CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Similar documents
How To Improve Your Ct Image Quality

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

Purchasing a cardiac CT scanner: What the radiologist needs to know

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Assessing Radiation Dose: How to Do It Right

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Multi-slice Helical CT Scanning of the Chest

American College of Radiology CT Accreditation Program. Testing Instructions

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Copyright March 1, 2016 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

Practical exercise: Effective dose estimate in CT

State-of-the-Art Technology in Cardiac CT

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Low-dose CT for Pulmonary Embolism

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

The Challenge of CT Dose Records

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

Pediatric Hospitals Bring Low-dose CT to the Middle East

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

GE Healthcare. Revolution EVO. More than just high tech. Higher purpose.

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35

Patient Exposure Doses During Diagnostic Radiography

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

DENTAL Cone beam 3D X-RAY SYSTEM with

Performance testing for Precision 500D Classical R/F System

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Diagnostic Exposure Tracking in the Medical Record

Fundamentals of Cone-Beam CT Imaging

WHERE IN THE WORLD JILL LIPOTI?

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Computed Tomography (CT) - Sinuses

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Cardiac CT for Calcium Scoring

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

3 CT Parameters that Influence the Radiation Dose

Computed Tomography (CT) - Chest

Prepublication Requirements

The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO

Brilliance CT 64-channel confi guration

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

CT Scan Thorax and Upper Abdomen. Respiratory Unit Patient Information Leaflet

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

GE Healthcare. Great treasures in small places. BrightSpeed Elite

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Computed Tomography (CT) - Spine

Sistemas de reducción de dosis en CT para pediatría. Nuevos avances Ángela de Pinto Siemens CT Business Manager

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!

AN INNOVATIVE & CUTTING-EDGE DIGITAL RADIOGRAPHY SYSTEM

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan

Physics testing of image detectors

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

INTRODUCTION. A. Purpose

Multi Detector Row CT Systems and Image- Reconstruction Techniques 1

Data. microcat +SPECT

Mx8000 IDT 16 provides dose control methodologies like DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation).

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Digital Breast Tomosynthesis QC Requirements

IAEA HUMAN HEALTH SERIES

Quality Control and Maintenance Programs

SITE IMAGING MANUAL ACRIN 6698

Staff Doses & Practical Radiation Protection in DEXA

Automation of a CT Acquisition: A System-based User-Support. 1. Introduction. Bärbel KRATZ 1, Frank HEROLD 1, Malte KURFISS 1

Protocol Management and Review Strategies The MD Anderson Experience

A d v a n c e d 1 2 8

Milling. COPYRIGHT 2008, Seco Tools AB 1/111

ABSTRACT INTRODUCTION

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction

Transcription:

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of Medicine & Public Health 1

Quick Overview TOPICS: Modern Computed Tomography CT Dosimetry CT Image Quality & Dose Optimization of CT Scan Techniques for Dose & Image Quality 8/8/13 2013, Frank Ranallo 2

Evolution to Helical/ Spiral CT Scanners Single Slice Helical/ Spiral CT 3

Evolution to Multislice Scanners 2, 4, 8, 16, 64,? Data Acquisition 4

Evolution to Multislice Scanners 2, 4, 8, 16, 64,? 5

Pitch = Evolution to Multislice Scanners 2, 4, 8, 16, 64,? Definition of Pitch for Single-slice Helical / Spiral Scanning: Pitch = Table travel per 360 tube rotation Nominal Slice Thickness Definition of Pitch for Multi-slice Helical / Spiral Scanning: Table travel per 360 tube rotation Total collimation width of all simultaneously collected slices 6

Dose in Computed Tomography 7

CT Dose CTDI w is a weighted value of CTDI 100 measurements, at the center and the surface of the 16cm or 32cm diameter CTDI phantom, that attempts to give the average dose throughout the volume of the phantom (for contiguous axial slices or for helical scanning at a pitch of 1). It is defined as: CTDI w = 1/3 CTDI 100 (center) + 2/3 CTDI 100 (surface) 8

CT Dose CTDI vol If you then take into account the effect of pitch on dose for a helical/ spiral scan then you have another version of CTDI: CTDI vol = CTDI w / Pitch This is an approximation of the dose averaged over the volume of the phantom. 9

