Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Similar documents
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

CORESTA RECOMMENDED METHOD N 72

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

A Complete Solution for Method Linearity in HPLC and UHPLC

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

How To Develop A Recipe Card

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)

ApplicationNOTE Introduction

INTRODUCTION. The principles of which are to:

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Simultaneous Analysis of Water-Soluble Vitamins in Vitamin-Enriched Beverages and Multivitamin Dietary Supplements by UHPLC-MS/MS

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Thermo Scientific Vanquish UHPLC System. Revolutionize Your. UHPLC Experience

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Determination of Tetracycline Antibiotics in Milk Using a Simple Strong Cation-Exchange SPE Cleanup Procedure and LC-MS/MS Analysis

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Optimize Your Process Operations by Improving Process Monitoring

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

WATERS QUANTITATIVE ANALYSIS solutions

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Pesticide Analysis by Mass Spectrometry

Appendix 5 Overview of requirements in English

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

S olid Ph ase Extraction (SP E) Method:

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Determination of Calcium and Magnesium in Brine

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

LC-MS/MS for Chromatographers

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS No. SSI-GCMS Shilpi Chopra, Ph.D.

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

How To Test For Contamination In Large Volume Water

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Environmental Analysis Using the ICS-3000 Ion Chromatography System

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

Thermo Scientific Dionex UltiMate 3000 Variable Wavelength Detectors

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

AMD Analysis & Technology AG

Signal, Noise, and Detection Limits in Mass Spectrometry

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Chromium Speciation in Drinking Water by LC-ICP-MS

Thermo Scientific Mass Spectrometric Immunoassay (MSIA)

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Gas Chromatography Liner Selection Guide

Development of an Automated Method for Monoclonal Antibody Purification and Analysis. Introduction. Gurmil Gendeh, 1 Wim Decrop, 2 and Remco Swart 2 1

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Electrospray Ion Trap Mass Spectrometry. Introduction

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Determination of Caffeine in Beverages HPLC-1

Samples RMN Analytical Report ( ) GC/FID Analysis (RM01) Nicotine Purity

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

Introduction. Methods. Sample Processing

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Transcription:

Application Note 242 Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS INTRODUCTION Tobacco-specific nitrosamines (TSNA) are a group of carcinogens found only in tobacco products. They are formed from nicotine and related alkaloids during the production and processing of tobacco and tobacco products. 1 Due to their carcinogenic properties, efforts have been made to reduce TSNA levels in tobacco products. The desired goal of this investigation is to develop a sensitive, high-throughput method to monitor TSNA levels in tobacco and tobacco products. This study focuses on N'-nitrosonornicotine (NNN), N'- nitrosoanatabine (NAT), N-nitrosoanabasine (NAB), and 4-(methylnitrosamino)-1-(3-pyridyl)-1-butanone (NNK). Conventional methods for TSNA analysis are based on gas chromatography with a thermal energy analyzer (GC-TEA) 2,3, or high-performance liquid chromatography (HPLC) with various detection techniques such as UV and mass spectrometry (MS). 4 6 This application note (AN) describes a robust and fast LC-MS/MS method developed to analyze the four TSNAs in tobacco cigarettes after a simple sample preparation of ammonium acetate liquid extraction and filtration. The chromatographic separation of the four TSNAs was achieved within 3.5 min and these analytes were detected with great sensitivity and selectivity by tandem mass spectrometry using Selected Reaction Monitoring (SRM) detection. This method has been evaluated with respect to linearity, detection limits, precision, and accuracy. The ruggedness of the methodology was proven with more than 1 replicate injections. EXPERIMENTAL Instrumentation HPLC P68 dual ternary pump ASI-1 autosampler TCC-1 thermostatted column compartment UVD34U detector Mass Spectrometer TSQ Quantum Access triple quadrupole mass spectrometer with heated electrospray ionization (HESI) interface. Software Xcalibur 2. Dionex DCMS Link for Xcalibur (version 2.)* * DCMS Link is a Chromeleon -based software module providing the interface for controlling a wide range of Dionex chromatography instruments from different mass spectrometer software platforms. Conditions Chromatographic Conditions Analytical Column: Acclaim RSLC PA2 (5. 2.1 mm, 2.2 µm) Column Temperature: 6 C Injection Volume: 1 µl Mobile Phase: 1% CH 3 CN in buffer (1 mm NH 4 OAc, ph adjusted to ph 8. by NH 4 OH) Flow Rate:.5 ml/min Detection: UV at 23 nm TSQ Quantum Access SRM

