Practical Course in Reference Dosimetry. Electron Beam Dosimetry and Codes of Practice

Similar documents
Characterisation of two new ionisation chamber types for use in reference electron dosimetry in the UK

High-Energy Electron Beam Therapy Dosimetry with Ionisation Chambers

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

AAPM protocol for kv x-ray beam dosimetry in radiotherapy and radiobiology

Challenges in small field MV photon dosimetry

Chapter 9 CALIBRATION OF PHOTON AND ELECTRON BEAMS

Worldwide Quality Assurance networks for radiotherapy dosimetry

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-energy X-Rays (APMP.RI(I)-K3)

Clinical Photon Beams

QA of intensity-modulated beams using dynamic MLC log files

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy

ACCELERATORS AND MEDICAL PHYSICS 2

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

Chapter XIII. PDD TAR TMR Dose Free Space

CHAPTER 9. CALIBRATION OF PHOTON AND ELECTRON BEAMS

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Scripps Proton Therapy Center: Configuration and Implementation

Use of 3D Printers in Proton Therapy

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Introduction to the Monte Carlo method

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2

Variance reduction techniques used in BEAMnrc

Secondary Neutrons in Proton and Ion Therapy

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

Manual for simulation of EB processing. Software ModeRTL

Therapy with protons and ion beams

Development and Experimental Performance Evaluation of a Dose-Rate meter for Pulsed Beam.

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi

External dosimetry Dosimetry in new radiotherapeutic techniques

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS

HADRON THERAPY FOR CANCER TREATMENT

Testing thermo-acoustic sound generation in water with proton and laser beams

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

Principles of dosimetry The ionization chamber

Quality Assurance for Treatment Planning Systems

Quality Assurance of Radiotherapy Equipment

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Solar Cell Parameters and Equivalent Circuit

Linear accelerator output, CT-data implementation, dose deposition in the presence of a 1.5 T magnetic field.

Quality Assurance of accelerators; the technologists responsibility

Monte Carlo Simulations in Proton Dosimetry with Geant4

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)

A dosimetry technique for measuring kilovoltage cone-beam CT dose on a linear accelerator using radiotherapy equipment

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass

Calorimetry in particle physics experiments

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

The photoionization detector (PID) utilizes ultraviolet

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

Quality Assurance in Stereotactic. Radiotherapy

Diagnostic Exposure Tracking in the Medical Record

The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.

Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations

PARTICLE THERAPY QA TOOLS. A complete range of tools for absolute dosimetry, machine QA and patient plan verification

The feasibility of a QA program for ISIORT Trials

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

arxiv:astro-ph/ v1 15 Sep 2005

TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS

The accurate calibration of all detectors is crucial for the subsequent data

Qualitätsmanagement-Handbuch

Proton tracking for medical imaging and dosimetry

Research on the Air Conditioning Water Heater System

The Calibration Chain: Role of BIPM,

DENSITY MEASURING SYSTEMS

Fibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Chapter 3: Radiation Dosimeters

METHODS OF BOLUSING THE TRACHEOSTOMY STOMA

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

Assessing Radiation Dose: How to Do It Right

DETERMINING THE ENTHALPY OF FORMATION OF CaCO 3

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: kv range

Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.

DISCRETE SEMICONDUCTORS DATA SHEET. General Temperature sensors Dec 05. File under Discrete Semiconductors, SC17

World-first Proton Pencil Beam Scanning System with FDA Clearance

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

Single Mode Fiber Lasers

Medical Applications of radiation physics. Riccardo Faccini Universita di Roma La Sapienza

Digital vs. Analogue Control Systems

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Hello and Welcome to this presentation on LED Basics. In this presentation we will look at a few topics in semiconductor lighting such as light

Transcription:

Practical Course in Reference Dosimetry Electron Beam Dosimetry and Codes of Practice

Electron beam dosimetry and Codes of Practice o cthe IPEM code of practice for electron beam dosimetry for radiotherapy beams of initial energy 4 to 25 MeV based on an absorbed dose to water calibration. Phys. Med. Biol. (2003) Vol 48 (No 18) 2929-2970

IPEM 2003 Code of Practice Overview Based on the NPL absorbed dose calibration service, which gives direct factors for electron chambers in terms of absorbed dose to water Very simple procedure to follow in the clinic, together with a significantly reduced uncertainty compared to IPEMB 1996 Code based on air kerma Electron dosimetry now on the same footing as MV photons

Outline of Code Section 1 - Introduction Section 2 Code of Practice Appendices: A Corrections B Chambers C Formalism D Data E Uncertainties Everything you need to measure absorbed dose to water in electron beams is in this document.

