WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE
|
|
- Dorcas Davis
- 8 years ago
- Views:
Transcription
1 WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales With offices around the world Maetrics, All Rights Reserved +
2 Introduction The medical device industry has over the last two years witnessed a phenomenal rise in innovative software applications. The proliferation of smart phone and tablet technology whether from Apple or her Android counterparts has fuelled the rise of innovative uses within the medical device industry. There are now apps available that contribute to the treatment of diabetes, apps that facilitate the transfer of lab results to be displayed at a nursing station, and apps that help detect cognitive disorders. This technology is contributing to major innovation efforts within the healthcare sector. And yet there is much confusion over how to deal with them for regulatory purposes. As regulatory specialists in this field, Maetrics has worked on a number of CE marking projects for medical device software app providers. In this paper, we offer our expertise and guidance from on the ground experience, to help developers remain compliant and navigate their way through the regulatory landscape. 2
3 When is software a medical device? Many conventional medical devices contain software to control their functions, whether they are intended to be used for therapy or diagnosis; the classification of the device depends on the degree of risk to the patient. But what happens when a software-based application is used and it operates on a platform such as an ipad with no controlling function on any other equipment? According the latest version of the Medical Devices Directive, which came into force in March 2010, stand alone software is considered to be an active medical device. This means that the classification rules for active devices apply and therefore this software could be classified as Class I, Class IIa or even Class IIb in cases where it is used to control or monitor the performance of Class IIb active devices. What about Apps for ipads which indicate therapeutic pathways? Are these diagnostic? There have been many examples recently of applications designed to run on the popular ipad platform, including tools for assisting in the diagnosis of disease, interpreting data generated during patient examinations and calculating dosages of medicine based on patient characteristics. The decision on whether these are to be regarded as Medical Devices depends on the Intended Use of the application. If an application is intended to diagnose a disease or even assist the clinician in reaching a diagnosis, it could easily be argued that this becomes a Medical Device by the definition in the Medical Devices Directive, 93/42/EEC, which states in the latest version as amended by 2007/47/EC: Medical device means any instrument, apparatus, appliance, software, material or other article, whether used alone or in combination, including the software intended by its manufacturer to be used specifically for diagnostic and/or therapeutic purposes..for the purpose of diagnosis, prevention, monitoring, treatment or alleviation of disease,. Therefore it can be seen that in the case of any software which is intended to be used for diagnostic purposes is a medical device. Likewise any software which is intended to be used to specify dosages of medicines is a medical device because it is being used for therapeutic purposes. 3
4 How to classify under the Medical Devices Directive? Under the Medical Devices Directive, Annex IX contains the rules for classifying the device as Class I, IIa, IIb or III: the higher the classification, the greater the risk to the patient. We would therefore expect that an application running on, for example, an ipad would not present a high degree of patient risk. However, it should be taken into account what mode of action of the software is intended. If the output from the software is to be interpreted by the clinician as part of a suite of diagnostic tools and is not used directly to determine a course of treatment, then it is more than likely classified in Class I, by Rule 12, which states that all other active devices are in Class I. Remember that stand-alone software is considered an active device. By reading the other foregoing rules 9, 10 and 11, it may sometimes be the case that the software has a greater impact, such as its use to monitor the performance of active therapeutic devices, which would then place the device in Class IIb. It is probably rare to find such applications at the current time, although in the future this may be the case. In the next year or two, there will be a recast of all three Medical Device Directives and the above situation could change. Note that in the case of a Class I device, there will be no requirement for Notified Body involvement of to verify that the device is in compliance with the directive prior to CE Marking. However, there is still the requirement to create a Technical File, which should contain all the items listed in Annex VII of the directive. 4
5 What is needed for a Product Technical File? There is no reference to a Technical File in the directive but, for any class of device, the manufacturer must prepare the following items in order to allow assessment of conformity to the Directive: A general description of the product, including any variants planned and its intended use(s); Design drawings, methods of manufacture envisaged and diagram of components, sub-assemblies, circuits, etc; The descriptions and explanations necessary to understand the above mentioned drawings and diagrams and the operations of the product; The results of the risk analysis and a list of the [harmonised] standards applied in full or in part, and descriptions of the solutions adopted to meet the essential requirements of the Directive; The results of the design calculations and of the inspections carried out, etc; The solutions adopted as referred to in Annex I [The Essential Requirements]; The pre-clinical evaluation; The clinical evaluation in accordance with Annex X; The label and instructions for use. For a device consisting entirely and solely of software, it is clear that the above requirements need to be interpreted carefully to ensure the correct information is recorded some examples of how to structure this information for software applications are given below. a. Description and Intended Use This must be provided in the Technical File so be sure to describe fully what the software is to be used for, by whom and under what conditions. Also specify what it is NOT intended to do, or who it is not intended to be used by. Are there any special training requirements as a precursor to allowing it to be released for use? Make sure that all variants of the product are defined, even if they are at the planning stage. b. What is required to satisfy the requirement to create Design Drawings and Methods of Manufacture? In the case of a pure software medical device, this would need to contain the software top-level description, the architecture definition, description of the modules, their functions and how they relate and function together. The more information, the better! It is not adequate to simply provide a programme listing. Make sure that any features which are not readily understood are fully described. The other vital feature to highlight is whether the software uses any operating system functions and how access to them is controlled. 5
