SBP270 BAROMETRIC PRESSURE SENSOR INSTRUCTION MANUAL



Similar documents
APPLICATION NOTE. Measuring Current Output Transducers with Campbell Scientific Dataloggers. App. Note Code: 2MI-B Revision: 1

LLAC4 Four-Channel, Low-Level AC Conversion Module Revision: 3/12

Absolute and Relative Gas Concentration: Understanding Oxygen in Air

Model SRMD Setra Remote Monitoring Display

CURS Ohm Current Shunt Terminal Input Module Revision: 5/07

INSTRUCTION MANUAL. MCR1 Memory Card Reader 5/93. Copyright (c) 1993 Campbell Scientific, Inc.


BroadBand PowerShield. User Manual

SDX Submersible Depth Transmitter User Manual

USER MANUAL. Model U ForeFront Alarm Card SALES OFFICE (301) TECHNICAL SUPPORT (301) An ISO-9001 Certified Company

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

User's Guide. Integrating Sound Level Datalogger. Model Introduction

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

Global Water Instrumentation, Inc.

RPS High Accuracy Resonant Pressure Sensor. GE Measurement & Control Solutions. Features:

Advantium 2 Plus Alarm

USER MANUAL. Kramer Electronics, Ltd. Models:

USER MANUAL. Kramer Electronics, Ltd. Models:

PMP 4000 Series. Druck Amplified Output Pressure Transducers. GE Sensing. Applications. Features

Global Water Instrumentation, Inc.

Single Station Remote Alarm

Global Water Instrumentation, Inc.

MACTek Corporation. USB HART Interface. Users Manual

Double Coated Urethane Foam Tapes

WHITE PAPER. Preventing and Attacking Measurement Noise Problems. White Paper Code: 2 Revision: 1. Copyright (C) 2001 Campbell Scientific, Inc.

Wireless Base Station Single and Multiple Headsets

Water Leak Detection System

INSTALLATION INSTRUCTIONS BOOST CONTROLLER. Pro Control Input (Optional) Tach. Signal. Speed. Gray. Signal. Green Blue. Orange.

T0118 T2118 T3118. Instruction Manual

Triple Expansion Chassis

PRODUCT PROFILE FEATURES

Global Water Instrumentation, Inc.

How To Use A Cdm250 Digital Multimeter

Quick Start Guide. Internet Conference Station X1-N6.

Signal Manager Installation and Operation Manual

Recommended Tools and Supplies: Small Flat Blade Screwdriver, 35mm x 7.5mm DIN Rail. Qwik Install

Digital echo-charge. Owner s Manual. Xantrex Digital echo-charge Battery Charger

Mini Digital Altimeter & Climb rate & Barometer & Thermometer & Compass & Weather forecast & Time

CHAPTER 1. GETTING STARTED... 1

SEMICONDUCTOR TECHNICAL DATA

Tipping Bucket Rain Gauge Models 6011-A 6011-B

About Your NetBotz 4-20mA Sensor Pod

Series 510 Submersible Level Transmitters

i410/i1010 AC/DC Current Clamp

Solar Power Systems Models: PVS220W-24 and PVS220W-48 Installation Manual

Powerware Relay-Serial Card User s Guide

Humidity Sensor. Model HU-224/225

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

RPS/DPS High Accuracy Resonant Pressure Sensor. GE Measurement & Control. Features:

LOSB9595 1/36 th Brushless ESC and 10250Kv Motor Combination

Part I - Installation

INTMOD485-LH Protocol Converter

Sip/Puff Switch User Guide

2-3 SAS/SATA II HDD Canister USER S MANUAL XC-23D1-SA10-0-R. Document number: MAN A

Installation and Operation Guide for PD5100 Automatic Transfer Switch

User's Manual. Research Isometric Transducer 0 to 5 grams and 0 to 50 grams

Indoor/Outdoor Color Camera with Built-in 2.4 GHz Wireless Transmitter, plus X10 controlled power supply, and Video Receiver.

ECH 2 O. Soil Moisture Sensor. Operator s Manual. For Models EC-20, EC-10, and EC-5. Version 5

ExStik TM ph Waterproof Meters


Installation Instructions

USB Rover port USB m Cat 5 Extender. User Guide

INSTRUCTION MANUAL MODEL PCD-88

16/32 Channel 1U Rack Mount CCTV Power Supply

Amplified High Speed Fiber Photodetectors

WINEGARD MOTORIZED SENSAR ANTENNA Models MA1055W & MA1055G MADE IN U.S.A. U.S. Patents D500,496 and 7,358,909 INSTALLATION MANUAL

