Figures 5-6 show scattering from the collimator of a Cobalt-60 source consistent with our expectations. Compton edges are visible at energy levels predicted by theory. More low energy scatter was found with the smaller collimators spectra, indicating proportionally more photons scattering occurred in smaller diameter collimators. The level of low-energy scatter is greater in the current model than in the author s previous studies. The most significant changes made from the previous model were 1) shielding surrounding the source, leaving photons only one unshielded exit from the source volume, and ) a larger initial photon cone angle. The previous model emitted photons in a 6 degree cone, whereas the current model uses a 45 degree cone. Many of the low-energy photons observed in the current model s spectra may have scattered in the source housing before proceeding down the collimator with reduced energy. By analyzing output factors as a function of measurement volume, as seen in Figure 7, we were able to determine a measurement volume to enable a best possible comparison for the OFs calculated here and those calculated by Elekta. As expected, the 4 mm collimator was most sensitive to changes in the OF measurement volume. The 8 mm collimator s output was constant until the measurement volume extended beyond the central high-dose plateau and started incorporating penumbra dose. The 16 mm collimator s output was constant up to the largest measurement volume, a 5 mm diameter sphere, consistent with expectations.
108 98 The calculated collimator OFs were within 1% of the values used by Elekta. The difference is greater than the uncertainty for both the 4 mm and 8 mm collimator OFs, but so small as to likely be due to micrometer-scale differences between the simulation models. The row OFs calculated by MC were no more than.5% different from the row OFs used by Elekta. The 16 mm row OFs relative to the output of row B were all higher than the Elekta values, with the row D value greatest at 1.05% high. The 8 mm row OFs showed the greatest differences from the standard. All were higher than expected, with the row E OF most of all at.44% high. The 4 mm OFs were nearly all less than 1% different from Elekta s values, with the greatest difference seen in row A at -1.60%. The OF measurement volume was selected with the 4 mm OF in mind since it is most susceptible to minute changes in measurement technique, so the strong agreement was expected. The 1D profile curves were the first tool used to compare the dose distributions. They provide a good qualitative measure of agreement between MC and TMR10 algorithms. In most cases, the dose disagreements were due to offsets in space rather than outright dose differences. The misalignments were smaller than the spatial resolution of the data, 0.5 mm, indicating strong agreement. Higher resolution sampling would allow better alignment of offset curves, but has not yet been attempted due to the large number of photon histories needed. The 3D dose distributions generated by TMR10, and our MC model enable direct comparison of 3D calculated dose distributions. The MADD algorithm allowed a
109 99 quantitative comparison between the dose distributions using the 3% / 0.5 mm criteria. Table V compares the results of the single sector simulations. While the overall pass rate was very high, >99.5%, in all cases the results were less convincing once the 10% dose threshold was applied. The 8 mm collimator s sector performed best with a pass rate of 91.5%. The 4 mm collimator sector s distribution has the fewest points above the threshold dose, and we found 70.9% of those points passed. The 16 mm collimator had the lowest pass rate after the threshold was applied. Only 56.9% of points above the threshold passed the MADD test. Examining the points of failure in Figure 35, it appears that many failing points are on the superior/anterior side of the dose distribution. This grouping may indicate a modeling problem with the sources in rows D and E, or perhaps a bias in the spatial registration of the two dose distributions for the comparison process. Since the dose distributions generated by LGP s TMR10 have been thoroughly examined by Elekta and outside parties, especially within the dosimetry sphere setup, the comparison of TMR10 dose distributions to film measurements provided a measure of validation for the film measurements. Film can be a difficult media to work with, so we were not surprised to see that our film results did not exactly match the TMR10 prediction. As seen in Table VI, when no threshold was applied, good pass rates were seen (>90%) for all but the axial plane of 16 mm and 8mm collimators; these two results had pass rates above 80%. With a 10% threshold applied, the 16 mm collimator s pass rates were 6% and 5% in the axial and coronal planes, respectively. The 8 mm collimator s pass rates were in the same range. The 4 mm collimator s pass rates were 87% and 63% in axial and coronal planes, respectively. This higher pass rate for the 4
