Determination of Benzodiazepines in Oral Fluid Using LC-MS-MS



Similar documents
The potential of oral fluid as a specimen in various drug testing programs

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Determination of Tapentadol and its Metabolite N

Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

AppNote 6/2011. Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS KEYWORDS ABSTRACT

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Interlaboratory Studies on the Analysis of Hair for Drugs of Abuse: Results from the Fourth Exercise

LC-MS/MS for Chromatographers

Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

CORESTA RECOMMENDED METHOD N 72

Opiates in Urine by SAMHSA GC/MS

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

How To Test For Contamination In Large Volume Water

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Overview. Purpose. Methods. Results

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

Drug Screening of Preserved Oral Fluid by Liquid Chromatography Tandem Mass Spectrometry

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Guidance for Industry

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

Oasis HLB Cartridges and 96-Well Plates

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Urine Drug Testing. Why drug test? Set the Standard!

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Supplementary Materials and Methods (Metabolomics analysis)

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING *

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow

CUT-OFF CONCENTRATIONS FOR DRUGS OF ABUSE IN SALIVA FOR DUI, DWI OR OTHER DRIVING-RELATED CRIMES

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

What Does the Drug Test Tell Us

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

Matrix Effect in Bio-Analysis of Illicit Drugs with LC-MS/MS: Influence of Ionization Type, Sample Preparation, and Biofluid

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Toxicology CPT Code Changes for 2016

APPLICATIONS MANUAL. Ifosfamide in blood serum Ingredients in blood serum Organophosphorus pesticides in tea leaf... 5

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Scientific Working Group for Forensic Toxicology (SWGTOX) Standard Practices for Method Validation in Forensic Toxicology

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

AMD Analysis & Technology AG

Ádám Tölgyesi Joerg Stroka. Report EUR EN

Patients with COPD were recruited according to the inclusion criteria; aged 40-75, current or

United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science

Application of Structure-Based LC/MS Database Management for Forensic Analysis

Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

A commitment to quality and continuous improvement

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

SOLID PHASE EXTRACTION APPLICATIONS MANUAL

FREQUENTLY ASKED QUESTIONS Oral Fluid Drug Testing with Oral-Eze

Transcription:

Determination of Benzodiazepines in Oral Fluid Using LC-MS-MS Technical Note I Christine Moore 1,', Cynthia Coulter 1, Katherine Crompton 1, and Michael Zumwalt 2 l lmmunalysis Corporation, 829 Towne Center Drive, Pomona, California 91767; and 2Agilent Technologies, 9780 S. Meridian Boulevard, Englewood, Colorado 80112 [Abstract I A rapid, simple, highly sensitive procedure for the simultaneous analysis of bromazepam, alprazolam, clonazepam, Iorazepam, oxazepam, diazepam, midazolam, flurazepam, flunitrazepam, nordiazepam, triazolam, temazepam, nitrazepam, and chlordiazepoxide in oral fluid is described, using liquid chromatography coupled to a triple quadrupole mass spectrometer in positive electrospray mode. Benzodiazepines in oral fluid samples (1 ml) were analyzed using the Quantisal TM collection device, quantified using solid-phase extraction, and detected using liquid chromatography with tandem mass spectrometric detection. For confirmation, two transitions were monitored and one ion ratio determined, which was within 20% of the ratio for known calibration standards. The limits of quantitation ranged from 0.5-5 ng/ml of neat oral fluid; the intraday precision of the assays (n = 5) ranged from 2.8-7.29%; and the interday precision ranged from 1.42-6.8% (n = 5). The percentage recovery of the drugs from the collection pads ranged from a low of 81.4% for midazolam to the highest of 90.17% for nitrazepam. The procedure can be applied to authentic oral fluid specimens. Introduction Benzodiazepines are the most commonly prescribed class of drugs in the USA for the treatment of anxiety and insomnia, particularly in the elderly (1). They are also used as muscle-relaxants and anti-convulsants. They are often detected in incidents of driving under the influence of drugs (DUID) and in combination with other medications (2,3). Oral fluid is becoming increasingly used as a specimen in many areas of forensic and clinical interest, including collection at the roadside during traffic stops. Its ease of collection, difficulty of adulteration, and applicability to routine testing have promoted its use as a valid test specimen. However, the detection of benzodiazepines is not without difficulty, particularly in oral fluid, because the saliva-plasma ratio for most of the drug class is low. *Author to whom correspondence should be addressed. Email: cmoore@immunalysis.com. One of the main issues with the quantitation of drugs in oral fluid is the difficulty of collection in terms of specimen volume. Many of the currently available devices do not give an indication of how much oral fluid is collected, thereby rendering any quantitative results questionable without further manipulation in the laboratory (4,5). Furthermore, devices incorporating a pad or material for the saliva collection do not always indicate how much of each drug is recovered from the pad before analysis, again calling into question any quantitative result. The drug concentration reported is dependent on the collection procedure used (6). This work employed the QuantisaV M oral fluid collection device, which collects a known amount of neat oral fluid. The efficiency of recovery of the benzodiazepines from the collection pad into the transportation buffer was determined in order to increase confidence in the quantitative value. Several publications have addressed the issue of the analysis of benzodiazepines in oral fluid. Quintela et al. (7) determined 9 benzodiazepines in neat oral fluid using a liquid chromatographic method with mass spectrometric detection (LC-MS). They included lormetazepam and tetrazepam, which were not in our profile; however, clonazepam, chlordiazepoxide, nordiazepam, temazepam, oxazepam, flurazepam, and nitrazepam were not included. A recent publication from Oiestad et al. reported the screening of oral fluid using tandem LC-MS for several drugs, including benzodiazepines (8). They analyzed fenazepam and some benzodiazepine metabolites, which we did not include (see later), but they did not include the commonly prescribed drugs triazolam, temazepam, midazolam, flurazepam, or chlordiazepoxide. Smink et al. (9) analyzed urine and oral fluid for 33 benzodiazepines using liquid chromatography with tandem mass spectrometric detection (LC-MS-MS). With the exception of diazepam, where a limit of quantitation (LOQ) of 0 ng/ml was reported, the lower LOQ for the other analytes was significantly higher than in our application. In our research, we did not include the metabolites such as 7-aminoflunitrazepam, 7-aminoclonazepam, 7-aminonitrazepam, a-hydroxy alprazolam, (~-hydroxytriazolam, or desalkyiflurazepam because the parent drug is found more often in higher concentration than metabolites in oral fluid. We did, 596 Reproduction (photocopying) of editorial content of this journal is prohibited without publisher's permission.

