IMAGO7, B1 and Local SAR evaluation



Similar documents
IRCCS FONDAZIONE STELLA MARIS La Risonanza Magnetica a 7 Tesla della Fondazione Stella Maris: stato dell arte

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

FDA Guidelines for Magnetic Resonance Equipment Safety

SITE IMAGING MANUAL ACRIN 6698

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Basic Principles of Magnetic Resonance

5 Factors Affecting the Signal-to-Noise Ratio

RICHARD T ELL ASSOCIATES,INC.

Musculoskeletal MRI Technical Considerations

MRI SEQUENCES. 1 Gradient Echo Sequence

advancing magnetic resonance imaging and medical data analysis

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

Numerical Modelling of E-M Occupational Exposures associated with MRI

Operating Frequency Selection for Loosely Coupled Wireless Power Transfer Systems with Respect to RF Emissions and RF Exposure Requirements

MGH Adult Diffusion Data Scanning Protocols

GE Medical Systems Training in Partnership. Module 12: Spin Echo

Generation and Detection of NMR Signals

An 8-channel RF Transceive Coil For Parallel Imaging at 7T/298MHz

Breast MRI Quality Control

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES

mri : Physics For anyone who does not have a degree in physics Evert J Blink Application Specialist MRI

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

RF EXPOSURE LIMITS AND TESTING REQUIREMENTS

Effects of Mobile Phone Radiation onto Human Head with Variation of Holding Cheek and Tilt Positions

How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless)

Selected Radio Frequency Exposure Limits

Magnetic Resonance Imaging

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging II

Effect of Metallic Materials on SAR

Prepublication Requirements

Body SAR and EU. B. Derat

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

RECOMMENDATION ITU-R F (Question ITU-R 157/9) b) that systems using this mode of propagation are already in service for burst data transmission,

In vivo stem cell tracking in the myocardium of small laboratory animals: Feasibility and acquisition strategies on clinical magnetic resonance

32-Channel Head Coil Imaging at 3T

ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System

Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint

Human Exposure to Outdoor PLC System

September 26, 2000 IMST GmbH Carl-Friedrich-Gauß-Str. 2 D Kamp-Lintfort

NMR Techniques Applied to Mineral Oil, Water, and Ethanol

Official Journal of the European Communities. (Acts whose publication is not obligatory) COUNCIL

Preface Acknowledgments Acronyms

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9

ARIB TR-T V Specific Absorption Rate (SAR) requirements and regulations in different regions

SAR Distribution in Test Animals Exposed to RF Radiation

Magnetic Resonance Imaging

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

Evaluating Cell Phone and Personal Communications Equipment and their EMC Effects on Automotive Audio and In-Cabin Modules

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

RADIOFREQUENCY RADIATION, (RFR): (RFR Information - Technology Newsletter, Full Version)

COUNCIL RECOMMENDATION of 12 July 1999 on the limitation of exposure of the general public to electromagnetic fields (0 Hz to 300 GHz) (1999/519/EC)

Specific Absorption Rate (SAR) in the head of Tablet user's

Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Valve Prostheses and Annuloplasty Devices

PHY114 S11 Term Exam 3

Electromagnetic Field Safety

Kap 8 Image quality, signal, contrast and noise

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

Introduction to Magnetic Resonance Imaging Techniques

Accurate Multislice Gradient Echo T 1 Measurement in the Presence of Non-ideal RF Pulse Shape and RF Field Nonuniformity

An equivalent circuit of a loop antenna.

Numerical Methods and Software Tools for Simulation, Design, and Resonant Mode Analysis of Radio Frequency Birdcage Coils Used in MRI

MRI SAFETY JEOPARDY (Technologist Edition) Questions

6 J - vector electric current density (A/m2 )

FDTD Computation of Sar Distributions in Human Head for Mobile Phones V.Purushothaman, N.Vignesh, S.Vijayakumar

Specific Absorption Rate (SAR) Induced in Human Heads of Various Sizes When Using a Mobile Phone

MRI DATA PROCESSING. Compiled by: Nicolas F. Lori and Carlos Ferreira. Introduction

13C NMR Spectroscopy

Human Exposure Limits

THE ANSI/IEEE RF SAFETY STANDARD AND ITS RATIONALE. Om P. Gandhi and Gianluca Lazzi Department of Electrical Engineering Salt Lake City, Utah 84112

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

NMR and MRI. Seppo Vahasalo Philips Medical Systems MR Finland

510(k) Summary. This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.