DLP A final dosimetry measure is the Dose Length Product (DLP) which is defined as: DLP = CTDI vol CT Dose Scan Length and has units of mgy cm CTDI vol and DLP are the CT dose units provided by the CT Scanner. 10

CT Dose DLP Estimates of an average size patient s effective dose (E) can be derived from the values of DLP for an examination. Use the following equation containing a coefficient E DLP appropriate to the examination: E = E DLP DLP 11

DLP CT Dose Values of E DLP for Adult scans: Region of Body E DLP (msv / mgy cm) Phantom Head 0.0023 0.0021 16 cm Head & Neck 0.0031 16 cm Neck 0.0054 0.0059 32 cm Chest 0.017 0.014 32 cm Abdomen 0.015 0.015 32 cm Pelvis 0.019 0.015 32 cm Two different sources 12

Converting DLP to Effective Dose for Adults CT Dose Quick method of getting the effective dose in msv from the DLP in mgy cm using previous table: Take the DLP and divide it by 100 For the Body, then multiply by 1.5 For the Head, then divide by 4 13

Image Quality in Computed Tomography 14

Image Quality and Dose: Artifacts Dose Image Sharpness Image Noise Modulation Transfer Function MTF (visibility of small high contrast objects) visibility of smaller, lower contrast clinical objects Low Contrast Detectability (visibility of large low contrast objects) 15

Image Reconstruction Filters, Algorithms, or Kernels Below are the measured MTF functions for the reconstruction algorithms available on a GE scanner. The algorithms that do not have a "hump" in the MTF are: soft, standard, detail, bone, and edge in order of increasing sharpness. A hump in the MTF indicates a form of edge enhancement of the image; these are evident in the lung reconstruction and the bone plus reconstruction. MTF (v) 1.8 Soft 1.6 Standard 1.4 Detail 1.2 Bone 1.0 Bone Plus 0.8 Edge 0.6 Lung 0.4 0.2 0.0 0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 v (cycles/cm) 16

Image Reconstruction Filters, Algorithms, or Kernels (Siemens) Kernel names have 4 positions. Example: B31s. Pos.1: kernel type (B=body, C=child head, H=head, U=ultra high resolution, S=special kernel, T=topo.) Pos. 2: resolution (1,...,9. Higher number -> higher resolution) Pos. 3: version (0,...,9) Pos. 4: scan mode (f=fast (no j-ffs, no UHR comb), s=standard (with j-ffs, no UHR comb), h=highres (with j-ffs, no UHR comb), u=ultrahighres (with j-ffs, with UHR comb)) The use of z-ffs is not coded to the kernel name. B10f, B18f, B19f, B20f, B25f, B29f, B30f, B31f, B35f, B36f, B39f, B40f, B41f, B45f, B46f, B47f, B50f, B60f, B65f, B70f, B75f, B80f, B08s, B10s, B18s, B19s, B20s, B25s, B29s, B30s, B31s, B35s, B39s, B40s, B41s, B45s, B46s, B47s, B50s, B60s, B65s, B70s, B75h, B80s. B30m/B40m (m=f,s) are standard kernels, B20m/B10m are more smooth. Br1m (r=3,4) have about the same visual sharpness as Br0m, but a finer noise structure (better image impression, improved LC). B25m correspond to kernels B30m with ASA. B35m is designed for Ca-scoring and quantitative analysis, B36f is a sharper version of B35f, B45m has intermediate resolution. B46m is for investigations of patency of stents and for quantitative investigations. B47m have resolution between B46m and B50m. B50m, B60m, B70m are sharper kernels (for cervical spine, shoulder, extremities, thorax). B65m is a special lunghr-kernel for quantitative evaluations. B80m is a special lunghr-kernel (corresponding to HCE with B40m); not as sharp as B70m. B18m are the kernels B10m with stronger de-ringing. Br9m (r=1,2,3) are the PET/SPECT versions of Br0m. B75m (m=f,h) are lunghr kernels with less overshooting at edges. H10f, H19f, H20f, H21f, H22f, H23f, H29f, H30f, H31f, H32f, H37f, H39f, H40f, H41f, H42f, H45f, H47f, H48f, H50f, H60f, H10s, H19s, H20s, H21s, H22s, H23s, H29s, H30s, H31s, H32s, H37s, H39s, H40s, H41s, H42s, H45s, H47s, H48s, H50s, H60s, H70h, H80h (Open only). H40m (m=f, s) is the standard kernel, H30m, H20m or H10m lead to softer images. Hr1m (r=2,3,4) yield the same visual sharpness as Hr0m, but have a finer noise structure (better image impression, improved LC). Therefore Hr1m are used in standard protocols. Hr2m are the kernels Hr0m without PFO. H23m serves for Neuro PBV. H37m, H47m, H48m are alternative kernels with different noise impression. H45m serve for intermediate resolution, H50m, H60m are sharper kernels. Hr9m (r=1,2,3) are the PET/SPECT versions of Hr0m. H70h gives highest resolution without comb. H80h gives the hires specification for Sensation Open. 17