however, include metabolites such as nordiazepam, temazepam, oxazepam, and Iorazepam (a metabolite of lormetazepam), because they can be prescribed as individual drugs. Methods and Materials Oral fluid collection devices Quantisal devices for the collection of oral fluid specimens were obtained from Immunalysis Corporation (Pomona, CA). The devices contain a collection pad with a volume adequacy indicator, which turns blue when 1 ml of oral fluid ( 10%) has been collected. The pad is then placed into a transport buffer (3 ml), allowing a total specimen volume available for the analysis of 4 ml (3 ml buffer + 1 ml oral fluid). This is specifically advantageous in cases where the specimen is positive for more than one drug and the volume of specimen available for analysis may be an issue. The oral fluid concentration is diluted 1:3 when using Quantisal collection devices, and detected drug concentrations were adjusted accordingly. Standards and reagents Deuterated internal standards: DS-diazepam, D5- temazepam, D5-alprazolam, D5-oxazepam, and D4-clonazepam, as well as unlabelled drug standards: bromazepam; clonazepam; nitrazepam; triazolam; alprazolam; flunitrazepam; flurazepam; lorazepam; midazolam; chlordiazepoxide; diazepam; oxazepam; nordiazepam; temazepam were purchased from Cerilliant (Round Rock, TX). Mixedmode solid-phase extraction columns (CSDAU020) were purchased from United Chemical Technologies (Bristol, PA). All solvents were of HPLC grade or better; all reagents were ACS grade and purchased from Spectrum Chemical (Gardena, CA). Calibrators and controls Calibration standards and controls were prepared from synthetic oral fluid, which comprised 25mM phosphate buffered saline (ph 7.0), 30ram sodium bicarbonate, 0.1% albumin, amylase, and 0.1% Proclin 300 as a preservative. The calibrators and controls prepared at concentrations of 0.5, 1, 5, 10, 20, and 40 ng/ml were then diluted by four with Quantisal transportation buffer. Throughout the development of the assay, multiple Quantisal collection devices were selected from different lots. The drug concentration used to fortify the synthetic oral fluid was adjusted according to the dilution factor for all calibration standards and controls. In this way, the final result obtained from the instrument did not need to be re-calculated for dilution factors. For each analysis, a six-point calibration cuwe (0.5, 1, 5, 10, 20, and 40 ng/ml of neat oral fluid; actually equivalent to 0.125, 0.25, 1.25, 2.5, 5, and 10 ng being measured) was run with each batch; the internal standard concentration was 40 ng/ml. The corresponding deuterated internal standard was used for the quantitation of clonazepam, oxazepam, diazepam, and temazepam; for all other benzodiazepines, d5-alprazolam was used as the internal standard. Extraction procedure In order to improve sensitivity and minimize matrix or ion suppression effects, the specimens were extracted prior to analysis. Sodium phosphate buffer (0.1M, ph 6.0, 1 ml) was added to the Quantisal buffer (lml) and the samples were mixed. Extraction tubes were placed onto the vacuum manifold and conditioned with methanol (3 ml), deionized water (3 ml), and 0.1M phosphate buffer (ph 6.0, 2 ml). The column bed was not allowed to dry. Each sample was poured through the column and allowed to dry, then rinsed with deionized water (3 ml), 0.1M phosphate buffer ph 6.0/acetonitrile (80:20; 2 ml) and allowed to dry (10 min; 30 psi). Hexane was allowed to flow through the column (1 ml), and the columns were again dried. Finally, the drugs were eluted in ethyl acetate + 2% ammonium hydroxide (2 ml). The eluates were evaporated to dryness under nitrogen (20 psi/37~ and reconstituted in deionized water (50 pl) for analysis. Analytical procedure An Agilent Technologies 6410 LC coupled to a triple quadrupole mass spectrometer (MS-MS) operating in positive electrospray mode (ESI) was used for the analysis. The LC pump was an Agilent 1200 Series and the column was a Zorbax Eclipse XDB C18 (4.6 x 50 mm x 1.8 pm). The column temperature was held at 35~ the initial flowrate was 0.2 ml/min, and the injection volume was 5 pl. Solvent A consisted of Table I. Benzodiazepine Multiple Reaction Monitoring (MRM) Acquisition Parameters Start Precursor Fragment Fragment Compound time (min) ion ion voltage CE (V)* Segment I 8romazepam 0 316 288 (209) 160 20 (30)* Segment 2 O4-Clonazepam 4.1 320 274 120 25 Clonazepam 4.1 316 270 (214) 120 25 (35) Lorazepam 4.1 321 275(229) 140 25(35) Nitrazepam 4.1 282 236 (I 80) 160 25 (35) D5-alprazolam 4.1 314 286 160 25 Alprazolam 4.1 309 281 (274) 160 25 (30) Chlordiazepoxide 4.1 300 283 (227) 120 15 (30) D5-Oxazepam 4.1 292 246 120 20 Oxazepam 4.1 287 241 (269) 120 20 (20) Triazolam 4.1 343 308 (239) 120 35 (35) Segment 3 Flunitrazepam 5.4 314 268 (239) 160 30 (35) Midazolam 5.4 326 291 (249) 200 30 (40) D5-Temazepam 5.4 306 260 120 25 Temazepam 5.4 301 255 (177) 120 35 (40) Nordiazepam 5.4 271 140 (165) 160 30 (30) Segment 4 D5-Diazepam 7.2 290 262 160 25 Diazepam 7.2 285 257 (222) 160 25 (25) Flurazepam 7.2 388 315 (288) 160 25 (25) * Values in parentheses indicate qualifying transition 597

