Individualized Quality Control Plan
|
|
|
- Hilary Blair
- 9 years ago
- Views:
Transcription
1 NOVEMBER 2014 CLIA Individualized Quality Control Plan WHAT IS AN IQCP? OVERVIEW Discovering, Deciphering and Designing an Individualized Quality Control Plan (IQCP) IQCP provides a framework for customizing a quality control (QC) program for your test systems and your laboratory s unique environment. By performing the steps in an IQCP, you will examine the potential sources of error in your pre-analytic, analytic and post analytic phases of testing, as well as establish the appropriate QC and quality practices which reduce the likelihood of errors occurring in your laboratory. After you complete this process, it is possible that you may determine that the amount of QC you have been doing all along is sufficient to achieve CLIA compliance. However, you could discover potential sources of error that you had not previously considered, and may need to implement additional QC activities. In either case you will have created a comprehensive QC program, which reflects your laboratory s unique operation, and the documentation which supports the rationale for your QC practices to ensure high quality testing. What are the parts of an IQCP? An IQCP requires: Risk Assessment (RA) Quality Control Plan (QCP) Quality Assessment (QA) SECTION 1: IQCP RISK ASSESSMENT WHAT IS RISK ASSESSMENT (RA?) Risk Assessment is the means of identifying and evaluating potential problems or errors that may occur in your testing process. The testing process begins with the specimen collection (preanalytic) and continues through the analysis of the specimen (analytic) until the final test result is reported (postanalytic). 1
2 WHAT ARE THE COMPONENTS OF A RISK ASSESSMENT (RA)? There are five components you must evaluate in your Risk Assessment (RA): Specimen Test System Reagents Environment Testing Personnel Consider the potential errors that might be attributable to these five components in your testing process. NOTE: You must evaluate these five components when performing your RA. However, you may identify additional risk factors to consider and are not limited to just these five components. WHERE CAN I FIND RESOURCES FOR INFORMATION FOR CONDUCTING A RA? The following list contains possible sources of information for conducting a risk assessment: Regulatory requirements Manufacturer s package insert (including intended use, limitations, environmental requirements, QC frequency, specimen requirements, reagent storage, maintenance, calibration, interfering substances, etc.) Manufacturer s operator manual Troubleshooting guide Manufacturer s alerts and bulletins Verification or establishment of performance specifications Testing personnel qualifications, training, and competency records QC data Proficiency testing data QA information, including corrective action Scientific publications Other information as appropriate 2
3 WHAT ARE SOME POTENTIAL SOURCES OF ERROR FOR THE FIVE RA COMPONENTS? To discover those unique features of your program, and decipher what their impact is in your laboratory, below is a general list of potential sources of errors. As you review this chart, ask yourself: What are the chances of this error happening? Is there a step in the process that helps reduce the chance of an error? If not, how do I minimize or reduce the likelihood that this error could occur anywhere in the testing process? Potential Sources of Error for the five Risk Assessment Components SPECIMEN Patient preparation Specimen collection Specimen labeling TEST SYSTEM Inadequate sampling Clot detection capabilities Capabilities for detection of interfering substances (e.g., hemolysis, lipemia, icterus, turbidity) Calibration associated issues Mechanical/electronic failure of test system Specimen storage, preservation, and stability Specimen transportation Optics Pipettes or pipettors Barcode readers Failure of system controls and function checks Built-in procedural and electronic controls (internal controls) Specimen processing Specimen acceptability and rejection Specimen referral External or internal liquid quality control (assayed vs. unassayed) Temperature monitors and controllers Software/Hardware Transmission of data to LIS Result reporting REAGENT Shipping/Receiving Storage condition requirements Expiration Date (may differ based on storage requirements) Preparation ENVIRONMENT Temperature Airflow/ventilation Light intensity Noise and vibration TESTING PERSONNEL Training Competency Humidity Altitude Dust Water Education and experience Staffing Utilities (Electrical failure/power supply variance or surge) Space Footnote: The above list of risk factors is not all-inclusive. You may find more factors in your laboratory needing assessment. 3
