The Use of Hydrogen Carrier Gas for GC/MS



Similar documents
The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

Hydrogen as an alternative to Helium for gas chromatography.

CHROMOCTANE O N LINE ANALYZER

23 The Thermal Conductivity Detector

Pesticide Analysis by Mass Spectrometry

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Addressing the World Helium Shortage For Gas Chromatography

COMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.

Background Information

AxION edoor. Web-Based, Open-Access Mass Spectrometry Software

Electrospray Ion Trap Mass Spectrometry. Introduction

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# RFQ)

Diffusion and Fluid Flow

Validation of process gas systems

Chapter 19 Purging Air from Piping and Vessels in Hydrocarbon Service

Appendix 5 Overview of requirements in English

Micro GC Fusion Pre-Installation Checklist

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles

Clarus 580 Gas Chromatograph

Tuning & Mass Calibration

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES. Randall Messman SW Regional Sales Manager

AUTOMATIC CHANGEOVER MANIFOLD Series 915

Expectations for GC-MS Lab

Separation of Amino Acids by Paper Chromatography

Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification

Handling Gas: Cylinders vs Gas Generators

How To Use Gc-Ms

OEM-EP Pressure Controllers

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

Compressed Gas Cylinder Program

Quantitative Evaluation of Perfume Fleuressence Samples using the znose

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012.

Opiates in Urine by SAMHSA GC/MS

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

Retention Parameters in Column Chromatography

Chemical Fume Hoods. Environmental Health and Safety

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

Fractional Distillation and Gas Chromatography

Gas Custody Transfer Calibration

Compressed Gas Cylinder Storage and Handling Guide

How To Test For Contamination In Large Volume Water

Laminar Flow Elements

AMD Analysis & Technology AG

Thermal Conductivity Detector

Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS

GC Troubleshooting 10. How to use this Troubleshooting Help system GC Troubleshooting - Top Level Generic problem descriptions...

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

5973 inert Mass Selective Detector

Compressed Gas Cylinder Safety

DO MORE AND DO IT WITH CONFIDENCE. Empower 3 Chromatography Data Software

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies

Online Infrared Flue Gas Analyzer Gasboard 3000

Agilent 5975 Series MSD

NexION 300 ICP-MS. ICP-Mass Spectrometry PREPARING YOUR LAB. Preparation Checklist

Refrigerant Changeover Guidelines R-22 to R-407C. Leading the Industry with Environmentally Responsible Refrigerant Solutions

Fume Hood Questions and Answers

Certain specific properties of compressed gases make them highly useful in various research activities.

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

VAV Laboratory Room Airflow The Lowdown on Turndown

Custody Transfer Measurement. with the V-Cone Flowmeter

Agilent Veriflow 500 Electronic Flowmeter. Flow Range: sccm. Operation Guide. Agilent Technologies

BTU ANALYSIS USING A GAS CHROMATOGRAPH

Paper Chromatography: Separation and Identification of Five Metal Cations

Micro Cryo-Trap Model 971

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter

Exhaust emissions of a single cylinder diesel. engine with addition of ethanol

Micromass LCT User s Guide

Agilent 7697A Headspace Sampler Data Sheet

ORGANIC SAMPLE PREPARATION

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas

Analysis of Liquid Samples on the Agilent GC-MS

ChromQuest 5.0 Chromatography Data System

Theory and Instrumentation of GC Introduction

PSAC/SAIT Well Testing Training Program Learning Outcomes/Objectives Level D

Battery Powered Digital Mass Flow Meters for Gases

C146-E087F. GCMS-QP2010 Plus. Shimadzu Gas Chromatograph Mass Spectrometer

COMPRESSED GAS SAFETY

MassLynx 4.1 Getting Started Guide

Stainless Steel Pipe Fabrication

Transformer Oil Gas Analysis

Analyzing Small Molecules by EI and GC-MS. July 2014

Chemical Analysis. Natural Gas Analyzer. A Family of optimized GC Solutions. Gas Chromatography. think forward

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder

Setting up a Quantitative Analysis MS ChemStation

RGC-IR Remote Gas Calibrator for IR400

Air Eliminators and Combination Air Eliminators Strainers

GAS CHROMATOGRAPHY. DETERMINING THE PERCENT COMPOSITION OF A MIXTURE OF VOLATILE LIQUIDS USING A GAS CHROMATOGRAPH. (Revised: )