Preserving Hyperpolarization of 129 Xe for Transport Submitted by: Zach Zytner Project Supervisor: Dr. Giles Santyr Medical Biophysics 3970 Date:

Automated Meter Reading Frequently Asked Questions. What is AMR?

Chapter 2 NMR in Inhomogeneous Fields

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Amplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B)

VBA Macro for construction of an EM 3D model of a tyre and part of the vehicle

runl I IUI%I/\L Magnetic Resonance Imaging

MRET-Shield Synopsis of Scientific Research Updated on June 1, 2007

REVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual

WIRELESS MAGNETIC CONTACT

Impedance Matching and Matching Networks. Valentin Todorow, December, 2009

Thermal Effects of Mobile Phones

Simulation of Mobile Phone Antenna Performance

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

Forum R.F.& Wireless, Milano il 14 Febbraio 2008 Dr. Emmanuel Leroux Technical Sales Manager for Italy

Real-life Applications of ICNIRP Guidelines to Various Human EMF Exposure Issues

MEASUREMENT AND ANALYSIS OF RF EME LEVELS FROM MOBILE TELEPHONE BASE STATIONS LOCATED AT LEICHHARDT, NSW

Phantom Test Guidance

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

Transcription:

Centro di ricercasullebiotecnologie di risonanza magnetica a campo ultra alto www.imago7.eu IMAGO7, the first Italian 7T MR scanner: B1 and Local SAR evaluation Gianluigi Tiberi INRIM, Febbraio 2014

THE IMAGO7RESEARCH FOUNDATION The Stella Maris Scientific Institute, Pisa IRCCS (Istituto di Ricerca e Cura a Carattere Scientifico in Developmental Neurology, Psychiatry and Rehabilitation, The University of Pisa Department of Physics Department of Chemistry The Pisa University General Hospital Department of Radiology Department of Neuroscience The MEDEA Scientific Institute, Bosisio Parini (Lecco) IRCCS (Istituto di Ricerca e Cura a Carattere Scientifico in Developmental Neurology, Psychiatry and Rehabilitation The CARIPI Foundation

Actual UHF MR worldwide distribution 16 15 16. 34 systems >= 7T 3

The evolution of MR images 1980 Zeugmatography 1974 PISA end of 90s end of 80s 15T 1.5 T 3 T 3 T 7 T

Delivery of the Magnet June 9, 2011

7 Tesla (& 7 Gauss) Reached on Nov 28 th 2011, 5pm NOTA: Il campo magnetico terrestre varia da poco più di 0,00002 Tesla all'equatore a 0,00007 Tesla nelle zone polari. A Milano è di circa 0,000047 Tesla. Questo significa che il campo ultra alto generato dal magnete è 150.000 volte più intenso!!!!

Imago7 Foundation Michela Tosetti IRCCS Stella Maris

Pisa, January 2012

1st Human on 30 March 2012!! ARC FSE (R=2.3) TE 92.6 ms TR 6 s 512x512 3 mm slice T2* weighted GRE 240/12.7/20 deg 1024x768 2Nex 4 mm slice Tx: shielded 16 element birdcage head coil; Rx: 32 array head coil (Nova Medical)

Physics of MRI Random orientation of the spins. Static field (~T) Excess magnetization along the direction of the static field. Linear varying fields Spatial information encoded d in different phase and frequency of spins. RF field in the transverse plane Magnetization rotated away from the equilibrium position RF field off Magnetization return to equilibrium with time scale T1 and T2*, both dependent from the sample, generating an RF field Receive coil on A signal is induced in the coil Read-out system and post-processing An image is formed

Hardware in MRI Main Magnet: static field Shimming i coils: improved static field homogeneity Gradient coils: Linear varying field, Fourier transform imaging RF coils: radiofrequency field Sample excitation Signal reception Read-out electronics

Magnetic Fields A. static (B0 + shims): ~ T B. db/dt (gradients): ~ mt/m C. radio frequency (B1): ~ T

RF coils (dedicated at 7T) Volume Coils ( T only or T/R) Solenoid Helmotz pair Saddle Coil Birdcage TEM Surface Coils (R only or T/R) Single loop Quadrature dual loop Butterfly coil Phased array

Two basic requirements must be fulfilled for obtaining high quality images: i) in the transmission mode, RF coils must be able to produce a uniform magnetic induction field B1 (in the transverse plane) in the volume of interest so that the nuclei can be properly excited; note that B1 μt T!!!! ii) in the receiving mode, a high signal to noise ratio (SNR) is needed, and the coil must be able to collect the signal emitted by the nuclei with a uniform amplitude throughout the volume of interest. Careful selection and design of the RF coil!! ARF coil can be seen an antenna which operates in the near- field region!!!

