Meeting the US Environmental Protection Agency Drinking Water Requirements with a UV/Persulfate Analyzer

Similar documents
Phoenix Apollo Laboratory TOC Analyzers and Services Designed To Meet Your Needs

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

Ecology Quality Assurance Glossary

The Sievers 900 Series TOC Analyzers

1 Quality Assurance and Quality Control Project Plan

How To Choose A Total Organic Carbon Analyzer

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

Where does it come from?

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

TECHNICAL MEMORANDUM TO: FROM: AUTHOR: DATE: SUBJECT: Introduction Experimental Design

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

METHOD Revision 1.0. May 2000

DR/4000 PROCEDURE. CHLORINE, Free

LZV585 (B) The 1-cm path is aligned with the arrow on the adapter

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

LACHAT METHOD NUMBER D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately g CN/L g CN/L. 09e

TRL-onlineTOC Monitor is designed for monitoring carbon content in

Cleaning Validation by TOC Analyzer

Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program

A Visual Method for the Detection of Arsenic 0 to 500 µg/l By Dan Kroll R&D Chemist. Hach Co., Loveland Colorado

Determination of Calcium and Magnesium in Brine

ph Alkalinity of Water

Microbiological Testing of the Sawyer Mini Filter. 16 December Summary

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE

METHOD DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

ETV Joint Verification Statement

NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility

Dissolved and precipitated oxalate

Spectrophotometer - Milton Roy Spectronic 21D or equivalent.

Measurement by Ion Selective Electrodes

MATHESON STACK EMISSION CALIBRATION PROGRAM. Experience the MATHESON Commitment to Supply Chain Excellence

Online Infrared Flue Gas Analyzer Gasboard 3000

Quality Assurance/Quality Control in Acid Deposition Monitoring

Work Assignment Prepared for: Office of Air Quality Planning and Standards U.S. Environmental Protection Agency Research Triangle Park, NC

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

ESS Method 130.1: General Auto Analyzer Procedures

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Nitrogen, Phosphorus, and Suspended Solids Concentrations in Tributaries to the Great Bay Estuary Watershed in 2011

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

High Purity Water Resistivity/ Conductivity Measurement

The First Step in Effluent Treatment

Drinking Water Analytical Method and Program Requirements: Roles and Responsibilities, Analytical Method Approval, and Effective Oversight

ORGANIC SAMPLE PREPARATION

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

FP628. Nitrogen / Protein Determination by Combustion. Rev 1-6/8/11

Samples RMN Analytical Report ( ) GC/FID Analysis (RM01) Nicotine Purity

Method OIA : Available Cyanide by Ligand Exchange and Flow Injection Analysis (FIA)

QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis

ASSURING THE QUALITY OF TEST RESULTS

Phosphorus, Reactive (Orthophosphate)

S olid Ph ase Extraction (SP E) Method:

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

How To Test For Contamination In Large Volume Water

Determination of Caffeine in Beverages HPLC-1

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

C391-E058K TOC-V. Ser ies. Shimadzu Total Organic Carbon Analyzer

Generation of Vapor-Phase Hydrogen Peroxide Test Atmospheres for the Evaluation of Respirator Cartridges and Air Samplers

Laboratory Procedure Manual

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Total Cyanide by Segmented Flow Injection Analysis, On-Line UV Digestion, and Amperometric Detection

Vitamin C quantification using reversed-phase ion-pairing HPLC

INTERNATIONAL OLIVE COUNCIL

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

CERTIFICATE OF ANALYSIS

Quality Assurance Manual

Online Total Organic Carbon (TOC)

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

Canadian Environmental Technology Verification Program Applicant Information Package CONTENTS

HS 1003 Part 2 HS 1003 Heavy Metals Test

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

Mercury Determination in House Dust, NIST SRM-2584, USEPA Method 7473, Using the Teledyne Leeman Labs Hydra II C Combustion CVAAS

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

Application Requirement

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Introduction. The following definitions may help you better understand the components of the data report.

