Instruction Manual Dräger MSI ALV

Size: px
Start display at page:

Download "Instruction Manual Dräger MSI ALV"

Transcription

1 Dräger Safety MSI GmbH Rohrstraße 32 D Hagen Tel.: / Fax: / info@draeger-msi.de D 914; Edition

2 Contents 1. Hints page 2 2. Pipe Leak Test page 3 3. Automatic Tightness Test page 4 4. Leak Rate Measurement page Determination of Gas Type and Pressure Reference Value 4.2 Preparing for Leak Rate Measurement page Stabilization page Measurement of the Leak Rate page Result of Leak Rate Measurement page Finish Leak Rate Measurement page Error Messages 6. Technical Data page Maintenance and Service page 14 D of 15 Edition

3 1. Hints Any use of a Dräger MSI 150 PRO2 with option " Pipe Leak Test ", requires a full understanding and strict adherence to this instruction manual, to the manual of the Dräger MSI 150 PRO2 and to national and international standards. Observe especially the regulations concerning the handling of explosive gases. The instrument is only to be used for the purposes specified in here. 1.1 Automatic Tightness Test The Dräger MSI ALV together with the stack gas analyzer Dräger MSI 150 PRO2 makes it possible to do a tightness test completely automatic. The user has to connect the gas pipe with the Dräger MSI ALV by only one pressure tube. The pressure in the gas pipe then will be pumped up to a little more than 110 mbar automatically. After a time of stabilization for temperature equalization etc. the pressure in the gas pipe will be measured for 10 min, without that the user has to open or close a valve. After the measurement the pressure at the beginning and at the end of the measurement (after 10 minutes) is displayed. The result together with date and time may be printed. 1.2 Leak Rate Measurement The Dräger MSI ALV together with the stack gas analyzer Dräger MSI 150 PRO2 makes it possible to measure the leak rate of a gas pipe at operating pressure. The used method is patent pending. After closing all valves of the gas consumer loads the user has to connect the gas pipe with the Dräger MSI ALV by only one pressure tube. After a time of stabilization for temperature equalization etc. the analyzer asks the user to close the gas supply (e. g. main valve direct behind the gas meter). The instrument measures the pressure in the gas pipe and opens after an automatically calculated period an comparison leak. The measured pressure variation with and without the comparison leak leads to the calculation of the leak rate of the gas pipe. From the leak rate follows the calculation of the volume of the gas pipe. The leak volume and the volume of the gas pipe are displayed and may be printed together with date and time. Edition of 15 D 914

4 2. Pipe Leak Test If this function is not part of the instrument, use the Dräger MSI Set-up-CD and a PC for installing the new firmware into the Dräger MSI 150 PRO2. See the instructions on the CD (pdf-file). After switching on the Dräger MSI 150 PRO2 (see Instruction Manual Dräger MSI 150 PRO2), call the menu of functions by pushing "! ", if the PRO2 shows the fuel type selection menu. The display will read: Functions Reset Pipe leak test Qa average Stack leak test Cont. With pushing " " you may mark the function " Pipe Leak Test ". with " F " you start this function and the display reads: Pipe leak test Leak rate Tightness End Sel. With pushing " " ( ) you may mark the wanted function. With " F " (Selection) you may start the marked function. Selectable functions are: End = calls "Menu of Functions" (see Instruction manual of Dräger MSI 150 PRO2) Tightness = calls "Automatic Tightness Test" (see 3.) Leak rate = calls "Automatic Leak Rate Test" (see 4.) D of 15 Edition

5 3. Automatic Tightness Test After choosing " Tightness test ", the display reads: Tightness Connect hoses Start With "!" () you may stop the "Tightness test " and skip to "Pipe Leak Test" (see 2). By pushing "F" (Start) you may start the " Tightness test ". Mount the Leak measurement adapter ALV onto the gas outlet and the pressure inlet of the PRO2 so, that you are able to read the words "Hauptprüfung" and "Tightness", during you are looking onto the display of the PRO2. Hauptprüfung Tightness Leak measurement adapter ALV Gas outlet Pressure inlet Put the hose coupling of the supplied pressure probe on the upper connector of the Dräger MSI ALV and mount the hose of the pressure probe on the nipple of the gas pipe. Now you may start the "Tightness Test". Edition of 15 D 914

6 Immediately after you started the test the flushing pump of the PRO2 begins to increase the pressure in the gas pipe. The display reads: Tightness Pumping P T 65.4 mbar 21 s The display shows that the pump is working and it informs you about the actual pressure in the tested gas pipe (P) and how much time has passed since the pump starts (T). With "!" () you may cancel the " Tightness Test" and skip to "Pipe Leak Test". If the pressure in the tested gas pipe exceeds 115 mbar, the flushing pump of the PRO2 is automatically stopped and the valve in the ALV closes. Now starts the stabilisation period (2 until 10 minutes) to get temperature balance. During this stabilisation period the display reads: Tightness Stabilisation P T mbar 20 s The display shows that the stabilisation time is running and it informs you about the actual pressure in the tested gas pipe and for how many seconds stabilisation is running. With "!" () you may cancel the " Tightness Test" and skip to "Pipe Leak Test". Is the temperature balance reached, the PRO2 switches to measurement mode. During this the display reads: Tightness Measurement Start mbar P T mbar 89 s The display shows that the measurement is running and it informs you about the pressure at the beginning (Start), the actual pressure (P) and how long the measuring is running. With "!" () you may cancel the " Tightness Leak Test" and skip to "Pipe Leak Test". D of 15 Edition

7 After 10 minutes of measuring the " Tightness Leak Test" is automatically finished and the next display screen is called and the display will read e.g.: Tightness Test time Start End Stab.time Cont. 10 min mbar mbar 2 min The display informs you how long the measuring has been done (Test time), about the pressure at the beginning (Start), the pressure at the end of the measurement (End) and how long stabilisation has needed (Stab.time). With "!" () you may cancel the " Tightness Test" and skip to "Pipe Leak Test". After pushing the key "F" (Continue) the display reads: Print NO YES With "F" (NO) you skip to the last display screen. After pushing the key "!" (Yes) the printout of the Tightness Test is done together with date and time. After that the instrument switches to the last display screen. Edition of 15 D 914