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

TEST REPORT for the ARID TECHNOLOGIES VAPOR RECOVERY UNIT installed in LANTANA, FL

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Micro Motion 3098 Gas Specific Gravity Meter

Transcription:

Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors for chemical laboratories Practical considerations for GC/MS method development Introduction Helium, as a limited natural resource, is increasingly expensive and, in some regions, in limited supply due to rationing. As such, the use of hydrogen as a carrier gas for gas chromatography mass spectrometry (GC/MS) is becoming more and more prevalent. Physical differences between hydrogen and helium give rise to chromatographic differences and the flammable nature of hydrogen increases safety concerns as well. In this document we will demonstrate the effective use of hydrogen with the PerkinElmer Clarus SQ 8 GC/MS and provide recommendations for ensuring laboratory safety. While the dangers of using hydrogen in a laboratory can be mitigated, every laboratory faces unique challenges and it is the responsibility of the laboratory manager and safety officer to address these issues to ensure the safety of their laboratory personnel. Extra caution must be taken in the design of both the laboratory and the experimental set-up. The development of safety conscious Standard Operating Procedures (SOP), including a thorough chemical hygiene plan, is a must. Polycyclic Aromatic Hydrocarbons by GC/MS Using Hydrogen Carrier Gas The determination of polycyclic aromatic hydrocarbon (PAH) compounds demonstrates the functionality of the PerkinElmer Clarus SQ 8 GC/MS using hydrogen carrier gas. The chromatograms presented in Figure 1 demonstrate the effectiveness of this system using both helium (Figure 1A) and hydrogen (Figure 1B) as carrier gas. A comparison of the two chromatograms clearly illustrates the faster elution times and improved peak sharpness one can expect when switching from helium to hydrogen.! Hydrogen is a flammable gas and when used incorrectly can pose a threat to both life and property. Prior to using hydrogen in a laboratory setting, safety precautions must be developed and subsequently observed to ensure safety.

A. 1 % B. 1 1,4-Dichlorobenzene 5.53 Naphthalene-d8 4.77 1,4-Dichlorobenzene % 3.42 Naphthalene-d8 7. Acenaphthene-d1 6.92 Acenaphthene-d1 9.24 Phenanthrene-d1 11.19 Chrysene-d12 14.9 Phenanthrene-d1 8.8 Chrysene-d12 12.21 Perylene-d12 14.2 Perylene-d12 17.6 3.3 5.3 7.3 9.3 11.3 13.3 15.3 17.3 19.3 Time Figure 1. Chromatogram of standard PAH mix using (A) helium and (B) hydrogen as the carrier gas with standards labeled with retention time in minutes and compound name. Scan EI+ TIC 1.94e8 Scan EI+ TIC 2.86e8 Guidelines for Using Hydrogen The following sections highlight the safety and performance considerations a laboratory will face when electing to use hydrogen as the carrier gas for their GC/MS experiments. The information presented here is provided as a guide only, and especially in the case of hydrogen safety, a full assessment of the dangers and benefits of using hydrogen should be explored prior to initiating any laboratory work. General Safety Considerations The main safety concern when using hydrogen in a laboratory is the ever-present danger of explosion stemming from its flammability. Hydrogen is flammable over a wide concentration range, and during the rapid expansion from a high pressure source, can self-ignite. While this is certainly possible when using a compressed gas cylinder as the hydrogen source, the development of a high pressure hydrogen pocket within the instrument is unlikely. The larger threat is the ignition of any accumulated hydrogen from an electrical spark. Carrier gas passes through a number of fittings and the thin-walled and somewhat brittle GC column before passing into the vacuum of the MS. A small leak could develop at any stage resulting in the hydrogen build-up in the GC oven. Accumulation of hydrogen within the MS is predominantly associated with the loss of functionality of the pumping system. While the pumping system is functioning properly all carrier gas that is introduced into the MS will be captured by the turbomolecular pump then vented to the exhaust by the rotary pump. Loss of function, through loss of electrical power or hardware failure, will cause the carrier gas to pool within the vacuum chamber at increasing concentrations resulting in a situation where ignition of hydrogen can occur. For hydrogen use, as with all procedures performed in the laboratory, a thorough review of local Environmental Health and Safety requirements is essential. Successful safety procedures in lab operations begin with a combination of design and planning, which is supported by a company culture that considers it to be mandatory and of the highest priority. The most effective chemical hygiene plans are ones that are built around a company culture of safety where all personnel strictly adhere to the guidelines and regulations. Some practical considerations to lessen the danger of hydrogen accumulation include: Always vent the roughing pump exhaust and injector vent lines to a fume hood. Always leak check the system after changing gas cylinders or installing/repairing the gas lines using a fully functioning hand-held leak sensor suitable for hydrogen applications. If you are using compressed gas cylinders always install the Hydrogen Snubber, as shown in Figure 2, and warning label (not shown) that comes with every Clarus GC. The Hydrogen Snubber attaches directly to the exit of the regulator and will prevent a sudden discharge of hydrogen into the lab should a gas line break. Consider using a hydrogen generator. Hydrogen generators produce high purity hydrogen at low pressure (relative to compressed gas cylinders), which eliminates the risk of high pressure discharge and subsequent selfignition. Consider using a hydrogen sensor in the GC oven or in the general area the instrument is located. It is highly recommended to utilize the optional Nitrogen Purge Vent kit if you are using hydrogen as a carrier gas. The Nitrogen Purge Vent kit will allow nitrogen to be introduced into the mass spectrometer upon venting. This will lower the risk of hydrogen build-up within the vacuum chamber should the carrier gas not be shut off along with the pumping system. Always turn off the hydrogen at the cylinder or hydrogen generator every time you shut down the GC or MS. Never depressurize a gas cylinder, especially a hydrogen gas cylinder, in the lab. The high pressure venting could result in self-ignition and subsequent explosions. 2