20mM ammonium formate ph 8.6; solvent B was acetonitrile. The mobile phase composition remained at 50:50 (v,v) although the flowrate changed. The flowrate was held at 0.2 ml/min for 6.5 min, then increased to I mumin. At 8 min, the flowrate was decreased back to 0.2 ml. The total run time was 10 rain, with a 4.5 rain post-run allowing mobile phase equilibrium. The mass spectrometer was operated at a capillary voltage of 4500 V; an initial nebulizer pressure of 15 psi, which was later raised to 50 psi for improved signal, was used; nitrogen gas flow of 6 L/min and a gas temperature of 300~ was used.the mass spectrometer was operated in multiple reaction monitoring (MRM) mode. The precursor and product ions, along with optimized fragment voltages and collision energy, are shown in Table I. Table II. Slope, Linearity, and Limit of Quantitation for the Assay Limit of Correlation quantitation Analyte Equation (r 2) (ng/mt) Alprazolam Y = 0.0298x + 0.0114 0.9995 0.5 Bromazepam Y = 0.0096x- 0.0129 0.9909 1 Chlordiazepoxide Y= 0.0146x- 0.0032 0.9998 1 Clonazepam Y = 0.0278x- 0.0108 0.9991 1 Diazepam Y = 0.0305x- 0.0004 0.9996 1 Flunitrazepam Y = 0.007x- 0,0002 0,9999 0.5 Flurazepam Y = 0.2984x- 0.0024 0,9993 0.5 Lorazepam Y = 0.0189x- 0.008 0,9986 5 Midazolam Y= 0.0156x- 0.0143 0,9960 0,5 Nitrazepam Y= 0,0551x + 0.018 0.9987 1 Nordiazepam Y = 0.011 x- 0.0013 0.9999 0.5 Oxazepam Y = 0,0228x- 0.0065 0.9996 5 Temazepam Y = 0.0149x- 0.0034 0.9998 0,5 Triazolam Y = 0.0225x + 0.0073 0.9995 0.5 Table III. Interday and Intraday Precision of the Assay Inter-day Intra-day precision (%) precision (%) Drug (n = 5) (n = 5) Alprazolam 2,03 2,8 Bromazepam 6.8 6.13 Chlordiazepoxide 2.26 6.7t Clonazepam 3.14 4.25 Diazepam 6.04 7.29 Flunitrazepam 5.11 4.64 Flurazepam 5.01 6.95 Lorazepam 3.68 3.64 Midazolam 5,94 3.29 Nitrazepam 1.42 4.41 Nordiazepam 3,27 3.71 Oxazepam 3.07 4.2 Temazepam 3 3.85 Triazolam 2.55 4.96 Method validation Linearity, sensitivity, and precision Calibration of the assay was calculated using linear regression analysis, forced through the origin, over a concentration range of 0.5-40 ng/ml for the benzodiazepines. Five deuterated internal standards were used in order to minimize the number of transitions being monitored per segment, and thereby improve the sensitivity of the assay. The LOQ for each drug was established by entering a signal-to-noise ratio (SNR) of 10 into the method analysis for the quantitative transition. If the response of the primary transition ion did not exceed an SNR of 10, it was not considered to be quantitative. The precision of the assay, both inter- and intraday, was determined at the concentration of 10 ng/ml of neat oral fluid: equivalent to 2.5 ng/ml drug diluted in buffer (n = 5). Drug recovery from the collection pad Extraction efficiency of the collection system for benzodiazepines was determined. Oral fluid was fortified with all the drugs at the concentration of 10 ng/ml (n = 6). A collection pad was placed into the fluid until the volume adequacy indicator turned blue, showing that 1 ml ( 10%) of oral fluid had been absorbed. The pads were placed into the Quantisal buffer (3 ml), capped, and allowed to remain at room temperature overnight, to simulate transportation to the laboratory. The following day, the pads were removed and an aliquot (1 ml) of the specimens was analyzed according to the described procedures. Results and Discussion The analytical method was validated according to standard protocols, whereby the LOQ, linearity range, correlation, and Table IV. Percentage Recovery of Benzodiazepines from Oral Fluid Collection System Following Overnight Incubation at Room Temperature (fortified at 10 ng/ml; n = 6) Drug Mean recovery (%) CV (%) Alprazolam 86.76 8.85 Bromazepam 88.42 14.01 Chlordiazepoxide 89.41 6.33 Clonazepam 88.1 2.97 Diazepam 82.82 4.42 Flunitrazepam 85.1 4.46 Flurazepam 81.57 2.85 Lorazepam 83.44 2.52 Midazolam 81.48 5,32 Nitrazepam 90.17 3.64 Nordiazepam 83.28 3.8 Oxazepam 84.65 2.82 Temazepam 84.19 2.96 Triazolarn 85.45 8.71 598

intra- and interday precision were determined. The validation results are presented in Tables II and III. Table II details the equation of the slope of the calibration curves (which were not forced through the origin), the nonweighted correlation coefficient, and the LOQ associated with each drug. The precision of the assays was excellent, with both withinday and between-day variations being below 7.3% for all drugs (Table III). The concentration detectable by a combination of solid-phase extraction and extremely sensitive tandem mass spectrometric detection is necessary for the determination of these drugs in oral fluid. The variables associated with the collection of oral fluid are not often considered. Some commercially available collection devices are unable to determine how much oral fluid has been collected for analysis, thereby calling into question any quantitative result. For any device incorporating a collection pad, the recovery of drugs from that pad is an important component of assay sensitivity and must be included in any experimental Figure 1O z * MRM [316.0 ~ 2~1,0) ~ 0F021,d 5.! Bromllzepam i~,/.~o + l 9 t~m t~+o.., ',e.,m m,,.,,o om~.~., 6 11 Td+zMama L ;']' 0J :': 103l * MM ize2,0.~ 2~ Ik.m 0~21 a q'-" 0 J 10 s/,, MRM t~l 0 + 2'R O} B~-,m 0r 0,!t Lorazepam 103 i + MRM (3200 -> 274 OJ Blnlo 0F0214 L_A_ 0.454 D4-cloc~lBepam ] ~ 10 3 I * MRM 13111,0.~ 2700} k OF'O~I d 1~1 Clofl'Zl~m I 04,, 104 I + MRM ~14~1.> "a~z.oj I~ 0~1,4 103 I + MRM ~IO~JD "~ ~1.0} Ikmm OIqB21,d 2-1 /Mprazolam? 0J... i: ~\... 10+ ] ~. M~M po0+.., 2~.0) e,,,,,,o O~,:r tlo 4 oj do ~ 1. Primary transitions for benzediazepines in oral fluid (5 ng/ml). + MRM 1287.0.,> 241.0) BtrwoOF4.d Figure 10 3.] 4.359 O~zepm 7-.I 6-.I..10~ 5-t ~ 5- ii 44 ] 4-3-1 < 3-24 2-1-1 ~ 1-0 I - - - 0- I 412 414 416 410 5 53 Acquisition time (rain) 2. Oxazepam extracted from oral fluid (10 ng/ml). <' MRM (29~.Q-> 24411.01 ~ OPO21xl A, I Ds-Oxazepam + MRM 12B? s ~ 241.0) ~ OPO21A '~' I OxazepMn + IdRM ~3~6~0.~ 291JD) Bz~ao OROQIA [ ]~,., Midazolam * MAM 1314-0,~ 288,0) ~ OFG21,d lol [ ~ e~,,=,p,,~ i10 4 1 + 14111,1 tso&o..~ 280.01 ~ OPO21.d zs-i I /~, I DS-Tmzepem o-i... I :10 ~l ~ MRM ~01.0.,~ 255.0 ~ OFO~l.d 21 I A I T'maz'pm OJI ~,10 3 + MI~M 12"/1.0-'* 140J~ Bir41o OFO21.d ;t I_A 1"-" :10 ~ + MRM [~le.o..~ 315,10) Bemao OPQ21.d + MRM 12~.0 J~ 262,01 Bmm 01q021 d :10 = L_.~ DS.D~zlpm, n :t :10 1 + MRM ~,Q,a 2S7.01BtNm OR~l.d :t LL ~ 207 ~ 241.0, 289.0 NlO 3 I~do'25.9 7" 6" 43 414 415 43 5 53 Acquisition t,~e (mtn) protocol. We have reported the efficient recovery of THC and its metabolite THCA from the Quantisal device (10,11), as well as the pain medications meperidine, tramadol, oxycodone, and propoxyphene (12,13); others have published the efficacy of the Quantisal device for various drugs, including oxazepam (14). Placing the pad in the Quantisal buffer overnight, and analyzing the samples the following day, allowed the recovery of benzodiazepines from the collection pad to be determined. In this way, losses occurring during overnight transportation to a laboratory were assessed. Six replicate analyses were analyzed to assess reproducibility of the extraction efficiency. The extraction recovery of all 14 benzodiazepines from the pad was over 81% (Table IV). The described procedure allowed the rapid determination of 14 benzodiazepines in oral fluid at an extremely low concentration, as is required for these drugs. The concentration of benzodiazepines present in oral fluid is very low due to the saliva-plasma ratio and the small dosage forms available. Additionally, the benzodiazepine class is wide, and an assay to incorporate all the possible members of the group is difficult. In their publication, Quintela et al. (7) detected a concentration of 0.57 ng/ml of midazolam I h after administration to a patient undergoing midazolam treatment; Smink et al. (9) reported only five positives for benzodiazepines in their comparison with urine specimens. In their study, 5 oral fluid samples were found to be positive: two for oxazepam (concentrations of 18 and 1659 ng/ml) and three for alprazolam (concentrations of 5, 6, and 9 ng/ml). All these concentrations were above the limits described in our method. There is only one description of 7- aminoflunitrazepam being present in slightly higher concentrations in oral fluid than its parent drug, flunitrazepam, which is reportedly unstable; the data was based on a single ingestion of flunitrazepam (15), and the concentration of parent drug detected (0.6-4 ng/ml) would have been identified using the described assay. From a forensic point of view, the chromatography afforded by the small particle analytical column allowed separation of the peaks in each of the four group segments. Segments allow specific transitions to be monitored over a given time period in order to improve the sensitivity of the assay. The primary transitions associated with the assay are shown in Figure 1 at a concentration of 5 ng/ml, which was the highest limit of quantitation for two of the drugs, lorazepam and oxazepam. The software is able to monitor a secondary transition from the precursor ion and automatically calculate the ratio to the primary ion. If the ratio is not within 20% of a calibration standard, determined at 10 ng/ml, the identification is rejected. This 599

