Principles and Techniques of Blood Pressure Measurement

Similar documents
Manual Blood Pressure Measurement

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Blood Pressure Guidelines and Screening Techniques

Plank 1 Tool: Hypertension Medical Assistant Training (Providence Medical Group)

MEASURING AND RECORDING BLOOD PRESSURE

How to measure blood pressure using digital monitors

Measuring Your Blood Pressure at Home. A Review of the Research for Adults

OEM MAXNIBP Frequently Asked Questions

MEASURING VITAL SIGNS TRAINING CHECKLIST

Indirect Blood Pressure Measurement

MEASURING AND RECORDING BLOOD PRESSURE

PTA 210 PTA Technique. Vital Signs

Blood pressure measurement devices. December 2013

Measurement of Blood Pressure

CHEM 1406 BSA/BMI Measurement Laboratory (revised 01/03//06) BSA, Body Surface Area

Automatic Wrist Blood Pressure Monitor with

Diagnostic Guidelines for Blood Pressure Screening Techniques

Responses to Static (Isometric) Exercise

Vasc Health Risk Manag. 2011; 7: Published online 2011 November 30. doi: /VHRM.S27193 PMCID: PMC

Prepared by: Westat, Inc Research Boulevard Rockville, MD September 1989

Body cavities. Body Planes

Blood Pressure. Blood Pressure (mm Hg) pressure exerted by blood against arterial walls. Blood Pressure. Blood Pressure

Vital and Diagnostic Signs

a Centre de Médecine Cardiovasculaire, Paris, France and b Lebanese Received 7 June 2009 Revised 6 September 2009 Accepted 17 September 2009

Pressure Sensor Based Estimation of Pulse Transit Time

AMBULATORY BLOOD PRESSURE MONITORING QUICK START GUIDE

INVESTIGATING HEART RATE AND BLOOD PRESSURE

An Overview of Transradial Patient Set-up. Susan R. Cooney RN,BSN,CCRN Durham VAMC Cardiac Catheterization Lab Duke University School of Nursing

Introduction. Provider Toolkit

Blood Pressure Assessment Program Screening Guidelines

Dialysis Professional

Arterial pressure monitoring Direct arterial pressure monitoring permits continuous measurement of systolic, diastolic, and mean pressures and allows

Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003

Section Two: Arterial Pressure Monitoring

The Body s Transport System

An Introduction to Blood Pressure

Circulation Stations

Project 4.2.1: Heart Rate

MY TYPE 2 DIABETES NUMBERS

Name: OMRON M3 HEM-7200-E Upper Arm Blood Pressure Monitor

Health Technician skills checklist Health Services Department Lincoln Public Schools TEMPERATURE

Cardiovascular Physiology

VENIPUNCTURE: PART 3 VEIN ASSESSMENT AND SELECTION

Affordable ABPM and strictly guideline-based SBPM for accurate out-of-office blood pressure measurement.

Upper Extremity Vein Mapping for Placement of a Dialysis Access

High Blood Pressure in People with Diabetes:

Blood Vessels and Circulation

Section Four: Pulmonary Artery Waveform Interpretation

CARDIAC REHABILITATION HOME EXERCISE ADVICE

PRACTICE GUIDELINE TITLE: INTRAVENOUS LINE INSERTION: PERIPHERAL AND CENTRAL

Blood Pressure and Your Health

HEALTH MANAGEMENT PLAN PROGRAMME

Blood Pressure Tester. Initial Project and Group Identification EEL 4914 Senior Design I Fall 2011

Vtial sign #1: PULSE. Vital Signs: Assessment and Interpretation. Factors that influence pulse rate: Importance of Vital Signs

Edwards FloTrac Sensor & Edwards Vigileo Monitor. Measuring Continuous Cardiac Output with the FloTrac Sensor and Vigileo Monitor

Note: The left and right sides of the heart must pump exactly the same volume of blood when averaged over a period of time

Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back

Upper Arm Blood Pressure Monitor SC Instructions for use

3 rd Russian-Bavarian Conference on Bio-Medical Engineering

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Making Simulation Easier. SimPad System. Laerdal Simulation Solutions. laerdal.com. laerdal.com

Anna M. Lamb DO FACOFP OMED 2010 San Francisco

Factors Affecting Blood Pressure. Vessel Elasticity Blood Volume Cardiac Output

Activity: Build Your Own Lung

Bioinstrumentation. Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University

INSTRUCTION MANUAL HEM-6052

Cardiovascular Disease Risk Factors

Common Causes of Fluctuating Blood Pressure. When Fluctuating Blood Pressure Readings are Normal

UW MEDICINE PATIENT EDUCATION. Aortic Stenosis. What is heart valve disease? What is aortic stenosis?