110 100 mm collimator reflects the fact that no flat plateau of high dose rate exists for the 4 mm collimator the high dose rate areas are also high gradient areas. The MADD algorithm assigns higher allowed-dose-differences to high gradient areas, so points with dose differences >3% will often pass. Though the film measurements were shown to be imperfect by the comparison with TMR10 results, film provided a method for directly comparing our MC measurements to experimental results. Table VII shows the MADD pass rates for these comparisons, again at the 3% / 0.5 mm pass level. Generally lower pass rates were expected for this D association versus the 3D comparisons handled in the tests with TMR10. In a 3D matrix, a given point is immediately adjacent to 6 points, conforming to the corners, edges, and faces of a cube (nearest neighbors, including slant distances). In a D matrix, only 8 points are one step away, as in the edges and corners of a square. A 3D matrix therefore provides over three times as many nearby points for comparison, increasing the odds of passing. The film test results were positive, but show the need for improvement. With no threshold applied, the lowest pass rate was 87.4% for the 16 mm collimator in the axial (XY) plane. The 16 mm collimator s coronal (XZ) plane was also fairly low at 9.8%. The 8 mm and 4 mm collimator s films all had greater than 97% pass rates with no threshold applied. Applying the 10% threshold lowered all of the film pass rates. The most significant drop in pass rate was seen in the coronal (XZ) plane of the 16 mm collimator, passing only
111 % of points. The 8 mm and 4 mm collimators passed approximately 70-85% points each. Looking at the points of failure in Figures 47, it is clear that the 16 mm collimator s areas of failure were concentrated in the flat regions of the 50-90% isodose levels. Pass rates were high in high-gradient areas. One effect of the MADD algorithm is higher tolerance of dose differences in the high-gradient areas. The flat, central regions are low-gradient in dose, and are therefore held to tighter dosimetric standards by this evaluation method. Comparing the MADD pass rates in Tables VI and VII allow a direct comparison of whether the MC or TMR10 dose distributions are closer to the film dose distributions. In every case but one, the 4 mm collimator XY plane, the MC algorithm produced higher pass rates than TMR10, indicating the MC algorithm more closely reflects the experimental dose distribution as reported by film measurements. However, as noted previously, the radiochromic film measurement technique and analysis would benefit from improvements in dosimetric fidelity, particularly for radiosurgical applications.
112 10 CONCLUSIONS Our MC-based model of the GK Perfexion represents a significant step toward a thirdparty implementation of GK PFX dosimetry. Our OF results showed strong conformance with the existing models and there is excellent agreement for MC versus TMR10 dose profiles for a single sector alone and for all eight sectors combined. The MADD analysis of our dose distributions using the manufacturer s dosimetry model and our own film measurements shows room for improvement, particularly for comparisons using radiochromic film. The film result showing the MC algorithm outperformed the TMR10 algorithm is promising, however, a more thorough, expert film dosimetry analysis is needed, including the possibility of an improved film scanner. The MC algorithm and possibly the Elekta algorithm could use significant improvements. All our analyses thus far have focused on the most idealized geometry. All simulations in this work were of the simplest dosimetry the Perfexion can generate and were carried out with a target in the center of a sphere of water, and compared to experiments in a sphere of solid water. Additionally, the Cobalt-60 sources were modeled as an individual stack of point sources at 1 mm intervals a future improvement would be to model the sources as linear, or a cylindrical volume. Future comparisons should be made to more complex, heterogeneous geometries and with more complex plans. Export of the PSFs generated by this work to third party academic researchers will allow new applications of our model to unique research geometries and animal anatomies, where the model should theoretically outperform the existing dosimetry models that assume homogenous volumes of materials. Given the high failure rates of the MADD analysis, significantly more work is needed to investigate both radiochromic film dosimetry techniques and the MADD analysis as
113 103 applied to radiosurgical dose distributions. This model is not recommended for use for human dosimetry in the clinical setting, however, investigations of the model s performance under conditions of inhomogeneities will be the focus of future work.
