Determination of PCDD/Fs in Environmental Samples using Accelerated Solvent Extraction (ASE) and GC-MS/MS
|
|
|
- Flora Lloyd
- 10 years ago
- Views:
Transcription
1 Determination of PCDD/Fs in Environmental Samples using Accelerated Solvent Extraction (ASE) and Paul Silcock, 1 David Gardner, 2 John Fardon, 2 Steve Moss 2 and Inge de Dobbeleer 1 1 Thermo Scientific POPs Centre of Excellence, Bremen, Germany 2 National Laboratory Service - Environment Agency, Leeds, United Kingdom Application Note 36 Key Words Dioxins, furans, PCDD, PCDF, environmental, fly ash, sediments, complex matrix, accelerated solvent extraction, isotope dilution, pressurized solvent extraction Introduction The measurement of PCDD/Fs in the environment is a widespread activity carried out by many regulatory agencies globally. The chronic toxicity of these compounds to humans and wildlife at extremely low concentrations requires that the techniques used in determination must be both sensitive and selective enough to allow high confidence results. This is especially true when measuring background levels in environmental matrices, such as soil and sediment or byproducts from waste incineration processes. Traditionally high resolution magnetic sector GC-MS () instrumentation has delivered the required analytical performance and has become the gold standard technique. In recent years, there has been more interest in GC triple-quadrupole instrumentation for this purpose, especially in the area of food safety control. 1,2 For this area and environmental analysis, it is necessary to deliver data that performs in the range of HRMS systems, which requires especially sensitive triple-quadrupole systems. It is also necessary to incorporate data-processing software specifically designed to handle the complex calculations associated with dioxins analysis. This application note describes the use of the Thermo Scientific TSQ Quantum XLS Ultra as applied to the analysis of PCDDs/PCDFs in sediments, soils, bottom, and fly ash (as incineration by-products) at the levels of interest and the level of agreement with gold standard analysis using. During this study, instrumental LOQs using were calculated in the low fg/µl concentration ranges. This, along with further analytical performance, is discussed alongside ; especially the degree of agreement between the techniques in some routine sample batches. Materials and Methods Extraction and Cleanup The standard spiking protocols, extraction, and cleanup process for all sediment and soil samples were performed using an ISO1725 accredited in-house variation of EPA 16B3. For incinerator ash samples, an in-house method based upon EN was applied. For both of these methods, accelerated solvent extraction was used as the preferred technique for extraction after initial sample processing. This is routinely used in the laboratory to reduce solvent consumption, allow for automation, and to generate efficient sample extractions within the performance limits required.
2 2 Accelerated Solvent Extraction Methods ASE cells were loaded with a base layer of copper filings then a layer of rinsed Ottawa sand (previously ashed at 45 C) after which 5 g of sample for soils/sediments (1 g for fly ash). Finally, another layer of sand was added to fill the cell and then extracted (Figure 1). GC-MS Measurement & Data Processing The measurements were performed using a Thermo Scientific TRACE GC Ultra coupled to a TSQ Quantum XLS Ultra system with quantitation performed using Thermo Scientific TargetQuan 3 software. Tables 1 and 2 show the instrument parameters used. Table 1. GC and injector conditions. Split/Splitless Injector (PCDD/Fs) Injection Temperature 26 C Toluene flow during extraction Extraction Solvent: Extraction Pressure: Oven Temperature: Heating Time: Static Time: Flush: Purge: Toluene 15 psi 175 C 8 minutes 5 minutes 6 % Cell volume N 2 24 Seconds Ottawa sand Soil/sediment/fly ash Ottawa sand Copper filings Figure 1. Cell schematic and method conditions for the extraction of PCDD/Fs from soil, sediment, and fly ash samples. Liner Injected Volume Splitless Time Surge Pressure GC Column GC Column Flow Table 2. Mass spectrometer parameters. Splitless straight liner (Siltek) mm (PN 453T2121) 2 µl (toluene) 1.5 min 22 psi (1 min) GC Program Initial Temperature 12 C Rate 1 TraceGOLD TG-5SilMS 6m.25mm.25um (PN ) 1.2 ml/min constant 18 C/min to 2 C (1 min) Rate 2 4 C/min to 29 C Final Temperature 29 C for 15 min Parameters Source Temperature 25 C Ionization Electron Energy EI 4 ev Emission Current 5 µa Q2 Gas Pressure (Argon) Collision Energy Q1 Peak Width FWHM Q3 Peak Width FWHM 1.5 mtorr 22 ev.7 Da.7 Da