Distance Learning Program Anatomy of the Human Heart/Pig Heart Dissection Middle School/ High School

Peripherally Inserted Central Catheter (PICC) for Outpatient

MEDICAL ASSISTING CLINICAL

9/19/2013. Standardizing Preventive Care for Hypertension. About the Speaker. About the Speaker

huidige en toekomstige hemodynamische bewaking

Efficient Heart Rate Monitoring

INSTRUCTION MANUAL. 7 series Blood Pressure Monitor with ComFit TM Cuff. Model BP760 ESPAÑOL ENGLISH

Best Practices in Hypertension

Percutaneous Transluminal Angioplasty (PTA) and Stenting For PVS Patients

MICROGRAVITY EFFECTS ON HUMAN PHYSIOLOGY: CIRCULATORY SYSTEM

Catheter Reduction Program: Creating the Ideal Vascular Access Culture. Presented by: Diane Peck, RN, CNN

HIGH BLOOD PRESSURE AND YOUR KIDNEYS

P R E S S U R E P O I N T S S E R I E S : Measuring your blood pressure at home

All About Your Peripherally Inserted Central Catheter (PICC)

April 3, Dear Autumn Liu,

INSTRUCTION MANUAL. Ultra Premium Blood Pressure Monitor featuring ComFit TM Cuff. Model HEM-775

HEART HEALTH WEEK 3 SUPPLEMENT. A Beginner s Guide to Cardiovascular Disease HEART FAILURE. Relatively mild, symptoms with intense exercise

HAND-ARM VIBRATION SYNDROME ASSESSMENT TESTS

Cardiac Catheterization

Effects of Caffeine on Cardiac and Skeletal Muscle Stimulation: A Noninvasive Study Based on a Single Dose of Caffeine

A Comparison of Forearm and Upper Arm Blood Pressure Measurements in a Sample of Healthy Young Adults


High Blood Pressure and Your Kidneys

Upper Extremity Arterial Duplex Evaluation

Human ECG Laboratory Experiment By

Healthy Blood Pressure Healthy Heart Beat. Initiated by the World Hypertension League

INSTRUCTION MANUAL. 10 series Blood Pressure Monitor with ComFit TM Cuff. Model BP785 ESPAÑOL ENGLISH

Variability of blood pressure No matter which measurement device is used, blood pressure will always be a variable haemodynamic

WRIST BLOOD PRESSURE MONITOR

Dressing and bandage

Training and Certification of Blood Pressure Observers GARY R. CUTTER, PH.D., AND C. MORTON HAWKINS, SC.D.

Transcription:

Principles and Techniques of Blood Pressure Measurement Introduction Blood pressure assessment is an integral part of clinical practice. Routinely, a patient s blood pressure is obtained at every physical examination, including outpatient visits, at least daily when patients are hospitalized before and during most medical procedures. Blood pressure, sometimes referred as arterial blood pressure, is the pressure exerted by circulating blood upon the walls of blood vessels, and is one of the principal vital signs. During each heartbeat, blood pressure varies between a maximum systolic and a minimum diastolic pressure. The blood pressure in the circulation is principally due to the pumping action of the heart. Blood pressure is recorded as two numbers, such as 120/80. The larger number indicates the pressure in the arteries as the heart pumps out blood during each beat. This is called the systolic blood pressure. The lower number indicates the pressure as the heart relaxes before the next beat. This is called the diastolic blood pressure. Blood Pressure Measurement Two methods for measuring a blood pressure exist, the direct ( invasive ) and indirect (noninvasive) method. The direct method is the criterion standard and consists of using an intra-arterial catheter to obtain a measurement. This method, however, is not practical due to its invasiveness and its inability to be applied to large groups of asymptomatic individuals for hypertension screening. Therefore, the indirect (noninvasive) method is typically used. The indirect method involves collapsing the artery with an external cuff, providing an inexpensive and easily reproducible way to measure blood pressure. The indirect method can be performed using a manual cuff and sphygmomanometer or with an automated oscillometric device. The manual method requires auscultation of the blood pressure, whereas the automated system depends on oscillometric devices. Usually the auscultatoric measurement is preferred by doctors, the oscillometric method is mainly used in home blood pressure measurement devices. 1