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 18

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 19

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 20

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 21

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 22

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 23

Image Sharpness: Images of a resolution pattern made with different Image Reconstruction Algorithms 24

Image Sharpness: Increasing the image sharpness Positive effects: Allows visualization of finer detail Negative effects: Increases the image noise Increases image artifacts 25

Image Noise: Images of a uniform water pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 26

Image Noise: Images of a uniform water pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 27

Image Noise: Images of a uniform water pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 28

Image Noise: Images of a uniform water pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 29

Image Noise: Images of a uniform water pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 30

Image Noise: Images of a uniform water pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 31

Image Noise: Image noise is reduced by: Using a less sharp or softer image reconstruction algorithm Increasing the dose to the detectors used in reconstructing the image slice. Noise 1/ Dose To reduce noise by a factor of 2, increase the effective mas by a factor of 4. Increasing the slice thickness by a factor of 4 decreases the noise by a factor of 2. 32

Low Contrast Detectability: Images of a low contrast detectability pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 50 mas 33

Low Contrast Detectability: Images of a low contrast detectability pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 34

Low Contrast Detectability: Images of a low contrast detectability pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 200 mas 35

Low Contrast Detectability: Images of a low contrast detectability pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 36

Low Contrast Detectability: Images of a low contrast detectability pattern imaged using mas values incrementing by a factor of 2, from 50 mas to 800 mas. 37

Artifacts from Data Acquisition Problems Ring Artifacts 38

Artifacts from Data Acquisition Problems Ring Artifacts 39

Artifacts from Data Acquisition Problems Ring Artifacts 40

Beam Hardening & Partial Volume Artifacts 41

Beam Hardening & Partial Volume Artifacts Reconstruction without iterative beam hardening correction Reconstruction with iterative beam hardening correction 42

Motion Artifacts No Motion With Motion 43

Artifact due to the patient extending outside the Scan Field of View; ALSO Stringy noise artifact 44

Effect of CT Protocols on Image Quality and Dose 45

Axial Scan Techniques Affecting Image Quality & Dose kv mas ma & rotation time Slice thickness 46

Helical Scan Techniques Affecting Image Quality & Dose kv mas ma & rotation time Slice thickness Pitch 47

Helical Scan Techniques Affecting Image Quality & Dose Definition of Pitch for Multislice Helical / Spiral Scanning: Pitch coll = Table travel per 360 tube rotation Total collimation width of all simultaneously collected slices 48

Helical Scan Techniques Affecting Image Quality & Dose The image noise and patient dose for helical scanning is generally a function of ma x rotation time / pitch which is often referred to as Effective mas : Effective mas = mas / pitch This is analogous to the use of CTDI vol : CTDI vol = CTDI w / pitch 49

Helical Scan Techniques Affecting Image Quality & Dose Effective mas = mas / pitch Siemens and Toshiba scanners use the term Effective mas in their scan techniques. Phillips uses the term mas/ slice, which means the same as effective mas. 50

Helical Scan Techniques Affecting Image Quality & Dose Effective mas = mas / pitch You may change the ma, rotation time, or pitch values, but if the effective mas remains constant, so does the CTDI vol and the patient dose. If the effective mas remains constant the image noise will also remain constant or nearly so. 51