CURS Ohm Current Shunt Terminal Input Module Revision: 7/13

Original instructions Tina 3A/Aps Adaptor unit

VERTECH II OPERATION AND SETUP SKI FEATURES SKI AND ALPIN OPERATION

MS900AC-B-LS User Information

VEHICLE SECURITY SYSTEM G25/G20

DPI 260 SERIES: Digital Pressure Indicators

VK-250 WARRANTY REGISTRATION FORM

NSI DDR DIMMING RACK POWER DISTRIBUTION UNIT

HIH-4000 Series Humidity Sensors

Propane Conversion Kit PROPANE CONVERSION KIT INSTALLATION MANUAL

TABLE 1: Wiring Terminals. Connect to... 1C 1H 2C 2H 1H1C 2H1C 2H2C 3H2C

AGS. Owner's Manual. Xantrex Automatic Generator Start Control System

Dual Laser InfraRed (IR) Thermometer with Color Alert

User guide Conference phone Konftel 100

Rackmount Intelligent Pressure Scanners

How To Size A Stenner Pump

Tina 2A/B Adaptor unit

AT&T. PARTNER Plus Door Phone. Installation and Operation Manual

Dual Laser InfraRed (IR) Thermometer

Installation Guide. WSD-100 Wind Speed and Direction Sensor For XR5 Data Loggers. February, 2011

1000 Series Data Loggers PRODUCT MANUAL

K Standard Grade Ni-Cr vs Ni-Al (Chromel vs Alumel)

PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller

ENGLISH FRANÇAIS DEUTSCH ESPAÑOL ÐÓÑÑÊÈÉ

TS-300-MSF TS-300 INSTALLATION AND CONFIGURATION MANUAL

Specifications - Installation and Operating Instructions Ø5 [127.00] 17/64 [6.91] 1-59/64 [48.89]

DCX300 - DCX400 - DCX600

UNIK 5800/5900. Flameproof/Explosion-Proof Pressure Sensing Platform. GE Measurement & Control. High Quality. Features. Bespoke as Standard.

Technical Manual. For use with Caller ID signaling types: Belcore 202, British Telecom, & ETSI

Subminiature Load Cell Model 8417

Transcription:

INSTRUCTION MANUAL REVISION: 29/11/01 COPYRIGHT (c) 1989, 1991 CAMPBELL SCIENTIFIC, INC.

WARRANTY AND ASSISTANCE The SBP270 BAROMETRIC PRESSURE SENSOR is warranted by CAMPBELL SCIENTIFIC (CANADA) CORP. ( CSC ) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. CSC's obligation under this warranty is limited to repairing or replacing (at CSC's option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CSC. CSC will return such products by surface carrier prepaid. This warranty shall not apply to any CSC products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CSC is not liable for special, indirect, incidental, or consequential damages. Products may not be returned without prior authorization. To obtain a Return Merchandise Authorization (RMA), contact CAMPBELL SCIENTIFIC (CANADA) CORP., at (780) 454-2505. An RMA number will be issued in order to facilitate Repair Personnel in identifying an instrument upon arrival. Please write this number clearly on the outside of the shipping container. Include description of symptoms and all pertinent details. CAMPBELL SCIENTIFIC (CANADA) CORP. does not accept collect calls. Non-warranty products returned for repair should be accompanied by a purchase order to cover repair costs.

TABLE OF CONTENTS PAGE Description...1 Specifications...1 Operation...1 Datalogger Connection...2 Multiplier and Offset Calculation...3 Maintenance...4 Conversion Factors...4 Resolution Considerations...4 Program Example 1 CR10, 21X...4 Program Example 2 CR7...5 FIGURES 1. SBP270 Barometric Pressure Sensor...2 APPENDICES A. Mean Sea Level Pressure (AES Standard)...7 B. Altimeter Setting Computations...8

INSTRUCTION MANUAL DESCRIPTION The SBP270 Barometric Pressure Sensor (SBP270) is a high accuracy barometer designed for use with the CR7, 21X, and CR10 dataloggers. Figure 1 shows the SBP270, which includes the Setra Model 270 variable capacitance barometer, interface circuitry, 5 foot cable, and a rain-tight enclosure. Mounting brackets allow the SBP270 to be mounted either to a flat surface, or 1-1.5" NPT pipe. SPECIFICATIONS Pressure Range: 800 to 1100 millibars (mbar) Operating Temperature: 0-175 F *Accuracy: +/- 0.2 mbar Zero Pressure Output: 0 mbar, internally adjustable, factory set within < +/- 0.3 mbar Long Term Stability: < +/- 0.1% FS over 6 months 70 F Thermal zero shift: < +/- 0.2% FS/100 F (30 to 120 F) Warm-up time: 10 milliseconds (ms) Power requirements: 9.6-16 VDC 35 ma @ 12 VDC active < 25 ua quiescent Output: 0-250 millivolts (mv) Enclosure: 9.5" (H) x 7.5" (W) x 5.0" (D) Weight: 6 lbs * Accuracy as RSS of non-linearity, hysteresis, and non-repeatability. OPERATION Interface circuitry contained in the SBP270 (Figure 1) has a "control" input which wires to a switched excitation or control port on the datalogger. When the datalogger applies voltage (+1.5 to +5.0 volts) to the "control" input, circuitry is enabled which doubles the 12 volt supply voltage and applies this voltage to the barometer. Output from the SBP270 is a differential 0-250mv. A 10ms warm-up period is required before the measurement is made. After the measurement the "control" input should be returned to 0 volts, which switches power off to the barometer. 1