Mass Spectrometric Conditions Needle Voltage: 2 V Vaporizer Temp.: 35 C Sheath Gas: Auxiliary Gas: Ion Sweep Gas: 6 (arbitrary unit) 5 (arbitrary unit) 2 (arbitrary unit) Collision Gas Pressure: 1.5 mtorr Capillary Temp.: 35 C SRM Transitions: NNN 178.1 148.1 m/z 11 V 178.1 12.1 m/z 18 V NNK 28.1 122.1 m/z 11 V 28.1 79.2 m/z 38 V NNK-d 4 212.1 126.1 m/z 11 V 212.1 83.2 m/z 38 V NAT 19.1 16.1 m/z 1 V 19.1 79.2 m/z 5 V NAB 192.1 162.1 m/z 1 V 192.1 133.1 m/z 24 V The 1st SRM transition of each compound is used for quantification and the 2nd SRM transition is used for confirmation. Sample Preparation Tobacco cuts from five brands of cigarettes were weighed to.25 grams into 2 ml glass vials, and extracted with 1 ml of 1 mm ammonium acetate solution. The vials were placed on a swirl table and agitated for 3 min. The extracts were filtered through.25 µm membrane syringe filters. A 1. ml aliquot of each extract was then spiked with 1 µl internal standard (IStd, NNK-d 4 ) in preparation for LC-MS/MS analysis. RESULTS AND DISCUSSION Chromatography As shown in Figure 1, the four TSNAs are retained and completely resolved within 3.5 min, with the minimum of the retention factors (K' NNN ) greater than 4 and minimum resolution (R snat ) greater than 2.. These demonstrated retention factors ensure analytes of interest are chromatographically separated from early eluted compounds that can suppress ionization. The total chromatographic resolution minimizes the possibility of ionization suppression and eliminates crosscontamination of SRM transitions between main analytes. Relative Abundance 1 1 1 1 1 7 Absorbance (mau) NNN: 1.15 AA:437725, S/N:2394 NNK: 1.55 AA:49382, S/N:2223 NNK-d4: 1.51 AA:13384245, S/N: 12322 (Rs: 2.34) NNK - 1.65 (Rs: 4.29) NAT: 2.45 AA:1338249, S/N:1465 NAB: 2.89 AA: 3918316, S/N:315 NAB - 3.2 NAT - 2.56 (Rs: 2.5) NL: 7.13E4 178.1 148.1 NL: 6.66E4 28.1 122.1 NL: 1.73E6 212.1 126.1 NL: 1.42E5 19.1 16.1 NL: 4.77E5 192.1 162.1.5 1. 1.5 2. 2.5 3. 3.5 4. MS/MS Conditions Needle Voltage: 2 V Vaporizer Temp.: 35 C Sheath Gas: 6 (arbitrary unit) Auxiliary Gas: 5 (arbitrary unit) Ion Sweep Gas: 2 (arbitrary unit) Collision Gas Pre.: 1.5 mtorr Capillary Temp: 35 C SRM Transitions: Analyte SRM Collision Energy NNN 178.1 148.1 11 V NNK 28.1 122.1 11 V NNK-d 4 212.1 126.1 11 V NAT 19.1 16.1 1 V NAB 192.1 162.1 1 V LC Conditions Column: Acclaim PA2 5 2.1mm, 2.2 µm Mobile Phase: 1% CH 3 CN in buffer (1 mm NH 4 OAc, adjusted to ph = 8. with NH 4 OH) Column Temp.: 6 C Flow Rate:.5 ml/min (backpressure ~ 13 bar) Inj. Volume: 1 µl Analyte Conc.: 1 ppb each analyte 2 ppb for internal standard 5 ppm for UV detection Detection: UV at 23 nm or TSQ Quantum Access SRM with HESI 25776 Figure 1. Comparison of SRM and UV chromatograms of TSNAs on the Acclaim RSLC PA2 column. 2 Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Mass Spectrometry Tandem mass spectrometry provides specific and sensitive detection. The SRM chromatograms in Figure 1 show that quantification can be performed with a high level of confidence even at low concentration (1 ng/ml,.1 ng injected amount). The use of SRM detection enabled simplification of sample preparation and a significant reduction in total process time. Method Performance Internal standard quantification and calibration were performed using isotope labeled NNK-d 4 (IStd). Calibration standards were prepared at 8 levels from 2 ng/ml to 1 ng/ml. Triplicate injections were performed to generate calibration curves, and a coefficient of determination (r 2 ) greater than.99 was achieved for each analyte. 1/X weighting provided best quantification for low concentration samples. Excellent linearity was achieved for NNK with r 2 =.9998, suggesting that the use of isotope-labeled analogs as internal standards could provide better quantification accuracy and precision. The calibration curve is shown in Figure 2. Method detection limits (MDLs) were statistically calculated using the equation MDL = S t (99%, n=7) where S is the standard deviation, and t is student s t at 99% confidence interval. Seven replicate injections of a standard solution at 2 ng/ml were performed, and used for MDL calculations. The MDLs for TSNAs range from.22 ng/ml (NAB) to.38 ng/ml (NNK). Precision and accuracy were evaluated at 1 ng/ml and 1 ng/ml. Seven replicate injections of calibration standards at both levels were performed, and quantification was calculated based on the calibration curves. The results in Table 1 show that the quantification of NNK was very accurate and precise (< 4 % difference, < 4.5% RSD) at both low (1 ng/ml) and medium levels (1 ng/ml). However, significant inaccuracy was observed for NAB and NAT at the low level, showing 54.2% and 36.6% quantification differences respectively. This observation suggests better quantification accuracy may be achieved using an isotope labeled internal standard for each analyte for low-level TSNA analysis, for example, analyzing TSNAs in biological matrices. Area Ratio 3.5 3. 2.5 2. 1.5 1..5 NNK Y =.11554+.339767*X r 2 =.9998 W: 1/X NNK Y =.11554+.339767*X r 2 =.9998 W: 1/X Area Ratio.35.3.25.2.15.1.5. 2 4 6 8 1 ppb 2 4 6 8 1 ppb 25777 Table 1. Precision and Accuracy Accuracy (%) 1 ng/ml 1 ng/ml % RSD Accuracy (%) % RSD NAB 154.2 2.74 13.7 1.49 NAT 136.6 4.69 1.7 1.77 NNK 13.9 4.38 1.9 1.88 NNN 95.3 3.56 94.4 1.56 Figure 2. Calibration curve of NNK using NNK-d 4 as internal standard. Application Note 242 3

Method Ruggedness Since the analytical column was running under the very harsh conditions of high temperature and high ph mobile phase, method ruggedness was evaluated by repeated injections of standard solutions and tobacco extracts. As shown in Figure 3, chromatographic retention and resolution were well-maintained after more than 1 injections. The column was cleaned with 5/5 CH 3 CN/H 2 O before injection 25 and a new inlet frit was installed before injection 13. (The use of an in-line filter is highly recommended to prevent column clogging.) 7 Absorbance (mau) Absorbance (mau) Absorbance (mau) 7 6 NNK - 1.65 (Rs: 4.29) (Rs: 2.34) NNK - 1.64 (Rs: 4.9) (Rs: 2.27) NNK - 1.62 (Rs: 3.68) NNN - 1.24 (Rs: 2.8) NAT - 2.56 (Rs: 2.5) NAT - 2.53 (Rs: 1.88) NAT - 2.49 (Rs: 1.63) NAB - 3.2 NAB - 2.98 NAB - 2.93 Injection #13 1 2 3 4 25778 Figure 3. Ruggedness of the Acclaim RSLC PA2 column. Injection #1 Injection #25 Determination of TSNAs in Tobacco Cigarettes Five brands of tobacco cigarettes were purchased from a local convenience store. Tobacco cuts from each brand of cigarettes were prepared (n = 3) following the procedure described previously. The results are shown in Table 2. The TSNA contents are significantly different between brands A and B and brands C, D, and E. The differences in TSNA content could be related to differences in the tobacco blended in each brand, such as type, origin, age of the tobacco, curing method, and storage conditions. The comparison of brands A and B (regular and light varieties of a premium US brand of cigarettes) shows no difference for TSNA contents except for a change in the amount of NAB. See Figure 4 for SRM chromatograms of TSNAs from brand A cigarette. Relative Abundance 1 1 1 1 1 NNN - 1.18 AA: 1591114, S/N: 1183 NNK - 1.58 AA: 494173, S/N: 348 NNK-d4-1.55 AA: 7125695, S/N: 218 NAT - 2.47 AA: 2186786, S/N: 798 NAB - 2.9 AA: 456429, S/N: 95 NL: 2.28E5 178.1 148.1 NL: 4.57E4 28.1 122.1 NL: 6.29E5 212.1 126.1 NL: 2.8E5 19.1 16.1 NL: 6.45E4 192.1 162.1.5 1. 1.5 2. 2.5 3. 3.5 4. 25778 Figure 4. SRM chromatograms of TSNAs from brand A cigarette. 4 Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

