ACR MRI Accreditation Program: The Technologist s Role

Similar documents
Breast MRI Quality Control

MRI QA Technologist s Tests

MRI QA Technologist s Tests

Phantom Test Guidance

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

SITE IMAGING MANUAL ACRIN 6698

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Quality Control of Full Field Digital Mammography Units

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

Prepublication Requirements

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

AT&T Global Network Client for Windows Product Support Matrix January 29, 2015

Musculoskeletal MRI Technical Considerations

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

Magnetic Resonance Imaging

COMPARISON OF FIXED & VARIABLE RATES (25 YEARS) CHARTERED BANK ADMINISTERED INTEREST RATES - PRIME BUSINESS*

COMPARISON OF FIXED & VARIABLE RATES (25 YEARS) CHARTERED BANK ADMINISTERED INTEREST RATES - PRIME BUSINESS*

GE Medical Systems Training in Partnership. Module 12: Spin Echo

INTRODUCTION. A. Purpose

Quality Control and Maintenance Programs

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

5 Factors Affecting the Signal-to-Noise Ratio

Analysis One Code Desc. Transaction Amount. Fiscal Period

Case 2:08-cv ABC-E Document 1-4 Filed 04/15/2008 Page 1 of 138. Exhibit 8

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Update on ACR Digital Mammography QC Manual

WHERE IN THE WORLD JILL LIPOTI?

Enhanced Vessel Traffic Management System Booking Slots Available and Vessels Booked per Day From 12-JAN-2016 To 30-JUN-2017

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions

Performance testing for Precision 500D Classical R/F System

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

The Dedicated MRI Factbook. Esaote MRI: A world of difference

Education. October, 1985 GE Medical Systems Milwaukee, WI Signa Operators Training Program

MRI Accreditation Program Clinical Image Quality Guide

The smart approach to digital MRI

MGH Adult Diffusion Data Scanning Protocols

Update on MQSA and ACR Mammography Accreditation

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Diagnostic Ultrasound Performance Testing & Accreditation

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

MRI for Paediatric Surgeons

MRI SEQUENCES. 1 Gradient Echo Sequence

FDA Guidelines for Magnetic Resonance Equipment Safety

Overview. Optimizing MRI Protocols. Image Contrast. Morphology & Physiology. User Selectable Parameters. Tissue Parameters

Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants

32-Channel Head Coil Imaging at 3T

Ashley Institute of Training Schedule of VET Tuition Fees 2015

ESTABLISHMENT OF A NEW BUSINESS

Whitepapers on Imaging Infrastructure for Research Paper 1. General Workflow Considerations

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

DOE LHC Quarterly Review. Magnet Production Cable Testing Cost to Complete issues. Mike Harrison, BNL July 26, 2004

Checklist for the Qualification of Digital Detector Array Systems

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

American College of Radiology Mammography Accreditation Program

Please follow all instructions carefully.

RF Coils... They ve Come a Long, Long Way

Schedule of VET FEE-HELP Tuition Fees & Census Dates

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010

CENTERPOINT ENERGY TEXARKANA SERVICE AREA GAS SUPPLY RATE (GSR) JULY Small Commercial Service (SCS-1) GSR

Configuration and Use of the MR Knee Dot Engine

FY 2015 Schedule at a Glance

CAFIS REPORT

General requirements. 1. Administrative Controls.

The multitude of symptoms following a whiplash injury has given rise to much discussion because of the lack of objective radiological findings.

Resource Management Spreadsheet Capabilities. Stuart Dixon Resource Manager

Purchased Services Areas of Opportunity:

Understanding Digital Modalities: System Integration and Use

Current Industry Neuroimaging Experience in Clinical Trials Jerome Barakos, M.D.

Accreditation Is Coming

a. mean b. interquartile range c. range d. median

1. Introduction. 2. User Instructions. 2.1 Set-up

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

University Children s Hospital Basel

Greece. Annual Observance Report of the Special Data Dissemination Standard for 2015 I. INTRODUCTION

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

SMRT Student Scope Submission

it s all about Choices School of Health Related Professions Diagnostic Imaging Technologies

ABHES BULLETIN. Subject: Final Radiologic Technology Standards Effective January 1, 2013

ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

Automatic Wheel Profile Measuring Device for Tramway Vehicles

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

American College of Radiology CT Accreditation Program. Testing Instructions

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

The IAC Standards and Guidelines for MRI Accreditation

The Impact of Medicare Part D on the Percent Gross Margin Earned by Texas Independent Pharmacies for Dual Eligible Beneficiary Claims

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

Emory University RT to Bachelor of Medical Science Degree Medical Imaging

Digital Breast Tomosynthesis QC Requirements

How To Write A Special Data Distribution Standard Report

Performance Monitoring on NERSC s POWER 5 System

Transcription:

ACR MRI Accreditation Program: The Technologist s Role Geoffrey D. Clarke, Ph.D. University of Texas Health Science Center at San Antonio Radiological Sciences Division

Overview ACR MRI Accreditation Program Application process & staff requirements ACR MRI QC Phantom Clinical Imaging Head, cervical, lumbar & knee ACR MRI QC Manual Technologist s Responsibilities for QC

ACR MRI Accreditation Program Overview Voluntary & Educational in Nature Evaluates qualifications of personnel Evaluates equipment performance Evaluates effectiveness of quality control measures Evaluates quality of clinical images

Two-Part Application Process Review of the entry application Credentials of physicians, physicists/mr scientists, and technologists Information common to the practice of MRI Acquisition of clinical and phantom images Required clinical images consist of routine brain, cervical spine, lumbar spine, and knee Must use a designated MRI phantom Data must be obtained from each full body general purpose magnet at the site Appeal process if failure

Technologist Requirements Technologists performing MRI should: 1. Be certified by ARRT as a MR Technologist, OR 2. Be certified by ARRT and/or state licensure and have 6 months of clinical MRI experience, OR 3. Have an associates degree in an allied health field or a bachelors degree and certification in another clinical imaging field and have 6 months supervised clinical MRI experience

Clinical Images - Acquisition 1.Routine Brain examination (for headache) Sagittal short TR/short TE with dark CSF Axial or coronal long TR/short TE (or FLAIR) and long TR/long TE (e.g., long TR double echo) 2.Routine Cervical Spine (for radiculopathy) Sagittal short TR/short TE with dark CSF Sagittal long TR/long TE or T2*W with bright CSF Axial long TR/long TE or T2*W with bright CSF

Clinical Images - Acquisition 3. Routine Lumbar Spine (for back pain) Sagittal short TR/short TE with dark CSF Sagittal long TR/long TE or T2*W with bright CSF Axial short TR/short TE with dark CSF and/or long TR/long TE with bright CSF 4. Complete Routine Knee examination (for internal derangement) To include sagittal(s) and coronal(s) with at least one sequence with bright fluid

Clinical Images - Evaluation Each set of clinical images will be evaluated for: Pulse sequences and image contrast. Filming technique. Anatomic coverage and imaging planes. Spatial resolution. Artifacts. Exam ID - All patient information annotated on clinical exams will be kept confidential by the ACR.

Clinical Images - Resolution Sequence Slice Thickness Gap Maximum Pixel Dimension All Brain < 5 mm < 2 mm <1.2 mm C-spine Sagittal < 3 mm <1 mm <1 mm C-spine Axial < 3 mm <1 mm <1 mm L-spine Sagittal < 5 mm <1.5 mm <1.5 mm L-spine Axial <4 mm <1 mm <1.5 mm Knee <4 mm <1 mm < 0.75 mm

ACR Phantom Goals Tests run in a short time Pulse sequences as compatible as possible with all commercial MRI scanners Specific measurements to account for : Geometric Distortion Slice Thickness & Position Factors Affecting Image SNR (resolution, bandwidth, ghosting) Image Uniformity Affordable ($730) Results easily evaluated

Documents for Using Phantom How to scan How to analyze $25 per set from ACR

ACR MRI Accreditation Program MRI Survey Agreement Official request for ACR Accreditation Site agrees to provide all documentation, including but not limited to quality control logs, films, records, or any necessary information requested by the survey team Agree to use the ACR MRI phantom

ACR Magnetic Resonance Imaging Quality Control Manual 2001 $57.50 from ACR

Technologist Responsibilities Designated and trained QC Technologist(s) Maintain QC Notebook QC policies and procedures data forms where QC test results are recorded notes on QC problems and corrective actions Review QC Data with QA Committee Only Use Alternative Phantoms & Procedures when documented by physicist or MRI Scientist Follow Action Limits Established by Physicist or MRI Scientist

Technologist s QC Tasks Perform image quality tests as appropriate (at least weekly): Central Frequency Signal-to-noise ratio Image Quality High contrast resolution Low contrast detectability Image Artifacts Perform weekly checks of hard copy fidelity (processor sensitometry) Weekly visual inspection of equipment

Setting Up Routine Image QC Daily (weekly) tests Develop cheat sheets to speed process System for rapid positioning of phantom TO SET ACTION LIMITS: Review data after first ten days to establish baseline values and variability.