6 The handling of fault conditions and fault reporting and listing of fault definitions must be included. The methods for checking for dead or redundant (sometimes development!) coding must be specified. With regard to methods of manufacture, it is suggested that all the specific procedures and tools used to design the software in question are fully defined, as well as the test methods employed. Make absolutely certain that the test methods are consistent with the test results reported in the pre-clinical evaluation. c. What is required for Risk Management? For any medical device, it is incumbent on the manufacturer to conduct a Risk Management exercise, and keep the Risk Management File up-to-date throughout the lifetime of the product. Unless there are compelling reasons to do otherwise (and these must be justified), the risk management process should follow the requirements of EN ISO 14971:2012 Medical Devices The application of Risk Management to Medical Devices. There is no exception to this for software! Following this standard will ensure that the Risk Management is in compliance with the Directive. In addition to this, careful thought must be given to the way the application interfaces with the user, especially with regard to usability and the possibility of use error giving rise to additional hazards for the user or patient. In this regard, a sister standard to EN ISO 14971:2012 is EN 62366:2008 Medical Devices The application of Usability Engineering to Medical Devices. This can be applied if the user interface has features which could give rise to critical use errors. EN is very closely linked to EN ISO 14971, and cannot be used alone without it. For assessment of how application software which has user interactions is to be specified and developed, the following are also useful: ISO :2006 Ease of operation of everyday products -- Part 1: Design requirements for context of use and user characteristics ISO TR : Ergonomics of human-system interaction Part 100: Introduction to standards related to software ergonomics Other standards in the ISO TR 9241 series can be consulted as appropriate The above standards are general to all user interfacing software products - not just medical devices. d. How are design calculations to be defined? This can be tricky for a software device! However, good practice is to show how the User Specification has been met by the software design and to make sure that all 6
7 verification and validation tests are directly related to a specific design input or group of inputs. A traceability matrix is a suggested method of providing this evidence. This is also a requirement of the FDA regulations (21 CFR part Design Controls) so the technical information created in this way can also be used to form a Design History File for FDA submissions. The V Model used for Systems Engineering Design is a suitable method to build into the software development and test procedures to achieve full traceability of tests to requirements. e. What about the Essential Requirements? In order to demonstrate that the Essential Requirements (ER) of Annex I have been satisfied, a check list should be compiled showing each ER in turn and specifying the following: Is this ER applicable to the (software) device? If not, specify why (justification rationale) If it is applicable, what standards, preferably Harmonised under the MDD have been applied? Have the standards been applied in part or in full? Specify where the evidence for compliance with the ER is to be found test report reference, risk management report, etc. It is essential that this checklist covers all 14 ERs. f. What about Clinical Evaluation? In the above list of the documentation requirements for a medical device, there are two items related to clinical: the pre-clinical evaluation and the clinical evaluation. The former can be defined as any testing of the device, in this case the software, prior to any exposure of it to real patients. Any testing conducted must be fully documented in terms of its purpose and expected outcomes, as well as any corrections to the design which result. The testing can usually be driven by the control measures adopted in the Risk Management exercises. The directive always specifies that a Clinical Evaluation be conducted and documented. This can start as a literature review to capture actual clinical performance of similar applications or software or other systems intended to be used in the same or similar therapeutic or diagnostic area of the intended use of the new device. In some cases, there will be insufficient data available in the literature and therefore a Clinical Investigation will be required. 7
8 Guidance on how to create a system of robust Clinical Evaluation is available in the Guidelines on Medical Devices document: MEDDEV Again, if the device is stand-alone software, there is no exception to the requirements! g. What about labelling and Instructions for Use? As software applications of this type are routinely downloaded from an apps store, there is no physical packaging, labelling or printed instructions. This means that the manufacturer must seek to ensure that the CE mark, identification of the variant and version of the software and any warnings or specific instructions are made available to the user in order to comply with the Directive. The favoured approach is to set the download so that this information is clearly displayed at start-up of the application. The manufacturer must ensure that all this information is subsequently available and easily accessible, for example the use of an icon which leads to a display of all the information. Care must be exercised that, when new versions are published, any changes to this information are immediately displayed and are available. It is also a requirement of the Essential Requirements of the Directive that the date or version of the Instructions for Use is identified on it. Is a Software Lifecycle Management process required? Yes, in short! The manufacturer will be expected to apply the requirements of IEC Medical Device Software Software Lifecycle Processes. This specifies that from the beginning of the software development, the manufacturer shall assign to each software system a software safety class (A, B or C) according to the possible effects on the patient, operator, or other people resulting from a hazard to which the software system can contribute. These classes are based on the following severities of harm: Class A: No injury or damage to health is possible Class B: Non-serious injury is possible Class C: Death or serious injury is possible This implies that some degree of preliminary hazard analysis should take place at the outset of the development, even before the full risk evaluation is conducted, as the classification into A, B and C has large implications on which elements are necessary to control and document during the software development cycle. 8
