SPECIALIZED GAMMA-RAY FLAW DETECTOR «DAR - Se75 FN R» WITH. RADIATION SOURCE Se-75



Similar documents
A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials

SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES

Cutting Edge Digital Radiography Technology in Daily NDT Use

SAFE MANAGEMENT OF SMOKE DETECTORS CONTAINING RADIOACTIVE SOURCES

PHYA5/1. General Certificate of Education Advanced Level Examination June Unit 5 Nuclear and Thermal Physics Section A

DEMONSTRATION ACCELERATOR DRIVEN COMPLEX FOR EFFECTIVE INCINERATION OF 99 Tc AND 129 I

Radiation Monitoring System DLMon

Radiation and the Universe Higher Exam revision questions and answers

W. C. Reinig. Savannah River Laboratory E. I. du Pent de Nemours and Company Aiken, South Carolina 298o1

For convenience, we may consider an atom in two parts: the nucleus and the electrons.

Eddy-current testing - Non-destructive testing for flaw detection of metals, coatings and carbon fibres

Basics of Nuclear Physics and Fission

Application Note. ipix A Gamma imager to support various applications

VAD. Variable Area Desuperheaters

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

MASS DEFECT AND BINDING ENERGY

ADVANCED NDT TECHNIQUES FOR PLASTIC PIPELINE INSPECTION

Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Basic Nuclear Concepts

PHYSICS PAPER 1 (THEORY)

[2] At the time of purchase of a Strontium-90 source, the activity is Bq.

RADON Although radon is agas, its decay products are not, and they occur either as unattached

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Intelligent Measurement and Diagnostic Techniques for Non-destructive Inspections

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

MATERIALS LICENSE. 1. Honeywell International, Inc. 3. License Number: SUB-526, Amendment 6-1

VAD Variable Area Desuperheaters

MATERIALS LICENSE. In accordance with the letter dated January 15, 2009,

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

Radiation Strip Thickness Measurement Systems

Main properties of atoms and nucleus

Nuclear accidents and radioactive contamination of foods 30 March 2011

The TVEL Fuel Company of Rosatom CREATING FUTURE TODAY

Nuclear Energy: Nuclear Energy

PART H RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT. Definitions. As used in this Part, the following definitions apply:

proper way. Therefore operator training and certification is key in every inspection program to ensure compliance and improve quality and integrity.

Optimization of Digital Industrial Radiography (DIR) Techniques for Specific Applications: An IAEA Coordinated Research Project

SECTION TWO PACKAGING, TRANSPORTATION AND STORAGE OF RADIOACTIVE MATERIALS

Technical Meeting on the Implications of the Fukushima Daiichi Accident on the Safety of Fuel Cycle Facilities. IAEA Headquarters Vienna, Austria

Cyber Security Design Methodology for Nuclear Power Control & Protection Systems. By Majed Al Breiki Senior Instrumentation & Control Manager (ENEC)

INNOVATIVE ELECTROMAGNETIC SENSORS

Objectives 404 CHAPTER 9 RADIATION

ILLUSTRATIVE EXAMPLE: Given: A = 3 and B = 4 if we now want the value of C=? C = = = 25 or 2

XRS-3 X-RAY SOURCE OPERATOR S MANUAL

Atomic and Nuclear Physics Laboratory (Physics 4780)

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Qualification of In-service Inspections of NPP Primary Circuit Components

Cambridge International Examinations Cambridge International General Certificate of Secondary Education

Using the Spectrophotometer

Zebra Technologies. Table of Contents. Healthcare Catalog. Lab. Pharmacy ISBT 128. Specimen Collection. Multipurpose. Specialty. Wristbanding.

READING DEVICE US-05

COMPENSATION MIT Flow Cytometry Core Facility

Training: NDT Access Conditions

A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS

NON DESTRUCTIVE TESTING & ASNT WELD INSPECTION & AWS

PROPOSALS FOR UNIVERSITY REACTORS OF A NEW GENERATION

BIS Metal Pipe Clamps with Lining

Accidents of loss of flow for the ETTR-2 reactor: deterministic analysis

PLATE HEAT EXCHANGER. Installation Manual. Customer Name: Serial number: Purchase order number: Project:

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

UX-30 PERFORMANCE TESTING SUMMARY REPORT Revision 0

Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N

Homework #10 (749508)

P O L Y T E C R.M. S.R.L.

Characteristic curves of a solar cell

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

HEAVY WATER LEAK DETECTOR

Syringe Calibration Factors and Volume Correction Factors for the NPL Secondary Standard Radionuclide Calibrator.

