Troubleshooting and Diagnostics Tips and Tricks

Similar documents
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Successful HPLC Operation. A Troubleshooting Guide Version 1.1

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

The Theory of HPLC. Gradient HPLC

[ Care and Use Manual ]

Protocol: HPLC (amino acids)

INSTRUCTIONS Edition AC

RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Dionex IonPac AS19-4µm

[ Care and Use Manual ]

Oasis HLB Cartridges and 96-Well Plates

Electrospray Ion Trap Mass Spectrometry. Introduction

[ Care and Use Manual ]

HPLC Basics, Equipment, and Troubleshooting

Capillary HPLC Introduction

Controlling Contamination in UltraPerformance LC /MS and HPLC/MS Systems , Rev. D

Chapter 28: High-Performance Liquid Chromatography (HPLC)

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

Purification of reaction mixtures using flash chromatography.

Determination of Calcium and Magnesium in Brine

Chromatograph Maintenance and Troubleshooting Class # 5050

Care and use guide. Contents. a handy solution made easy. page

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Optimizing the Post-implementation Monitoring of a LC-MS/MS Method

ChromQuest 5.0 Chromatography Data System

ÄKTA system Training Guide

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

23 The Thermal Conductivity Detector

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Analytical Test Report

Waters 2695 Separations Module

High Sensitivity MS Determination of Carryover in a New Autosampler Design

Tamsulosin Hydrochloride Capsules

The basic division explosives for forensic purposes is shown in Fig. 1

Amino Acid Analyzer L-8900

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Thermal Conductivity Detector

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0)

[ Care and Use Manual ]

Determination of Caffeine in Beverages HPLC-1

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

An In-Gel Digestion Protocol

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Gas Chromatography Liner Selection Guide

Smartline HPLC by design

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna

Analysis of Vegetable Oils by High Performance Liquid Chromatography

HPLC Troubleshooting Guide

AMD Analysis & Technology AG

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

How To Test For Contamination In Large Volume Water

Appendix 5 Overview of requirements in English

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Paper Chromatography: Separation and Identification of Five Metal Cations

PAPER CHROMATOGRAPHY

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

A Novel Approach to Low Volume Sample Preparation

Automation of Sample Preparation in Ion Chromatography

Supplementary Materials and Methods (Metabolomics analysis)

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

ÄKTA Laboratory-scale Chromatography Systems

ACQUITY UPLC H-Class System Guide

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

GC Troubleshooting 10. How to use this Troubleshooting Help system GC Troubleshooting - Top Level Generic problem descriptions...

AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

A Practical Guide to Ion Chromatography AN INTRODUCTION AND TROUBLESHOOTING MANUAL. SeQuant. Innovators in Chemical Analysis

Micromass LCT User s Guide

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Trouble Shooting. Pump

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

ACQUITY UPLC H-Class System Guide

Troubleshooting Common HPLC Problems.

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Transcription:

Troubleshooting and Diagnostics Tips and Tricks Cerdanyola del Vallès, Enero 2014 2013 Waters Corporation 1

Troubleshooting and diagnostics Troubleshooting System Pressure Problems Incorrect Retention time Loss of precision Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 2

Troubleshooting and diagnostics Troubleshooting System Pressure Problems Incorrect Retention time Loss of precision Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 3

System Pressure problems To identify a pressure change from normal operation, create a pressure reference point System Pressure is affected by the column, mobile phase, flow rate, temperature and can vary greatly with different methods 2013 Waters Corporation 4

System Pressure problems Erratic flow rates/pressure pulsations Overpressure No or low pressure 2013 Waters Corporation 5

System Pressure problems Erratic flow rates/pressure pulsations Air in system o Prime the pump (methanol or IPA to remove air) Air in solvent lines. Not enough solvent in bottle o Replace the solvent bottle Air in solvent lines. Bottle filters dirty o Remove the bottle filters Air in solvent lines. Not enough degas o Degas the mobile phase Problem with check valves o Sonicate or replace the check valves Problem with seals or plungers 2013 Waters Corporation 6

