Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
|
|
|
- Myles Weaver
- 10 years ago
- Views:
Transcription
1 Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical Chemistry Associate Medical Director of Clinical Operations Vanderbilt University School of Medicine Nashville, TN
2 Overview Define Quality Control Define risk management and its application to clinical laboratory testing Understand common sources of error in the laboratory and mechanisms to reduce risk. Use the CLSI EP23 document as a resource for developing a quality control plan 2
3 Audience Poll Physician calls questioning a positive ethanol result of 135 mg/dl in a patient on the liver transplant list. A positive result would remove the patient as a candidate. This was first occurrence and patient is an adamant teetotaler! Repeat analysis of the specimen was negative (<10 mg/dl). Further investigation revealed that the technologist calibrated the assay backwards by switching positive and negative calibrators (Hi cal labeled #1 and Low cal labeled #2). Controls did not pick up, as they were switched as well. What should you do? A. Chastise the technologist B. Contact manufacturer to determine how the assay calibrated without error C. Seek better quality control D. Run for cover under risk management umbrella 3
4 Quality in Laboratory Results Clinicians rely on laboratory results for medical decision making. Quality is an essential component of laboratory testing. Errors in laboratory results can lead to missed diagnosis, medical mismanagement and increased cost of care. 4
5 Definitions Quality assurance in pathology and laboratory medicine is the practice of assessing performance in all steps of the laboratory testing cycle including pre analytic, analytic, and post analytic phases to promote excellent outcomes in medical care. Quality control is an integral component of quality assurance and is the aggregate of processes and techniques to detect, reduce, and correct deficiencies in an analytical process. Quality improvement is the practice of continuously assessing and adjusting performance using statistically and scientifically accepted procedures. College of American Pathologists 5
6 Historical Quality Control Quality control historically used to document stability of an analytical system (environment, operator, and analyzer). Born from the 1950 s industrial model of quality in analytical process Quality control is a stabilized surrogate sample analyzed like a patient sample containing known amount of measured analyte. If the analytical system can achieve the desired result using the QC sample, then the system is stable and quality patient results are being produced. 6
7 ISO QC Recommendations ISO The laboratory shall design internal control systems that verify the attainment of the intended quality of results. Special attention should be paid to elimination of mistakes in the process of handling samples, requests, examinations, reports, etc. Documentation should include quality control procedures based on manufacturer instructions for use. Internal Quality Control (internal to the laboratory) is defined as a set of procedures undertaken by laboratory staff for the continuous monitoring of operation and the results of measurements in order to decide whether results are reliable enough to be released. The regular analysis of QC materials can serve as an essential component of a laboratory s internal control system. 7
8 Cape Clinic Laboratory L-J Chart for XYZ Chemistry Analyzer Analyte: Glucose Control Material: Level 1 Units: mmol/l Lot #: Exp Date: 19/4/20XX assign 3.48 SD assign 0.05 From: June 1, 20XX Through: June 30, 20XX Target Value 3.39 TEa 6.90% SD C o n t r o l V a l u e SD + 1 SD X - 1 SD SD SD Run
9 Quality Control A stabilized surrogate sample of known concentration analyzed like a patient sample to determine assay recovery and result stability over time Advantages QC has target values, if assay recovers target, then everything is assumed stable (instrument, reagent, operator, sample) QC monitors the end product (result) of the entire test system Disadvantages Failure to analyze QC periodically or take action when QC fails Possible to release patient results before a problem is detected particularly with autoverification or continuous release automated tests With autoverification or continuous release of results When problem detected must go back and reanalyze patients since last good QC and possibly send out corrected results. Clinical action could be taken on erroneous results in the interim of correcting an unrecognized issue! 9
10 QC and Error Detection Systematic errors affect every test in a constant and predictable manner Can occur from one point forward or for a limited period of time QC samples do a good job at detecting systematic errors, like: Reagent deterioration or preparation Improper storage or shipment conditions Incorrect operator technique (dilution, pipette setting) Calibration errors wrong setpoint, factors Random errors can also be detected by increased QC variability, but QC does a poor job overall at detecting random errors! 10
