"Quiet-Zero TE" ASL brain MR angiography Can Outperform 3D TOF at 1.5T

Similar documents
Comparison of radiation dose from X-ray, CT, and PET/ CT in paediatric patients with neuroblastoma using a dose monitoring program

Determination of bone age using MRI of hand/wrist: a pilot study

Utilization management for successful process optimization in radiology

MRI scanning of the claustrophobic patients

CT findings in Differential Diagnosis between Tuberculous Pleurisy and Malignant Effusion

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

Diffuse infiltration in multiple myeloma treatment response assessment with "total-spine" contrast enhanced MR imaging

OA related pain medication intake in subjects from the OAI incidence cohort - association with focal knee lesions and cartilage T2 measurements

Preoperative evaluation of future remnant liver function by the contrast enhance ratio in hepatocellular image.

Patterns of nodal spread in thoracic malignancies

An Image Based Semiautomatic Online Cancer Staging Program for Common Female Pelvic Malignancies

Eye lens dose measurements in Interventional Cardiology

Needle crystal detector technology in mammography further dose reduction and clinical image quality with

Normal vascular variants of the upper extremity

Musculoskeletal MRI Technical Considerations

Acute abdominal pain in the elderly patient: Impact of early MDCT examination on diagnosis and management

Role of 3D volumetry CT in the correlation between postoperative gastric volume and weight loss in obese patients undergoing gastric sleeve surgery.

SITE IMAGING MANUAL ACRIN 6698

DICOM metadata-mining in PACS for computed radiography X-Ray exposure analysis: a mammography multisite study

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

Development and implementation of a help desk system in a radiology department in a high complexity hospital: a South American experience

5 Factors Affecting the Signal-to-Noise Ratio

Pancreatic masses: What is there besides cancer

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

32-Channel Head Coil Imaging at 3T

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

Angio-CT preoperative planning of inferior limb free flaps in plastic surgery

GE Medical Systems Training in Partnership. Module 12: Spin Echo

Basic Principles of Magnetic Resonance

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Metastatic malignant melanoma revisited

MGH Adult Diffusion Data Scanning Protocols

MEDIMAGE A Multimedia Database Management System for Alzheimer s Disease Patients

Prepublication Requirements

Differential diagnosis of vertebral compression fracture using in-phase/opposed-phase and Short TI inversion recovery imaging

Magnetic Resonance Imaging

Comparison between photoacoustic mammography images of breasts cancers and histological staining patterns of CD31 and carbonic anhydrase IX

Abdominal CT Perfusion: Effects of Breath Control Technique

S-822. Noise and Hearing Conservation RISK MANAGEMENT

MRI SEQUENCES. 1 Gradient Echo Sequence

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

The MRI Study Guide for Technologists

PET/CT-MRI First clinical experience

Noise at the Work Site

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

OAK GROVE SCHOOL DISTRICT. Hearing Conservation Program

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Hearing Conservation Program

DURATION OF HEARING LOSS

FDA Guidelines for Magnetic Resonance Equipment Safety

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

COST AID ASL post- processing Workshop

Stanford University. Hearing Conservation Program

NEURO MRI PROTOCOLS TABLE OF CONTENTS

Kap 8 Image quality, signal, contrast and noise

Breast MRI Quality Control

How To Improve Your Ct Image Quality

Diagnostic performance of MRI in differentiating metastatic from acute osteoporotic compression fractures of the spine

runl I IUI%I/\L Magnetic Resonance Imaging

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

MRI Physics for Radiologists

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)

REVIEW. Magnetic Resonance in Medicine 00:00 00 (2014)

In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer

University Children s Hospital Basel

Noise. CIH Review PDC March 2012

Adolescent Brain Development and Effects of Alcohol Use

Unwarping Echo Planar Images Using CMTK 1 Release 1.2

Current Industry Neuroimaging Experience in Clinical Trials Jerome Barakos, M.D.

