IC-EMC v2 Application Note. A model of the Bulk Current Injection Probe



Similar documents
A NEAR FIELD INJECTION MODEL FOR SUSCEPTIBILITY PREDICTION IN INTEGRATED CIRCUITS

IC-EMC Simulation of Electromagnetic Compatibility of Integrated Circuits

MEASUREMENT SET-UP FOR TRAPS

Current Probes. User Manual

S-PARAMETER MEASUREMENTS OF MEMS SWITCHES

Cable Analysis and Fault Detection using the Bode 100

LS RS. Figure 1: Assumed inductor model

Measurement of Inductor Q with the MSA Sam Wetterlin 3/31/11. Equation 1 Determining Resonant Q from Inductor Q and Capacitor Q

Capacitor Self-Resonance

Standex-Meder Electronics. Custom Engineered Solutions for Tomorrow

Impedance Matching and Matching Networks. Valentin Todorow, December, 2009

Rated Power(W) 8W 2. EG-LED W 3. EG-LED W

UNDERSTANDING AND CONTROLLING COMMON-MODE EMISSIONS IN HIGH-POWER ELECTRONICS

RFID Receiver Antenna Project for Mhz Band

Crosstalk effects of shielded twisted pairs

VJ 6040 Mobile Digital TV UHF Antenna Evaluation Board

When designing. Inductors at UHF: EM Simulation Guides Vector Network Analyzer. measurement. EM SIMULATION. There are times when it is

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD OF INSERTION LOSS MEASUREMENT

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term

Extending Rigid-Flex Printed Circuits to RF Frequencies

EMC / EMI issues for DSM: new challenges

AN2866 Application note

DRAFT. University of Pennsylvania Moore School of Electrical Engineering ESE319 Electronic Circuits - Modeling and Measurement Techniques

Impedance Analysis of Automotive High Voltage Networks for EMC Measurements

Predicting radiated emissions from cables in the RE02/RE102/DO- 160/SAE J test set up, using measured current in NEC and simple TX equations.

PCB Radiation Mechanisms: Using Component-Level Measurements to

EMC Supplemental Information and Alternative Component Requirements

Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer. Application Note

Applications in EMC testing. Outline. Antennas for EMC Testing. Terminology

Why test results can t compare

Amplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B)

EM Noise Mitigation in Circuit Boards and Cavities

Radiated Emission and Susceptibility

Software for Design NMR Probes Using the Shielded Split Ring and the Shielded Symmetrical Band Resonators

Experiment 7: Familiarization with the Network Analyzer

RigExpert AA-30 Antenna Analyzer (0.1 to 30 MHz) AA-54 Antenna Analyzer (0.1 to 54 MHz) User s manual

5. MINIMIZATION OF CONDUCTED EMI. Chapter Five. switch-mode power supply decrease approximately linearly with the increase of the switching

Yaesu FT MHz PA output and filter simulations Marc Vlemmings, PA1O - Eindhoven, The Netherlands

Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters

APN1001: Circuit Models for Plastic Packaged Microwave Diodes

Changes PN532_Breakout board

Tire pressure monitoring

Reproducing system-level bulk current injection test in direct power injection setup for multiple-port DUTs

Connectivity in a Wireless World. Cables Connectors A Special Supplement to

An equivalent circuit of a loop antenna.

Agilent Advanced impedance measurement capability of the RF I-V method compared to the network analysis method. Application Note

Minimizing crosstalk in a high-speed cable-connector assembly.

Cable Impedance and Structural Return Loss Measurement Methodologies

The Critical Length of a Transmission Line

100 ADS Design Examples A Design Approach Using (ADS)

EMC STANDARDS STANDARDS AND STANDARD MAKING BODIES. International. International Electrotechnical Commission (IEC)

Agilent Ultra-Low Impedance Measurements Using 2-Port Measurements. Application Note

Application Note AN:005. FPA Printed Circuit Board Layout Guidelines. Introduction Contents. The Importance of Board Layout

Digital Systems Ribbon Cables I CMPE 650. Ribbon Cables A ribbon cable is any cable having multiple conductors bound together in a flat, wide strip.