11 Audience Poll Oncology clinic calls to question 6 patients from earlier this day. Patients showed low sodiums ( mmol/l) when repeated later in day, sodiums were higher by at least 10 mmol/l ( mmol/l) [reference range mmol/l]. Quality control showed a low trend earlier that day after morning calibration, but QC was acceptable. Medical director review noted QC ranges (120 +/ 3 mmol/l) were wider than CAP PT acceptability limits of +/ 3 mmol/l. This allowed for a 4 SD shift of nearly 12 mmol/l and still be acceptable! Analyzer had been in use for past 6 months without issue, and had adopted QC ranges from previous analyzer. What should you do as laboratory director? A. Resign B. Contact risk management C. Rerun all patient samples from period of low QC trend and correct any results outside CAP limits of +/ 4 mmol/l D. Revise the QC ranges 11
12 Quality Control Laboratory QC has limitations: Range for QC must be established by laboratory Setting ranges too wide can allow imprecision or too much method variability Incorrect establishment of control means can allow test result biases Biases in a single laboratory would go undetected without inter laboratory comparison (PT or External Quality Assessment Schemes EQAS) 12
13 New CMS QC Recommendations Analyze at least 2 levels of QC (different concentrations) with each batch of patient samples Minimum of 2 levels of QC each day of testing Hold patient results and troubleshoot test performance when QC fails to recover expected range of results. OR Develop an Individualized QC Plan (IQCP) based on risk management (US Centers for Medicaid and Medicare Services recommendations starting in 2014) 13
14 History CLSI EP23 document introduces industrial and ISO risk management principles to the clinical laboratory CMS adopted key risk management concepts to develop the IQCP option for quality control IQCP allows laboratories to develop a plan that optimizes the use of engineered, internal control processes on a device and the performance of external liquid QC 14
15 Why Should Labs Care About CLSI EP23? New generation of laboratory devices contain a variety of internal control processes. Labs are challenged with deciding appropriate liquid QC interval given QC engineered into newer devices. Package inserts provide little help from the manufacturer.
16 Package Inserts Current Good Laboratory Practice includes the daily use and documentation of either liquid controls or electronic internal controls to assure that the calibration of the diagnostic device is maintained within acceptable limits. (ffn test) Good Laboratory Practice suggest that external controls should be tested with each new lot or shipment of test materials, and as otherwise required by your laboratory s standard quality control procedures. (hcg test)
17 Package Inserts For quality control use the control materials listed. Other suitable control material can also be used in addition. The control intervals and limits should be adapted to each laboratory s individual requirements. (CO2 reagent) For best results, performance of reagent strips should be confirmed by testing known negative and positive samples, or controls whenever a new bottle is opened. Each laboratory should establish its own goals for adequate standards of performance and should question handling and testing procedures if these standards are not met. (Urine dipsticks)
18 Definitions Risk management is a way for labs to develop an individualized QC plan for their tests. A Quality Control Plan from CLSI EP23 a document that describes the practices, resources, and sequences of specified activities to control the quality of a particular measuring system or test process to ensure requirements for its intended purpose are met. An Individualized Quality Control Plan (IQCP) from CMS CLIA Interpretive Guidelines anew quality control option based on risk management for CLIA laboratories performing nonwaived testing. 18
19 Risk Management Definition Systematic application of management policies, procedures, and practices to the tasks of analyzing, evaluating, controlling, and monitoring risk (ISO 14971) 19
20 Risk Definition Risk the chance of suffering or encountering harm or loss (Webster s Dictionary and Thesaurus, 1993 Landoll, Ashland, Ohio) Risk can be estimated through a combination of the probability of occurrence of harm and the severity of that harm (ISO/IEC Guide 51) Risk essentially is the potential for an error to occur 20
21 Quality There is no perfect device, otherwise we would all be using it! Any device can and will fail under the right conditions Any discussion of risk must start with what can go wrong with a test (errors) Lab tests are not fool proof!
