CT simulator. Professor and Director Department of Radiation Oncology. University of Pittsburgh Cancer Institute Pittsburgh, Pennsylvania, USA

Similar documents
Quality Assurance of Radiotherapy Equipment

INTRODUCTION. A. Purpose

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

Key words: treatment planning, quality assurance, 3D treatment planning

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

WHERE IN THE WORLD JILL LIPOTI?

Quality Control and Maintenance Programs

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

How To Improve Your Ct Image Quality

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Assessing Radiation Dose: How to Do It Right

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

AAPM REPORT NO. 39 SPECIFICATION AND ACCEPTANCE TESTING OF COMPUTED TOMOGRAPHY SCANNERS

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

First Three Years After Project Proton Therapy Facility:

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

HDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

American College of Radiology CT Accreditation Program. Testing Instructions

GE Healthcare. pet/ct for simulation. Precision in motion.

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Prepublication Requirements

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Scripps Proton Therapy Center: Configuration and Implementation

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Physics testing of image detectors

Purchasing a cardiac CT scanner: What the radiologist needs to know

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

THE DOSIMETRIC EFFECTS OF

Accreditation Is Coming

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Our Department: structure and organization

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

IAEA HUMAN HEALTH SERIES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Quality Control of Full Field Digital Mammography Units

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

Elements of required physical infrastructures: space, schielding, and patient flow..

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

QA of intensity-modulated beams using dynamic MLC log files

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA USA phone NOMOS

Digital Breast Tomosynthesis QC Requirements

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

DICOM Correction Item

Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

Performance testing for Precision 500D Classical R/F System

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

Page: 1 of 6 Page: 1 of 6

MRI QA Technologist s Tests

Data. microcat +SPECT

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

Fundamentals of Cone-Beam CT Imaging

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS

Use of 3D Printers in Proton Therapy

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Recommendations for a technical quality control program for diagnostic X-ray equipment

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

Quality Assurance in Stereotactic. Radiotherapy

Failure Modes and Effects Analysis (FMEA)

Transcription:

CT simulator M. Saiful Huq,, Ph.D., FAAPM, FInstP Professor and Director Department of Radiation Oncology University of Pittsburgh Cancer Institute Pittsburgh, Pennsylvania, USA

Outline What is a CT simulator? CT simulation process Functional performance characteristics of CT equipment QA of CT simulators QA for CT simulation software Evaluation of CT simulation process Conclusion

CT simulator CT scanner with flat table top Laser positioning and marking system (preferable with external lasers) Virtual simulation software Various hardcopy output devices

CT Simulation Process Varies from center to center 3 major categories: 1. CT-scan, patient positioning and immobilization 2. Treatment planning and CT-simulation 3. Treatment setup

CT Simulation Process 1. CT-scan, patient positioning and immobilization Scanner acquires volumetric scan of a patient, which represents a virtual or digital patient Scan is acquired with the patient immobilized in treatment position treatment specific scan protocols often increased scan limits use of contrast placement of localization marks on the patient skin

CT Simulation Process 2. Treatment planning and CT-simulation The virtual simulation software provides virtual representations of the geometric capabilities of a treatment machine It also allows import, manipulation, display, and storage of images from CT and/or other imaging modalities

CT Simulation Process 2. Treatment planning and CT-simulation Beam placement and treatment design is performed using virtual simulation software Contouring of target and OAR using information obtained from different imaging modalities Treatment isocenter placement» can be a final isocenter marked (needs physician presence)» a reference point marked (does not need physician s presence)» localization marks are placed on the patient s skin Placement of the beams, design of treatment portals and communication of information to TPS Printing of DRRs and patient setup instruction

CT Simulation Process 3. Treatment setup Patient is setup on the treatment machine according to instructions created from the CT-sim software Port films are acquired and compared with CT-sim DRRs

CT simulation process CT simulation includes the CT scanner and components of treatment planning system and provides input for dose calculation 3 4 5 1 2 CT Simulator CT Scanner Virtual Simulation Dose Calculation Treatment Planning System

