Blue Sepharose 6 Fast Flow



Similar documents
RESOURCE Q, 1 ml and 6 ml RESOURCE S, 1 ml and 6 ml

HiTrap Heparin HP, 1 ml and 5 ml

INSTRUCTIONS Edition AC

MabSelect SuRe. GE Healthcare Life Sciences

Process-scale purification of monoclonal antibodies polishing using Capto Q

HiTrap Chelating HP, 1 ml and 5 ml

How To Test For Cleaning Efficiency With A Predictor 96 Well Filter Plate

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration

His GraviTrap. GE Healthcare. Operation

ReadyKart Mobile Processing Station Instructions for Use

UniFlux 10: hollow fiber conversion kit ( ) installation instruction

HisTrap HP, 1 ml and 5 ml

GE Healthcare. UNICORN 5.2 vs Software Change Description

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

Amersham High Molecular Weight Calibration Kit for native electrophoresis

HisTrap HP, 1 ml and 5 ml

GE Healthcare. UNICORN 6.x. SQL Server 2008 configuration for UNICORN installations

How to Download a GE Healthcare License File

Page finder. 1. Legal Handling Safety warnings and precautions Storage Expiry 4

TECHNICAL BULLETIN. HIS-Select Nickel Affinity Gel. Catalog Number P6611 Storage Temperature 2 8 C

Gain efficiency in your process development with ÄKTA avant

CU-950 Installation Guide

soft ware change description

UNICORN control software

Summary of GE Healthcare's Quality Management System (QMS) Covering BioProcess chromatography media, equipment, software, and spare parts

Size Exclusion Chromatography

Instrument services for WAVE Bioreactor systems

His Mag Sepharose excel. Ni Sepharose excel. HisTrap excel. gelifesciences.com

A novel AIEX chromatography medium (resin) to remove IgA and IVIG purification process

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Purification of GST-tagged Proteins

UNICORN 7.0 software. Description. gelifesciences.com

ÄKTA system Training Guide

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

Purification of reaction mixtures using flash chromatography.

Diffusion and Fluid Flow

Affi-Prep Protein A Matrix Instruction Manual

UNICORN. Hardening Guide

Glutathione Resin. User Manual. User Manual. Cat. Nos , , PT (071414)

GAMP5 - a lifecycle management framework for customized bioprocess solutions

For in vitro preparation of human mononuclear cells from peripheral blood, bone marrow, and umbilical cord blood. Not for in vitro diagnostic use.

GE Healthcare Life Sciences. Validation Services. Compliance support through life cycle management

Bio-Gel P Polyacrylamide Gel Instruction Manual

ÄKTA Laboratory-scale Chromatography Systems

SUCRALOSE. White to off-white, practically odourless crystalline powder

Fast Trak Training & Education

The Theory of HPLC. Gradient HPLC

HiPer Ion Exchange Chromatography Teaching Kit

illustra MicroSpin G-25 Columns

Prepacked chromatography columns for ÄKTA systems

High-throughput Process Development with PreDictor Plates

GE Healthcare. Hydrophobic Interaction and Reversed Phase Chromatography. Principles and Methods

Gas Chromatography Liner Selection Guide

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

Affinity Chromatography

Elevated temp up to 150 C. HiT

Single-Use Flow-Through Cell ph FTC-SU-HP8

HiTrap Affinity columns

2D gel electrophoresis

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

[ Care and Use Manual ]

Evaluation of an alkali stable Protein A matrix versus Protein A Sepharose Fast Flow and considerations on process scale-up to 20,000L.