ACOUSTICAL CONSIDERATIONS FOR EFFECTIVE EMERGENCY ALARM SYSTEMS IN AN INDUSTRIAL SETTING

Noise: Impact on Hearing; Regulation

Voxar 3D TM. A suite of advanced visualization and analysis software tools

The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology

MRI Department Goals. Effective Ways to Improve Patient Cooperation and Safety in MRI Exam. Lecture Topics

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

Program No Section Heading Page # 6.0 Monitoring Training Audiometric Testing Noise Exposure Control 6

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010

MRI for Paediatric Surgeons

The File-Card-Browser View for Breast DCE-MRI Data

UCSC HEARING CONSERVATION PROGRAM Environment, Health & Safety (EH&S) August 2007

Imaging Markers of Brain Network Dysfunction in Multiple Sclerosis

Respiratory dynamic MRI for determining aortic invasion of thoracic neoplasms

A Study on Relationship between Power of Adapter and Total Harmonic Distortion of Earphone s leaking Sound

Questions and Answers for Parents

Video-Based Eye Tracking

What Is an Arteriovenous Malformation (AVM)?

Second look ultrasound examination for breast lesions: MRI and pathologic correlation

Measure #195 (NQF 0507): Radiology: Stenosis Measurement in Carotid Imaging Reports National Quality Strategy Domain: Effective Clinical Care

noise induced Working Together to Prevent Hearing Loss

Hemorrhagic venous infarction Heather Borders, MD

Shorter Scanning Times Through Efficient Software. Annette Tuffs. Article from the customer magazine Medical Solutions, December 2010

THE WIDEST WIDE-BORE IN THE INDUSTRY

Noise at work a guide for health and safety representatives

Configuration and Use of the MR Knee Dot Engine

2. MATERIALS AND METHODS

In vivo stem cell tracking in the myocardium of small laboratory animals: Feasibility and acquisition strategies on clinical magnetic resonance

Human Exposure Limits

Transcription:

"Quiet-Zero TE" ASL brain MR angiography Can Outperform 3D TOF at 1.5T Poster No.: C-2666 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit I. Corcuera Solano 1, W. Sun 2, G. Wu 2, L. N. Tanenbaum 1 ; 1 New York, NY/US, 2 Waukesha, WI/US Neuroradiology brain, MR-Angiography, Safety, Imaging sequences, Quality assurance 10.1594/ecr2013/C-2666 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myesr.org Page 1 of 11

Purpose Background: Acoustic noise generated during magnetic resonance (MR) imaging is an unwanted side effect that may cause discomfort in patients and healthcare professionals. The problems associated with acoustic noise include simple annoyance, heightened anxiety, verbal communication difficulties (1), temporary hearing loss and, in extreme cases, the potential permanent hearing impairment (2-4). Temporary shifts in hearing thresholds have been reported in 43% of the patients scanned without ear protection and patients with improperly fitted earplugs (2). Additionally, acoustic noise may pose a particular hazard to specific patient groups who may be at increased risk (e.g., patients with psychiatric disorders, elderly, pediatric and sedated patients, newly born children) (4). Furthermore, noise exposure for the fetus could be of concern for both patients and interventional MRI staff (5). The gradient magnetic field is the primary source of acoustic noise in MR procedures (4, 6). This noise occurs during the rapid alterations of currents within the gradient coils. These currents, in the presence of the strong static magnetic field of the MR system, produce significant (Lorentz) forces that act upon the gradient coils. Acoustic noise is produced when the forces cause motion or vibration of the gradient coils as they impact against their mountings, which, in turn, flex and vibrate (7, 8). The acoustic noise varies due to the alteration of the gradient output (rise time or amplitude) by modifying MR imaging parameters. Noise tends to be enhanced by decreases in section thickness, field of view, repetition time, echo time, and as field strentgh arises. Gradient magnetic field-induced noise levels have been measured during a variety of pulse sequences for MR systems with static magnetic field strengths reporting sound pressure levels that can run as high as 100-120 db. (6, 9-11). The situation is exacerbated in ultra-high speed imaging because of the very high switching rates used in these techniques. Noise levels can be as high as 140 db, which is well above generally accepted safety level permitted in the workplace (7). Hearing protection is applied routinely to all children and adults undergoing MR imaging. Earplugs, when properly used, can abate noise by 10 to 30 db, which is usually an adequate amount of sound attenuation for the MR environment. Unfortunately, passive noise control methods suffer from a number of limitations. They affect the verbal communication with patients during the operation of the MR system, standard earplugs are often too large for the ear canal of adolescents and infants, and passive noise control devices offer non-uniform noise attenuation over the hearing range. While high Page 2 of 11