RED BLACK BLUE RED BLACK BLUE YELLOW GREEN WHITE YELLOW GREEN WHITE CLEAR Figure 1. SBP270 Barometric Pressure Sensor NOTE: The black outer jacket of the cable is Santoprene rubber. This compound was chosen for its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude its use inside buildings. DATALOGGER CONNECTIONS A 5 foot cable is included with the SBP270 which connects to the CR7, 21X, or CR10 datalogger. Lead wires are connected to the datalogger as shown in Figure 1. Note that when used with the CR10, the black lead wire (of the white/black pair) connects to an Analog Ground (AG); the other 2 lead wires labeled GND connect to power ground (G). 2

MULTIPLIER AND OFFSET CALCULATION Output from the SBP270 is 0 to 250mv, which corresponds to 800 to 1100 mbars of pressure. The multiplier and offset used to convert voltage to mbars of pressure is: mbar = (mv * 1.2) + 800mbar Multiplier = 1.2 Offset = 800 The measurement result using a multiplier of 1.2 and an offset of 800 is true barometric pressure in mbars. The weather service and most airports, radio stations and TV stations correct the pressure recorded at a particular station to what it would be if the station was located at sea level. This is done so that weather forecasters can obtain a clearer picture of what is happening as a storm crosses over mountains or high plateaus. Use Equation 1 to determine the pressure correction factor, which is added to the 800 mbar offset to have the datalogger output barometric pressure corrected to sea level. Because barometric pressure is strongly affected by elevation, the elevation of the site should be as accurate as possible. 5.253283 ELEVATION P = 1013. 25 1 1 44307. 69231 Equation 1** where elevation is in meters above sea level. For example, the correction factor (P) for Logan Utah at 4450 ft elevation is: 4450 ft * m/3.281 ft = 1356.29 m pressure correction factor (P) = 152.67 mbar transducer offset = 800 mbar pressure correction factor (P) = 152.67 mbar offset for sea level correction = 952.67 mbar ** Wallace, John M., and Peter V. Hobbs. 1977. Atmospheric Science an Introductory Survey. Academic Press, New York, NY, p 59-61. 3

MAINTENANCE The SBP270 requires minimal maintenance. The foam diffuser (see Figure 1) should be kept clean and unobstructed. Desiccant packs inside of the enclosure must be replaced as necessary. The desiccant packs can be reactivated by placing in an oven at 250 degrees F for 16 hours, or extra desiccant packs can be ordered from CSI. CONVERSION FACTORS mbar * 0.0145 PSI/mbar = PSI mbar * 0.75006 mm of Hg/mbar = mm of Hg mbar * 0.02953 in. of Hg/mbar = in. of Hg mbar * 0.00102 kg/cm 2 /mbar = kg/cm 2 mbar * 0.1 kpa/mbar = kpa RESOLUTION CONSIDERATIONS The value stored in final storage with the low resolution format is reduced to 3 significant digits when the first (left most) digit is 7 or greater. This rounding will occur when the SBP270 measurement result is in mbars or mmhg. For example, 860.65 mbars would be stored as 861. One way to preserve the resolution of the measurement is to specify the high resolution format using Instruction 78. For example, 860.65 mbars would be stored as 860.65 in final storage. The disadvantage of using high resolution is that it takes twice as much memory per data point. An alternative to high resolution would be to use an offset in the measurement instruction to keep the result less than 700. For example, 400 mbars could be used as an offset instead of 800 mbars, and then compensated for in the computer. This method would use low resolution, maximizing the number of data values which could be stored in final storage. PROGRAMMING EXAMPLE 1 - CR10, 21X Programming Example 1 uses Instruction 8, contained in the CR10 and 21X instruction set, to switch power to the SBP270 and to measure its output. Instruction 8 applies 2500 mv to excitation Channel 1 (wired to the "control" input to the SBP270), delays for 10ms, and then measures the output on differential Channel 1. The multiplier and offset are applied, and the result placed in Input Location 1. Every 15 minutes the output flag is set, and the average barometric pressure is stored in final storage using the high resolution format. * 1 Table 1 programs 01: 300 Sec. Execution Interval 01: P8 Excite,Delay,Volt(DIFF) 01: 1 Rep 02: 4 250 mv slow Range CR10, 500 mv slow range 21X 03: 1 IN Chan 04: 1 Excite all reps w/exchan 1 Switches power on to barometer 05: 1 Delay (units=.01sec) 10 ms delay 06: 2500 mv Excitation 07: 1 Loc : Corrected BP in mbars 08: 1.2 Mult 09: 952.67 Offset (800 @ sea level corrected to Logan Utah elevation) 4

