Guidance on safe use of Autoclaves
|
|
|
- Marvin Norman
- 10 years ago
- Views:
Transcription
1 Safety Office Reviewed: Dec 2013 Reviewed: DD Month Year Guidance on safe use of Autoclaves 1. Hazards & Uses There are several different types of hazard associated with the use of autoclaves. The main ones are associated with: Exposure to steam under pressure. Contact with hot surfaces and liquids Failure of the pressure vessel itself Manual handling whilst loading and unloading Exposure to infectious substances within the load Failure of the makesafe/discard cycle to sterilise contaminated waste. Autoclaves are used to sterilise equipment, media and to inactivate waste and therefore it is vitally important that sterility is assured. The following describes some of the main factors that should be taken into account in risk assessments, the types of control measures to be implemented and provides information on the key requirements for ensuring effective sterilisation. 2. Choice and performance The relevant British Standard covering use of autoclaves in laboratories is BS 2646 Autoclaves for Sterilisation in Laboratories 1993 and BSEN 12347: 1998 Performance criteria for steam sterilisers and autoclaves. When Schools are purchasing autoclaves or arranging maintenance work they should always check that the autoclave and the associated installation and servicing complies with the British Standard specifications. The need for any additional controls such as HEPA filters on the exhaust for autoclaves in Containment Level 3 laboratories should be determined by risk assessment. 3. Maintenance & Validation Regular maintenance is essential for the continued efficiency and safety of laboratory autoclaves and ancillary equipment. Schools should have a continuing programme of planned preventative maintenance throughout the life of the autoclave, in accordance with the manufactures recommendations. Where the autoclave is used to make safe waste it is also necessary to validate and monitor the performance of the autoclave. Further details on this are given in section 14 below. 4. Periodic examination In addition to ongoing maintenance and validation arrangements, it is a requirement under the Pressure Systems Safety Regulations that autoclaves be periodically examined by a competent person. This examination is necessary prior to the autoclave being brought in to service and thereafter at appropriate regular intervals. Within the University this is carried out by the University's Engineering Insurers Notification In order to ensure these requirements are met the details of all autoclaves must be notified at the time of installation or purchase to the Responsible Person within the
2 School, who will arrange for them to be added to the insurer s database, in order that they are assessed, insured and regularly inspected. The procedure for this can be found on the Commercial Services Website (follow the links to Insurance; Engineering and Statutory Inspections; Standard Inspections for Pressure Vessels and Lifting Equipment). The Safety Office should also be notified about autoclaves that are to be used for biological/gm waste inactivation so that appropriate checks can been made of validation procedures. These notification requirements also apply to any autoclave that is relocated within the University. 5. Management, Information and Training Autoclaves should be under the managerial control of a nominated individual who will be the responsible person for ensuring that the autoclave is safely managed and operated by trained and competent individuals. Operation instructions should be provided for the operator and should include details of action to be taken by the operator in the event of a fault or any abnormality in autoclave performance. It is recommended that simple, easy to understand instructions are provided close to the equipment. This should be drawn up with reference to the manual provided by the manufacturer. The responsible person /manager of the facility must ensure that operators receive training in accordance with the written instructions and are deemed competent before being allowed to operate the autoclave. Training and attainment of competence must be noted in the individuals training record. There should also be a notice adjacent to the autoclave which shows the name of the responsible person and a list of people trained and authorised to use the equipment. 6. Routine Inspection & Maintenance by Operators In addition to the maintenance and inspection detailed above there are a series of routine checks that should be carried out by the responsible person or a nominated individual. After each run the chart recorder should be inspected to ascertain that the operating cycle has been achieved. The following procedure should be carried out after each operation or where autoclaves are in constant use, at a suitable interval [minimum weekly]. When performing these tasks ensure that the autoclave is in a cool and safe condition Clean the chamber internally including fittings, shelves and brackets, as recommended by the manufacturer. Clean the drain filter if fitted. Clean the door seal with a damp cloth and examine to ensure there are no cuts or abrasions. Weekly Check the operation of indicator lamps. During an operating cycle, check the pressure gauge against the temperature gauge to ensure correct operation.
3 7. Failure during the discard cycle When a fault occurs during a discard/makesafe cycle an assessment of risk should be made and appropriate action taken. It may be necessary to disinfect those chamber attachments on which engineering work is to be carried out. During a makesafe process, chamber condensate should be considered to be contaminated with viable microorganisms. Therefore adequate time must be allowed to allow the autoclave to cool down completely and for aerosol to dissipate before opening the door. A contaminated laboratory autoclave should never be returned to the manufacturer for servicing or repair. Decontamination should be carried out in accordance with local safety rules before a service engineer works on the autoclave and suitable decontamination certification provided. The use of permit to work systems for routine maintenance by service engineers is strongly recommended. 8. Protective Clothing & Equipment Protective clothing with long cuffed sleeves worn in the autoclave loading/unloading area(s). Additional protective clothing should be available in the loading/unloading area(s) to protect the operator. This should include: An impervious apron Heatresistant gauntlet gloves A fullface visor Suitable nonslip heavyduty safety shoes 9. Loading the Autoclave The hazards on loading include spills of biohazardous material, broken glass and dropped load contents. Controls All materials awaiting autoclaving should be stored safely. Bagged waste materials should be supported in a robust, leakproof container, bags should never be placed directly on the floor. It should be possible to identify the source of all materials to be autoclaved. A labelling system should be used in order that the waste can be tracked back to the laboratory or area from where it came. This serves two important purposes. Firstly to be able to identify the contents of the waste and what hazards may be present in the event of someone having an accident in which they are exposed to any infectious agents that may be present. Secondly to be able to identify the producer of the waste if there is any problem with it, such as a container is leaking, overfilled, or contains inappropriate items etc., to prevent repetition of bad practice. Items should be packed in a way that ensures that steam will penetrate the load. Bags should not be sealed; if the tops have been taped or tied the bags should be opened or slashed to allow effective steam penetration. The following tin has been specifically designed to facilitate safe transport and ensure effective steam penetration and it is strongly recommended that these be used wherever possible.