4 IS THERE A SPECIFIC FORMAT SUCH AS A PROCESS MAP, DIAGRAMS, ETC. THAT MY LABORATORY SHOULD USE WHEN DEVELOPING AN IQCP? CLIA does not require the use of any specific tools or format in the development of an IQCP. CLIA is not prescriptive as to what tools or resources are to be used for a laboratory to meet the regulatory requirements of IQCP. It is the responsibility of the laboratory director to determine how to meet these requirements, i.e. acceptable tools, resources. The new IQCP option gives laboratories the flexibility to determine the appropriate tool(s) for their test system and environment. WILL I NEED TO PERFORM ALL NEW STUDIES TO GATHER DATA/INFORMATION FOR THE RA AND THE DEVELOPMENT OF THE QCP FOR EXISTING TESTS IN MY LABORATORY? Much of the data/information needed by your laboratory to perform the RA for each test will most likely be data that has been accumulated in the process of routine operations in the laboratory while meeting CLIA regulations and implementing quality systems. For example, verification of manufacturer s performance specifications, QC performance, maintenance records, and documents of corrective actions taken can all be used in your RA. You must have data that demonstrates the stability of the test system and supports the QC type and frequency in the QCP. I HAVE SEVERAL DEVICES IN MY LABORATORY THAT ARE IDENTICAL. THEY ARE THE SAME MAKE AND MODEL NUMBER AND ARE LOCATED AT DIFFERENT SITES, BUT ARE UNDER THE SAME CLIA CERTIFICATE. HOW DOES THIS AFFECT MY IQCP FOR ALL OF THESE DEVICES? The RA will be performed for the test system. Any variables of the five components should be taken into consideration when conducting the RA, e.g., education and experience of the testing personnel, environment, and use of the test system. The QCP applies to the individual device. It is your laboratory s choice whether to have one QCP which includes all of the sites or individual QCPs for each site. Each device must be monitored in some way, as well as each location. Data must indicate that the QC established for each device is acceptable. I HAVE COMPLETED THE RA. WHAT DO I DO WITH THE RESULTS OF MY RA? Now that you have completed the RA, determine whether these identified risks need to be monitored or controlled regularly in the testing process or if they may already be addressed by the manufacturer in the design of the test system. This information will help you in developing your QCP. 4
5 SECTION 2: IQCP QUALITY CONTROL PLAN WHAT SHOULD I CONSIDER WHEN CREATING A QUALITY CONTROL PLAN (QCP)? A strong, well-documented QCP will establish control procedures that reduce the likelihood of providing an inaccurate patient test result. Your QCP must at least include the number, type and frequency of testing and criteria for acceptable result(s) of the quality control(s). Your data must support the rationale for the number, type and frequency of testing. It s possible that you may find your customized QCP will be less than the CLIA control requirements, but more than the manufacturer s instructions for controls. However, at a minimum, your QCP must not be less stringent than the manufacturer s instructions for testing QC. The QCP may also describe the use of electronic controls, procedural controls, training and competency assessment and all other QC activities. DOES MY LABORATORY DIRECTOR HAVE TO REVIEW, APPROVE AND SIGN THE QCP? Yes, it is the responsibility of the laboratory director to review, sign and date the QCP before patient testing begins and results are reported. The laboratory director may assign, in writing, specific duties for the IQCP to qualified laboratory personnel, but is still responsible overall for the entire testing process. CAN ONE QCP ADDRESS MULTIPLE TEST PROCEDURES IF ALL TESTS ARE PERFORMED IN THE SAME MANNER? For those cases in which similar methodologies are used, it is likely that some potential sources of error will be similar, while others will be unique to each test. Therefore, you must perform the RA and develop a QCP for each individual test system. The QCPs for these test systems may, or may not, be included in one QCP. The QCP must at least include the number, type, frequency of testing and criteria for acceptable result(s) of the quality controls(s). CAN MY LABORATORY CHOOSE TO USE IQCP FOR SOME ANALYTES ON A PLATFORM AND NOT OTHERS? Yes, you can choose to use IQCP for some or all of your test systems. However, you should indicate which test(s) use IQCP in your QCP. 5