02: P92 If time is 01: 0 minutes into a 02: 15 minute interval 03: 10 Set high Flag 0 (output) 03: P77 Real Time 01: 0110 Day,Hour-Minute 04: P78 Resolution 01: 1 High Resolution 05: P71 Average 01: 1 Reps 02: 1 Loc PROGRAMMING EXAMPLE 2 - CR7 Programming Example 2 uses a control port to switch power to the SBP270, which is necessary when the SBP270 is used with the CR7 datalogger (the CR7 doesn't contain Instruction 8, used in example 1). In the following example, control port 1 (wired to the white "control" input to the SBP270) is set high to switch power on to the barometer. Instruction 22 delays for 10ms to allow the sensor to warm-up. Output from the SBP270 is measured using Instruction 2; the multiplier and offset are applied, and the result placed in Input Location 1. Control Port 1 is set low, which switches power off to the barometer. Every 15 minutes the output flag is set, and the average barometric pressure is stored in final storage using the high resolution format. * 1 Table 1 Programs 01: 300 Sec. Execution Interval 01: P20 Set Port Switches power on to barometer 01: 1 Set high 02: 1 EX Card 03: 1 Port Number 02: P22 Excitation with Delay 10ms delay 01: 1 EX Card 02: 1 EX Chan 03: 0 Delay w/ex (units=.01sec) 04: 1 Delay after EX (units=.01sec) 05: 0 mv Excitation 03: P2 Volt (DIFF) Measure output 01: 1 Rep 02: 6 500 mv slow Range 03: 1 IN Card 04: 1 IN Chan 05: 1 Loc : Corrected BP in mbars 06: 1.2 Mult 07: 952.67 Offset (800 @ sea level corrected to Logan UT elevation) 5

04: P20 Set Port Switches power off to barometer 01: 0 Set low 02: 1 EX Card 03: 1 Port Number 05: P92 If time is 01: 0 minutes into a 02: 15 minute interval 03: 10 Set high Flag 0 (output) 06: P77 Real Time 01: 0110 Day,Hour-Minute 07: P78 Resolution 01: 1 High Resolution 08: P71 Average 01: 1 Reps 02: 1 Loc 6

Appendix A. Mean Sea Level Pressure Symbol: P MSL (millibars) Input: P stn - station pressure (mb) H - station elevation (m) T - station temperature (degrees C) T 12 - station temperature 12 hours prior to observation time (degrees C) Process: P MSL = P stn * EXP(0.0341636 * H / T mv ) where T mv = 273.16 + t + (a*h/2) + e s Ch + F(t) t = (T + T 12 )/2 a = lapse rate (0.0065 degrees C/metre) e s = (273.16+T)**(-0.00014t 2 + 0.0116t + 0.279) Ch = 2.8322E-9 * H 2 + 2.225E-5 * H + 0.10743 F(t) Plateau correction F(t) = at 2 + bt + c a, b, c are empirically derived and site specific (available from AES). NOTE: Reference: The Plateau Correction can cause a significant error in cold temperatures. Savdie, I. AES Barometery Program A Technical Record, Data Acquisition Branch Publication TR9, August 1982. 7

Appendix B. Altimeter Setting Computations Two Setra 270 pressure transducers are required to do altimeter setting computations for aviation interests. The reading should consist of an average of at least 60 samples taken uniformly from the sensors over a period of 40 to 120 seconds. If the maximum scatter of sample points around the mean is greater than 3 millibars, the pressure reading will default to MISSING. If the DC/DC converter voltage falls below the specified minimum sensor supply voltage of 22 volts, the pressure reading will be considered MISSING. The procedure for computing altimeter setting follows: 1. The lower of the two pressure readings is taken provided the difference between the two readings is less than or equal to 1.4 mb. 2. If the difference is unacceptable, both the altimeter setting and the station pressure are set to MISSING. 3. If the pressure readings are acceptable, the altimeter setting is calculated as follows: AS = 0.029530 * (P stn 0.19026 + 8.41717*10-5 H P ) 5.25593 where AS Altimeter Setting (inches) P stn Stations Pressure (millibars) Station Elevation (Geopotential Metres) H P 8