8 4. Leak Rate Measurement After choosing " Leak rate ", the function "Determination of Gas Type and Pressure Reference Value" is started immediately. 4.1 Determination of Gas Type and Pressure Reference Value After the start of this function the display reads: Select gas Air Propane Natural Gas Cont. With " " ( ) or "!" ( ) you mark the gas type. With pushing "F" (Continue) the marked gas type is chosen The display now reads: Select nom. pressure Pn mbar Cont. + - With " " or "!" you may change the pressure reference value. With pushing "F" (Continue) the displayed pressure reference value is accepted, the purge pump of the PRO2 starts working and closes the pneumatic valve in the ALV. Now the function "Preparing for Leak Rate Measurement " (see 4.2) is called D of 15 Edition

9 4.2 Preparing for Leak Rate Measurement If "Preparing for Leak Rate Measurement" has been called, the display reads: Connect ALV P 0.08 mbar Cont. With "!" () you may cancel the "Leak Rate Measurement " and skip to "Finish Leak Rate Measurement" (see 4.6). Mount the leak rate measurement adapter ALV onto the gas outlet and the pressure inlet of the PRO2, so that you are able to read the words "Leckmenge" and "Leak volume" if you are looking onto the display of the PRO2. Leak measurement adapter ALV Leckmenge Leak volume Gas outlet Pressure inlet Put the coupling of the pressure probe on the lower connector of the ALV and mount the hose of the pressure probe on the nipple of the gas pipes plug. Open the plug, the PRO2 now displays in the second line the current pressure in the gas pipe. Observe national and international regulations concerning the handling of explosive gases. Edition of 15 D 914

10 After you have pushed "F" (Cont.) the display reads: Purge P t mbar 5 sek. After 5 seconds the purge pump of the PRO2 is stopped and the pneumatic valve in the Dräger MSI ALV opens. Now gas exhausts from the gas pipe through the capillary tube in the ALV. The gas gets to the air through the small holes on top of the Dräger MSI ALV. (Observe national and international regulations concerning the handling of explosive gases). After 15 seconds the purge pump of the PRO2 starts again and the pneumatic valve in the Dräger MSI ALV is closed. The capillary tube (comparison leak) now is filled with gas from the gas pipe. The Dräger MSI 150 PRO2 now skips to " Stabilization " (see 4.3). 4.3 Stabilization The measuring system waits for stabilization of the gas pressure (ca. 2 to 6 min.) et al. to get temperature balance. The display reads now: Stabilise P t dp mbar 75 sec mbar The display informs you about the current pressure in the gas pipe, how much time has passed since the stabilization period started and the decrease of pressure (negative values mean the pressure in the gas pipe has risen). With "!" () you may cancel the "Leak Rate Measurement " and skip to "Finish Leak Rate Measurement" (see 4.6). If a sufficient stabilization is achieved (dp < 1% of the gas pressure) the stabilization period will be finished and the measuring of the leak rate is started (4.4). D of 15 Edition

11 4.4 Measurement of the Leak Rate The end of the stabilization period will be announced acoustically, the display reads now: Close main valve P 18,62 mbar dp 0,29 mbar T 25,3 C Cont. The measuring system asks to close the gas supply (e. g. main valve direct behind the gas meter), and informs about the current pressure, the pressure decrease and the ambient air temperature. With "!" () you may cancel the "Leak Rate Measurement " and skip to "Finish Leak Rate Measurement" (see 4.6). If the gas supply has been closed and the gas pipe has a leak, the PRO2 with ALV will recognize a pressure decrease. If the pressure decrease is > 0,4 mbar, the measurement will start immediately. If the gas pipe is tight or the volume of the pipe very large and the leak very small ( > very small pressure decrease), you may push "F" (Cont.). The display now reads: Start measurement P mbar dp 0.29 mbar 55 sec. After 60 seconds the leak measurement is automatically started. The start of the measurement will be announced acoustically, the display reads: Leakage measurement P mbar t 75 sek. dp 0.04 mbar With "!" () you may cancel the "Leak Rate Measurement " and skip to "Finish Leak Rate Measurement" (see 4.6). The display informs you about the current pressure (P) in the gas pipe, how much time has passed since the measurement has been started (t) and the decrease of pressure dp. Edition of 15 D 914

12 If the pressure decrease in the gas pipe is > 0,9 mbar or the measurement lasts longer than 5 minutes, then the purge pump is stopped and the pneumatic valve in the Dräger MSI ALV is open. Now gas from the gas pipe will exhaust through the comparison leak. The leak rate of the comparison leak is about 5 l/h. The begin of the comparison leak measurement is announced acoustically. The gas gets to the air through the small holes on top of the Dräger MSI ALV (Observe national and international regulations concerning the handling of explosive gases). The display of the Dräger MSI 150 PRO2 reads: Comparison P mbar t 15 sec. dp 0.14 mbar Abbr. After the comparison leak measurement (dp > 0,9 mbar or t > 5 minutes) the purge pump is started, the pneumatic valve closes and the "Result of the Leak Rate Measurement" is called. The end of the comparison leak measurement is announced acoustically. 4.5 Result of the Leak Rate Measurement If the Result of the Leak Rate Measurement has been called the display reads: Result P 21.8 mbar L(P) 0.01 l/h L(n) 0.02 l/h Vol 13.5 l End New Print You are informed about the mean pressure value "P" during the leak rate measurement, the measured leak rate "L(P)", the calculated leak rate for the pressure reference value "L(n)" and the measured volume "Vol" of the gas pipe. With "F" (End) you may finish the "Leak Rate Measurement" and skip to "Finish Leak Rate Measurement" (see 4.6). With " " (New) you may start a new leak rate measurement. The PRO2 will ask you to open the gas supply (e. g. main valve) and skips after pushing "F" ( Cont.) to "Stabilization" (see 4.3). By pushing "!" (Print) the printout of the measurement results is started. After the printing the above readout of the display is called again. D of 15 Edition