Manual vs. Automatic Exposure One deficiency of CT Scanners before 2001 They did not contain any type of phototimer or automatic exposure control (AEC) to assure a proper patient dose. Therefore, manual technique charts were needed for different patient sizes. Usually this was not done so that techniques more suited for larger patients were used on all patients resulting in unneeded radiation exposure. 52

Automatic Exposure Control in CT Scanners Modern CT scanners have some type of automatic exposure control (AEC) that changes the ma during the scan. There are two basic types of AEC that can be used separately or together: The scanner varies the ma at different axial positions of the patient. The scanner varies the ma as the tube rotates around the patient. It is optimal to use both types together if the scanner allows it (Most do allow it). 53

Automatic Exposure Control in CT Scanners The scanner varies the ma at different axial positions of the patient. The scanner varies the ma at different axial positions of the patient and also varies the ma as the tube rotates around the patient. 54

Automatic Exposure Control in CT Scanners Caution: The methods used by different manufacturers to perform AEC in CT are very different and may achieve very different clinical results. 55

Automatic Exposure Control in CT Scanners Some scanners (GE, Toshiba) try to keep the image noise constant as patient size increases: the automatic exposure control is adjusted by selecting the amount of noise that you wish in the image. This is done by selecting a Noise Index or SD (standard deviation). Typical values of Noise Index are 2.5 to 3.5 for a standard adult head scan and 12 to 20 for the body (for a 5 mm slice thickness). The scanner attempts to keep the image noise constant by adjusting the ma within set limits. 56

Automatic Exposure Control in CT Scanners For scanners that use a Noise Index or SD for AEC: The dose for a scan depends both on the Noise Index or SD AND the slice thickness selected for the first image reconstruction. Let s say you want to view reconstructed slice thicknesses of both 5 mm and 1.25 mm: Suppose the first image reconstruction has a slice thickness of 5 mm with a Noise Index of 12. If the first image reconstruction is switched to a slice thickness of 1.25 mm, the Noise Index needs to be changed to 24 to keep the dose constant. 57

Automatic Exposure Control in CT Scanners For scanners that use a Noise Index or SD for AEC: Example: The same ma values and the same patient dose will result from the following settings: Noise Index or SD Slice Thickness 12 5.0 17 = 12 x 2 2.5 24 = 12 x 2 1.25 58

Automatic Exposure Control in CT Scanners GE Example: Smart ma adds rotational variation of the ma to the axial variation performed in Auto ma without Smart ma. Therefore always press the Smart ma button when using Auto ma With GE scanners you must select whether you will be using manual techniques Manual ma or AEC techniques Auto ma. Manual ma uses an actual ma setting, Auto ma uses a Noise Index setting. Having one set correctly in a protocol does nothing to insure the other is properly set. 59

Automatic Exposure Control in CT Scanners Other scanners (Siemens, Philips) allow you to select the mas, Effective mas, or the mas/ slice that you would use for an reference size patient. For Siemens scanners this selection is called the Quality reference mas. In AEC mode the scanner then automatically increases or decreases the effective mas for larger or smaller patients. This is done by varying the ma. Effective mas mas/ slice = (ma x rotation time) / pitch 60

Automatic Exposure Control in CT Scanners Siemens: With Siemens scanners you select the Eff. mas whether you will be using manual techniques OR AEC techniques. In manual mode this is the actual eff. mas used and in AEC mode it is the eff. mas that you would desire for an reference size patient. There is not the use of 2 different parameters for manual & AEC mode. 61

Automatic Exposure Control in CT Scanners Scanners that try to keep the image noise constant have the problem that they can quickly reach the maximum ma ceiling before getting to very large patients. Scanners that use a reference mas setting will generally allow the ma to increase only modestly with increased patient size, allowing the image noise to increase substantially for large patients. What is needed is a new AEC approach and the use of higher kv for larger patients. 62

Automatic Exposure Control in CT Scanners 63

Automatic Exposure Control in CT Scanners A Concern with All CT Scanner: Proper centering of the patient is very important for the proper operation of the AEC system. A common problem is mis-centering the patient too low in the scan field. 64