IPEM 2003 Theory (1) (1) define reference depth in water for beams at NPL z ref,w = 0.6 R 50,w 0.1 cm (2) use range scaling to get depth in graphite z ref,g = z ref,w R 50,g / R 50,w (3) calibrate chamber against the calorimeter at NPL, in graphite, at z ref,g N D, ref, g M D g ref, g

IPEM 2003 Theory (2) (4) theoretical conversion from graphite to water N N D, ref, w D, ref, g p p ref, w ref, g s s w/ air g/ air (5) compare user and reference chambers at NPL, at z ref,w in water N N D, user, w D, ref, w M M ref, w user, w The formalism is described in detail in Appendix C.

IPEM 2003 Code of Practice Section 1 Validation Trial calibrations carried out in 1996, 1997 and 1999 17 UK radiotherapy centres, supplied 46 chambers NACP, Markus, Farmer and Roos types, electrometers also supplied Calibrated at 7 energies in the range 3 to 17 MeV Comparison in clinic with IPEMB 1996 CoP The change in stated dose was no more than 2% for individual chambers of Farmer, NACP and Roos designs and generally less than 1% Based on this work, the Markus design is not recommended as a designated chamber for absolute dosimetry (see section 2.1.2 and appendix B).

Continued Validation 1.05 Dose (NPL) / Dose (IPEMB) 1.04 1.03 1.02 1.01 1.00 0.99 0.98 1 2 3 4 5 6 7 R 50,D (cm)

IPEM 2003 Code of Practice Section 2 2. Code of practice 2934 2.1. Instrumentation 2934 2.2. Measurements in an electron beam 2936 2.2.1. General points 2936 2.2.2. Beam quality specification 2937 2.2.3. Calibration reference depth 2938 2.2.4. Determination of central-axis depth dose curves 2938 2.2.5. Determination of absorbed dose at the reference point 2941 2.2.6. Transfer of dose from calibration reference point to the depth of dose maximum 2942 2.2.7. Determination of relative output factors 2942 2.2.8. Use of non-water phantoms 2943 2.2.9. Use of other chambers as field instruments 2946 2.2.10. Extension to beam qualities outside the range available at the

Section 2 in more detail 2.1 2.1.1 Equipment required Chamber & electrometer Temperature & Pressure sensors Scanning tank 2.1.2. Designated chambers Farmer: NE2571, PTW30004, PTW30012 Parallel plate: Scanditronix NACP-02, PTW Roos 34001 Note: Manufacturers specified not copies! 2.1.3. Effective point of measurement

Section 2 in more detail 2.2.1 Reference material Calibration in a water phantom recommended Solid phantoms allowed, especially for low energy beams (see 2.2.8) Order of preference: 1. Water 2. Epoxy resin materials 3. PMMA, polystyrene N.B. Uncertainties increase

Section 2 in more detail 2.2.2 Beam quality specification Electron energy as a specifier can be confusing what does it refer to? Mean energy emerging from linac? Mean energy at phantom surface? Choose R 50,D as the specifier: Defines reference depth Calibration factors on the NPL certificate given as a function of R 50,D R 100 depth in water, z/cm R 85 R 50 R p

% depth-dose Section 2 in more detail 2.2.3 Calibration reference depth z ref = 0.6 R 50,D - 0.1 (cm) 100 80 What s wrong with the peak? 60 40 20 0 0 2 4 6 8 10 12 NPL beam clinical beam z (cm)

Relative dose R 50 and D ref Depth-dose curves for NPL Clinical Linac 120 100 80 60 40 4 MeV 6 MeV 8 MeV 10 MeV 12 MeV 15 MeV 18 MeV 20 MeV 22 MeV 20 0 0 20 40 60 80 100 120 Depth,mm