3 The collision cell (Q2) gas pressure and collision energy were optimized for PCDD/F measurement at 22 ev. The monitored selected reaction monitoring (SRM) transitions as well as the MS conditions are given below in Table 3. 3 Table 3. Target congener groups SRM transitions. Compound Name Abrev. Formula Nominal Mass Exact Mass Presursor Ion m/z Product Ion m/z Collision Energy [ev] Tetrachloro-dibenzodioxin TCDD H 4 Cl Pentachloro-dibenzodioxin PeCDD Cl Hexachloro-dibenzodioxin HeCDD Cl Heptachloro-dibenzodioxin HpCDD Cl Octachloro-dibenzodioxin OCDD Cl Tetrachloro-[ ]dibenzodioxin TCDD H 4 Cl Pentachloro-[ ]dibenzodioxin PeCDD Cl Hexachloro-[ ]dibenzodioxin HeCDD C l Heptachloro-[ ]dibenzodioxin HpCDD Cl Octachloro-[ ]dibenzodioxin OCDD Cl Tetrachloro-dibenzofuran TCDF H 4 OCl Pentachloro-dibenzofuran PeCDF OCl Hexachloro-dibenzofuran HeCDF OCl Heptachloro-dibenzofuran HpCDF OCl Octachloro-dibenzofuran OCDF OCl Tetrachloro-[ ]dibenzofuran TCDF H 4 OCl Pentachloro-[ ]dibenzofuran PeCDF OCl Hexachloro-[ ]dibenzofuran HeCDF OCl Heptachloro-[ ]dibenzofuran HpCDF OCl Octachloro-[ ]dibenzofuran OCDF OCl
4 4 Results and Discussion Sensitivity for PCDD/F Analysis The prerequisites for a technique to be applicable to low level PCDD/F determinations is sensitivity and selectivity. The concentration levels where these substances are required to be measured often exceed the performance capability of bench-top GC-MS systems. In order to test the sensitivity of the TSQ Quantum XLS Ultra for this application, a serial dilution of commonly used EPA 16B CS1 standard was performed, and increasingly low levels were injected onto the system. The limit of detection (LOD) was then calculated using statistical methods. Figure 2 shows the lowest level CS1 dilution (x1 vial concentrations given in Table 4) used for this study. Quantitative SRM ions are clearly detected along with the confirmatory SRM ions for all PCDD/Fs tested within the QC ion ratio criteria. Precision studies at this level enabled an LOD to be calculated for the methodology. Table 4 gives the precision data obtained and calculated LOD at 99 % confidence limits. The LOD obtained from the TSQ Quantum XLS Ultra was found at a level that is highly applicable for environmental dioxins analysis in a range of sample types. Table 4. Instrumental LOD (pg/µl) given to 2 d.p. (99% confidence) and precision data (n=1) for PCDD/Fs from analysis of a ten times diluted EPA 16 CS1. Concentration pg/µl RSD % LOD (99) pg/µl 2,3,7,8 TCDF ,3,7,8 TCDD ,2,3,7,8 PeCDF ,3,4,7,8-PeCDF ,2,3,7,8 PeCDD ,2,3,4,7,8-HxCDF ,2,3,6,7,8-HxCDF ,3,4,6,7,8-HxCDF ,2,3,4,7,8-HxCDD ,2,3,6,7,8-HxCDD ,2,3,7,8,9-HxCDD ,2,3,7,8,9-HxCDF ,2,3,4,6,7,8-HpCDF ,2,3,4,6,7,8-HpCDD ,2,3,4,7,8,9-HpCDF OCDD OCDF TCDD/F PeCDD/F 1 RT: 32.7 RT: RT: RT: > RT: RT: > > RT: RT: > > RT: > > > RT: 4.29 HxCDD/F RT: 41.8 RT: RT: HpCDD/F RT: OCDD/F RT: RT: > RT: 41.8 RT: > > RT: RT: 52.4 RT: RT: > RT: > RT: RT: > RT: > RT: RT: > RT: > > > Minutes Minutes Minutes > Figure 2. SRM chromatograms of PCDD/Fs after injection of a 1x diluted EPA16B CS1 standard. (TCDD/F.5 pg/µl, PeCDD/F thru HpCDD/F.25 pg/µl, OCDD/F.5 pg/µl)
