TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS
|
|
|
- Edwina Hood
- 9 years ago
- Views:
Transcription
1 TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified by measuring the accuracy, precision, linearity, limit of detection, and limit of quantification. The Sievers* M9 Analyzers used in this study included one laboratory, two portable, and five online Sievers M9 e units. 2. BACKGROUND INFORMATION AND CALCULATIONS 2. Accuracy The accuracy performance specification results are presented here as percent recoveries [,2]. The averages of replicates of sucrose standards were determined at the following concentrations: 25 ppb, 5 ppb, ppm, 5 ppm, ppm, 25 ppm, and 5 ppm. The percent recovery for each standard was calculated using the following equation: % Recovery = X % Percent recoveries close to % indicate a high degree of accuracy for the Analyzer. 2.2 Precision The precision performance specification data are presented as a % relative standard deviation (%RSD). The standard deviation of replicates and %RSD are calculated as follows: Standard Deviation: Where: x = Each result n = Number of measurements Relative Standard Deviation (%RSD) = X % A %RSD value close to % indicates a high degree of precision for the analyzer [-3].
2 2.3 System Suitability System suitability is a test commonly performed in the pharmaceutical industry to confirm the instrument s ability to recover both an easy and a difficult to oxidize compound [4,5]. A response efficiency of % indicates that the instrument recovered both compounds equally. Response Efficiency is calculated using the following equation: Response Efficiency = % Where Rss is the measured concentration of 5 ppb,4-benzoquinone (BQ), Rs is the measured concentration of 5 ppb sucrose, and Rw is the concentration of the reagent water blank. For USP/EP System Suitability, the acceptance criterion is a Response Efficiency between 85% and 5% [5]. 2.4 Limit of Detection/Limit of Quantification The limit of detection (LOD) is defined as the minimum concentration of TOC that can be reliably detected by the analyzer [,2]. It is calculated by the following: LOD = 3s Where: s = standard deviation of replicates of online DI water The limit of quantification (LOQ) is defined as the lowest concentration of TOC that can be determined quantitatively with an acceptable uncertainty [,2]. It is calculated via the equation: LOQ = s 2.5 Linearity Linearity is defined as the ability of the TOC analyzer to generate data that are directly proportional to the concentration of multiple TOC standards across the dynamic range of the instrument. Linearity is measured by calculating a correlation coefficient (R 2 ), slope of a regression line, and y-intercept. R 2 values close to. represent a high degree of correlation. 3. EXPERIMENTAL TEST PLAN 3. Sucrose and BQ Standards This study was divided into two separate parts in order to quantify performance specifications using various calibrations on three different instruments. One laboratory Sievers M9 unit and two portable Sievers M9 units were used in this study. Using the different calibrations, a wide range of sucrose standards were run to encompass the full 2
3 operating range of the analyzers and quantify differences (if any) between calibrations. Standard solutions of sucrose were produced at the following concentrations: 25 ppb, 5 ppb, ppb, 2 ppb, 25 ppb, 5 ppb, ppm, 5 ppm, ppm, 25 ppm, and 5 ppm. For accuracy and precision performance specifications, the standards in the concentration range of 25 ppb to 5 ppm were used. For linearity, all standards were used. To determine system suitability and response efficiency, 5 ppb BQ standards were analyzed. These BQ standards and the associated reagent water blanks were part of the System Suitability Standards Set made at GE Analytical Instruments. The table below outlines the order in which the samples were run on each instrument, the acid and oxidizer flow rates, and the number of repetitions. TOC concentration Testing for Acid (µl/min) Oxidizer (µl/min) Repetitions Reagent Water All.5 / 5 ppb Acc/Prec.5 / Rw for BQ System Suitability.5 / 5 ppb BQ System Suitability.5 / 25 ppb Linearity.5 / 5 ppb Linearity.5 / ppb Linearity.5 / 2 ppb Linearity.5 / 25 ppb Linearity/Acc/Prec.5 / 5 ppb Linearity/Acc/Prec.5 / ppm Linearity/Acc/Prec.. / 5 ppm Linearity/Acc/Prec..2 / ppm Linearity/Acc/Prec..6 / 25 ppm Linearity/Acc/Prec. 3.4 / 5 ppm Linearity/Acc/Prec. 6.8 / 3.2 All Analyzers with ppm The three analyzers were each calibrated using a ppm single-point calibration using KHP and Na 2 CO 3 as outlined in GE protocols [6]. Once the analyzers were calibrated at ppm and verified at 5 ppb, all of the sucrose and BQ standards were analyzed as described above. 