114 104 REFERENCES 1. Attix, F.H. Introduction to Radiological Physics and Radiation Dosimetry, (John Wiley & Sons, New York, 1986).. Das, I., J., George, X.D. & Anders, A. Small fields: Nonequilibrium radiation dosimetry. Medical Physics 35, (008). 3. Khan, F.M. The Physics of Radiation Therapy, (Lippincott Williams & Wilkins, Philadelphia, PA, 003). 4. Task Group 1, R.T.C., American Ass'n of Physicists in Medicine. A protocol for the determination of absorbed dose from high-energy photon and electron beams. Medical Physics 10, (1983). 5. Almond, P.R., et al. AAPM's TG-51 protocol for clinical reference dosimetry of highenergy photon and electron beams. Medical Physics 6, (1999). 6. Purdy, J.A. & Klein, E.E. Photon External-Beam Dosimetry and Treatment Planning. in Principles and Practice of Radiation Oncology (eds. Halperin, E.C., Perez, C.A. & Brady, L.W.) (Lippincott Williams & Wilkins, Philadelphia, PA, 008). 7. Li, X.A., Soubra, M., Szanto, J. & Gerig, L.H. Lateral electron equilibrium and electron contamination in measurements of head-scatter factors using miniphantoms and brass caps. Medical Physics, (1995). 8. Turner, J.E. Atoms, Radiation, and Radiation Protection, (John Wiley & Sons, Inc., New York, 1995). 9. Berger, M.J., et al. XCOM: Photon Cross Section Database (version 1.5). [Online] (National Institute of Standards and Technology, Gaithersburg, MD, 010). 10. Salvat F, F.-V.J.M.a.S.J. PENELOPE - A Code System for Monte Carlo Simulation of Electron and Photon Transport, (NEA-OECD, Paris, 003). 11. Singh, K. & vardhan V, A. Chemical Sciences : A manual for CSIR-UGC National Eligibility Test for Lectureship and JRF [UGC-CSIR NET] CSIR-UGC National Eligibility Test (NET) for Junior Research Fellowship and Lecturer-ship. (ed. Singh, K.) (Wikibooks, 01). 1. Sauter, F. Über den atomaren Photoeffekt in der K-Schale nach der relativistischen Wellenmechanik Diracs. Ann. Phys. 11, (1931). 13. Knoll, G.F. Radiation Detection and Measurement, (Wiley & Sons, Ann Arbor, MI, 000).
115 Sauer, O.A. & Wilbert, J. Measurement of output factors for small photon beams. Medical Physics 34, (007). 15. Seuntjens, J. & Verhaegen, F. Comments on 'Ionization chamber dosimetry of small photon fields: A Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams'". Physics of Medicine and Biology 48, L43-45 (003). 16. Kijewski, P.K., Bjangard, B.E. & Petti, P.L. Monte Carlo calculations of scatter dose for small field sizes in a 60C beam. Medical Physics 13, p74-77 (1986). 17. Ding, G.X., Duggan, D.M. & Coffey, C.W. Commissioning stereotactic radiosurgery beams using both experimental and theoretical methods. Phys Med Biol 51, p (006). 18. Ding, G.X. Dose discrepancies between Monte Carlo calculations and measurements in the buildup region for a high-energy photon beam. Medical Physics 9, p (00). 19. Paelinck, L., Reynaert, N., Thierens, H., Neve, W.D. & Wagtner, C.D. Experimental verification of lung dose with radiochromic film: Comparison with Monte Carlo simulations and commercially available treatment planning systems. Phys Med Biol 50, p (005). 0. Kahn, H. Random sampling (Monte Carlo) techniques in neutron attenuation problems--i. Nucleonics 6, 7 (1950). 1. Rogers, D.W. Fifty years of Monte Carlo simulations for medical physics. Phys Med Biol 51, R (006).. Andreo, P. Monte Carlo techniques in medical radiation physics. Phys Med Biol 36, (1991). 3. Berger, M.J., Wang, R. Multiple-scattering angular deflections and energy-loss straggling. in Monte Carlo Transport of Electrons and Photons (ed. T M Jenkins, W.R.N., A Rindi) p1-56 (Plenum, New York, 1988). 4. Berger, M.J., Wang, R. Monte Carlo calculation of the penetration and diffusion of fast charged particles. Methods in Computational Physics v1(1963). 5. Baro J, S.J., Fernandez-Varea J M, Salvat F. PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter. Nuclear Instruments and Methods in Physics Research B 100, p31-46 (1995). 6. Sempau, J., et al. Monte Carlo simulation of electron beams from an accelerator head using PENELOPE. Phys Med Biol 46, (001). 7. Amponsah, K., et al. Staged Gamma Knife Radiosurgery for Large Cerebral Arteriovenous Malformations. Stereotactic and functional neurosurgery 89(011).