Sampling Solutions. Fuel and Lube Oil sampling Equipment. Convenient, Representative, Approved.

Regulatory Guide. Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

Radiation Protection s Benefits in the Production of Iodine-131

Report on heat production during preliminary tests on the Rossi Ni-H reactor.

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor

Voltage nominal maximum for test. Inverse voltage. nominal maximum for test. Focal spots (IEC 60336:2005) small large

Rosatom Standard Reference Data System

Industrial tracers - Finding leaks and blockages

Masses in Atomic Units

ABSORPTION OF BETA AND GAMMA RADIATION

Confocal μ-xrf for 3D analysis of elements distribution in hot environmental particles

RC-17. Alejandro V. Nader National Regulatory Authority Montevideo - Uruguay

IAEA Research Reactor Operations & Maintenance Support 2014 TRTR Meeting August 3rd-7th, 2014 Benson Hotel Portland Oregon

Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)

mgr inż. Rafał Kaczmarek (MSc Eng.), dr inż. Ryszard Krawczyk (PhD (DSc) Eng.) Częstochowa University of Technology, Welding Technology Department

XR200 X-RAY SOURCE OPERATOR S MANUAL

THERMA-TUBE preinsulated pipes and fittings Catalogue of products

Ultrasonic Technique and Device for Residual Stress Measurement

Light as a Wave. The Nature of Light. EM Radiation Spectrum. EM Radiation Spectrum. Electromagnetic Radiation

RADIATION SAFETY STANDARD

NON-DESTRUCTIVE TESTING OF PIPELINES. L. Annila. Abstract

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

UniPlate. Solar Collectors SOLAR SYSTEMS.

Transcription:

SPECIALIZED GAMMA-RAY FLAW DETECTOR «DAR - Se75 FN R» WITH RADIATION SOURCE Se-75 Khoroshev V.N.; Volchkov J.E.; Dekopov A.S.; Kositsyn E.M., Fedotov V.I. «Open Joint Stock Company Scientific Research Institute of Physics and Tehncheskoy avtomatitsii»; Zlobin N.N. «Machine-Building Plant name Ordzhenikidze Podolsk» 10 - th European NDT Conference (7-11 June 2010, Moscow) This work was carried out under the Program for the production of heat-exchange equipment for the FN800 reactor under construction and the full realization of the requirements of the Federal Rules "Equipment and pipelines of nuclear power plants. Welded joints and surfacing. Control Rules RRNEG - 7-010 - 89 "in relation to radiographic examination of welded joints" pipe - pipe board "heat exchangers of the reactor FN800. Aim was to increase the sensitivity of the method in two-fold compared with currently used equipment with a source of iridium - 192. In the process of implementation of works created gamma-defectoscope "DAR- Se75 BN R" for a panoramic raying welds, heat exchangers under serial acute focal source of selenium - 75 activity to 5Ki with the size of the active part 1,0 x 1,0 mm, manufactured by JSC "SSC Nuclear Reactors ", Dimitrovgrad. Gamma - defectoscope "DAR - Se75FN R" meets the following regulatory documents governing the requirements for safety, design and ecology in the design, manufacture and operation of gamma - flaw:. 1. GOST 23764-79 " Gamma - defectoscope. IS.

2. Sanitary rules SP2.6.1.1284-03 "Radiation Safety in radionuclide flaw". 3. OSPORB - 99 "Basic sanitary rules for radiation safety. 4. NRB - 99/2009 "Radiation Safety Standards". 5. GOST 16327-88 - "Complete packaging for the transport of radioactive substances". 6. ISO 3999-2000 "Apparatus for radiographic TU". 7. IAEA, Vienna, 1996 "Regulation on the transport of radioactive substances". 8. NP-053-04 "Safety in the Transportation of Radioactive Material Photo prototype gamma defectoscope is shown in Figure 1. Fig.1 Gamma - defectoscope "DAR - Se75 FN R" (nominated by the holder of the source). 1. Drive 2. Container 3 - channel 3. Head radiation 4. Holder

Object of control - heat exchanger is schematically shown in Figure 2. Pipe 16 х 3 (2,5) up to 100 ps Fig.2 Heat exchanger. Pipes welded in the tube plate thickness up to 100mm. Site welding is shown schematically in Figure 3. Fig.3 Node weld tube plate with the pipe 16 x 3 (2,5) heat exchanger. Development of gamma - ray flaw detectors for the panoramic weld "pipe - tube plate" heat exchangers on the basis of radionuclide iridium - 192 and ytterbium -169 was conducted in the USSR «USRIRT» All-Union Scientific Research Institute for Radiation Technology (apparatus DAR - 3), and iridium - 192 in Germany by Sauerwein (apparatus B3) 1975-1976gg. Schematic diagram of the panoramic raying the weld is shown in Figure 4.