6 SECTION 3: IQCP QUALITY ASSESSMENT HOW DO I IDENTIFY AND SELECT QUALITY MONITORS FOR A QUALITY ASSESSMENT (QA)? Quality control records are a great place to start; but, you are not limited to just these records. Other documents may include: Proficiency testing records (test score, testing failures, trends), Patient results review, Specimen rejection logs, Turnaround time reports, Records of preventive measures, Corrective actions and follow-up and, Personnel competency records. You may also create specific monitors, unique to the test system, that answer how often or how many times (within a time period for example) did a particular error occur? You are not limited to the number of monitors used to verify the continued performance of a testing process. Good monitors ensure that you continue to identify and minimize problems and that accurate test results are being reported. Keep in mind that your QA monitors may indicate a need to reevaluate the effectiveness of your IQCP. WHERE CAN I FIND ADDITIONAL GUIDANCE ON IQCP? For the most current information and additional brochures on IQCP, please visit the CLIA website at 6
Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)
Topic: Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016) Click on the links below to be taken to a specific section of the FAQs. General
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED
IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED Linda C. Bruno, M.A., MT(ASCP) Director, Microbiology and Molecular Labs ACL Laboratories, Rosemont, IL June 1, 2015 Disclosures - No disclosures 2 IQCP
Laboratory Director Responsibilities
Clinical Laboratory Improvement Amendments (CLIA) Laboratory Director Responsibilities What Are My Responsibilities As A Laboratory Director NOTE: Congress passed the Clinical Laboratory Improvement Amendments
2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests
2010 Laboratory Accreditation Program Audioconference Accreditation Requirements for Waived Vs. Non-waived Tests Objectives: After participating in this audioconference you will be able to: Define the
TEST METHOD VERIFICATION AND VALIDATION
1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical
Examples: collecting the right sample on the right patient; labeling the sample accurately;
Quality Assessment: Basics & EMR is called by several names: and are just two of them. Quality Assessment (QA) is an on-going program designed to monitor and evaluate the quality or accuracy of all laboratory
Content Sheet 3-1: Equipment Management Overview
Content Sheet 3-1: Equipment Management Overview Role in quality management system Equipment management is one of the essential elements of a quality management system. Proper management of the equipment
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver
Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver When is a CLIA Certificate of Waiver Required? NOTE: Congress passed the Clinical Laboratory Improvement Amendments
Flow Cytometry. CLIA Compliance Manual*
Flow Cytometry CLIA Compliance Manual* Revised for the International Society for Advancement of Cytometry (ISAC) and Clinical Cytometry Society (CCS) By Michael Keeney ART, FIMLS Teri Oldaker B.A., CLS(NCA)QCYM
CLIA & Individualized Quality Control Plan (IQCP)
& Individualized Quality Control Plan (IQCP) Judith Yost, M.A., M.T.(ASCP) Director Division of Laboratory Services Centers for Medicare & Medicaid Services 1 Objectives Provide Background & History of
General. What is a laboratory test? What is a waived test?
General What is a laboratory test? What is a waived test? Laboratory tests are medical procedures that involve testing samples of blood, urine, or other tissues or substances from the body. Information
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Checklist. Standard for Medical Laboratory
Checklist Standard for Medical Laboratory Name of hospital..name of Laboratory..... Name. Position / Title...... DD/MM/YY.Revision... 1. Organization and Management 1. Laboratory shall have the organizational
Calibration Verification
Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in
Quality Requirements of a Point of Care Testing Service
Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal
Standards for Laboratory Accreditation
Standards for Laboratory Accreditation 2013 Edition cap.org Laboratory Accreditation Program Standards for Accreditation 2013 Edition Preamble Pathology is a medical specialty essential to patient care
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
Center for Clinical Standards and Quality/Survey & Certification Group
DEPARTMENT OF HEALTH & HUMAN SERVICES Centers for Medicare & Medicaid Services 7500 Security Boulevard, Mail Stop C2-21-16 Baltimore, Maryland 21244-1850 Center for Clinical Standards and Quality/Survey
Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition
JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR CliniCAl laboratories 2nd Edition Effective 1 April 2010 International Patient Safety Goals (IPSG) Goals The following is a list of all goals.
Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?
Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency? http://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/clia_compbrochure_508.pdf
Using and Maintaining Documents and Records. Finding the information when you need it
Using and Maintaining Documents and Records Finding the information when you need it Problem Scenario You are the laboratory supervisor in a moderate-sized hospital. A physician calls you and says that
Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren
Urinalysis Compliance Tools POCC Webinar January 19, 2011 Dr. Susan Selgren Learning Objectives Be able to review and improve upon a laboratory plan for compliance including: Competency Documentation Proficiency
CHAPTER 7 QUALITY ASSESSMENT
CHAPTER 7 QUALITY ASSESSMENT Chapter 7 QUALITY ASSESSMENT 7.1 OVERVIEW OF CBB-SPECIFIC QUALITY ASSURANCE/QUALITY CONTROL PROGRAM An extensive, site-specific Quality Assurance/Quality Control (QA/QC) program
Canadian Environmental Technology Verification Program Applicant Information Package CONTENTS
Canadian Environmental Technology Verification Program CONTENTS Contents... 1 Overview... 2 Performance Verification... 2 Application Information... 2 How much does verification cost?... 2 What is the
Medical Cannabis Laboratory Approval Program
Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,
LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009
LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009 Paul Bachner, MD, FCAP Professor & Chair Department of Pathology & Laboratory Medicine University of Kentucky
Automation is Here. Now How Do You Handle All the Data? Swedish Covenant Hospital Lab. Life IS Good
Automation is Here Now How Do You Handle All the Data? Susan Dawson, MBA, MT (ASCP) Swedish Covenant Hospital Chicago, IL Swedish Covenant Hospital Lab Automation in 2002 (Chem/IA) Autovalidation for Chemistry
MEDICAL DEVICE GUIDANCE
May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained
Going Beyond Laboratory Automation: Do Less Accomplish More. Swedish Covenant Hospital
Going Beyond Laboratory Automation: Do Less Accomplish More Susan Dawson, MBA, MT(ASCP) Swedish Covenant Hospital Chicago, IL Swedish Covenant Hospital Open Heart Oncology Birthing Unit Community Outreach
CHAPTER ONE TABLE OF CONTENTS 1.0 INTRODUCTION... 1
CHAPTER ONE TABLE OF CONTENTS Section Page 1.0 INTRODUCTION... 1 2.0 QA PROJECT PLAN... 1 2.1 DATA QUALITY OBJECTIVES... 2 2.2 PROJECT OBJECTIVES... 2 2.3 SAMPLE COLLECTION... 3 2.4 ANALYSIS AND TESTING...
Is Your Lab s Competency Assessment a Competent Assessment?
Every number is a life. Is Your Lab s Competency Assessment a Competent Assessment? Presented by Jean Ball Inspections Team Lead College of American Pathologists www.cap.org Objectives Understand the reasons
COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS
COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS Primer CONTENTS INTRODUCTION...3 QUALITY AND COMPLIANCE IN QUALITY CONTROL LABORATORIES...5 Compliance
APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN
APPENDIX 7-B SUGGESTED OUTLINE OF A QUALITY ASSURANCE PROJECT PLAN This outline is recommended for use by UST consultants/contractors in preparing a generic Quality Assurance Project Plan (QAPP) for use
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD
Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD Director, Clinical Chemistry Molecular Diagnostics / Immunology Department of Laboratory Medicine Lahey Clinic Medical
PROCEDURES FOR HANDLING OOS RESULTS
SOP # A-195--0100 STANDARD OPERATING PROCEDURES Page: 1 of 6. PROCEDURES FOR HANDLING OOS RESULTS 1. PURPOSE The purpose of this Standard Operation Procedure is to establish a procedure for the routine
Dedicated and calibrated pipettes can reduce variability. The Most Expensive Test is a Retest By Ruth Noe, UK and Ireland Sales Manager
LALreview LAL REVIEW: PUBLISHED BY LONZA 2008 Issue 3 EDITOR S COLUMN What is the most expensive endotoxin test that you run in your QC lab? If I were to guess, I would say the retest. How often do you
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
How To Get Blood Test Records From A Blood Alcohol Test
CAUSE NO. THE STATE OF TEXAS IN THE COUNTY CRIMINAL V. COURT AT LAW NO. HARRIS COUNTY, TEXAS STANDING DISCOVERY ORDER ON COPYING AND PRODUCTION OF BLOOD TESTING RECORDS THE COURT ORDERS the District Attorney
Mayo Central Laboratory for Clinical Trials A Division within Mayo Clinical Trial Services
Mayo Central Laboratory for Clinical Trials A Division within Mayo Clinical Trial Services LABORATORY TESTING SERVICES YOUR PARTNER IN LABORATORY MEDICINE The key note of progress in the 20th century is
STILLWATER BILLINGS CLINIC. Job Description Medical Technologist
STILLWATER BILLINGS CLINIC Job Description Medical Technologist General Description: The Medical Technologist is responsible for performing high complexity clinical laboratory testing in one or more sections
Audits must be conducted with due concern for employee safety and environmental protection.