System pressure problems /Correct pressure Same distance between pump strokes Same height of pump stroke 2013 Waters Corporation 7

System Pressure problems Overpressure Check if pressure has risen gradually or suddenly. o If pressure has risen gradually particulates are accumulated in inline filter, columns frits or column. o If pressure has risen suddenly something could be a obstruction in system or column. 2013 Waters Corporation 8

System Pressure problems Overpressure Check if something has changed (column, mobile phase, temperature) If nothing has changed, remove the column and replace it with a union to check if the system pressure is the usual. If system pressure is high loosen fittings beginning with the last connection in line and working backward to the pump o After loosening each fitting observes if pressure stays the same or reduces. o Replace or clean the appropiate part Caution: Do not loosen fittings under high pressure 2013 Waters Corporation 9

System Pressure problems Low pressure Check if something has changed (column, mobile phase, temperature, method) If nothing has changed, check for leaks. No pressure Air in system o Prime the pump (methanol or IPA to remove air) Air in solvent lines. Not enough solvent in bottle o Replace the solvent bottle Problem with check valves o Sonicate or replace the check valves Problem with seals or plungers 2013 Waters Corporation 10

Troubleshooting and diagnostics Troubleshooting System Pressure Problems Retention time Loss of precision Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 11

Retention Time Retention time changed to a new constant value Erratic retention time Increasing retention time Decreasing retention time 2013 Waters Corporation 12

Retention Time Retention time changed to a new constant value Check column, mobile phase, temperature, method, flow rate Increasing/decreasing retention times Column contaminated, degraded System not equilibrated Mobile phase contaminated Erratic retention times Check if system also has erratic pump pressure/pressure fluctuations System not equilibrated Check for leaks Improper solvent blending Temperature fluctuations 2013 Waters Corporation 13

Troubleshooting and diagnostics Troubleshooting System Pressure Problems Retention time Loss of precision Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 14

Loss of precision Incorrect peak integration Check loss of precision is for all peaks in the cromathogram. If it is only for some of them does not seem a injector problem Check if reproducibility lack is for areas or also for Retention times. Check for leaks Check injection volume and sample concentration Don t overload the column 2013 Waters Corporation 15

Loss of precision Check injector wash solvents. Are appropriate for the method? Check injection volume is inside system specifications Are area values erratics or are increasing or decreasing? Injector problem. Pass injector test It is important to have a system suitability to check the system. In case you don`t have one, Waters supplies specific solutions for each system to check the system. This solution is called ASR or QCRM. 2013 Waters Corporation 16

Troubleshooting and diagnostics Troubleshooting System Pressure Problems Incorrect Retention time Loss of precision Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 17

Definitions Contamination the presence of any unwanted substance in a chromatographic system that appears either as peaks or high background noise. Carryover is a specific type of contamination 0.00 UV Baseline without Sample Injection A U -0.02 0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 Minutes 2013 Waters Corporation 18

Background and Contamination Do these two terms mean the same thing? If they are different, how are they different? Is background ever zero? 2013 Waters Corporation 19

Background and Contamination Do these two terms mean the same thing? Not exactly If they are different, how are they different? Background is the baseline noise and will always be present. Contamination is background at an unacceptably high level. Is background ever zero? No. The sample and liquids used in HPLC and LC/MS are chemicals which the detectors may detect. 2013 Waters Corporation 20

Definition of Background What is background in chromatography for UV and for mass spectrometry? How does it appear in a chromatogram? When does it appear? 2013 Waters Corporation 21

Definition of Background What is background in chromatography for UV and for mass spectrometry? Changes in baseline How does it appear in a chromatogram? Deviation from zero response When does it appear? Always present. The mobile phases are chemicals and contain additives or contaminants. 2013 Waters Corporation 22