5 Quality Control in Routine Sample Batches To evaluate the quantitative performance of the TSQ Quantum XLS Ultra when applied to analytical quality control samples, data was taken and compared with that obtained on currently implemented systems (see Figure 3). These quality control samples included certified reference materials (CRMs) for sediments SETOC 738 and CRM 49 incinerator fly ash. These were all processed through the entire method procedure which included accelerated sample extraction. All of the compound recovery QC criteria specified in EPA 16 and EN1948 were satisfied. This gave confidence that the extraction yielded high recoveries throughout the study. CRM 49 is a highly contaminated incinerator fly ash. This type of sample and level of contamination allows for a great opportunity for interference to occur. The MS/MS system measured consistently higher concentrations for TCDD/F and some penta and hexa furans. This was indeed closer to the true CRM consensus value, so the QC check passed. SETOC 738 is a sediment CRM sample that is much lower level CRM and had good agreement in the calculated concentrations between the and HRMS data. The data on these three types of QC sample all fell within acceptable performance limits for the current methodology suggesting that the selectivity and quantitative performance of the technique is applicable for reporting PCDD/F data in the routine environmental lab. 5 5 Par CRM 49 SETOC 738 % Deviation from HRMS Result TCDF 2378-TCDD PeCDF PeCDF PeCDD HxCDF HxCDF HxCDD HxCDD HxCDD HxCDD HxCDF HpCDF HpCDD HpCDF OCDD OCDF Figure 3. Mean results for routine laboratory QC samples for the in-house PCDD/Fs method over 3 routine batches of samples. This is plotted as difference to result obtained on the system.the QCs include a precision and recovery standard (PAR), Incinerator fly ash certified reference material (CRM 49) and a sediment certified reference material (SETOC 738).
6 6 Routine Samples The QC performance was reflected in the real sample batches with generally very good agreement in results between the and approach. Larger deviations in calculated value were reported for incinerator fly ash samples, particularly for the lower chlorinated PCDFs. These differences are thought to arise from selectivity differences in the two techniques. The result remained valid within the quality control criteria specified within the methodology. SRM chromatograms for TCDDs in three different routine environmental sample types are given in Figure 4. These include both quantifying and confirming SRM transitions. The selective detection of the highly significant 2,3,7,8- TCDD congener was achievable in all samples. Calculated concentrations for each congener are given in Table 5 for both the and analyses for routine soil, sediment, and a matrix QC sample. Again, these show a good level of agreement in the calculated result. For the soil samples, the TSQ Quantum XLS Ultra showed confirmed detections of PCDD/Fs for the low concentration samples when no result was reported from. This was due to the lower performance of the system used in this study and not directly related the more sensitive systems available today. Relative Abundance 1 Fly Ash Contaminated Soil Contaminated Sediment > > > ,3,7,8-TCDD 2,3,7,8-TCDD 2,3,7,8-TCDD Relative Abundance > > > RT: RT: RT: RT: 32.2 RT: RT: RT: RT: Minutes Minutes Minutes Figure 4. TCDD SRM chromatograms for three different sample types typically analyzed in the environmental laboratory. Left: Fly ash (2,3,7,8-TCDD 31.4 ), Center: Contaminated soil (2,3,7,8-TCDD 1.85 ), Right: Contaminated sediment (2,3,7,8-TCDD 17.9 ). Table 5. Calculated concentrations of TCDD/F congeners in soil and sediment samples run on both and. (ND=not detected) Soil Sediment 1 Sediment 2 SETOC 738 (CRM 2,3,7,8 TCDF ,3,7,8 TCDD ND ND ,2,3,7,8 PeCDF 1.26 < ,3,4,7, 8 PeCDF ,2,3,7,8, PeCDD.436 ND ,2,3,4,7,8,-HxCDF ,2,3,6,7,8,-HxCDF ,3,4,6,7,8,-HxCDF ,2,3,4,7,8,-HxCDD.271 ND ,2,3,6,7,8,-HxCDD ,2,3,7,8,9-HxCDD.469 ND ,2,3,7,8,9-HxCDF.932 ND ,2,3,4,6,7,8-HpCDF ,2,3,4,6,7,8-HpCDD ,2,3,4,7,8,9-HpCDF OCDD OCDF
7 Conclusion The TSQ Quantum XLS Ultra is a highly sensitive and selective system applicable to dioxins and furans analysis in a range of environmental sample types and generates results that perform within current in-house QC criteria. Some differences arise in the calculated concentration between the and analysis. These are thought to be related to the differences in the selectivity mechanism of each system. applied to PCDD/Fs still allows for full isotope dilution quantitation as currently used by systems. TargetQuan 3 software can be used with the TSQ Quantum XLS Ultra to provide the specific calculations required. The Thermo Scientific Dionex ASE Accelerated Solvent Extractor system allows for unattended, efficient extractions from all samples and enabled recoveries well within QC criteria. References 1. Ingelido, A. M.; Brambilla, G.; Abballe, A.; di Domenico, A.; Fulgenzi, A. R.; Iacovela, N.; Lamiceli, L.; Valentini, S.; De Felip, E. Rapid Commun. Mass Spectrom. 