22 What Could Go Wrong?
23 Sources of Laboratory Error Test System: Reagent contamination, deterioration, lot variation Reaction fluctuations Inadequate sampling Improper or loss of calibration Electronic or mechanical failure Power supply Environment: Temperature and airflow Humidity Light intensity Altitude Operator: Improper specimen prep, handling Incorrect test interpretation Failure to follow test system instructions CLIA Interpretive Guidelines for Laboratories. Appendix C, Subpart K Standard: Control Procedures. Interpretive Guidelines (a) (c). Guidance/Legislation/CLIA/Interpretive_Guid elines_for_laboratories.html
24 Fishbone Diagram from CLSI EP23 Refer to Appendix A in CLSI EP18 for more comprehensive list of error sources
25 Managing Risk with a Control Process Once we identify the risks, we need to detect and prevent those errors from harming the patient. Control processes reduce risk by enhancing detection of errors or limiting harm if errors go undetected. Control processes can take many forms from liquid quality control to engineered checks within a device
26 Types of Quality Control On Board or Analyzer QC built in device controls or system checks (IL GEM, Radiometer ABL80, i stat) Internal QC laboratory analyzed surrogate sample controls. (Manufacturer controls performed on kit) External QC blind proficiency survey, samples sent a few times a year to grade an individual laboratory s performance against other labs (CAP or Wisconsin State) Other types of QC Control processes either engineered by manufacturer or enacted by laboratory to ensure result reliability (reagent barcoding to prevent use of expired reagent) 26
27 27
28 Quality Control No single quality control procedure can cover all devices, since devices may differ in design, technology, function, and intended use. QC practices developed over the years have provided labs with some degree of assurance that results are valid. Newer devices have built in electronic controls, and onboard chemical and biological controls. Quality control information from the manufacturer increases the user s understanding of device overall quality assurance requirements so that informed decisions can be made regarding suitable control procedures. ISO 15198:2004 Clinical laboratory medicine: In vitro diagnostic medical devices Validation of user quality control procedures by the manufacturer. 28
29 CLSI Project: EP23 Laboratory Quality Control Based on Risk Management. James H. Nichols, Ph.D., Chairholder EP23 describes good laboratory practice for developing a quality control plan based on manufacturer s information, applicable regulatory and accreditation requirements, and the individual healthcare and laboratory setting 29
30 EP23 Laboratory QC Based on Risk Management Input Information Medical Requirements for Test Results Regulatory and Accreditation Requirements Test System Information: Provided by the manufacturer Obtained by the Laboratory Information about Health Care and Test-Site Setting Process Risk Assessment Continuous Improvement Output Laboratory Director s QC Plan Post Implementation Monitoring CLSI EP23 Table 30
31 Individualized Quality Control Plan Quality Control Plan Risk Assessment Quality Assessment Individualized Quality Control Plan 31 CLIA
32 Benefits of IQCP to Lab Single use cartridge based methods use engineered controls in lieu of 2 levels QC/day POCT analyzers perform QC by lot of cartridge (using subset of devices) rather than every lot and each device. Core lab analyzers higher complexity tests helps labs identify weaknesses and appropriate actions to reduce risk of error (may need >2 levels QC/day) 32
33 Audience Poll As laboratory director, you are reviewing monthly temperature logs and notice a problem. The 20 C freezer logs showed temperatures out of range for past several days (ranging 16 C to 19 C for expected temps of < 20 C), with no corrective actions. Lead signed off the monthly log. When approached, the lead tech looks baffled and indicates that 16 C is colder than 20 C so no corrective action. You explain this problem to the manager who approaches the other techs involved with similar amazement. What do you do? A. Fire the techs involved and hire more competent staff B. Retrain the techs on negative numbers, put a number line as reminder on the freezer door, and change < 20 C to a range. C. Discard all the controls and calibrators stored in the freezer and call for maintenance D. Move the controls and calibrators to a working freezer and continue to use them. If controls are in, they are still good. 33
