How to Pass a CAP Inspection.

Similar documents
2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)

Is Your Lab s Competency Assessment a Competent Assessment?

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

Transfusion Medicine: Checklists and Challenges

College of American Pathologists Accreditation for Public Health Laboratories

General. What is a laboratory test? What is a waived test?

Calibration Verification

Quality Requirements of a Point of Care Testing Service

2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan

Laboratory Computer Systems

How to Verify Performance Specifications

2009 LAP Audioconference Series. Laboratory Computer Systems

Examples: collecting the right sample on the right patient; labeling the sample accurately;

Medical Cannabis Laboratory Approval Program

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

EVALUATION AND ACCREDITATION OF SUPPLIER TEST FACILITIES GP 10

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3

XL-I 2011: Understanding and Explaining the CAP POCT Checklist: A Tool for Laboratorians to Use With POCT Staff

Standards for Laboratory Accreditation

Prepublication Requirements

A Roadmap for Competency Assessment ASCLS-NJ Spring Seminar April 24, 2014

Laboratory Accreditation. Personnel Qualifications. What s New? March 17, 2010

XV.Quality Assurance in the Blood Bank

Internal Audit Checklist

Individualized Quality Control Plan

Answers to Your Questions About Blood Bank Proficiency, Competency and QC Follow Up to April 28, 2016 Webinar

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

TEST METHOD VERIFICATION AND VALIDATION

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

AS9100 Quality Manual

Content Sheet 3-1: Equipment Management Overview

Meaningful Competency Assessment. Kathy Nucifora, MPH, MT(ASCP) COLA 1

Content Sheet 16-1: Introduction to Documents & Records

PERFORMANCE EVALUATION AUDIT CHECKLIST EXAMPLE. EIIP Volume VI

PROCEDURES FOR HANDLING OOS RESULTS

Laboratory Director Responsibilities

Using and Maintaining Documents and Records. Finding the information when you need it

LABORATORY PERSONNEL QUALIFICATION APPRAISAL and APPLICATION FOR LICENSURE

Self-Audit Checklist

Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing

Document and Record Control in a Hospital Clinical Laboratory

Flow Cytometry. CLIA Compliance Manual*

AUDITOR GUIDELINES. Responsibilities Supporting Inputs. Receive AAA, Sign and return to IMS with audit report. Document Review required?

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver

1 Quality Assurance and Quality Control Project Plan

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Checklist. Standard for Medical Laboratory

New York State Department of Health (NYSDOH) Bureau of Immunization Vaccine Storage and Handling Plan

Copies of this document shall be classified either as Controlled or Uncontrolled, and shall be identified as such on the cover sheet.

UCLA PATHOLOGY & LABORATORY MEDICINE QUALITY MANAGEMENT PLAN

CHAPTER 7 QUALITY ASSESSMENT

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

Software Verification and Validation

Record Keeping. Guide to the Standard for Professional Practice College of Physiotherapists of Ontario

Point of Care Testing and Informatics: How to Prepare for the POCT Explosion

Chapter. Guaiac Screening CHAPTER 4: GUAIAC TESTING SCREENING FOR OCCULT BLOOD. Page 1 of 5 Guaiac doc 6/24/2005

Inform II Glucose Meter

Content Sheet 18-1: Organizational Requirements for a Quality Management System

IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED

CHAPTER 6 OUTPATIENT SURGICAL CENTERS 6. 1

AN AUDIT QUESTIONNAIRE THAT EXAMINES SPECIFICALLY THE MANAGEMENT OF TECHNICAL ACTIVITIES CLAUSES IN ISO 15189

Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?

CHAPTER 5 - SAFETY ASSESSMENTS, LOG OF DEFICIENCIES AND CORRECTIVE ACTION PLANS

General Information. Our Background Automated Phone Services Holiday Coverage Licenses & Accreditations

Hemoglobin Testing using the Data Management HemoCue Analyzer

Document Control for Control Freaks. Kristina Martin, MLS(ASCP) CM,MS

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

1.0 INTRODUCTION 1.1 Overview

Managing POCT: Trying to Control Testing in an Out-of-Control Environment. William Clarke, PhD, MBA, DABCC Johns Hopkins School of Medicine 3/24/11

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

UNCONTROLLED COPY FOR REFERENCE ONLY

NIST HANDBOOK CHECKLIST CONSTRUCTION MATERIALS TESTING

NC SBI QUALITY ASSURANCE PROGRAM

Content Sheet 10-1: Overview of External Quality Assessment (EQA)

Installation and Operational Qualification Protocol (Reference: SOP )

U. S. Department of Energy Consolidated Audit Program Checklist 5 Laboratory Information Management Systems Electronic Data Management

Module 5: Assuring the Quality of HIV Rapid Testing

City of De Pere. Halogen How To Guide

Content Sheet 12-1: Overview of Personnel Management

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

SECTION 1.11 RISK MANAGEMENT / QUALITY ASSURANCE POLICY

State Water Resources Control Board Environmental Laboratory Accreditation Program

Recall and Reminder Policy and Procedure Manual Best Practice

BQ-9000 Quality Management System. Marketer Requirements

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

1 For more information T: / E: DAP@ul.com / W: ul.com/dap

The NBT Online Banker PERSONAL FINANCIAL MANAGEMENT TOOL HOW TO GUIDE

Competency Assessment (both of employees and of the laboratory as a whole)

Working Copy - 1,- / t&,1 r. Responsibilities of the Laboratory Quality Assurance Officer. EffectiveDate: 7/ltl/ l'rritials: tuwt

May 2005 Lear Corporation For use by Lear Corporation and its Suppliers only. Copies or reproduction for any other purpose is permitted only by prior

To help you understand the role documents and records play in the quality system and the monitoring of programs.

COORDINATED PROGRAM REVIEW CORRECTIVE ACTION PLAN (To be completed by collaborative)

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

Transcription:

How to Pass a CAP Inspection www.cap.org

Don t Pull out your hair; It s Just Point of Care www.cap.org

Learning Objectives After participating in this session, you will be able to: 1. List the CAP requirements for waived and non waived testing 2. Explain new or changing requirements for POCT 3. Understand how best to prepare your POCT for inspection 4. Recognize common POCT deficiencies and what you need to do to correct. 2010 College of American Pathologists. All rights reserved. 3

Summary of Waived Testing Requirements Quality Control o Follow manufacturer instructions o Verify QC is acceptable before reporting patient results o Document corrective action if QC exceeds tolerance limits o Internal QC results need not be documented for instruments using such controls, if an unacceptable instrument control automatically locks the instrument & prevents the release of patient results. o Frequency of QC is defined by the manufacturer o External controls run as required by manufacturer 2010 College of American Pathologists. All rights reserved. 4

Summary of Waived Testing Requirements Correlation o Initial correlation between waived instruments not required o Correlations between waived instruments & main lab instruments not required o Multi-instrument comparison not required 2010 College of American Pathologists. All rights reserved. 5

Summary of Waived Testing Requirements Method Performance Specifications o Verification not required except for reference range Calibration/Calibration Verification o Follow manufacturer instructions o Calibration verification every 6-month not necessary unless required by manufacturer Analytical Measurement Range (AMR) o Follow manufacturer instructions o Initial AMR validation & 6-month interval validation not necessary unless required by manufacturer 2010 College of American Pathologists. All rights reserved. 6

Summary of Waived Testing Requirements Reagents o Lot-to-lot reagent validation not required o Follow manufacturer instructions for handling & validating Competency o During 1 st yr that an individual is performing patient testing, competency assessed at least semiannually o Competency must at least be reassessed annually 2010 College of American Pathologists. All rights reserved. 7

UNDERSTANDING THE POC CHECKLIST www.cap.org

CAP Point of Care Requirements are the same for waived & nonwaived testing in the following areas: o Proficiency Testing o Instrument maintenance o Procedure manuals o Personnel o Specimen Handling o Results reporting o Safety 2010 College of American Pathologists. All rights reserved. 9

New edition released June 17, 2010 Current edition for each checklist Change document for each checklist showing revisions since previous editions o New checklist questions o Revised questions o Deleted questions o Migrated questions List of new, deleted and revised questions at beginning of each checklist, with date of change (items on list for 18 mos.) 2010 College of American Pathologists. All rights reserved. 10

New edition released June 17, 2010 Remember the NOTE o Valuable information o Clarification o Additional guidelines Evidence of Compliance ROAD 2010 College of American Pathologists. All rights reserved. 11

Proficiency Testing Enrollment in CAP Surveys or another proficiency testing program accepted by CAP This applies to all tests, waived & non-waived o List of analytes requiring enrollment available on the CAP web site [http://www.cap.org] o Analytes NOT requiring enrollment may use alternative assessments 2010 College of American Pathologists. All rights reserved. 12

Proficiency Testing POC.03225 For tests that CAP doesn t require PT, lab can participate in external PT/alternative assessment o Perform semiannually o Perform on tests for which PT isn t available Ungraded proficiency surveys Split sample (with reference or other labs, established in-house method) Assayed material Regional pools Clinical validation by chart review Other suitable & documented means 2010 College of American Pathologists. All rights reserved. 13

Proficiency Testing POC.03250 Testing personnel must participate in PT o Don t have to get everyone every year Recommend rotation schedule Integrate PT samples within routine workload Can repeat if patient would be repeated Must be in PT policy o No interlaboratory communication o No referral of PT specimens 2010 College of American Pathologists. All rights reserved. 14

Proficiency Testing POC.03275 Review results in a timely manner o Within 1 month (prompt and timely) o Investigate unacceptable results o Solve any problems/document corrective action o Retain records for 2 years (unless specific discipline requires longer) o Acceptable or good 2010 College of American Pathologists. All rights reserved. 15

Quality Management Program Requirements The Point of Care Laboratory must have: o Indicators and/or Quality Measures defined o Evidence of review of the indicators and /or quality measures o Thresholds for each indicator and /or quality measures o Corrective action when the indicator or quality measure falls below the threshold. o Changes to operations as a result of the QM activities as specified in the program o POC QM activities reported as specified in the plan 2010 College of American Pathologists. All rights reserved. 16

Quality Management Program Requirements The Point of Care Laboratory must have a program to identify, document, and correct problems that affect patient care services. o Examples would be areas not notifying physicians of a critical value or not notifying the laboratory of a high glucose value that needs tested by the laboratory if required by policy. o An example in Point of Care when a root cause analysis needs to be performed is when a PT-INR result is reported incorrectly and the patient is treated based on the result and experiences an adverse outcome. 2010 College of American Pathologists. All rights reserved. 17

Procedure Manual POCT Procedures should include: o Principle o Clinical Significance o Specimen Type o Required Reagents o Test Calibration o Quality Control o Procedural Steps o Calculations o Reference Intervals o Interpretation of Test Result 2010 College of American Pathologists. All rights reserved. 18

Procedure Manual Other Considerations: Copy of the POCT Procedure Manual at each area o Physical manual or electronic access (i.e. via an intranet). o If electronic procedures, back-up plan in case of a system failure o If paper manuals distributed, mechanism to ensure they contain the current versions (i.e. a document control system) Archiving process must be established Annual review process 2010 College of American Pathologists. All rights reserved. 19

Quality Control General Requirements (waived & non-waived) o Performed by testing personnel o QC reviewed at least monthly (includes temp & instrument checks) o QC performed & acceptable before patient results reported o When out-of-range, document corrective actions 2010 College of American Pathologists. All rights reserved. 20

Quality Control (waived) POC.07037 Follow manufacturer instructions for frequency QC must be acceptable/documented prior to release of patient data Evidence of corrective action when QC exceeds tolerance limits Internal control results need not be documented for instruments using such controls, if there is a lock out feature 2010 College of American Pathologists. All rights reserved. 21

Quality Control (Non-waived) POC.07300 Controls are run daily for quantitative and qualitative tests. External only: o Qualitative (positive & negative) daily o Quantitative (2 levels) daily External (liquid) and Internal o External QC required each new reagent lot / shipment and every 30 days (at a minimum) o Internal QC (built-in/on-board or electronic) 2010 College of American Pathologists. All rights reserved. 22

Quality Control (Non-waived) Specific QC Requirements: o 2 levels of QC run daily o For ph, pco2, po2 1 level every 8 hrs of operation o For coagulation 2 levels required every 8 hrs o For qualitative tests: run positive & negative QC daily 2010 College of American Pathologists. All rights reserved. 23

Quality Control (Non-waived) Evidence of Compliance: o Records of QC results including external and internal (electronic/procedural/built-in) controls AND o Records documenting in-house validation of electronic/procedural/built-in control systems, if used 2010 College of American Pathologists. All rights reserved. 24

Method Comparison POC.07568 Must be performed for all non-waived instruments or methods twice/year o QC comparison for tests on the same instrument, using same QC lot#/manufacturer o Use of fresh patient samples is preferred, but not required *Remember this only applies to instruments within a single laboratory number. 2010 College of American Pathologists. All rights reserved. 25

Frequently Asked Question / Method Comparison The POCT Department has 50 i-stat meters. Does the lab have to correlate all meters? 2010 College of American Pathologists. All rights reserved. 26

Single-Use Device / Method Comparison POC.07568 Perform twice/year Use fresh human samples unless specimen stability is an issue If must use QC or reference materials, verify they respond the same as a patient sample Single-use devices can use alternative approach 2010 College of American Pathologists. All rights reserved. 27

Single-Use Device / Method Comparison For identical single-use non-waived devices using the same reagent lot: o Place analyzers into groups & select a representative device from each group o compare to each other and/or main lab instrument o If devices correlate, compare remaining devices to the subgroup using QC data 2010 College of American Pathologists. All rights reserved. 28

Analytical Measurement Range (AMR) POC.08450 AMR range of values that a method can measure without any dilution, concentration, or other pretreatment not part of the usual assay process Must be defined Revalidate at least every 6-months & reagent lot changes Does not apply to coagulation tests 2010 College of American Pathologists. All rights reserved. 29

Analytical Measurement Range (AMR) POC.08500 If the materials used for the calibration/cal verification include low, midpoint, & high values that are near the AMR, the AMR has been validated 2010 College of American Pathologists. All rights reserved. 30

Analytical Measurement Range (AMR) POC.08500 AMR Validation of Single-use Devices (non-waived): Initial validation & following service/repair Revalidation performed on sample of devices Devices not sampled, revalidation of AMR may be inferred using other approaches o review of QC o comparison of POCT results with near-simultaneously collected specimens analyzed in main lab. (This type of comparison is facilitated when POCT results are downloaded to central data management computer.) 2010 College of American Pathologists. All rights reserved. 31

Provider-Performed Testing (PPT) Testing that is personally performed by a physician or midlevel practitioner credentialed by the institution s medical staff. Includes both waived and PPM tests - 13 on list 2010 College of American Pathologists. All rights reserved. 32

Provider-Performed Testing (PPT) PPT subject to CAP inspection if: The Laboratory Director is responsible for competency assessment. and PPT is performed under the same CLIA# as the laboratory. 2010 College of American Pathologists. All rights reserved. 33

Frequently Asked Question How do I know if the physicians are credentialed? Credentialing is handled by the facility medical staff credentialing committee Documentation Policy in manual could assist with the inspection 2010 College of American Pathologists. All rights reserved. 34

PREPARATION IS KEY www.cap.org

Where do I start? Identify all POCT sites and the testing performed Review the POCT Checklist included with your reapplication well before your inspection Prepare an Audit Form Perform a self/mock inspection to determine if any deficiencies currently exist Implement corrective actions for any deficiencies identified in the self/mock inspection 2010 College of American Pathologists. All rights reserved. 36

What will Inspectors look for? General Items: Reagents labeled Patient results and reporting process QC, maintenance, and function checks Mechanism to determine the testing personnel Training & competency records Procedure manual in each POC testing area o Evidence personnel are knowledgeable of where procedure is located and contents Proficiency testing 2010 College of American Pathologists. All rights reserved. 37

Inspection Scenarios You are preparing for your upcoming inspection, and decide to do a walk-through of several point of care testing areas to assess their inspection-readiness. During your walk-through, you check for the following: o Current procedure manual available o List of authorized testing personnel available o No unauthorized testing kits o Reagents are in-date and properly labeled o Observe patient testing procedure o Observe patient result reporting practices o Review QC results for acceptability and documented corrective actions 2010 College of American Pathologists. All rights reserved. 38

Scenario 1 You maintain a master manual, and distribute copies of procedures to each testing site annually and upon revision. A cover memo accompanies each procedure instructing the testing site personnel to return the old procedure to you, replace it with the enclosed procedure and have all testing personnel sign the review sheet for the new procedure. The master manual has documentation of annual review for each procedure, and the signature pages are sent to the testing sites annually following review. You visit 5 testing sites during your walk-through. Four of the 5 testing sites have old versions of the procedures, one of which is for a glucose meter that was replaced18 months ago. The fifth site could not locate their procedure manual 2010 College of American Pathologists. All rights reserved. 39

Scenario 1 / What do you do? 1. Get the manuals up-to-date Replace outdated procedures with current procedures Replace missing procedure manuals Ensure staff are aware of contents of new procedures and are adequately trained 2. Review process for updating procedures with the responsible parties 2010 College of American Pathologists. All rights reserved. 40

Scenario 2 You ask for the list of authorized personnel. The list contains the names, ID numbers, and due dates for the next annual recertification. You note the list is dated January 2011 and several employees are past due for their recertification. 2010 College of American Pathologists. All rights reserved. 41

Scenario 2 / What do you do? 1. Set up recertification for overdue employees 2. Verify there are no new employees that were not on the list and set up training if needed 3. Update list 2010 College of American Pathologists. All rights reserved. 42

Is This a Deficiency? During the inspection of the glucose testing performed in the ER you ask the ER staff member performing the test to show you the procedure manual. The employee doesn t know, and says she hasn t seen the procedure before. Another nurse overhears the conversation and proceeds to show you where the manual is kept. Is this a deficiency? 2010 College of American Pathologists. All rights reserved. 43

Is This a Deficiency? Yes POC. 04200 The POCT program has a system documenting that all personnel are knowledgeable about the contents of procedure manuals (including changes) relevant to the scope of their testing activities. NOTE: This does not specifically require annual procedure sign-off by testing personnel. The form of this system is at the discretion of the laboratory director. 2010 College of American Pathologists. All rights reserved. 44

Is This a Deficiency? While observing a urine dipstick being performed you notice the specimen is unlabeled. The staff member tells you that labeling the specimen is unnecessary because the patient is with the specimen the entire time. Is this a deficiency? 2010 College of American Pathologists. All rights reserved. 45

Is This a Deficiency? Yes POC.04300 There is a documented procedure describing methods for patient identification, patient preparation, specimen collection and labeling, specimen accessioning, and specimen preservation (if applicable) before testing. GEN.40491 Primary specimen containers are labeled by at least 2 identifiers. 2010 College of American Pathologists. All rights reserved. 46

Is This a Deficiency? The laboratory has 20 waived glucose meters. You have asked for semi-annual instrument correlations. The laboratory has stated that they don t do them. Is this a deficiency? 2010 College of American Pathologists. All rights reserved. 47

Is This a Deficiency? No What if the instruments are I-stats? Why or why not? 2010 College of American Pathologists. All rights reserved. 48

Is This a Deficiency? Your current glucose meter control lot number has a manufacturer expiration date on the bottle of March 2011. You find glucose meter controls currently in use marked opened on 1/15/2011. Controls are stable for 90 days after opening. Would this be a deficiency? 2010 College of American Pathologists. All rights reserved. 49

Is This a Deficiency? No Material expires on last day of month when only month/year are indicated The control vials were opened less than 90 days ago, so they have not exceeded the open stability Controls vials are dated with the open date, and the manufacturer label indicates that the materials are stable for 90 days once opened 2010 College of American Pathologists. All rights reserved. 50

Thank You! For additional information: www.cap.org e-mail: accred@cap.org 800-323-4040 2010 College of American Pathologists. All rights reserved. 51