212, 26(3), Kotz, A.;, Malisch, R.; Wahl, K.; Bitomsky, N.; Adamovic, K.; Gerteisen, I.; Leswal, S.; Schächtele, J.; Tritschler, R.; Winterhalter, H. Organohalogen Compd. 211, 73, Silcock, P.; Krumwiede, D.; de Dobbeleer, I.; Huebschmann, H. J.; Kotz, A. 211, Thermo Scientific Application Note Application Note Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Austin, TX USA is ISO 91:2 Certified. Africa-Other Australia Austria Belgium Canada China Denmark Europe-Other Finland/Norway/Sweden France Germany India Italy Japan Latin America Middle East Netherlands New Zealand Russia/CIS South Africa Spain Switzerland UK USA AN36_E 7/S
TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet
TargetQuan 3 Software Leading the way in regulatory POPs quantification Bullet Bullet Bullet Leading the way in regulatory POPs quantification Analyse Samples Open Sequence Process Sequence Analyse Samples
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Application Note. Abstract. Authors. Food Safety
Determination of Dioxin-Like and Non-Dioxin-Like Polychlorinated Biphenyl Congeners in Foodstuffs and Animal Feed Using the Agilent 7000 Triple Quadrupole GC/MS System Application Note Food Safety Authors
MICROORGANIC POLLUTANTS: ACTION NEEDED TO SIMPLIFY THEIR ANALYSIS
MICROORGANIC POLLUTANTS: ACTION NEEDED TO SIMPLIFY THEIR ANALYSIS Mosca S., Guerriero E., Rotatori M. The analysis of PCDD/Fs, PCBs and other related compounds requires complex analytical procedures using
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data
Low cost ICP-OES analysis high quality data The NEW Thermo Scientific icap 7000 Series ICP-OES provides low cost multi-element analysis for measuring trace elements in a diverse sample range. The instrument
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use Confidence in Results With Data Integrity 2 Make the World Healthier With the LC-MS Tests You Run Confidence in Test Results With Data
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Thermo Scientific TSQ Quantum XLS Triple Quadrupole GC-MS/MS
c h r o m a t o g r a p h y Thermo Scientific TSQ Quantum XLS Triple Quadrupole GC-MS/MS The Next Evolution in GC Triple Quadrupoles Higher Sample Throughput Higher Productivity Detection Limits Beyond
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID
Thermo Scientific Compound Discoverer Software A New Generation of integrated solutions for small molecule structure ID Transforming Small Molecule Research Solutions for Small Molecule Structural ID Thermo
Thermo Scientific SIEVE Software for Differential Expression Analysis
m a s s s p e c t r o m e t r y Thermo Scientific SIEVE Software for Differential Expression Analysis Automated, label-free, semi-quantitative analysis of proteins, peptides, and metabolites based on comparisons
Gas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA
Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers Rob Grothe Thermo Fisher Scientific, San Jose, CA Overview Purpose: Robust and accurate calibration of mass positions
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,
Thermo Scientific ConFlo IV Universal Interface
m a s s s p e c t r o m e t r y Thermo Scientific ConFlo IV Universal Interface The Universal Interface Part of Thermo Fisher Scientific ConFlo IV Universal Interface The Universal Interface The development
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping
Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping For Conclusive Characterization of Biologics Deep Protein Characterization Is Crucial Pharmaceuticals have historically been small
GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs
GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs Kinet C, De Pauw E, Focant JF CART, Mass Spectrometry Laboratory, Chemistry Department, University of Liège, Allée de la Chimie 3,
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)
