Druck DPI 142/150. GE Infrastructure Sensing. Pressure Indicator Calibration manual - K382



Similar documents
RPS/DPS High Accuracy Resonant Pressure Sensor. GE Measurement & Control. Features:

RPS High Accuracy Resonant Pressure Sensor. GE Measurement & Control Solutions. Features:

The PACE5000 Modular Pressure Controller/Indicator

Druck DPI 104. GE Industrial Sensing. Digital pressure indicator User manual - K394

PACE1000 Precision Pressure Indicator

PACE Modular Pressure Controller

PACE1001 Barometer Precision Barometric Indicator and Recorder

UNIK 5800/5900. Flameproof/Explosion-Proof Pressure Sensing Platform. GE Measurement & Control. High Quality. Features. Bespoke as Standard.

Precision digital pressure gauge Model CPG1500

DPI 260 SERIES: Digital Pressure Indicators

Specifications - Installation and Operating Instructions Ø5 [127.00] 17/64 [6.91] 1-59/64 [48.89]

Druck ADTS 505 Air Data Test Set User Manual K0260

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

Druck DPI 104-IS Intrinsically safe digital pressure indicator

Precision Pressure Indicator Model CPG2500

T0118 T2118 T3118. Instruction Manual

GE Sensing. SiCalPro. Monitor and control software. User manual - K0452

PIEZOMETER AIRFLOW MEASURING RING

UNIK Pressure Sensing Platform. GE Measurement & Control. High Quality. Features. Bespoke as Standard. Expertise

CALIBRATION PRINCIPLES

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002

= 800 kg/m 3 (note that old units cancel out) J 1000 g = 4184 J/kg o C

Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version

JOFRACAL Temperature and Pressure Calibration Software

PMP 4000 Series. Druck Amplified Output Pressure Transducers. GE Sensing. Applications. Features

RUSKA PRODUCTS - PRICE LIST

PPC4E Pressure Controller/Calibrator

GE Sensing. Druck PC6-IDOS. Pressure calibrator/indicator User manual - K0355

Digital Pressure Measuring Instrument MDR480

Calibrating Differential Pressure Sensors

Druck ADTS 405. Air Data Test Systems. User Manual K114. GE Sensing. Ps mbar Aim < Pt mbar Aim 3200.

Vaisala Weather Transmitter WXT510 The Most Essential of Weather

Rackmount Intelligent Pressure Scanners

Druck PV 62x pneumatic/hydraulic pressure stations

«The compact and mobile testing device for ventilators.»

Hand-held thermometer Model CTH7000

EXPERIMENT 15: Ideal Gas Law: Molecular Weight of a Vapor

Gas Custody Transfer Calibration

RTX 1000 Series. Versatile Transmitters for a World of Pressure. GE Measurement & Control Solutions. Features

! Ranges 0 to 2.5 mbar through to 700 bar. ! Accuracy 0.025% FS all ranges. ! Integral combined pressure/vacuum pump. !

KNC Model 3666 Automatic Pressure Calibration System

3-step setting. 2-Color Display High-Precision Digital Pressure Switch. Series ZSE30A(F)/ISE30A. Added vacuum range.

BW-500 BED WEIGHER USER MANUAL

GE Druck. Hand Pumps Pneumatic/Hydraulic hand pump Product Guide. Pneumatic hand pumps from -1 to 60 bar. Hydraulic hand pumps from 0 to 1000 bar

USB-Based Temperature & Humidity Sensor

Laminar Flow Elements

G-100/200 Operation & Installation

TELEDYNE HASTINGS INSTRUMENTS

Side-Trak 830/840. Process Gas Mass Flow Meters and Controllers.

High-End Pressure Controller Model CPC8000

Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate

Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual

Measurement Solutions for Refrigeration Technology.

Schedule of Accreditation issued by United Kingdom Accreditation Service High Street, Feltham, Middlesex, TW13 4UN, UK

BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER. Installation, Operation, and Maintenance Instructions. Manual No.

Instruction Manual. T3111 T3111L T3111P T3111Ex

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

PPM Users Manual Signature Software

Model 43AP Pneumatic Controller, Style B

WIKA INSTRUMENT CORPORATION

2/ These vacuum switches are used in measuring ranges between -1 and 0 bar.

