Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS



Similar documents
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

Thermo Scientific ConFlo IV Universal Interface

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Thermo Scientific SIEVE Software for Differential Expression Analysis

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

How To Test For Contamination In Large Volume Water

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Introduction. Methods. Sample Processing

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare

LIFECYCLE. Thermo Scientific. Enterprise Solutions

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

LC-MS/MS Solutions for the Pharmaceutical and CRO Industries TSQ Series -based Solutions for Drug Discovery and Development

WATERS QUANTITATIVE ANALYSIS solutions

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Overview. Purpose. Methods. Results

A Complete Solution for Method Linearity in HPLC and UHPLC

High Sensitivity MS Determination of Carryover in a New Autosampler Design

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

What is Eluent Generation?

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Appendix 5 Overview of requirements in English

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

Pesticide Analysis by Mass Spectrometry

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

LC-MS/MS for Chromatographers

AMD Analysis & Technology AG

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

Development and Validation of Methods for Chemotherapy Drugs on the New Prelude SPLC LC-MS/MS System

Thermo Scientific ExactFinder Software

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Determination of Calcium and Magnesium in Brine

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

ProteinChip Energy Absorbing Molecules (EAM)

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

Determination of PCDD/Fs in Environmental Samples using Accelerated Solvent Extraction (ASE) and GC-MS/MS

How To Control A Record System

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

EQuan MAX Getting Connected Field Service Guide

FT-NIR for Online Analysis in Polyol Production

Gas Chromatography Liner Selection Guide

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

ApplicationNOTE Introduction

[ Care and Use Manual ]

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

Amino Acid Analyzer L-8900

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Agilent 4339B/4349B High Resistance Meters

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES

Transcription:

Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific, Ottawa, ON, Canada Key Words TSQ Quantum Ultra Clinical analysis High-throughput Selectivity Overview Current high throughput clinical assays utilizing triple quadrupole mass spectrometry for the quantitation of testosterone can be further enhanced through the use of FAIMS (high-field Asymmetric waveform Ion Mobility Spectrometry) coupled to a TSQ Quantum triple quadrupole mass spectrometer. This application note describes how a FAIMS-enabled TSQ Quantum improves the performance of a testosterone assay. Introduction Testosterone is the androgenic hormone primarily responsible for normal growth and development of male reproductive organs. Although testosterone production declines naturally with age, testosterone production may be compromised by diseased or damaged organs. Women biosynthesize very low levels of testosterone, which makes quantitation extremely difficult. Estrogen replacement therapy may further reduce testosterone production, resulting in additional complications in its quantitation. In addition, endogenous interferences may prevent accurate and precise testosterone measurement. In this study, an LC-MS/MS method was used together with the selectivity offered by FAIMS to quantify testosterone in human serum. FAIMS acts to remove chemical background and endogenous interferences resulting in more accurate and precise determinations for clinical samples than LC-MS/MS alone. Experimental Conditions Chemicals and Reagents Testosterone and testosterone-d 3 (internal standard-is) were purchased from Sigma-Aldrich (St. Louis, MO). HPLC grade methanol and formic acid were acquired from Burdick and Jackson (Muskegon, MI). All chemicals were used as received. Sample preparation: Stripped human serum (Golden West Biologicals, Temecula, CA) was fortified with testosterone at the following concentrations: 2.5, 5, 1, 25, 5, 1, 25, 5, 1, and 25 pg/ml. Internal standard in 5% formic acid was added to a final concentration of 5 pg/ml. No further sample preparation was required. Sample analysis: LC-MS/MS analyses were performed on a Thermo Scientific Surveyor LC system. The method used mobile phases A (.1% formic acid in water) and B (.1% formic acid in methanol) at a flow rate of.5 ml/min. Serum samples (9 µl) were injected onto an LC- MS/MS extraction column. The analyte was back-flushed to the 2.1 5 mm, 3 µ, Thermo Scientific Hypersil GOLD analytical column. The entire LC effluent from the sample injections was directed to the Ion Max source on a Thermo Scientific TSQ Quantum Ultra. Additional gas-phase separation prior to entry of ions into the mass spectrometer was achieved by including FAIMS in the analysis. FAIMS Conditions Dispersion voltage 5 V Outer bias voltage 35 V Compensation voltage 12.5 V Temperature (inner electrode) 6 C Temperature (outer electrode) 6 C FAIMS gas composition 5% He in N 2 FAIMS gas flow rate 3.8 L/min MS Conditions Ionization mode and source Positive APCI Spray current 1. µa Vaporizer temperature 4 C Sheath gas 35 Transfer tube temperature 25 C Transfer tube offset 35 V Tube lens offset 1 V Collision energy 22 ev Scan time 5 ms Q1 Resolution.7 Da FWHM Q3 Resolution.7 Da FWHM Testosterone m/z 289.2 m/z 97.1, 19.1 Testosterone-d 3 m/z 292.2 m/z 97.1, 19.1