3.3 Analyzers with Three Different s After the completion of all samples using the ppm calibration, the three analyzers were recalibrated as follows: Lab 39: Multi-Point (25 ppb, ppm, 5 ppm, ppm, 25 ppm, 5 ppm TOC as KHP, and ppm IC as Na 2 CO 3 ) Portable 42: 5 ppm Single-Point Portable 43: ppm Single-Point New aliquots of the same sucrose and BQ standards were then run on each instrument with these new calibrations. 3
4 4. TEST EQUIPMENT USED Sievers M9 Laboratory TOC Analyzer Serial Number: Sievers M9 Portable TOC Analyzer Serial Number: 442 Sievers M9 Portable TOC Analyzer Serial Number: 443 GE Autosampler Serial Number: 424 GE Autosampler Serial Number: 323 GE Autosampler Serial Number: 945 DataPro2 Software 5. CALIBRATION AND VERIFICATION Please refer to sections 3.2 and RESULTS AND DISCUSSION All results presented here are for replicates of each sucrose standard (25 ppb-5 ppm) with no rejects. There was one statistical outlier on the 25 ppm sucrose standard on Lab 39 using the ppm calibration, which was determined using a Grubb s statistical test for outliers [-3] at the 99% confidence level. This data point was discarded. 6. Precision Lab 39 ppm Cal 25 ppb 2.2 ppb.9% ppm Cal 5 ppb 2.7 ppb.5% ppm Cal ppm 6.7 ppb.6 % ppm Cal 5 ppm 9.3 ppb.4 % ppm Cal ppm 78.9 ppb.8 % ppm Cal 25 ppm 22 ppb.5 % ppm Cal 5 ppm 362 ppb.7 % Multi-Point Cal 25 ppb.9 ppb.8% Multi-Point Cal 5 ppb 3.3 ppb.7 % Multi-Point Cal ppm 4.8 ppb.5 % Multi-Point Cal 5 ppm 5.2 ppb.3% Multi-Point Cal ppm 7 ppb.7% Multi-Point Cal 25 ppm 5 ppb.6 % Multi-Point Cal 5 ppm 36 ppb.7 % Portable 42 ppm Cal 25 ppb 2. ppb.8 % ppm Cal 5 ppb 3.9 ppb.8 % ppm Cal ppm 9.5 ppb.9 % 4
5 ppm Cal 5 ppm 8.4 ppb.4 % ppm Cal ppm 48.3 ppb.5 % ppm Cal 25 ppm 9.9 ppb.4 % ppm Cal 5 ppm 399 ppb.8 % 5 ppm Cal 25 ppb 4.7 ppb.% 5 ppm Cal 5 ppb 4. ppb.7% 5 ppm Cal ppm 7.4 ppb.7 % 5 ppm Cal 5 ppm 9. ppb.4 % 5 ppm Cal ppm 42.2 ppb.4 % 5 ppm Cal 25 ppm 67.5 ppb.3 % 5 ppm Cal 5 ppm 323 ppb.6 % Portable 43 ppm Cal 25 ppb 2. ppb.8 % ppm Cal 5 ppb 2.3 ppb.4% ppm Cal ppm 6.7 ppb.6 % ppm Cal 5 ppm 3.3 ppb.6 % ppm Cal ppm 63.2 ppb.6 % ppm Cal 25 ppm 7 ppb.3 % ppm Cal 5 ppm 27 ppb.5 % ppm Cal 25 ppb.8 ppb.7% ppm Cal 5 ppb 2.3 ppb.4% ppm Cal ppm 8.4 ppb.8% ppm Cal 5 ppm 2.2 ppb.4% ppm Cal ppm 85. ppb.8% ppm Cal 25 ppm 78.9 ppb.3% ppm Cal 5 ppm 272 ppb.6% 6..2 Precision Statistics Using a two-tailed Student s t-test, it was assessed whether or not instrument precision (as %RSD) was statistically different depending on the calibration or depending on the individual unit [,2]. At the 99% confidence level, there is no statistical difference (p>.) between the precision of the instruments regardless of calibration nor was there a statistically significant difference between the precision of each individual analyzer. The following table outlines the p-value results of the Student s t-tests for the precision of each analyzer. Instrument Average %RSD ppm Cal Standard Deviation Average %RSD Alternate Cal Lab 39.62%.7%.6% (Multi-Point cal) Portable 42.64%.23%.6% (5 ppm cal) Standard Deviation p-value (Student s t-test).6%.85.25%.62 5
6 Portable 43.57%.5%.58% ( ppm cal) All Units.6%.8%.58% (all alternate cals).2%.95.2% Precision Performance Specification Based on the data reported here, the precision performance specification for the Sievers M-Series TOC Analyzers is %. 6.2 Accuracy 6.2. Percent Recoveries TOC Concentration %Recovery Lab 39 ppm % Recovery Lab 39 Multi-Point Accuracy Performance Specification % Recovery Portable 42 ppm % Recovery Portable 42 5 ppm % Recovery Portable 43 ppm % Recovery Portable 43 ppm 25 ppb 98.6% 98.%.% 4.7% 99.% 99.6% 5 ppb 99.3% 97.8% 99.5% 4.2% 99.5% 99.4% ppm.9% 99.5% 99.6% 4.6%.4%.8% 5 ppm.6% 99.9% 99.% 3.7%.5%.4% ppm 2.% 99.2% 99.2% 3.%.2% 99.2% 25 ppm.7% 98.8% 96.5%.6% 98.5% 98.5% 5 ppm 99.5% 97.% 95.4%.