Application Note 340 Determination of Fat in Dried Milk Products INTRODUCTION Many extraction techniques for the determination of fat in food are labor-intensive or require long extraction times. The Roese-Gottlieb
Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare
Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare HEALTHCARE SUPPLYSIDE SOLUTIONS Focused on your unique needs for managing assets, costs, and suppliers Part of Thermo Fisher Scientific
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
LC-MS/MS, the new reference method for mycotoxin analysis
LC-MS/MS, the new reference method for mycotoxin analysis What is necessary for the implementation of the multi-residue mycotoxin method in an analytical laboratory? Azel Swemmer [email protected] AFMA
European Union Reference Laboratory for Dioxins and PCBs in Feed and Food
for Dioxins and s in Feed and Food Chemisches und Veterinäruntersuchungsamt Freiburg PO Box 46 D-793 Freiburg Germany CEN Ring Trial - Animal Feed Ring trial on determination of PCDD/Fs and s in animal
MIDLAND AREA SOIL SAMPLING MIDLAND, MICHIGAN
QUALITY ASSURANCE PROJECT PLAN MIDLAND AREA SOIL SAMPLING MIDLAND, MICHIGAN Revised May 2013 URS Corporation 9400 Amberglen Boulevard Austin, Texas 78729 (512) 454-4797 TABLE OF CONTENTS Section 1 Objectives
METHOD SUMMARY & DATA QUALITY OBJECTIVES
METHOD SUMMARY & DATA QUALITY OBJECTIVES TestAmerica West Sacramento Method 1613B: Polychlorinated Dibenzo-p-dioxins (PCDDs) and Polychlorinated Dibenzofurans (PCDFs) by High Resolution Gas Chromatography/High
Thermo Scientific ExactFinder Software
mass spectrometry Thermo Scientific ExactFinder Software Workflow software for routine targeted and general unknown screening Unifies qualitative and quantitative high-resolution accurate mass workflows
AutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?
AutoSpec Premier Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements? Looking for sensitivity, selectivity and dynamic range? The AutoSpec Premier is the latest
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)
Application Note: 52020 Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Timothy O. Deschaines, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words Array Automation
HRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform
Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform Charles Yang, Leo Wang, Dipankar Ghosh, Jonathan Beck,
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
A Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH
RAPID MARKER IDENTIFICATION AND CHARACTERISATION OF ESSENTIAL OILS USING A CHEMOMETRIC APROACH Cristiana C. Leandro 1, Peter Hancock 1, Christian Soulier 2, Françoise Aime 2 1 Waters Corporation, Manchester,
HORIZONTAL concerning the layout, editorial and technical content, this document needs revising. In order to harmonize all documents within
Template for s and secretariat Date: 006-06-30 Document: CEN/BT/TF 151/TG 4 N0013 Horizontal 4.7 BFR Author s: ALL; Replies: TG 4 1 (3) 4 5 (6) (7) on each submitted NL ge This document is poorly attuned
Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software
Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
LIFECYCLE. Thermo Scientific. Enterprise Solutions
Thermo Scientific LIFECYCLE Enterprise Solutions Support Plans Improve productivity and quality Increase production uptime Extend instrument lifetime Lower overall cost-of-ownership Keep instruments up-to-date
FT-NIR for Online Analysis in Polyol Production
Application Note: 51594 FT-NIR for Online Analysis in Polyol Production Key Words Acid Number Ethylene Oxide FT-NIR Hydroxyl Value Polyester Polyols Abstract Hydroxyl value and other related parameters
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
Opiates in Urine by SAMHSA GC/MS
application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health
Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software
Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software White Paper 70885 Executive Summary Over the last few decades chromatographic instrumentation and separations have
[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process
- Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
What is Eluent Generation?