Define Carryover What is carryover? Analyte remaining from a previous injection that appears as peaks in subsequent injections and compromises quantification. How do you measure carryover? Sample or standard followed by one or more blanks Does every equipment have a carryover specification? Yes. It should be define at the equipment user guide. Compound and detector specific. Is the injector always the problem when there is carryover? No 2013 Waters Corporation 23

Example of Carryover 0.04 Blank Injection before Sample AU 0.02 0.00 0.04 High Concentration Sample AU 0.02 0.00 0.04 Blank Injection after Sample AU 0.02 0.00 0.00 0.75 1.50 2.25 3.00 Minutes 2013 Waters Corporation 24

How to Measure Carryover? Blank Injection(s) Sample diluent/matrix without the analyte Follow the injection of the sample with a blank injection o Carryover is expressed as the response observed in the blank divided by the response of the sample injection and expressed as a percentage Follow the injection of the sample by multiple blank injections o Does the response in the blanks diminish with each injection or does it remain constant? If the response does not decrease with multiple blank injections, the problem is likely contamination 2013 Waters Corporation 25

Critical Factors Affecting Carryover Injection Type / Injection Technique / Wash Solvent Selection Different injection modes Choosing proper wash solvents Analytical Method Column carryover? Precipitation in the injector? Hardware Issues Is something broken? Are there materials issues? 2013 Waters Corporation 26

Contamination It is easier to prevent contamination than to remove it. Sources of information Source of contamination Prevention Diagnostic scheme Cleanup 2013 Waters Corporation 27

Potential Sources of Contamination Greater sensitivity means greater chance of seeing contamination Sources Solvents and additives (water is a big problem) Sample matrix Sample preparation chemicals (detergents, salts) Dirty glassware (Do not send solvent bottles to dish washer.) Plastic containers or tubing Detergents HPLC systems, tubing Hand creams Manufacturing process WARNING Contaminates will adsorb and concentrate on C18 columns 2013 Waters Corporation 28

Questions to Ask Should you assume the solvents are free of contaminants? No Should you assume the equipment is clean? No You should not assume anything First determine if the solvents or their bottles contain any contamination. Then determine if the equipment is contributing contamination. 2013 Waters Corporation 29

Contamination from the Sample Inject a pure standard Inject a sample with matrix. If there are other peaks they came from the matrix. Inject a blank of the sample diluent alone. Is it different from a water blank? Inject volume zero 2013 Waters Corporation 30

Contamination Diagnostic: Injector-Pump Zero Volume Injection Have the injector go through the injection sequence without injecting any volume o Vial contamination No Injection or Disconnect the autosampler and connect pumps to detector Run gradient without injection ( Inject Immediate samples in Empower) If the contamination peak is present without an injection, it is not from the injector o System contamination o Solvent contamination o Method Carryover (column) Change pump solvents to check the pump If profile changes, contaminations could be related to the pump. 2013 Waters Corporation 31

Solvent Contamination With a C18 column in the system, a blank gradient is run and there are peaks in the chromatogram, there is contamination from somewhere. If it is in the water, the longer the reequilibration at high aqueous, the larger the peaks. Steps to eliminate Find a better supply of solvents and or clean bottles Strip column at 100% organic until baseline is low and stable. Run blank gradient again Organic solvent contamination. Change solvent bottles 2013 Waters Corporation 32

Solvent Quality What is good quality solvent? What is good quality water? 2013 Waters Corporation 33

Solvent Quality What is good quality solvent? Low background for the detector to be used What is good for UV may not work for MS Particle-free o Filtered by supplier to 0.2µm filter before bottling o Do not filter again because filters will add contamination. What is good quality water? Is on-demand water purification system (e.g. Milli-Q) good? Is bottled water better? Never stored for long periods of time (things grow in it) 2013 Waters Corporation 34

Mobile Phase Quality Highest quality of solvents available Highest quality of additives Buffer salts Acids Bases Use additives at the lowest concentration that gives good results WARNING Failure to use the highest quality solvents leads to higher backgrounds, contamination, loss of sensitivity 2013 Waters Corporation 35