frequencies may be well attenuated, attenuation is poor at low frequencies. This is problematic because, for certain pulse sequences, the low frequency range is where the peak MR imaging-related acoustic noise is generated. Several investigators have described the development of "quiet" pulse sequences, which substantially decrease acoustic noise and are acceptable for MR imaging and functional MRI examinations. To date two methods have been used to reduce noise - dampen/ isolate the gradient coil from the patient bore, or reduce switching rate. Both methods have drawbacks; the first resulting in reduction of bore space and the second reducing performance. Quieter techniques are feasible which minimize gradient switching over the TR period by utilizing a 3D projection reconstruction with no excitation, pre-winding, phase-encoding, refocus, or crusher gradients. This technique can reduce acoustic noise levels by more than 30dBA while generating an isotropic volumetric acquisition. In addition, the nearly zero-te nature of the "Silent" protocol has the potential to provide superior rendering due to reduced dephasing related signal loss. "Quiet" ASL MRA is a new technique aimed for central cerebral artery imaging by combining ASL with 3D radial acquisition together. The ASL technique is optimized to create a constant labeling state with derated gradient switching, in order to reduce the acoustic noise. The 3D acquisition uses a spiral radial sampling scheme. It does not have an in-plane phase and frequency direction, as well as not having a slice selection phase. Due to the K Space trajectory and data sampling scheme, the gradient steps are relatively small, this accounts for the quiet behavior of this acquisition. Furthermore, the 3D acquisition acquires isotropic voxels and may be reformatted at the native resolution for no loss of spatial acuity. The acoustic noise is mainly created from the labeling component for "Quiet" ASL MRA. The purpose of this study is to compare the quality of "Quiet" arterial spin labeling based (ASL-MRA) acquisitions to clinical 3D TOF techniques in current day to day practice for MRA imaging of the brain. Methods and Materials This is an on-going trial prospective study. Our aim is to include 20 subjects in this project and include two readers for the assessment of the images. Subjects will be scanned with the "Quiet" ASL MRA technique in addition to 3D-TOF on a 1.5T MR. To date, 14 subjects have been included. 1 healthy volunteer (one women, 30 years old) and 13 patients (seven women, six men, age range 85-25) diagnosed with headache (n=2), toxoplasmosis (n=1), viral meningitis (n=1), sarcoidosis (n=1), AIT (n=1), stroke Page 3 of 11

(n=4), myoclonus (n=1) and seizures (n=1) were imaged after obtaining Institutional Review Board approval and informed consent. All exams were performed on a clinical 1.5 T magnetic resonance system (Optima 450W; GE Healthcare, Waukesha, WI) with an 8-channel head coil. Subjects were imaged with a clinical standard 3D TOF scan and a whole brain "Quiet" ASL MRA scan. The standard 3D TOF scan was acquired with the following parameters: TR/TE 15/3.2 flow compensated; field of view 18 x 18 cm; matrix 384 x 224; slice thickness 1 mm; flip angle 20º; bandwidth 31.25 khz; number of slabs 4 with an 8 partition overlap; slice thickness 1 mm/0.5 mm. The examination time was 3:59 min. For the "Quiet" ASL MRA scan, the labeling plane was approximately placed at C1 segment of the internal carotid arteries such that the major feeding arteries of the brain vasculature including internal carotid arteries, external carotid arteries, and vertebral arteries will be labeled. A 16 cm3 volumetric imaging slab was prescribed immediately above the labeling plane. ASL parameters include: labeling duration 3 s; matrix size 160 x160 ; 3D isotropic resolution 1 mm; readout bandwidth 31.25 khz; TR/TE (ms) 590/0.016; fractional echo 1; flip angle 4º. The scan time was 6:56 min. Images from both 3D TOF and "Quiet" ASL MRA was presented, on a commercial imaging workstation, to one experienced neuroradiologist who qualitatively the images. The grading was not blinded as the 3D TOF and "Quiet" ASL exams can be readily distinguished by the presence or absence of slab artifact and residual background signal on the 3D TOF studies. 3D TOF and "Quiet" ASL MRA examinations were evaluated using three criteria: vessel segment visualization, saturation effect, and luminal uniformity (Table 1). Four vessel segments were evaluated for each subjects (segment precavernous =1, cavernous= 2, clinoid= 3, distal= 4) (fig. 1) A non-parametric paired Wilcoxon test was used to compare the following parameters between 3DTOF and "Quiet" ASL MRA: quality of the vessel segment visualization, luminal rendering uniformity, and saturation effect. Statistical analysis was performed using commercially available software (SPSS version 20). A p < 0.05 indicated a statistically significant difference. Images for this section: Page 4 of 11

