Lasers and Specturbility of Light Scattering Liquids - A Review



Similar documents
Evaluation of the image quality in computed tomography: different phantoms

Using light scattering method to find The surface tension of water

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND

SUITABILITY OF RELATIVE HUMIDITY AS AN ESTIMATOR OF LEAF WETNESS DURATION

DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR

DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY

Fiber Optics: Fiber Basics

DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION

Will Your Fiber Optic Cable Plant Support Gigabit Ethernet?

INTRODUCTION FIGURE 1 1. Cosmic Rays. Gamma Rays. X-Rays. Ultraviolet Violet Blue Green Yellow Orange Red Infrared. Ultraviolet.

P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)

Experiment #5: Qualitative Absorption Spectroscopy

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Continuing Medical Education Program: a credit system evaluation

Fiber Optic Specifications

Optical Communications

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

FIBER OPTIC ISOLATORS

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

5-Axis Test-Piece Influence of Machining Position

DIODE PUMPED CRYSTALASER

104 Practice Exam 2-3/21/02

ENERGETIC EFFICIENCY OF AN AGRICULTURAL TRACTOR IN FUNCTION OF TIRE INFLATION PRESSURE

Diffraction of Laser Light

PIPELINE LEAKAGE DETECTION USING FIBER-OPTIC DISTRIBUTED STRAIN AND TEMPERATURE SENSORS WHITE PAPER

Temperature. LumaSense Technologies, Inc. supplies a unique selection of very precise calibration

Diffraction of a Circular Aperture

NEAR FIELD OPTICAL MICROSCOPY AND SPECTROSCOPY WITH STM AND AFM PROBES

Optical Fiber Switches

MICROSTRUCTURAL AND MECHANICAL CHARACTERIZATION OF GRAY CAST IRON AND AlSi ALLOY AFTER LASER BEAM HARDENING

Module 13 : Measurements on Fiber Optic Systems

Polarization of Light

Lab 9: The Acousto-Optic Effect

SENSORS. Miniature Sensors - S3Z. Advanced line of miniature Asian style of photoelectric sensors mm background suppression

AS COMPETITION PAPER 2008

Digital Television Broadcasting Systems Testing in Brazil

View of ΣIGMA TM (Ref. 1)

Single Mode Fiber Lasers

OPTICAL FIBERS INTRODUCTION

HYBRID INTELLIGENT SUITE FOR DECISION SUPPORT IN SUGARCANE HARVEST

Determination of the Effective Energy in X-rays Standard Beams, Mammography Level

Infrared Viewers. Manual

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL 1

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

DOE Solar Energy Technologies Program Peer Review. Denver, Colorado April 17-19, 2007

XXIX ENFMC - Annals of Optics D shadowgram projection using a simple diffractive screen José J. Lunazzi, Noemí I. Rivera Rodríguez.

It has long been a goal to achieve higher spatial resolution in optical imaging and

Experiment 5. Lasers and laser mode structure

Smart Electromagnetic Flowmeter Open channel Flowmeter Detector

Fraunhofer Diffraction

CREOL, College of Optics & Photonics, University of Central Florida

Measuring Laser Power and Energy Output

v = fλ PROGRESSIVE WAVES 1 Candidates should be able to :

Measuring of optical output and attenuation

Accurately Testing fibre Optic Cables

Alignement of a ring cavity laser

Interference. Physics 102 Workshop #3. General Instructions

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Scanning Near Field Optical Microscopy: Principle, Instrumentation and Applications

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

PROVE, the next generation registration metrology tool, status report

Harmonic oscillations of spiral springs Springs linked in parallel and in series

arxiv:cond-mat/ v1 [cond-mat.soft] 25 Aug 2003

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

SINGLEMODE OR MULTIMODE FIBER OPTIC PATCHCORDS

Explanation of Reflection Features in Optical Fiber as Sometimes Observed in OTDR Measurement Traces

SpectraTec II. Polarized Multi-Laser Source BLUE SKY RESEARCH WAVELENGTHS. The SpectraTec II

Raman Spectroscopy Basics

Acousto-optic modulator

Owner s Manual

Using the Spectrophotometer

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Beam shaping to generate uniform Laser Light Sheet and Linear Laser Spots

ArcHC_3D research case studies (FCT:PTDC/AUR/66476/2006) Casos de estudo do projecto ArcHC_3D (FCT:PTDC/AUR/66476/2006)

Theremino System Theremino Spectrometer Technology

Example of SOP for IR Laser

Lecture 3: Fibre Optics

Asian Journal of Food and Agro-Industry ISSN Available online at

Color holographic 3D display unit with aperture field division

Optical coherence tomography as a tool for non destructive quality control of multi-layered foils