Results and Discussion LC-MS/MS is a highly selective technique for analyz ing drugs from biological matrices. As shown in Figure 1, samples are loaded to the extraction column via an autosampler and LC pump combination. After a short washing time, the central valve is rotated 6 to the injection position, which allows a second pump to elute the analytes from the extraction column onto the analytical column and into the FAIMS-enabled mass spectrometer. In cases where background or co-eluting interferences appear, the limiting factor is selectivity. FAIMS is a unique selectivity enhancing tool for LC- MS. With FAIMS, gas-phase ions are purified after LC analysis but before they are mass analyzed. The waveform shown in Figure 2 separates the gas-phase ions as they are transferred into the high vacuum region of the MS. The interference ions in orange are filtered out from the ion beam, while the analyte ions in blue pass into the MS. The basic experiment that establishes FAIMS conditions is shown in Figure 3. While infusing a reference standard of testosterone into the mobile phase, the CV is scanned over a specific range. The black trace appears for the transition due to testosterone, and a maximum signal for testosterone appears at CV 12.5V. To ensure that mobile phase components with the same transition will be eliminated, stop the infusion of the reference standard and repeat the CV scan over the same specified range. The red trace indicates that chemical background in the mobile phase emerges from FAIMS not at the CV for testosterone, but rather in a broad range between CV 15 and 25V. For subsequent LC-FAIMS-MS/MS assays, the CV set to 12.5V will exclude mobile phase contributions to the analysis of testosterone. Other endogenous components present in the human serum samples may also be excluded if their FAIMS behavior is different from that of testosterone. Regression analysis based on a linear model with 1/[concentration] 2 weighting was used. The average accuracy, as deviation from theo retical is less than 5% at all concentrations. The precision of the standards, as relative standard deviation (%RSD), is less than 17% at the lower limit of quantitation and less than 11% at all other concentrations. Despite the excellent performance of LC at cleaning up samples, many interferences are still present from the matrix. Other extraction techniques might remove these Load sample to extraction column Analytical Column Autosampler Extraction Column Waste FAIMS-enabled MS/MS Inject sample to analytical column for FAIMS-MS/MS analysis Analytical Column Autosampler Extraction Column Waste FAIMS-enabled MS/MS Figure 1: Schematic diagram of LC-FAIMS-MS/MS system

High field Inner electrode Source Side Mass Analyzer Side Outer electrode Figure 2: Compensation voltage set to transmit testosterone (blue ion, solid line). Testosterone and some of the interferences behave like the blue ion, while most of the interferences behave like the orange ion. 1 95 9 85 8 75 7 65 Testosterone CV = 12.5. 48.49. 77.85.9.92. 94.97.98.99 Mobile Phase Relative Abundance 6 55 5 45 4 1.2 1.5 35 3 25 1.12 2 1. 14 15 1.15.65 1.4.6.62 1.17 1.2 5.46.53.33.62 1.25 1.27.4.18.29.74.81.89. 98 1. 8 1.19 1.41 1.46 1. 12.5 15. 17.5 Compensation voltage Figure 3: Compensation voltage scan for the infusion of testosterone reference standard in black. The red overlaid trace represents the CV scan from mobile phase alone.

