CAUSTIC SOLUTION REPLACEMENT (Mud Safe CR) STABILITY TESTING



Similar documents
AUTOCLAVE CORROSION INHIBITOR EVALUATION

Dew Point Tester. Instruction Manual. CVS Regular Chiller Model A-2. CVS Regular Chiller Model A-2

Distillation Experiment

Lab Exercise 3: Media, incubation, and aseptic technique

STANDARD OPERATING PROCEDURES Safe Autoclave Operations

Full Technical Catalog For Closed-loop Sampling System

23 The Thermal Conductivity Detector

Hydrogen Sulfide in Petroleum. Mike Nicholson/Tim O BrienO Baker Petrolite Corporation

MOISTURE (Karl Fischer, Buffered)

105 Adopted:

Acids & Bases: Using Purple Cabbage as a ph indicator. Grade 9 Activity Plan

Chemistry 212 VAPOR PRESSURE OF WATER LEARNING OBJECTIVES

Standard Hydraulic Piping Flushing Procedure

ph Measurements of Common Substances

Test Methods for the Colour Fastness of Leather & Dyes. TFL Leather Technology Ltd. Quality and Environment February 2004 Version 1.

Experiment 6 Coffee-cup Calorimetry

Mixing Warm and Cold Water

Sample Collection Procedures for Insulating Fluid Samples

WIKA INSTRUMENT CORPORATION

Index Page. Pumping Systems Intensifiers Gas Boosters High Pressure Generators

VAPOR PRESSURE AS A FUNCTION OF TEMPERATURE. This laboratory covers material presented in section 11.8 of the 9 th Ed. of the Chang text.

Installing a Probe. Typical Installation

Pre-Lab Notebook Content: Your notebook should include the title, date, purpose, procedure; data tables.

EQUIPMENT CALIBRATION/MAINTENANCE

Procedure Code A Insulin pen needles A4230 Infusion set for external insulin pump, non needle cannula type A4231 Infusion set for external

REGULATORS 175/153 Series

USER MANUAL. 10 GPM Small Single Line Cryogenic Coupling

ph and Acidity in Wine and Fruit Juice

Sterilization methods and equipment Lab 1-2

TUTTNAUER OLD AUTOCLAVE OPERATING PROCEDURE

The NEW Advanced Range of Portable Temperature Calibrators PROVISIONAL. Portable Dry Blocks, Liquid Baths and Multi-functional Heat Sources

P O B o x S t. L o u i s, M O P h o n e : ( ) ~ E m a i l : w m h u I t a o l.

AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP

BB-18 Black Body High Vacuum System Technical Description

Originators of the Flexible Water Tube design

Isolation of Caffeine from Tea

Gas Custody Transfer Calibration

UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE ABSORPTION OF CARBON DIOXIDE INTO WATER

Preview of Period 5: Thermal Energy, the Microscopic Picture

INSTRUCTIONS Edition AC

AUTOCLAVE STEAM STERILIZERS

Enzyme Lab. DEFINITIONS: 1. Enzyme: 2. Catalase: 3. Catalyze: 4. Hydrolysis: 5. Monomer: 6. Digestion: BACKGROUND INFORMATION

Module 1: History of Fuels. Lecture 6: Fundamental definitions, properties and various measurements

Calibrating Differential Pressure Sensors

COMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.

Field Test Protocol NANOTAC

HyLife TM HVAC Line Corrosion testing. Precision Tubing Technology Centre, 2012

PP / PE X X X X (X) X PA (X) X

Stainless Steel Pipe Fabrication

AIRMoN, ph and Conductivity Field Measurement

1. The Determination of Boiling Point

KGI MEDICAL WASTE MANAGEMENT PLAN

Cells and cell cultivation

Sanitary Type Electronic Flow Switches

MOLAR AND PARTIAL MOLAR VOLUMES IN AQUEOUS SOLUTIONS (10/21/03)

PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE

PREPARATION FOR CHEMISTRY LAB: COMBUSTION

AIRFREE TECHNIQUE AND SENSITIVE REAGENTS S ECTI O N 1: GLASS W ARE A ND E Q UIP M ENT. A. Using a manifold

Care and use guide. Contents. a handy solution made easy. page

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *

Trace Gas Exchange Measurements with Standard Infrared Analyzers

Modulex. Gas Fired Boiler Systems PIPING GUIDE GF-115-P. Modular, Condensing, Hot Water Boilers Models: 303, 454, 606, 757, 909, Instruction No.

Transformation Protocol

CHEM 105 HOUR EXAM III 28-OCT-99. = -163 kj/mole determine H f 0 for Ni(CO) 4 (g) = -260 kj/mole determine H f 0 for Cr(CO) 6 (g)

Selective Catalytic Reduction (SCR) and Diesel Exhaust Fluid (DEF) Training Module

By Vince Paulauskas, Technical Service Manager, Winpak December 2002

Liquid level measurement using hydrostatic pressure and buoyancy

Catalase Enzyme Lab. Background information

COMMON LABORATORY APPARATUS

System Diagnosis. Proper vehicle diagnosis requires a plan before you start

Dissolved Oxygen Sensors

Mustang Series PRESSURE REDUCING CONTROL VALVE WITH PRESSURE SUSTAINING FEATURE. M115-2 (Globe) M (Angle) Schematic. Standard Components

PRESSURE REDUCING CONTROL VALVE

Oil Sight Glass BENEFITS

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

APPENDIX N BULK MILK TANKER SCREENING TEST FORM

Catalase. ***You will be working with hot water, acids and bases in this laboratory*** ****Use Extreme Caution!!!****

ENGINE COOLING SYSTEM

INSTRUCTION MANUAL. Boil-Out TABLE OF CONTENTS SAFETY WARNINGS 2-3 RESPONSIBILITY 3 INTRODUCTION 3 PRE-BOIL OUT PROCEDURES 3-5 BOIL OUT PROCESS 5-9