Near-field scanning optical microscopy (SNOM)

ESCOAMENTO DO AR EM ARMAZÉM GRANELEIRO DE GRANDE PORTE COM SISTEMA DE AERAÇÃO SIMULATION OF AIR FLOW IN LARGE AERATED GRAIN STORAGE

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

L-LAS-TB-CL serie. laser light curtains for inline measuring tasks

Optical laser beam scanner lens relay system

A Guide to Acousto-Optic Modulators

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab

Transcription:

Artigo Original Revista Brasileira de Física Médica.2011;5(1):57-62. Reproducibility of radiant energy measurements inside light scattering liquids Reprodutibilidade de medidas de energia radiante dentro de líquidos espalhadores de luz Ana C. de Magalhães and Elisabeth M. Yoshimura Institute of Physics of São Paulo University/Nuclear Physics Department, São Paulo (SP), Brazil. Abstract The aim of this project is to evaluate the uncertainty associated with measurements performed with laser beams in scattering liquids with optical fibers. Two lasers with different wavelengths were used, 632.8 nm and 820 nm, to illuminate a cuvette with Lipovenos PLR, the scattering liquid. A mask at the top of the cuvette was used to control the positioning of 250 µm optical fiber and the laser entrance point. The light energy was measured with an optical power meter, and the integration time was 60 s, measured with a chronometer. There were no systematic errors associated with the integration time. Two tests were done to evaluate the uncertainty associated with the positioning of the components of the experimental arrangement. To evaluate the uncertainty of the positioning of the cuvette, seven series of measurements with the optical fiber 5 mm far from the beam were performed. Between two series, the cuvette was removed from the holder, the liquid was mixed and the cuvette was put back in the same position. A second test was done to evaluate the reproducibility of fiber positioning and the relative positioning of mask and cuvette. Three distances between the beam and the fiber were used: 4 mm, 5 mm and 6 mm, through the positioning of the fiber at three different holes in the mask. Six series of measurements were performed. Between two series, the cuvette was removed from the holder, the liquid was mixed and the cuvette was put back in the same position. Each series was done in a different order, permuting the three positions. The uncertainty associated with the positioning of the experimental elements was of the order of 7%. The variation of incident laser energy was also evaluated, resulting in 6.5% and 4.0% for the red and infrared lasers respectively. Keywords: laser, scattering liquids, light energy. Resumo O objetivo deste projeto é avaliar a incerteza associada às medidas realizadas com feixes de laser em líquidos espalhadores com fibras ópticas. Dois lasers com diferentes comprimentos de onda foram utilizados, 632,8 e 820 nm, para iluminar uma cubeta com Lipovenos PLR, o líquido espalhador. Uma máscara no topo da cubeta foi utilizada para controlar o posicionamento da fibra óptica de 250 µm e o ponto de entrada do laser. A energia da luz foi medida com um medidor de potência óptica (Optical Meter), o tempo de integração foi de 60 segundos, medido com um cronômetro. Não houve erros sistemáticos associados com o tempo de integração. Dois testes foram realizados para avaliar a incerteza associada com o posicionamento dos componentes do arranjo experimental. Para avaliar a incerteza do posicionamento da cubeta, foram realizadas sete séries de medidas com fibra óptica a 5 mm de distância do feixe. Entre duas séries, a cubeta foi removida do recipiente, o líquido foi misturado e a cubeta foi colocada na mesma posição novamente. Um segundo teste foi realizado para avaliar a reprodutibilidade do posicionamento da fibra e o relativo posicionamento da máscara e da cubeta. Três distâncias entre o feixe e a fibra foram utilizadas: 4, 5 e 6 mm, por meio do posicionamento da fibra em três diferentes orifícios na máscara. Seis séries de medidas foram realizadas. Entre duas séries, a cubeta foi removida do recipiente, o líquido foi misturado e a cubeta foi posta novamente na mesma posição. Cada série foi realizada em uma ordem diferente, mudando as três posições. A incerteza associada com o posicionamento dos elementos experimentais foi da ordem de 7%. A variação da energia de laser incidente também foi avaliada, resultando em 6,5 e 4,0% para os lasers vermelho e infravermelho, respectivamente. Palavras-chave: laser, líquidos espalhadores, energia da luz. Introduction The determination of the uncertainty associated with an experimental arrangement is very important. It determines the reliability of the experiment. In our experiment, with the uncertainty, is possible to distinguish measurements and to investigate variations of light energy inside an illuminated volume. The aim of this project was to evaluate the reproducibility of light energy measurements on scattering liquids with the use of an optical fiber coupled in an optical meter, in order to determine the uncertainty associated to the positioning of the experimental components and analyze the main uncertainty components. Corresponding author: Ana Carolina de Magalhães Instituto de Física Rua do Matão, Travessa R, 187 Cidade Universitária São Paulo (SP), Brazil CEP: 05508-090 E-mail: anamagalhaes@usp.br Associação Brasileira de Física Médica 57