Mass Flow Meters & Controllers

Quick Reference Guide VT55 & OBDII. Customer Service TPMS Version: QSG /12

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

CONTENTS USING THE LIGHTS KEY...31 NETWORK ALARMS...32 FAULT AND ERROR MESSAGES...33

REGTRONIC OPERATING INSTRUCTIONS

Models SFP33 & SFP32 Pressure Transmitters

User's Guide. Integrating Sound Level Datalogger. Model Introduction

DM Series CONTENTS. 1. Features Applications General Calibration Dimensions Specifications Calibration procedure 7

Multi-Range Programmable DC Power Supplies 9115 Series

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

Rosemount 333 HART Tri-Loop

Pressure transmitter For general industrial applications Model A-10

4Sight Calibration Management Software

Pressure calibration equipment

Vacuum drying oven for non-flammable solvents

FiberOptix TM. AP Sensor Zeroing and Calibration Guide INTRA-AORTIC BALLOON PRODUCT SUPPORT HOTLINE. U.S. and Canada: IABP (4227)

XP Series Balances Operation Manual

Nokeval. User's Manual Models 301 / 302 / 305

PN-L702B LCD MONITOR TOUCH PANEL DRIVER OPERATION MANUAL. Version 2.1

DPI 610/615 Series. Druck Portable. Pressure Calibrators. GE Sensing. Features. Setting the Standard for Portable. Pressure Calibrators

Online Infrared Flue Gas Analyzer Gasboard 3000

PRODUCT PROFILE FEATURES

Transmitter Interface Program

Instruction Manual Dräger MSI ALV

RS Stock No Instruction Manual RS-1340 Hot Wire Anemometer

Thermistor Thermometer MODEL NO

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

FLOWMETER INSTALLATION INSTRUCTIONS

WARNING CANADIAN D.O.C. WARNING

M 140i. Multifunction Calibrator. Operation Manual

Mitos P-Pump. product datasheet

RAINCONTROL LCD Special

EURAMET 1197 SUPPLEMENTARY BILATERAL COMPARISON OF HYDRAULIC GAUGE PRESSURE STANDARDS UP TO 50 MPa

BlueSolar Pro Remote Panel For BlueSolar PWM-Pro charge controllers 12/24V 5, 10, 20, 30A Article number SCC

Diagnosing Hydraulic Problems

Operating instructions Diffuse reflection sensor with background suppression O1D101 O1D / / 2012

Transcription:

GE Infrastructure Sensing 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Druck DPI 142/150 Pressure Indicator Calibration manual - K382

Introduction This technical manual provides calibration instructions for the Druck DPI 142 and DPI 150 Pressure Indicators. Scope This technical manual provides calibration instructions for the calibration technician of this equipment series. Safety The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. The user must not use this equipment for any other purpose than that stated. This manual contains safety and operating instructions that must be followed to make sure of safe operation and to keep the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use suitably qualified* calibration technicians and good engineering practice for all procedures in this manual. Pressure Do not apply pressure greater than the maximum safe working pressure to the equipment. Technical advice For technical advice contact the manufacturer or subsidiary. *A qualified calibration technician must have the necessary technical knowledge, documentation, special test equipment and tools to carry out the required calibration on this equipment. Associated Publications K343... DPI 142 User Manual K344... DPI 150 User Manual K381...DPI 142/150 SCPI User Manual i K382 Issue No. 1

ABBREVIATIONS The following abbreviations are used in this publication. Note: Abbreviations are the same in the singular and plural. abs absolute m metre C degrees Celsius mbar millibar const constant mm millimetre DPI digital pressure indicator (GE Druck mmh 2 O millimetres of water product) ape NPT National Pipe Thread F degrees Fahrenheit PIN personal identification number g gauge psi pounds per square inch (h) hour Ref. reference IEEE 488 institute of electrical and electronic engineers standard 488 data RS232 serial interface communication standard inhg inches of mercury SCPI standard commands for programmable instruments kg kilogram (s) seconds kts knots TBA to be advised lbs pounds CONTENTS Title page Introduction...1 Preliminary Operations...3 Calibration Check DPI 142 and DPI 150 Option A...4 Calibration Adjustment...5 Calibration Check DPI 150...6 Calibration Adjustment...7 K382 Issue No. 1 ii

Glossary Terminology The terminology used in this manual is specific and individual interpretation must not be introduced. The terms are defined as follows: Adjust Align Calibrate: Check: Disconnect: Examine: Inspect: Maintain: Operate: Readjust: Reconnect: Reset: To bring to a more satisfactory state; to manipulate controls, levers, linkages, etc. to return equipment from an out-of-tolerance condition to an in-tolerance condition. To bring into line; to line up; to bring into precise adjustment, correct relative position or coincidence. To determine accuracy, deviation or variation by special measurement or by comparison with a standard. Make a comparison of a measure of time, pressure, temperature, resistance, dimension or other quality with a known figure for that measurement. To detach the connection between; to separate keyed or matched equipment parts. To perform a critical visual observation or check for specific conditions; to test the condition of. Review the work carried out by Specialists to ensure it has been performed satisfactorily. To hold or keep in any particular state or condition especially in a state of efficiency or validity. Make sure that an item or system functions correctly as far as possible without the use of test equipment or reference to measurement. To adjust again; to move back to a specified condition; to bring back to an in-tolerance condition. To rejoin or refasten that which has been separated. To put back into a desired position, adjustment or condition. iii K382 Issue No. 1