Standard Operating Procedure Title: GMP Audit Procedure Related Documents TEM-080 Internal Audit Report Template TEM-120 Vendor Audit Report Template Form-385 Vendor Audit Questionnaire Form-445 EHS Workplace
2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan
2009 LAP Audioconference Series How to Prepare and Comply with Your Quality Management Objectives: As a result of participating in this session, you will be able to: Explain the reasons why the QM is important
Quality Assurance (QA) & Quality Control (QC)
Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580
Bio-Rad Laboratories. QC data management solutions. Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions
Bio-Rad Laboratories QC data management solutions QC Data Management Solutions Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions Bio-Rad Laboratories QC data management
Document and Record Control in a Hospital Clinical Laboratory
Document and Record Control in a Hospital Clinical Laboratory Linda A. Chambers, MD Transfusion Service Medical Director Riverside Methodist Hospital and Grant Medical Center OhioHealth, Columbus, Ohio,
National Survey of Pre- and Post- Analytical Performance Measures Used in Newborn Screening Programs
National Survey of Pre- and Post- Analytical Performance Measures Used in Newborn Screening Programs Simran Tiwana, MBA* Susan Tanksley, PhD* Brad Therrell, PhD Donna Williams, BS* Mirsa Douglass, MBA*
Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3
Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3 Prepared by Date Adopted Supersedes Procedure # Estelle Piwowar-Manning Quality Assessment vers 2 Sarah Dawson Paul Richardson
Biochemistry Validation Form
Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August
PROPOSED UPDATED TEXT FOR WHO GOOD MANUFACTURING PRACTICES FOR PHARMACEUTICAL PRODUCTS: MAIN PRINCIPLES (JANUARY 2013)
January 2013 RESTRICTED PROPOSED UPDATED TEXT FOR WHO GOOD MANUFACTURING PRACTICES FOR PHARMACEUTICAL PRODUCTS: MAIN PRINCIPLES (JANUARY 2013) DRAFT FOR COMMENTS Please address any comments on this proposal
XL-I 2011: Understanding and Explaining the CAP POCT Checklist: A Tool for Laboratorians to Use With POCT Staff
Case Study, Presentation John Tran is a medical technologist who has worked in an automated laboratory but is newly hired as a hospital point-of-care coordinator (POCC) for the Rapidly Growing Health System
Unique Standards and Documentation Required for Accredited CLT/MLT Programs
Unique Standards and Documentation Required for Accredited CLT/MLT Programs Section III (CLT/MLT) Page 1 UNIQUE STANDARDS AND THE REQUIRED DOCUMENTATION Objectives Clinical Laboratory Technician/Medical
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
Chapter 2: Quality Assurance and Legal Issues
Objectives Chapter 2: Quality Assurance and Legal Issues 1. Define the key terms and abbreviations listed at the beginning of this chapter. 2. Identify national organizations, agencies, and regulations
Electronic Raw Data and the Use of Electronic Laboratory Notebooks
7th WRIB- M-2 Workshop Electronic Raw Data and the Use of Electronic Laboratory Notebooks Ron Shoup Outline for Today Raw Data- defining it, creating it, and keeping it for posterity Hybrid Data- paper,
NC SBI QUALITY ASSURANCE PROGRAM
NC SBI QUALITY ASSURANCE PROGRAM for the SBI Reviewed by: Deputy Assistant Director Bill Weis Date: Approved by: Assistant Director Jerry Richardson Date: Originating Unit: SBI Effective Date: July 25,
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
INNOVATION TO MEET YOUR NEEDS
INNOVATION TO MEET YOUR NEEDS 100% High Throughput The throughput is optimized splitting the pipetting phases of Reagents and Samples into different cycles. The innovative technical solutions adopted by
Lean Strategies Used to Optimize Automation. Why Lean Six Sigma? Laboratory Goals. Decreased TAT. Accurate Results. LEAN Goals.