Table 1: Criteria for qualitative evaluation Page 5 of 11

Results Preliminary Results: When the total vessel was evaluated (four segments) a higher mean image quality score was found for 3D TOF in comparison with "Quiet" ASL MRA (p=0.04). No significant differences in the means scores for the complete vessel were found for saturation effect and luminal rendering uniformity between the two sequences (Table 1). When a segment vessel analysis was performed, a higher mean image quality score was found for 3D TOF that with "Quiet" ASL for segment 3 (clinoid segment). For all the other vessel segments (segment 1, 2, 4) no significant differences were found for image quality, saturation effect and luminal rendering uniformity between the two techniques. "Quiet" ASL MRA was slightly better in the segment 2 for image quality, saturation effect and luminal rendering uniformity that 3D TOF. However, there was no statistically significant difference between the two methods in any of these groups. Images for this section: Page 6 of 11

Fig. 2: Sagittal MIP images of "Quiet" ASL MRA and 3D TOF demonstrate "Quiet" ASL MRA seems to be superior in image quality to conventional 3D TOF for proximal artery. For distal artery, 3DTOF is still better. Page 7 of 11

Table 2 Page 8 of 11

Fig. 1: AP view (A) MIP of "Quiet" ASL MRA and (B) MIP of 3D TOF. Images show complete filling of the entire arterial vasculature of the brain. Signal intensity decreases from proximal to distal vessels. Page 9 of 11

Conclusion "Quiet" ASL MRA seems to produce superior image quality to conventional 3D TOF acquisitions with a lower tendency to turbulence related signal loss for proximal artery. For distal artery, 3DTOF is still better due to high signal noise ratio and high resolution. "Quiet" ASL MRA provides superior image quality for proximal artery at reduced sound pressure levels and can be used as an additional method beside conventional 3D TOF for routine MRA of the brain. References 1. Moelker A, Maas RA, Pattynama PM. Verbal communication in MR environments: effect of MR system acoustic noise on speech understanding. Radiology. 2004;232(1):107-13. 2. Brummett RE, Talbot JM, Charuhas P. Potential hearing loss resulting from MR imaging. Radiology. 1988;169(2):539-40. 3. Radomskij P, Schmidt MA, Heron CW, Prasher D. Effect of MRI noise on cochlear function. Lancet. 2002;359(9316):1485. 4. McJury M, Shellock FG. Auditory noise associated with MR procedures: a review. Journal of magnetic resonance imaging : JMRI. 2000;12(1):37-45. 5. Reeves MJ, Brandreth M, Whitby EH, et al. Neonatal cochlear function: measurement after exposure to acoustic noise during in utero MR imaging. Radiology. 2010;257(3):802-9. 6. Shellock FG, Morisoli SM, Ziarati M. Measurement of acoustic noise during MR imaging: evaluation of six "worst-case" pulse sequences. Radiology. 1994;191(1):91-3. 7. Mansfield P, Haywood B, Coxon R. Active acoustic control in gradient coils for MRI. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 2001;46(4):807-18. 8. Bowtell RW, Mansfield P. Quiet transverse gradient coils: Lorentz force balanced designs using geometrical similitude. Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1995;34(3):494-7. 9. Moelker A, Wielopolski PA, Pattynama PM. Relationship between magnetic field strength and magnetic-resonance-related acoustic noise levels. Magma (New York, NY). 2003;16(1):52-5. 10. Hurwitz R, Lane SR, Bell RA, Brant-Zawadzki MN. Acoustic analysis of gradient-coil noise in MR imaging. Radiology. 1989;173(2):545-8. Page 10 of 11

11. Moelker A, Maas RA, Lethimonnier F, Pattynama PM. Interventional MR imaging at 1.5 T: quantification of sound exposure. Radiology. 2002;224(3):889-95. Personal Information LN Tanenbaum MD FACR, I Corcuera-Solano MD. Department of Radiology. Neuroradiology. Icahn School of Medicine at Mount Sinai One Gustave L. Levy Place New York, NY 10029-6574 Nuromri@gmail.com Page 11 of 11