13 4.6 Finish Leak Rate Measurement If the Leak Rate Measurement has been finished or cancelled the display reads: Please disconnect hose and shut gas pipe. Cont. Close the plug of the nipple of the gas pipe and remove the hose. Observe national and international regulations concerning the handling of explosive gases. By pushing "F" (Cont.) you may leave the leak rate measurement and the menu for choosing the functions leak rate or tightness (see 2.) is called. 5. Error messages The Dräger MSI 150 PRO2 with the option Pipe Leak Test recognizes the following errors: Message Error Remedy P-Sensor Signal of pressure sensor too high during zero calibration. Disconnect hose from ALV, switch off Dräger MSI 150 PRO2 and switch it on again. Pump time overflow or Pressure cannot be developed pressure generation lasts too long Service Connect hose between ALV and gas pipe Tighten the plug of the gas pipe Service Edition of 15 D 914

14 6. Technical Data 6.1 Technical Data ALV Humidity: barometric pressure: Operating temperature: +5 C C Storage temperature: -20 C C % RF not condensing hpa Dimensions: 105 mm x 50 mm x 15 mm (L x H x W) Weight: 140 g 6.2 Technical Data Tightness Test Pump power: Test pressure: Resolution: Stabilisation time: Measuring time: 0,3 bis 0,5 Liter / min 110,1 bis 119,9 mbar 0,1 mbar 2 10 min, depends of pressure change 10 min 6.3 Technical Data Leak Rate Measurement Gas types: natural gas, propane, air Leak rate Range: l / h Resolution: 0,01 l / h Volume Range: l Resolution: 0,1 l Pressure Range: mbar Resolution: 0,1 mbar D of 15 Edition

15 7. Maintenance and Service 7.1 Service In order to assure accurate measurements and the reliability of the functions the Dräger MSI 150 PRO2 together with the leak test adapter ALV should be checked and if requested be calibrated by an authorised service point once a year. (See instruction manual of PRO2 too) 7.2 Consumables and accessories Printer with infra-red data transmission Paper for IR-Printer Consumable set consisting of: 10 x Disc filter 20 x Filter fleece ø 26 Edition of 15 D 914

16 Remarks: D of 15 Edition

Highly efficient flue gas analyzer

Highly efficient flue gas analyzer Highly efficient flue gas analyzer testo 320 Just a few clicks away from a heating system diagnosis O 2 High-resolution colour graphic display Quick and easy menu structure Storage space for 500 measurement

More information

Thermo Top - Troubleshooting Tree

Thermo Top - Troubleshooting Tree Thermo Top - Troubleshooting Tree 07-15-2002 CAUTION Troubleshooting requires comprehensive knowledge about the structure and theory of operation of the Thermo Top heater. Troubleshooting and repairs may

More information

Online Infrared Flue Gas Analyzer Gasboard 3000

Online Infrared Flue Gas Analyzer Gasboard 3000 Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume

More information

Valve Proving System VPS 504 Series S06 (120 VAC) Series S05 (24 VDC)

Valve Proving System VPS 504 Series S06 (120 VAC) Series S05 (24 VDC) Valve Proving System VPS 504 Series S06 (120 VAC) Series S05 (24 VDC) VPS 504 S06 (120 VAC) is: CSA Certified File # 1637485 CSA Requirement No. 4-01 (USA) Technical Information Letter R-15 (Canada) UL

More information

820 NOVA mv THERMOPILE POWER SUPPLY THERMOELECTRIC SAFETY DEVICE SERVO-CONTROLLED PRESSURE REGULATOR SHUT-OFF SOLENOID VALVE.

820 NOVA mv THERMOPILE POWER SUPPLY THERMOELECTRIC SAFETY DEVICE SERVO-CONTROLLED PRESSURE REGULATOR SHUT-OFF SOLENOID VALVE. SIT Group 820 NOVA mv MULTIFUNCTIONAL CONTROL FOR GAS THERMOPILE POWER SUPPLY THERMOELECTRIC SAFETY DEVICE SERVO-CONTROLLED PRESSURE REGULATOR SHUT-OFF SOLENOID VALVE PIN 3AP700/2 SINGLE-KNOB MULTIFUNCTIONAL

More information

Instruction Manual. T3111 T3111L T3111P T3111Ex

Instruction Manual. T3111 T3111L T3111P T3111Ex www.cometsystem.cz Instruction Manual T3111 T3111L T3111P T3111Ex Programmable transmitter of temperature, relative humidity and other calculated humidity values with external probe and 4-20 ma outputs

More information

Continuous Monitoring of Flue and Combustible or Hazardous Gases and Vapors

Continuous Monitoring of Flue and Combustible or Hazardous Gases and Vapors Continuous Monitoring of Flue and Combustible or Hazardous Gases and Vapors www.static-control.in 1 Introduction The detection or monitoring of flue gases and or hazardous gases has been a complex subject

More information

Software Features of Auto Gas Gas Analyzer Features HC, CO, CO2, Lambda, Air Fuel, NOx. Has optional RPM and Oil Temp.