Electronic filters design tutorial -2

Embedded FM/TV Antenna System

Nano Stepping Notch Filter Jim Hagerman

Time and Frequency Domain Analysis for Right Angle Corners on Printed Circuit Board Traces

Impedance Matching of Filters with the MSA Sam Wetterlin 2/11/11

RLC Series Resonance

Pre-Compliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions

ELECTRICAL CHARACTERISATION OF SEMI-RIGID COAXIAL CABLES WITH SMA AND K CONNECTORS. Carmen Diez Rafael García Juan Daniel Gallego.

Consideration of a high-capacity foil cable:

CONDUCTED EMISSION MEASUREMENT OF A CELL PHONE PROCESSOR MODULE

CONCEPT-II. Overview of demo examples

Model BiConiLog Antenna MANUAL EMC TEST SYSTEMS, L.P. MARCH 2002 REV C PN

Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements

Tuning a Monopole Antenna Using a Network Analyzer

75 Ω Transmission System

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER

Circuits with inductors and alternating currents. Chapter 20 #45, 46, 47, 49

EMC Basics. Speaker : Alain Lafuente. Alain.lafuente@we-online.com

Probes and Setup for Measuring Power-Plane Impedances with Vector Network Analyzer

Techniques for precise cable and antenna measurements in the field

ATE-A1 Testing Without Relays - Using Inductors to Compensate for Parasitic Capacitance

The N2PK Vector Network Analyzer (N2PK VNA) Original by Paul Kiciak, N2PK, 2007 * (*Pages edited or added by VE7WRS, 2009)

3.3 Calibration standards

A BROADBAND COAXIAL LINE FOR ELECTRICAL CHARACTERIZATIONS OF DIELECTRICS JAFER OMER SULIMAN AZZABI

155 Mb/s Fiber Optic Light to Logic Receivers for OC3/STM1

Utilizing Time Domain (TDR) Test Methods For Maximizing Microwave Board Performance

Planar Inter Digital Capacitors on Printed Circuit Board

Coupling/Decoupling Networks (CDN)

AN3353 Application note

Issues and Solutions for Dealing With a Highly Capacitive Transmission Cable

Consulting. IEEE Joint Meeting Rockford, March 28, ROY LEVENTHAL

AN003. Evaluation of Anechoic Chambers For EMC Measurements Using A CNE. Overview. CNE III 9kHz to 2GHz Conducted and radiated output

How to design SMPS to Pass Common Mode Lightning Surge Test

RF Design Guidelines: PCB Layout and Circuit Optimization

2. The Vector Network Analyzer

ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER

AVR2006: Design and characterization of the Radio Controller Board's 2.4GHz PCB Antenna. Application Note. Features.

MABINDRA RAI THE INVESTIGATION OF MINIVNAPRO VECTOR NETWORK ANALYZER

Modeling Physical Interconnects (Part 3)

ESD Line Ultra-Large Bandwidth ESD Protection

Pulsed Power Engineering Diagnostics

IFI5481: RF Circuits, Theory and Design

An Ethernet Cable Discharge Event (CDE) Test and Measurement System

Coupling- / Decoupling Network. 150 khz 300 MHz. 150 khz 230 MHz. 10 khz 230 MHz. IEC and CISPR 15 / CISPR 22 IEC

APN1009: A Varactor Controlled Phase Shifter for PCS Base Station Applications

Transcription:

IC-EMC v2 Application Note A model of the Bulk Current Injection Probe This work has been conducted by S. Akue Boulingui and A. Cisse Ndoye within the French project EPEA-Aerospace Valley funded by the French Ministry of Industry. This section illustrates how to build an electrical model of the Bulk Current Injection probe (BCI), which is part of the international standard IEC 62132 [Ref 62132]. We present the BCI probe under test, make a description of the impedance measurement set-up and describe the method for BCI probe modelling. I. BCI probe under test The BCI test method is part of the IEC standard 62132, under reference 62132-3 Bulk Current Injection (BCI) [1] (Table 1). Standard Description Stage in 2010 IEC 62132-1 General Conditions and Definitions International standard IEC 62132-2 Radiated Immunity - TEM Cell and Wideband TEM Cell Method Committee draft for voting IEC 62132-3 Bulk Current Injection (BCI) International standard IEC 62132-4 Direct RF Power Injection (DPI) International standard IEC 62132-5 Work Bench Faraday Cage (WBFC) International standard IEC 62132-6 Local Horn Injection Antenna (LIHA) New proposal IEC 62132-7 Mode stirred chamber New proposal IEC 62132-8 Micro strip-line New Proposal IEC 62132-9 Near field scan immunity (NFSI) New proposal TABLE I: Status of IEC 62132 standards for measurement of integrated circuit immunity The Bulk Current Injection Probe we used is provided by ETS-Lindgren, model number 95242-1 [2]. It has only one N port connexion. The probe is used to inject a radio-frequency parasitic current into conductors and cables which are connected to electronic equipments. The probe is presented in 0. The central hole allows inserting cables. 17/05/2010 etienne.sicard@insa-toulouse.fr 1

Opening sytem for cabes introduction RF injection signal (N-connector) Tests cables Figure 1: BCI injection probe description Probe reference The BCI Probe provides a mean to inject a controlled RF stress to a device under test (DUT) through interconnecting or power cables without requiring a direct connection to that cables. RF disturbance can be injected through a single or multiple cables, grounding and bonding straps, outer conductors of shielding conduits and coaxial cables, etc. The bulk current injection probe series is especially designed to provide minimum insertion loss between 2 MHz and 400 MHz (Fig. 1). II. BCI Test Bench The BCI test bench requires at least one BCI probe, the equipment under test and the line for power injection as shown in 0. RF source Amplifier RF receiver Load(s) or other device Injection probe Line under test Device under test RF disturbance source Criterion analysis Loads, electronic system, Device under test Figure 2: Simple BCI test bench : principles (top) and test example (bottom). A RF source is used to generate the disturbance signal. A power amplifier increases the level of the signal at the desired value. The amplified signal is injected in the BCI probe. A coupler and a RF receiver (such as a powermeter) help controlling the level of forward power. The part of the forward power which is transferred to the device under test remains unknown. To monitor the power at the far end of the cable, near the device under test, another BCI probe may be used as monitoring current probe (0). 17/05/2010 etienne.sicard@insa-toulouse.fr 2

RF source Amplifier RF receiver Load(s) or other device Injection probe Line under test Monitoring current probe Device under test Figure 3: BCI test bench with monitoring current probe. III. Impedance measurement setup and results BCI probe impedance over frequency is characterized through S parameter measurements. From that measurement, we can derive an electrical model which fits the probe behaviour. We used a Vector Network Analyzer (VNA) and the port 1 for the measurement (Fig. 4). The measurement system is calibrated using appropriate short-circuit, open-circuit and 50-Ohm loads in N-connector format, so that the reference plane is at the input of the probe connector. The calibration compensates the cable effects, allowing the characterization of the probe impedance only. Measurement reference plane BCI probe Vector Nectwork Analyzer Port 1 Figure 4: S parameters measurement set-up To download the measurements, proceed as follows: 1. Click the icon Impedance vs. frequency 2. Click Add measurement and select *.s50 format 3. In EMC_lib, select the file Meas_Sparam_BCI-probe_INSA.s50 17/05/2010 etienne.sicard@insa-toulouse.fr 3

4. The following screen appears. It can be seem that the coil features a very low impedance at low frequencies (10 Ohm around 1 MHz) and ranges between 10 and 500 Ohm in the whole frequency bandwidth. The impedance profile features an inductance effect (+ 20 db/decade) at low frequencies. Figure 5 : Measurement of the BCI impedance from 1kHz to 6 GHz (Meas_Sparam_BCIprobe_INSA.s50) 17/05/2010 etienne.sicard@insa-toulouse.fr 4