Test: Ascertain by using the appropriate test equipment that a component or system functions correctly. Uncertainty: The uncertainty of a measurement is defined as a parameter, associated with the result of a measurement, that characterises the dispersion of values that could be reasonably be attributed to the measurand. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com

Pressure units and conversion factors Druck DPI 142/150 Calibration Manual Pressure unit Factor (Pascals) Pressure unit Factor (Pascals) bar 100000 lbf/ft 2 47.8803 lbf/in 2 (psi) 6894.76 inhg 3386.39 mh 2 O 9806.65 inh 2 O [1] 249.089 mbar 100 fth 2 O [1] 2989.07 kgf/cm 2 98066.5 atm 101325.0 kgf/m 2 9.80665 pdl/ft 2 1.48816 mmhg 133.322 dyn/cm 2 0.1 cmhg 1333.22 hbar 10000000 mhg 133322.0 tonf/ft 2 (UK) 107252.0 mm/h 2 O [1] 9.80665 tonf/in 2 (UK) 15444300 cm/h 2 O [1] 98.0665 inh 2 O (USA) [2] 248.64135 N/m 2 1 fth 2 O (USA) [2] 2983.6983 hpa 100 kp/mm 2 9806650 kpa 1000 kp/cm 2 98066.5 MPa 1000000 kp/m 2 9.80665 torr 133.322 Table of pressure units and conversion factors Unit Conversion To convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS 2, calculate as follows: VALUE 2 = VALUE 1 x FACTOR 1 FACTOR 2 Note: The conversion factor for pressure units referenced [1] are calculated for a water temperature of 4 C. Pressure units referenced [2] are calculated for a water temperature of 68 F these units are normally used in the USA. v K382 Issue No. 1

Introduction At manufacture this instrument is calibrated against precision calibration equipment which is traceable to the U.K. National Physical Laboratory. The DPI 142 and DPI 150 incorporate a calibration ; for the indicator to stay accurate, a calibration check should be carried out at chosen intervals (the recommended maximum interval is 12 months). If the accuracy of the indicator is not within the permissible deviation, carry out a calibration adjustment. Using the Set-up/Status/Calibration History, the calibration status of the indicator can be displayed on the front panel screen. The Calibration History gives a list of dates of the stored calibration corrections. Table 1 Absolute Pressure Calibration Equipment and Requirements DPI 142 and DPI 150 Option A Note: Equivalent substitutes may be used. Range Uncertainty required Recommended Equipment Type 35 mbar to 3500 mbar ±0.005% Rdg RUSKA 2465 Piston gauge 1 K382 Issue No. 1

Table 2 Gauge Pressure Calibration Equipment and Requirements DPI 150 Note: Equivalent substitutes may be used. Range Uncertainty required Recommended Equipment Type 25 mbar to 70 mbar ±0.005% FS, ±0.005%Rdg Ruska 7050 or 7250 Indicator/ controller 200 mbar to 1 bar ±0.005%Rdg Pressurements 6000 or Ruska 2465 Dead weight/ piston gauge 2 bar to 20 bar ±0.005%Rdg Ruska 2465 Piston gauge 35 bar to 200 bar ±0.005%Rdg Ruska 2465 Ruska 2485 Piston gauge 350 bar 700 bar ±0.01%Rdg Ruska 2485 Piston gauge K382 Issue No. 1 2

Preliminary Operations Review and become familiar with the whole procedure before beginning a calibration process. Allow at least one hour for the calibrator to thermally stabilize after switching on and before calibration. Before starting a calibration procedure: Carry out a leak test. If necessary and as a separate procedure from the calibration procedure: Notes on calibration The pressure standard outlet port and the indicator inlet port must be at the same level. On gauge measurements, applying positive pressure to the reference port produces a negative pressure. Set the calibrator units of pressure to one of the following recommended units of pressure measurement: Pa, hpa, kpa, MPa, bar, mbar, torr, kgf/cm 2. Calibration Security GE Druck strongly advise protection of the set-up s in this equipment. Unauthorised access to the calibration can result in degraded performance and inaccuracy. The factory set PIN is: 4321 press MENU. The code can be changed to another code of 4 digits; entering 0000 disables this security facility. Note; Refer to the user manual to change PIN codes. General 1. Carry out a leak check, refer to the User Manual. 2. Connect the instrument and carry out a calibration check. 3. If necessary, carry out a calibration adjustment. 4. After completing the calibration adjustments, adjust the calibration standard to atmospheric pressure. 5. Carry out a calibration check to verify this procedure. 6. Disconnect calibration standard from the instrument. 3 K382 Issue No. 1