4 For discard/waste inactivation, the load within the autoclave must be positioned in accordance with the load validation arrangements outlined in Appendix A and the validation report [see section 13]. Ensure that the volume of liquid in the vessel containing the reference probe is at the correct level as detailed in the validation report. When some chemicals (including disinfectants) or materials are autoclaved they are likely to produce vapours that could harm persons exposed when opening the door or they may be corrosive [e.g. Virkon or Chloros] to the autoclave. A risk assessment should be made of all items to be autoclaved taking account of this possibility. Radioactive waste must not be autoclaved. It must be sealed in a yellow burn bin and sent for incineration. If the waste is genetically modified it will need to go to an incinerator registered to accept GM Waste, if it cannot be effectively inactivated by disinfection. Consult the Safety Office for advice. 10. Unloading the Autoclave The hazards on unloading include splashes and spillage of hot material from the load, hot condensate, hot equipment, broken glass, dropped load contents, and vapour from volatile chemicals. Controls Temperature and pressure indicators and warning lights should be checked to ensure that the autoclave has successfully completed the operating cycle. If a fault is indicated, attempts to open the autoclave should only be made with the authority of the responsible person. It is dangerous to attempt to release the autoclave door mechanism before the chamber is vented to atmosphere or whilst the load contents are at high temperature. No attempt should be made by the operator to override door interlocking safety devices. The operator should stand clear when opening the door as hot liquid or vapour may escape from the chamber. The operator should also be aware that containers of liquid could be pressurised and may explode, volatile liquids may produce harmful vapour and liquids spilled on unloading may cause scalding. The following measures are designed to minimise the unloading hazard: temperature activated door interlocks; timer activated door interlocks; training and supervision of operators; protective clothing; and load transfer systems.
5 After autoclaving, waste containers should be emptied in a safe manner and their contents disposed of or reclaimed as specified by local laboratory rules. IN the event of there being a breakage inside the autoclave ensure that cut resistant gloves are worn and there is some means of removing debris from the rear of the chamber [e.g. long tongs/litter picker] 11. Waste Containers All waste should be placed in waste containers which should be easily transportable, leak proof and of a robust design with solid sides and bottoms. They should allow adequate steam penetration to the contents. If autoclave plastic bags are used they should be supported in a waste container whilst in the laboratory, during transport to the autoclave and also whilst in the autoclave. The bag should be open during the autoclave cycle so that steam can penetrate its contents. It is recommended that transparent (rather than opaque) autoclave bags be used as this allows the autoclave operator to see if any hazardous materials have been placed in the bag by mistake. 12. Operating Cycles Typical operating cycle conditions are listed in the table below. TYPICAL OPERATING CYCLE CONDITIONS Process Sterilising Temperature 0 C Holding Time (Minutes) Min 0 C Max 0 C Min Max Liquids sterilisation Equipment/glassware sterilisation Makesafe (for inactivation waste) 134 recommended The operating cycle should take account of heat up times. The time required for the load to reach sterilizing temperature should be determined during validation tests.
6 12.1 Operating Cycles for Liquids Sterilisation When liquids are sterilised the possibility of adverse effects on the liquid caused by the heat treatment must also be taken into account on selecting a cycle. The operating cycle should be selected to ensure that sterilisation is achieved with minimum damage to the liquid. Microbiological culture media are particularly heat sensitive; the degree of deterioration is related to the length of time the medium is maintained at sterilising temperature; the heatup and cooling stages also contribute significantly to this deterioration. Heatup times should be as short as possible, achieved by uniformly filling the chamber with steam at sterilising temperature. Large volumes of fluids will heat up slowly, therefore volumes of liquids should be kept small; a maximum container volume of 500ml is recommended, larger volumes taking considerably longer to heat up (and cool down). Cooling loads quickly helps to protect heatsensitive constituents and also shortens operating cycle times. Air at high pressure may be admitted to ballast the chamber, minimise boiling and prevent bottles exploding. Containers should be loosely capped unless they are specifically designed for sealing. However, sealing bottles can increase the likelihood of explosion during autoclaving and slows cooling Operating Cycles for Waste Inactivation When sterilising loads which contain infectious waste it is important that the sterilising temperature is attained and maintained for the required time at the centre of each part of the load. In order to ensure the minimum sterilising temperature at the centre of each part of the load is achieved the air must be completely removed from the load and then steam must then be injected into the chamber at specific pressure in order to ensure it penetrates the load. The temperature must then be held for the required sterilising time. Because contents of a waste inactivation load varies, this holding time may be in excess of the minimum required time this is to ensure that all parts of the load achieve that temperature for the minimum time. Key factors affecting this are: Load materials Plastic and card (consider the release of formaldehyde from card) containers will act as an insulator increasing the equilibration time. Metal boxes are the preferred option. Plastic coated bottles also act as an insulator, taking longer to heat up that normal glass bottles. Load Configuration The validated load must be documented, repeatable and based on a worst case scenario. Ideally validated loads should be photographed and logged on Appendix A.. Any new waste loads will need to be validated before the work commences. Waste should only be loaded and autoclaved in an approved and validated load pattern. Porous Loads Air removal is a key factor, failure to achieve direct steam contact with Dry Saturated Steam can lead to failure, regardless of whether the required temperature was achieved within the load.