Bench Top Reaction System

V47 Series Temperature Actuated Modulating Water Valves

UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE CONVECTIVE HEAT TRANSFER

Installation Questions

C. starting positive displacement pumps with the discharge valve closed.

Product Bulletin INTEROX HYDROGEN PEROXIDE

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Optimization Design for Sulfur Dioxide Flow Monitoring Apparatus in Thermal Power Plants Hao-wei Hu 1, a, Xue Yang 1, b and Xiao-wei Song 1, c

Guidance on safe use of Autoclaves

FIXED DISPLACEMENT HYDRAULIC VANE PUMPS BQ SERIES

Pressure Coring, A New Tool for Unconventional Oil & Gas Characterization. Matt Bjorum Global Product Line Manager

FIXED DISPLACEMENT HYDRAULIC VANE PUMPS BQ SERIES

Validation Guide. USTR 805j (5) Steam sterilization of Pall filter assemblies which utilize replaceable filter cartridges

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P Thermodynamic cycle Heat pump

Bio-Clean. for Septic Systems. Tips for Using Bio-Clean. Page Index METRIC CONVERSIONS

Fall 2004 Ali Shakouri

Model 43AP Pneumatic Controller, Style B

Experiment 1: Colligative Properties

Catalytic Activity of Enzymes

Combustion Calorimetry

Safety Data Sheet SECTION 1: IDENTIFICATION OF THE SUBSTANCE/PREPARATION AND OF THE COMPANY/UNDERTAKING

Transcription:

CAUSTIC SOLUTION REPLACEMENT (Mud Safe CR) STABILITY TESTING Cormetrics Job #: 13-207 Prepared for: Heartland Energy Group Ltd. Date: March 20, 2013 Revision 2 Lab: Bay 4 2280 39 th Avenue NE, Calgary, AB. T2E 6P7 Phone: 258 2853 Fax: 291 1423

Heartland Energy Group Ltd. Page 2 of 5 1. Introduction A synthetic caustic solution replacement product was submitted for thermal stability evaluation by Heartland Energy Group Ltd. The caustic solution replacement was evaluated in the autoclave apparatus according to the protocol received from Heartland Energy. The product tested is listed in Section 2.1. 2. Test Conditions 2.1. Synthetic Caustic Replacement Product Name Mud Safe CR Table 1 Synthetic Caustic Product 2.2. Autoclave Test Apparatus The autoclaves used by Cormetrics Limited are constructed of Hastelloy 276-C and have a capacity of approximately 300mL. The tests were carried out with 200 ml of synthetic caustic product in each cell (approximately two-thirds full). A Teflon sleeve was inserted into the base of the autoclave. The synthetic caustic fluid was then poured into this Teflon cup. The temperature of the fluid in the autoclave is sensed by a thermistor probe, held at the center of the cell by a Hastelloy sleeve. Charging of the autoclave is by means of an offset Hastelloy tube, fitted with a pressure gauge and sour-service valve. Each cell is also equipped with a pressure relief valve which is used when purging the test liquids directly in the cells. 2.3. Mud Safe CR Stability Test Protocol The Mud Safe CR product was evaluated at 149 C (300 F) under a Nitrogen gas cap. The temperature was chosen to determine the thermal stability of the product at maximum field temperatures. The autoclave was brought to temperature and held for the test duration of 3 hours. The autoclave cell was then allowed to cool for half an hour prior to venting the pressure. Visual observations and ph measurements were taken on the post-test fluids to determine if they were still caustic.

Heartland Energy Group Ltd. Page 3 of 5 2.4. Autoclave Pressuring Once the cell had been sealed it was then filled with 500 psi of pure nitrogen. Once pressurized, the autoclave cell was then placed inside a heating mantle and brought to the desired test temperature via proportional temperature controllers. The fluid was not stirred. 3. Results Summary 3.1. Stability Test Data 160 Heartland Energy Group Mud Safe CR Temperature Stability Autoclave Evaluation Mud Safe CR Temperature Mud Safe CR Pressure 900 140 850 120 800 Temperature ( C) 100 80 60 750 700 650 Pressure (psi) 40 600 20 0 Date: February 22, 2013 550 Test: AC 432 Target Temp: 149 C (300 F) Initial N 2 Charge: 500 psi 500 0 50 100 150 200 250 300 Time (mins) Figure 1 Temperature and Pressure Data at 149 C (300 F)

Heartland Energy Group Ltd. Page 4 of 5 3.2. Visual and ph Data Figure 2 Fluid Visual and ph Pre Test at 149 C (300 F) Figure 3 Fluid Visual and ph Post Test at 149 C (300 F)

Heartland Energy Group Ltd. Page 5 of 5 4. Discussion The Mud Safe CR product is clear at room temperature with a very high ph. The ph as measured by 0-14 ph strips is fourteen. In the 149 C (300 F) test the product reached temperature after an hour of heating. The pressure increased during this heating period to 855 psi. When the cell was at temperature the pressure remained stable. The stability of the pressure at temperature indicates the rise in pressure is due to heating and not due to any changes in the fluid. After 3 hours exposure at 149 C (300 F) the Mud Safe CR fluid remained clear. The fluid retained its alkalinity with a ph of 14 as determined by 0-14 ph strips. 5. Conclusions In the conditions as tested Mud Safe CR is thermally stable and maintains its alkali nature up to a temperature of 149 C (300 F). The thermal stability threshold of this fluid was not reached in the conditions as tested. Sincerely, Cormetrics Limited Mike Koldijk, B.Sc. Frank Hornsby Please note, all fluid samples are stored for 6 months prior to disposal.