AG 501-X8 and Bio-Rex MSZ 501(D) Mixed Bed Resin Instruction Manual. Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules CA LIT205 Rev B

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

Aurum Ion Exchange Mini Kits and Columns. Instruction Manual

Anion chromatography using on-line recycled eluents

illustra puretaq Ready-To-Go PCR Beads

GE Healthcare s cell culture media, sera and process liquids

AFFINITY CHROMATOGRAPHY

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

DENSITY MEASURING SYSTEMS

Running protein gels and detection of proteins

ProteoMiner Protein Enrichment Kits

105 Adopted:

POROS CaptureSelect affinity columns for rapid, small-scale purification and sample preparation of recombinant proteins

Monitor UPC-900. User Manual

Western Blotting: Mini-gels

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

UltraClean PCR Clean-Up Kit

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Tamsulosin Hydrochloride Capsules

An In-Gel Digestion Protocol

Troubleshooting Guide for DNA Electrophoresis

ÄKTAexplorer, ÄKTApurifier

HiTrap affinity columns

Chromatin Immunoprecipitation (ChIP)

SODIUM CARBOXYMETHYL CELLULOSE

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

INSTRUCTION Probemaker

Protein Purification Handbook

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Dissolved and precipitated oxalate

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

DOWEX Resins as Organic Solvent Desiccants

Gel Filtration Standard

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

Transcription:

GE Healthcare Life Sciences Instructions 71-7055-00 AI Affinity chromatography Blue Sepharose 6 Fast Flow Blue Sepharose 6 Fast Flow is Cibacron Blue 3G covalently attached to the Sepharose 6 Fast Flow matrix by the triazine coupling method. The blue dye binds many proteins, such as albumin, interferon, lipoproteins and blood coagulation factors. It also binds several enzymes including kinases, dehydrogenases, and most enzymes requiring adenyl-containing cofactors e.g., NAD+. The highly cross-linked matrix provides a stable, rigid medium. Blue Sepharose Fast Flow belongs to the BioProcess media family. BioProcess media are separation media developed, made and supported for industrial scale especially the manufacture of healthcare products. With their high physical and chemical stability, very high batch-to-batch reproducibility, and Regulatory Support File back-up, BioProcess media are ideal for all stages of an operation from process development through scale-up and into production. Large quantities can be delivered at short notice.

Table of contents 1 Product description...3 2 Column packing...4 3 Evaluation of packing...6 4 Operation...9 5 Regeneration, Cleaning, Sanitization and Storage... 10 6 Ordering information... 12 7 Further information... 13 2 71-7055-00 AI

1 Product description Table 1. Medium characteristics Total binding capacity Ligand density Average particle size Bead structure Linear flow velocity 1 ph stability 2 long term short term (CIP) Chemical stability >18 mg human serum albumin/ml drained medium approx. 7 μmol Cibacron Blue 3G/ml drained medium 90 μm (45 to 165 μm) 6% highly cross-linked agarose 300 cm/h (25 C, XK 50/30 column, 15 cm bed height) 4 to 12 3 to 13 7 days at 40 C in: 70% ethanol 6 M guanidine hydrochloride 8 M Urea Temperature stability 4 C to 40 C Autoclavable 121 C for 15 min in distilled water Storage 2 C to 8 C in 0.1 M KH 2PO 4, ph 8.0 and 20% ethanol 1 Linear flow rate = volumetric flow rate (cm 3 /h)/column cross-sectional area (cm 2 ) 2 The ranges given are estimates based on our knowledge and experience. Please note the following: ph stability, long term refers to the ph interval where the medium is stable over a long period of time without adverse effects on its subsequent chromatographic performance. ph stability, short term refers to the ph interval for regeneration, Cleaning-In-Place and sanitization procedures. 71-7055-00 AI 3

2 Column packing Blue Sepharose 6 Fast Flow is supplied preswollen in 20% ethanol, 0.1 M KH 2PO 4, ph 8.0. Decant the 20% ethanol solution and replace it with binding buffer. The binding buffer should not contain agents which significantly increase the viscosity, but the column may be equilibrated with viscous buffers at reduced flow rates after packing is completed. Table 2. Recommended lab-scale columns for Blue Sepharose 6 Fast Flow Empty Column 1 Packing flow rate (ml/min) First step Second step Tricorn 10/20 0.9 4.7 2 Tricorn 10/50 0.9 4.7 2 Tricorn 10/100 0.9 4.7 2 XK 16/20 2.5 8.7 5 XK 26/20 6.6 23 13 XK 50/20 24.5 85 49 XK 50/30 24.5 85 49 Max. recommended flow rate for chromatography (ml/min) 1 For inner diameter and maximum bed volumes and bed heights, see Ordering information. Table 3. Recommended process-scale columns for Blue Sepharose 6 Fast Flow Column Inner diam (mm) Bed volume (L) Bed height max (cm) BPG 100/500 100 up to 2.0 L 26 BPG 140/500 140 up to 4.0 L 26 BPG 200/500 200 up to 8.2 L 26 BPG 300/500 300 up to 18.0 L 26 BPG 450/500 450 up to 36.0 L 23 Chromaflow 400/100-300 400 13 to 37 L 30 Chromaflow 600/100-300 600 28 to 85 L 30 4 71-7055-00 AI