7 A vacuum cycle should be used where possible and when not available a free steam cycle should be set up along with the use of approved discard containers. 13. Validation, Calibration and Routine Monitoring of Autoclave Performance 13.1 Validation & Calibration Validation is testing the ability of the autoclave to effectively perform the autoclaving process when loaded with userdefined loads. An autoclave cycle can be split into three phases, heating up, holding at pressure and cooling down. Validation should demonstrate that in the holding part of the cycle the necessary temperature has been attained throughout the load and this is held for the minimum time. Calibration, which should not be confused with validation, compares the temperature set on the dial to the temperature in the chamber The validation process For discard/make safe cycles where loads contain pathogens or harmful genetically modified microorganisms then validation must demonstrate that the process is effective in achieving a 100% kill. The validated load must be documented, repeatable and based on a worst case scenario, validated loads should be photographed and logged on Appendix A To do this a worst case load of simulated waste (i.e. worst case volumes, materials and equipment but not contaminated with pathogens) should be placed in the autoclave for the validation test.. Any new waste loads will need to be approved and validated before the work commences. Waste should only be loaded and autoclaved in an approved and validated load pattern. Appendix A is a Load Validation Form which can be used to ensure the validation load is suitably set up and recorded. Probes are then inserted at various 12 points in the waste and connected to recording equipment. This then shows the temperature at the various points throughout the cycle. The holding part of the cycle commences once all sensors indicate the prescribed sterilizing temperature has been reached. All sensors must then maintain at least that temperature for the prescribed holding time. Validation must be carried out by a competent trained person using thermometric test equipment that is calibrated and traceable to national standards. The test method is described in British Standard The use of biological or chemical indicators for this purpose is not acceptable. The use of bench top autoclaves for waste is not recommended because these cannot usually be validated. Loads comprising a variety of items and containers do not heat uniformly. Short holding times are therefore subject to large proportionate variations and should be avoided if possible. Following validation it will be possible to ascertain the settings required on the autoclave to achieve the above conditions. This will vary considerably with many factors including the load characteristics. For example, one particular autoclave may need to be set at 121 o C for 50 minutes in order to achieve cycle (i) above whereas another may need to be set at 123 o C for 75 minutes to achieve the same. Where dials can be altered by the user, the temperature and time requirements for a waste cycle should be clearly displayed on the autoclave.
8 13.3 The Validation Report The validation report must contain information on the temperature achieved for each probe and the time held along with the location of each probe. This is typically displayed in a table format along with a diagram of the chamber layout. Chamber set up and validation notes Channel Number Location Max Temp Min Temp Holding Time Comments 13.4 Frequency of validation Validation and calibration must be carried out at the frequencies shown below and at any other times when the previous test may no longer be valid (such as part of recommissioning after maintenance work). Waste containing Group 1 & 2 organisms Group 3 organisms at least annually six monthly The person responsible for the autoclave should check certificates and printouts received for the validation testing to satisfy themselves that the autoclave meets the criteria specified above for the required cycle(s). Validation records should be kept for 5 years. These will be inspected during audits by the Safety office and may be requested by Health and Safety Executive Inspectors during visits to the University Routine monitoring by users
9 Following validation, autoclave performance should be routinely monitored. Where autoclaves are used for waste containing pathogens, the autoclave should be fitted with chart recorders or print out facilities and these should be checked after each run and kept as part of the autoclave process record. Autoclave tape and indicators printed on bags do not show the load has been sterilised. These are used only to indicate that a load has been processed since the stripes appears very quickly even at low temperatures. Use of these would not be regarded as an adequate means of monitoring autoclave performance. Biological indicators are vials containing Geobacillus stearothermophilus which can be used to confirm that sterilisation is achieved. These are not a substitute for validation as described in section 13.1 above but are a useful means of confirming effective sterilisation in between formal validation cycles.
10 APPENDIX A Load Validation Form Autoclave Name and Serial Number Location Chamber (If applicable) Load Description / Ref Required cycle temperature and time Full Description of load at each point (note points 1 & 2 in are drain and load probe) Photo of load Notes:
Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:
Autoclave Safety Purpose: Sterilization refers to the complete killing of all living organisms, including spores. Common sterilization techniques include the application of wet heat, dry heat, chemicals,
EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use
Please post or circulate Fact heet nvironment, ealth and afety Information for the Berkeley Campus No. 33 Revised 04/04/11 afe and ffective Use of Autoclaves Autoclaves are easy to use but can pose a safety
Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.
Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam
Fundamental Autoclave Techniques
Fundamental Autoclave Techniques Version 3, September 2010 Veronika Tatarinoff GSBmE University of NSW Brandy Nelson University of Kentucky The following information is directed at aseptic/sterile surgical,
STANDARD OPERATING PROCEDURES Safe Autoclave Operations
Department of Environmental Health & Safety Biosafety & Biosecurity STANDARD OPERATING PROCEDURES Safe Autoclave Operations The purpose of this document is to provide standard operating procedures for
Arrangements for the Maintenance and Testing of Laboratory Autoclaves
Arrangements for the Maintenance and Testing of Laboratory Autoclaves Royal Free NHS Trust & Royal Free and UC Medical School (Hampstead Campus) This document has been produced by the Safety Office, Royal
KGI MEDICAL WASTE MANAGEMENT PLAN
1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:
Steam Sterilization Cycles for Lab Applications
Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain
AUTOCLAVE PROGRAM. SOP Bio-006 FOR THE USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE SOP
ENVIRONMENTAL AND EMERGENCY MANAGEMENT ENVIRONMENTAL HEALTH AND SAFETY 175 Cabot St. Wannalancit Suite 311 Lowell MA 01854 Ruth Medina, Ph.D. Tel: 978-934-2778 Senior Biosafety Specialist/Biosafety Officer
FACT SHEET : Using Autoclaves Safely
CSULA Environmental Health and Safety Biosafety Office FACT SHEET : Using Autoclaves Safely Most science research laboratories on campus require the use of autoclaves. The primary purpose of the autoclave
THE UNIVERSITY OF NEWCASTLE- SCHOOL of BIOMEDICAL SCIENCES
Page: 1 of 7 1. Purpose: 1.1. To describe the procedures to be used when dealing with chemical or microbiological spills. 2. Equipment: 2.1. Spill Kit 2.2. Miscellaneous items as listed 3. Materials: 3.1.
Decontamination and Waste Management www.biosecurity.sandia.gov
Decontamination and Waste Management www.biosecurity.sandia.gov SAND No. 2006-3684C Sandia is a multiprogram laboratory operated by Sandia Corporation, a Lockheed Martin Company, for the United States
Infectious Waste Management Plan
Infectious Waste Management Plan Infectious Waste Management Plan USC Health & Safety Programs Unit 777-5269 POLICY: A. In keeping with the University of South Carolina's policy of providing protection
Biosafety Level 2 (BSL-2) Safety Guidelines
BLS-4 Biosafety Level 2 (BSL-2) Safety Guidelines BSL-3 BSL-2 BSL-1 BSL-2 builds upon BSL-1. If you work in a lab that is designated a BSL-2, the microbes used pose moderate hazards to laboratory staff
A GUIDELINE FOR THE SAFE USE OF AUTOCLAVES UNIVERSITY OF OTTAWA. Environmental Health and Safety Service
A GUIDELINE FOR THE SAFE USE OF AUTOCLAVES UNIVERSITY OF OTTAWA Environmental Health and Safety Service Last date of revision July 9, 2003 TABLE OF CONTENT 1.0 PURPOSE...1 2.0 INTRODUCTION...1 3.0 PERSONAL
OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE
OSEH Occupational Safety & Environmental Health OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE Subject: Biohazardous (Medical) Waste Disposal Date: 08/19/09 Revision: 03 Page: 1 of 7 TABLE OF Section
Checklist for a Physical Containment Level 2 (PC2) Laboratory Facility
Checklist for a Physical Containment Level 2 (PC2) Laboratory Facility (The Gene Technology Regulator s Guidelines for Certification of Facilities/Physical Containment Requirements applies) Facility Name:.
UCCS Biosafety Management Policy
UCCS Biosafety Management Policy SAFE STORAGE AND HANDLING OF BIOLOGICAL LABORATORY WASTE (For assistance, please contact Environmental Health & Safety) It is the intention of the University of Colorado
Required Biosafety Level Three (BSL-3) Practices, Procedures, Facilities, and Safety Equipment For BSL-3 and BSL3/2 Laboratories
Required Biosafety Level Three (BSL-3) Practices, Procedures, Facilities, and Safety Equipment For BSL-3 and BSL3/2 Laboratories 1) LABORATORY OPERATIONS a) Responsibilities: A clear organization is set
Biological Safety Program
Risk Management & Safety Main Office, Wyoming Hall Phone: (307) 766-3277 Fax: (307)766-6116 Regulated Materials Management Center Phone: (307)766-3696 Fax: (307)766-3699 Web: www.uwyo.edu/ehs Email: [email protected]
PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3)
PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3) Yes No 1. Is access to the laboratory limited or restricted at the discretion of the laboratory director when experiments are in progress? Yes No
2.3. The management in each HCF shall be responsible for ensuring good waste management practices in their premises.
1. PURPOSE Health-care activities lead to production of medical waste that may lead to adverse health effects. Most of this waste is not more dangerous than regular household waste. However, some types
RISK ASSESMENT: HANDLING, TRANSPORTATION AND STORAGE OF LIQUID NITROGEN AND OTHER CRYOGENIC MATERIAL
RISK ASSESMENT: HANDLING, TRANSPORTATION AND STORAGE OF LIQUID NITROGEN AND OTHER CRYOGENIC MATERIAL More information can be found in the liquid nitrogen section of the departmental Health & Safety webpages:
THE SAFE DISPOSAL OF CLINICAL/DOMESTIC WASTE
Section V THE SAFE DISPOSAL OF CLINICAL/DOMESTIC WASTE The Trust is currently reviewing the requirements of the recent guidelines Health Technical Memorandum Safe Management of Healthcare Waste (HTML 07-01).
