PVD 75 DC and RF SPUTTERING SYSTEM SOP

Similar documents
Balzers Sputter Coater SCD 050

Electron Beam and Sputter Deposition Choosing Process Parameters

Schneps, Leila; Colmez, Coralie. Math on Trial : How Numbers Get Used and Abused in the Courtroom. New York, NY, USA: Basic Books, p i.

Reactive Sputtering Using a Dual-Anode Magnetron System

Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models)

POWER TRIM 5 F AUTO TRIM AND AUTO TRIM

VCR Ion Beam Sputter Coater

Vacuum Evaporation Recap

HP Pavilion All-in-One MS200 series PC. Upgrading and Servicing Guide. Printed in

Kurzweil KORE 64. ROM Expansion Board for the PC3 and PC3K

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater

High Rate Oxide Deposition onto Web by Reactive Sputtering from Rotatable Magnetrons

EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board.

Coating Technology: Evaporation Vs Sputtering

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter

Dell P Series Monitor VESA Mounting Bracket Installation Instructions

Product End-of-Life Disassembly Instructions Product Category: Consumer Display

WHITEPAPER ENHANCED REACTIVELY SPUTTERED AL 2 O 3 DEPOSITION BY ADDITION OF ACTIVATED REACTIVE OXYGEN

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

CEILING CASSETTE FAN COIL UNIT

LT SERIES User s Manual

Instrument Software Update Instructions. Keysight X-Series Signal Analyzers (PXA/MXA/EXA/CXA) Upgrading an older version to A.12.

Connecting External Resources

Memory Modules User Guide

Atomic Force Microscope

Thermo Top - Troubleshooting Tree

Neuere Entwicklungen zur Herstellung optischer Schichten durch reaktive. Wolfgang Hentsch, Dr. Reinhard Fendler. FHR Anlagenbau GmbH

A Remote Plasma Sputter Process for High Rate Web Coating of Low Temperature Plastic Film with High Quality Thin Film Metals and Insulators

Kodak Home Page on Internet. Intranet. Table of Contents. Eastman Kodak Company, Publication No. SM NOV97 Supersedes SM MAY97

Lecture 12. Physical Vapor Deposition: Evaporation and Sputtering Reading: Chapter 12. ECE Dr. Alan Doolittle

RVR Hybrid System Installer Instructions RVR Hybrid System Controller Installation Instructions

Operation Manual For NDJ-8S Digital Rotary Viscometer

OPL BASIC. Dosing System for Professional Laundry machines. Contents

RF Etch Monitor. Introduction

The Black Box Push Button Control Box

SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE

Internal Modem Installation with Windows 95

COMMERCIAL GAS DRYER

1. Installing The Monitoring Software

Heat Surge Model X5C Fire Place Insert Service Manual Applies to all units w/ circuit board

EVBIKE LCD Display Control User Guide. EVBIKE LCD Display Control User Guide

TECHNICAL SERVICE DEPARTMENT Technical Service Bulletin Tankless Electric (RTE) Troubleshooting

USER MANUAL WARNING! CONTENTS MODEL 1 SPECIFICATIONS READ ALL INSTRUCTIONS BEFORE PROCEEDING. Non-Programmable Single Stage Heat/Cool Thermostat

Vent Valve Output. Vent Valve Feedback. High Vac Valve Output. High Vac Valve Feedback. Full Range Gauge #1 Crossover Interlock

Vertical Display and Storage B SKOPE Gen2: Three Door Chiller

INSTALLATION INSTRUCTIONS

OPERATING INSTRUCTIONS

Express5800/120Ed. Rack Mount Kit Installation Procedures PN:

CHECK WITH LOCAL CODE AUTHORITIES ABOUT INSTALLATION REQUIREMENTS FOR THIS PRODUCT INSTALLATION INSTRUCTIONS

Western Blotting For Protein Analysis

PC1131 Electric Air Compressor

MDHearingAidAIR Diagram

User Manual. Windows 7 / Vista / XP. iotablet Driver

Be careful not to scratch or hit front edge of the side viewing micro prisms onto hard objects!