116 Foote, R.L., et al. Stereotactic radiosurgery using the gamma knife for acoustic neuromas. International journal of radiation oncology, biology, physics 3, (1995). 9. Elekta, A.B. The Convolution algorithm in Leksell GammaPlan 10. (ed. Elekta, A.B.) 8 (Stockholm, Sweden, 011). 30. Das I.J., K.A., Verhaegen F, Moskvin, V.P. Interface Dosimetry: measurements and Monte Carlo simulations of low-energy photon beams. Radiation Physics and Chemistry 61, p (001). 31. Das I.J., M., V.P., Kassaee A, Tabata T, Verhaegen F. Dose perturbations at high-z interfaces in kilovoltage photon beams: comparison with Monte Carlo simulations and measurements. Radiation Physics and Chemistry 64, p (00). 3. Moskvin, V., et al. Monte Carlo simulation of the Leksell Gamma Knife: I. Source modelling and calculations in homogeneous media. Phys Med Biol 47, (00). 33. Cheung, J.Y. & Yu, K.N. Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies. Med Phys 33, (006). 34. Novotny Jr J, B.J., Niranjan A, Quader MA, Saiful Huq M, Bednarz G, Flickinger J, Kondziolka D, Lunsford LD. Dosimetric comparison of the Leksell Gamma Knife Perfexion and 4C. Journal of Neurosurgery 109, p8-14 (008). 35. Lundquist C, P.I. The Leksell Gamma Knife Perfexion and comparisons with its predecessors. Operative Neurosurgery 61, p (007). 36. Wiant, D., et al. Gamma knife radiosurgery treatment planning for small animals using high-resolution 7T micro-magnetic resonance imaging. Radiat Res 17, (009). 37. Cheung, J.Y., Yu, K.N., Ho, R.T. & Yu, C.P. Monte Carlo calculated output factors of a Leksell Gamma Knife unit. Phys Med Biol 44, N47-49 (1999). 38. Leskell Gamma Plan 8. Online Reference Manual. (ELEKTA Instruments AB, Stockholm, Sweden, 007). 39. Elekta, A.B. A new TMR dose algorithm in Leksell GammaPlan. (ed. Elekta, A.B.) 8 (Stockholm, Sweden, 011). 40. Novotny, J., Jr., et al. Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry. Med Phys 36, (009).
117 Bednarz, G., Huq, M.S. & Rosenow, U. Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams. Phys Med Biol 47, (00). 4. Ekstrand, K.E. & Bourland, J.D. A film technique for the determination of output factors and end effect times for the Leksell Gamma Knife. Phys Med Biol 46, (001). 43. Kurjewicz, L. & Berndt, A. Measurement of Gamma Knife helmet factors using MOSFETs. Med Phys 34, (007). 44. Gasparian, P.B.R., et al. Demonstrating the use of optically stimulated luminescence dosimeters (OSLDs) for measurement of staff radiation exposure in interventional fluoroscopy and helmet output factors in radiosurgery. Radiation Measurements 45, Moutsatsos, A., et al. Gamma knife output factor measurements using VIP polymer gel dosimetry. Med Phys 36, (009). 46. Elekta. Elekta Bulletin , Leksell Gamma Knife "New 4-mm Helmet Output Factor". (1998). 47. Kjäll, P. & Johansson, J. Geometry information for Gamma Knife Perfexion. (ed. Best, R.) 11 (Elekta Instrument AB, Stockholm, 010). 48. Elekta, A.B. Safety Evaluation of Sealed Source - Medical Teletherapy Source. 10 (Elekta A.B., 007). 49. Salvat F, F.-V.J.M.a.S.J. PENELOPE - A Code System for Monte Carlo Simulation of Electron and Photon Transport, (NEA-OECD, Paris, 008). 50. Moskvin, V., et al. Monte Carlo simulation of the Leksell Gamma Knife: II. Effects of heterogeneous versus homogeneous media for stereotactic radiosurgery. Phys Med Biol 49, (004). 51. Bush, K., Zavgorodni, S.F. & Beckham, W.A. Azimuthal particle redistribution for the reduction of latent phase-space variance in Monte Carlo simulations Physics in Medicine and Biology 5(007). 5. Miller, T., David Chin. Information for Grant Writing. in WFU DEAC Support Wiki, Vol. 010 (Wake Forest University, Winston-Salem, NC, 010). 53. Niroomand-Rad, A., et al. Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 Medical Physics 5, 3 (1998). 54. Butson, M.J., Yu, P.K.N., Cheung, T. & Metcalfe, P. Radiochromic film for medical radiation dosimetry. Materials Science and Engineering: R: Reports 41, (003).