Software Features of Auto Gas Gas Analyzer Features HC, CO, CO2, Lambda, Air Fuel, NOx. Has optional RPM and Oil Temp. Auto Gas Ultimate Model Shown Above Software Features of Auto Gas Gas Analyzer Features HC, CO, CO2, Lambda, Air Fuel, NOx. Has optional RPM and Oil Temp. Durable high strength aluminum case with all of

More information

LANCOM4 PORTABLE FLUE GAS MONITORING. NO NO2 NOx CO2 H2S SO2 CxHy. CO low CO high O 2 QUALITY CUSTOMER SOLUTIONS WWW.LANCOM4.COM

LANCOM4 PORTABLE FLUE GAS MONITORING. NO NO2 NOx CO2 H2S SO2 CxHy. CO low CO high O 2 QUALITY CUSTOMER SOLUTIONS WWW.LANCOM4.COM PORTABLE FLUE GAS MONITORING WWW.LANCOM4.COM i CO low CO high O 2 NO NO2 NOx CO2 H2S SO2 CxHy QUALITY CUSTOMER SOLUTIONS AMETEK LAND HAS BEEN BUILDING PRECISION MEASURING EQUIPMENT SINCE 1947. WE ARE SPECIALISTS

More information

Installation and maintenance instructions. Danfoss Vent

Installation and maintenance instructions. Danfoss Vent Installation and maintenance instructions Danfoss Vent 1 Unit design... 2 2 Project planning and installation instructions... 3 2.1 Ventilation...3 2.2 Piping installation....3 2.3 Condensation line...4

More information

4-Channel Thermometer / Datalogger

4-Channel Thermometer / Datalogger USER GUIDE 4-Channel Thermometer / Datalogger RTD and Thermocouple Inputs Model SDL200 Introduction Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter

More information

How To Fix A Car With A Carbide Engine

How To Fix A Car With A Carbide Engine Section 12 Adjustment, Repair, and Replacement Page 12-1 Section 12 Adjustment, Repair and Replacement Section Contents Page Overview...12-5 Belts Belt Guard Removal/Installation...12-7 Belt Removal/Installation...12-11

More information

COMBIMASS. Technical Data COMBIMASS eco-bio +

COMBIMASS. Technical Data COMBIMASS eco-bio + COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments

More information

Measurement Solutions for Refrigeration Technology.

Measurement Solutions for Refrigeration Technology. Measurement Solutions for Refrigeration Technology. Digital manifold, leak detection, vacuum measurement, and lots more. Testo refrigeration technology Digital is better: Refrigeration technology expertise

More information

Lambda Meter Measurement of parameter λ (Lambda) air / fuel ratio (AFR)

Lambda Meter Measurement of parameter λ (Lambda) air / fuel ratio (AFR) Lambda Meter Measurement of parameter λ (Lambda) air / fuel ratio (AFR) Wide band lambda probe measures the numerical value of the ratio Air Fuel Ratio AFR or parameter of λ (Lambda) by measuring the oxygen

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

Calibrating Differential Pressure Sensors

Calibrating Differential Pressure Sensors Calibrating Differential Pressure Sensors Additel Corporation 22865 Savi Ranch Parkway, Ste F Yorba Linda, CA 92887 Introduction Differential pressure sensors are common in the process industry and cover

More information

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications P.O. Box 26, FIN-00421 Helsinki, FINLAND Tel: +358 9 894 91 Fax: +358 9 8949 2485 Email: industrialsales@vaisala.com www.vaisala.com The Vaisala

More information

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration

More information

Drayton Digistat +2RF/+3RF

Drayton Digistat +2RF/+3RF /+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer

More information

Pressure Switch for gas and air

Pressure Switch for gas and air Switch for gas and air GAO-A4 GMH-A4 GML-A4 UL Listed UL 353 File # MH 16628 CSA Certified CSA C22.2 No. LR 53222 Certification file # 201527 FM Approved Class 3510, 3530 File # J.I. 1Y919.AF Commonwealth

More information

SELECTION, APPLICATION AND MAINTENANCE

SELECTION, APPLICATION AND MAINTENANCE DIESEL PROTECTION SYSTEMS Automatic Diesel Engine Shut Down System for Safe Area Applications SELECTION, APPLICATION AND MAINTENANCE Series 300 Series 310 SYSTEM DESCRIPTION Suitable for attended engine

More information

Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version

Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version Electronic pressure measurement Digital pressure gauge for industrial applications Model DG-10-S, Standard version Model DG-10-E, Enhanced version WIKA data sheet PE 81.66 Applications Machine building

More information

Via Raffaello,33/A - 42100 Reggio Emilia Italia Tel +39 0522.514.461 Fax +39 0522 514469 - E-mail: aftersales@landi-gas.

Via Raffaello,33/A - 42100 Reggio Emilia Italia Tel +39 0522.514.461 Fax +39 0522 514469 - E-mail: aftersales@landi-gas. Trouble Shooting Solving problems with LSI-NSI software Update to version 2.1.0 1/14 Rev.0 2/14 Rev.0 CONTENTS 1.1. Switching from petrol to gas PAGE 4 1.2. Drop at medium-high speeds PAGE 5 1.3. Functioning

More information

Automatic Bypass Control AVDO

Automatic Bypass Control AVDO Application AVDO is a self-acting constant flow control primarily used either to maintain minimum flow rates through e.g. a low-capacity gas boiler or to control the differential pressure in a central

More information

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL

NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Esenboğa Yolu 22 km Akyurt - ANKARA / TURKEY Tel : (00 90 312)

More information

ON-VEHICLE INSPECTION

ON-VEHICLE INSPECTION Sight Glass I11244 ONVEHICLE INSPECTION 1. INSPECT REFRIGERANT VOLUME Observe the sight glass on the liquid tube. AC3 AC22F05 Test conditions: Running engine at 1,500 rpm Blower speed control switch: HI

More information

PACE Modular Pressure Controller

PACE Modular Pressure Controller GE Measurement & Control Solutions PACE Modular Pressure Controller A new generation of high precision Druck pressure controller, designed for test bench, bench top & rack mount calibration & automated

More information

CONTOIL DFM 8EDM. Table of contents. Mounting and operating instructions

CONTOIL DFM 8EDM. Table of contents. Mounting and operating instructions Mounting and operating instructions CONTOIL DFM 8EDM Table of contents Safety instructions 2 - Designed use - Installation, commissioning and operation - Operational safety - Return of the instruments

More information

Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains

Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains Drayton Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains Invensys Controls Europe Technical Helpline: +44 (0) 845 130 7722 www.draytoncontrols.co.uk Installation