Heat Pump Training System

ipad for Attorneys 366 South Oyster Bay Road Hicksville, NY Phone: (516) Fax: (516)

OVERVIEW Playbacks: Shortcuts: Memories: Data Entry Wheels: Touchpad: Master and Blackout:

A) What Web Browser do I need? B) Why I cannot view the most updated content? C) What can we find on the school website? Index Page Layout:

OWNER S OPERATION AND MAINTENANCE MANUAL OF THE MAK SPUTTERING SOURCES

Oil and Coolant Circulating Heating System. Model - OCSM

Essential Electrical Concepts

1 DESCRIPTION OF THE APPLIANCE

Step-by-step instructions:

Replacing Redundant AC Power Supplies in an EX4500 Ethernet Switch Without Disrupting Service

Memory Modules. Document Part Number: This guide explains how to replace and upgrade memory in the computer.

Boiler Preparation, Start-Up and Shutdown

C&A AR Online Credit Card Processor Installation and Setup Instructions with Process Flow

Name: Due: September 21 st Physics 7230 Laboratory 3: High Resolution SEM Imaging

Air Conditioning Sign-Off Sheet

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P Thermodynamic cycle Heat pump

INSTALLATION & SERVICE MANUAL. Display Panel

Lasers Design and Laser Systems

PowlVac Vacuum Integrity Tester

The RIDZ 8x2 Audio Switcher

Permanent Magnet Motor Kit, Magnetic Reed Type. (SKY-ReedMotorKit) Instructions

Applied Electronics. Commercial Dimming System UPDATE NOTICE

Introducing Your Cisco IP Phone

EasyPen i405x. MousePen i608x. User Manual For Windows 7/Vista/XP

Installation Instructions

PowerPoint Presentation

Cisco Unified IP Phone 8961, 9951, and 9971 User Guide for Cisco Unified Communications Manager 8.0 (SIP)

Installation, operation and maintenance manual TX 35A

For Touch Panel and LCD Sputtering/PECVD/ Wet Processing

Win 7 - STIMA CLS / CMP Driver Installation Quick Guide

Deposition of Thin Metal Films " (on Polymer Substrates)!

Mustang Series PRESSURE REDUCING CONTROL VALVE WITH PRESSURE SUSTAINING FEATURE. M115-2 (Globe) M (Angle) Schematic. Standard Components

PERFECT FIT SERIES IN-DASH HEAT/ COOL/ DEFROST 1967 CHEVROLET IMPALA

Service Guide. Gateway M275

Experiment 5. Strain Gage Measurements

Smart Cam, CC-Smart-Cam, and Smart Cam Packages Installation and Quick Start Operating Instructions

Operation Manual. 150 Watt Halogen Light Source for Endoscopy Model # s ESS-150, ESS-150A ESS-220, ESS-220A

Memory Modules. Document Part Number: This guide explains how to replace and upgrade memory in the computer. May 2006

2A.1 Features of Chiller Control Components A.3 Chilled-Water Supply Temperature Control A.4 Cooling-Water Supply Control...

ILLUSTRATED PARTS LIST XTREME ICE MAKER (1230 SERIES) MODEL NO.

Service Overview. Chapter

The Strange Effects of a Vacuum

Table of Contents Getting Started... 3 The Motors... 4 The Control Board... 5 Setting up the Computer with Mach Starting up the Equipment...

AOYUE. INT 738H (Centigrade / Fahrenheit switchable) Multi-Function Re-Working System INSTRUCTION MANUAL. Replacement Air Nozzles

Transcription:

PVD 75 DC and RF SPUTTERING SYSTEM SOP May 2013 System Start Up 1. The sputterer and utilities should be on and the chamber should be under vacuum. 2. Simply touch the screen or mouse pad and the computer will awake. 3. If the computer or entire system is off contact cleanroom staff. Chamber Vent 1. Select the VACUUM page at the top left of the screen 2. Note the current vacuum found at the bottom of the screen in Torr. See Figure 1. If the vacuum is greater than 1 x 10-5 Torr there may be a problem. 3. To open the chamber press START PC VENT located at the middle right of the screen. Target Loading 1. Once vented, select the DEPOSITION button to allow one to open the target and substrate

shutters. See Figure 2. 2. Press the SOURCE SHUTTER button (near the top left corner of the screen) corresponding to the source you would like to use. Observe that the shutter within the vacuum chamber corresponding to your selection has opened. Source 1 and 2 are 3 sputter heads and source 3 is for 4 sputter targets. 3. Change the material listed in the Source Configuration section of the deposition screen to reflect the target you are loading. Click (or press) in the Material name box that corresponds to your source number, and then enter the correct abbreviation for the material you are loading. 4. Remove the gas inlet tube from the dark space shroud if it is connected. Note: it may be helpful to have the tube connected to the shroud when one is doing low pressure depositions (< 5 mtorr). See Figure 3. 5. Remove the dark space shroud around the sputter head by loosening the 3 screws around the periphery of the sputter head. 6. Remove the ring(s) holding the target in place (remove 4 screws) 7. Remove old target and replace with the new target. CAUTION: The magnets in the sputter head, a.k.a. cathode, are strong and can snap magnetic targets to the head with substantial force. Watch your fingers when inserting new targets.

8. Replace target holder ring (3 screws). For 1/4 targets use 2 additional spacer rings, 1/8 use 1 ring and for 1/16 targets use no rings. 9. Replace dark space shroud. For thicker targets use the shallow set of notches on the shroud and for thin use the deeper set of notches. The screws should be tightened so that the shroud does not move at all. 10. Check for continuity (using the digital multi-meter hanging outside the sputterer) between the target and shroud. These should be electrically isolated from one another. If there is continuity re-seat the dark space shroud or increase the height of the shroud by placing it on a shallower set of notches. For dielectric targets, it is impossible to perform the continuity check. Users are advised that it is common for a short to occur between the dark space shroud and the target holder rings therefore it is important to leave ample space between these electrodes. For RF dielectric depositions a short may not be discovered until one has failed to light the plasma. 11. Close the source shutter. Press the SOURCE SHUTTER button (near the top left corner of the screen) corresponding to the source you would like to use. 12. Pump the system down if finished or proceed to substrate loading below. Press the START PC PUMP button located at the right center of screen.

Substrate Loading 1. With the system vented and the deposition screen selected, press the SUBSTRATE SHUTTER button located at the top left of the screen. 2. Remove the substrate access panel above the substrate holder and shutter. 3. Remove the 4 substrate fixture by pushing it out from the shutter side. 4. Load your substrate into the fixture and replace the substrate access panel. 5. Pump down the chamber. Press the START PC PUMP button located at the right center of screen. DC and RF Deposition 1. Verify that both the targets and substrate have been loaded and the system is at least 5 x 10-5 Torr. 2. Select the VACUUM page at the top left of the screen. 3. Turn on the PC HIGH VAC THROTTLE valve near the center of the screen. See Figure 1. 4. Select the PLATEN MOTION screen at the top right. 5. Start platen rotation by pressing the FWD button under DRIVE MOTOR CONTINUOUS. 6. Select the GAS page from the top center (roughly) of the screen. 7. Start argon flow. Argon flow may be controlled via constant flow rate or via constant pressure. For most applications constant pressure is the preferred method of gas introduction. a. To set Ar flow subject to a target pressure: i. Open the source valve for gas flow, for example SOURCE 1 GAS for sputtering off of source 1 ii. Enter 4 in the mode window of MFC1 iii. Enter a target pressure in mtorr in the widow below CAPMAN PRESSURE SP in the upper left of the screen. Argon flow should begin and the pressure displayed in CAPMAN PRESSURE should approach the set point. b. To set Ar flow for constant flow: i. Open the source valve for gas flow, e.g. SOURCE 1 GAS for sputtering off of Source 1 ii. Enter a target flow rate in the box below SETPOINT SCCM for MFC 1 in the lower left of the screen.

8. Return to the deposition window. Press the DEPOSITION button at the top left of the screen. 9. Start the deposition. In the Power Supply section of the deposition window the user can choose POWER SUPPLY 1 or POWER SUPPLY 2 for RF or POWER SUPPLY 3 for DC depositions. Typically DC sputtering is used for conductive materials and RF is used to deposit dielectrics. a. RF Deposition POWER SUPPLY 1 is used for source 1 and source 3, while POWER SUPPLY 2 is used for source 2. Using independent supplies allows the user to perform co-depositions of two materials at the same time. Co-depositions are not covered in this SOP. For purposes of this document we assume the user has loaded a target into either source 1 or 3. i. In the row corresponding to POWER SUPPLY 1 select the desired target by pressing the button above the source number, e.g. SOURCE SW1 (SRC 1) for source 1 ii. Set the target power in Watts by clicking in the box adjacent to the Power Supply 1 ON/OFF button. If a ceramic or other dielectric target is used it is wise to start the plasma at some low power, e.g. 50 W, then slowly ramp the power up to avoid cracking the target due to thermal expansion. iii. Turn on the power. Click on the ON/OFF button next to POWER SUPPLY 1 label near the center of the screen. Verify that the plasma is lit by looking in the vacuum chamber window. Note the reflected power in the box labeled RFL POWER W in the row corresponding to the power supply being used. Caution: If the reflected power is above 10% of the set point, stop the deposition as this can damage the power supply and there is likely a problem with the sputter cathode. iv. Set ramp rate for reaching the final power (optional). Enter a ramp rate in the box below the label RAMP RATE U/S found in the POWER SUPPLY 1 row. Typical values are 0.5 to 0.15. v. Set the final power (optional). Enter the final value for the power in Watts by clicking in the box adjacent to the Power Supply 1 ON/OFF button. The set point should slowly rise to the set point value. vi. Note the plasma electrical properties. After the power has reached the set point and plasma is lit, note the reflected power (number below RFL POWER W ) and the DC bias (number below DC BIAS V at the right side of the screen) associated with the plasma in the displays at the center right of the screen. If the reflected power is above 10% of the set point, there may be a problem. The DC bias is an indicator of the plasma s impedance, the energy of arriving Argon ions and is a useful diagnostic for the

deposition. vii. Open the substrate shutter. Press the SUBSTRATE SHUTTER button at the top left of the deposition screen. viii. Open the source shutter. Press the SOURCE SHUTTER # button associated with the source at the top left of the screen. ix. Begin timing the deposition. x. Turn off plasma. Turn off the plasma by pressing the Power Supply 1 ON/OFF button xi. Vent the system. Press the START PC VENT button. This will close all shutters, turn off the gas flow, and switch off all power supplies before actually venting. xii. After unloading the substrate pump the system back down and leave it at vacuum b. DC Deposition i. Select the source number corresponding to the desired target by pressing the button above the source number, e.g. SOURCE SW3 (SRC 1) for source 1. ii. Set the target power in Watts by clicking in the box adjacent to the Power Supply 3 ON/OFF button. iii. Turn on the power. Click on the ON/OFF button next to POWER SUPPLY 3 label near the center of the screen. Verify that the plasma is lit by looking in the vacuum chamber window and note the voltage and current associated with the plasma in the displays at the center right of the screen. This current and voltage is an indicator of the plasma s impedance and is a useful diagnostic for the deposition. iv. Open the substrate shutter. Press the SUBSTRATE SHUTTER button at the top left of the deposition screen. v. Open the source shutter. Press the SOURCE SHUTTER # button associated with the source at the top left of the screen. vi. Begin timing the deposition. vii. Stop deposition. Close the source shutter and turn off the plasma. Turn off the plasma by pressing the Power Supply 3 ON/OFF button. viii. Vent the system. Press the START PC VENT button. This will close all shutters, turn off the gas flow, and switch off all power supplies before actually venting. ix. After unloading the substrate pump the system back down and leave it at vacuum.