IV. First-order model of the BCI probe We have no precise information about the internal structure of the BCI, such as the conductor width and length, number of loops, material permittivity, transformer characteristics, etc.. This does not allow us to use physical formulations in order to calculate the corresponding impedance. We consider here the probe mainly as an inductor (0) with parasitic capacitive and resistive effects.. input Figure 6 : BCI probe internal struture 1. Click «Add» 2. Select «L» 3. Click around here 4... or enter the value 5. Insert the corresponding symbol Figure 7 : Extraction of BCI R,L,C elements from impedance measurement To identify the electrical equivalent elements, proceed as follow: 1. In the Z(f) sub-menu, Select Add 2. Select L (inductance) 3. Click along the measurement curve in the +20 db slope (1-30 MHz) 4. The tool computes the equivalent L value (here: 700 nh). Alternatively, you may enter the value and observe the matching. 17/05/2010 etienne.sicard@insa-toulouse.fr 5

5. Click Insert symbol. A 700 nh inductance is added to the schematic diagram for further use in the BCI model The resonant behaviour around 50 MHz is modelled by a RLC equivalent circuit (Erreur! Source du renvoi introuvable.). We use a combination of L_BCI=700 nh, R_P=200 Ω and C_P=15 pf capacitor. The serial resistance R_BCI=1 Ohm is mandatory to avoid WinSpice simulation problems. As shown in Fig. 8, the schematic diagram should also include: Connections to ground A probe symbol Z(f) Convert the schematic diagram into a SPICE netlist (Command and icon Generate Spice file ), execute the simulation using WinSpice and click the icon Impedance Window to compare the simulated Z(f) with measured Z(f) as shown in Fig. 8 bottom. The fit is quite satisfactory upto 400 MHz. Figure 8: Schematic diagram (top) and Spice conversion (bottom) for BCI impedance simulation 17/05/2010 etienne.sicard@insa-toulouse.fr 6

Figure 9: A simple BCI probe model valid upto 400 MHz (BCI_Model_HF.sch) V. Accurate model of the BCI probe up to 6 GHz In Figure 10 a more accurate model of the BCI probe is presented. Apart from the HF model described previously (L_BCI, C_P1, R_P1), a set of R,L,C elements are added to account for s around 200, 400, 900 and 4000 MHz. We propose to include 3 RLC cells. HF model Main inductance of the BCI probe Figure 10 : BCI probe model 17/05/2010 etienne.sicard@insa-toulouse.fr 7

Table II describes the BCI probe electrical elements. It shows extracted and optimized values to improve the matching with measurements. The major modification concerns L_P1 and C_P2. Element Extracted values Optimized values Description L_BCI 0.7 uh 0.7 uh Probe inductance value C_P1 15 pf 15 pf Parasitic capacitance for 50 MHz L_P2 19 nh 6 nh Parasitic inductance for 200 MHz C_P2 11 pf 35 pf Parasitic capacitance 200 MHz L_P3 18 nh 18 nh Parasitic inductance for 1 GHz C_P3 1.5 pf 2 pf Parasitic capacitance for 1 GHz TABLE II: Description of the model elements Figure 11 shows the comparison between the measured and simulated impedance of the BCI probe. We can notice a very good correlation between measurement and simulation from 1 MHz to 6 GHz. Figure 11: Comparison between measurement and simulations for the accurate model (BCI_Model_xHF.sch) 17/05/2010 etienne.sicard@insa-toulouse.fr 8

VI. References [1] IEC62132-, Integrated Circuits, Measurement of Electromagnetic Immunity, 150 KHz 1 GHz, International Electrotechnical Commission, Geneva, Switzerland, 2007 www.iec.ch [2] Datasheet: ETS Lindgreen, "Bulk Current Injection Probes manual", February 2005, online: http://www.ets-lindgren.com. [3] F. Lafon, Y. Benlakhouy, F. De Daran, "Modélisation d une pince d injection pour le test BCI sur ligne de transmission multi conducteur", CEM 2008, Paris. 17/05/2010 etienne.sicard@insa-toulouse.fr 9