Equipment Cleanup What should you do before starting the cleaning process? Cleaning solvents must be contaminant-free. Check them! Do not start to clean the UPLC/HPLC system until you prevent future contamination New and/or clean bottles Never wash UPLC/HPLC glassware in the dishwasher Rinse only with the HPLC-grade solvents If the contaminant is known, what determines your choice of solvents? Solvents must be compatible with the UPLC/HPLC materials of construction (no THF with PEEK parts) Solvents and additives must be compatible with the required detector or they will become contaminants. 2013 Waters Corporation 36

UPLC/HPLC Cleanup Choices of cleanup solvent mixtures Acetonitrile and isopropanol dissolve different things Some compounds require a mixture of water and organic ph will affect solubility Strong acids or bases for cleaning can cause other problems 6N Nitric acid could dissolve particles from bottle filters. Do not use with these filters. Don t use 6N Nitric acid for UPLC If necessary, use 30% phosphoric acid with UPLC. 1% Ammonium hydroxide will dissolve silica and glass releasing Si ions 2013 Waters Corporation 37

UPLC/HPLC Cleanup Several Mixtures Starting place Isopropanol (IPA) 50:50 acetonitrile-water + 0.1% formic acid Basic mixture good for PEG, amides, esters 50:50 acetonitrile-water + 1% ammonium hydroxide Organic mixture good for hydrophobic compounds 25:25:25:25 acetonitrile-methanol-isopropanol-water + 0.1% formic acid Organic mixture Isopropanol-water + 1% acetic acid Acid cleanup 30% phosphoric acid (~4.4N) for UPLC. Nitric acid for HPLC. Use a last resort. Removal of all traces of acid with water is required. 2013 Waters Corporation 38

UPLC/HPLC Cleanup Cleaning the injector Cleaning suggestion Remove column. Put wash lines in the cleaning mixture. Fill a vial with cleaning mixture. Inject multiple full loops with the cleaning mixture. Replace the parts if the cleaning of the injector does not work. 2013 Waters Corporation 39

Summary Prevent contamination is easier than troubleshooting and cleaning up. 2013 Waters Corporation 40

Troubleshooting and diagnostics Troubleshooting Pressure related Problems Retention Time Reproducibility Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 41

Split and distorted Peaks Why does my column yield split/distorted peaks? As in traditional liquid chromatography, several possible factors could be contributing to a split or distorted peak: Poor tubing connections can result in voids forming, giving distorted peaks. Wrong system volume. Check tube diameters and length. Blocked in-line filter What are you using as your needle wash? when choosing the needle wash use an apropiate composition 2013 Waters Corporation 42

Split and distorted Peaks What are the specific wash volumes, set in the instrument method? If any strong wash is leftover, peak distortion can occur What is the sample diluent? It might need to be similar to the mobile phase What is the injection volume? Is the sample overloaded? Are you using a mobile phase pre-heater for UPLC? Thermal mismatch of mobile phase and column Have you allowed for proper column equilibration? 2013 Waters Corporation 43

Effect of Contaminated/blocked In-line Filter on Peak Shape/Efficiency ciency Debris from seal shedding, particulates from buffer, particulates from sample Contaminated Frit P = 6400 psi* N = 7575 Sample Band Column New Frit P = 5100 psi* N = 9349 2013 Waters Corporation 44

Effect of a Poor Fitting/Poorly Cut Peak Tubing Suspect Issue: Observed kink in needle or poor fittings Remedy: Check all fittings and Replace kinked needle with new needle Note: Never try to bend needle to original state. 1. Problem not fixed with new needle 2. Fittings checked and tested on column position 1 3. Switched column to position 2 thru 4 4. Problem still observed during each change!!! 2013 Waters Corporation 45

Effect of a Poor Fitting/Poorly Cut Peak Tubing Injection: Column manager left valve in-line Remedy: Since earlier tests on each column position resulted in the split peaks, proceeded to check fitting and tubing on outlet of valve to the detector Problem resolved. 2013 Waters Corporation 46