Packing lab-scale columns 1 Assemble the column (and packing reservoir if necessary). 2 Remove air from the end-piece and adapter by flushing with water. Make sure no air has been trapped under the column bed support. Close the column outlet leaving the bed support covered with water. 3 Resuspend the medium and pour the slurry into the column in a single continuous motion. Pouring the slurry down a glass rod held against the column wall will minimize the introduction of air bubbles. 4 If using a packing reservoir, immediately fill the remainder of the column and reservoir with water. Mount the adapter or lid of the packing reservoir and connect the column to a pump. Avoid trapping air bubbles under the adapter or in the inlet tubing. 5 Open the bottom outlet of the column and set the pump to run at the desired flow rate. Ideally, Sepharose 6 Fast Flow media are packed in XK or Tricorn columns in a two-step procedure: Do not exceed 0.5 bar (0.05 MPa) in the first step and 1.5 bar (0.15 MPa) in the second step. If the packing equipment does not include a pressure gauge, use a packing flow rate of 2.5 ml/min (XK 16/20 column) or 0.9 ml/min (Tricorn 10/100 column) in the first step, and 8.7 ml/min (XK 16/20 column) or 4.7 ml/min (Tricorn 10/100 column) in the second step. See Table 2 for packing flow rates for other columns. If the recommended pressure or flow rate cannot be obtained, use the maximum flow rate your pump can deliver. This should also give a well packed bed. Note: For subsequent chromatography procedures, do not exceed 75% of the packing flow rate. See Table 2 for flow rates for chromatography. 6 Maintain packing flow rate for at least 3 bed volumes after a constant bed height is reached. Mark the bed height on the column. 7 Stop the pump and close the column outlet. 8 If using a packing reservoir, disconnect the reservoir and fit the adapter to the column. 71-7055-00 AI 5

9 With the adapter inlet disconnected, push the adapter down into the column until it reaches the mark. Allow the packing solution to flush the adapter inlet. Lock the adapter in position. 10 Connect the column to a pump or a chromatography system and start equilibration. Re-adjust the adapter if necessary. Packing large scale columns For general process-scale column packing instructions, please visit support section at www.gelifesciences.com/protein-purification. 3 Evaluation of packing To check the quality of the packing and to monitor this during the working life of the column, column efficiency should be tested directly after packing, prior to re-use, and when separation performance is seen to deteriorate. The best method of expressing the efficiency of a packed column is in terms of the height equivalent to a theoretical plate, HETP, and the asymmetry factor, A s. These values are easily determined by applying a sample such as 1% (v/v) acetone solution to the column. Sodium chloride can also be used as a test substance. Use a concentration of 2.0 M NaCl in water with 0.5 M NaCl in water as eluent. The calculated plate number will vary depending on the test conditions and it should therefore be used as a reference value only. It is also important that conditions and equipment are kept constant so that results are comparable. Changes in solute, solvent, eluent, sample volume, flow rate, liquid pathway, temperature, etc., will influence the results. For optimal results, the sample volume should be at max. 2.5% of the column volume and the flow velocity between 15 and 30 cm/h. If an acceptance limit is defined in relation to column performance, the column plate number can be used as part of the acceptance criteria for column use. 6 71-7055-00 AI