118 Devic, S., et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Medical Physics 3, 9 (005). 56. Abramoff, M.D., Magalhaes, P.J. & Ram, S.J. Image Processing with ImageJ. Biophotonics International 11, 36-4 (004). 57. Andreo, P., Izewska, J., Shortt, K. & Vatnitsky, S. Commissioning and Quality Assurance of Computerized Planning Systems for Radiation Treatment of Cancer - IAEA TRS-430. in Technical Report Series (International Atomic Energy Agency, Vienna, 1999). 58. Jiang, S.B., et al. On dose distribution comparison. Phys Med Biol 51, (006). 59. van Elmpt, W., et al. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiotherapy and Oncology 88, (008). 60. Flampouri, S., et al. Estimation of the delivered patient dose in lung IMRT treatment based on deformable registration of 4D-CT data and Monte Carlo simulations Phys. Med. Biol. 51(006). 61. Baldock, C., et al. Polymer gel dosimetry Phys. Med. Biol. 55(010). 6. Wendling, M., et al. A fast algorithm for gamma evaluation in 3D Medical Physics 34(007). 63. Park, D.H., et al. Optimized matching of film dosimetry with calculated doses for IMRT quality assurance. Physica Medica 3, (007). 64. Anjum, M.N., Parker, W., Ruo, R. & Afzal, M. Evaluation criteria for film based intensity modulated radiation therapy quality assurance. Physica Medica 6, (010). 65. Dyk, J.V., Barnett, R.B., Cygler, J.E. & Shragge, P.C. Commissioning and quality assurance of treatment planning computers. International Journal of Radiation Oncology*Biology*Physics 6, (1993). 66. Low, D.A., Harms, W.B., Mutic, S. & Purdy, J.A. A technique for the quantitative evaluation of dose distributions Medical Physics 5, 6 (1998). 67. Gloi, A., Buchana, R., Zuge, C. & Goettler, A. RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys. 1(011). 68. Ezzell, G.A., et al. IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 Medical Physics 36, 15 (009).
119 69. Best, R.C.M. Monte Carlo Modeling of the Gamma Knife Perfexion. (Wake Forest University, Winston-Salem, NC, 010). 109
120 110 APPENDIX I Single Sector Dose Profiles with Error Range Single sector dose profile curves incorporating error propagated through the simulation are presented here. Compare to Figures (A) (B) (C) (D) (E) (F) (G) (H) (I) Figure 53: Dose profile curves generated by the MC algorithm in blue with error bars compared to dose profile curves generated by LGP s TMR10 algorithm in green. The first row (A, B, C) are the 16 mm collimator s x, y, and z axis profiles, respectively. The second row (D, E, F) are the 8 mm collimator s x, y, and z axis profiles, respectively. The third row (G, H, I) are the 4 mm collimator s x, y, and z axis profiles, respectively.
121 111 APPENDIX II Derivation of Compton Energy Equations Derivation of Equation 1, maximum energy transferred in a Compton scattering event. Equation 1 reproduced as Equation A1 below. T max hvi = (A1) + m c / hv e i A Compton scattering event is diagrammed as shown in Figure 54 Figure 54: Diagram of Compton scattering event. The incident photon from the left interacts with a free electron, scattering both. Equation A1 can be derived from the conservation of momentum (Equation A) and the conservation of energy (Equation 8), where p i indicates initial momentum of the photon, and p f indicates final photon momentum p = p + p (A) i Rearranging and squaring both sides results in Equations A3-4, and moves the terms from vector to scalar forms. p e p i f e ( p p ) e i f = (A3) f ( θ ) = p + p p p cos (A4) Photon momentum is its energy divided by its speed, p=e/c. The energy of a photon is its frequency times Planck s constant, or E=hν. Replacing momentum in A4 with these new terms results in A5, then A6 once c is moved to the left side. i f