Automatic Exposure Control in CT Scanners A Concern with All CT Scanner: Patient positioned 6 cm too low Patient positioned properly Midpoint of scout and scan field Midpoint of scout and scan field 65

Automatic Exposure Control in CT Scanners A Concern with Older Siemens Scanners: After performing the topo scan, Siemens scanners warn you if the available effective mas is lower than the effective mas requested by the AEC system, which would result in unacceptable image quality. However the scanner does not let you increase the kv, for example, from 120 to 140 which could solve the problem! This means a work-around is required: temporarily reduce the Quality Ref mas to a very low value. You can then raise the kv to 140. Then increase the Quality Ref mas to at least ½ of its original value, if possible. 66

Automatic Exposure Control in CT Scanners A Concern with GE and Toshiba Scanners: Since the GE and Toshiba scanners require two separate parameters for determining the ma in manual and AEC mode, one must understand the proper use of the Noise Index or SD parameter when using AEC. When switching from manual to AEC mode or from AEC to manual mode one must be sure that the exposure parameter of manual ma or Noise Index/ SD is properly adjusted. When one of these modes is the manufacturer s default mode, one should not assume that correct settings will result when switching modes. 67

Manual vs. Automatic Exposure - kv Adjustment - Increasing the kv will have different effects when using manual exposure mode and different types of automatic exposure modes. 68

Manual vs. Automatic Exposure - kv Adjustment - For all CT scanners: In a manual mode increasing the kv will always increase the patient dose, if all other scan parameters are kept constant. 69

Manual vs. Automatic Exposure - kv Adjustment - In AEC Mode: With GE and Toshiba scanners, increasing the kv will decrease the patient dose, if all other adjustable scan parameters are kept constant (ma will decrease) With Siemens and Philips scanners, increasing the kv will increase the patient dose, if all other adjustable scan parameters are kept constant (ma will remain nearly constant) 70

Optimizing CT Protocols: Misconceptions and Recommendations for Scan and Imaging Parameters 71

kv Misconceptions: Scanning at 140 kv will reduce patient dose for any type of CT scan: head, body, adult or pediatric. For head scans, 140 kv should be used through the posterior fossa region to reduce image artifacts from bone. 72

kv Recommendations: If we ignore beam hardening artifact limitations and CT scanner power limitations: The theoretical optimal kv, for any CT imaging, is the kv that will give the highest ratio of contrast to noise at a given patient dose. 73

kv Recommendations: For all Head CT scans and all Head or Body Pediatric scans this theoretical optimal would be close to 80 kv. For Adult Body CT scans this theoretical optimal will range from 80 kv up to 140 kv. 74

kv Recommendations: Modern CT scanners now have higher x-ray power & much more efficient use of this power through multi-slice design. They also have improved beam hardening/ bone correction algorithms. These improvements allow you to use lower kv settings closer to the theoretical optimal. 75

Optimal kv Technique Setting for Axial or Helical Scanning kv - Head CT Peds and Adult Use 80 kv for Peds Head 0-2y w/wo IV contrast. Use 80 kv for Peds Head 2-6y w IV contrast. Use 100 kv for Peds Head 2-6y wo IV contrast. Use 100 kv for Adult Head w IV contrast. Use 120 kv for Adult Head w/o IV contrast. 76

Optimal kv Technique Setting for Axial or Helical Scanning kv Body CT - Peds Use 80 kv for Peds Body for whom the sum of lateral and AP dimensions is less than 44 cm. Use 100 kv for Peds Body for whom the sum of lateral and AP dimensions is between 44-55 cm. Use Adult protocols for larger patients. 77

Optimal Technique Setting for Axial or Helical Scanning kv Body CT Adults wo IV contrast Use 100 kv for Small Adults for whom the sum of lateral and AP dimensions is less than 55 cm. Use 120 kv for Medium Size Adults. Use 140 kv for Large Adults for whom the sum of lateral and AP dimensions is greater than 75 cm. 140 kv for Large Adults reduces image noise and provides better image quality without large exposure increases. 78