Functional performance characteristics of CT equipment A. CT scanner and virtual simulation system X-ray tube» Large no of images and rapid data acquisition time requires the tube to withstand high heat and dissipate heat quickly Collimator and attenuator» pre-patient patient collimator used to produce a narrow beam» Post-patient collimator used to reduce scatter to detector Patient support table» Must have a flat table top

Functional performance characteristics of CT equipment A. CT scanner and virtual simulation system Computer workstation» Multiple microprocessors coordinate setup of the scan parameters, x-ray x production, data acquisition, processing of transmission data and display of reconstructed image External patient marking/positioning lasers» Used for marking for patients» important that sagittal laser be mobile because the CT table cannot move laterally

Functional performance characteristics of CT equipment B. Conventional and spiral CT Conventional CT acquires data one slice at a time Spiral CT allows data to be acquired while the table translates and the tube rotates continuously. Path of the tube forms a helical pattern around the patient C. Multislice scanners Projection data from multiple slice is acquired simultaneously Uses multiple row of detectors in the z axis Enables acquisition of imaging studies faster than single slice scanners

Image Generation Single Slice One Rotation One Image

Multislice CT 1 2 3 4 One Rotation Multiple Images

CT scanner Single and multi-slice scanning Wider collimator widths Post patient collimation Multiple area detectors 1992 - Elscint CT Twin - first CT scanner capable of simultaneously acquiring more than one transaxial slice 1998 Four major manufacturers introduce scanners capable of scanning 4 slices simultaneously Today - 64+ slice scanners commercially available

CT scanner Multi-slice scanning Faster scan times 4 slice scanner example (8 times faster):» multi: 0.5 sec/rotation and 4 slices/rotation» single: 1 sec/rotation and 1 slice/rotation Lower tube heat loading Longer volume covered per rotation Improved temporal resolution - faster scan times Improved spatial resolution thinner slices Decreased image noise more ma available

Resolution The lower limit on slice thickness for most single slice scanners in radiotherapy is 3 mm Often 5 and 8 mm slices were used Multislice scanners can acquire 0.75 mm thick slices with equivalent scan parameters Everything else being equal thinner slice thickness produces better DRRs

DRR image quality Everything else being equal, thinner slices produce better images Balance between large amounts of data and image quality 5mm Slices 3mm 0.8mm

DRR Image Quality 5mm Slices 3mm 0.8mm

DRR Image Quality 800 Images 0.8 mm slice thickness

Multi-slice CT - Speed Single slice scanner 30 seconds Multi slice scanner 2 to 4 seconds Dynamic CT 4D CT 5D CT Gating Tumor motion

Multi-slice CT - Dynamic CT Approximately 7000 images

Functional performance characteristics of CT equipment D. Large bore scanners 85 cm Bore Opening 70 cm Bore Opening

CT scanner Bore size Patient Size

CT scanner Bore size Mantle Field

Functional performance characteristics of CT equipment E. CT performance parameters X-ray system calibration Collimator assessment Localization laser alignment Slice width and sensitivity profile Radiation exposure and dose E. CT performance parameters Image uniformity and noise Spatial resolution Contrast resolution CT number calibration and linearity Artifact evaluation

Quality assurance of CT simulators

Commissioning Ensure that the equipment performs within the specifications agreed on by the vendor and user Establish a record of baseline performance An opportunity for a physicist to become an expert user Develop clinical procedures and policies Develop a quality assurance program

CT Sim QA AAPM TG66 QA recommendations and guidelines for diagnostic radiology and radiation therapy for CT-Simulators Equipment description Tests for CT scanners Virtual simulation software Image registration Overall CT-Simulation process Test frequencies, methods, and tolerances

Quality assurance of CT scanners The AAPM Report Number 39- Specification and acceptance testing of computed tomography scanners Radiation safety CT Dosimetry Electromechanical tests X-ray Generator Image performance tests