122 11 ( ) ( ) ( ) θ ν ν ν ν cos c h c h c h p f i f i e + = (A5) ( ) ( ) ( ) θ ν ν ν ν cos f i f i e h h h c p + = (A6) Shifting gears to conservation of energy, the energy conservation of Figure 54 before and after the event can be written as Equation A7. 4 c m c p h m c h e e i e i + + = + ν ν (A7) Equations A8-13 show rearrangement of terms with the goal of isolating p e c. 4 m c c p h m c h e e f e i + + = + ν ν (A8) ( ) ( ) 4 c m c p h m c h e e f e i + = + ν ν (A9) ( ) ( ) ( ) ( ) ( ) 4 4 m c c p h m c h h h m c h mc h h m c h e e i e f f i e f e i f e i + = ν ν ν ν ν ν ν ν (A10) ( ) ( ) ( ) ( ) ( ) c p h m c h h h m c h m c h h h e i e f f i e f e i f i = ν ν ν ν ν ν ν ν (A11) ( ) ( ) c p h m c h m c h h h e i f e f e i f i = + + ν ν ν ν ν ν (A1) ( ) ( ) ( ) c p h h h m c h h e i f f i e f i = + + ν ν ν ν ν ν (A13) The right side of Equation A13 is equal to the left side of Equation A6. Setting these two equal produces Equation A14. ( ) ( ) ( ) ( ) ( ) ( ) θ νν ν ν ν ν ν ν ν ν cos f i f i i f f i e f i h h h h h h m c h h + = + + (A14) Canceling out various terms and simplifying: ( ) ( ) θ νν ν ν ν ν cos f i i f f i e h h h h c m = (A15) ( ) ( ) θ ν ν ν ν ν ν cos f i i f f i e h h h h c m = (A16) Cross-multiply to move terms to opposite side denominators: ( ) cos 1 m c h h h e i f f i θ ν ν ν ν = (A17) ( ) cos 1 m c h h h h e i f f i f i θ ν ν ν ν ν ν = (A18)
123 113 1 hν f ( θ) 1 1 cos = + (A19) hν m c i e hν hν 1 1 cos ( θ) 1 hν i f = + hν i mec (A0) hν i 1 hν ( 1 cos( θ) ) 1 hν i f = + i mc (A1) e hν i hν = f hν i( 1 cos( θ) ) (A) 1+ m c Equation A is the one most commonly seen for calculating the Compton scattered photon s final energy hν f after scattering through angle θ. Conservation of energy requires that the sum of the final kinetic energies of the photon (hν f ) and electron (T) equal the initial kinetic energy of the photon (hν i ), per equation A3. Substituting A3 into A is a step toward Equation A1. e hν = hν T (A3) f i hν i hν i T = (A4) hν i( 1 cos( θ) ) 1+ m c e 1 T = h ν ( ( )) i 1 hν i 1 cosθ (A5) 1+ mec m ec T = hν i 1 (A6) mec + hν i ( 1 cos( θ) ) ( 1 cos( θ) ) m ec ( 1 cos( θ) ) m + ν ( 1 cos( θ) ) ec h i mec + hν i T = hν i (A7) mec + hν i ( 1 cos( θ) ) hν ( 1 cos( θ) ) i hν i T = hν i (A8) mec +
124 114 1 cos( θ) T = hν i + 1 cos( θ) (A9) mec hν i The maximum energy is delivered to the electron when the photon scatters backward at 180 degrees. Setting θ=180 in Equation A9 produces equation A30. 1 ( 1) Tθ = 180 = hν i + 1 ( 1) (A30) mec hν i hν i Tθ = 180 = (A31) mec + hν Equation A31 is equivalent to Equation A1, proving the equation for the maximum energy a photon can deliver to an electron in a Compton scattering event. i
125 115 VITA Ryan Best is from Ashland, Virginia. He did his undergraduate work at Guilford College in Greensboro, NC, and in 003 received Bachelor of Science degrees in Physics with honors and Computing & Information Technology. While at Guilford College he received the Jeglinski Physics Award and membership in Sigma Pi Sigma, the physics honors society. After college he worked for Gamma Corporation, a Medical/Health Physics consulting company in Honolulu, Hawaii. During that time he took graduate courses in Medical Physics from Georgia Tech in Atlanta, Georgia through a distancelearning program. In 010 he earned a Master s Degree in Physics from Wake Forest University in Winston-Salem, NC where he was supported by a clinical assistantship from the Department of Radiation Oncology at Wake Forest University Baptist Medical Center. His previous research is published in the Journal of Applied Clinical Medical Physics 1, (JACMP) and the Journal of Computer Assisted Tomography 3 (JCAT), and was presented at the conferences of the American Association of Physicists in Medicine (AAPM) in , and American Society for Radiation Oncology (ASTRO) in 010. He has accepted a Clinical Medical Physics Residency position in Radiation Oncology at the University of Virginia in Charlottesville, VA. Vita References 1. Gersh, J.A. et al. Improved volumetric imaging in tomosynthesis using combined multiaxial sweeps. Journal of Applied Clinical Medical Physics 11(010).. Bennett, M.C., et al. Mechanisms and prevention of thermal injury from gamma radiosurgery headframes during MR imaging. Journal of Applied Clinical Medical Physics (01). 3. Bharkhada, D. et al. Demonstration of Dose and Scatter Reductions for Interior Computed Tomography. J. of Computer Assisted Tomography 33, (009).