Autoclave Quality Control Policy
Northwestern University Autoclave Quality Control Policy Office of Research Safety Office of the Vice President for Research Oct 2010, Aug 2005, Aug 2004, May 2002 1 Table of Contents 1.0 INTRODUCTION
Laboratory Waste Disposal
Laboratory Waste Disposal 1. Purpose This guideline details the procedures to follow in disposing of hazardous waste that is generated in the laboratory in order to minimise risks associated with the disposal
Private Water Supplies Sampling Manual. A Field Guide
Private Water Supplies Sampling Manual A Field Guide Foreword This sampling manual details standard procedures for the collection, storage and transportation of samples at private water supplies in accordance
Waste Management. Course Description
Waste Management Course Description After completing this training course you will be familiar with the correct handling and disposal of waste materials generated through the provision of PoCT (Point of
SPILLS & SPILL KITS. Spills -General Guidelines:
Spills -General Guidelines: If handled properly, a spill may be nothing more than a nuisance. If handled improperly, a spill can seriously disrupt your activities and the work of your colleagues. At worst,
Waste Management Policy
Waste Management Policy Revised April 2013 1 Contents Page Content Page No. Clinical Waste 3 - The handling and disposal of Clinical and Soiled 3 - Policy 3 - Warning - The collection of Clinical Waste
STANDARD OPERATING PROCEDURES SPILL RESPONSE AND CLEAN-UP OUTSIDE BIOSAFETY CABINET
BIOLOGICAL SPILL KIT IN A 5 GALLON BUCKET WITH LID Spill response and cleanup procedures (SOP) 1 Notepad 1 Pen 6 Business cards 1 Permanent marker 1 trash bag 6 Biohazard stickers 1 roll duct tape 1 roll
North Carolina State University Emergency Facilities Closure Checklist- Part I
North Carolina State University Emergency Facilities Closure Checklist- Part I Unplanned or spontaneous events often disrupt daily operations on campus. In the event that an incident may interrupt your
3039878000or8009926372
3039878000or8009926372 bi s a l e s @me s a l a bs. c om Regulatory officials and sterilization experts have voiced concerns regarding the appropriateness of using a Biological Indicator (BI) Ampoule interchangeably
Laboratory Biosafty In Molecular Biology and its levels
Laboratory Biosafty In Molecular Biology and its levels Workshop 16-17 Oct..2012 Guidelines Does not mean optional Laboratory Biosafety The Laboratory Biosafety Manual is an important WHO publication
Medical or Biological Waste: Storage, Treatment, Disposal and Transportation Plan
Medical or Biological Waste: Storage, Treatment, Disposal and Transportation Plan 1. Scope This program covers all departments at Wellesley College who generate medical or biological waste to include Health
Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools
APPENDIX II Guide on the Segregation, Packaging, Labelling and Storage of Laboratory Chemical Wastes for Schools (Prepared by Environmental Protection Department) 1. Introduction The Waste Disposal (Chemical
Revision 5. Calvin College Medical Waste Management Plan. Date: Health and Safety
Calvin College Medical Waste Management Plan Prepared by: Date: Environmental 6/10/1 Health and Safety Approved By: Date: Revision 1.0 Policy The following medical waste management plan has been established
Effective Use of Autoclaves
Effective Use of s Overview An autoclave is a commonly used piece of equipment in microbiology laboratories. s (steam sterilizers) are metal pressure vessels that are used for steam sterilization of media,
UNIVERSITY OF RICHMOND REGULATED MEDICAL WASTE MANAGEMENT GUIDELINES
UNIVERSITY OF RICHMOND REGULATED MEDICAL WASTE MANAGEMENT GUIDELINES November 2003 Table of Contents Section Page I. Introduction.... 1 II. Characteristics of Regulated Medical Waste 1-2 III. Exclusions...2-3
Procedures for the Effective Use of Biological Safety Cabinets
Procedures for the Effective Use of Biological Safety Cabinets APPROVAL DATE BY BIOHAZARDS SUBCOMMITTEE: August 10, 2012 SUPERSEDES: March 1, 2012 This corporate guideline/procedure is intended as a minimum
Administrative Procedure
Administrative Procedure Number: 410 Effective 1/30/98 Supercedes: 1 of Subject: MEDICAL WASTE MANAGEMENT PLAN 1.0. PURPOSE: The purpose of the California State University Los Angeles Medical Waste Management
/ Clinical Waste & Offensive Waste Disposal Procedures
/ Clinical Waste & Offensive Waste Disposal Procedures Document Control Document Created by Last Updated by Shane McAteer 26/01/2011 Paul Monk 29/10/2013 1 Introduction This clinical and offensive waste
MEDICAL WASTE MANAGEMENT
MEDICAL WASTE MANAGEMENT Biological Safety INTRODUCTION PURPOSE Regulated medical waste is a designation for wastes that may contain pathogenic microorganisms which was previously termed infectious waste.