34 Summary Anything can and will go wrong without robust control plans! A quality control plan simply summarizes the risk for potential errors with a device and how the lab intends to address those risks. A quality control plan can be high level or very detailed depending on the device, the laboratory, and clinical application of the test result and may vary from one lab to next. Risk management and developing QC plans are generally accepted by the industry and required for FDA test approval Once implemented, the quality control plan is monitored for effectiveness and modified as needed to maintain risk to an acceptable level. 34
35 Where s the Quality Control? 35
36 Review Questions Which device would best benefit from an individualized quality control plan? A A glucose meter where manufacturer recommends daily QC B A high volume chemistry analyzer requiring QC 3 times daily C A molecular test with 500 reactions occurring on a single use cartridge D A batch analyzer for infrequent therapeutic drug testing Which errors would most likely be detected by traditional liquid QC? A A mistake where the assay pipette is set to the incorrect volume B Drug interference in a patient s sample C Hemolysisin a neonatal specimen D A clot in a dialysis patient sample Staff approach the medical laboratory director with a problem. During maintenance on Monday morning, a brown precipitate was found in the acid wash buffer which cleans all the cuvettes. It appears sometime over the weekend, staff poured an assay reagent into the wash buffer. What is the best course of action? A Fire the weekend staff B Call the manufacturer to determine how to flush the analyzer C Verify the validity of the weekend quality controls and reanalyze all the samples, correcting results as necessary D Contact risk management 36
Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren
Urinalysis Compliance Tools POCC Webinar January 19, 2011 Dr. Susan Selgren Learning Objectives Be able to review and improve upon a laboratory plan for compliance including: Competency Documentation Proficiency
Individualized Quality Control Plan
NOVEMBER 2014 CLIA Individualized Quality Control Plan WHAT IS AN IQCP? OVERVIEW Discovering, Deciphering and Designing an Individualized Quality Control Plan (IQCP) IQCP provides a framework for customizing
2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests
2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED Linda C. Bruno, M.A., MT(ASCP) Director, Microbiology and Molecular Labs ACL Laboratories, Rosemont, IL June 1, 2015 Disclosures - No disclosures 2 IQCP
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
Content Sheet 10-1: Overview of External Quality Assessment (EQA)
Content Sheet 10-1: Overview of External Quality Assessment (EQA) Role in quality management system Assessment is a critical aspect of laboratory quality management, and it can be conducted in several
Quality Requirements of a Point of Care Testing Service
Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal
XL-I 2011: Understanding and Explaining the CAP POCT Checklist: A Tool for Laboratorians to Use With POCT Staff
Case Study, Presentation John Tran is a medical technologist who has worked in an automated laboratory but is newly hired as a hospital point-of-care coordinator (POCC) for the Rapidly Growing Health System
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver When is a CLIA Certificate of Waiver Required? NOTE: Congress passed the Clinical Laboratory Improvement Amendments
Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians
Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Prepared by the Advisory Committee on Laboratory Medicine College of Physicians & Surgeons of Alberta Serving the Public by guiding
General. What is a laboratory test? What is a waived test?
General What is a laboratory test? What is a waived test? Laboratory tests are medical procedures that involve testing samples of blood, urine, or other tissues or substances from the body. Information
SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)
WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory
CHAPTER 13. Quality Control/Quality Assurance
CHAPTER 13 Quality Control/Quality Assurance Quality Control/Quality Assurance (QC/QA) can be defined as the set of planned and systematic activities focused on providing confidence that quality requirements
Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)
Topic: Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016) Click on the links below to be taken to a specific section of the FAQs. General
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD TABLE OF CONTENTS Introduction Applicability General Topics Proficiency Testing
College of American Pathologists Accreditation for Public Health Laboratories
College of American Pathologists Accreditation for Public Health Laboratories Romesh Gautom, PhD Director, Washington State Public Health Laboratory Garry McKee, PhD, MPH Director, Oklahoma State Public
Standards for Laboratory Accreditation
Standards for Laboratory Accreditation 2013 Edition cap.org Laboratory Accreditation Program Standards for Accreditation 2013 Edition Preamble Pathology is a medical specialty essential to patient care
TEST METHOD VERIFICATION AND VALIDATION
1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components
General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations
General Information Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations Our Background Foundation Laboratory is an advanced, accredited and certified clinical diagnostic
Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.
LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009
LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 Paul Bachner, MD, FCAP Professor & Chair Department of Pathology & Laboratory Medicine University of Kentucky
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
LabGuide 71. Incident Management: Developing a Plan
LabGuide 71 Incident Management: Developing a Plan INTRODUCTION In light of the Institute of Medicine studies focusing on reducing medical errors, COLA developed Accreditation criterion QA 20 to focus
Basic Lessons in Laboratory Quality Control
Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and
Mayo Central Laboratory for Clinical Trials A Division within Mayo Clinical Trial Services
Mayo Central Laboratory for Clinical Trials A Division within Mayo Clinical Trial Services LABORATORY TESTING SERVICES YOUR PARTNER IN LABORATORY MEDICINE The key note of progress in the 20th century is
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Director, Clinical Chemistry Molecular Diagnostics / Immunology Department of Laboratory Medicine Lahey Clinic Medical
2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan
2009 LAP Audioconference Series How to Prepare and Comply with Your Quality Management Objectives: As a result of participating in this session, you will be able to: Explain the reasons why the QM is important
Flow Cytometry. CLIA Compliance Manual*
Flow Cytometry CLIA Compliance Manual* Revised for the International Society for Advancement of Cytometry (ISAC) and Clinical Cytometry Society (CCS) By Michael Keeney ART, FIMLS Teri Oldaker B.A., CLS(NCA)QCYM
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency? http://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/clia_compbrochure_508.pdf
In vitro diagnostic reagent, calibrator and control material stability
In vitro diagnostic reagent, calibrator and control material stability Pauline Armstrong Randox Laboratories, Ltd. Crumlin The stability of in vitro medical diagnostic devices whether they are reagents,
Center for Clinical Standards and Quality/Survey & Certification Group
DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality/Survey
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Document and Record Control in a Hospital Clinical Laboratory
Document and Record Control in a Hospital Clinical Laboratory Linda A. Chambers, MD Transfusion Service Medical Director Riverside Methodist Hospital and Grant Medical Center OhioHealth, Columbus, Ohio,
Blood Glucose Monitoring: The Facts about Accuracy
Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined
Incident Management: Developing a Plan
LABFACTS 71 Incident Management: Developing a Plan INTRODUCTION In light of the Institute of Medicine studies focusing on reducing medical errors, COLA criterion #285 has been changed effective May 1,
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories () Edited by Richard Pang, PhD, FACB Senior Consultant Pro Q Consulting Services Hong Kong For Hong Kong
MEDICAL DEVICE GUIDANCE
May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained
Calibration Verification
Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in
PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL
Managing POCT: Trying to Control Testing in an Out-of-Control Environment. William Clarke, PhD, MBA, DABCC Johns Hopkins School of Medicine 3/24/11
Managing POCT: Trying to Control Testing in an Out-of-Control Environment William Clarke, PhD, MBA, DABCC Johns Hopkins School of Medicine 3/24/11 Learning Objectives Participants should be able to: Discuss
ADMINISTRATIVE MANUAL Policy and Procedure
ADMINISTRATIVE MANUAL Policy and Procedure TITLE: Point of Care Testing NUMBER: CH 30-111 (Laboratory Diagnostic Bedside Testing) Effective Date: January 2014 Page 1 of 6 Applies To: Holders of Administrative
2. Contact Person: Garo Mimaryan, MS., RAC 7 Technical Regulatory Affairs Specialist III 3. Phone Number: (914)-524-3270
SIEM EN S Abbreviated 510(k) Premnarket Notification IMMULITE 2000 PSA Calibration Verification Material 6. 510(k) Summary as Required by 21 CFR 807.92 K( - is-1 3 (v A. 5 10(k) Number: B. Purpose for
Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
CLIA & Individualized Quality Control Plan (IQCP)
& Individualized Quality Control Plan (IQCP) Judith Yost, M.A., M.T.(ASCP) Director Division of Laboratory Services Centers for Medicare & Medicaid Services 1 Objectives Provide Background & History of
Quality Assurance for the Clinical Laboratory
EP Evaluator Quality Assurance for the Clinical Laboratory Simplified process, powerful results While we (clinical labs) are but a small component of the total healthcare system, our data control a large
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K092353 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY B. Purpose for Submission: This is a new 510k application for a new indication for the MONOLISA Anti-HAV IgM EIA
Onboard Quality Controls for the CoaguChek XS System
Onboard Quality Controls for the CoaguChek XS System The new CoaguChek XS Anticoagulation Monitoring System incorporates a new onboard quality control system that eliminates the need for liquid controls
Transfusion Medicine: Checklists and Challenges
SEABB Annual Meeting 2014 Transfusion Medicine: Checklists and Challenges Denise Driscoll, MS, MT(ASCP)SBB Director, Accreditation and Regulatory Affairs March 19, 2014 cap.org v. # Transfusion Medicine:
Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO
Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban
Laboratory Director Responsibilities
Clinical Laboratory Improvement Amendments (CLIA) Laboratory Director Responsibilities What Are My Responsibilities As A Laboratory Director NOTE: Congress passed the Clinical Laboratory Improvement Amendments
Answers to Your Questions About Blood Bank Proficiency, Competency and QC Follow Up to April 28, 2016 Webinar
Answers to Your Questions About Blood Bank Proficiency, Competency and QC Follow Up to April 28, 2016 Webinar Anne Chenoweth, MBA, MT(ASCP)CM, CQA(ASQ) Director, Accreditation and Quality AABB Stacy Olea,
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
ILLINOIS DEPARTMENT OF CENTRAL MANAGEMENT SERVICES CLASS SPECIFICATION CLINICAL LABORATORY TECHNICIAN SERIES
ILLINOIS DEPARTMENT OF CENTRAL MANAGEMENT SERVICES CLASS SPECIFICATION CLINICAL LABORATORY TECHNICIAN SERIES CLASS TITLE POSITION CODE CLINICAL LABORATORY TECHNICIAN I 08215 CLINICAL LABORATORY TECHNICIAN
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved
A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved Overview and Objectives This module provides an introduction to
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM
For In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date
Is Your Lab s Competency Assessment a Competent Assessment?
Every number is a life. Is Your Lab s Competency Assessment a Competent Assessment? Presented by Jean Ball Inspections Team Lead College of American Pathologists www.cap.org Objectives Understand the reasons
Simoa HD-1 Analyzer Troubleshooting Guide. USER-104-28 17 Sep 2014 Software Version 1.3 Beta
Simoa HD-1 Analyzer Troubleshooting Guide USER-104-28 17 Sep 2014 Software Version 1.3 Beta Customer Support Customer support is available 8 AM to 8 PM, EDT. Phone: 1-877-786-8749 Email: [email protected]
GUIDELINES ON PREVENTING MEDICATION ERRORS IN PHARMACIES AND LONG-TERM CARE FACILITIES THROUGH REPORTING AND EVALUATION
GUIDELINES GUIDELINES ON PREVENTING MEDICATION ERRORS IN PHARMACIES AND LONG-TERM CARE FACILITIES THROUGH REPORTING AND EVALUATION Preamble The purpose of this document is to provide guidance for the pharmacist
Medical Laboratory Technician AAS Program
Medical Laboratory Technician AAS Program New Program Proposal State Submission Steve Buchholz, Dean of Accreditation and Advancement July 2015 TABLE OF CONTENTS Executive Summary... 2 Identification and
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
Everything you want to know about ISO 15189:2012 Medical Laboratories Requirements for Quality and Competence
Everything you want to know about ISO 15189:2012 Medical Laboratories Requirements for Quality and Competence Michael A Noble MD FRCPC CMPT and POLQM University of British Columbia Vancouver BC Canada
Using and Maintaining Documents and Records. Finding the information when you need it
Using and Maintaining Documents and Records Finding the information when you need it Problem Scenario You are the laboratory supervisor in a moderate-sized hospital. A physician calls you and says that
Content Sheet 5-1: Overview of Sample Management
Content Sheet 5-1: Overview of Management Role in quality management system management is a part of process control, one of the essentials of a quality management system. The quality of the work a laboratory
SAMPLE. Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision)
August 2012 Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision) This document provides guidance on how to verify comparability of quantitative
Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit
Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Catalog No: E0479r 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
SMF Awareness Seminar 2014
SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD
Automation is Here. Now How Do You Handle All the Data? Swedish Covenant Hospital Lab. Life IS Good
Automation is Here Now How Do You Handle All the Data? Susan Dawson, MBA, MT (ASCP) Swedish Covenant Hospital Chicago, IL Swedish Covenant Hospital Lab Automation in 2002 (Chem/IA) Autovalidation for Chemistry
Point of Care Testing and Informatics: How to Prepare for the POCT Explosion
Point of Care Testing and Informatics: How to Prepare for the POCT Explosion Pathology Informatics Summit 2014 David McClintock, MD May 14, 2014 Disclosures No conflicts of interest with any content presented
Vitamin B12. ADVIA Centaur System
Vitamin B12 Principle of the Test ADVIA Centaur System The ADVIA Centaur VB12 assay is a competitive immunoassay using direct chemiluminescent technology in which vitamin B 12 from the patient sample competes
State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist
This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to
Laboratory Quality Metrics
Laboratory Quality Metrics Laila O. Abdel-Wareth. MBBCh, FCAP, FRCPC, FRCPath, EMHCA Chief Medical Officer- Pathology & Laboratory Medicine Services SEHA Chair of Pathology & Laboratory Medicine Institute
FMEA and FTA Analysis
FMEA and FTA Analysis Why it is Coming to Your Hospital and Your Laboratory Tina A. Krenc Director, R&D Phase Systems Abbott Laboratories 1 Agenda Background on requirements for risk management Tools to
ArtisanLink Staining System is an automated special stains slide
Technical Tips Tips on using the ArtisanLink Special Staining System Jamie Nowacek, BS, HT(ASCP) CM, QIHC, PMP Dako North America, Inc. Carpinteria, CA, USA ArtisanLink Staining System is an automated
Validation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
Preparation "Clinical Laboratory Technologist and Technician Overview"
Clinical Laboratory Technologist and Technician Overview The Field - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Clinical laboratory
accreditation will it still be needed or will other schemes show up?