Larmor Frequency: = B 0 nuclues (MHz/T) f @ 1.5T f @ 3T f @ 7T 1 H 42.58 63.87 127.74 298.06 13 C 10.71 16.07 32.13 74.97 17 O 577 5.77 865 8.65 17.31 40.39 19 F 40.08 60.12 120.24 280.56 23 Na 11.26 16.89 33.78 78.82 31 P 17.23 25.85 51.69 120.61 129 Xe 11.77 17.65 35.31 82.39

Research activities at IMAGO7 Coils realization and testing Local SAR evaluation Human protocols and sequences development 1.5 T

E Local SAR: why bother? SAR 2 2 One of the most crucial safety issues during magnetic resonance (MR) acquisition iiti at ultra high field is the local Specific Absorption Rate (SAR), which describes the potential for heating, hence damaging, g, patient s tissue. The RF power required to obtain B1 μt can be equal to some kw!!!!

E Local SAR: why bother? SAR 2 2 During MR acquisition average SAR and Local SAR must be monitored MR systems can measure the B1 but not the E field!!! Thus, local SAR cannot be measured directly!!!

Local SAR: why bother? Routine SAR monitoring procedures rely on the measurement of reflected power at the transmitting coil port, however this information has two main limitations: 1. it does not discriminate among: o loss within the sample o ohmic losses of the coil o radiated power 2. it quantifies only the average SAR irrespectively of local spatial inhomogeneities 3. Impossible to push the performance (within regulatory limits!) because actual SAR is unknown

Field inhomogeneities increase when the frequency increases!! B1+ distribution for a head in an ideal birdcage coil at several frequencies Note: B1+ represents the RF B field in the rotating frame!

Field inhomogeneities increase when the dielectric constant of the sample increases oil water B1+ distribution at 7T when using an oil and water cylindrical phantom in a realistic birdcage coil B1+ inhomogeneities, thus: i) Image brightening!! ii) Local SAR variability!!

A possible approach: We derive the distribution of the local SAR by the combination of EM simulation (CST MW Suite and/or line source) and B1 MR acquisition obtained by the GE MR950 7T human system (GE HealthCare, Milwaukee, WI, USA)

Local SAR evaluation, human calf Element γ/2π (MHz/T) f 0 (MHz) 1 H 42.5756 298 31 P 17.2348 120.64 HUGO calf model Voxels size: 4x4x4mm^3 Realistic dielectric properties of tissues SAR 1H channel only 2 E 2

Local SAR, 10 g avg, at 298 MHz for unit RF cycle and 1W input power Simulations performed in collaboration with Dr. N. Fontana, Department of Information Engineering, Pisa

avg (B 1+ ) μt max SAR W/Kg Hugo, M, adult 0.52 0.68 Maps of B1+ magnitude [μt] and SAR [W/Kg] in the axial slice crossing the coil center obtained by using CST MW Suite with the human model HUGO. We noted that modification of the matching capacitor leads to a modification of the scale of both B1+ magnitude and SAR maps, but the shape of these maps will be not affected.

The previous results have been obtained using 1W on input power and CW avg (B + 1+ ) max SAR μt W/Kg Hugo, M, adult 0.52 0.68 In MR, the Flip Angle (FA) indicates the Magnetization rotation (away from the equilibrium position). Relation between B1+ and FA: /2 2 B a( t) dt /2, peak 1 In our system (i.e. sinc-shaped shaped pulse a(t) having a duration τ=3.2 ms), we have B 7.2 μt, peak 1 FA 90 2 1.5 1 This value can be used to scale the simulations!! 05 0.5 0-0.5-1 -0.01-0.008-0.006-0.004-0.002 0 0.002 0.004 0.006 0.008 0.01 t [second]

avg max avg(b 1+ ) for max SAR max SAR (B 1+ ) SAR FA=90 μt W/Kg μt for FA=90 CW W/Kg for FA=90 RF pulse W/Kg Hugo, M, adult 0.52 0.68 7.2 130.3 20.9 The fourth and the fifth columns show the average B1+ magnitude and the maximum of the local SAR after scaling the simulations so to achieve the B1+ slice average value of 7.2μT; note that such value leads to an (average) FA θ =90 (thus, this scaling is quite similar to what conventional autoprescan routines do). Scaling to others FAs can be performed The SAR values given in the fifth columns refer to CW (i.e. unit RF cycle). The correspondent SAR per unit FA θ =90 RF pulse is given in the sixth column, which appropriately accounts for the duty cycle of the pulse a(t). 2 1.5 1 0.5 0-0.5-1 -0.01-0.008-0.006-0.004-0.002 0 0.002 0.004 0.006 0.008 0.01 t [second]