Method for measuring HETP and A s To avoid dilution of the sample, apply it as close to the column inlet as possible. Conditions Sample volume: 2.5% of the bed volume Sample conc.: 1.0% (v/v) acetone Flow velocity: 15 cm/h UV: 280 nm, 1 cm, 0.1 AU Calculate HETP and A s from the UV curve (or conductivity curve if NaCl is used as sample) as follows: HETP = L/N where N = 5.54(V e/w h) 2 L = Bed height (cm) N = number of theoretical plates V e = Peak elution distance W h = Peak width at half peak height V e and W h are in the same units. To facilitate comparison of column performance, the concept of reduced plate height is often used. The reduced plate height is calculated: where HETP/d d = diameter of the bead As a guideline, a value of <3 is normally acceptable. The peak should be symmetrical, and the asymmetry factor as close as possible to 1 (values between 0.8 and 1.5 are usually acceptable). A change in the shape of the peak is usually the first indication of bed deterioration due to use. Peak asymmetry factor calculation: where A s= b/a a = 1st half peak width at 10% of peak height b = 2nd half peak width at 10% of peak height 71-7055-00 AI 7

Figure 1 shows a UV trace for acetone in a typical test chromatogram in which the HETP and A s values are Column: BPG 300 Media: Sepharose 6 Fast Flow Bed height: 57.5 cm Bed volume: 40.6 litres Eluent: Distilled water Sample: 1.05 litres (1% acetone) Flow velocity: 19 cm/h V e =18.7 W h =0.9 HETP =0.024 cm a: 0.90 b: 0.85 A s : 0.94 UV absorption Ve W h a b Volume Fig 1. UV trace for acetone in a typical test chromatogram showing the HETP and A s value calculations. 8 71-7055-00 AI

4 Operation Binding Different substances differ in their affinity for Blue Sepharose 6 Fast Flow. The available capacity will depend upon parameters such as flow rate, ph, buffer composition and temperature. 1 Sample ph should be the same as that of the binding buffer. Filter the sample through a 0.22?m or 0.45?m filter to prolong the working life of the medium. 2 After the sample has been loaded, wash the medium with binding buffer until the base line is stable. Elution 1 Elution conditions vary with the sample. Elution may be accomplished by a change in ph, polarity (e.g., ethylene glycol) or ionic strength of the buffer. Enzymes can often be eluted at less than 1 M NaCl. 2 Competitive elution with low concentrations of the cofactor is required for very specifically bound proteins. Either step or continuous gradients may be used. 71-7055-00 AI 9

5 Regeneration, Cleaning, Sanitization and Storage Regeneration Depending on the nature of the sample, reversibly bound material can be eluted with 4 to 5 washing cycles of alternate high ph (0.1 M Tris-HCl, 0.5 M NaCl, ph 8.5) and low ph (0.1 M sodium acetate, 0.5 M NaCl, ph 4.5) buffers, followed by re-equilibration in binding buffer. In some applications, substances like denaturated proteins or lipids do not elute in the regeneration procedure. These can be removed by Cleaning-In-Place. Cleaning-In-Place (CIP) Remove precipitated proteins by washing the column with 4 bed volumes of 0.1 M NaOH solution at a low linear flow rate (40 cm/h), followed by washing the column with 3 to 4 bed volumes of 70% ethanol or 2 M potassium thiocyanate. Alternatively, wash the column with 2 bed volumes of 6 M guanidine hydrochloride. In both cases, wash immediately with at least 5 bed volumes of sterile filtered binding buffer at ph 8.0. Remove strongly bound hydrophobic proteins, lipoproteins and lipids by washing the column with 3 to 4 bed volumes of up to 70% ethanol or 30% isopropanol. (Apply increasing gradients to avoid air bubble formation when using high concentrations of organic solvents.) Alternatively, wash the column with 2 bed volumes of detergent in a basic or acidic solution, e.g., 0.1% non-ionic detergent in 1 M acetic acid. Wash at a flow rate of 40 cm/h. Remove residual detergent by washing with 5 bed volumes of 70% ethanol. In both cases, wash immediately with at least 5 bed volumes of sterile filtered binding buffer at ph 8.0. 10 71-7055-00 AI

Sanitization Sanitize the column by equilibrating it with a buffer consisting of 2% hibitane digluconate and 20% ethanol and allow to stand for 6 hours. Alternatively, equilibrate the column with 70% ethanol and allow to stand for 12 hours. In both cases wash with at least 5 bed volumes of sterile filtered binding buffer at ph 8.0. Column performance is not significantly changed by the Cleaning-In-Place and sanitization procedures described above. All of the above recommended procedures can be performed directly on the packed column. Storage For longer periods of storage, for example weeks, we recommend that the medium be stored at 2 C to 8 C in 20% ethanol, 0.1 M KH 2PO 4, ph 8.0. The medium must not be frozen. 71-7055-00 AI 11

6 Ordering information Product Pack size Code No. Blue Sepharose 6 Fast Flow 50 ml 17-0948-01 Blue Sepharose 6 Fast Flow 500 ml 17-0948-02 Blue Sepharose 6 Fast Flow 1 L 17-0948-03 Blue Sepharose 6 Fast Flow 5 L 17-0948-04 Related product Pack size Code No. Tricorn 10/20 column, 10 mm i.d., max 2.2 ml bed volume or 2.8 cm bed height 1 28-4064-13 Tricorn 10/50 column, 10 mm i.d., max 4.5 ml bed volume or 5.8 cm bed height Tricorn 10/100 column, 10 mm i.d., max 8.5 ml bed volume or 10.8 cm bed height XK 16/20 column, 16 mm i.d., max 30 ml bed volume or 15 cm bed height XK 26/20 column, 26 mm i.d., max 65 ml bed volume or 12.5 cm bed height XK 50/20 column, 50 mm i.d., max 270 ml bed volume or 14 cm bed height XK 50/30 column, 50 mm i.d., max 550 ml bed volume or 28.5 cm bed height 1 28-4064-14 1 28-4064-15 1 18-8773-01 1 18-1000-72 1 18-1000-71 1 18-8751-01 1 Process-scale quantities are available. Please contact your local representative. 12 71-7055-00 AI

Accessories Pack size Code No. HiTrap /HiPrep, 1/16" male connector for ÄKTA design For connection of columns with 1/16" fittings to ÄKTA design 8 28-4010-81 Union 1/16" male/1/16" male with 0.5 mm i.d. For connecting two columns with 1/16" fittings in series Fingertight stop plug, 1/16" 1 For sealing a HiScreen column 2 18-1120-93 5 11-0003-55 1 One fingertight stop plug is connected to the inlet and the outlet of each HiScreen column at delivery. Data files Code No. BPG columns 18-1115-23 BPG 450 column 18-1160-59 CHROMAFLOW columns 18-1138-92 Blue Sepharose 6 Fast Flow 18-1060-75 Literature Code No. Affinity Chromatography Handbook 18-1022-29 Handbook of Process Chromatography: 18-1121-56 A Guide to Optimization, Scale-Up, and Validation (Academic Press, 1997, Sofer, G. and Hagel, L.) 7 Further information Please read these instructions carefully before using Blue Sepharose 6 Fast Flow medium. For further information visit www.gelifesciences.com/protein-purification or contact your local GE Healthcare representative. 71-7055-00 AI 13

14 71-7055-00 AI

71-7055-00 AI 15

For local office contact information, visit www.gelifesciences.com/contact GE Healthcare Bio-Sciences AB Björkgatan 30 751 84 Uppsala Sweden www.gelifesciences.com/protein-purfication GE Healthcare Europe GmbH Munzinger Strasse 5, D-79111 Freiburg, Germany GE Healthcare UK Ltd Amersham Place Little Chalfont Buckinghamshire, HP7 9NA UK GE Healthcare Bio-Sciences Corp 800 Centennial Avenue P.O. Box 1327 Piscataway, NJ 08855-1327 USA GE Healthcare Japan Corporation Sanken Bldg. 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073 Japan GE, imagination at work and GE monogram are trademarks of General Electric Company. ÄKTA, BioProcess, BPG, Chromaflow, HiTrap, HiPrep, HiScreen, Sepharose and Tricorn are trademarks of GE Healthcare companies. All third party trademarks are the property of their respective owners. 2005-2011 General Electric Company All rights reserved. Previously published Mar. 2005. All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information. imagination at work 71-7055-00 AI 02/2011