Absorbed Dose Determination in Small Fields of High Energy Photon Beams

Similar documents
RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Challenges in small field MV photon dosimetry

High-Energy Electron Beam Therapy Dosimetry with Ionisation Chambers

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

Introduction to the Monte Carlo method

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Manual for simulation of EB processing. Software ModeRTL

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

PARTICLE THERAPY QA TOOLS. A complete range of tools for absolute dosimetry, machine QA and patient plan verification

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Process Optimization in LINAC QA

Clinical Photon Beams

Chapter XIII. PDD TAR TMR Dose Free Space

Use of 3D Printers in Proton Therapy

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

7üVNORD. TÜV NORD Röntgentechnik

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Proton tracking for medical imaging and dosimetry

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Qualitätsmanagement-Handbuch

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

RITTER Multiple Sonic Nozzle Calibration System

Chapter 16. Mensuration of Cylinder

SURFACE AREA AND VOLUME

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Removing chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches.

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

Sound Power Measurement

Material test chamber with mechanical convection

Atomic and Nuclear Physics Laboratory (Physics 4780)

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES

Quality Assurance for Treatment Planning Systems

THE DOSIMETRIC EFFECTS OF

Modern Physics Laboratory e/m with Teltron Deflection Tube

MECHANICAL PRINCIPLES HNC/D MOMENTS OF AREA. Define and calculate 1st. moments of areas. Define and calculate 2nd moments of areas.

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

R-Value and Thermal Conductivity of PEX and PE-RT TR-48/2014

The feasibility of a QA program for ISIORT Trials

Worldwide Quality Assurance networks for radiotherapy dosimetry

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

THE DØ Luminosity Monitor (LM) consists of two array of

The Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS)

INTRODUCTION. A. Purpose

Calorimetry in particle physics experiments

Fundamentals of Cone-Beam CT Imaging

APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-energy X-Rays (APMP.RI(I)-K3)

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

MECHANICS OF SOLIDS - BEAMS TUTORIAL TUTORIAL 4 - COMPLEMENTARY SHEAR STRESS

Physics 221 Experiment 5: Magnetic Fields

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

R&S R-Line Compact Test Chamber Precise measurements of radiated emissions of wireless terminals 800 MHz to 18 GHz

External dosimetry Dosimetry in new radiotherapeutic techniques

METHODS OF BOLUSING THE TRACHEOSTOMY STOMA

arxiv:astro-ph/ v1 15 Sep 2005

Pipeline External Corrosion Analysis Using a 3D Laser Scanner

Quality Assurance of Radiotherapy Equipment

CHAPTER 29 VOLUMES AND SURFACE AREAS OF COMMON SOLIDS

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July Reprinted from Photonics Spectra, Laurin Publishing

DEUTSCHE NORM June Forms and dimensions of undercuts

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Edmund Li. Where is defined as the mutual inductance between and and has the SI units of Henries (H).

Ampere's Law. Introduction. times the current enclosed in that loop: Ampere's Law states that the line integral of B and dl over a closed path is 0

Variance reduction techniques used in BEAMnrc

A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS

Monte Carlo Simulations in Proton Dosimetry with Geant4

Perturbation correction factors for the NACP-02 plane-parallel ionization chamber in water in high-energy electron beams

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

On the first Townsend coefficient at high electric field

Quality Assurance in Stereotactic. Radiotherapy

SAR Distribution in Test Animals Exposed to RF Radiation

The accurate calibration of all detectors is crucial for the subsequent data

Mensuration. The shapes covered are 2-dimensional square circle sector 3-dimensional cube cylinder sphere

Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters

GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2

Automatic and Objective Measurement of Residual Stress and Cord in Glass

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

DRAFTING MANUAL. Gears (Bevel and Hypoid) Drafting Practice

Solid State Detectors = Semi-Conductor based Detectors

Status and perspective of emission imaging techniques for ion beam therapy in Lyon

Measurement of Length, Mass, Volume and Density

IDEAL AND NON-IDEAL GASES

Conversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Transcription:

Absorbed Dose Determination in Small Fields of High Energy Photon Beams Based on the German Standard DIN 6809 8 1. Introduction This document constitutes an excerpt of procedures and data as published in the German standard DIN 6809 8. The aim is to determine absorbed dose to water in the center of small high energy photon fields with dimensions of 4 cm x 4 cm or smaller. Although this document provides the reader with a concise overview of equations and factors it shall not replace pertinent standards, protocols and publications, nor is it intended to give all of the details that are important for accurate dosimetry. Also, the procedures outlined in this document are not the only ones described in the referenced literature, they constitute only one of several possibilities for absorbed dose determination in small fields. 2. General Instructions 2.1 Measuring phantom This document assumes that all measurements are made in a water phantom. 2.2 Detector types Dose measurements in a 4 cm x 4 cm field shall be made with a cylindrical vented ionization chamber that was calibrated in a 60 Co beam. Dose measurements at field sizes 4 cm x 4 cm shall be made with detectors featuring a high spatial resolution such as small size ionization chambers, silicon diodes or diamond detectors. In this document such detectors are referred to as small field detectors. 2.3 Detector positioning General rules The dosemeter reading is obtained by positioning the detector s effective point of measurement at the point of interest in the phantom. The effective point of measurement may not coincide with the reference point of the detector. The reference point is defined by the manufacturer of the detector and serves for positioning the detector in the phantom. Usually, the reference point is a point on the central axis of a cylindrical detector with a stated distance from the detector tip. For solid state detectors such as silicon diodes or diamond detectors the manufacturer may select the effective point of measurement as the reference point. D560.210.03/02 Schü 2014 05 1

For cylindrical ionization chambers the effective point of measurement is defined as a point shifted by 0.5 from the axis of the chamber towards the focus ( is the inner radius of the measuring chamber volume) [DIN 6800 2]. To measure dose at a focus distance of cm, the axis of the cylindrical chamber must be positioned at a focus distance of cm + 0.5, i. e. the chamber must be shifted away from the focus, i.e. downstream. The location of the effective point of measurement of solid state detectors can be taken from the data sheet of the detector. The TRUFIX system The task of positioning various types of radiation detectors precisely in their effective point of measurement can be quite challenging. The patented TRUFIX system (see Figure 1) facilitates this task considerably. TRUFIX can be used on automated PTW water phantoms (MP2, MP3 etc.) in connection with most PTW therapy detectors. A plastic tip lets you easily locate the water surface the coordinate system is set to (0,0,0). Then the plastic tip is replaced by a holding device specific to each detector type, and the effective point of measurement is automatically placed at the tip's earlier position. The radius of cylindrical chambers, the water equivalent window thickness of plane parallel chamber and solid state detector windows as well as the detector centers are automatically accounted for. Figure 1: The TRUFIX detector positioning system. 2.4 Corrected reading In this document detector readings are designated as corrected readings. For ionization chambers the reading of the dosemeter must be corrected for influence quantities as described in [DIN 6800 2]. The reader must compute the corrected reading from the uncorrected reading and the reading without irradiation by (2 1) The correction factors,, and correct for air density, incomplete charge collection, polarity effect and humidity, respectively. They are described in detail in [DIN 6800 2]. For solid state detectors some of the correction factors in equation (2 1) may not apply. For absorbed dose determination in small fields, additional factors are to be applied to the corrected reading as described in the following chapters. D560.210.03/02 Schü 2014 05 2

3. Formalism for dose measurements in small fields 3.1 Overview The determination of absorbed dose to water in small fields consists of three steps: (1) Measure the absorbed dose to water in a 4 cm x 4 cm field using a vented ionization chamber calibrated in a 60 Co beam. Make sure the ionization chamber is not too large for the field. See chapter 3.2. (2) Apply the same dose in the 4 cm x 4 cm field to a detector with high spatial resolution ( small field detector ), e.g. a silicon diode, diamond detector or PinPoint chamber, and calculate the cross calibration factor for the small field detector by referring to the dose measured in (1). See chapter 3.3. (3) Use the small field detector for dose measurements in fields smaller than 4 cm x 4 cm, applying the correction factors described in chapter 3.4. 3.2 Step 1: Ionization chamber measurement at 4 cm x 4cm The absorbed dose shall be measured in a water phantom at a depth of 10 cm with a source tosurface distance of 100 cm. The field size is definded at the measuring depth. If a filed size of 4 cm x 4 cm cannot be adjusted any filed size between 3.5 cm x 3.5 cm and 5 cm x 5 cm is allowed. Dose homogeneity within 0.5 cm distance from the beam axis must be good enough to ensure that dose differences do not exceed ± 0.5 %. The absorbed dose to water, measured with a vented cylindrical ionization chamber, is determined by the following equation:,, (3 1),, absorbed dose to water = 1 0,03 (replacement correction factor, is the inner radius of the measuring volume of the cylindrical chamber, given in cm) energy dependant correction factor accounting for measuring in a high energy photon beam instead of a 60 Co beam. This factor can be taken from [DIN 6800 2] and is most accurate for the reference conditions described therein, especially for a field size of 10 cm x 10 cm. Note that this factor is denoted by in [DIN 6800 2]. correction factor accounting for the change of radiation quality when measuring in a 4 cm x 4 cm field instead of a 10 cm x 10 cm field correction factor accounting for the size of the ionization chamber, causing a volume effect in a 4 cm x 4 cm radiation field, see Table 1 calibration factor for absorbed dose to water for 60 Co corrected reading of the dosemeter as described in chapter 2.4 D560.210.03/02 Schü 2014 05 3

PTW chamber type at 6 MV at 15 MV PinPoint 31015 (0.03 cm³) 1.000 1.001 Semiflex 31010 (0.125 cm³) 1.001 1.002 Semiflex 31013 (0.3 cm³) 1.004 1.008 Table 1: Correction factor for ionization chamber measurements in a 4 cm x 4 cm field, for 6 MV and 15 MV. Data is taken from Table 3 and equation (C.11) in [DIN 6809 8]. In order to determine the correction factor, the radiation quality index must be known. can be obtained by depth dose measurements in a 10 cm x 10 cm field at depths of 10 cm and 20 cm according to the following equation [DIN 6800 2]: 1.2661 0.0595 (3 2) 10, 20 radiation quality index dose values at 10 cm and 20 cm depth, respectively. Source to surface distance 100 cm, field size at the phantom surface 10 cm x 10 cm. If a field size of 10 cm x 10 cm cannot be adjusted for the determination of the quality index, the quality index is determined by [DIN 6809 8]:, (3 3) TPR 20,10 s t Tissue phantom ratio 20 / 10 at 100 cm source to detector distance field dimension in one direction, in cm = 19.5 cm A 1 = 0.625 A 2 = 0.679 b 1 = 0.208 b 2 = 1.213 Equation (3 3) is valid for 0.62 < < 0.8 and field dimensions 0.4 cm. For circular fields the field dimension can be taken as 0.9, is the diameter of the circular field. D560.210.03/02 Schü 2014 05 4

Hint: DIN 6800 2 offers an iteration process to increase the accuracy of equation (3 3). However, the increase of accuracy is smaller than the uncertainties of the parameters in equation (3 3). Therefore, an iteration process is not suggested in this document. The correction factor, is taken from [DIN 6800 2]. Note that the notation in DIN 6800 2 is. In a 4 cm x 4 cm field, the correction factor, can be set to 1.000 for both beams with and without flattening filter. 3.3 Step 2: Cross calibration of a small field detector in a 4 cm x 4 cm field Detectors with high spatial resolution ( small field detectors ) are calibrated in a 4 cm x 4 cm field for which the dose has been determined with a calibrated ionization chamber as described in chapter 3.2. If is the corrected reading of the small field detector (see chapter 2.4) and the dose measured with the calibrated ionization chamber, the calibration factor for the small field detector is / (3 4) This calibration factor is valid for the radiation quality and the measuring depth used for the crosscalibration, and for a field size of 4 cm x 4 cm. Deviations from these calibration conditions are accounted for by applying correction factors to the reading as described in chapter 3.4. Hint: If PinPoint chambers are used as small field detectors, and if a calibration factor at 60 Co as well as, factors are known, a cross calibration may not be necessary down to a field size of 1.5 cm x 1.5 cm. 3.4 Step 3: Measurements with small field detectors at field sizes 4 cm x 4 cm The absorbed dose to water, measured with a detector of high spatial resolution ( small field detector ), is determined by (3 5) D W absorbed dose to water correction factor accounting for variation of detector response when changing the field size from 4 cm x 4 cm to smaller fields, see Tables 2a and 2b correction factor for the size of the small field detector, causing a volume effect in small fields, see Tables 3a and 3b calibration factor for the small field detector under cross calibration conditions, see chapter 3.3 corrected reading of the dosemeter as described in chapter 2.4 Hint: Due to their size Semiflex chambers with a volume of 0.3 cm³ or 0.125 cm³ are not recommended for dosimetry at field sizes smaller than 4 cm x 4 cm. D560.210.03/02 Schü 2014 05 5