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

An Introduction to Standards and Quality Control for the Laboratory Barbara Martin

Elemental Scientific. Stability. Elemental Scientific

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

Analysis of Total Organic Carbon

MIDLAND AREA SOIL SAMPLING MIDLAND, MICHIGAN

Transcription:

Abstract The rapid and precise measurement of organic carbon in trace levels of water is of interest to the pharmaceutical, drinking water and environmental industries where regulations restrict the amount of discharge or contamination. Total Organic Carbon (TOC) analysis is considered an effective indicator of organic contamination in water. Since the US EPA set limits on Disinfection By-Products (DBPs) in 1979 and subsequently required testing, TOC analysis has slowly taken the forefront of accurate and rapid testing for DBPs water contamination. Municipalities and other drinking water facilities follow various recognized national and international methodologies to ensure compliance with the US EPA regulations including US EPA 415.3, 9060A, and SM5310. This study will focus on using a UV/ Persulfate analyzer to meet US EPA drinking water regulations using SM5310. QC Requirements A typical laboratory quality control program for TOC analysis may include the following: Laboratory Reagent Blank (LRB) Calibration Standards (STDs) Independent Quality Control Samples (QCS) Laboratory Fortified Sample Matrix (LFM) Laboratory Duplicates (LD) End Calibration Check (ECC) Calibration Standards analyzed as Samples

QC Type Batch Requirement Acceptance Criteria Laboratory Reagent Blank(LRB) Calibration Standards (STDs) Second Source Quality Control Sample (QCS) Laboratory Fortified Sample Matrix (LFM) Laboratory Duplicate (LD) End Calibration Check (ECC) Calibration Standards Analyzed as Samples At the beginning of each batch and every 10th sample 4 point calibration standards under 10mg/L C Every 10 th sample < 0.350mg/L C Fit first order regression (R 2 0.99) Within acceptance range set by the manufacturer 1 per batch 80 120 % recovery 1 per batch Table 1: Laboratory QC Requirements At the end of each run. Concentration at calibration curve mid-range. At the end of each run. Relative % difference (RPD) <20% 90-110% recovery 90-110% recovery Calibration A calibration curve was generated on the Fusion TOC, using standard points of 0.30, 2.0, 4.0 and 10mg/L carbon (C), by manual dilution from a 1000mg/L C potassium hydrogen phthalate source standard. Each calibration point was analyzed in triplicate measurements. Reagent water was used as the zero point on the curve and the linearity was excellent with an R 2 value of 0.99995. (Figure 1)

Y NDIR Absorbance (ABS) 300 250 200 150 100 50 0 Fusion TOC Calibration y = 27.27x + 12.13 R² = 0.99995 0 2 4 6 8 10 12 X expected (mg/l C) Figure 1: Multi-point Calibration Curve of the Fusion TOC Method Parameters TOC analyzers utilize a number of techniques to oxidize organic carbon in the sample to CO 2, the final form of the carbon in the sample that is detected and quantified. The Fusion TOC utilizes UV/ Persulfate oxidation of carbon. UV/ Persulfate oxidation is highly efficient and reliable for ground water samples and waste water samples without large amount of particulates. To detect and quantify CO 2 an NDIR detector is used. The TOC Fusion uses SPC technology in conjunction with NDIR detection. SPC technology is a process by which a single measurement of the CO 2 inside a pressurized NDIR detector is taken. This is achieved by first oxidizing the carbon by UV-Persulfate. During the oxidation phase, the detector outlet is sealed allowing the CO 2 to be transferred inside the detector to a predetermined pressure set-point. Once the pressure setting has been achieved and all the CO 2 is pressurized within the detector, a single CO 2 measurement is taken. The amount of CO 2 detected correlates to the amount of carbon in the sample 1.

Fusion Method Parameters Parameter Value Sample Volume 4.0mL Dilution 1:1 Acid Volume 1.0mL Reagent Volume 1.3mL UV Reactor Prerinse On UV Reactor Prerinse Volume 6.0 Number Of UV Reactor Prerinses 3 Needle Rinse Volume 5.0mL Vial Prime Volume 2.0mL IC Sample Prime Volume 2.0mL IC Sparge Rinse Volume 10.0mL Baseline Stabilize Time 0.70 min Detector Pressure Flow 200mL/min Syringe Speed Waste 10 Syringe Speed Acid 7 Syringe Speed Reagent 7 Syringe Speed DI Water 7 NDIR Pressurization 45 psig Syringe Speed Sample Dispense 7 Syringe Speed Sample Aspirate 4 Syringe Speed UV Dispense 7 Syringe Speed UV Aspirate 5 Syringe Speed IC Dispense 7 Syringe Speed IC Aspirate 5 NDIR Pressure Stabilize 0.50 min Sample Mixing Off Sample Mixing Cycles 1 Sample Mixing Volume 10.0 Low Level Filter NDIR Off Table 3: The method parameters for analysis of ground water samples. The use of a 4.0mL 6.0mL sample volume is ideal for ground water sample analysis. It provides ample sample carbon mass to be deposited onto the detector and sample volume for multiple replicates.

The use of a long Pre-Sparge Time and low Detector Pressure Flow allows for the analysis of samples without particulates to those with large amounts of very fine particulates. Sample ID Fusion TOC Results (mg/l C) Check Standard - (1.80 mg/l C certified value) 1.9350 Blank 0.0942 Basin #7 Treated 07/07/10 2.5861 Basin #7 Treated 07/07/10 Spiked @ 4mg/L C (LFM1) 6.6756 Basin #7 Treated 08/04/10 1.9647 Basin #4 Treated 07/07/10 2.1431 Basin #4 Treated 07/07/10 Spiked @ 4 mg/l C (LFM2) 6.5952 Basin #1 Treated 07/07/10 2.2215 Basin #3 Treated 08/10/10 1.8317 Basin #4 Untreated 07/07/10 3.0078 Basin #5 Treated 08/10/10 4.8528 Basin #6 Treated 08/10/10 3.5537 Blank 0.1066 Check Standard - (4.32mg/L C certified value) 4.3788 Basin #4 Treated 07/07/10 2.0717 Basin #1 Untreated 07/07/10 3.3079 Basin #1 Untreated 08/04/10 3.2957 Basin #1 Treated 08/04/10 2.3404 Basin #2 Treated 8/10/10 3.3351 Basin #4 Treated 07/07/10 Spiked @ 4mg/L C (LFM2 Duplicate) 6.8022

Basin #4 Untreated 08/10/10 2.0552 Check Standard (4.32mg/L C certified value) 4.5501 ECC4 3.9443 4 ppm Calibration Standard 4.2410 10 ppm Calibration Standard 9.9385 Table 4: The sample analyses of ground water samples Method Tests Requirement Fusion TOC Performance Fusion Exceeds Specification Laboratory Reagent Blank (LRB) Calibration Standards (STD) < 0.350mg/L C 0.0942mg/L C 0.1066mg/L C Fit first order regression 0.99995 (R 2 0.99) Second Source QCS Laboratory Fortified Sample Matrix (LFM) 1.5205-2.0808mg/L C 3.9312-4.8816mg/L C 1.9350mg/L C 4.3788mg/L C 80 120 % recovery 99 % recovery Laboratory Duplicates (LD) RPD < 20% 3 % End Calibration Check (ECC) Calibration Standards run as Samples 90-110% recovery 99 % 90-110% recovery 106 % 99 % Table 5: Fusion TOC Results for Laboratory QC Requirements

Conclusions The Fusion TOC utilizes UV/ Persulfate oxidation and NDIR detection with enhanced SPC technology that directly and specifically measures CO 2 gas generated from the organic carbon in the sample. This method offers the only interference-free detection of TOC. Since the sample is in a gaseous form, no direct contact is made between the sample gas and the detector adding years of life to the detector. Automatic calibration and a self-calibrating detector make the Fusion a great choice in analyzing TOC, per SM5310. The Fusion can effectively help your facility implement USEPA s 415.3 new regulations by providing: Outstanding Precision and Accuracy +/- 1% RSD Limit of detection: 0.2 ppb Easy, Fast and Linear Auto-Calibration and Auto-Calibration Verifications Default Method Parameters Can be used to meet most laboratory analysis needs without modification NDIR Detection coupled with SPC Technology Provides reliable, rugged and interference free CO 2 detection for TOC analysis References 1. TOC Fusion Manual 2. Standard Methods for the Examination of Water and Wastewater, 19 th Edition 1995 3. Potter, B. and Wimsatt, J. USEPA Method 415.3 Determination of TOC and SUVA at 254nm in Source Water and Drinking Water. National Exposure Research Laboratory, Office of Research and Development, USEPA, Cincinnati, OH, June 2003.

Abbreviations TOC Total Organic Carbon SM5310 Standard Methods 5310 US EPA United States Environmental Protection Agency NDIR Non-Dispersive Infrared ABS Absorbance SPC Static Pressure Concentration DBPs Disinfection By-Products UV Ultra Violet IC Inorganic Carbon min Minutes ml Milliliters HTC High Temperature Combustion CO 2 Carbon Dioxide mg/l parts per million C Carbon QCS Quality Control Sample ECC End Calibration Check LD Laboratory Duplicates LFM Laboratory Fortified Sample Matrix STDs Calibration Standards LRB Laboratory Reagent Blank