Auscultatoric Measurement Devices Auscultatoric measurement devices determine blood pressure by using the sphygmomanometer. A sphygmomanometer, an instrument that measures pressure, is needed in both methods. Each sphygmomanometer consists of a cuff (containing a "bladder") which is connected by lengths of tubing to an inflating bulb with a needle valve and to an aneroid pressure gauge as showed in figure 1. Figure 1: Sphygmomanometer with aneroid pressure gauge 1.1 Positioning of The Cuff The center of the sphygmomanometer bladder should be placed over the brachial artery. Many cuffs have some sort of marking scheme so that placement over the brachial artery is under, or just medial to the biceps tendon is facilitated [2]. The lower border of the cuff should be about 2cm proximal to the antecubital fossa and the cuff should be firmly wrapped around the arm. Figure 2 show the position of checking the brachial artery. 2

Figure 2: Checking the position of brachial artery The subject will be asked to relax and sits on a chair with the lower arm supported as in Figure 3. The blood pressure cuff is placed on the subject's right arm, allowing 1 inch between the bottom of the cuff and the crease of the elbow as show in figure 3. The brachial pulse is palpated just above the angle of the elbow (the "antecubital fossa"). Figure 3: The blood pressure cuff is placed on the subject s right arm Figure 4 show the diaphragm is placed over the brachial artery in the space between the bottom of the cuff and the crease of the elbow. At this point no sounds should be heard. The cuff pressure is inflated quickly to a pressure about 30 mm Hg higher than the systolic pressure determined by the method of palpation [3]. Then the air is let out of the cuff at a rate such that cuff pressure falls at a rate of about 5 mm Hg/sec. 3

Figure 4: The diaphragm is placed over the brachial artery in the space between the bottom of the cuff and the crease of the elbow At some point the person listening with the stethoscope will begin to hear sounds with each heartbeat. Then the pressure in the cuff is slowly released. When blood starts to flow in to the artery, the turbulent flow creates a pulse synchronic pounding (first Korotkoff sound). Korotkoff sound is the arterial sounds heard through a stethoscope applied to the brachial artery distal to the cuff of a sphygmomanometer that change with varying cuff pressure and that are used to determine systolic and diastolic blood pressure. The pressure at which this sound is first detected is the systolic blood pressure. The cuff pressure is further released until no more sound can be detected at the diastolic arterial pressure. The laminar flow that normally occurs in arteries produces little vibration of the arterial wall and therefore no sounds. However, when an artery is partially constricted, blood flow becomes turbulent, causing the artery to vibrate and produce sounds. The figure 5 show between two different blood flow : laminar flow and turbulent flow, Figure 5:Laminar flow and turbulent flow 4

When measuring blood pressure using the auscultation method, turbulent blood flow will occur when the cuff pressure is greater than the diastolic pressure and less than the systolic pressure. The "tapping" sounds associated with the turbulent flow are known as Korotkoff sounds. These sounds are not same with the heart sounds produced by the opening and closing of the heart valves. Oscillometric Measurement Devices Blood pressure monitor operation is based on the oscillometric method. This method takes advantage of the pressure pulsations taken during measurements. Oscillometric measurement devices use an electronic pressure sensor with a numerical read out of blood pressure [3]. In most cases the cuff is inflated and released by an electrically operated pump and valve, which may be fitted on the wrist (elevated to heart height), although the upper arm is preferred. Initially the cuff is inflated to a pressure in excess of the systolic arterial pressure, and then the pressure reduces to below diastolic pressure. Once the blood flow is present, but restricted, the cuff pressure will vary periodically in synchrony with the cyclic expansion and contraction of the brachial artery. The values of systolic and diastolic pressure are computed from the raw data, using an algorithm. Conclusion Initially the cuff is inflated to a level higher than the systolic pressure. Thus the artery is completely compressed, there is no blood flow, and no sounds are heard. The cuff pressure is slowly decreased. At the point where the systolic pressure exceeds the cuff pressure, the Korotkoff sounds are first heard and blood passes in turbulent flow through the partially constricted artery. Korotkoff sounds will continue to be heard as the cuff pressure is further lowered. However, when the cuff pressure reaches diastolic pressure, the sounds disappear. Now at all points in time during the 5

cardiac cycle, the blood pressure is greater than the cuff pressure, and the artery remains open. Prepared by, Wee Soo Lee CMET Date: 23 October 2013 REFERENCE [1]Caro, Colin G. (1978). The Mechanics of The Circulation. Oxford [Oxfordshire]: Oxford University Press. ISBN 0-19-263323-6. [2]Jump up ^ Klabunde, Richard (2005). Cardiovascular Physiology Concepts. Lippincott Williams & Wilkins. pp. 93 4. ISBN 978-0-7817-5030-1. [3]Jump up to: a b "Understanding blood pressure readings". American Heart Association. 11 January 2011. Retrieved 30 March 2011. 6