Calibration Check DPI 142 and DPI 150 with Option A After a successful leak test, set the instrument indicator mode: 1. Connect the pressure standard for the pressure range to be checked. Note: In DPI 150 instruments, Option A connects internally to the negative (reference) pressure port. 2. Adjust the pressure standard to the first pressure value (A) in the Table 3. 3. Compare the pressure value of the pressure standard to the value displayed and record the difference. 4. Repeat steps 2 and 3 for pressure value (B) in the Table 3. 5. If the recorded difference exceeds the tolerance in the table, the instrument needs a calibration adjustment. 6. Adjust the pressure standard to the first pressure value (A) in the Table 3. Applied Pressure A B Tolerance ± 0.01%FS ± 0.01%FS Pressure Ranges The DPI 142 instrument and the DPI 150 instrument with Option A* have the following pressure ranges. Table 3 Applied Barometric Pressures Barometric Range Pressure Values A B *750 to 1150 mbar absolute 750 mbar 1150 mbar 35 to 1310 mbar absolute 35 mbar 1310 mbar 35 to 2620 mbar absolute 35 mbar 2620 mbar 35 to 3500 mbar absolute 35 mbar 3500 mbar Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com

Calibration Adjustment To adjust a calibration range of the DPI142 and DPI 150 Option A, proceed as follows.1 user supervisor calibration status enter PIN: 0000 change PIN sensor correction + - 3.65 mbar g 750.000 mbar a 1150.00 mbara 20000.00 mbarg 21000.00 mbara range do cal. 3.65 start calibration of this range? press ok to startmbar g to cancel range do cal. 0.00 Apply positive full scale pressure press `` when applied value mbar g is stable set pressure standard set pressure standard Apply nominal zero pressure press 3.65 when applied value is stablembar g 1150.00 mbar a Sampling please wait 750.00 mbar a 1150.00 enter actual pressure value produced by pressure standard Sampling please wait applied value: 750.00 applied value: + - Do you wish to accept cal? Press 1150.00mbar ok to accept, g to cancel enter actual pressure value produced by pressure standard + - 5 K382 Issue No. 1

Calibration Check DPI 150 After a successful leak test, set the instrument to indicator mode: 1. Connect the pressure standard for the pressure range to be checked. 2. Set the pressure standard to zero pressure, if necessary press the zero key so that the instrument display shows zero. 3. Adjust the pressure standard to the 0% span pressure value. 4. Compare the pressure value of the pressure standard to the value displayed and record the difference. 5. Repeat steps 3 and 4 for the 100% span pressure value. 6. If the recorded difference exceeds the tolerance in the table, the instrument needs a calibration adjustment. 7. Adjust the pressure standard to zero pressure. Applied Pressure % span Tolerance 0 ± 0.01%FS 100 ± 0.01%FS The DPI 150 instrument has the following pressure ranges. Table 4 Transducer Equivalent Pressure Units range bar mbar psi Pa hpa kpa MPa 0.025 25 036 2500 25.0 2.5 0.0025 0.07 70 1 7000 70 7 0.007 0.2 200 3 20000 200 20 0.02 0.35 350 5 35000 350 35 0.035 0.7 700 10 70000 700 70 0.07 1 1000 15 100000 1000 100 0.1 2 2000 30 200000 2000 200 0.2 3.5 3500 50 350000 3500 350 0.35 7 7000 100 700000 7000 700 0.7 10 10000 150 1000000 10000 1000 1 20 20000 300 2000000 20000 2000 2 35 35000 500 3500000 35000 3500 3.5 70 70000 1000 7000000 70000 7000 7 100 N/A 1500 N/A 100000 10000 10 135 N/A 2000 N/A 135000 13500 13.5 200 N/A 3000 N/A 200000 20000 20 350 N/A 5000 N/A 350000 35000 35 700 N/A 10000 N/A 700000 70000 70 K382 Issue No. 1 6

Calibration Adjustment To adjust a calibration range of the DPI 150, proceed as follows.1 user supervisor calibration status enter PIN: 0000 change PIN sensor correction + - 1003.65 mbar a 0.0000 mbar g 1150.00 mbara 20000.00 mbarg 21000.00 mbara range do cal. start calibration of this range? press ok to startmbar g to cancel range do cal. 3.65 Apply positive full scale pressure press `` when applied value mbar g is stable 0.00 set pressure standard to 100% span set pressure standard to 0% span Apply nominal zero pressure press 3.65 when applied value is stablembar g 20000.00 mbar g Sampling please wait 0.0000 mbar g 20000 enter actual pressure value produced by pressure standard Sampling please wait applied value: 0.0000 applied value: + - Do you wish to accept cal? Press 1150.00mbar ok to accept, g to cancel enter actual pressure value produced by pressure standard + - Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com