% 96.9% 96.4% The results presented here demonstrate a high degree of accuracy of the Sievers M9 Analyzers regardless of the calibration. Several recommendations on the optimal calibration for particular sample concentrations include the following: ) The ppm calibration produced recoveries of 98.6% to.9% on all three instruments in the 25 ppb to 5 ppm range. On two of the instruments, using the ppm calibration produced recoveries of 99.2% and.2% at the ppm concentration indicating that using this calibration is still sufficient even at concentrations times higher than the single calibration point. 2) The Multi-Point provided excellent results (98.8% to 99.5%) in the ppm to 25 ppm TOC concentration range. Customers regularly running samples in this range can use this type of calibration if desired, though a single-point ppm calibration produces equivalent recoveries. Below ppm and above 25 ppm, the recoveries were lower by 2.% to 2.9% using the multi-point calibration. 3) The 5 ppm single-point calibration provided the best results only at 5 ppm. A slightly lower calibration slope was likely produced using this type of calibration than the other calibration types leading to higher percent recoveries on all of the lower TOC concentrations (.6% to 4.7%). 4) The ppm calibration had excellent recoveries (98.5% to.8%) in the range of 25 ppb to 25 ppm. Recovery was slightly lower (96.4%) at 5 ppm, but that TOC concentration was five times higher than the single calibration point, so the results are not unexpected. On Portable 43 the unit that had both the ppm and the ppm calibrations there was no 6
7 statistical difference between the percent recoveries using either calibration in the 25 ppb to 5 ppm range (p>.). 5) As a general guideline, measurements on the Sievers M9 Analyzers should be either at or below the calibration point. Based on the recoveries discussed above, the accuracy performance specification of the Sievers M-Series TOC Analyzers are ±2% when using the appropriate calibration based on the TOC range of the samples. 6.3 Linearity To demonstrate linearity, least-squares regression lines for a wide range of TOC concentrations were plotted (25 ppb to 5 ppm). Results of these regressions as well as the correlation coefficients (R 2 ) are reported in this section of the results. These results are separated into two subsections with one demonstrating linearity of the full instrument range (up to 5 ppm) and the other showing linearity of the instrument across lower TOC concentrations (25 ppb ppm). Each of these subsections includes linearity plots for the different calibrations on each instrument Linearity Across the Full Instrument Range (up to 5 ppm) Lab 39 ppm y =.9967x R² = Lab 39 Multi-Point y =.974x R² = Portable 42 ppm y =.9555x R² = Portable 42 5 ppm y =.28x +.72 R² =
8 Portable 43 ppm y =.977x R² = Portable 43 ppm y =.9668x R² = Linearity Across Low TOC Concentrations (up to ppm) Lab 39 ppm y =.2x -.57 R² = Lab 39 Multi-Point y =.995x -.29 R² = Portable 42 ppm y =.9957x +. R² = Portable 43 ppm y =.67x -.4 R² = Portable 42 5 ppm y =.45x +. R² = Portable 43 ppm y =.94x -.35 R² =
9 6.3.3 Linearity Performance Specification All three instruments demonstrated highly linear responses (R ) across the full range of the instrument as well as at low TOC levels with the four different calibration options. 6.4 System Suitability Instrument Response Efficiency Lab 39 ppm.9% Lab 39 Multi-Point 99.7% Portable 42 ppm.2% Portable 42 5 ppm 99.6% Portable 43 ppm.6% Portable 43 ppm 99.3 % System suitability results from this experiment (99.3% to.9% response efficiency) demonstrate the excellent response efficiency of the Sievers M9 Analyzers for all calibration types. These response efficiencies are well within the acceptable USP/EP System Suitability range of 85% to 5% [5]. 6.5 LOD/LOQ To determine the LOD and LOQ of the Sievers M-Series Analyzers, replicates of DI water on five Sievers M9 e On-Line units were used. These units had serial numbers: 5, 5, 52, 54, and 59. In order to eliminate variability in TOC measurements due to water system changes, the Sievers M9 e units were run for 24 hours. The replicates for the LOD and LOQ calculations were selected during the most stable hour-long time span of the water system. The table below outlines the standard deviation for these reps on each instrument, the average standard deviation, as well as the LOD and LOQ. SIEVERS M9e Serial Number Standard deviation ( reps) Average.84 LOD.3 LOQ.8 Based on the standard deviations of ten replicates for each of these units, the LOD and LOQ for the Sievers M-Series Analyzers are: LOD =.3 ppb and LOQ =.8 ppb 9