More information

MAKE OF AUTOMOBILE: NUMBER : 076/0800901 DATE : 2008-12-08 Copyright Prins Autogassystemen B.V. 2008 VERSION NR : 2008-10-24 B

MAKE OF AUTOMOBILE: NUMBER : 076/0800901 DATE : 2008-12-08 Copyright Prins Autogassystemen B.V. 2008 VERSION NR : 2008-10-24 B MAKE OF AUTOMOBILE: HONDA TYPE: CIVIC PISTON DISPLACEMENT: 1800 NUMBER OF VALVES: 16 ENGINE NUMBER: R18A TYPE VSI INJECTOR ( COLOR ) ORANGE MODEL YEAR: 2008 ENGINE SET NUMBER 348/0250200.07 NUMBER : 076/0800901

More information

REGTRONIC OPERATING INSTRUCTIONS

REGTRONIC OPERATING INSTRUCTIONS REGTRONIC OPERATING INSTRUCTIONS 1.FEATURES Electrical connection: M12 8-pin connector Preset pressure range 0.05-10 bar with possible full scale regulation. 10 100 mbar adjustable deadband Supply pressure:

More information

Doc.No. NDP 192U-02. Electric Pump Controller CE-124P. Instruction Manual

Doc.No. NDP 192U-02. Electric Pump Controller CE-124P. Instruction Manual Doc.No. NDP 192U-02 Electric Pump Controller CE-124P Instruction Manual 1. Parts Name and Function 1) POWER LED Indicate that CE-124P is powered up. 2) START/STOP KEY (SELECTOR KEY) Used to start and stop

More information

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater SALES SPECIFICATION SC7640 Auto/Manual High Resolution Sputter Coater Document Number SS-SC7640 Issue 1 (01/02) Disclaimer The components and packages described in this document are mutually compatible

More information

Evaporative emissions system

Evaporative emissions system just a test. Evaporative emissions system 20-48 Function description of EVAP canister system Depending upon the air pressure and ambient temperature, fuel vapor will form above the level of fuel in the

More information

Fluxi 2000/TZ Turbine Gas Meter

Fluxi 2000/TZ Turbine Gas Meter Gas Fluxi 2000/TZ Turbine Gas Meter Turbine gas meters are flow meters. The flow of gas turns a turbine wheel, and thus, the rotating speed of the turbine is proportional to the linear speed of the gas.

More information

OASIS SERVICE & TROUBLESHOOTING MANUAL (WITH DM10 OR DM12)

OASIS SERVICE & TROUBLESHOOTING MANUAL (WITH DM10 OR DM12) OASIS SERVICE & TROUBLESHOOTING MANUAL (WITH DM10 OR DM12) Revision 1.0 Page 1 of 50 THE INFORMATION CONTAINED IN THIS DOCUMENT IS THE SOLE PROPERTY OF INTERNATIONAL THERMAL RESEARCH, LTD. ANY REPRODUCTION

More information

Adding Pneumatic Preset Counter. Type 497. Continuously visible preset Integrated pneumatic reset 3 or 5-digit display Convenient button setting

Adding Pneumatic Preset Counter. Type 497. Continuously visible preset Integrated pneumatic reset 3 or 5-digit display Convenient button setting Adding Pneumatic Preset Counter Type 497 Continuously visible preset Integrated pneumatic reset 3 or 5-digit display Convenient button setting VARIABLE PILOT SIGNAL Using two pneumatic preset counters

More information

Introduction. Refrigerant Leak Detecting System. Feature and Benefits

Introduction. Refrigerant Leak Detecting System. Feature and Benefits PSC European Refrigeration Controls Catalogue Catalog Section 9 Product Bulletin PD-GAS-E Refrigerant Leak Detecting System Introduction This range of refrigerant leak detecting systems is designed for

More information

80-115 80-140 80-180 95-199 95-260 70-360 65-400 65-500

80-115 80-140 80-180 95-199 95-260 70-360 65-400 65-500 Release R.2.0 Uk 80-115 80-140 80-180 95-199 95-260 70-360 65-400 65-500 Installation User and Service Manual www.state-waterheaters.co.uk your installer Warning Read this manual carefully before first

More information

SLAM-SHUT VALVE. Type OSE. Europe, Middle East, Africa and Asia Pacific Documents Only. Process Management TM

SLAM-SHUT VALVE. Type OSE. Europe, Middle East, Africa and Asia Pacific Documents Only. Process Management TM SLAM-SHUT VALVE Type OSE Process Management TM DESCRIPTION The purpose of the OSE slam shut valve is to totally and rapidly cut off gas flow when the outlet pressure exceeds or drops below the setting.

More information

BB-18 Black Body High Vacuum System Technical Description

BB-18 Black Body High Vacuum System Technical Description BB-18 Black Body High Vacuum System Technical Description The BB-18 Black Body is versatile and is programmed for use as a fixed cold target at 80 K or variable target, at 80 K- 350 K no extra cost. The

More information

Portable Air Conditioner

Portable Air Conditioner Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE

More information

13" Real Time Quad Monitor with Camera System

13 Real Time Quad Monitor with Camera System 3" Real Time Quad Monitor with Camera System MODEL NO.: QSM-300 CTRAST BRIGHT VOLUME EXIT MENU AUTO PIP CH CH 3 CH POWER Complete Ready- to- use, Easy-to-manage, Commercialized Package Set. Real Time High

More information

Outages Happen 24/7 So Should Monitoring / SAFEGUARD YOUR HIGH VOLTAGE ASSETS WITH VAISALA ONLINE MONITORING

Outages Happen 24/7 So Should Monitoring / SAFEGUARD YOUR HIGH VOLTAGE ASSETS WITH VAISALA ONLINE MONITORING Outages Happen 24/7 So Should Monitoring / SAFEGUARD YOUR HIGH VOLTAGE ASSETS WITH VAISALA ONLINE MONITORING A Small Investment to Generate Measurable Benefits Sometimes, solving challenges is surprisingly

More information

G670 Intermittent Pilot Ignition Controls

G670 Intermittent Pilot Ignition Controls Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin Issue Date 0300 Intermittent Pilot Ignition Controls Figure 1: Intermittent Pilot Ignition Control

More information

SunMaxx Solar Filling Station Operating Instructions

SunMaxx Solar Filling Station Operating Instructions SunMaxx Solar Filling Operating Instructions Content 1. Declaration of conformity... 2 2. Introduction... 2 3. Transportation and unpacking... 4 4. Mounting and commissioning... 5 5. End of operation...