Magalhães AC, Yoshimura EM Materials and methods A cylindrical cuvette was used, with 25 mm diameter, with a plane window with 10 mm of width, parallel to the cylinder axis, as it is possible to see at Figure 1-A and 1-B. Lipovenos PLR, which is a lipid emulsion for intravenous use and has scattering behavior 1, was used as scattering liquid. An optical fi ber with a diameter of 250 μm was used to collect the scattered light. The fi ber was guided by a metallic tube (internal diameter of 0.85 mm, external diameter of 1.5 mm and 47.65 mm length) in order to make sure that the fi ber was vertical and that its entering position in the liquid could be measured. The fi ber depth in the liquid was 5 mm and this portion of fi ber was without protective cover. A mask, detailed in Figure 1-C, was used in order to allow the controlled positioning of the optical fi ber in relation to the laser beam. The central hole, for the laser beam, has a diameter of 3 mm, and each one of the other holes, for the optical fi ber, has a diameter of 1.5 mm. The fi rst hole is at a distance of 4 mm of the center, the second is at 5 mm and so on, until the last one that is at 11 mm of the center. The fi nal experimental arrangement is represented at Figure 2. Lasers with two different wavelengths were used, one in red range (He-Ne, 632.8 nm) and the other in infrared range, 820 nm. The circular beams reached the center of the cuvette, passing through the mask central hole. In the case of infrared laser, an aperture was used to ensure that the laser beam reaching the liquid was circular. An optical meter (Newport Hand-Held Optical meter, model 1918-C) with a connector to optical fi bers was used in integration mode, and the signal was integrated in 60 s. This measurement was taken with a chronometer, as the optical meter doesn t have this function. In order to determine the uncertainty associated with the positioning of the cuvette and the optical fi ber, two reproducibility tests were carried out. The objective of the fi rst one (R1) was to evaluate the reproducibility of the positioning of the cuvette in the holder for illumination. Seven series of seven measurements with the optical fi - ber 5 mm far from the beam were performed. Between two series, the cuvette was removed from the holder, the liquid was mixed and the cuvette was put back in the same position. The objective of the second test (R2) was to evaluate the reproducibility of the optical fi ber positioning and the relative positioning of mask and cuvette. Three distances between the beam and the fi ber were used: 4 mm, 5 mm and 6 mm, through the positioning of the fi ber at three different holes in the mask. Six series of seven measurements with the fi ber at each of the three positions were performed. Between two series, the cuvette was removed from the holder, the liquid was mixed and the cuvette was put back in the same position. Each series was composed by a different permutation of the three positions. Data of the incident energy from the lasers were collected in order to determine the stability of this value. These measurements were done with the same optical meter, but without the optical fi ber attachment, in order to determine how the lasers, the mask and the aperture affect the results obtained. For red laser, measurements were taken at two different positions: position 1 - before the mask; and position 2 - behind the mask. For infrared laser, measurements at three different positions were taken: position 0 - before the aperture; position 1 - between the aperture and the mask; and position 2 - behind the mask. The positions 1 and 2 are the same for both lasers. 50 mm B C A Figure 1. Figure of the cuvette used at the experiment. A) Threedimensional view; and B) cross-section view. At C, is shown the mask used at the experiment to position the optical fi ber and the laser beam. Figure 2. Figure of the experimental arrangement used. The laser is represented by the red line. The optical fi ber is represented by the green line, the fi ber depth in the liquid is 5 mm. The mask used is represented in A, it has a central hole for the laser beam and eight radial holes for the fi ber. 58 Revista Brasileira de Física Médica.2011;5(1):57-62.