avg max avg(b 1+ ) for max SAR max SAR (B 1+ ) SAR FA=90 μt W/Kg μt for FA=90 CW W/Kg for FA=90 RF pulse W/Kg Hugo, M, adult 0.52 0.68 7.2 130.3 20.9 Requirements: maximum local SAR in the extremities <20 W/Kg Averaging time equal to 6min has to be applied, as specified in [CEI EN 60601-2-33] Since the SAR depends on the characteristics of the sequence adopted during the MR exposure, all the parameters related to the sequence itself have to be taken into account in the calculation (number of RF pulses, FAs)

avg max avg(b 1+ ) for max SAR max SAR (B 1+ ) SAR FA=90 μt W/Kg μt for FA=90 CW W/Kg for FA=90 RF pulse W/Kg Hugo, M, adult 0.52 0.68 7.2 130.3 20.9 Example: GRE, FA=90, TE=8, TR=500 500, 512 512, Nex=1, Phase FoV=1, #slices=12 Nrf = Number of rf pulses for 90 =Phase Phase FoV Nex #slices = 6144 Sequence_scan_time= Phase FoV Phase TR =256 second 2 Sequence_duty_cycle= (τ*nrf Nrf)/ Sequence_scan_time It follows MAX SAR= 1.6 W/Kg 1.5 1 0.5 0-0.5-1 -0.25-0.2-0.15-0.1-0.0505 0 005 0.05 01 0.1 015 0.15 02 0.2 025 0.25 t [second]

In EM simulations, an appropriate power scaling must be performed to evaluate the SAR correctly Up to now, we performed such scaling on theoretical basis We can proceed for applying a scaling to B1+ scanner based measurements!! Measured B1+, expressed in Gauss 0.14 0 Simulated simulated B1+ obtained B+ [1W using power] 1 W, expressed in T x 10-6 50 0.12 50 7 6 100 0.1 100 5 0.08 150 4 150 0.06 0.04 200 0.02 200 250 3 2 1 250 300 50 100 150 200 250 150 100 50 0-50 -100-150 There is a good agreement between the two results!

0.14 0 simulated B+ [1W power] x 10-6 50 0.12 50 7 6 100 0.1 100 5 0.08 150 4 150 0.06 200 3 200 0.04 002 0.02 250 2 1 250 50 100 150 200 250 300 150 100 50 0-50 -100-150 measured B1+ map (performed using P,1 1 W Pin=0.04KW=40W, corresponding to TG=0) avg _ slice( B ) 27 2.7 T 1, measured input W avg max (B 1+ ) SAR μt W/Kg Hugo, M, adult 0.52 0.68 Scaling factor can calculated as: k avg _ slice( B ) 1, measured 1,1 WSIMULATED avg _ slice( B ) 5.2

Steps for local SAR monitoring Scaling for realistic operative conditions: 1. we 2. we 3. we we perform ab1+ mapping using different FAs, i.e. FA=90 we extract the average on a slice, i.e. avg slice B 1, measured FA 90 _ ( ) we use this value for scaling the simulations, following the procedure described before Simulations can be performed and stored; it would be preferable to perform simulations using many different anatomical models so to allow a good match with the patient

We conclude showing some pictures and images..

HIPPOCAMPUS

Research people Michela Tosetti physicist IRCCS Stella Maris Mirco Cosottini Neuroradiologist University of Pisa Daniela Frosini Neurologist University of Pisa Ilaria Pesaresi Neuroradiologist Pisa General Hospital Massimo Del Sarto Engineer IRCCS Stella Maris Laura Biagi Physicist IRCCS Stella Maris Mauro Costagli Engineer IMAGO7 Gianluigi Tiberi Engineer IMAGO7 Eleonora Maggioni Engineer IRCCS Md Medea Luca Cecchetti Psychologist University of Pisa Danilo Scelfo Physicis IRCCS Stella Maris and University of Pisa Alessandra Retico Physicist INFN Riccardo Stara Physicist University of Pisa GE Applied Science Laboratory EUROPE Marie Curie HiMR PhD Program Nunzia Fontana Engineer University of Pisa Alessandra Toncelli Physicist Univesity of Pisa Maria Evelina Fantacci Physicist Univesity of Pisa Valentina Domenici Chemist University of Pisa Agostino Monorchio Engineer University of Pisa Marcello Alecci Physicist Univesity of Aquila James Tropp, physicist GE ASL USA Doug Kelley GE ASL 7T UCSF USA