9 What differences are there between European (CE Marking) and other Regulatory requirements? In the United States the medical devices industry is regulated by the Food and Drug Administration (FDA). The FDA has published guidelines on Mobile Applications this guidance is to be found at:- s/ucm htm In other worldwide markets expect that the authorities will be following either the European Medical Devices directive (for example, the Therapeutic Goods Act in Australia is similar in content to it) or the guidance issues by the FDA. How is Post Market information to be collected? One key requirement of any CE-marked medical device once on the market is to gather real data on clinical performance, both to ascertain whether there are any trends which require action and to inform the regular updating of the Risk Management File. In the case of software downloads, this should in theory be simple to manage. Any application which is downloaded from an apps store should be able to be traced to a user, and fault and comments collected on-line. The use of customer surveys will be facilitated by maintenance of a user database. In order for this to be effective, there must be traceability controls in place. There must also be a procedure in place specifying how the manufacturer responds to incidents as defined in MEDDEV
10 How might operating system upgrades affect the validated state of the application? A key requirement for software running on mobile platforms is that it retains its tested, validated status. This can be difficult to achieve against a background of updates to operating systems. The manufacturer must ensure the following: The application software should be as far as possible completely independent of operating system features. For example, calculations and management of databases must occur within the application and not rely on any functions supplied with the platform; Update regimes for the platform must be fully understood and any operation system updates must be fully tested with the application before released. For the most part, the greatest risk is for the application to cease running rather than to partially fail. However, it is incumbent upon the manufacturer to carry out testing, update the Risk File (if appropriate) and warn all users if there is a potential for the application to fail or, even worse, to appear to function correctly but deliver erroneous results. 10
11 Conclusion Whenever software is used to provide a therapeutic or diagnostic outcome, then all the requirements specified in the Medical Devices Directive must be fulfilled. Special care must be taken when the software is stand-alone and can be downloaded, especially when considering what technical information is required and how it should be presented to the regulators. There must be systems in place to monitor how the device performs once on the market and all the regional requirements for incident reporting must be adhered to. Careful interpretation of the requirements is needed to ensure that the device is compliant, and this can be done with the support of regulatory specialists who understand the ins and outs of the Medical Devices Directive. They can take into account the uniqueness of your device and support you through the entire process, from concept to delivery to market. 11
12 Global Acumen From A Single Source - EU Authorized Representative - Quality Remediation - Global Regulatory Issues, 483s, Warning Letters, - CAPA - Regulatory Compliance - Regulatory Submissions - Quality Audits & Assessments - Quality System Implementation & Process Improvement - Validation Strategies, Planning, & Execution - For all types -process, software, test method, etc. - Complaints, Adverse Events & Recalls - Sterilization/contamination control - UDI - Supply chain management - Make/buy, distribution, test/quality, functions & delivery - Supply Chain Risk Assessment - Supplier Quality Auditing, Qualification & Management - Commissioning Of Facilities & Utilities - Manufacturing engineering services - Post-market surveillance - Mock FDA Inspections - Implementation & Validation Of Software Systems - Organizational change management - Maetrics U Quality Training White paper produced by Maetrics For more information, please contact global sales globalsales@maetrics.com With offices around the world Maetrics, All Rights Reserved +
How To Know If A Mobile App Is A Medical Device
The Regulation of Medical Device Apps Prepared for West of England Academic Health Science Network and University of Bristol June 2014 June 2014 1 Table of Contents 1 Purpose...3 2 Scope...3 3 The Regulation
More informationGUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES
Foreword GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES These guidance notes do not aim to be a definite interpretation of National Laws and/or regulations and are for guidance purpose only.
More informationGUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound
More informationMedical Device Software Do You Understand How Software is Regulated?
Medical Device Software Do You Understand How Software is Regulated? By Gregory Martin Agenda Relevant directives, standards, and guidance documents recommended to develop, maintain, and validate medical
More informationRegulatory Considerations for Medical Device Software. Medical Device Software
Medtec Ireland 2015 Wireless Medical Devices Regulatory Considerations for Medical Device Software Kenneth L. Block, RAC October 7, 2015 Galway, Ireland Offices: Dallas, Texas (12 employees) Tokyo, Japan
More informationSafeguarding public health The Regulation of Software as a Medical Device
Safeguarding public health The Regulation of Software as a Medical Device Dr Head of New and Emerging Technologies Medicines and Healthcare Products Regulatory Agency (MHRA) Crown Copyright 2013 About
More informationMedical Software Development. International standards requirements and practice
Medical Software Development International standards requirements and practice Food and Drug Administration What? A public health agency Why? Protect American consumers How? By enforcing the Federal Food,
More informationNavigating the New EU Rules for Medical Device Software
Navigating the New EU Rules for Medical Device Software Mathias Klümper and Erik Vollebregt examine how European Directive 2007/47/EC will affect requirements for medical device software. Directive 2007/47/EC
More informationMedical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?
Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? CEMO Congress, Paris, February 3rd, 2011 Dr. Martine Dehlinger-Kremer Vice President Global Regulatory Affairs Agenda
More informationGUIDELINES ON MEDICAL DEVICES
EUROPEAN COMMISSION DIRECTORATE GENERAL for HEALTH and CONSUMERS Consumer Affairs Health technology and Cosmetics MEDDEV 2.12/2 rev2 January 2012 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP
More informationMedical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day
Medical Devices Notified Bodies and the CE certification Process for Medical Devices European Surgical Robotics Demonstration Day 27 th March 2014, Leuven, Belgium 1 Bart Mersseman Product Manager Medical
More informationAn introduction to the regulation of apps and wearables as medical devices
An introduction to the regulation of apps and wearables as medical devices Introduction Phones are increasingly equipped with a range of devices capable of being used as sensors, including gyroscopes,
More informationGuide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical SUR-G0014-1 12 NOVEMBER 2013 This guide does not purport to be an interpretation of law and/or regulations
More informationNOTICE TO MANUFACTURERS OF RADIOTHERAPY MEDICAL DEVICES
French health products safety agency Medical Devices Evaluation Directorate Market Surveillance Department Matter handled by Pascal Di Donato Email: pascal.di-donato@afssaps.sante.fr August 2007 NOTICE
More informationMedical Device Training Program 2015
Medical Device Training Introduction Supplementary training and education is often overlooked by medical device professionals until it is triggered by an upcoming FDA or Notified Body and/or ISO 13485
More informationQMS for Software as a Medical Device [SaMD] Lessons Learned from a Quality Perspective
QMS for Software as a Medical Device [SaMD] Lessons Learned from a Quality Perspective MedCon 2015 Francis Blacha, Robert Banta Evolution from Physical to Digital Devices From To icloud Platform Mechanical
More informationGUIDELINES ON MEDICAL DEVICES EVALUATION OF CLINICAL DATA : A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES
EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.7.1 April 2003 GUIDELINES
More informationSchweppes Australia Head Office Level 5, 111 Cecil Street South Melbourne Victoria 3205. www.schweppesaustralia.com.au
Schweppes Australia Head Office Level 5, 111 Cecil Street South Melbourne Victoria 3205 www.schweppesaustralia.com.au Quality Management Systems 1. Quality Management Systems develop, implement, verify
More informationHow To Write Software
1 Medical Device Software - Software Life Cycle Processes IEC 62304 2 Credits John F. Murray Software Compliance Expert U.S. Food and Drug Administration Marcie R. Williams Medical Device Fellow Ph.D.
More informationFrequently Asked Questions. Unannounced audits for manufacturers of CE-marked medical devices. 720 DM 0701-53a Rev 1 2014/10/02
Frequently Asked Questions Unannounced audits for manufacturers of CE-marked medical devices 720 DM 0701-53a Rev 1 2014/10/02 What is an unannounced audit?... 6 Are unannounced audits part of a new requirement?...
More informationConformity assessment certification
TGA use only This form, when completed, will be classified as 'For official use only'. For guidance on how your information will be treated by the TGA see: Treatment of information provided to the TGA
More informationGuideline on good pharmacovigilance practices (GVP)
1 2 20 February 2012 EMA/541760/2011 3 4 Guideline on good pharmacovigilance practices (GVP) Module I Pharmacovigilance systems and their quality systems Draft finalised by the Agency in collaboration
More informationMed-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH
Med-Info International expert information for the Medical Device industry Council Directive 93/42/E on Medical Devices Practice-oriented summary of the most important aspects and requirements contained
More informationCE Marking: Your Key to Entering the European Market
CE Marking: Your Key to Entering the European Market N. Gerard Zapiain U.S. Department of Commerce International Trade Administration Office of Microelectronics, Medical Equipment and Instrumentation I.
More informationMedical Device Software Standards for Safety and Regulatory Compliance
Medical Device Software Standards for Safety and Regulatory Compliance Sherman Eagles +1 612-865-0107 seagles@softwarecpr.com www.softwarecpr.com Assuring safe software SAFE All hazards have been addressed
More informationImplementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes.
Implementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes.. www.pharmout.net Page 1 of 15 Version-02 1. Scope 1.1. Purpose This paper reviews the implementation of the ANSI/AAMI/IEC
More informationEPF Position Statement on the European Commission s proposal for a Regulation on In Vitro Medical Devices
EPF Position Statement on the European Commission s proposal for a Regulation on In Vitro Medical 03/09/2013 In Vitro Diagnostic medical devices include all tests performed to provide a diagnosis by assessing
More informationPreparing for Unannounced Inspections from Notified Bodies
Preparing for Unannounced Inspections from Notified Bodies Europe has introduced further measures for unannounced audits of manufacturers by notified bodies. With this in mind, James Pink, VP Europe-Health
More informationIntroduction into IEC 62304 Software life cycle for medical devices
Introduction into IEC 62304 Software life cycle for medical devices Christoph Gerber 4. September 2008 SPIQ 9/5/2008 1 Agenda Current Picture Regulatory requirements for medical device software IEC 62304
More informationMedical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011
Medical Certification: Bringing genomic microcores to clinical use VoltedFlow GmbH Industriestrasse 23, 6055, Alpnach Dorf, Switzerland 1 Summary VoltedFlow has developed solutions to vastly speed up DNA
More informationGuidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device
Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device List of Contents Page 1. Introduction 1.1 The Medical Device Administrative Control System and the 3 proposed legislation
More informationGuideline on good pharmacovigilance practices (GVP)
22 June 2012 EMA/541760/2011 Guideline on good pharmacovigilance practices (GVP) Module I Pharmacovigilance systems and their quality systems Draft finalised by the Agency in collaboration with Member
More informationLegal compliance for developers. Training materials (prepared by Tilburg University)
Legal compliance for developers using FI-STAR ehealth platform Training materials (prepared by Tilburg University) Target audience: Target audience and objectives developers & other potential users of
More informationSoftware Development for Medical Devices
Software Development for Medical Devices Overcoming the Challenges of Compliance, Quality and Cost Software is fast becoming the differentiator for manufacturers of medical devices. The rewards of software
More informationSoftware Development for Medical Devices
Overcoming the Challenges of Compliance, Quality and Cost An MKS White Paper Introduction Software is fast becoming the differentiator for manufacturers of medical devices. The rewards available from software
More informationSoftware within the medical device regulatory framework in the EU
Software within the medical device regulatory framework in the EU E-HEALTH WEEK (AMSTERDAM, 10 JUNE 2016) Erik Hansson, Deputy Head of Unit, DG GROWTH - Health Technology and Cosmetics, European Commission
More informationEffective Software Verification for Medical Devices
STERLINGTECH AND KLOCWORK WHITE PAPER NOVEMBER 2009 Effective Software Verification for Medical Devices Achieving compliance and meeting productivity goals with static analysis In addition to producing
More informationBSI Road Show: September 8 th to 15 th, 2014
BSI Road Show: September 8 th to 15 th, 2014 Post Market Surveillance (including PMCF): common non compliances Ibim Tariah Ph.D Technical Director, Healthcare Solutions Itoro Udofia Ph.D Global Head, Orthopaedics
More informationRegulated Mobile Applications
Regulated Mobile Applications Sion Wyn Conformity +[44] (0) 1492 642622 sion.wyn@conform-it.com Waters, Wilmslow, November 2014 1 Introduction According to industry estimates, 500 million smartphone users
More informationDemystifying the European Machinery Directive and SEMI Requirements for the Industrial Automation and Semiconductor Markets
Demystifying the European Machinery Directive and SEMI Requirements for the Industrial Automation and Semiconductor Markets An Overview of Current Changes and Directions By Andras Szende Senior Engineer
More informationGuidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC August 2013 Contents 1 Introduction...3 2 Scope of the directive...3 2.1 What is an in vitro diagnostic medical device?... 3 2.2 Specimen
More informationISO 13485:201x What is in the new standard?
ISO 13485:201x What is in the new standard? Eric Finegan, Quality Mgr, BTE Technologies, Inc. 2015-09-10 1 Presentation Slides This slide deck is the presentation performed on 2015-09-10. A more detailed
More informationUNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES
UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES A WHITE PAPER ON THE REQUIREMENTS, REGULATIONS AND OPPORTUNITIES CONTAINED IN EC DIRECTIVE 98/79/EC AND A BRIEF COMPARISON
More informationReporting Changes to your Notified Body
Reporting Changes to your Notified Body Vicky Medley - Head of QMS, Medical Devices September 2015 1) Requirements & Guidance 2) Reporting Changes - What, When, Why & How 3) Some examples 4) Summary &
More informationMedical Devices: CE Marking Step-by Step
Medical Devices: CE Marking Step-by Step Sylvia Mohr Standards Specialist U.S. Mission to the European Union 1 Introduction CE marking in 8 Steps Overview intended purpose - technical documentation classification
More informationRAPS ONLINE UNIVERSITY
RAPS ONLINE UNIVERSITY Practical education and training for business success. For regulatory professionals, there is only one name to know and trust for online education and training RAPS Online University.
