Installation and User Guide. UpIR Out-of-Compartment Diffuse Reflectance Accessory

Similar documents
Installation and User Guide. RotatIR Automated Rotating Sample Stage Accessory

Introduction to Fourier Transform Infrared Spectrometry

QUANTITATIVE DETERMINATION OF COMMON TYPES OF ASBESTOS BY DIFFUSE REFLECTANCE FTIR

Wall-Mounting your HP computer. User Guide

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Elo Touch Solutions Wall-mounting Kit for the 5501L IDS Touchmonitors

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory

Dual Laser InfraRed (IR) Thermometer

Dual Laser InfraRed (IR) Thermometer with Color Alert

Char-Lynn Spool Valve Hydraulic Motors. Repair Information. W Series Geroler Motors

WM002 LCD Display Wall Mounting Kit Installation Guide. Welcome

CHARISMA SHOWER/TUB DOOR INSTALLATION INSTRUCTIONS

Installation Instructions / Warranty

Features: Mounting the Optic

Infrared Viewers. Manual

SEQUEL ASSEMBLY INSTRUCTIONS Compact Desk

4RU High Density Shelf Installation Instructions

Experiment 5. Lasers and laser mode structure

TABLE OF CONTENTS 11.0 WARRANTY

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes

RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G

Compaq Presario Desktop Products. Upgrading and Servicing Guide

Universal Vehicle Power Supply 9007AX01. Installation Instructions

Triple Threat 3-in-1 Game Table 3 IN 1 GAME TABLE

INSTALLATION INSTRUCTIONS for Bifold Doors (JII103)

Hygro-Thermometer + InfraRed Thermometer Model RH101

Surface Mount LEDs - Applications Application Note

el-720 Carbon Dioxide Gas Leak Detector User s Guide

2-3 SAS/SATA II HDD Canister USER S MANUAL XC-23D1-SA10-0-R. Document number: MAN A

Lunette 2 Series. Curved Fixed Frame Projection Screen. User s Guide

V-500-PRE. 500mm F/8.0 Telephoto Lens INSTRUCTION MANUAL

Round Housing with Side Ports

Owner s Guide and Installation Manual. Vancouver Model Name , Model No. English Español

Mini-led spotlight with magnetic base

INSTALLATION AND SERVICE MANUAL

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions

E8887 First Edition February 2014

ATS Overhead Table Shelf System INSTRUCTION MANUAL

ROUTER TABLE INSERT ASSEMBLY AND OPERATING INSTRUCTIONS

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

EnduraLAM 150 User Manual

ANI-64S-xRU. Single Section Wall Cabinet INSTRUCTION MANUAL

MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC

Alignement of a ring cavity laser

OPERATION MANUAL INFRARED THERMOMETER

FTW 175 VERIZON Wireless Monitoring System Reference Manual Part Number VERIZON

Detector transparent with Color Inserts. FAA 500 TR P Trim Ring transparent with Color Inserts. FCA 500 / FCA 500 E Detector Bases

Strut Spring Compressor

Installation and Operation Manual Document Number 000-vivo

OAK WORKBENCH WITH 2 DRAWERS

Installation Instructions / Warranty

Tipping Bucket Rain Gauge Models 6011-A 6011-B

MODEL T10023 ACCESSORY KIT #1 FOR T10010/T10097 WET GRINDER

Chapter 4 COATINGS Full Reflective Coatings:

Installation Instructions / Warranty

lamp post light Set up and Operating Instructions

1500 Follow Spot Yoke

ST-80 MICROSCOPE With Electronic Eyepiece #6810. User Guide

High Temperature InfraRed Thermometer with Laser Pointer

F-Series Camera Mounting Accessories

GENUINE PARTS INSTALLATION INSTRUCTIONS

LUCCI AIRFUSION QUEST II CEILING FAN

P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)

Installation Guide. WSD-100 Wind Speed and Direction Sensor For XR5 Data Loggers. February, 2011