What is Eluent Generation? Monika Verma Thermo Fisher Scientific, Sunnyvale, CA, USA White Paper 70607 Executive Summary Eluent generation is an electrolytic process that allows the production of high
Chemical Profiling and Differential Analysis of Whiskies Using Orbitrap GC-MS
Chemical Profiling and Differential Analysis of Whiskies Using Orbitrap GC-MS Dominic Roberts, 1 Jana Hajslova, 2 Jana Pulkrabova, 2 and Paul Silcock 1 1 Thermo Fisher Scientific, Runcorn, UK 2 University
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015
A Navigation through the Tracefinder Software Structure and Workflow Options Frans Schoutsen Pesticide Symposium Prague 27 April 2015 Kings day in The Netherlands 1 Index Introduction Acquisition, Method
How To Control A Record System
Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) Software for 21 CFR Part 11 Compliant Laboratories Technical Note 43106 Key Words Compliance, Electronic Records, 21 CFR Part 11 Goal
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
Pesticide residue analysis: New Trends. Michal Godula, Ph.D. EU Marketing Manager Food Safety & Environmental Thermo Fisher Scientific
Pesticide residue analysis: New Trends Michal Godula, Ph.D. EU Marketing Manager Food Safety & Environmental Thermo Fisher Scientific Why pesticide analysis? Extensively used in agriculture to control
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity
For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,
Thermo Scientific Vanquish UHPLC System. Revolutionize Your. UHPLC Experience
Thermo Scientific Vanquish UHPLC System Revolutionize Your UHPLC Experience We feel that achieving better performance is not the same as spending more effort. We know that every breakthrough starts with
Oilfield Services Measurement Solutions
Oilfield Services Measurement Solutions Superior flow and density measurement Best-in-class measurement How can I streamline my operations for efficiency without compromising environmental, safety and
Thermo Scientific Qtegra Intelligent Scientific Data Solution Software. Delivering quality. Driving productivity
Thermo cientific Qtegra Intelligent cientific Data olution oftware Delivering quality Driving productivity There can be economy only where there is efficiency The Thermo cientific Qtegra Intelligent cientific
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
Benzo[a]pyrene Method Development
Benzo[a]pyrene Method Development Dr Michael Intorp CORESTA Cooperation Center for Scientific Research Relative to Tobacco FDA Center for Tobacco Products Scientific Workshop Tobacco Product Analysis 30
Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps
Application Note: ANTV10 Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Abstract In order to regulate product safety and quality
Which analytical tools/instruments/techniques are used for POPs. A general walk through the analysis
Which analytical tools/instruments/techniques are used for POPs A general walk through the analysis Kay Kelterer Eurofins GfA Lab Service GmbH Großmoorbogen 25, 21079 Hamburg, Germany 23.04.2015 Miljøgifter
New drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas
Determination of Calcium and Magnesium in Brine
Application Note Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS
aurora M90 A new era in ICP-MS Innovation with Integrity ICP-MS Your Partner in ICP-MS Solutions Bruker continues to find new and novel ways to meet your changing needs. As a leader in elemental analysis
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum
Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in uman Serum Technical verview Clinical Research Authors Anthony Macherone, Ph.D. Agilent Technologies,
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
Accepted 26 May 2006 Available online 17 January 2007
Chemosphere 67 (27) 1822 183 www.elsevier.com/locate/chemosphere A study on PCB, PCDD/PCDF industrial contamination in a mixed urban-agricultural area significantly affecting the food chain and the human
Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley
Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening
Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Application Note Forensics Author Erin Shonsey Director of Research Alabama Department of Forensic Science Abstract
Application Note # MS-14 Fast On-site Identification of Drugs with the mobile GC/MS system E²M
Bruker Daltonics Application Note # MS- Fast n-site Identification of Drugs with the mobile GC/MS system E²M For the detection of drugs a couple of quick tests are available which were used by the police
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative