RISK MANAGEMENT FOR MEDICAL LABORATORIES BY TATI KUBHEKA

Similar documents
Individualized Quality Control Plan

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Laboratory Director Responsibilities

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009

Chapter 2: Quality Assurance and Legal Issues

MEDICAL DEVICE GUIDANCE

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED

2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan

Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing

MANITOBA PATIENT SERVICE CENTRE STANDARDS

How to Verify Performance Specifications

Content Sheet 3-1: Equipment Management Overview

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)

Blood Glucose Monitoring: The Facts about Accuracy

AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189

Quality Requirements of a Point of Care Testing Service

Content Sheet 5-1: Overview of Sample Management

Examples: collecting the right sample on the right patient; labeling the sample accurately;

Checklist. Standard for Medical Laboratory

Center for Clinical Standards and Quality/Survey & Certification Group

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Procedure for Equipment Calibration and Maintenance

Content Sheet 16-1: Introduction to Documents & Records

Content Sheet 12-1: Overview of Personnel Management

Canadian Environmental Technology Verification Program Applicant Information Package CONTENTS

Using and Maintaining Documents and Records. Finding the information when you need it

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

In vitro diagnostic reagent, calibrator and control material stability

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

CHAPTER 13. Quality Control/Quality Assurance

LabGuide 71. Incident Management: Developing a Plan

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1

Internal Quality Management System Audit Checklist (ISO9001:2015) Q# ISO 9001:2015 Clause Audit Question Audit Evidence 4 Context of the Organization

Flow Cytometry. CLIA Compliance Manual*

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

State Water Resources Control Board Environmental Laboratory Accreditation Program

FINAL DOCUMENT. Global Harmonization Task Force. Title: Label and Instructions for Use for Medical Devices

AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK

FMEA and FTA Analysis

Medical Cannabis Laboratory Approval Program

NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory

Jonathan Wilson. Sector Manager (Health & Safety)

AS9100 Quality Manual

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

Incident Management: Developing a Plan

General. What is a laboratory test? What is a waived test?

TG TRANSITIONAL GUIDELINES FOR ISO/IEC :2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES

ISO/IEC QUALITY MANUAL

ASSURING THE QUALITY OF TEST RESULTS

Validation and Calibration. Definitions and Terminology

TEST METHOD VERIFICATION AND VALIDATION

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers.

Content Sheet 10-1: Overview of External Quality Assessment (EQA)

The Top 5 Software Capabilities for Ensuring Food Safety

Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?

Performing Audit Procedures in Response to Assessed Risks and Evaluating the Audit Evidence Obtained

APPENDIX N. Data Validation Using Data Descriptors

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations

Unlocking value across the payment continuum. Enhancing performance in a changing healthcare environment

ISO 9001:2015 Internal Audit Checklist

Measuring equipment in a laboratory.

XL-I 2011: Understanding and Explaining the CAP POCT Checklist: A Tool for Laboratorians to Use With POCT Staff

Software-based medical devices from defibrillators

A Stability Program for the Distribution of Drug Products

APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN

DRAFT RESEARCH SUPPORT BUILDING AND INFRASTRUCTURE MODERNIZATION RISK MANAGEMENT PLAN. April 2009 SLAC I

QUALITY MANUAL ISO 9001:2015

FAILURE INVESTIGATION AND ROOT CAUSE ANALYSIS

DNV GL Assessment Checklist ISO 9001:2015

Sutter Health Support Services Shared Laboratory Position Description. Incumbent: Laboratory Date: October 23, 2006 Written By: Michele Leonard

Effective Root Cause Analysis For Corrective and Preventive Action

CAPABILITY 12: Public Health Laboratory Testing

User checks and maintenance of laboratory balances

THE IMPORTANCE OF PRE-ANALYTICAL PHASE IN LABORATORY TESTING AND DIAGNOSIS. Tg Jiu, September, 2010

Quality Assurance (QA) & Quality Control (QC)

CLTAC SUBCOMMITTEE ON CLS EDUCATION AND TRAINING POSITION PAPER

Content Sheet 13-1: Overview of Customer Service

7 Directorate Performance Managers. 7 Performance Reporting and Data Quality Officer. 8 Responsible Officers

Quality Management Systems Manual

EQR PROTOCOL 4 VALIDATION OF ENCOUNTER DATA REPORTED BY THE MCO

calibrate confidence with Risk Management: The Critical Role of Measurement and Calibration WHITE PAPER SERVICES