10 7. CONCLUSIONS In summary, the performance specifications for the Sievers M-Series TOC Analyzers have been experimentally verified as follows: Accuracy ±2% (25 ppb to 5 ppm) Precision Linearity LOD LOQ % (25 ppb to 5 ppm) R (25 ppb to 5 ppm).3 ppb.8 ppb 8. REFERENCES. Barwick, V., S. Burke, R. Lawn, P. Roper, and R. Walker. (2). Applications of Reference Materials in Analytical Chemistry. Cambridge: LGC Limited. 2. Ellison, S.R., V.J. Barwick, and T.J. Duguid Farrant. (29). Practical Statistics for the Analytical Scientist. Cambridge: RSC Publishing. 3. Taylor, J.K. (987). Quality Assurance of Chemical Measurements. Florida: CRC Press LLC. 4. ICH Harmonized Tripartite Guideline Validation of Analytical Procedures: Text and Methodology Q2(R). (25). 5. USP <643> Total Organic Carbon. 6. Sievers M9/M9 e TOC Analyzers Operation and Maintenance Manual..4Rev Firmware. DLM 77- EN Rev D. * Trademark of General Electric Company; may be registered in one or more countries. The Americas GE Analytical Instruments 66 Spine Road Boulder, CO USA T T F [email protected] Europe/Middle East/Africa GE Analytical Instruments Unit 3, Mercury Way Urmston, Manchester UK M4 7LY T +44 () F +44 () [email protected] Asia Pacific GE Analytical Instruments 7/F, Building, No. Hua Tuo Rd, ZhangJiang Hi-Tech Park,, Pudong Shanghai, China 223 T + (862) F + (862) [email protected] 25, General Electric Company. All rights reserved. MC Rev. B
The Sievers 900 Series TOC Analyzers
GE Water & Process Technologies Analytical Instruments The Sievers 900 Series TOC Analyzers imagination at work Delivering Next-Generation Ease of Use, Productivity, and Reliability The Sievers* 900 Series
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
To avoid potential inspection
Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
A practical and novel standard addition strategy to screen. pharmacodynamic components in traditional Chinese medicine using
Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2015 A practical and novel standard addition strategy to screen pharmacodynamic components in traditional
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods
Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. [email protected] Overview GMP requirements for Quality Control laboratories
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE
Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND
Design of Experiments for Analytical Method Development and Validation
Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Validating Methods using Waters Empower TM 2 Method. Validation. Manager
Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Analytical Methods for Cleaning Validation
Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2011, 3 (6): 232-239 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4
Online Total Organic Carbon (TOC)
Reprinted from PHARMACEUTICAL ENGINEERING The Official Magazine of ISPE January/February 2008, Vol. 29 No. 1 This article offers an alternative method to cleaning validation using online total organic
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)
Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:
Implementing New USP Chapters for Analytical Method Validation
Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: [email protected] Today s Agenda Handling Method Changes vs. Adjustments
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:
How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
QbD Approach to Assay Development and Method Validation
QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Assay Development and Method Validation Essentials
Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
Procedure for Writing Technical Procedures
Procedure for Writing Technical Procedures 1.0 Purpose This procedure specifies when a technical procedure is needed and outlines the required elements for writing the technical procedure. 2.0 Scope This
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
Quality Assurance/Quality Control in Acid Deposition Monitoring
Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
Magruder Statistics & Data Analysis
Magruder Statistics & Data Analysis Caution: There will be Equations! Based Closely On: Program Model The International Harmonized Protocol for the Proficiency Testing of Analytical Laboratories, 2006
American Association for Laboratory Accreditation
Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number
1.0 XRF Analyses Performance Evaluation
1.1 Scope and Application 1.0 XRF Analyses Performance Evaluation Due to the time and monetary costs associated with conventional analytical methods for measuring the concentrations of inorganic elements
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
Signal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
FT-NIR for Online Analysis in Polyol Production
Application Note: 51594 FT-NIR for Online Analysis in Polyol Production Key Words Acid Number Ethylene Oxide FT-NIR Hydroxyl Value Polyester Polyols Abstract Hydroxyl value and other related parameters
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Using R for Linear Regression
Using R for Linear Regression In the following handout words and symbols in bold are R functions and words and symbols in italics are entries supplied by the user; underlined words and symbols are optional
Beräkningar enligt EN 14211, EN 14212, EN 14625 och EN 14626.
Beräkningar enligt EN 14211, EN 14212, EN 14625 och EN 14626. 8.4.5 Repeatability standard deviation After waiting the time equivalent of one independent reading 20 individual measurements both at zero
Qualification Study CHO 360-HCP ELISA (Type A to D)
Short Report Qualification Study CHO 360-HCP ELISA (Type A to D) Presented by: http://www.biogenes.de Version 01 Issue date: 27.11.2013 Version 01 Page 1 of 8 Table of Content TABLE OF CONTENT... 2 1 INTRODUCTION...
Sample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
Calibration Verification
Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS
DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS Bernie B. Bernard, Heather Bernard, and James M. Brooks TDI-Brooks International/B&B Laboratories Inc. College Station,
Measurement Systems Correlation MSC for Suppliers
Measurement Systems Correlation MSC for Suppliers Copyright 2003-2007 Raytheon Company. All rights reserved. R6σ is a Raytheon trademark registered in the United States and Europe. Raytheon Six Sigma is
Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2
ENERGY RESEARCH at the University of Oulu 117 Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2 1 University of Oulu, Department
Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10
Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan SECTION 10 DATA QUALITY ASSURANCE AND QUALITY CONTROL (QA/QC) Hawai`i HEER TGM 10-1 December 31, 2008 CONTENTS Acronyms
Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument
Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Technical Note INTRODUCTION Direct measurements of nucleic acid samples at OD 260 or protein samples at OD
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
Evaluation of Quantitative Data (errors/statistical analysis/propagation of error)
Evaluation of Quantitative Data (errors/statistical analysis/propagation of error) 1. INTRODUCTION Laboratory work in chemistry can be divided into the general categories of qualitative studies and quantitative
INTERNATIONAL OLIVE COUNCIL
INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: [email protected]
Pearson's Correlation Tests
Chapter 800 Pearson's Correlation Tests Introduction The correlation coefficient, ρ (rho), is a popular statistic for describing the strength of the relationship between two variables. The correlation
ANALYTICAL DETECTION LIMIT GUIDANCE. & Laboratory Guide for Determining Method Detection Limits
ANALYTICAL DETECTION LIMIT GUIDANCE & Laboratory Guide for Determining Method Detection Limits Wisconsin Department of Natural Resources Laboratory Certification Program April 1996 PUBL-TS-056-96 introduction
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
Fairfield Public Schools
Mathematics Fairfield Public Schools AP Statistics AP Statistics BOE Approved 04/08/2014 1 AP STATISTICS Critical Areas of Focus AP Statistics is a rigorous course that offers advanced students an opportunity