Time to Perform QC Set-up Phantom Center Frequency Table Positioning Setup & Scanning Geometric Accuracy High Contrast Resolution Low Contrast Resolution Artifact Analysis Film Quality Control Visual Checklist 2 minutes 1 minute 3 minutes 5 minutes * 2 minutes * 1 minute 1 minute 1 minutes 10 minutes 5 minutes 15 minutes * Geometric accuracy analysis on sagittal is done while scanning axial

Central Frequency & Transmitter Gain Record center frequency value on ACR phantom or manufacturer s phantom Reflects power required to optimize RF pulse: Depends on coil, phantom, pulse sequence, etc. Should remain constant over time if nothing in pulse sequence or hardware has changed

Tranasmitter Gain Terminology Varies GE: displayed on screen (db) Philips: under system performance parameters rf_act_drivescale Siemens: options adjustments Frequency Transmitter amplitude (temp) (V) Toshiba: acquisition window Center frequency (MHz) RF level

Geometric Accuracy Measure distance along main axes of phantom Compare with known values

Geometric Accuracy - Axial BW = 7.4 khz BW = 3.6 khz Slice #1 Failure Due To: Miscalibrated Gradients Low Receiver Bandwidth High B o Inhomogeneities

10.00% Percent Geometric Distortion T1-Weighted, Central Slice 8.00% 6.00% 4.00% 2.00% Poor Gradient Calibration 0.00% 1-31- 30-30- 29-28- 29-28- 28-27- 27-26- 25-25- Oct Oct Nov Dec Jan Feb Mar Apr May Jun Jul Aug Sep Oct Date of Measurement System #1 System #2 System #3

Assessment of MR Image Quality Using ACR Phantom High Contrast Resolution Low Contrast Detectability (also used for SNR)

High Contrast Spatial Resolution UL 1.1 mm 1.0 mm 0.90 mm LR Evaluate visibility of holes arranged in two square arrays Avoid partial-volume artifact from tilting.

High Contrast Resolution Specific but not sensitive Action Criteria: Any reduction in # of holes seen Suggestive of: Increased eddy currents Poor gradient calibration Poor B o uniformity Reduced stability of system

Low Contrast Detectability Four sets of plastic membranes with holes 1.5 mm to 7 mm in diameter Only look at one slice: - determined by Physicist/MR Scientist 1.4% Contrast Detectability 2.4% Contrast Detectability 7mm 6mm 5mm 4mm 3.5mm Slice #8 Slice #9

Low Contrast Detectability 1.5 T System: T1-Weighted Scan 25.00 Hole Size Visualized (mm) 20.00 15.00 10.00 5.00 1.4% contrast 2.4% contrast 3.7% contrast 5.1% contrast 0.00 1-May 31-May 30-Jun 30-Jul 29-Aug 28-Sep 28-Oct 27-Nov 27-Dec 26-Jan 25-Feb 27-Mar 26-Apr Date

Low Contrast Detectability ACR Slice #8 (1.4% Contrast) 10 System #3 Grad Amp Bad Another New Gradient Amp Number of Hole Sets 8 6 4 2 0 7/1/1998 8/20/1998 10/9/1998 11/28/199 8 1/17/1999 3/8/1999 4/27/1999 6/16/1999 8/5/1999 Date System #1 System #2 System #3 Bad Slice Positioning Mini Upgrade

Measurement of System SNR Can be performed by technologist Method to be used Manufacturer s method ACR method Automated analysis and recording often available on modern MRI systems

LCD and Signal-to-Noise Total Number of Spokes 40 35 30 25 20 15 10 5 0 0 50 100 150 200 250 300 350 Signal-to-Noise Ratio

Artifact Evaluation Check for: Distortion? Ghosts in phantom or background? Streaks? Bright or dark spots? New features? Ghosting in phantom

Artifacts Good Scan Spike Drop Out DC Offset

ACR MRI Standard Hardcopy Image QC Tests Sensitometric Measurements for Film Processors Hard Copy of SMPTE test pattern Similar to process used for mammography program Laser camera film less sensitive to temperature changes

Laser Film QC Weekly: View SMPTE pattern Verify gray levels 0/5% & 95/100% patches Film 6 on 1 4 on 1 if necessary Plot OD of 10%, 40% & 90& patches Observe film for artifacts SMPTE Test Pattern

Action Limits SMPTE patch 0 10% 40% 90% OD 2.45 2.10 1.15 0.30 Control Limit ±0.15 ±0.15 ±0.15 ±0.08

Quality Control Program Records Data form for daily (weekly) equipment quality control ACR MRI QC manual, pg. 64 MRI Facility quality control visual checklist ACR MRI QC manual, pg. 65 Laser film printer control chart ACR MRI QC manual, pg. 66

If QC Test Fails. Common errors- Check for magnetic objects in bore Check connections and reseat head coil Reposition & landmark phantom Make sure scan room door securely closed Repeat QC scan procedures Record results again in QC notebook

Successful MRI QC Program 1. Technologists run QC scans on a daily basis 2. Set action criteria repeat QC procedure 3. Physicist reviews QC data quarterly or semiannually 4. Record data - report problems to service 5. Have service record problems and solutions in a service log