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Introduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
GAMMA-RAY SPECTRA REFERENCES
GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.
Appendix A. An Overview of Monte Carlo N-Particle Software
Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation
Challenges in small field MV photon dosimetry
Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland [email protected] Why are
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
ACCELERATORS AND MEDICAL PHYSICS 2
ACCELERATORS AND MEDICAL PHYSICS 2 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 2-28.10.10 - U. Amaldi 1 The icone of radiation therapy Radiation beam in matter EPFL 2-28.10.10 - U.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
Radiation Detection and Measurement
Radiation Detection and Measurement June 2008 Tom Lewellen [email protected] Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron
Advanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
Measurement of Charge-to-Mass (e/m) Ratio for the Electron
Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
Variance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range The radiation from orthovoltage units is referred to as x-rays, generated by bombarding
Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:
Gamma Rays OBJECT: To understand the various interactions of gamma rays with matter. To calibrate a gamma ray scintillation spectrometer, using gamma rays of known energy, and use it to measure the energy
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
Geometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The
Chapter NP-5. Nuclear Physics. Nuclear Reactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 NUCLEAR REACTIONS 2.0 NEUTRON INTERACTIONS
Chapter NP-5 Nuclear Physics Nuclear Reactions TABLE OF CONTENTS INTRODUCTION OBJECTIVES 1.0 2.0 NEUTRON INTERACTIONS 2.1 ELASTIC SCATTERING 2.2 INELASTIC SCATTERING 2.3 RADIATIVE CAPTURE 2.4 PARTICLE
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
Fiber Optics: Fiber Basics
Photonics Technical Note # 21 Fiber Optics Fiber Optics: Fiber Basics Optical fibers are circular dielectric wave-guides that can transport optical energy and information. They have a central core surrounded
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] [email protected] [e-mail] This
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
GRID AND PRISM SPECTROMETERS
FYSA230/2 GRID AND PRISM SPECTROMETERS 1. Introduction Electromagnetic radiation (e.g. visible light) experiences reflection, refraction, interference and diffraction phenomena when entering and passing
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
Section 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved.
Section 5. : Horn Physics Section 5. : Horn Physics By Martin J. King, 6/29/8 Copyright 28 by Martin J. King. All Rights Reserved. Before discussing the design of a horn loaded loudspeaker system, it is
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
WAVELENGTH OF LIGHT - DIFFRACTION GRATING
PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant
5-Axis Test-Piece Influence of Machining Position
5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,
1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy
Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
THE BOHR QUANTUM MODEL
THE BOHR QUANTUM MODEL INTRODUCTION When light from a low-pressure gas is subject to an electric discharge, a discrete line spectrum is emitted. When light from such a low-pressure gas is examined with
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
E/M Experiment: Electrons in a Magnetic Field.
E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
Proton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical
Information about the T9 beam line and experimental facilities
Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions
Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues
Journal of mathematics and computer science 9 (2014), 41-45 Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues S. A. Mahdipour 1, H. R. Abdi Roknabadi
DO PHYSICS ONLINE FROM QUANTA TO QUARKS QUANTUM (WAVE) MECHANICS
DO PHYSICS ONLINE FROM QUANTA TO QUARKS QUANTUM (WAVE) MECHANICS Quantum Mechanics or wave mechanics is the best mathematical theory used today to describe and predict the behaviour of particles and waves.
ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner
PUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
O6: The Diffraction Grating Spectrometer
2B30: PRACTICAL ASTROPHYSICS FORMAL REPORT: O6: The Diffraction Grating Spectrometer Adam Hill Lab partner: G. Evans Tutor: Dr. Peter Storey 1 Abstract The calibration of a diffraction grating spectrometer
Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry
Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry John Keightley NSUF 18 May 2010 Acknowledgement: Several discussions with Stefaan Pommé, IRMM, EC-JRC, Geel Some diagrams
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
AP2 Magnetism. (c) Explain why the magnetic field does no work on the particle as it moves in its circular path.