More informationProcesses for the Development of Healthcare Applications. Christian Johner
Processes for the Development of Healthcare Applications Christian Johner Process/Life-Cycle Models: The true history Directives and Standards: Requirements with Respect to Processes Conclusion, Recommendations
More informationHow to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters
environmental failure analysis & prevention health technology development How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters Kevin L. Ong, Ph.D., P.E. Managing
More informationTotal Product Lifecycle Solutions from NSF Health Sciences Medical Devices
Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices Experts in medical device quality systems, compliance, regulatory affairs, auditing and training www.nsf.org The Right People.
More informationOverview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC
Overview of Medical Device Controls in the US By Nandini Murthy, MS, RAC 18 controls are a regulatory requirement for medical devices. In the US, compliance with the design controls section of 21 Code
More informationIVD Regulation Overview. Requirements to Assure Quality & Effectiveness
IVD Regulation Overview Requirements to Assure Quality & Effectiveness CLIAC Jan. 2002 Statutory and Regulatory Requirements Statute: Food, Drug, and Cosmetic Act Food and Drugs Act of 1906 Food and Drug
More informationEmpowering the Quality and Regulatory Compliance Functions
Empowering the Quality and Regulatory Compliance Functions Management must take steps to ensure that regulatory and quality compliance is everyone s responsibility. By: J. Glenn George, Kenneth Imler,
More informationIntroduction Processes and KPIs Health Care Value Chain. Rafael Provencio
Introduction Processes and KPIs Health Care Value Chain Rafael Provencio Table of contents 1. Regulatory Framework EU 2. Hospital Value Chain 3. Processes 4. KPIs 2 What is the goal of the regulatory framework?
More informationDeveloping a Mobile Medical App? How to determine if it is a medical device and get it cleared by the US FDA
Developing a Mobile Medical App? How to determine if it is a medical device and get it cleared by the US FDA In this presentation: App stats: Explosive growth Examples already cleared by the US FDA Is
More informationUpdate: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)
Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR) Paul Brooks Healthcare Solutions Disclaimer The new regulations are not finalized and subject to change http://medicaldevices.bsigroup.com/en-us/resources/whitepapers-and-articles/
More informationDIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
29.3.2014 Official Journal of the European Union L 96/149 DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 26 February 2014 on the harmonisation of the laws of the Member States relating
More informationRegistration of Class B Medical Devices
Registration of Class B Medical Devices 1 INTRODUCTION Medical devices are classified based on a rule based risk classification system into four risk classes Class A to D with Class A being the lowest
More informationThe Medical Device Industry in Korea: Strategies for Market Entry
The Industry in Korea: Strategies Seth J. Goldenberg, Senior Principal Scientist, NAMSA; Yongha Na, Director, Medipert Regulatory NAMSA Whitepaper #09 06/2014 Seth J. Goldenberg, PhD (sgoldenberg@namsa.com)
More informationRegulation of Medical Apps FDA and European Union
JHW Comment - The content of below article is intended to provide a general guide to the subject matter. Specialist advice should be sought about your specific circumstances. Erik Vollebregt Erik is a
More informationASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION
PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 038-1 26 March 2012 AIDE-MEMOIRE ASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION PIC/S March 2012 Reproduction
More informationDevelopment of a Process Assessment Model for Medical Device Software Development
Development of a Process Assessment Model for Medical Device Software Development Marion Lepmets, Paul Clarke, Fergal McCaffery, Anita Finnegan, Alec Dorling Regulated Software Research Centre, Dundalk
More informationDesign Controls: Are They Worth the Effort?
Design Controls: Are They Worth the Effort? Compliance-Alliance conducted a survey to measure the effects of FDA s design control regulation on the industry. Most respondents believe the controls have
More informationRegulation and Risk Management of Combination Products
Regulation and Risk Management of Combination Products Katherine Ulman Associate Scientist Global Regulatory Compliance Manager Dow Corning Healthcare Jim Curtis Senior Specialist, Healthcare Applications
More informationALL Medical Devices regardless of Classification
TÜV SÜD Product Service GmbH 30.09.2013 http://omowizard.files.wordpress.com/2010/07/fit-for-purpose.jpg Folie 1 Medical Device Requirements for Clinical Data Dir. 2007/47/EC Which medical devices require
More informationKonformitätsbewertung 3.9 B 17. Guidance for Notified Bodies auditing suppliers to medical device manufacturers
Antworten und Beschlüsse des EK-Med Konformitätsbewertung 3.9 B 17 Guidance for Notified Bodies auditing suppliers to medical device manufacturers Herkunft Notified Body Operations Group Quellen NBOG BPG
More informationBoost the Success of Medical Device Development With Systematic Literature Reviews
FOR PHARMA & LIFE SCIENCES WHITEPAPER Boost the Success of Medical Device Development With Systematic Literature Reviews Biomedical literature supports medical Device development Before the ideation and
More informationPLM: Privat Label Manufacturer (Customer of the OEM-PLM relation) OEM: Original Equipment Manufacturer (Supplier of the OEM-PLM relation)
1 Area of application of the product documentation in case of a Privat Label Manufacuter according to OEM procedure. To apply with new certification applications and sample evas in the context of the regular
More informationitac solutions for the medical industry Quality assurance of the highest standard FDA-compliant. Reliable. Productive.