Permit Number: 7.0/048904/JT
POLLUTION PREVENTION AND CONTROL ACT 1999 ENVIRONMENTAL PERMITTING (ENGLAND & WALES) REGULATIONS 2010 (as amended) Permit Number: 7.0/048904/JT Installation Address: St Michaels Dry Cleaners 54 St Michaels
Biosafety Level 2 Criteria
Biosafety Level 2 Criteria Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5th Edition Biosafety Level 2 (BSL-2): Biosafety Level 2 builds upon BSL-1. BSL-2 is suitable for work involving
University of Colorado at Boulder
University of Colorado at Boulder Department of Public Safety Division of Environmental Health and Safety Environmental Health and Safety Center 1000 Regent Drive, 413 UCB Boulder, Colorado 80309-0413
b) Exhaust Selector and Sterilization Times... 26 c) Unloading the Autoclave... 27 Appendix 1 - Autoclave Logbook...28
Autoclave Program This Autoclave Program (User Guide and Maintenance Guide) was written in collaboration with the Autoclave Working Group, Faculty of Science, Biosafety Advisory Committee, Health, Safety
Appendix J IBC Biohazard Spill Management Plan
OVERVIEW Prevention is the most important part of any spill management plan. Be sure to read and understand standard operating procedures (SOP s) and protocols for safe manipulation of biohazards before
"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH
"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH REGULATED MEDICAL WASTE "Infectious medical wastes" includes solid
Biohazardous Waste Management Plan
Central Michigan University Biohazardous Waste Management Plan This document has been prepared to provide guidance to Central Michigan University (CMU) employees in the use and disposal of biohazardous
Biosafety Spill Response Guide
Yale University Office of Environmental Health & Safety Biosafety Spill Response Guide Office of Environmental Health & Safety 135 College Street, 1 st Floor, New Haven, CT 06510 Telephone: 203-785-3550
The following standard practices, safety equipment, and facility requirements apply to BSL-1:
Section IV Laboratory Biosafety Level Criteria The essential elements of the four biosafety levels for activities involving infectious microorganisms and laboratory animals are summarized in Table 1 of
SPECIAL MEDICAL WASTE PROGRAM
SPECIAL MEDICAL WASTE PROGRAM Department of Environmental Health & Safety Phone: (410) 704-2949 Fax: (410) 704-2993 Emergency: (410) 704-4444 Email: [email protected] Website: www.towson.edu/ehs/index.html
Policies. Prep Room Policies
Introduction INTRODUCTION The Microbiology Prep Room is located in 531A Life Sciences Building. The telephone number is 372-8609. It is open from 7:30 a.m. to 4:30 p.m. during the fall and spring semesters.
Biosafety Program University of Colorado. BSL-1 Audit Checklist
PI first name: PI last name: username: IBC application #: Inspection Date: Lab Location: Department: email: best lab contact lab contact email Inspected by: Time BSL-1 Yes BSL-2 Yes A. Contamination Control
JABSOM KAKA AKO STANDARDIZED AUTOCLAVE PROTOCOL
Page 1 JABSOM KAKA AKO STANDARDIZED AUTOCLAVE PROTOCOL GENERAL INFORMATION The autoclave uses high pressure and high temperature steam to effectively kill microorganisms or render a bio hazardous material
BOSTON COLLEGE ENVIRONMENTAL HEALTH AND SAFETY LABORATORY GUIDE FOR RENOVATIONS, REMODELS, MOVES AND TERMINATIONS
BOSTON COLLEGE ENVIRONMENTAL HEALTH AND SAFETY LABORATORY GUIDE FOR RENOVATIONS, REMODELS, MOVES AND TERMINATIONS I. Policy for Termination of Laboratory Use of Hazardous Materials The Principal Investigator
Biosafety Guidelines. 5. Pipetting Mechanical pipetting devices must be available and used. Mouth pipetting is prohibited.
Biosafety Guidelines Biosafety Level 1 Laboratory Biosafety Level 1 is suitable for work involving well-characterized agents not known to consistently cause disease in immunocompetent adult humans, and
Laboratory Line. Vertical and Bench-Top Life Science Autoclaves
Laboratory Line Vertical and Bench-Top Life Science Autoclaves The Advanced Laboratory Line The Full Spectrum of Laboratory Sterilizers Sterilization in a laboratory environment has its unique requirements.