Hemostasis analyzer system

Calibration Verification

Validation of measurement procedures

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Going Beyond Laboratory Automation: Do Less Accomplish More. Swedish Covenant Hospital

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

UNCONTROLLED COPY FOR REFERENCE ONLY

Karas Engineering AS9100 QUALITY MANAGEMENT SYSTEM MANUAL

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

June 2008 Report No An Audit Report on The Texas Education Agency s Oversight of Alternative Teacher Certification Programs

STANDARD OPERATING PROCEDURES

USE OF REFERENCE MATERIALS IN THE LABORATORY

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

Implementing ISO 9001

References to Computer and Software in FDA s CGTPs (Final Rule and Draft Guidance) for HCT/Ps Preamble to the CGTP Final Rule

Transcription:

Newsletter of the South African National Accreditation System Issue 25, July 2016 RISK MANAGEMENT FOR MEDICAL LABORATORIES BY TATI KUBHEKA WHY RISK MANAGEMENT? Risk management is a requirement of ISO 15189:2012 To ensure laboratories understand the concept of risk management 4.14.6 The laboratory shall evaluate the impact of work processes and potential failures on examination results as they affect patient safety, and shall modify processes to reduce or eliminate the identified risks and document decisions and actions taken. Risk analysis, assessment and management is not new to the clinical laboratories. FORMAL risk analysis, assessment and management is NEW. INTRODUCTION In the delivery of quality patient care, a clinical laboratory prides itself on performing extensive, effective, and documented quality control (QC), giving healthcare professionals confidence that test results obtained on patient specimens are consistently accurate and reliable. By definition, risk is a measure of the severity of the impact of a potential error, multiplied by the probability of how likely it is that the error will occur and the ability to detect the error if it should re - occur. Correspondingly, risk management is the sequential process of risk identification, risk assessment, and risk mitigation. Simply stated, risk management includes: Risk identification: identifying potential sources of error; Risk assessment: evaluating errors to determine their impact on patient test results; and Risk mitigation: controlling errors in such a way that residual risk is manageable. RISK IDENTIFICATION Risk identification is the first and perhaps the most important step in the risk management process. Correctly identifying potential sources of error for a particular test reveals valuable information 1

for developing an individualized quality control plan. Identifying the potential risks embedded in laboratory testing processes allows for the implementation of an overarching quality control plan that effectively mitigates errors not addressed by the test system s internal quality checks and external quality control. Figure 1. A process map showing key steps in the clinical diagnostic testing process. Medical decisions and effective treatment planning are dependent on laboratory results. Therefore, it is important to understand that in performing an effective risk assessment, one size does not fit all! Every diagnostic test carries varied risks and may require varied control measures. Accurate risk identification is essential when developing and implementing a comprehensive and effective individualized quality control plan. There are several tools available to help laboratory staff perform risk identification. The most effective tool will vary from laboratory to laboratory. Therefore, it is important that all options are considered, and that the most effective tool is selected that will positively impact the uniqueness of each laboratory. The following are examples: Process mapping: A process map is a graphical representation of all the steps in a testing process. This tool is used to analyze a particular testing process by breaking it down into small steps from start to finish (see Figure 1). Fishbone diagram: A fishbone diagram outlines the cause and effect of a testing process. It may be considered to be a graphical representation of all the major elements in a process. This diagram can help labs identify and organize potential errors in a test system (see Figure 2). Risk identification table: A risk identification table is a simple table that lists all the errors identified in the different testing phases for a specific test (see Table 1). Regardless of the tool selected, it is important to consider all phases of sample testing: the preanalytic, analytic, and postanalytic phases. The key objectives in implementing an effective risk management process and, hence, an IQC Plan, are identifying the potential errors in all phases of testing, and ensuring optimal mitigation by putting in place effective and documented processes. Results obtained in performing the risk assessment will be used to develop an IQC Plan. This plan will serve to describe QC practices, resources, and procedures that will be used to monitor and ensure continued quality laboratory testing. As with all laboratory procedures, the laboratory director remains ultimately responsible for the proper development and implementation of each IQC Plan. The laboratory director is required to approve and sign each IQC Plan before it is implemented. RISK ASSESSMENT Although the goal for all healthcare professionals is to optimize the delivery of quality patient care, error rates are evident and distributed among the three phases of testing: preanalytic, analytic, and postanalytic. For example, studies have shown that 46% to 68.2% of lab errors occur during the preanalytic phase. Such errors may include, but are not limited to, inappropriate test requests, order entry errors, misidentification of patients, utilization of improper containers, inadequate sample collection and transport procedures, inadequate sample/anticoagulant volume ratio, insufficient sample volume, sorting and routing errors, and labeling errors. By contrast, roughly 7% to 13% of lab errors occur during the analytic phase. Such errors may include, but are not limited to, equipment malfunction, sample mix interference, undetected failure in quality control, and incorrect testing procedure. And finally, 18.5% to 47% of errors occur in the postanalytic phase. Such errors may include, but are not limited to, failure in reporting, erroneous validation of analytical data, and improper data entry and reporting. 2