By Leif Madsen, DELTA, President of Eurolab Denmark Member of ISO WG 44 We re facing the revision of ISO 17025, that may entail significant changes in how the laboratories are operated. Other developments
116 The Future Role of Laboratory Inspections and Accreditation. Sharon Ehrmeyer PhD, MT(ASCP)
116 The Future Role of Laboratory Inspections and Accreditation Sharon Ehrmeyer PhD, MT(ASCP) 2011 Annual Meeting Las Vegas, NV AMERICAN SOCIETY FOR CLINICAL PATHOLOGY 33 W. Monroe, Ste. 1600 Chicago,
Quality Assurance (QA) & Quality Control (QC)
Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580
DAIDS Guidelines for Good Clinical Laboratory Practice Standards
DAIDS Guidelines for Public Health Services National Institutes of Health Bethesda, Maryland 20892 www.nih.gov Dear Site Principal Investigator and Laboratory Director: All National Institutes of Health,
Bio-Rad Laboratories. QC data management solutions. Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions
Bio-Rad Laboratories QC data management solutions QC Data Management Solutions Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions Bio-Rad Laboratories QC data management
MEDICAL DIAGNOSTIC LABORATORIES, L.L.C. 2439 Kuser Road * Hamilton, NJ 08690 Toll Free (877) 269-0090 * Fax (609) 570-1050 www.mdlab.
FAQ's For Corporate Q1. What are MDL's hours of operation? A1. You may reach us Monday through Friday from 8:00 am to 12:00 midnight EST and Saturday from 9:00 am to 5:00 pm EST. Q2. How long has MDL been
Biochemistry Validation Form
Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
A Stability Program for the Distribution of Drug Products
A Stability Program for the Distribution of Drug Products Teresa I. Lucas*, Rafik H. Bishara, and Robert H. Seevers Drug products must be transported in a manner that ensures products will be maintained
James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract
Clinical Chemistry / ANALYTIC QUALITY TODAY The Quality of Laboratory Testing Today An Assessment of σ Metrics for Analytic Quality Using Performance Data From Proficiency Testing Surveys and the CLIA
What Does Having a FDA Cleared Pregnancy Test Mean?
What Does Having a FDA Cleared Pregnancy Test Mean? Clinical Trials Transformation Initiative (CTTI) July 15-16, 2013 Denise N. Johnson-Lyles, Ph.D. Toxicology Branch Chief Division of Chemistry and Toxicology
Laboratory medicine as one of the pharmacists competencies
2014 EAFP Annual Conference Laboratory medicine as one of the pharmacists competencies Borut Božič Examinations of the body fluids in order to recognize the state of health of the individuals Greco-Roman
PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.
Department Of Pathology POC.514.01- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 1 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.01 LBH 1 OF 7 Printed copies are
ST. VINCENT'S. MEDICAL CENTER St. Vincent's Healthcare
ST. VINCENT'S MEDICAL CENTER St. Vincent's Healthcare Medical Technology St. Vincent s Schools of Medical Science Throughout Northeast Florida and Southern Georgia, St. Vincent s HealthCare is well known
4/24/2015. Urgent, STAT, Super STAT, ASAP! Achieving timely lab testing for the Emergency Department. Learning Objectives.
Urgent, STAT, Super STAT, ASAP! Achieving timely lab testing for the Emergency Department Nichole Korpi-Steiner, PhD, DABCC, FACB University of North Carolina Chapel Hill, NC Learning Objectives Describe
Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit
Mouse Keyhole Limpet Hemocyanin antibody(igm) ELISA Kit Catalog No. MBS702810 (96 tests) This immunoassay kit allows for the in vitro semi-quantitative determination of mouse KLH(IgM)antibody concentrations
CHAPTER 7 QUALITY ASSESSMENT
CHAPTER 7 QUALITY ASSESSMENT Chapter 7 QUALITY ASSESSMENT 7.1 OVERVIEW OF CBB-SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL PROGRAM An extensive, site-specific Quality Assurance/Quality Control (QA/QC) program
IMPACTS OF BLUE DROP CERTIFICATION ON WATER QUALITY TESTING LABORATORIES
IMPACTS OF BLUE DROP CERTIFICATION ON WATER QUALITY TESTING LABORATORIES Speaker(s) / Author(s): MJ Dlamini Rand Water Analytical Services PO Box 3526, Vereeniging, South Africa e-mail: [email protected]