University of Otago. Laboratory Biohazard Waste Disposal. Guidelines
University of Otago Laboratory Biohazard Waste Disposal Guidelines Approved by the IBSC on the 25 th January, 2011 1. Scope... 2 2. Definition... 3 3. Responsibilities... 4 4. Labelling... 4 5. Approved
Infection Control Manual - Section 8 Sharps & Clinical Waste. Infection Prevention Control Team
Title Document Type Document Number Version Number Approved by Infection Control Manual - Section 8 Sharps & Clinical Waste Policy IPCT001/10 3 rd Edition Infection Control Committee Issue date August
Laboratory- and sterilisation equipment HMC EUROPE
Laboratory- and sterilisation equipment HMC PROFILE HMC Laboratory- and sterilisation equipment Consistent marketing of technologies and the conversion of new ideas create product ranges, that are tailor-made
BSL 1 Laboratory Biosafety Manual
BSL 1 Laboratory Biosafety Manual Version 1.0 Idaho State University, Office for Research Institutional Biosafety Committee (IBC) 1651 Alvin Ricken Drive, Pocatello, ID 83201-8046 Phone: 208-282-2179 Fax:
MEDICAL WASTE DISPOSAL POLICY. EFFECTIVE DATE: January 31, 1991 REVISED DATE: April 1, 2008 I. AUTHORITY TO ESTABLISH DISPOSAL POLICY:
MEDICAL WASTE DISPOSAL POLICY EFFECTIVE DATE: January 31, 1991 REVISED DATE: April 1, 2008 I. AUTHORITY TO ESTABLISH DISPOSAL POLICY: The authority by which the Municipality of Anchorage, Solid Waste Services
CHAPTER V: DISPOSAL OF WASTES CONTAMINATED WITH INFECTIOUS AGENTS
CHAPTER V: DISPOSAL OF WASTES CONTAMINATED WITH INFECTIOUS AGENTS These biohazard waste disposal guidelines are designed to not only protect the public and the environment, but also laboratory and custodial
Top 10 considerations when validating an autoclave
Top 10 considerations when validating an autoclave Validating an autoclave is a daunting and time-consuming task. This white paper details the tricks, tips and traps to such a validation project from how
Carnegie Mellon University s Policy and Procedures for Recombinant and Synthetic Nucleic Acid Materials Spills
Carnegie Mellon University s Policy and Procedures for Recombinant and Synthetic Nucleic Acid Materials Spills Background In accordance with Section IV-B-2-b-(6) of the NIH Guidelines for Research Involving
To provide direction for the safe handling, administration and disposal of hazardous drugs.
Subsection: MEDICATION Related terms: Cytotoxic Drugs, Antineoplastic Drugs Authorized by: Clinical Directors CS-04-02-01 Page 1 of 9 Date Established: October 2006 Date For Review: September 2014 Dates
GEORGIA INSTITUTE OF TECHNOLOGY ENVIRONMENTAL HEALTH AND SAFETY PERSONAL PROTECTIVE EQUIPMENT
PURPOSE To protect the health and welfare of GEORGIA TECH employees in areas where there may be a risk of injury or exposure to hazardous substances or conditions employees who work in areas where physical
Standard Operating Procedure for Dacarbazine in Animals
Standard Operating Procedure for Dacarbazine in Animals 1. Health hazards Dacarbazine, also known as DTIC, DIC, and Imidazole Carboxamide, is an antineoplastic chemotherapy drug used in the treatment of
Protocol for Disinfection of Cell Culture and Tissue Culture in Media:
Protocol for Disinfection of Cell Culture and Tissue Culture in Media: Location: Hickory Hall 001 Director: Dr. Guido Verbeck DECONTAMINATION OF CELL CULTURE WASTE Cell culture has become a common laboratory
LABORATORY SAFETY INSPECTION CHECKLIST
LABORATORY SAFETY INSPECTION CHECKLIST As part of EH&S's laboratory safety survey program, we have compiled a list of the most common safety problems. This is not a list of every possible hazard, but can
PERSONAL PROTECTIVE EQUIPMENT (PPE)
Page 1 of 6 (PPE) Personal protective equipment (PPE) is used to protect an individual from hazards associated with their work tasks or environment. Specific types of personal protective equipment include
Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance
Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam
Beware that ordinary prescription glasses do not provide adequate protection. This is also true with poorly fitting safety glasses.
Ethidium Bromide Introduction Ethidium bromide (EtBr) is widely used for visualization of nucleic acids in electrophoretic gels. EtBr forms fluorescent complexes, by intercalation of DNA, which are readily
Sterilization methods and equipment Lab 1-2
Sterilization methods and equipment Lab 1-2 PHT 434 Sterilization Sterilization a process that by which all viable M.O are removed or destroyed, based on a probability function. Sterilization concept Sterilization
Laboratory- and sterilisation equipment HMC EUROPE
Laboratory- and sterilisation equipment HMC EUROPE PROFILE HMC EUROPE Laboratory- and sterilisation equipment Consistent marketing of technologies and the conversion of new ideas create product ranges,
Knowledge and Awareness of Proper Waste Disposal and Routine Biosafety Measures Among HealthCare Workers in Karachi, Pakistan (JUW-BEP-BSAP Project)
Knowledge and Awareness of Proper Waste Disposal and Routine Biosafety Measures Among HealthCare Workers in Karachi, Pakistan (JUW-BEP-BSAP Project) 11/01/2011 1 Shazia Tabassum Hakim 1,4, Syed Muhammad
Infection Control in the Dental Practice through Proper Sterilization
Midmark Corporation 60 Vista Drive P.O. Box 286 Versailles, Ohio 45380-0286 937-526-3662 midmark.com Infection Control in the Dental Practice through Proper Sterilization In today's world, the need for
MUNICIPALITY OF ANCHORAGE
MUNICIPALITY OF ANCHORAGE Solid Waste Services Department MEDICAL WASTE DISPOSAL POLICY EFFECTIVE DATE: January 31, 1991 REVISED DATE: November 1, 2013 I. AUTHORITY TO ESTABLISH DISPOSAL POLICY: The authority
STERILIZATION AND DISINFECTION
STERILIZATION AND DISINFECTION Importance of hand washing shown by Semmelweis STERILIZATION A physical or chemical process that destroys or eliminates all forms of microbial life, including spores. A satisfactory
THE UNIVERSITY OF MAINE BIOMEDICAL WASTE MANAGEMENT PLAN
THE UNIVERSITY OF MAINE BIOMEDICAL WASTE MANAGEMENT PLAN Department: The University of Maine Safety and Environmental Management Department Page i TABLE OF CONTENTS Section Title Page 1. Purpose... 1 2.