QA program Purpose To assure proper equipment and program operation To assure safety of patient, staff, and public Should include specification of tests test frequency tolerance limits corrective actions QA assignments

QA components Radiation and patient safety CT dosimetry Evaluation of electromechanical components Evaluation of image quality

Radiation Safety/Interlocks Exposure Measurements Two shielding design methods commonly employed» direct use of isoexposure contours»ncrp Report No. 49 & 147 Interlocks (x-ray interruption, dead man) Patient Dose (CTDI, MSAD)

Survey Sites G F L E K C D B A I J H

Staff and Public Weekly Exposure NCRP 49: Controlled Areas: 100 mr/week Non-Controlled Areas: 10 mr/week Brilliance AcQSim Location Weekly Weekly Measured Exposure Exposure Instantaneous Occupancy Based on Based on Exposure Rate Factors Workload Workload (mr/hr) (mr/week) (mr/week) A. Room 450 Door 0.18 1.00 0.026 0.645 B. Room 450 Window 0.30 1.00 0.043 1.075 C. Room 450 Countertop 0.33 1.00 0.047 1.183 D. Room 450 Chair 0.22 1.00 0.031 0.788 E. Room 451 0.23 0.25 0.008 0.206 F. Corridor 474 Door 0.10 0.25 0.004 0.090 G. Corridor 474 0.17 0.25 0.006 0.152 H. Corridor 362 0.15 0.25 0.005 0.134 I. Room 454 0.10 0.25 0.004 0.090 J. Room 448 0.10 1.00 0.014 0.358 K. Room 447 0.10 0.25 0.004 0.090 L. First floor lobby background

Patient Dose CTDI - Computed Tomography Dose Index represents average dose per slice MSAD - Multiple Scan Average Dose represents average dose per scan Body Dose from CT Scans ~ 1 cgy Continually evolving concept due to multislice technology

CTDI

CTDI Measurement Technique Pencil Ionization Chamber (long cylindrical ionization chamber) TLD array Body and Head phantom Technique kvp mas Slice Width Index Other parameters

CTDI CT dose Index (CTDI) For axial scan only Ideally from ±,, in reality 10 cm f med = 0.78 cgy/r (acrylic) n is number of slices T is slice width CTDI = 100 Rdg * Ctp * kel * N x * fmed *100mm cgy n*t (mm) [ ] CTDI 100 1 + 50mm = nt 50mm ( ) D z dz

Electromechanical Tests Localization Laser Accuracy Table/Image Plane Orthogonality Gantry Tilt Slice Localization from Scout Image (Topogram, Pilot Image) Table Indexing Collimation Radiation Profile Width Sensitivity Profile Width X-ray Generation kvp HVL Timer accuracy mas Linearity mas Reproducibility

Electromechanical components Laser Alignment/Accuracy Gantry lasers should accurately identify scan plane within the gantry opening Gantry lasers should be parallel and orthogonal with the scan plane and should intersect in the center of scan plane Vertical side-wall lasers should be accurately spaced from imaging plane Wall lasers should be parallel and orthogonal with the scan plane, and should intersect at a point which is coincident with the center of the scan plane The overhead (sagittal) laser should be orthogonal to the imaging plane The overhead (sagittal) laser movement should be accurate, linear and reproducible

Electromechanical components Table Alignment/Accuracy The tabletop should be level and orthogonal with respect to the imaging plane Table vertical and longitudinal motion according to digital indicators should be accurate and reproducible Table indexing and position under scanner control should be accurate Flat tabletop should not contain any objectionable artifact producing objects

Electromechanical components Gantry Alignment/Accuracy The angle of the gantry tilt with respect to nominal vertical imaging plane should be accurate After the tilt, the gantry should return to the nominal vertical imaging plane

Electromechanical Components x-ray Generator Need a non-invasive meter kv accuracy Timer accuracy malinearity HVL measurements