4. 5 2 3 6 1 7 F = 4 5 сm Fig.4 Scheme of the panoramic raying welds "tube - tube plate. 1. The Head radiation 2. Compensator 3. Special cassette 4.Plastmassovy liner 5. Source gamma - radiation 6. Pipe 7. Blocks of protection..

Table 1 Sources of gamma - radiation Sources of Gamma - radiation is used to control weld pipe boards are shown in Table 1. Sources Т 1/2 Еγ ef. MtV 1 2 3* Iridium- 192 Ytterbium -169 FSUE PA «MAYA K» JSC State Scientific Center Research Institute of Atomic Reactors JSC State Scientific Center Research Institute of Atomic Reactors Selenium- 75 Ir 192 74 days Yb 169 30 days Se 75 124 days Sources,mm External capsule Active part D L d l Activit y Ku 0,40 4,0 5,0 1,0 1,0 4 5 0,08 4,0 5,0 0,5 0,5 3,0 0,200 6,0 +0,5 12,0 +0,5 1,0 1,0 5,0 Manufac turer * Acute focal sources selenium - 75 utilized in the production of high specific activity of 1200Ki g. The sensitivity of radiographic control in half (from 04 mm to 0.2 mm) compared with Ir 192 source with a possible reduction in the energy spectrum - the radiation source and the size of the focal spot. The use of ytterbium - 169 of - for the small half-life T1 / 2 in-line production is not acceptable and useful in research. Source on the basis of selenium - 75 has the following advantages compared to Ir 192: energy gamma radiation is less than 2 times, and half-life greater than that of Ir 192 in 1,6 times. That gives rise to the use of Se75 to the task.

Table 2 lists the technical characteristics of gamma defectoscope "DAR - Se75 FN R." Table 2 Characteristics of gamma defectoscope "DAR - Se75 FN R" Parameter Weight, kg Sources Signaling Locks EDR, msv / hour (mr / h) not more 1 2 3 4 5 6 7 8 Work set 1 Radiation head Remote manual drive with 5m. Protection - tungsten. Not more than 6,0 Selenium - 75 Aktiv.do 2 5Ki 0,7 x0, 7mm Active part or 1 x 1 mm. Vial 6 +0.5 x 12 +0.5 mm Color Green OFF. (storage) Red OPEN (translucence) Lock, Accession drive manual Surface 1.0 (100) From the surface At 0,05 m 0,5 (50); At 1m 20 msv (2.0 mr / h) Note Set packing Transport Type "A" 2. Drive manual remote Support Kit 1 Container transport - charge exchange (TPC) Protection - depleted uranium 2. Spare parts, tools; Supplies; Documentation 2,0 Distance control radiation head of 5 m. No more than 20 3 - channel to 15Ki sources Selenium - 75 -------- -------- -------- Labeling channels Lock,blockin g the source holders, filling of. Surface 2,0 msv / h (200mr / h) On 1 m 0,1 msv / h (10.0 mr / h) Set packing Transport Type B (U) ------ ------- ------ ------ ------- Manual, Formulyar

The experimental work on radiographic examination of welded joints "pipe - tube plate" heat exchangers FN800 reactor prototype gamma - flaw "DAR Se75FN" sources selenium -75 (OJSC SSC NIIAR ") does not yield the expected results. This is due to the fact that the source Se75 did not meet the technical requirements. Spectral analysis of radiation sources and digital image processing of the active part of the source showed that the serial Acute focal sources of selenium -75 with the size of 1 X 1 mm primary capsule of vanadium did not have a fixed size of the active part, and the use of primary titanium capsule additionally contain in the emission spectrum of the source of "hard" line of 0.8-1.0 MeV. The calculation of the protection of gamma-flaw detector was carried out on the spectrum of Se75, which has no hard lines, the appearance in the spectrum of the source lines 0,8-1 MeV led to a reduction of the protection device. Strict requirements to the parameters of the radiation source Selenium - 75 are needed to provide the required sensitivity of the control of welding pipe-pipe board FN800 reactor in limited quantities focal length (F = 4,0-5,0 cm) and require the "purity" of the emission spectrum characteristic of Selenium - 75, rather than the primary material for the capsule. Steps are being taken to address the deficiencies identified sources based on selenium - 75 and its manufacture, as part of a tool for controlling weld "pipe - tube plate" heat exchangers of the reactor FN800. We have shown that the source of Se75 production has flaws that are discovered in the setting of targets with high requirements to the sensitivity of radiographic control.