More information

Service manual. Website: www.andico.com.au CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL.

Service manual. Website: www.andico.com.au CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL. Website: www.andico.com.au Service manual CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL. - ONLY FOR AUTHORISED SERVICE PERSONNEL. MODELS: MPK1-09CR-QB8 MPK1-12ER-QB6

More information

T0118 T2118 T3118. Instruction Manual

T0118 T2118 T3118. Instruction Manual Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity

More information

Experiment HE-6: Breath by Breath Measurement of Respiratory Exchange Ratio (RER)

Experiment HE-6: Breath by Breath Measurement of Respiratory Exchange Ratio (RER) Experiment HE-6: Breath by Breath Measurement of Respiratory Exchange Ratio (RER) Background The two sources of energy available for human metabolism are carbohydrates (CHO) and fats. These molecules are

More information

- mark gas burners Cuenotherm NC.4 - NC.6 - NC.9 14.5-90kW.

- mark gas burners Cuenotherm NC.4 - NC.6 - NC.9 14.5-90kW. - mark gas burners Cuenotherm NC.4 - NC.6 - NC.9 14.5-9kW. Low NO x Comprehensive control panel Gas rack Low NOx combustion head Strong light alloy pressuremoulded housing Electronic ignition transformer

More information

SOUND LEVEL METER 94405

SOUND LEVEL METER 94405 SOUND LEVEL METER 94405 INSTRUCTION MANUAL REC FAST A/C 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter only as specified

More information

Installing a Probe. Typical Installation

Installing a Probe. Typical Installation for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assembly

More information

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05 RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,

More information

OEM-EP Pressure Controllers

OEM-EP Pressure Controllers Pressure Controllers Typical Applications Carrier Gas Pressure Control Air over Liquid Control Mass Spectrometer Process Gas Supply Pressure Control Miniature ler The Miniature ler converts a variable

More information

Quinta Range. Condensing Boilers REMEHA QUINTA RANGE. High efficiency wall hung condensing boiler with ultra low NOx emission

Quinta Range. Condensing Boilers REMEHA QUINTA RANGE. High efficiency wall hung condensing boiler with ultra low NOx emission Quinta Range Condensing Boilers REMEHA QUINTA RANGE High efficiency wall hung condensing boiler with ultra low NOx emission QUINTA RANGE OUTPUTS Q30: 5.6-28 kw Q45: 7-40 kw Q65: 11-61 kw Q85: 14.4-84.2

More information

Maxi Pressurisation Units (Maxi, Maxi Plus & HP)

Maxi Pressurisation Units (Maxi, Maxi Plus & HP) Maxi Pressurisation Units (Maxi, Maxi Plus & HP) Installation and Maintenance Instructions Contents 1 Dimensions 3 1.1 Installation 4 1.1 Mains/boosted cold water inlet 4 1.2 Connection to the system 5

More information

GREISINGER electronic GmbH

GREISINGER electronic GmbH T38.0.0X.6C-03 CO2 - Transmitter Operating Manual GT10 - CO2-1R GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 Tel.: +49 9402 / 9383-0, Fax: +49 9402 / 9383-33, email: info@greisinger.de

More information

CORONA E / FLYPPER ELECTRONIC METER MULTI-JET

CORONA E / FLYPPER ELECTRONIC METER MULTI-JET APPLICATION Fully electronic compact water meter with impeller scanning for recording volume. Highly accurate recording of all billing data at medium temperatures up to 90 C. FEATURES 4 Electronic sensor

More information

PORTABLE AIR CONDITIONER

PORTABLE AIR CONDITIONER VERY IMPORTANT! MODEL: GDC-AC9RW / GDC-AC9RCS GDC-AC9RCW / GDC-AC12RW GDC-AC12RB / GDC-AC12RCB GDC-AC12RCW INSTRUCTIONS FOR USE PORTABLE AIR CONDITIONER Do not install and use your portable air conditioner

More information

FLOWMETER INSTALLATION INSTRUCTIONS

FLOWMETER INSTALLATION INSTRUCTIONS SAFETY FIRST: Follow ALL safety precautions when working on motor equipment-wear safety glasses! Fuel vapors and liquid are extremely flammable and explosive, personal injury and/or property damage could

More information

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510

User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510 User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Signature and ISX CM870 Fuel System

Signature and ISX CM870 Fuel System Signature and ISX CM870 Fuel System Cummins Ontario Training Center HPI-TP Fuel System Heavy Duty High Pressure Injection - Time Pressure Fuel System The fuel system developed for the Signature and ISX

More information

Coolflo Safe Touch Shower Mixing Valve ST20015CP

Coolflo Safe Touch Shower Mixing Valve ST20015CP Coolflo Safe Touch Shower Mixing Valve ST20015CP Installation and Maintenance Instructions In this procedure document we have endeavoured to make the information as accurate as possible. We cannot accept

More information

Ambiente Quick Reference Manual

Ambiente Quick Reference Manual Ambiente Quick Reference Manual Schaerer USA Corporation 2900 Orange Avenue, Suite 102 Signal Hill, CA 90755 Phone (888) 989-3004 Fax (562) 989-3075 TABLE OF CONTENTS Machine Overview Page 1 Button Panel

More information

Section 7 Adjustment, Repair and Replacement

Section 7 Adjustment, Repair and Replacement Section 7 Adjustment, Repair, and Replacement Page 7-1 Section 7 Adjustment, Repair and Replacement Section Contents Page Overview...7-5 Belts Belt Guard Removal/Installation...7-7 Belt Removal/Installation...7-11