X-ray Generation As many possible modes as practical should be verified: kvp, time, HVL - non-invasive tests device ma Linearity - inferred indirectly using a mas linearity measurement. For a constant tube potential and slice width, the integral exposure should be a linear function of mas. mas Reproducibility

CT simulator mechanical alignment

Image quality indicators Quantitative Phantom Measurements»High Contrast»Low Contrast»Uniformity» Spatial Integrity»Artifacts»Slice thickness» CT # accuracy Qualitative Physician Preferences» Tumor» Normal Structures» DRR/DCR Objects» Workflow» Customized protocols

Image performance Catphan 500 CTP401 CTP515 CTP445 CTP528 CTP486

Ability of the system to record separate images of small objects that are placed very close together Resolution (High Contrast)

Subject contrast (Low Contrast) Ability of a system to resolve adjacent objects with small density differences Noise limited

Field uniformity The average CT number of water should always be 0, independent of the position within the homogeneous phantom Measurements are performed using head and body uniformity phantoms Measure average CT numbers in center and at the periphery (several locations) of the phantom

Uniformity

Image performance evaluation

Quantitative CT Phantom with sections of known electron densities Determine CT numbers for each section Compare measured CT numbers with vendors specifications Should be independent of phantom size CT numbers are a function of kvp Edge enhancement algorithms should be avoided

CT Number Accuracy (MIR with Manufacturer s s Phantom) Material Big Bore (HU) Std Bore (HU) Manufacturer (HU) Polyethylene 70.9 75.1 75 ± 15 Water 2.5 2.5 0 ± 4 Nylon 100.0 100.5 100 ± 15 Lexan 114.0 117.5 116 ± 15 Acrylic 136.0 137.5 140 ± 15 Teflon 1020.0 1033.0 1016 ± 50 Both scanners exhibited CT number accuracy well within manufacturer s s specifications

Spatial accuracy Spatial accuracy within the image plane can be verified by imaging a convenient object of known dimensions

Virtual simulation software tests Beam geometry Target localization Target contour verification (using BEV) Isocenter calculation Isocenter movement Virtual field size Virtual gantry, collimator, and table rotation Evaluation of digitally reconstructed radiographs (DRRs)

Virtual simulation software QA Fraass et al. AAPM TG 53: Quality assurance for clinical radiotherapy treatment planning. Med Phys,25:1773-1829,1998 Spatial and geometric accuracy contour delineation isocenter localization treatment port definition virtual treatment machine operation Evaluation of DRRs and DCRs Accuracy of the multimodality image fusion and registration

QA for the CT simulation process CT scan transfer of images isocenter localization treatment plan design multimodality image fusion transfer of virtual simulation data for calculation of dose distributions DRR generation transfer of the plan to treatment machine treatment of the patient treatment verification

Quality assurance frequency Daily (Therapists) Weekly (Therapists) Monthly (Clinical Engineering) Annual (Medical Physics)

Daily QA (Sim Therapists) Radiographic Uniformity Most important - look for noise and distortions in CT image Laser Alignment and Tracking Critical for patient setup Most critical - Couch in/out 500 mm tracking Audio/Visual System

Weekly Inspection (ok, not ok) (Sim Therapists) Scanner couch Couch and gantry controls Gantry lasers Wall lasers Overhead laser General image quality Control console Software functions Networking and archiving Audio communication Machine operation during scanning Room condition Treatment aids condition

Monthly QA (Clinical Engineering) Couch Digital Position Accuracy Laser Digital Position Accuracy Laser Alignment Slice Thickness Spatial Resolution Contrast resolution CT# Accuracy, Uniformity, and Noise MTF Distance Measuring Tool Accuracy

Conclusions CT will remain the primary imaging modality in radiotherapy CT simulator only departments Large bore multislice CT will become standard scanners for radiotherapy Existing documents (Report #39, TG66, TG53) are a good foundation New CT scanning capabilities/parameters can be evaluated with the existing recommendations

Acknowledgements I would like to thank Sasa Mutic for giving me many of the slides that I presented in this talk