Calibration Verification



Similar documents
How to Verify Performance Specifications

Laboratory Director Responsibilities

2010 Laboratory Accreditation Program Audioconference. Accreditation Requirements for Waived Vs. Non-waived Tests

Document Review: Centers for Medicare and Medical Service (CMS) What Do I Need to Do to Assess Personnel Competency?

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

Prepublication Requirements

Individualized Quality Control Plan (IQCP) Frequently Asked Questions Date: May 5, 2015 (last updated 05/13/2016)

CAP Point of Care Checklist Frequently Asked Questions Roger D. Klein, MD Sheldon Campbell, MD, PhD Peter J. Howanitz, MD

Individualized Quality Control Plan

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate

Flow Cytometry. CLIA Compliance Manual*

Clinical Laboratory Improvement Amendments (CLIA) How to Obtain a CLIA Certificate of Waiver

TEST METHOD VERIFICATION AND VALIDATION

Procedure for Equipment Calibration and Maintenance

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Incident Management: Developing a Plan

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

General. What is a laboratory test? What is a waived test?

COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.

ASSURING THE QUALITY OF TEST RESULTS

A Roadmap for Competency Assessment ASCLS-NJ Spring Seminar April 24, 2014

Examples: collecting the right sample on the right patient; labeling the sample accurately;

Content Sheet 3-1: Equipment Management Overview

2009 LAP Audioconference Series. Laboratory Computer Systems

Laboratory Computer Systems

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

Quality Assurance/Quality Control in Acid Deposition Monitoring

IQCP GUIDELINES and TEMPLATE FOR GETTING STARTED

LabGuide 71. Incident Management: Developing a Plan

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Is Your Lab s Competency Assessment a Competent Assessment?

Competency Assessment (both of employees and of the laboratory as a whole)

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

PROFICIENCY TESTING. Clinical Laboratory Improvement Amendments (CLIA) DOs and DON Ts. Brochure # 8

Validation and Calibration. Definitions and Terminology

Basic Lessons in Laboratory Quality Control

Center for Clinical Standards and Quality/Survey & Certification Group

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

Answers to Your Questions About Blood Bank Proficiency, Competency and QC Follow Up to April 28, 2016 Webinar

Best Practices for Maintaining Quality in Molecular Diagnostics Gyorgy Abel, MD, PhD

CHAPTER 7 QUALITY ASSESSMENT

LAP Audioconference How to Prepare and Comply with Your Quality Management Plan* February 18, 2009

Procedure: Quality Assurance Policy Version 5 Quality Assessment Policy Version 3

Level 1 Hematology Training Module For AcT Diff and AcT Diff 2

PROCEDURES FOR HANDLING OOS RESULTS

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Standards for Laboratory Accreditation

rtech laboratories began working toward ISO 9002:1994 in ISO 17025: Practical Benefits of Implementing a Quality System STATISTICAL ANALYSIS


CLIA & Individualized Quality Control Plan (IQCP)

Subpart H--Participation in Proficiency Testing for Laboratories Performing Nonwaived Testing

Meaningful Competency Assessment. Kathy Nucifora, MPH, MT(ASCP) COLA 1

2009 LAP Audioconference Series. How to Prepare and Comply with Your Quality Management Plan

EZValidation Online Tool. The easy way to apply guidelines. The EZValidation Online Tool. The implementation of new tests

LABORATORY SUPERVISOR JOB CODE: 2701 DATE: 11/8/95

Medical Cannabis Laboratory Approval Program

Document and Record Control in a Hospital Clinical Laboratory

Bio-Rad Laboratories. QC data management solutions. Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

How To Get Blood Test Records From A Blood Alcohol Test

1 Quality Assurance and Quality Control Project Plan

CLIA & Your Laboratory A Guide for Physicians and Their Staff

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

College of American Pathologists Accreditation for Public Health Laboratories

Preparation "Clinical Laboratory Technologist and Technician Overview"

Building Commissioning

NC SBI QUALITY ASSURANCE PROGRAM

Medical Laboratory Technology Program. Student Learning Outcomes & Course Descriptions with Learning Objectives

Best Practice POS Cash Collections. Ryan Wheeler, President and GM Early Out and Patient Finance Division

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Equipment Qualification

Ecology Quality Assurance Glossary

LAB 37 Edition 3 June 2013

Measurement Systems Correlation MSC for Suppliers

2013 Laboratory Accreditation Overview Guide

XL-I 2011: Understanding and Explaining the CAP POCT Checklist: A Tool for Laboratorians to Use With POCT Staff

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

MEDICAL LABORATORY TECHNICIAN COMPETENCY PROFILE

Welcome to Hibbing Community College s Online MLT Brochure!