Reproducibility of radiant energy measurements inside light scattering liquids Results He-Ne laser The results obtained for the wavelength 632.8 nm for the test R1 are represented at the graphs of Figures 3 and 4. The data sequence for test R2 is shown at the graph of Figure 5, in which it is possible to see a change at the data for all distances, because of the same reasons of test R1. At Table 1 are the values for average, standard deviation and coeffi cient of variation for each test. The results for the measurements of laser incident energy are shown at Figure 6. The energy is relative to the average of the measurements of each position (position 1 and 2). There are no irregularities at the data, showing that the laser output is stable. Results for average, standard deviation, coeffi cient of variation and sample size are shown at Table 1. Figure 5. Graph with data for test R2, red laser. The graph shows data in chronological order for each distance from the fi ber to the center of the cuvette. 380 370 Energy (nj) 360 350 340 330 59,85 59,90 59,95 60,00 60,05 60,10 60,15 60,20 Integration time (s) Figure 3. Graph with data for test R1, red laser. The graph shows integration time dispersion data, measured with manual chronometer. Energy (nj) 380 370 360 350 340 330 0 7 14 21 28 35 42 49 Data number Figure 4. Graph with data for test R1, red laser. The graph shows data in chronological order. The arrows correspond to the change of measure series. Figure 6. Graph with data for laser incident energy, red laser. The energy is relative to the average of the measurements of each position. Table 1. Average, standard deviation and coeffi cient of variation of the radiant energy for each test (R1 and R2) realized with red laser and data for laser incident energy at positions 1 and 2 (lines Pos 1 and Pos 2, respectively). Test/ distance Average (nj) Standard deviation (nj) Coeffi cient of variation (%) Sample size R1 353 10 2.9 49 R2/4 mm 475 24 5.0 42 R2/5 mm 400 30 7.6 42 R2/6 mm 229 9 3.9 42 Pos 1 333 10 6 7 10 6 2.2 11 Pos 2 362 10 6 23 10 6 6.5 10 Revista Brasileira de Física Médica.2011;5(1):57-62. 59

Magalhães AC, Yoshimura EM Infrared laser The data obtained for test R1, in chronological order, are shown at the graph of Figure 7. It is possible to note, at graph of Figure 7, that data oscillate a lot, without any clear change, neither tendencies. For test R2, the results are shown at the graph of Figure 8, in which is possible to note that there is a big change at energy for all distances after the seventh measurement. The reason for this alteration is a variation of the incident laser energy, which is changed by a factor of 2.6. After this big change, other variations with this magnitude did not occur. Table 2. Average, standard deviation and coeffi cient of variation of the radiant energy for each test (R1 and R2) realized with infrared laser and data for laser incident energy at positions 0, 1 and 2 (lines Pos 0, Pos 1 and Pos 2, respectively). Test/ distance Average (nj) Standard deviation (nj) Coeffi cient of variation (%) Sample Size R1 88.4 2.1 2.3 49 R2/4 mm 196.8 3.4 1.8 21 R2/5 mm 152.5 8.0 5.2 35 R2/6 mm 113.9 3.6 3.2 35 Pos 0 262.2 10 6 6.1 10 6 2.3 11 Pos 1 89.3 10 6 3.4 10 6 4.0 11 Pos 2 81.3 10 6 0.61 10 6 0.75 11 The results for the measurements of laser incident energy are shown at Figure 9. The energy is relative to the average of the measurements of each position (position 0, 1 and 2). There are no irregularities at the data, what means that there are not big variations of incident energy and there is not any important effect of the aperture and the mask. Results for average, standard deviation, coeffi cient of variation and sample size are shown at Table 2. Figure 7. Graph with data for test R1, infrared laser. The graph shows data in chronological order. Figure 9. Graph with data for laser incident energy, infrared laser. The energy is relative to the average of the measurements of each position. Figure 8. Graph with data for test R2, infrared laser. The graph shows data in chronological order for each distance measured. At Table 2 are the values for average, standard deviation and coeffi cient of variation for each test. Probably, the low standard deviation obtained for test R2, at distance of 4 mm, occurred because three series of measurements at this position were lost. Conclusions Through the analysis of the data, it was concluded that there are no tendencies related to the integration time, what is seen at integration time dispersion graphs. Therefore, it is possible to say that this source of error is considered when the energy analysis is done. There is an uncertainty associated to the optical fi - ber and cuvette positioning, shown by the standard 60 Revista Brasileira de Física Médica.2011;5(1):57-62.

Reproducibility of radiant energy measurements inside light scattering liquids deviation obtained for each group. As they have similar magnitudes, varying roughly between 2 and 8%, we will adopt 7% as a value of the uncertainty associated with the data collection. It is equivalent to the largest standard deviation found for the series data. A large portion of the radiant energy data variation is due to the laser stability, as the coefficients of variation of the incident energy are similar to the standard deviation of the reproducibility experiments. Acknowledgment The authors thank the personnel of the Optical Laboratory of Institute of Physics, where the measurements were done. References 1. Pogue BW, Patterson MS. Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry. J Biomed Opt. 2006;11(4):041102. Revista Brasileira de Física Médica.2011;5(1):57-62. 61