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA
STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1002.9 Date Issued: 10/98 Date Revised: 5/16 Page 1 of 5 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: Animal Care
Training on Standard Operating Procedures for Health Care Waste Management Swaziland 12 May, 2011
Training on Standard Operating Procedures for Health Care Waste Management Swaziland 12 May, 2011 Safe Infectious Waste Handling and Transport Objective Waste Overview Roles and Responsibilities of Waste
GUIDELINES FOR THE MANGAGEMENT OF BIO- MEDICAL WASTE AT THE SCHOOL OF LIFE SCIENCES
Basic Safety Rules SV vademecum Safety information Version 1.2 17.04.2013 GUIDELINES FOR THE MANGAGEMENT OF BIO- MEDICAL WASTE AT THE SCHOOL OF LIFE SCIENCES Table of contents 1. General Information...
Wine Cooler with Thermo-Electric Cooling & Heating Technology
Wine Cooler with Thermo-Electric Cooling & Heating Technology Model number: WC-0888H (8-bottles capacity) Instruction Manual Please read carefully and follow all safety rules and operating instructions.
DESERT RESEARCH INSTITUTE MOLD ASSESSMENT AND REMEDIATEION PROCEDURE
1.0 Introduction DESERT RESEARCH INSTITUTE This Desert Research Institute (DRI) Mold Assessment and Remediation Procedure was established to assure that consistent and uniform information is provided to
The University of Texas at San Antonio Office of Environmental Health, Safety and Risk Management. Part A. Biological Waste Management Safety Plan
The University of Texas at San Antonio Office of Environmental Health, Safety and Risk Management Part A Biological Waste Management Safety Plan i. SIGNATURE PAGE This Biological Waste Management Safety
Federal Wage System Job Grading Standard For Laboratory Working, 3511. Table of Contents
Federal Wage System Job Grading Standard For Laboratory Working, 3511 Table of Contents WORK COVERED... 2 WORK NOT COVERED...2 TITLES... 2 GRADE LEVELS... 2 NOTE TO USERS... 3 LABORATORY WORKER, GRADE
APPENDIX D INFECTIOUS WASTE MANAGEMENT GUIDE
APPENDIX D INFECTIOUS WASTE MANAGEMENT GUIDE TABLE OF CONTENTS INTRODUCTION... PAGE 1 SUMMARY OF REGULATIONS... PAGE 1 WRIGHT STATE UNIVERSITY INFECTIOUS WASTE STRATEGY... PAGE 5 PROCEDURES FOR WSU INFECTIOUS
INSTITUTIONAL POLICY AND PROCEDURE (IPP)
HOSPITAL NAME INSTITUTIONAL POLICY AND PROCEDURE (IPP) Department: Manual: Section: TITLE/DESCRIPTION POLICY NUMBER HAZARDOUS WASTE: HANDLING AND DISPOSAL EFFECTIVE DATE REVIEW DUE REPLACES NUMBER NO.
Each has the following responsibilities under this SOP:
Page.: 1 1. Purpose The Standard Operating Procedure (SOP) for sharps disposal was developed by the Department of Environmental Health & Safety in accordance with the University s Policy Statement on Health
Fact Sheet on Steam Sterilizers at Stanford University
Fact Sheet on Steam Sterilizers at Stanford University By: Environmental Quality and Water Efficiency Group, Stanford University; December 2013 Jennifer Fitch, Marty Laporte, Julia Nussbaum, Adam Kern,
FLAMMABLE LIQUIDS COMPLIANCE NOTE
Issued by: Heather Rice Page: 1 of 7 COMPLIANCE NOTE INTRODUCTION This document provides additional guidance to support the Arrangement on Flammable Liquids (Ref No. CYC/HS/A27) CONTENTS Section Topic
Fumigation of Safety Cabinets
Fumigation of Safety Cabinets General Microbiological safety cabinets should always be fumigated if a large spillage of infectious material occurs within them, before filters are changed or before any
UNH SAFETY AUDIT Office of Environmental Health and Safety
UNH SAFETY AUDIT Office of Environmental Health and Safety Administrative Information Customized for each location. Ensure contact information on door caution signs is up to date. Doors YES NO N/A 1. The
Response to Biological Spills in the Laboratory (Intentional or Accidental)
Response to Biological Spills in the Laboratory (Intentional or Accidental) Exposure Management For splash to eyes, mucous membranes, or broken area of the skin Irrigate eyes with clean water, saline or
Appendix H IBC Managing Biohazardous Waste SOP
Biohazardous waste is managed under the State of Rhode Island s Regulated Medical Waste Regulations (Regulation DEM-OWM-MW-1-2009, amended July, 2010). http://www.dem.ri.gov/pubs/regs/regs/waste/medwaste10.pdf
Bloodborne Pathogens Program Revised July, 5 2012
Bloodborne Pathogens Program Revised July, 5 2012 Page 1 of 16 Table of Contents 1.0 INTRODUCTION...3 1.1 Purpose...3 1.2 Policy.3 2.0 EXPOSURE CONTROL METHODS 4 2.1 Universal Precautions.4 2.2 Engineering