More information

Specifications - Installation and Operating Instructions Ø5 [127.00] 17/64 [6.91] 1-59/64 [48.89]

Specifications - Installation and Operating Instructions Ø5 [127.00] 17/64 [6.91] 1-59/64 [48.89] Bulletin P-DM-1100 Series DM-1000 DigiMag Digital Differential Pressure Gage Specifications - Installation and Operating Instructions PRESSURE CONNECTIONS 21/64 [8.59] LONG Ø5 [127.00] 4-31/64 [113.89]

More information

Thermistor Thermometer MODEL NO. 93210-00

Thermistor Thermometer MODEL NO. 93210-00 Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:

More information

G100 Range. Operating Manual

G100 Range. Operating Manual Operating Manual G100 CO2 0-20% G110 CO2 0-100% G150 CO2 0-10,000ppm Geotechnical Instruments (UK) Ltd Sovereign House Queensway Leamington Spa Warwickshire CV31 3JR England Tel: +44 (0)1926 338111 Fax:

More information

Valve systems Base plates, according to standard ISO 5599-1. Brochure

Valve systems Base plates, according to standard ISO 5599-1. Brochure ISO 5599-1 Brochure 2 ISO 5599-1 Single subbase Single subbase, thread connections on the side standard: ISO 5599-1 Frame size: ISO 1 - ISO 4 compressed air connection output: G 1/4 - G 1 Reversed pressure

More information

SHOWER WATER HEATER MODEL BS 35 / 45 / 60 BS 35 E / 45 E / 60 E OPERATION AND INSTALLATION INSTRUCTIONS

SHOWER WATER HEATER MODEL BS 35 / 45 / 60 BS 35 E / 45 E / 60 E OPERATION AND INSTALLATION INSTRUCTIONS SHOWER WATER HEATER MODEL BS 35 / 45 / 60 BS 35 E / 45 E / 60 E OPERATION AND INSTALLATION INSTRUCTIONS 2 This water heater must be installed (water and electrical installation), commissioned and serviced

More information

Kane Automotive. Gas Analyser Manual for models; Auto 2-2; 4-2 & 5-2; Auto 4-1 & 5-1

Kane Automotive. Gas Analyser Manual for models; Auto 2-2; 4-2 & 5-2; Auto 4-1 & 5-1 Kane Automotive Gas Analyser Manual for models; Auto 2-2; 4-2 & 5-2; Auto 4-1 & 5-1 Description The Kane Automotive range of emission analysers covering the models Auto 2-2, 4-2, 5-2, 4-1 & 5-1 has been

More information

1. SAFETY INFORMATION

1. SAFETY INFORMATION RS-232 Sound Level Meter 72-860A INSTRUCTION MANUAL www.tenma.com 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter

More information

NAVIEN AMERICA INC. 17855 FITCH AVENUE

NAVIEN AMERICA INC. 17855 FITCH AVENUE Navien America Inc. Specification Details for the Navien models: CC-240 CC-240A For Potable Water Heating and Space Heating NAVIEN AMERICA INC. 17855 FITCH AVENUE IRVINE, CALIFORNIA, USA 92614 Toll Free:

More information

Druck DPI 142/150. GE Infrastructure Sensing. Pressure Indicator Calibration manual - K382

Druck DPI 142/150. GE Infrastructure Sensing. Pressure Indicator Calibration manual - K382 GE Infrastructure Sensing 99 Washington Street Melrose, MA 02176 Fax 781-665-0780 TestEquipmentDepot.com Druck DPI 142/150 Pressure Indicator Calibration manual - K382 Introduction This technical manual

More information

PARTS NEEDED FOR CONVERTING FROM LP TO NG

PARTS NEEDED FOR CONVERTING FROM LP TO NG 1 6 7 8 3 5 4 10 9 2 PARTS NEEDED FOR CONVERTING FROM LP TO NG Number Description 1 30" NG Manifold 2 NG side burner valve 3 NG Regulator 4 Rear I/R NG orifice 4 Rear I/R NG orifice 5 Side burner NG orifice

More information

Please keep this manual for f uture reference.

Please keep this manual for f uture reference. GAS WATER/AGUA Congratulations! You've just purchased a new Marey Portable tankless water heater and will soon begin to enjoy the benefits of going tankless. Take the time to thoroughly read and understand

More information

Car and commercial cial air conditioning service solutions

Car and commercial cial air conditioning service solutions Car and commercial cial air conditioning service solutions THE STANDARD IN INFORMATION AND DIAGNOSTICS SYSTEMS INTRODUCTION For the KoolKare range of vehicle air conditioning service equipment we went

More information

TGF11 Tracer Gas Filler

TGF11 Tracer Gas Filler Operating Instructions Type no. TGF.305.115 TGF11 Tracer Gas Filler Order No.: 590-558 (TGF11) 590-559 (TGF11 LP) Software version: 1.06 Reprinting, translation, and duplication require written confirmation

More information

ADU 5. Automatic Distillation Unit

ADU 5. Automatic Distillation Unit ADU 5 Automatic Distillation Unit Atmospheric Distillation Distillation tests are used to characterize the volatility of petrochemical products. These volatility characteristics are a critical measurement

More information

Quik Lining Systems Heater Owners Guide

Quik Lining Systems Heater Owners Guide Quik Lining Systems Heater Owners Guide Quik Lining Systems, Inc is pleased to present the Quik Heater, an innovative heater used to cure cured in place pipe for sewer laterals. The unit is the most cost

More information

BRAZED PLATE INSTALLATION MANUAL

BRAZED PLATE INSTALLATION MANUAL www.brazedplate.com BRAZED PLATE INSTALLATION MANUAL S.E.C. Heat Exchangers P.E.I. CANADA C0A 1A0 tel; 902-659-2424 fax; 902-659-2800 Table of Contents Description...2 Mounting position...3 Piping connections......3

More information

Signature and ISX CM870 Electronics

Signature and ISX CM870 Electronics Signature and ISX CM870 Electronics Cummins West Training Center System Description General Information The Signature and ISX CM870 engine control system is an electronically operated fuel control system

More information

Your safety and the safety of others are very important.