Lean Strategies Used to Optimize Automation Linda Stubbs, ASQ CSSBB Senior Workflow Specialist Why LEAN Six Sigma? Why Lean Six Sigma? Laboratory Goals Decreased TAT Accurate Results LEAN Goals Increase
Help Desk Review of Quality Assurance Project Plans
Help Desk Review of Quality ssurance Project Plans May 2, 2011 pproved by Beverly H. van Buuren, Surface Water mbient Monitoring Program Quality ssurance Officer, on May 2, 2011 1. Purpose This document
DAIDS Guidelines for Good Clinical Laboratory Practice Standards
DAIDS Guidelines for Public Health Services National Institutes of Health Bethesda, Maryland 20892 www.nih.gov Dear Site Principal Investigator and Laboratory Director: All National Institutes of Health,
Guidance for Industry
Guidance for Industry Investigating Out-of-Specification (OOS) Test Results for Pharmaceutical Production U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation
Cartel Electronics. AS 9100 Quality Systems Manual
Cartel Electronics AS 9100 Quality Systems Manual 1900 C Petra Lane Placentia, California 92870 Introduction Cartel Electronics, as a global supplier to the aviation, space, and space industries, has developed
Basic Lessons in Laboratory Quality Control
Bio-Rad Laboratories QC education Basic Lessons in Laboratory Quality Control QC Workbook Basic Lessons in Laboratory Quality Control Written by Greg Cooper, CLS, MHA Manager of Clinical Standards and
IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.
IgM ELISA For the quantitative determination of IgM in human serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calibrator
AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189
Indian Journal of Clinical Biochemistry, 2010 / 25 (1) 92-98 LABORATORY DOCUMENTATION AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189 T F Hartley
Sutter Health Support Services Shared Laboratory Position Description. Incumbent: Laboratory Date: October 23, 2006 Written By: Michele Leonard
Sutter Health Support Services Shared Laboratory Position Description Position Title: Clinical Laboratory Scientist, Microbiologist Incumbent: Entity: SHSS Shared Laboratory Reports To: Director of Operations,
MEDICAL LABORATORY TECHNICIAN COMPETENCY PROFILE
Description of Work: Positions in this banded class support or perform laboratory tests that are used in the diagnosis and treatment of patients and animals. Duties performed include: receiving or procuring
Haulsey Engineering, Inc. Quality Management System (QMS) Table of Contents
Haulsey Engineering, Inc. Quality Management System (QMS) Table of Contents 1.0 Introduction 1.1 Quality Management Policy and Practices 2.0 Quality System Components 2.1 Quality Management Plans 2.2 Quality
1 For more information T: +1.847.664.1499 / E: [email protected] / W: ul.com/dap
DATA ACCEPTANCE PROGRAM UL Client Test Data Program Investigations of products by UL typically involve the testing of product samples. While UL maintains extensive facilities for these purposes, customers
Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians
Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Prepared by the Advisory Committee on Laboratory Medicine College of Physicians & Surgeons of Alberta Serving the Public by guiding
1.0 INTRODUCTION 1.1 Overview
Guidance document for the writing of standard operating procedures (Taken from United States Environmental Protection Agency Guidance for Preparing Standard Operating Procedures (SOPs) EPA QA/G- 6 [EPA/600/B-
General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations
General Information Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations Our Background Foundation Laboratory is an advanced, accredited and certified clinical diagnostic
Meaningful Competency Assessment. Kathy Nucifora, MPH, MT(ASCP) COLA 1
Meaningful Competency Assessment Kathy Nucifora, MPH, MT(ASCP) COLA 1 Objectives Differentiate between Competency Assessment and the traditional performance evaluation Outline the six CLIA-required components
References to Computer and Software in FDA s CGTPs (Final Rule and Draft Guidance) for HCT/Ps Preamble to the CGTP Final Rule