Poorly Cut Peak Tubing UPLC Tube Poorly Cut Peak Tubing 2013 Waters Corporation 47

HPLC/UPLC Fittings Classic Waters HPLC fittings UPLC fittings 0.062 0.060 0.130 0.063 0.062 0.087 2013 Waters Corporation 48

Over tightened Finger tight Fittings 0.040 Reusable Fitting Installed Correctly 0.030 au 0.020 Normal Ferrule 0.010 0.000 0.025 Overtightened Reusable Fitting 0.020 0.015 au 0.010 Deformed Ferrule 0.005 0.000 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 2013 Waters Corporation 49 Minutes

In-line filter installation problem N 5sigma = 5025 Tailing factor = 1.31 Wrong position of ferrule N 5sigma = 6959 Tailing factor = 1.17 Correct position of ferrule 2013 Waters Corporation 50

Troubleshooting and diagnostics Troubleshooting Pressure related Problems Retention Time Reproducibility Carryover/Contamination Split and Distorted Peaks Baseline Noise 2013 Waters Corporation 51

Baseline noise 2013 Waters Corporation 52

Baseline noise characterization Non-cyclic (erratic) baseline noise Cyclic (synchronous) baseline noise Short term cycling Long term cycling Baseline drift Noise spikes on baseline 2013 Waters Corporation 53

Troubleshooting procedure To isolate the source of the baseline noise (detector or not detector): Stop the flow Monitor the baseline for a few minutes* and observe : If there is a significant improvement in the baseline noise the problem is within the fluid path (pump/mobile phase/flow path/column) If the noise continues the problem is within the detector or its electrical connections. *Some flow sensitive detectors (such as RI, electrochemical) may require a significant time to stabilize once flow is stopped. 2013 Waters Corporation 54

Fluid path-related noise Short term (seconds to minutes) cyclic noise: Most often related to pump pressure/flow fluctuation Air in pump Remove air degas solvents Faulty check valve Replace check valve Wrong plunger seals Replace seals Broken plunger Replace plunger Inadequate solvent blending Increase mixing volume 2013 Waters Corporation 55

Fluid path-related noise Long term (minutes to hours) cyclic noise : Ambient temperature fluctuations Stabilize column temp. 5 C > ambient temp. Solvent recycling? Avoid recycling if not absolutely necessary 2013 Waters Corporation 56

Fluid path-related noise Non-cyclic (erratic) noise : Air bubble trapped in detector flow cell Remove air in flow cell o To prevent air in flow cell add 50-100 cm of 0.23mm ID tubing to the detector outlet* Small air bubbles traveling through the flow path Degas mobile phase remove air from pump System not stabilized Equilibrate system Low Detector Energy * Keep in mind that not all detectors (such as Fluorescence, RI, Conductivity and Electrochemical) can tolerate backpressure on the flow cell. Consult the manual. 2013 Waters Corporation 57

Fluid path-related noise Non-cyclic (erratic) noise (continued): Mobile phase contaminated Prepare fresh mobile phase. Clean solvent filters Detector flow cell leaking Check for leaks - repair Column contaminated Remove column and see if noise disappears o Replace/clean column 2013 Waters Corporation 58

Fluid path-related noise Baseline drift : System not equilibrated Equilibrate system Temperature fluctuations Stabilize column temperature Mobile phase contamination Prepare fresh mobile phase. Clean solvent filters 2013 Waters Corporation 59

Fluid path-related noise Baseline drift : Contaminated column Remove column and see if noise disappears o Replace/clean column Stationary phase bleed (ligand hydrolysis) Remove column and see if noise disappears o Check ph of mobile phase (<2?) o Select different ph o Select different column type (trifunctional) Function of gradient and difference in UV absorbance of solvents Does drift follow gradient curve/profile? o Use higher wavelength o Replace methanol w. acetonitrile 2013 Waters Corporation 60