A Review of Statistical Outlier Methods
Page 1 of 5 A Review of Statistical Outlier Methods Nov 2, 2006 By: Steven Walfish Pharmaceutical Technology Statistical outlier detection has become a popular topic as a result of the US Food and Drug
FP628. Nitrogen / Protein Determination by Combustion. Rev 1-6/8/11
FP628 Nitrogen / Protein Determination by Combustion Rev 1-6/8/11 Presentation Outline LECO Background N Determination FP628 Theory of Operation Specifications Advantages Software Data LECO Background
Mass Spectrometry Signal Calibration for Protein Quantitation
Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of
THE LABORATORY NOTEBOOK
THE LABORATORY NOTEBOOK In scientific work keeping a permanent record of all raw data, observations, calculations, et cetera obtained during an experiment is important. Therefore, a student must become
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014 This experiment explores quality assurance practices which are commonly
SOLUBILITY, IONIC STRENGTH AND ACTIVITY COEFFICIENTS
SOLUBILITY, IONIC STRENGTH AND ACTIVITY COEFFICIENTS References: 1. See `References to Experiments' for text references.. W. C. Wise and C. W. Davies, J. Chem. Soc., 73 (1938), "The Conductivity of Calcium
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
General and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation
IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
A Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet
TargetQuan 3 Software Leading the way in regulatory POPs quantification Bullet Bullet Bullet Leading the way in regulatory POPs quantification Analyse Samples Open Sequence Process Sequence Analyse Samples
Spectrophotometer - Milton Roy Spectronic 21D or equivalent.
3M COMPANY OCCUPATIONAL HEALTH AND ENVIRONMENTAL SAFETY DIVISION DETERMINATION OF FORMALDEHYDE VAPORS IN AIR USING 3M 3721 FORMALDEHYDE MONITORS May, 2002 SCOPE This procedure covers the method of collecting
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
RARITAN VALLEY COMMUNITY COLLEGE ACADEMIC COURSE OUTLINE MATH 111H STATISTICS II HONORS
RARITAN VALLEY COMMUNITY COLLEGE ACADEMIC COURSE OUTLINE MATH 111H STATISTICS II HONORS I. Basic Course Information A. Course Number and Title: MATH 111H Statistics II Honors B. New or Modified Course:
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Quantifying measurement error from digital instruments
Quantifying measurement error from digital instruments W. BLAKE LAING AND SEAN BRYANT SOUTHERN ADVENTIST UNIVERSITY CHAT TANOOGA, TN What I m doing HELPING STUDENTS LEARN TO CONSTRUCT KNOWLEDGE First lab:
Methods of Dealing with Values Below the Limit of Detection using SAS
Methods of Dealing with Values Below the Limit of Detection using SAS Carry W. Croghan, US-EPA, Research Triangle Park, NC, and Peter P. Egeghy US-EPA, Las Vegas, NV ABSTRACT Due to limitations of chemical
3.4 TEST FOR BACTERIAL ENDOTOXINS. Final text for revision of The International Pharmacopoeia
Document QAS/11.5 FINAL July 01 3. TEST FOR BACTERIAL ENDOTOXINS Final text for revision of The International Pharmacopoeia This monograph was adopted at the Forty-sixth WHO Expert Committee on Specifications
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
The Fitness for Purpose of Analytical Methods
The Fitness for Purpose of Analytical Methods EURACHEM Guide The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method Validation and Related Topics This document has been developed by
Determination of the Sensitivity Range of Biuret Test for Undergraduate Biochemistry Experiments
77 Determination of the Sensitivity Range of Biuret Test for Undergraduate Biochemistry Experiments Gerardo Janairo * ; Marianne Linley Sy; Leonisa Yap; Nancy Llanos-Lazaro; Julita Robles Chemistry Department
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting Method Number: ELAP method code 7037 SOP Number: Revision Number: SOP Date: Personnel / Data
Experiment #1, Analyze Data using Excel, Calculator and Graphs.
Physics 182 - Fall 2014 - Experiment #1 1 Experiment #1, Analyze Data using Excel, Calculator and Graphs. 1 Purpose (5 Points, Including Title. Points apply to your lab report.) Before we start measuring