Preamble to the CGTP Final Rule Pages 68626-68627 Computers Proposed 1271.160(e) would require establishments to validate computer software used as part of manufacturing or tracking or for maintaining
THE KOCOUR COMPANY 4800 S. St. Louis Avenue, Chicago, IL 60632
THE KOCOUR COMPANY 4800 S. St. Louis Avenue, Chicago, IL 60632 Metal Finishing Instrumentation, Equipment and Supplies ISO 9001: 2008 QUALITY MANAGEMENT MANUAL ISSUED: 03/01/02(REV. 0) REVISION NO: 4 DATE:
Pharmaceutical Wholesaler Site Inspection Checklist
Pharmaceutical Wholesaler Site Inspection Checklist Date and time of inspection: Legal business name and plant name: Location (address): Phone number: Fax number: Email address: Web site URL: Contact Person:
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC August 2013 Contents 1 Introduction...3 2 Scope of the directive...3 2.1 What is an in vitro diagnostic medical device?... 3 2.2 Specimen
Inspection, Measuring and Test Equipment (IMTE) Calibration
Inspection, Measuring and Test Equipment (IMTE) Calibration Purpose and Scope 2 Responsibilities 2 About IMTE 3 Master IMTE Log 4 Identification of IMTE Calibration Status 5 Handling and Storage of IMTE
Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit
ab108650 Immunoglobulin E (IgE) Human ELISA Kit Instructions for Use For the quantitative measurement of Immunoglobulin E (IgE) concentrations in Human serum. This product is for research use only and
ISO 13485:201x What is in the new standard?
ISO 13485:201x What is in the new standard? Eric Finegan, Quality Mgr, BTE Technologies, Inc. 2015-09-10 1 Presentation Slides This slide deck is the presentation performed on 2015-09-10. A more detailed
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD
CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD TABLE OF CONTENTS Introduction Applicability General Topics Proficiency Testing
Quality Management Systems Manual
Washington Division Quality Management Systems Manual ISO 13485:2003 DOCUMENT NUMBER 01-QM-MED REVISION LEVEL 01 PAGE 1 of 40 This manual describes the quality management systems structure at GM Nameplate
CLSI TRAINING COURSES CATALOG
CLSI TRAINING COURSES CATALOG A Laboratory Development Resource CLINICAL AND LABORATORY STANDARDS INSTITUTE TRAINING COURSES CATALOG The Clinical and Laboratory Standards Institute (CLSI) is known for
US EPA REGION III QUALITY MANAGEMENT PLAN REVIEW CHECKLIST
US EPA REGION III Organization: EPA Organization: EPA Program: Contact: EPA Contact: EPA Contract/Grant/IAG Number: Phone Number: Phone Number: Reviewer: Date Reviewed: Reviewer Organization: Check New
LABORATORY PERSONNEL QUALIFICATION APPRAISAL and APPLICATION FOR LICENSURE
LABORATORY PERSONNEL QUALIFICATION APPRAISAL and APPLICATION FOR LICENSURE WEST VIRGINIA DEPARTMENT OF HEALTH AND HUMAN RESOURCES OFFICE OF LABORATORY SERVICES 167 11th Avenue South Charleston, West Virginia
Everything you want to know about ISO 15189:2012 Medical Laboratories Requirements for Quality and Competence
Everything you want to know about ISO 15189:2012 Medical Laboratories Requirements for Quality and Competence Michael A Noble MD FRCPC CMPT and POLQM University of British Columbia Vancouver BC Canada
GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION
International Document OIML 1 st CD GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Organization of Legal Metrology Contents 01 Scope 2 12
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Quality Manual. DuraTech Industries, Inc. 3216 Commerce Street La Crosse, WI 54603 MANUAL SERIAL NUMBER 1
Quality Manual Approval Page Document: QA1000 Issue Date: 5/29/1997 Page 1 of 17 Revision Date: 5/20/2013 DuraTech Industries, Inc. 3216 Commerce Street La Crosse, WI 54603 MANUAL SERIAL NUMBER 1 This
Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program
Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly
Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.
K-ASSAY Mouse IgM ELISA For the quantitative determination of IgM in mouse biological samples Cat. No. KT-407 For Research Use Only. 1 Rev. 072309 K-ASSAY PRODUCT INFORMATION Mouse IgM ELISA Cat. No. KT-407