Quality assurance of the highest standard FDA-compliant. Reliable. Productive. Market Requirements 1 Statutory regulations and stringent requirements laid down by the American regulatory authority, FDA
More informationORACLE CONSULTING GROUP
ORACLE CONSULTING GROUP An Official United States Agent Firm for Foreign Establishments CONSULTING MEMORANDUM: DEALING WITH A MEDICAL DEVICE IN THE U.S. 5398 Golder Ranch Rd., Ste. 1 Tucson, Arizona 85739
More informationThis is Document Schedule 5 Part 1 referred to in this Contract SCOTTISH MINISTERS REQUIREMENTS SCHEDULE 5 PART 1 QUALITY MANAGEMENT SYSTEM
This is Document Schedule 5 Part 1 referred to in this Contract SCOTTISH MINISTERS REQUIREMENTS SCHEDULE 5 PART 1 QUALITY MANAGEMENT SYSTEM CONTENTS Page No 1 GENERAL REQUIREMENTS 1 1.1 Requirements 1
More informationWorking Party on Control of Medicines and Inspections. Final Version of Annex 16 to the EU Guide to Good Manufacturing Practice
Version 8 (final) EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 S\common\legal-legislation\75-319nd81-851\91-356\eudralexvol4\Annex
More informationSoftware-based medical devices from defibrillators
C O V E R F E A T U R E Coping with Defective Software in Medical Devices Steven R. Rakitin Software Quality Consulting Inc. Embedding defective software in medical devices increases safety risks. Given
More informationGuideline on good pharmacovigilance practices (GVP)
9 April 2013 EMA/816573/2011 Rev 1* Guideline on good pharmacovigilance practices (GVP) Module II Pharmacovigilance system master file (Rev 1) Draft of first version finalised by the Agency in collaboration
More informationVigilance Reporting. Vicky Medley - Head of QMS, Medical Devices. September 2015. Copyright 2015 BSI. All rights reserved.
Vigilance Reporting Vicky Medley - Head of QMS, Medical Devices September 2015 2 Why? 3 protecting and improving public health https://www.gov.uk/government/organisations/medicines-and-healthcareproducts-regulatory-agency/about
More informationPHARMACEUTICAL QUALITY SYSTEM Q10
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE PHARMACEUTICAL QUALITY SYSTEM Q10 Current Step
More informationFrequently Asked Questions (FAQ) About Patient Management Software Licensing
Frequently Asked Questions (FAQ) About Patient Management Software Licensing V1.0 April 12, 2010 About ITAC The Information Technology Association of Canada (ITAC) is the voice of the Canadian information
More informationCopyright, Language, and Version Notice The official language of this [Certification Protocol] is English. The current version of the [Certification
Copyright, Language, and Version Notice The official language of this [Certification Protocol] is English. The current version of the [Certification Protocol] is maintained on the Bonsucro website: www.bonsucro.com.
More informationORACLE CONSULTING GROUP
ORACLE CONSULTING GROUP 9 Golder Ranch Rd., Ste. 1 Tucson, Arizona 9 Web Site: E-mail: 20-2-0 20-2-0 (FAX) CONSULTING MEMORANDUM QUALITY SYSTEM INSPECTION TECHNIQUE
More informationDECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL
L 218/82 EN Official Journal of the European Union 13.8.2008 DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL DECISION No 768/2008/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of
More informationUsability of Medical Applications Ved Line Kagenow Svenstrup, lks@delta.dk
Usability of Medical Applications Ved Line Kagenow Svenstrup, lks@delta.dk What is usability? The user, rather than the system, at the center of the process. Risk of operating errors that can cause injury
More informationGuidance for Industry. Q10 Pharmaceutical Quality System
Guidance for Industry Q10 Pharmaceutical Quality System U.S. Department of Health and Human Services Food and Drug Administration Center for Drug Evaluation and Research (CDER) Center for Biologics Evaluation
More informationRegulation of Medical Devices involving Software in Australia an Overview
Regulation of Medical Devices involving Software in Australia an Overview John Jamieson Conformity Assessment Branch Therapeutic Goods Administration PO Box 100, Woden ACT 2606 John.Jamieson@health.gov.au
More informationMobile Medical Application Development: FDA Regulation
Mobile Medical Application Development: FDA Regulation Mobile Medical Applications: Current Environment Currently over 100,000 mobile health related applications are available for download; there is an
More information2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002
STATUTORY INSTRUMENTS 2002 No. 618 CONSUMER PROTECTION The Medical Devices Regulations 2002 Made - - - - - 20th May 2002 Laid before Parliament 21st May 2002 Coming into force - - 13th June 2002 ARRANGEMENT
More informationBest Practice In A Change Management System
Quality & Compliance Associates, LLC Best Practice In A Change Management System President Quality & Compliance Associates, LLC Change Control and Its Role in a Continuous Improvement Environment 3 Benefits
More informationHealth Informatics Application of clinical risk management to the manufacture of health software Formerly ISO/TS 29321:2008(E) DSCN14/2009
Health Informatics Application of clinical risk management to the manufacture of health software (formerly ISO/TS 29321:2008(E)) DSCN14/2009 Programme Sub-Prog / Project NPFIT Clinical Safety Document
More informationWhen Your Life Depends on Software
Published on Quality Digest (http://www.qualitydigest.com) Home > Content When Your Life Depends on Software IEC 62304 imposes requirements on software for medical devices. We re all aware of the importance
More informationClinical evaluation Latest development in expectations EU and USA
Clinical evaluation Latest development in expectations EU and USA Medical Devices: staying ahead of regulatory developments Gert Bos BSI Israel 22 April - Herzliya Copyright 2012 BSI. All rights reserved.