interferences, but much time can be spent changing extraction selectivity. A representative LC-MS/MS chromatogram for testosterone in a human serum clinical sample is shown in the upper trace of Figure 4. Multiple interferences prevent accurate integration of the analyte. Note that the peak(s) at retention time 2.7 are not due to testosterone but rather multiple interferences. If this were the true concentration of testosterone the patient would not be able to survive having such a high endogenous level. The lower trace of Figure 4 shows the IS in human serum. Multiple interferences and an elevated baseline due to chemical background make peak integration difficult. Figure 5 shows the same sample analyzed with FAIMS included in the method. LC-FAIMS-MS/MS of testosterone in the upper trace shows that many of the interferences of Figure 4 are removed. Correct peak integration for testosterone is now possible. The lower trace for the IS shows that the chemical background and interferences were eliminated. The use of FAIMS together with LC and tandem MS has improved the selectivity of the assay, resulting in a very accurate and precise method. The lower level of quantitation was improved four-fold more than the LC-MS/MS method. 5 45 4 35 3 25 2 15 1 5 6 5 Testosterone without FAIMS Testosterone-d 3 internal standard without FAIMS 2. 6 2. 7 4 3 2 1..5 1. 1.5 2. 2.5 3. 3.5 4. Time (min) Figure 4: Representative LC-MS/MS chromatogram for testosterone in a human serum clinical sample. The upper trace is testosterone (retention time 2.7 min), the lower trace is IS (testosterone-d 3 ).

3.23 7 6 5 Testosterone with FAIMS 2.75 3.26 3.29 4 3 2 1 14 2.74 12 1 Testosterone-d 3 internal standard with FAIMS 8 6 4 2..5 1. 1.5 2. 2.5 Time (min) 3. 3.5 Figure 5: Using FAIMS. Representative LC-FAIMS-MS/MS chromatogram for testosterone and the lower trace is for the internal standard. Note that the absolute signal from FAIMS analysis is lower than without FAIMS because of the removal of interferences. Conclusions A FAIMS-equipped TSQ Quantum Ultra triple quadrupole mass spectrometer provides more accurate results by eliminating the chemical noise that arises from mobile phase and sample matrix. In cases where interferences prevent accurate and precise determination, FAIMS removes interferences. The resulting chromatograms accurately represent the concentration of testosterone in the samples. The LLOQ was improved four-fold compared to the LC-MS/MS method. References 1 Simultaneous analysis of prostanoids using liquid chromatography/highfield asymmetric waveform ion mobility spectrometry/tandem mass spectrometry. Kapron et al. Rapid Commun. Mass Spectrom. 26; 2: 154-151. 2 Determination of steroids from serum and/or plasma using turbulent flow chromatography. Cohesive Technologies; Application Note 9.-35; 25.

Laboratory Solutions Backed by Worldwide Service and Support Tap our expertise throughout the life of your instrument. Thermo Scientific Services extends its support throughout our worldwide network of highly trained and certified engineers who are experts in laboratory technologies and applications. Put our team of experts to work for you in a range of disciplines from system installation, training and technical support, to complete asset management and regulatory compliance consulting. Improve your productivity and lower the cost of instrument ownership through our product support services. Maximize uptime while eliminating the uncontrollable cost of unplanned maintenance and repairs. When it s time to enhance your system, we also offer certified parts and a range of accessories and consumables suited to your application. To learn more about our products and comprehensive service offerings, visit us at www.thermo.com. In addition to these offices, Thermo Fisher Scientific maintains a network of represen - tative organizations throughout the world. Africa-Other +27 11 57 184 Australia +61 2 8844 95 Austria +43 1 333 5 34 Belgium +32 2 482 3 3 Canada +1 8 53 8447 China +86 1 8419 3588 Denmark +45 7 23 62 6 Europe-Other +43 1 333 5 34 Finland/Norway/ Sweden +46 8 556 468 France +33 1 6 92 48 Germany +49 613 48 114 India +91 22 6742 9434 Italy +39 2 95 591 Japan +81 45 453 91 Latin America +1 68 276 5659 Middle East +43 1 333 5 34 Netherlands +31 76 579 55 55 South Africa +27 11 57 184 Spain +34 914 845 965 Switzerland +41 61 716 77 UK +44 1442 233555 USA +1 8 532 4752 www.thermo.com Legal Notices 28 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. View additional Thermo Scientific LC/MS application notes at: www.thermo.com/appnotes Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. AN62867_E 11/8S Part of Thermo Fisher Scientific