Your safety and the safety of others are very important. NATURAL GAS TO PROPANE CONVERSION KIT 090 INSTALLATION INSTRUCTIONS FOR ALTITUDES 0 -,00 FT. ONLY PROPANE CONVERSION KIT SAFETY... INSTALLATION REQUIREMENTS... Tools and Parts... LP Gas Requirements...

More information

Streaming Potential System (SPT1000, SPD1000)

Streaming Potential System (SPT1000, SPD1000) Streaming Potential System (SPT1000, SPD1000) User's Manual Sentrol Systems, Inc. 3949 Cotswold Dr. SW, Lilburn, GA 30047-2371 Tel: 770-564-1541, Fax: 770-564-8605 www.sentrolsystems.com Since the streaming

More information

2/8.10.01. These vacuum switches are used in measuring ranges between -1 and 0 bar.

2/8.10.01. These vacuum switches are used in measuring ranges between -1 and 0 bar. CATALOGUE > Release 8.5 > Electronic vacuum switches Series SWM Electronic miniature vacuum switches Series SWM These vacuum switches are used in measuring ranges between -1 and 0 bar. Small dimension

More information

Oil - kerosene burners

Oil - kerosene burners Installation, use and maintenance instructions Oil - kerosene burners One stage operation CODE MODEL TYPE 3747469 G0 474T5 90 (0) - 05/008 TECHNICAL FEATURES Thermal power output 95 3 kw 8 8 kg/h Fuel

More information

CARING FOR YOUR WATER HEATER

CARING FOR YOUR WATER HEATER http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html Water Heater Inspections CARING FOR YOUR WATER HEATER Venting System (Direct Vent Only) The venting system should be inspected

More information

Integrating Sound Level Meter and Datalogger

Integrating Sound Level Meter and Datalogger USER GUIDE Integrating Sound Level Meter and Datalogger Model 407780A Introduction Thank you for selecting the Extech Instruments Model 407780A. This device is shipped fully tested and calibrated and,

More information

The testo 340 What makes it The Ultimate Combustion Tuner?

The testo 340 What makes it The Ultimate Combustion Tuner? The testo 340 What makes it The Ultimate Combustion Tuner? Zeichen setzen für die Zukunft Design 2/27 Specification 4 Gas Analyzer- CO,NO,NO2 or SO2 Compact Tough Plug n Play Sensor w/ NEW Replicable Interference

More information

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1

More information

DUAL SENSING DIGITAL THERMOSTAT PRODUCT INSTRUCTIONS. Construction Automotive Industry

DUAL SENSING DIGITAL THERMOSTAT PRODUCT INSTRUCTIONS.  Construction Automotive Industry DUAL SENSING DIGITAL THERMOSTAT PRODUCT INSTRUCTIONS www.rehau.com Construction Automotive Industry SCOPE This guide gives instruction regarding REHAU Programmable Digital Thermostat installation and operation.

More information

Geyser-R Heat pump water heater Installation manual

Geyser-R Heat pump water heater Installation manual Geyser-R Heat pump water heater Installation manual 1 Version 2.9 SAFETY INFORMATION Please read carefully to prevent serious accidents or injury. The Geyser-R Heat Pump Water Heater must be installed

More information

36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve

36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION

More information

CDS TROUBLESHOOTING SECTION I. VACUUM. 1.0. Weak vacuum at wand. Gauge reads normal (10hg to 14hg)

CDS TROUBLESHOOTING SECTION I. VACUUM. 1.0. Weak vacuum at wand. Gauge reads normal (10hg to 14hg) CDS TROUBLESHOOTING SECTION I. VACUUM 1.0. Weak vacuum at wand. Gauge reads normal (10hg to 14hg) 1.1. Clogged hoses or wand tube. Disconnect hoses and carefully check for an obstruction. 1.2. Excessive

More information

United Kingdom en. ErP information. Gas fired condensing boiler Gas 210 ECO PRO

United Kingdom en. ErP information. Gas fired condensing boiler Gas 210 ECO PRO United Kingdom en ErP information Gas fired condensing boiler Gas 210 ECO PRO Dear customer, Thank you for purchasing this appliance. Please read this manual carefully before using the product and keep

More information

VDL 53 - Vacuum drying oven for flammable solvents

VDL 53 - Vacuum drying oven for flammable solvents Vacuum drying ovens VDL 53 - Vacuum drying oven for flammable solvents The VDL series ensures maximum safety when drying organic solvents in accordance with TÜV/GS guidelines. The interior chamber is designed

More information

Reduced Pressure Principle Assembly 3 Valve Procedure

Reduced Pressure Principle Assembly 3 Valve Procedure Reduced Pressure Principle Assembly 3 Valve Procedure A. Initial Setup 1. Flush test cocks by: a. opening test cock #4 to establish flow through the unit. (Leave this test cock open until others are flushed)

More information

BW-500 BED WEIGHER USER MANUAL

BW-500 BED WEIGHER USER MANUAL BW-500 BED WEIGHER USER MANUAL 0 1 TABLE OF CONTENTS SPECIFICATION... 3 Equipment... 4 INSTALLING WEIGH BEAM... 5 PREPARATION FOR MEASUREMENT... 6 DP3800 Indicator... 8 POWER SUPPLY... 8 PANEL... 9 KEY

More information

MP-4000 Alarm List (Software version 2.4.3 or later)

MP-4000 Alarm List (Software version 2.4.3 or later) Service Bulletin SUBJECT: MP4000 Alarm s BULLETIN: C 100 DATE: June 19, 2013 ALARM LIST Where it is possible the alarm number is kept the same as for MP-3000. MP-3000 holds alarm number from 0 to 127.

More information