R50 and Dref

Section 2 in more detail 2.2.4 Ion chamber or diode Need to convert ionisation to dose: a) Equation 2.1 R 50,I to R 50,D (Ding et al, 1995) R 50,D = 1.029R 50,I 0.063 cm Based on evaluations of measurements and matched Monte Carlo calculations b) Equation D.2 (the Burns equation ) Based on evaluations by Burns et al 1996,of the work by Ding et al, 1995 c) Table 1 (p2940) Based on the Burns equation Measurement of depth dose curves d) Could also use stopping power ratios as calculated directly using Monte Carlo codes such as SPRRZ

Section 2 in more detail 2.2.4

Section 2 in more detail 2.2.5 Determination of absorbed dose at the reference point For user chamber with NPL calibration factor N D,user,w (given as a function of R 50,D ) : (1) Measure R 50,D for nominal 1 m SSD (2) Position chamber at z ref = 0.6 R 50,D - 0.1 (cm) (3) Interpolate certificate data to derive N D,w,user (R 50,D ) (4) At z ref : D w = M w N D,w,user

Typical NPL certificate

Section 2 in more detail 2.2.8 Non-water phantoms Solid phantoms offer significant advantages: Robust Dry! Mechanically stable Easy to set up but we need absorbed dose to water

Section 2 in more detail 2.2.8 Non-water phantoms h m = ratio of fluences in the water and non-water phantoms at the same water equivalent depth D w (z w,eff ) = M ch,non-w h m N D,w Also need range scaling factor C pl to get the depth in solid where the electron spectrum is (approximately) the same: z w,eff = z non-w C pl Data given in CoP for h m and C pl (Tables 2 & 3, p2945)

Section 2 in more detail 2.2.8 Non-water phantoms

Appendix A Chamber corrections D ( z )= M( z ) N ( w ref ref R50 D,w ) M( z )= M f f f f ref raw TP pol ion elec

Recombination correction: It is important that a chamber is operated in the linear region: 1.016 1.014 1.012 1/I (normalised) 1.010 1.008 1.006 1.004 1.002 1.000 0.998 0.000 0.005 0.010 0.015 0.020 0.025 0.030 0.035 1/V (V -1 )

Appendix D Data D.2. Stopping power ratios D.4. Non-water phantoms D.5. Determination of energy parameters of electron beams

Appendix E: Uncertainties Uncertainties in the calibration of an ionisation chamber in terms of absorbed dose to water Source of of uncertainty Uncertainties in the determination of absorbed dose to water Standard Uncertainty ( %) Note that the uncertainties Type A in Type the A certificate are slightly different (and larger) than those in tables E.1 and E.2 of the CoP: Type B Graphite calibration factor for NPL chambers 0.25 0.25 Ratio of chamber measurements 0.1 Ratio of stopping powers 0.5 Ratio of perturbation factors 0.7 Recombination correction 0.17 Polarity correction 0.1 Electrometer calibration 0.05 Quadratic sum 0.34 0.90 Combined uncertainty 0.96 Standard Uncertainty ( %) Type B Water calibration factor for user chambers 0.34 0.90 Matching of NPL and clinical beam qualities 0.3 Practical set and chamber measurement per MU [a] 0.50 Influence factors 0.25 Measuring chamber stability 0.25 Electrometer calibration 0.05 Quadratic sum 0.7 0.95 Combined uncertainty 1.2

Summary It s all there! It is prescriptive but requires some interpretation You will need to read the entire CoP to implement it

Electron practical session The practical session will focus on the IPEM 2003 Code of Practice through a combination of paper exercises and measurements. Paper exercises: Measurements: Determination of range parameters, N D,w factors and other data Output factors, recombination and polarity

The End

Polarity corrections Farmer chambers E (MeV) PTW NE2505/3 NE2571 10 1.0008 0.9881 0.9961 12 0.9926 0.9982 16 1.0013 0.9951 0.9990 19 1.0011 0.9948 0.9981 Beware - all graphite-walled Farmer chambers are not the same! Significant differences of up to 0.5% Difference between NE2505/3 and NE2571 is due to different guarding arrangements