A charged particle is projected from point P with velocity v at a right angle to a uniform magnetic field directed out of the plane of the page as shown. The particle moves along a circle of radius R.
MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics
Microsoft Excel 2010 Charts and Graphs
Microsoft Excel 2010 Charts and Graphs Email: [email protected] Web Page: http://training.health.ufl.edu Microsoft Excel 2010: Charts and Graphs 2.0 hours Topics include data groupings; creating
Intensity-Modulated Radiation Therapy (IMRT)
Scan for mobile link. Intensity-Modulated Radiation Therapy (IMRT) Intensity-modulated radiotherapy (IMRT) uses linear accelerators to safely and painlessly deliver precise radiation doses to a tumor while
ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.
1 ATOMIC SPECTRA Objective: To measure the wavelengths of visible light emitted by atomic hydrogen and verify the measured wavelengths against those predicted by quantum theory. To identify an unknown
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
TIME OF COMPLETION NAME SOLUTION DEPARTMENT OF NATURAL SCIENCES. PHYS 3650, Exam 2 Section 1 Version 1 October 31, 2005 Total Weight: 100 points
TIME OF COMPLETION NAME SOLUTION DEPARTMENT OF NATURAL SCIENCES PHYS 3650, Exam 2 Section 1 Version 1 October 31, 2005 Total Weight: 100 points 1. Check your examination for completeness prior to starting.
Use of 3D Printers in Proton Therapy
Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology [email protected] 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing
1. Units of a magnetic field might be: A. C m/s B. C s/m C. C/kg D. kg/c s E. N/C m ans: D
Chapter 28: MAGNETIC FIELDS 1 Units of a magnetic field might be: A C m/s B C s/m C C/kg D kg/c s E N/C m 2 In the formula F = q v B: A F must be perpendicular to v but not necessarily to B B F must be
CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas
How To Understand Light And Color
PRACTICE EXAM IV P202 SPRING 2004 1. In two separate double slit experiments, an interference pattern is observed on a screen. In the first experiment, violet light (λ = 754 nm) is used and a second-order
Optical Design Tools for Backlight Displays
Optical Design Tools for Backlight Displays Introduction Backlights are used for compact, portable, electronic devices with flat panel Liquid Crystal Displays (LCDs) that require illumination from behind.
Transmission Line and Back Loaded Horn Physics
Introduction By Martin J. King, 3/29/3 Copyright 23 by Martin J. King. All Rights Reserved. In order to differentiate between a transmission line and a back loaded horn, it is really important to understand
Nuclear Physics. Nuclear Physics comprises the study of:
Nuclear Physics Nuclear Physics comprises the study of: The general properties of nuclei The particles contained in the nucleus The interaction between these particles Radioactivity and nuclear reactions
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
How To Use A Proton For Radiation Therapy
Proton Cancer Therapy: Using Laser Accelerate Protons for Radiation Therapy A Thesis Submitted in Partial Fulfillment of the Requirements for Graduation with Research Distinction in Engineering Physics
Copyright 2011 Casa Software Ltd. www.casaxps.com. Centre of Mass
Centre of Mass A central theme in mathematical modelling is that of reducing complex problems to simpler, and hopefully, equivalent problems for which mathematical analysis is possible. The concept of
Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.
T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM
Force measurement Forces VECTORIAL ISSUES In classical mechanics, a force is defined as "an action capable of modifying the quantity of movement of a material point". Therefore, a force has the attributes
Laue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
Interference. Physics 102 Workshop #3. General Instructions
Interference Physics 102 Workshop #3 Name: Lab Partner(s): Instructor: Time of Workshop: General Instructions Workshop exercises are to be carried out in groups of three. One report per group is due by
The Phenomenon of Photoelectric Emission:
The Photoelectric Effect. The Wave particle duality of light Light, like any other E.M.R (electromagnetic radiation) has got a dual nature. That is there are experiments that prove that it is made up of
THE DØ Luminosity Monitor (LM) consists of two array of
FERMILAB-CONF-6-469-E Radiation Damage to Scintillator in the DØ Luminosity Monitor Brendan Casey, Kayle DeVaughan, Yuji Enari, Richard Partridge, and Sahal Yacoob Abstract We report the result of evaluating
3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
Physics Lab Report Guidelines
Physics Lab Report Guidelines Summary The following is an outline of the requirements for a physics lab report. A. Experimental Description 1. Provide a statement of the physical theory or principle observed