PTW detector type = 2.5 cm = 4 cm = 5 cm Semiflex 31010 (0.125 cm³) 1.007 1.000 1.007 PinPoint 31014 (0.015 cm³) 1.006 1.000 1.007 PinPoint 3D 31016 (0.016 cm³) 1.005 1.000 1.006 Dosimetry Diode P 60016 1.002 1.000 1.003 Dosimetry Diode 60017/60018 (PTW) 1.001 1.000 0.998 Dosimetry Diode 60017/60018 (MC) 1.006 1.000 1.003 Diamond Detector 60003 1.000 1.000 1.000 microdiamond 60019 1.000 1.000 1.000 Table 2a: Correction factor for measurements in the center of small fields at 10 cm depth. The field dimensions are designated by, the radiation quality is 6 MV photons. Data is taken from Table E.1 in [DIN 6809 8]. All values are based on Monte Carlo simulations with the exception of the line labeled (PTW). The values for the diamond detectors are preliminary as no measurements or simulations have been published so far. PTW detector type = 2.5 cm = 4 cm = 5 cm Semiflex 31010 (0.125 cm³) 1.006 1.000 0.996 PinPoint 31014 (0.015 cm³) 1.003 1.000 0.996 PinPoint 3D 31016 (0.016 cm³) 1.003 1.000 0.995 Dosimetry Diode P 60016 1.002 1.000 0.991 Dosimetry Diode 60017/60018 (PTW) 1.000 1.000 0.997 Dosimetry Diode 60017/60018 (MC) 1.003 1.000 0.991 Diamond Detector 60003 1.000 1.000 1.000 microdiamond 60019 1.000 1.000 1.000 Table 2b: Correction factor for measurements in the center of small fields at 20 cm depth. The field dimensions are designated by, the radiation quality is 6 MV photons. Data is taken from Table E.2 in [DIN 6809 8]. All values are based on Monte Carlo simulations with the exception of the lined labeled (PTW). The values for the diamond detectors are preliminary as no measurements or simulations have been published so far. PTW detector type = 1 cm = 2 cm = 3 cm = 4 cm Semiflex 31013 (0.3 cm³) 1.570 1.067 1.012 1.004 PinPoint 1.076 1.007 1.001 1.000 Table 3a: Correction factor for 6 MV photons as a function of field dimension. Data is taken from Figure C.3 in [DIN 6809 8]. PTW detector type = 1 cm = 2 cm = 3 cm = 4 cm Semiflex 31013 (0.3 cm³) 1.543 1.091 1.023 1.008 PinPoint 1.075 1.009 1.002 1.001 Table 3b: Correction factor for 15 MV photons as a function of field dimension. Data is taken from Figure C.3 in [DIN 6809 8]. D560.210.03/02 Schü 2014 05 6

References [DIN 6800 2] [DIN 6809 8] DIN 6800 2, Deutsche Norm, Dosismessverfahren nach der Sondenmethode für Photonen und Elektronenstrahlung, Teil 2: Dosimetrie hochenergetischer Photonenund Elektronenstrahlung mit Ionisationskammern, März 2008 E DIN 6809 8, Entwurf zur Deutschen Norm, Klinische Dosimetrie Teil 8: Dosimetrie kleiner Photonen Bestrahlungsfelder, März 2014 PTW FREIBURG Physikalisch Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055 0 Fax: +49 761 49055 70 info@ptw.de www.ptw.de D560.210.03/02 Schü 2014 05 7