1 Samples 2 Operator 4 Laboratory Environment Sample Integrity - Lipemia - Hemolysis - Interfering subtances - Clotting - Incorrect tube Sample Presentation - Bubbles - Inadequate volume Identify Potential Hazards Operator Capacity - Training - Competency Reagent Degradation Quality Control Material Degradation Operator staffing - Short staffing - Correct staffing 3 Reagents Atmospheric Environment - Dust - Temperature - Humidity Calibrator Degradation Utility Environment - Electrical - Water quality - Pressure 5 Measuring System Instrument Failure - Software failure - Optics drift - Electronic instability Incorrect Test Result Inadequate Instrument Maintenance - Dirty optics - Contamination - Scratches Figure 2. A fishbone diagram showing potential sources of error associated with five major components of the diagnostic testing process. From these examples, one can conclude that greater emphasis has traditionally been placed on risk identification and risk mitigation for the analytic phase of testing versus the preanalytic and postanalytic phases. However, the high error rates apparent in the preanalytic and postanalytic phases demonstrate precisely the need for risk management in all phases of testing. Risk management is an all-inclusive quality management process that guides the laboratory staff to evaluate potential risks in all phases of testing, which is the essence of IQC Plan. Element Specimen Personnel Preanalytical Sample quantity insufficient Lack of technical or phlebotomy training Analytical Interfering substances Clotted specimen Competency assessment not peformed in past 2 years Postanalytical Inadequate training in how to report critical value results Element Reagents Postanalytic Activities Preanalytical Lack of inventory control Analytical Improper reagent mixing Postanalytical LIS not verfied after recent update Table 1. A risk identification table enables laboratory staff to identify sources of error in each of the five major areas of diagnostic testing, and to further divide errors according to the phase of testing in which they occur. When performing a risk management process and, hence, evaluating the entire testing process, it is important to consider the following key areas within the five components of the diagnostic process that affect the quality patient test results: environment, testing personnel, specimen, testing process, and reagents. Lab Environment Test Systems Heavy traffic in speciment receiving area Calibration overdue Fluctuating temperatres Tst system lacks mechanism for error detection In doing so, how the test is performed, how often relevant errors or undesirable conditions occur, the potential impact of those errors, what control activities are in place to detect or prevent those errors, how the test is used by the physician, patient population, and volume 3

of testing can all be considered in developing the risk assessment. The following sections offer some key examples of risk assessment considerations in each of the five areas. Environment. Areas of importance that require focus include the following: Where is testing performed, and what other activities occur nearby? Are room temperature and humidity stable? Are testing areas level, and draft- and vibration-free? Does altitude affect the testing? Have lighting, electricity, water quality, and other utilities been considered? Testing Personnel. Individuals performing testing must be evaluated to ensure that their training and competency assessment records can be used to validate the qualifications of testing personnel to accurately perform testing. Key questions to be addressed in this area include the following: Do testing personnel have laboratory medicine education and experience? Have testing personnel been adequately trained to perform the test? Have competency assessments been performed on testing personnel? Although testing personnel may have been trained to perform specific types of testing, it cannot be assumed that all testing personnel maintain the superior level of performance they initially demonstrated. It is therefore essential that competency assessments be performed on a continuing basis. Specimen. Ensuring the collection of the right specimen from the right patient is essential. Review specimen collection, handling, and storage procedures. Review all instructions that are provided to patients regarding preparation for the self-collection of specimens. It is important to ensure that all specimens are handled and stored appropriately, and are suitable for testing. Testing Process. It is important to gather information regarding the measuring system, the instrument, or other test device. Varied types of information can contribute to risk assessment in this area. For commercialized instruments, tests, and test kits, a great deal of such information is supplied by the manufacturer through a package insert or operator s manual. A test system may have safeguards such as lockout functions and error codes that detect and prevent errors. Laboratory historical data can be considered. Maintenance logs, QA and QC records, verification of performance specifications, and calibration verification records are a few examples of documentation to review. In addition, an assessment of risk should consider the utilization of test results by the ordering physician when developing a treatment plan for the patient. Some test results represent only a portion of the information that contributes to clinical decisions. Others are used immediately, as the sole decision-making criterion. In the latter situation, an inaccurate result has a much greater potential to cause harm, making an understanding of the clinical use of test results an important element for determining the quality control measures a lab needs to perform. Reagents. Test reagents can be compromised during shipment, handling, storage, and processing. In addition, consideration should be given to the quality and stability of reagents used directly as part of the QC process: Calibrators Controls It is important that package inserts and storage records are gathered and reviewed to assess these potential risks. RISK MITIGATION For any risk that is deemed unacceptable, the lab should identify ways to reduce the probability of harm, using prevention and detection methods, in order to bring the risk down to an acceptable level. This may be through a variety of means such as revised operator training, posted warnings, more robust QC rules, greater surveillance of the process, or even repeat testing for values exceeding a specified threshold. In cases where the test manufacturer indicates that certain aspects of the test are monitored through built-in controls, the lab 4

may wish to conduct a study to verify the effectiveness of the control before including it as a control measure. Evaluate residual risk. Once controls are assigned, the team then reviews the risks again and re-determines the risk level. Any risk not prevented or detected 100% of the time is considered residual risk. Further risk control procedures. If any risks still remain unacceptable, the team must decide how to bring that test into an acceptable level for continued operation. In cases where it is not possible to mitigate the risk to an acceptable level, the team should work with clinicians to determine if the benefits of performing the test outweigh the residual unacceptable risk. Determine a monitoring plan. Since the goal of the risk management plan is to maintain or improve the quality of the test, it s important to define a means to measure the quality to ensure that it meets the needs of the laboratory and the intended use of the test. One could argue that monitoring clinician complaints is an adequate means of monitoring quality, but this may only be true in cases where there is a close relationship between the lab and the clinical staff. In large institutions, it may be that the lab is made aware of only the most significant errors and some incorrect results are never reported back to the lab. It is better to have measures in place that are more likely to be logged or otherwise recorded. Write the quality control plan. A good plan does not need to be voluminous to be effective, but it should be complete enough that staff (and auditors) understand the factors that were considered in its creation. The lab may wish to include the process map and risk tables, but at a minimum they should be referenced and maintained separately. They will be invaluable later should the lab need to perform a failure investigation, evaluate a new test method, or look to improve the process of the current method. In contemplating the risk management process, a few of the steps may seem daunting to document, and certainly time consuming, but the process, if performed diligently, will provide insights that in the end will improve the quality of the test and improve patient safety. CONCLUSION Although risk management may appear menacing, laboratorians will soon realize that risk management simply encompasses management processes focused on the delivery of quality laboratory medicine, which have historically been performed on a daily basis. Having performed risk identification and risk assessment sequentially, laboratorians can move on to framing the ultimate outcome of the process risk mitigation which will result in the successful completion of an IQC Plan. Developing a risk management process does not ensure the complete elimination of risk. However, implementing a thorough risk management process that includes detailed risk identification and assessment will contribute to the reduction of risk and result in the continued delivery of quality patient care. REFERENCES 1. Clinical Laboratory Improvement Amendments, Subpart A, Section 493.1 Basis and scope. 2. Centers for Medicare and Medicaid Services. Clinical Laboratory Improvement Amendments, Interpretive Guidelines for Laboratories. 3. Hammerling JA. A review of medical errors in laboratory diagnostics and where we are today. LabMed. 2012;43(2): 41 44. 4. CLSI EP23-A Guidelines CONTACT INFORMATION SANAS website: www.sanas.co.za Main Switchboard No: +27 (0) 12 394-3760 General Fax No: +27 (0) 12 394-0526 Physical Addresses: the dti Campus, Block G, Ground Floor 77 Meintjies Street, Sunnyside, Pretoria SANAS Knowledge Transfer Centre: 121 Muckleneuck Street Nieuw Muckleneuk Pretoria 0002 Private Bag X23 Sunnyside Pretoria 0132 5