Table 2. TSNAs in Tobacco Cigarettes* NAB NAT NNK NNN Note Brand A 5.3 1145.1 676.5 2221.1 US brand, regular Brand B 42.1 1218.1 696. 2472.5 US brand, lights Brand C 4.1 16.8 225.2 213. US brand, all natural Brand D 21.2 13.9 7.2 155.4 International brand Brand E 32.7 78.2 87.8 114.3 Asian brand *Amounts shown in Table 2 are ng/g tobacco cuts. CONCLUSION A rugged and ultrafast method for TSNA analysis was developed using HPLC-MS/MS on an Acclaim RSLC PA2 column. Four TSNAs were chromatographically retained and resolved within 3.5 min. Tandem mass spectrometry ensured selectivity and sensitivity. The use of isotope-labeled intenal standards for each analyte may improve the quantification accuracy, as was observed for NNK in this study. The ruggedness of the method was shown by more than 1 injections of standards and tobacco extracts. The applicability of the method for the determination of TSNAs in tobacco cuts from five brands of cigarettes was demonstrated. REFERENCES 1. Hecht, S.S.; Hoffmann, D. Carcinogenesis 1988, 9, 875 884. 2. Adams, J.D.; Brunnemann, K.D.; Hoffmann, D. J. Chromatogr., A 1983, 256, 347 351. 3. Krull, I.S.; Swartz, M.; Hilliard, R.; Xie, K.H.; Driscoll, J.N. J. Chromatogr., A 1983, 26, 347 362. 4. Wu, W.; Ashley, D. L.; Watson, C. H. Anal. Chem. 23, 75, 4827 4832. 5. Wagner, K.A.; Finkel, N.H.; Fossett, J.E.; Gillman, I.G. Anal. Chem. 25, 77, 11 16. 6. Wu, J.; Joza, P.; Sharifi, M.; Rickert, W.S.; Lauterbach, J.H. Anal. Chem. 28, 8, 1341 1345. TSQ Quantum Access is a trademark and Xcalibur is a registered trademark of Thermo Fisher Scientific. DCMS Link is a trademark and Acclaim and Chromeleon are registered trademarks of Dionex Corporation. Passion. Power. Productivity. Dionex Corporation North America Europe Asia Pacific 1228 Titan Way U.S./Canada (847) 295-75 Austria (43) 1 616 51 25 Benelux (31) 2 683 9768 (32) 3 353 4294 P.O. Box 363 Denmark (45) 36 36 9 9 France (33) 1 39 3 1 1 Germany (49) 6126 991 Sunnyvale, CA Ireland (353) 1 644 64 Italy (39) 2 51 62 1267 Sweden (46) 8 473 338 Taiwan (886) 2 8751 6655 South America 9488-363 Switzerland (41) 62 25 9966 United Kingdom (44) 1276 691722 Brazil (55) 11 3731 514 (48) 737-7 www.dionex.com Australia (61) 2 942 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 258 Singapore (65) 6289 119 LPN 2372 PDF 1/9 29 Dionex Corporation