After a power failure, the mass spectrometer may have collected hydrogen regardless of whether or not the pumping system was automatically reactivated. It is recommended to utilize the Nitrogen Purge Vent kit function for 3 minutes prior to returning the system to full operation. If the purge kit is not installed the source may be removed to allow the vacuum chamber to vent to atmosphere. Table 1. Required inlet carrier gas pressures, in psig, for 3 cm/s helium at 5 C. Column Length (m) Column i.d. (mm) 1 15 3 6 1.1 31.6 49.1 >1. >1. >1..18 9.1 13.9 29. 61.9 >1..25 5.7 7. 14.4 3.2 52.7.32 <5. <5. 8.6 17.8 3.7.53 <5. <5. <5. 6.3 5. Figure 2. Hydrogen Snubber attached to the exit of the dual-stage stainless steel gas regulator. The installation of a Hydrogen Snubber will prevent the rapid discharge of hydrogen into the laboratory should a gas line break. Table 2. Required inlet carrier gas pressures, in psig, for 4 cm/s hydrogen at 5 C Column Length (m) Column i.d. (mm) 1 15 3 6 1.1 18.5 28.6 6.8 >1. >1..18 5.5 8.3 17. 35.9 62.8.25 <5. <5. 8.6 17.7 3.6.32 <5. <5. <5. 1.6 18..53 <5. <5. <5. <5. 6.3 Experimental Set-up (Configuration, Pressures, and Flows) Column Compatibility Because hydrogen has a different viscosity from helium and the maximum column efficiency occurs at a different gas velocity, the applied carrier gas pressure will be different for the two gases when run under optimum conditions. In some instances it may enable or constrain column types when used with an MS detector. Tables 1 and 2 indicate the carrier gas inlet pressure necessary to run under near optimum conditions on columns connected to an MS with a variety of geometries for the two gases respectively. These tables show that most of the commonly used columns are suitable for use with either helium or hydrogen. At the extremes, helium is better with shorter and/or wider-bore columns whereas hydrogen is better with longer and/or narrower-bore columns. When swapping carrier gases, refer to these tables to ensure that the column being used will still work with practical pressures with the new gas. Setting up the instrumentation to use hydrogen was quite straightforward. The installation included: Ultra high purity hydrogen gas cylinder (99.999%) Hydrogen Snubber on the hydrogen gas cylinder (and label) Nitrogen Purge Vent kit Roughing pump exhaust and injector vent lines vented to the laboratory fume hood No further modifications of the hardware were necessary to perform this work. Setting up the GC to operate using hydrogen as the carrier gas required the selection of hydrogen on the touch screen of the GC. This was achieved by navigating into the Programmed Pneumatic Control (PPC) configuration window on the touch screen and selecting hydrogen as the Gas under the correct Channel tab as illustrated in Figure 3. Selecting a different gas in this window changes the mass-flow calibration parameters used by the firmware. The calibration parameters are traceable to the National Institute of Standards and Technology (NIST ) standards; however, it is prudent to check and recalibrate locally if necessary using a certified flow meter periodically. 3

H2 A. B. C. D. Figure 3. Procedure to change the carrier gas to hydrogen on the GC touchscreen. A) Access the Configuration screen from the Tools menu, B) Select the pneumatics control using the Pneumatics button, C) Choose PPC Configure, D) In the Carrier tab select the proper Channel and select H2 as the Gas and click OK. Carrier Velocity (cm/s) 14 12 1 8 6 4 2 Helium Hydrogen Vacuum System Performance Two pumping systems are available with Clarus mass spectrometers; the standard S model equipped with a 55 L/sec turbomolecular pump and the larger capacity T and C models equipped with a 255 L/sec turbomolecular pump. The compression of the pumped gas, and thus pumping efficiency, varies according to the square of its molecular mass such that heavier compounds are pumped more efficiently. The resulting compression ratios decrease with molecular weight, see Table 3, and it is recommended to only use the T and C models if hydrogen is to be used as the carrier gas. 1 1.5 2 2.5 3 3.5 4 4.5 5 Column Flow (ml/min) Figure 4. Carrier Velocity vs. Column Flow for hydrogen and helium. Using a 3 m x.25 mm x.25 μm column the following carrier velocity versus flow setting was generated (see Figure 4). The velocity numbers were obtained from the GC screen. Hydrogen achieves a much higher carrier velocity and demonstrates more linear behavior over the flow range. This can result in shorter elution times and better chromatography (sharper peaks) however the performance of the mass spectrometer pumping system must also be taken into consideration. All Clarus mass spectrometers use a single wide range gauge (WRG) that operates using a combined Pirani/inverted magnetron ionization sensor. The pressure reading of the WRG is dependent on the type of gas pumped and is factory set to operate with helium. Converting to hydrogen does not require further actions, however, due to the relative molecular mass (RMM) of hydrogen relative to helium the absolute vacuum reading can be up to a factor of 2 off. In Figure 5 we report vacuum readings relative to flow rate using the same 3 m x.25 mm x.25 μm column mentioned above. As expected, as more gas is pushed into the mass spectrometer the pressure increases. This effect is more drastic with hydrogen due to the effects previously described. Even when taking up to a factor of 2 difference in the reading, the level of hydrogen raises rapidly relative to helium. Because of this, higher flow rates combined with larger diameter columns are not recommended for use with hydrogen. 4

Vacuum Reading (*e5) 3 25 2 15 1 5 Helium Hydrogen appropriate for their environment. Hydrogen cylinders offer maintenance free operation but can be expensive and in the event of release, can present a safety concern. Hydrogen generators require regular upkeep but operate at low pressures thus improving safety. Additionally, the operation of the unit can be tied to the operation of the GC/MS system so in the event of a loss of electrical power or hardware failure, the generator can be automatically shut off. The initial cost of a hydrogen generator must be weighed against the both improved safety and long term savings versus compressed gas cylinders. 1 1.5 2 2.5 3 3.5 4 4.5 5 Figure 5. Wide range gauge reading. Column Flow (ml/min) Table 3. Compression ratios of nitrogen, helium, and hydrogen gas for the large capacity T and C model turbomolecular pump. Compound Compression ratio N 2 >1 x 1 11 He 3 x 1 5 H 2 1 x 1 4 Sources of Hydrogen In general, two sources of hydrogen are available high pressure gas cylinders and hydrogen generators. Both sources have advantages and disadvantages and it falls upon the laboratory manager to select the source most Conclusion The use of hydrogen as a carrier gas for GC/MS-based experimentation offers many advantages in both cost and performance yet does not come without risks. The flammable nature of hydrogen presents researchers with specific challenges, which when addressed using clear planning and stringent standard operating procedures, can be mitigated such that the safety of both laboratory personnel and property can be reasonably assured. In all cases a regular review of standard operating procedures and a thorough chemical hygiene plan is required. While the dangers of working with hydrogen can never be fully eliminated, many inherently dangerous processes are already commonly performed in lab operations, whose risks are mitigated with the development and observance of well thought out and implemented SOPs and chemical hygiene plans. Additional information available on the OSHA web site: http://www.osha.gov. Keyword: Hydrogen. PerkinElmer, Inc. 94 Winter Street Waltham, MA 2451 USA P: (8) 762-4 or (+1) 23-925-462 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 212, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 1288_1