Control Charts and Trend Analysis for ISO Speakers: New York State Food Laboratory s Quality Assurance Team

A.1 Sensor Calibration Considerations

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

In vitro diagnostic reagent, calibrator and control material stability

QUALITY ASSURANCE PROGRAM (QAP) City Rohnert Park

Instrument/Method Validation. Lebah Lugalia Lab QA/QC Coordinator UNC Project Lilongwe, Malawi

Software Verification and Validation

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Assay Qualification Template for Host Cell Protein ELISA

EDUCATIONAL COMMENTARY RED BLOOD CELLS AND WHITE BLOOD CELLS IN URINALYSIS

Enhanced calibration High quality services from your global instrumentation partner

Health Care Job Information Sheet #13. Laboratory

XV.Quality Assurance in the Blood Bank

Monitoring the Quality and Performance of Analytical Process Testing

Transcription:

Calibration Verification INTRODUCTION On January 24, 2003, the Centers for Medicare and Medicaid Services (CMS) published updated laboratory regulations that became effective April 24, 2003. Included in the updated CLIA requirements are revised standards for calibration verification of nonwaived (moderate and high complexity) test systems. For analytes that require calibration, laboratories must continue to perform calibration as specified in the manufacturer s instructions. The updated CLIA regulations now specify that laboratories must also perform calibration verification at least every six months, using at least three levels of materials that are within the reportable range of the test. COLA has developed new accreditation criteria to ensure COLA laboratories comply with the updated CLIA requirements for calibration verification. This LabGuide begins by defining calibration and calibration verification, and explaining the difference between the two related processes. Detailed information is provided about calibration verification to help you understand the process and meet the requirements of CLIA and COLA. WHAT IS CALIBRATION? Calibration is the process of testing and adjusting the instrument or test system readout in order to establish a correlation between the instrument s measurement of the substance being tested (the analyte) and the actual concentration of the substance. In other words, calibration sets the instrument to give an accurate result for each analyte. When is Calibration Required? The laboratory is responsible for performing calibration for each analyte at the frequency specified by the manufacturer of the test and when calibration verification does not give acceptable results. If calibration for your test system is less stable than the manufacturer s recommended frequency, then more frequent calibration or additional calibration materials may be required, as determined by your laboratory. Are There Exceptions? Calibration is not required for: Manual procedures that do not involve an instrument Microscopic procedures These tests use visual interpretations rather than instrument measurements, so there is nothing to calibrate. Examples of manual procedures: Microbiology cultures Tilt-tube prothrombin times ABO group and D (Rho) typing Examples of microscopic procedures: Urine sediment examination KOH and pinworm preparations Manual WBC differentials Other tests that do not require calibration are: Most prothrombin time devices Point-of-care or unit-use devices that are factory calibrated and do not permit user calibration, or calibration is performed internally by the instrument. Nonwaived devices of these types are required to have calibration verification performed What are Calibration Materials? The test system s instructions describe the process for performing calibration, as well as the number, type, and concentration of calibration material to use for each analyte. Calibration materials (calibrators) are solutions that contain a known amount of the substance (analyte) being measured. Some calibrators contain, and can be used to calibrate, more than one analyte. Previously, the term standards was used when referring to calibration materials. 15-1