More informationINTEGRATED MANAGEMENT SYSTEM MANUAL IMS. Based on ISO 9001:2008 and ISO 14001:2004 Standards
INTEGRATED MANAGEMENT SYSTEM MANUAL IMS Based on ISO 9001:2008 and ISO 14001:2004 Standards Approved by Robert Melani Issue Date 30 December 2009 Issued To Management Representative Controlled Y N Copy
More informationResponse of the German Medical Association
Response of the German Medical Association to the European Commission proposal for a regulation of the European Parliament and of the Council on clinical trials on medicinal products for human use, and
More informationICH guideline Q10 on pharmaceutical quality system
September 2015 EMA/CHMP/ICH/214732/2007 Committee for Human Medicinal Products Step 5 Transmission to CHMP May 2007 Transmission to interested parties May 2007 Deadline for comments November 2007 Final
More informationMEDICAL DEVICES INTERIM REGULATION
Issued by the Saudi Food and Drug Authority Board of Directors decree number 1-8-1429 and dated 27 December 2008 The Interim was published in Umm Al-QURA Journal year 86 Issue No:4249 dated in 17 April
More informationCertification Authorities Software Team (CAST) Position Paper CAST-26
Certification Authorities Software Team (CAST) Position Paper CAST-26 VERIFICATION INDEPENDENCE COMPLETED January 2006 (Rev 0) NOTE: This position paper has been coordinated among the software specialists
More informationGuidelines for conformity assessment of In Vitro Fertilisation (IVF) and Assisted Reproduction Technologies (ART) products
EUROPEAN COMMISSION DG HEALTH & CONSUMERS Directorate B - Consumer Affairs Unit B2 Health Technology and Cosmetics MEDICAL DEVICES: Guidance document MEDDEV 2.2/4 January 2012 Guidelines for conformity
More informationThis interpretation of the revised Annex
Reprinted from PHARMACEUTICAL ENGINEERING The Official Magazine of ISPE July/August 2011, Vol. 31 No. 4 www.ispe.org Copyright ISPE 2011 The ISPE GAMP Community of Practice (COP) provides its interpretation
More informationREFERENCE DOCUMENT. (DRAFT for Public Comments) Title: White Paper on Medical Device Software Regulation Software Qualification and Classification
AHWP/WG1/R001:2014 REFERENCE DOCUMENT (DRAFT for Public Comments) Title: White Paper on Medical Device Software Regulation Software Qualification and Classification Author: Work Group 1, Pre-Market Submission
More informationEVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR STUDIO on the Cloud Data Management Software DECISION SUMMARY
A. DEN Number: DEN140016 EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR STUDIO on the Cloud Data Management Software B. Purpose for Submission: DECISION SUMMARY De novo request for adjunct data management
More informationGFMAM Competency Specification for an ISO 55001 Asset Management System Auditor/Assessor First Edition, Version 2
GFMAM Competency Specification for an ISO 55001 Asset Management System Auditor/Assessor First Edition, Version 2 English Version PDF format only ISBN 978-0-9871799-5-1 Published April 2014 www.gfmam.org
More informationABSTRACT. The Guidelines Section F is related to the Purchasing requirements of NSQ100 (Chapter 7.4). Summary
Page 1 / 9 ABSTRACT This document is the Section F to NSQ-100 Guidelines. Its objective is to help to the understanding of NSQ-100 requirements through some examples or recommendations and descriptions
More informationGOOD DISTRIBUTION PRACTICE FOR MEDICAL DEVICES (GDPMD)
MEDICAL DEVICE CONTROL DIVISION Ministry of Health, Malaysia Regulatory Requirements for Medical Device Safety & Performance GOOD DISTRIBUTION PRACTICE FOR MEDICAL DEVICES (GDPMD) DRAFT DRAFT 1 September
More information