Optimal Technique Setting for Axial or Helical Scanning kv Body CT Adults w IV contrast Use 80 kv for Small Adults for whom the sum of lateral and AP dimensions is less than 55 cm. Use 100 kv for Medium Size Adults. Use 120 kv for Large Adults for whom the sum of lateral and AP dimensions is greater than 80 cm. Note: the use of lower kv produces a significant increase in the contrast of iodine, with better optimization of contrast to noise. 79

kv Recommendations: For scanning the neck or upper thorax, the amount of lateral attenuation through the shoulders is a serious problem. It will cause some degree of horizontal streaking artifact through the shoulder, which is actually a noise effect. 80

kv Recommendations: Here the solution is to increase the kv from 120 kv to 140 kv to reduce the amount of lateral attenuation through the shoulders as much as possible and thus reduce this noise streaking artifact. 81

kv and Pitch - Pediatric Misconceptions: Using 140 kv for children to reduce dose. On the contrary this will generally raise the dose for equal image quality and is not recommended. Using a pitch greater than 1.0 for children is often strongly recommended to reduce radiation dose. This is usually totally misguided, as we will see shortly. 82

Pitch Misconceptions Scanning at higher pitch should be used as a strategy to reduce adult or pediatric patient dose and is the best way to reduce scan time and motion artifact and blur. WRONG!!! 83

Pitch Misconceptions: A pitch of less than one over-irradiates the patient due to scanning overlap, and thus wastes radiation dose. Thus one should avoid using a pitch less than one, particularly in pediatric scans. WRONG!!! 84

Pitch Recommendations: Changing the pitch from 1.0 to 0.5 increases the patient dose by a factor of 2 but also decreases image noise. The effects on dose and noise are the same as increasing the ma or the rotation time by a factor of 2, but with the added advantage of decreasing helical artifacts. 85

Pitch Recommendations: The effect of increased dose at lower pitch is easily countered by reducing the rotation time or ma in manual mode. There is NO increase in dose when decreasing pitch in AEC mode since the AEC mode in all scanners will keep the dose constant. 86

Pitch Recommendations: Lowering the pitch and decreasing the exposure time by the same factor will keep the patient dose and exam time constant, but provide better image quality you get something for nothing! 87

Pitch Recommendations: Example: Change a 1.0 sec rotation time and a pitch of 1.6 to a 0.5 sec rotation time and a pitch of 0.8 88

Pitch Recommendations: For head scanning ALWAYS use a pitch of less than 1.0 to minimize helical artifact. Best results are usually obtained with a pitch just above 0.5: 1. 89

Pitch Recommendations: For body scanning use a pitch of less than 1.0 whenever possible to minimize helical artifact and allow more radiation for the adequate imaging of larger patients. When decreasing pitch in body scans, you need to be aware of breath hold limitations and contrast considerations. 90

Dose Reduction Recommendations: Instead of increasing pitch, the proper dose reduction strategy is: 1. Reduce the rotation time (will reduce dose in manual mode and is the first step in AEC mode). 2. Reduce the effective mas (in manual or AEC mode); reduce the ma (in manual mode); or increase the noise index or SD (in AEC mode). 3. Only then increase pitch if required to reduce total exam time. 91

Axial vs. Helical Scanning Misconceptions: Heads should always be scanned using the axial rather than the helical mode or you will get a lower quality image. 92

Axial vs. Helical Scanning Recommendations: Helical scanning will almost always allow an exam with equal or better image quality than an axial scan if you have a CT scanner with 16 or more slices and select proper scan techniques. Axial scanning is still useful if required for positioning of the patient to avoid artifacts, since tilting the gantry is not allowed with helical scanning. 93

Axial vs. Helical Scanning and slice reconstruction interval Recommendations: Advantages of Helical scanning: Shorter total scan time with less chance for patient motion during the scan. The ability to reconstruct slices at intervals less than the slice thickness. VERY IMPORTANT! 94

Axial vs. Helical Scanning and slice reconstruction interval Recommendations: With axial scanning, the slice reconstruction incrementation is normally equal to the slice thickness. 95

Axial vs. Helical Scanning and slice reconstruction interval Recommendations: With helical scanning, the slice reconstruction incrementation can be set at any value. The best z-resolution is obtained by reconstructing at intervals ½ of the actual slice thickness this particularly helps with multiplanar reformatting. This is a significant advantage of helical scanning that is often not utilized. 96