Summary of Calibration Calibration: Sets the instrument to give an accurate result for the analyte being tested Is performed: o At the frequency specified by the manufacturer of the test, or more frequently if determined necessary by the laboratory o Using the number, type, and concentration of calibration materials specified by the manufacturer o When calibration verification does not give acceptable results LINEARITY Before the final CLIA regulations were published in 2003, laboratories were required to perform linearity studies every six months for quantitative high complexity tests. These studies involved: Replicate testing of five samples spanning the reportable range Determining a mean for each sample Plotting the points to produce a linear curve Linearity studies, as such, are no longer required by CLIA, but they are considered good laboratory practice. Calibration verification has basically replaced the term (and process) of linearity studies. REPORTABLE RANGE You will see the term reportable range used in this LabGuide. Reportable range is the range of results for an analyte, from minimum to maximum, that the test system can accurately measure. Determining reportable range and the other requirements for verification of performance specifications is covered in detail in another LabGuide. WHAT IS CALIBRATION VERIFICATION? Calibration verification means testing materials with a known concentration of the analyte in the same manner as patient specimens. At least three samples spanning the reportable range of the test are tested as unknowns (in normal test mode, not calibration mode) and the values obtained are compared to the known value of the materials. If the values obtained are acceptable, calibration of the analyte is verified. If not, a new calibration may be required. In other words, calibration verification checks the analyte s calibration to verify if the calibration is still valid. A successful calibration verification confirms that the test system is providing accurate results for the analyte throughout the reportable range of the test. When is Calibration Verification Required? The laboratory is responsible for performing calibration verification, for each analyte measured by the test system, every six months, or more frequently if: Specified in the manufacturer s test system instructions The lab has determined (when establishing or verifying performance specifications) that the calibration for this test system or analyte should be checked more frequently The laboratory must also perform calibration verification when: There is a change in the reagent lot number, reagent formulation, or reagent manufacturer used for the test Note: A lab is not required to perform calibration verification after lot number changes if they have documented several instances where the calibration verification was acceptable after a lot number change There is major preventive maintenance or replacement of critical parts Control materials do not perform as expected Are There Exceptions? The requirement for calibration verification is automatically met if the test system s calibration procedure for the analyte uses three or more levels of calibration material and: Includes a low, mid-point, and high value Is performed at least once every six months If you perform an acceptable calibration at least every six months for an analyte, then you don t also need to do calibration verification. Calibration at least every six months satisfies the requirement for calibration verification. The calibration verification requirements for automated hematology cell counters (CBC instruments) are met if the laboratory: Follows the manufacturer s instructions for instrument operation 15-2

Tests at least two levels of control materials each day of testing Obtains acceptable results for the daily controls What Is Used to Perform Calibration Verification? To perform calibration verification, test materials of known concentration in the same manner as patient specimens. What are suitable materials of known concentration? Calibration Verification Materials There are various materials with known concentrations that can be used for calibration verification: Commercially available calibration materials or linearity sets o Use the acceptable limits established by the manufacturer Proficiency testing samples with known results o Use the CLIA-defined acceptable limits in the PT summary Control materials with known results (you must use a different lot number for calibration verification than you use for your routine quality control) o Use the acceptable limits established by the manufacturer Patient specimens with known results o Establish your own acceptable limits, within reason (PT limits are a good guide) Since calibration verification is used to check if the test system is providing accurate results across the reportable range of the test, it is important to use calibration verification materials that include at least: A low value at or near the lower limit of the instrument s reportable range for the analyte A mid-point value for the analyte A high value at or near the upper limit of the instrument s reportable range for the analyte Many commercially available calibration verificationmaterials contain more than one analyte. You may need to use more than one set of calibration verification materials in order to perform calibration verification for all of the analytes on all of the test systems used in your laboratory. How Do You Perform Calibration Verification? To perform calibration verification: 1. Document the date, the analyte, and the calibration verification materials used. 2. Run at least three levels of materials, in the same manner and test mode that patient specimens are tested, and record the results. One repetition for each level is adequate, but additional repetitions will supply more data that the laboratory may find useful. 3. Compare the results obtained to the acceptable limits. 4. Document the results and indicate if the results are acceptable. Calibration Verification Example The following is an example calibration verification to help demonstrate how to evaluate the data. Analyte: Total Cholesterol Instrument: ChemXYZ Calibration Verification Material Used: XYZ Brand Commercial Calibrators Source of Acceptable limits: Manufacturer limits The low value sample result falls within the range of acceptable limits, so this point is acceptable. The midpoint value sample result is within the limits, so it is also acceptable. The high value sample result is below the acceptable limit, so this point fails. This data means the calibration verification for this analyte failed. The instrument calibration for this analyte may no longer be valid and corrective action is needed. Known Value of Material Low 20 Mid 150 High 300 Acceptable limits for material 18-22 135-165 270-330 Result obtained 20 152 269 Results acceptable? Yes Yes No Comments May need to recalibrate. Use troubleshooting checklist first. 15-3

> What if Calibration Verification Fails? If your calibration verification fails (results for one or more levels of material are not acceptable) for any analyte, you may need to recalibrate the instrument for that analyte. A troubleshooting checklist is provided at the end of this LabGuide to help uncover potential problems that require corrective action. Proceed through the troubleshooting checklist and attach the completed form to your calibration verification worksheet. Document all activities in your investigation and all corrective actions taken. If the test system is factory-calibrated by the manufacturer and cannot be recalibrated by the laboratory, then contact the manufacturer for advice if calibration verification fails. Corrective Action The troubleshooting checklist may uncover a problem that you can correct (replacing expired material, for example) and then simply repeat the calibration verification. It may also uncover a problem that indicates a new calibration (recalibration) is necessary. Document all activities in your investigation and all corrective actions taken. Summary The final CLIA regulations (published in 2003) specify requirements for calibration verification of nonwaived testing, and the new COLA criteria for calibration verification are derived from these requirements. Calibration verification checks the test system calibration at a point in time to determine if it is still providing accurate results for the analyte throughout the reportable range Perform calibration verification for each analyte at least once every six months, or whenever certain situations occur Perform calibration verification using materials of known concentration Properly document all calibration verifications and any corrective actions taken Retain records of all calibration verification activities for at least two years. REFERENCES CLIA Regulations, Subpart K, 493.1255. CMS Brochure #3 Calibration and Calibration Verification. COLA Criteria for Quality Laboratory Performance, Evaluation Grouping: Calibration. When the troubleshooting checklist does not uncover the cause of the problem, you will also want to take corrective action by recalibrating the instrument for that analyte, following the manufacturer s instructions for calibration. Document the calibration and indicate that it was performed in response to unsuccessful calibration verification. Always run controls after a calibration and document the results. If the instrument still does not perform as expected, call the instrument manufacturer for further troubleshooting advice. 15-4

Worksheet and Documentation Form for Calibration Verification Analyte Date Instrument Serial Number Reagent/Strip/Cassette Lot# Expires: Materials used Source of Acceptable Limits Lot Number Low Level Mid Level High Level Expiration Date Expected Value Acceptable Limits Calibration Verification Results Low Level Mid Level High Level Results Obtained Repetition 1 Optional: Repetition 2 Repetition 3 Mean Results Acceptable? Comments and/or Corrective Actions Performed by Reviewed by Date 15-5

Troubleshooting Checklist for Calibration Verification Analyte Date Instrument Serial Number 1. Ch eck your quality control (QC) results for th e analyte Are there any patterns seen in the control results? Are all values below the mean? Are all values above the mean? Are there any noticeable shifts or trends over time? Are accuracy and precision acceptable? 2. Ch eck your calibrati on verifi cation mat eri al Are the materials used appropriate and in-date? Have you properly determined the acceptable limits for the calibration verification material? 3. Ch eck your reag ents Have there been any reagent changes? Is there a new lot number of reagent? Has there been a change in manufacturer? Has there been a new formulation (check the package insert) of current reagent? Are the reagents in date? 4. Ch eck inst rum ent m aintenance Review the daily, weekly, monthly, quarterly, etc. logs. Is there any missing maintenance, problems, or changes? 5. Ch eck the environment Has the instrument been moved recently? Have there been any changes to the environment or surroundings of the instrument? 6. Ch eck the servi ce record Has the instrument been serviced recently? Has there been any software or hardware upgrades or changes? 7. Ch eck inst rum ent op erati on Are there new instrument operators? Are all operators following established procedures for instrument operation? Has there been any recent modification to the technique used to run the test? 8. Ch eck a comp arative meth od Is there another laboratory nearby that can run your calibration verification material so you can compare the results? 9. Will recalibrati on b e p erform ed for the analyt e? 15-6