Gradient baseline drift 10mM Phosphate - Methanol gradient 5-80% in 10 min. 10mM Phosphate - THF gradient 5-80% in 10 min. 25mM Amm. acetate - Methanol gradient 5-80% in 40 min. 50mM Amm. bicarbonate - Methano gradient 5-60% in 10 min. 2013 Waters Corporation 61

Detector related noise Spikes on the baseline Defective lamp Air bubble in flow cell Non-cyclic noise Contaminated flow cell 2013 Waters Corporation 62

Detector Tips and Tricks Optimizing Noise and Resolution Performance Independent optimization of data rates and digital filtering on detector allows for optimization of data rate without sacrificing resolution Detector set up Data rate Filter constant 2013 Waters Corporation 63

Importance of Sampling Rate Must ensure enough points are collected across a peak to adequately define the peak shape. Peak detection algorithms require a minimum number of points across a peak to distinguish it from baseline noise and correctly determine peak lift off and touch down. A peak which does not have enough data points will be difficult to integrate and therefore have irreproducible peak areas and heights. We aim at collecting 25-50 points across a peak. 2013 Waters Corporation 64

Effect of Sampling Rate AU 0.080 0.070 0.060 0.050 0.040 1 pt/s 2 pts/s 5 pts/s 10 pts/s 20 pts/s 40 pts/s 0.030 0.020 0.010 0.000 0.50 0.52 0.54 0.56 0.58 0.60 0.62 0.64 0.66 0.68 0.70 0.72 0.74 Minutes La Cuantificación Reproducible Requiere un Mínimo de 15-20 Puntos por Pico 2013 Waters Corporation 65

Importance of Sampling Rate 2013 Waters Corporation 66

What is Digital Filtering? Digital Filtering is a mathematical algorithm applied to a data set that smoothes out higher frequency noise. The desired result is reduced baseline noise with minimal impact on peak intensities so as to increase signal-to-noise. The degree of filtering (time constant) is crucial, too much filtering can dramatically impact peak shapes and resolution. 2013 Waters Corporation 67

Optimizing Digital Filter Settings Ensure that the data rate is set to collect 25 50 points Start with the FAST, NORMAL, or SLOW settings depending upon the requirements for resolution versus sensitivity Optimize for the method Start with the inverse of the sampling rate Filtering Time Constant = 1/Sampling Rate Is increased sensitivity needed? Is baseline noise interfering with integration? Filtering Time Constant Is increased resolution needed? Filtering Time Constant 2013 Waters Corporation 68

Effect of Filter Time Constant Setting Other = 1.0 sec. Other = 0.5 sec. Slow = 0.2 sec. Normal = 0.1 sec. Fast = 0.05 sec. No Filter 2013 Waters Corporation 69

Tips and tricks to prevent problems Solvent considerations 2013 Waters Corporation 70

Tips and Tricks General Recommendations - Mobile Phase Preparation Use High quality organic solvents, buffers and additives Don t use metallic container solvents Always make sure that there are no insoluble particles in the eluents Usually High Grade organic solvents are filtered through a membrane (read the label on the bottle) Use high quality salts to prepare buffered eluents Use fresh Milli-Q water Change aqueous mobile phases often. Every 24 48 hr 2013 Waters Corporation 71

Tips and Tricks General Recommendations - Mobile Phase Preparation All bottles must be capped Do not use Parafilm Use clean glassware (bottles and for filtration) Who takes care of this? How is it managed? o Never use communal, department dishwashers o Never use any detergents for cleaning, especially for mass spectrometer applications What is the quality of glass? Is it compatible with the mobile phase (ph)? Bottles must be capped Do not top-off the bottles(can promote microbial growth) 2013 Waters Corporation 72

Tips and Tricks General Recommendations - Mobile Phase Preparation Flush buffers out of system with water after use (after cleaning use 10-20% organic in water for storage). Keep all solvent lines primed (use 10-20% organic in water for unused lines, MeOH or IPA are fine) Keep seal wash primed. (90-95% water) Re-prime solvent lines before starting 2013 Waters Corporation 73

Questions? 2007 Waters Corporation