Axial vs. Helical Scanning and slice reconstruction interval Recommendations: When creating slices for reformating of axial images to a modified axial plane, or for sagittal or coronal images, ALWAYS use thin slices as the source images, if this is not done automatically by the scanner. DO NOT USE 5 mm slices! For soft tissue recons use 1.0 to 1.5 mm slice thickness. For bone or high res recons use 0.5 to 0.75 mm slice thickness. 97

Detector Configuration Misconceptions: The acquisition slice width (acquisition detector configuration) can be equal to the reconstructed slice thickness without causing image degradation. 98

Detector Configuration Recommendations: Streaking artifacts off of bone and air are due to both beam hardening and partial volume artifacts. Thus it is important to use scan techniques to reduce partial volume artifacts. 99

Detector Configuration Recommendations: To minimize partial volume artifacts in head scans always use the smallest detector width in acquiring the scan data, regardless of the imaged slice thickness. You may be restricted in some body scans since using the smallest detector width can also reduce the total beam width and increase the exam time. 100

Detector Configuration Recommendations: For head scans this means using 16 x 0.5 mm, 16 x 0.6 mm, or 16 x 0.625 mm for a 16 slice scanner and using 32, 40, or 64 x 0.5, 0.6, or 0.625 mm for 32 to 64 slice scanners. Do not use 2.5 mm or 5 mm acquisition for example. 101

Detector Configuration Recommendations: With a GE 16 slice scanner you can use a 16 x 0.625 acquisition for the best quality in the head, but this only gives you a 10 mm beam width. In the body you may need to go to a 16 x 1.25 acquisition which provides a 20 mm beam width and allows you to scan at twice the speed. 102

Detector Configuration Recommendations: However the 16 slice GE scanner also allows you to use it in an 8 slice mode: 8 x 1.25 mm or 8 x 2.5 mm - which should NEVER BE USED. Likewise the GE 8 slice scanner can also be used in a 4 slice mode - which again should NEVER BE USED. 103

CT Protocols Abdomen/ Pelvis --Medium Adult-- From Recon 2: Sa & Co Reformat: Ave., 5.0 mm thick 2.5 mm interval CT 1 CT 2 CT 3 CT 4 & East & RP CT Scanner GE Opt CT580W GE LS 16 GE LS 16 Pro GE LS VCT 64 GE LS 8 Scan Type Helical Helical Helical Helical Helical Detector Coverage (mm) Beam Collimation (mm) 20 20 20 40 10 Detector Rows 16 16 16 64 8 Detector Configuration 16 x 1.25 16 x 1.25 16 x 1.25 64 x 0.625 8 x 1.25 Scan FOV Large Large Large Large Body Large Pitch 0.938 0.938 0.938 0.516 1.35 Speed (mm/rot) 18.75 18.75 18.75 20.64 13.5 Rotation Time (sec) 0.7 0.7 0.6 0.4 0.6 kv 120 120 120 120 120 Smart/ Auto ma or Manual ma Smart ma Auto ma Smart ma Smart ma Auto ma Smart ma/ Auto ma Range 60-660 50-440 60-660 50-400 80-440 Noise Index 12 12 12 15 12 (Manual ma) (470) (380) (440) (330) (440) % Dose Reduction (ASiR) None Slice Thickness (mm) 5.0 5.0 5.0 5.0 5.0 Interval (mm) 3.0 3.0 3.0 3.0 3.0 104

CT Protocols Recon 1: DFOV 36 36 36 36 36 Recon Type Standard Standard Standard Standard Standard WW/ WL 325/15 325/15 325/15 325/15 325/15 Recon Option Plus Plus Plus Plus Plus ASiR Setup Slice 40% Recon 2: DFOV 36 36 36 36 36 Recon Type Standard Standard Standard Standard Standard WW/ WL 325/15 325/15 325/15 325/15 325/15 Recon Option Plus Plus Plus Plus Plus Recon Option IQ Enhance ASiR Setup Slice 40% Slice Thickness (mm) 1.25 1.25 1.25 1.25 1.25 Interval (mm) 0.625 0.625 0.625 0.625 0.625 105