M A R C O. CP-670 AUTOMATIC CHART PROJECTOR Instruction Manual

Measurement & Analytics Measurement made easy

Survival Laser SL-001BB Laser Parts Bundle Assembly & Operation Instructions

Express5800/120Ed. Rack Mount Kit Installation Procedures PN:

LED Security Spotlight User Manual

DUBOX TM LOOSE PIECE HAND TOOL

Dual Laser InfraRed (IR) Thermometer

AM/FM ANTENNA RELOCATION KIT

Infrared Optical Fiber. Datasheets and Price list. JTIngram Sales and Marketing

FirstView 3 Reflector Telescope Owner s Manual

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

Model: HD11-ARC. Installation Guide

Dual Laser InfraRed (IR) Thermometer

CAST IRON THE BASICS. Heatline - Cast Iron Radiators SMOOTH FLAT FILE TO REMOVE ANY SWARF. ONE TIME. ASSEMBLY. JOINTS SHOULD BE TIGHTENED.

Installing the Broadband Global Area Network (BGAN) Fixed Mount Kit

Owner s Manual Read and keep this manual. Patents World Wide

SC, FC and ST * Hot Melt Fiber Optic Connectors

Arctic Leash Instruction Manual Alaskan Products Company LLC.

Owner s Manual & Safety Instructions

Optical Lab Room 1122 Nexus 670 and Magna 560 Instructions Feb. 26, 2013

How To Use A Ds340 (Dsp)

81000 Series Aluminum Gate Valve Installation and Repair Manual

Pet hair clipper. Model Diagrams within this manual may not be drawn proportionally.

3/4 PIPE CLAMP WITH STAND

Curved Quadrant Instruction Manual

USB-ISS-UV/VIS Integrated Sampling System Installation and Operation Instructions

Retractable Screen. Replacement Screen Instructions

FTIR and UV-Vis Accessories and Supplies

Accessory Manual Ventilation Section Model Comfort Circle

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions

Digital Fingerprint safe

TL-12 Passenger Car Alignment System

CETAC Z-Drive Assembly

Transcription:

Installation and User Guide UpIR Out-of-Compartment Diffuse Reflectance Accessory

The information in this publication is provided for reference only. All information contained in this publication is believed to be correct and complete. PIKE Technologies, Inc. shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this material. All product specifications, as well as the information contained in this publication, are subject to change without notice. This publication may contain or reference information and products protected by copyrights or patents and does not convey any license under the patent rights of PIKE Technologies, Inc. nor the rights of others. PIKE Technologies, Inc. does not assume any liability arising out of any infringements of patents or other rights of third parties. This document contains confidential or proprietary information of PIKE Technologies, Inc. Neither this document nor the information herein is to be reproduced, distributed, used or disclosed, either in whole or in part, except as specifically authorized by PIKE Technologies, Inc. PIKE Technologies, Inc. makes no warranty of any kind with regard to this material including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Copyright 1991-2013 by PIKE Technologies, Inc., Madison, WI 53719. Printed in the United States of America. All world rights reserved. No part of this publication may be stored in a retrieval system, transmitted, or reproduced in any way, including but not limited to, photocopy, photograph, magnetic or other record, without the prior written permission of PIKE Technologies, Inc. Address Comments to: PIKE Technologies, Inc. 6125 Cottonwood Drive Madison, WI 53719 Phone (608) 274-2721 Fax (608) 274-0103 E-mail info@piketech.com Web Site www.piketech.com Jan. 1, 2013

Contents Introduction 1 Unpacking Your Accessory 2 Packing List 2 Optical Description 3 Installation and Alignment 4 Preparing the Spectrometer 4 Preparing the Accessory 5 Alignment of the Accessory 5 Sample Placement 6 Typical Spectra 7 Precautions 8 Replacement Parts and Options 8

Introduction The PIKE Technologies Upward Looking Diffuse Reflection Accessory is an ideal device to measure the wide angle reflectance of samples with an easy face-down sample placement. The design employs a high efficiency fixed ellipsoidal reflector to collect the maximum amount of diffusely reflected energy from the sample. The optical elements of the accessory are all reflective, thus the spectral range covers both the mid and near-infrared range. The 044-10XX series UpIR has aluminum reflective elements. The optical elements of the 044-60XX series UpIR are gold coated to provide the maximum signal possible for both mid and near IR applications. The sample position is on the top surface of the accessory. The sample position can be covered by an optional IR or NIR transparent window, upon which a sample may be placed for the analysis of powders. An optional vial holder is available for the analysis of the Upward Looking Diffuse Reflection Accessory is available for most FT-IR Spectrometers. Sample Position Sample Focus Adjustment Thumb Screws (to remove front panel) Purge Seals to Interferometer Purge Tubing Connection Figure 1. The UpIR Accessory PN 350-044000-02 Page 1

Unpacking Your Accessory In order for you to quickly verify receipt of your accessory, we have included a packing list. Please inspect the package carefully. Contact PIKE Technologies if any discrepancies are noticed. Packing List User Manual UpIR Accessory Gold-Coated Alignment Mirror PN 350-0044000 PN 044-10XX PN 300-0002 Quantity 1 Quantity 1 Quantity 1 Window Mount PN 044-3030 Powders Sampling Insert with ZnSe Window Quantity 1 PN 044-3040 Quantity 4 Quantity 1 Sample Masks, various sizes PN 044-3050 Purge Tubing Kit Purge Seals Hex Wrenches PN 025-3055 PN 055-0020, 055-0021 Quantity 1 Set Quantity 1 Kit Quantity 8 PN 350-044000-02 Page 2

Optical Description The optical system of the Upward Looking Diffuse Reflection Accessory is symmetrical with identical mirrors being used for the optical path from the spectrometer interferometer to the sample and the sample to the spectrometer detector. Two flat mirrors and an ellipsoidal mirror are used in each path. A diagram of the optical path is shown below. The two lower mirrors are adjustable and, if necessary, are used for aligning the optical system during installation. The upper flat mirrors and ellipsoid mirror are fixed in position. The ellipsoid is a large solid angle monolithic mirror which focuses the beam onto the sample with a power of 3X. This means that if the beam in your spectrometer sample compartment with no accessory in place is 9mm in diameter, then the size of the focused spot at the sample in the Upward Looking Diffuse Reflection Accessory will be 3 mm. Note that the size of the beam in the spectrometer is dependent on manufacturer. Refer to your spectrometer user guide for the correct size. Sample Position Figure 2. Optical diagram of the UpIR accessory PN 350-044000-02 Page 3

Installation and Alignment Inspect the drawing below and locate the indicated items on your accessory. Sample Position Micrometer Focus Adjustment Purge Tube Purge Tube Mirror Alignment Screws Figure 3. Front interior view of the UpIR To install the accessory in your sample compartment perform the following: Preparing the Spectrometer 1. Remove any sample holder present in the sample compartment. 2. Scan a background with your FTIR software and save either in memory or on disk, depending on the software you are using. This single beam spectrum can be used to verify the performance of the system. PN 350-044000-02 Page 4

Preparing the Accessory 1. Place the accessory into the sample compartment. Using the 3/32 wrench provided, loosen the screws on the two purge seal tubes located on the sides of the accessory. Push the tubes toward the side walls of the spectrometer. This will seal the accessory in the sample compartment. Purge tubing provided with the instrument can be connected on the front of the accessory for purging the accessory housing. 2. Place the accessory in the sample compartment. Some configurations of the accessory include their own instrument sample compartment mounting plate. This plate may be secured in place without requiring access to the inside of the accessory. 3. If your accessory does not contain an instrument sample compartment baseplate, remove the front cover of the accessory. This cover is held in place by six thumbscrews. Inside the accessory are two mirror mounts and a base fixing position. Tighten the screw in place. Alignment of the Accessory 1. Place the gold-coated alignment mirror onto the top surface of the accessory with the mirror surface facing down. 2. Set up the FTIR software to alignment or monitor mode and note the size of the interferogram. Turn the focus adjust micrometer to raise and lower the sample stage to achieve maximum throughput. The gap between the sample plate and body of the accessory should be about ¼. T Figure 4. Typical throughput curve obtained with a gold mirror as sample. The background was the instrument open beam without any accessories. Typical throughput of a well-aligned UpIR should be better than 40% T. PN 350-044000-02 Page 5

If the throughput of the accessory is above 40%T at 1000 cm -1, the alignment is acceptable. If the throughput is less than 40%T, the mirrors inside the accessory should be adjusted: 1. Remove the six thumbscrews holding the front cover in place and remove the cover. Behind this cover, there are two adjustable mirrors which are used for alignment. 2. Adjust the output adjustable mirror first. This is the mirror that is the closest to the instrument detector. Turn both screws on this mirror mount to maximize throughput. 3. Adjust the focus micrometer again to maximize throughput. 4. Adjust the input adjustable mirror. This is the mirror that is the closest to the instrument interferometer. Turn both screws on this mirror mount to maximize throughput. 5. Adjust the focus micrometer to maximize throughput. 6. Repeat the last four steps until there is no further increase in signal. Typically, with this alignment procedure you should achieve a signal throughput which is at least 40% of the signal that was present before the accessory was placed into the sample compartment. Replace the front cover and secure in place with the six thumb screws. Your accessory is ready for sample measurements. Sample Placement The UpIR accessory is typically used by placing the sample directly on top of the accessory. When the sample is smaller than the opening on top of the accessory, a sample holder mask should be used. A sample mask is included in your accessory kit. For powders or other samples with smaller particles, a sample cup can be used. A ZnSe powder holder is enclosed with the accessory. Powder holders can be manufactured with different window materials. The standard holder has a ZnSe window. There is an opening with two finger access slots to help insert and remove sample holders. Please make sure that the holders are properly positioned in the opening, so that the sample is situated in the plane of the top of the UpIR top plate. PN 350-044000-02 Page 6

Typical Spectra Using the gold-coated mirror as reference, a matte beige color paint panel was measured. The test panel was placed directly on top of the accessory without any sample holder. Measurement time was less than 1 minute using 4 cm -1 resolution. Figure 5. Spectrum of coated panel collected directly on UpIR accessory For best results, dilute powdered samples in KBr. Caffeine crystals in approximately 1-5% w/w result in excellent diffuse reflectance spectra. The background was the same ZnSe holder (enclosed with the accessory) with KBr. The sample was thoroughly ground in a mortar with an agate pestle and placed on the ZnSe window. The cup does not need to be filled, but the window should be covered for best results. Figure 6. Spectrum of 2% caffeine sample diluted in KBr powder collected on UpIR accessory PN 350-044000-02 Page 7

Precautions In order to provide the maximum transmission in the infrared and near infrared, with the minimum spectral interference, the mirrors used in the 044-60XX version of the UpIR device are uncoated (bare) gold on the mirror substrates. Since the coatings are soft, care must be taken to avoid damage. Normally, these mirrors will not need cleaning since they are contained within the housing of the accessory. If they do need cleaning, use a clean dry supply of air to blow off any dust. Under no circumstances must the mirrors be rubbed with paper products such as Kleenex since this will scratch the mirror coating. Replacement Parts and Options Part Number Description 044-60XX UpIR Accessory with Gold-Coated Optics 044-3030 Solids Sampling Insert 044-3040 Powders Sampling Insert 044-3010 21 mm Glass Vial Holder 044-3020 Sample Vials with Threaded Caps, 21 mm x 70 mm, (200 per package) 160-1155 25 x 2 mm ZnSe Window 160-1307 25 x 2 mm Ge Window 160-1201 25 x 2 mm AMTIR Window 160-5000 25 x 2 mm Sapphire Window 160-8010 KBr Powder, 100 g 161-5050 Mortar and Pestle, 50 mm Some options are available for purchase directly on-line at www.piketech.com. PN 350-044000-02 Page 8

6125 Cottonwood Drive Madison, WI 53719-5120 (608) 274-2721 (TEL) (608) 274-0103 (FAX) info@piketech.com www.piketech.com