xlo 3 + MRM [320 0 o 274 O) ~ OFO21 d D4-Clonazeparn method that includes qualifying ions for the identification of benzodiazepines at low concentration in oral fluid, and is currently in routine use in our laboratory. References xlo 3 094 O.B..,I f17-1 0.64 0.54 D.4.-I 0.3-1 0.24 o14 x10z I 12-1 0.8-.I O.64 04-1 0.2-1 I]-4 MRM 1321.0 -:' 275.0) ~ OPO21.d Lorazepam: Quantdyincj transition MRM (321.0..>229s Lorazepam: Qualifying transition OPO21.d Figure 3. Primary and secondary transitions for Iorazepam extracted from oral fluid (5 ng/ml): no interference from D4-clonazepam on the qualifying trace. is important in forensic analysis where court challenges to laboratory data are frequent. Monitoring a second transition gives additional confidence in the result; applying a ratio to that second transition compared to the primary product ion is a further enhancement to the identification of drugs in oral fluid. The software plots the ratio in the chromatographic window, so the operator is able to visually assess positivity (Figure 2). The utility of this is can be specifically noted in the MRM for lorazepam. In order to improve the sensitivity of the assay, the resolution for the first quadrupole is set to "Wide", allowing a significant range around the selected ion to be monitored. Unit resolution in the third quadrupole generally eliminates all but the selected transition; however, in this case, the transition from m/z 321 to m/z 275 allowed two ions to be detected. The lorazepam is separated from the D4-clonazepam on the basis of retention time, but also on the presence of the secondary qualifying transition: m/z 321 to m/z 229 (Figure 3). There is no corresponding qualifying transition from m/z 321 to m/z 229 in the MRM trace of D4-clonazepam. Summary The procedure described is suitable for the detection of benzodiazepines in oral fluid using a triple quadrupole LC-MS-MS system. The sensitivity of the assay is a significant improvement over that found in other publications. This is the first 1. J.M. Cook, R. Marshall, C. Masci, and J.C. Coyne. Physicians' perspectives on prescribing benzodiazepines for older adults: a qualitative study. J. Gen. Intern. A4ed. 22(3): 303-307 (2007). 2. A.W. Jones, A. Holmgren, and F.C. Kugelberg. Concentrations of scheduled prescription drugs in blood of impaired drivers: considerations for interpreting the results. Ther. Drug Monit. 29(2): 248-260 (2007). 3. E.W. Schwilke, M.I. Sampaio dos Santos, and B.K. Logan. Changing patterns of drug and alcohol use in fatally injured drivers in Washington State. J. Forens. 5ci. 51(5): 1191-1198 (2006). 4. C. Moore and D. Lewis. Comment on oral fluid testing for drugs of abuse: positive prevalence rates by Intercept immunoassay screening and GC-MS-MS confirmation and suggested cutoff concentrations. J. Anal. ToxicoL 27(3): 169 (2003). 5. G.E Kauert, S. Iwersen-Bergmannand, and S. Toennes. Assay of gg-tetrahydrocannabinol (THC) in oral fluid-evaluation of the OraSure oral specimen collection device. J. Anal. Toxicol. 30(4): 274-277 (2006). 6. P. Kintz and N. Samyn. Use of alternative specimens: drugs of abuse in saliva and doping agents in hair. Ther. Drug Monit. 24(2)' 239-246 (2002). 7. O. Quintela, A. Cruz, A. de Castro, M. Concheiro, and M. Lopez- Rivadulla. Liquid chromatography-electrospray ionization mass spectrometry for the determination of nine selected benzodiazepines in human plasma and oral fluid. J. Chromatogr. 825: 63-71 (2005). 8. E.L. Oiestad, U. Johnsen, and A.S. Christophersen. Drug screening of preserved oral fluid by liquid chromatography-tandem mass spectrometry. Clin. Chem. 53(2): 300-309 (2007). 9. B.E. Smink, M.P.M Mathijssen, K.J. Lusthof, J.J. de Gier, A.C.G. Egberts, and D.R.A. Uges. Comparison of urine and oral fluid as matrices for screening of thirty-three benzodiazepines and benzodiazepine-like substances using immunoassay and LC-MS(-MS). J. Anal. Toxicol. 30:478-485 (2006). 10. C. Moore, M. Vincent, S. Rana, C. Coulter, A. Agrawal, and J. Soares. Stability of A9-tetrahydrocannabinol (THC) in oral fluid using the Quantisal collection device. Forensic Sci. Int. 164(2-3): 126-130 (2006). 11. C. Moore, C. Coulter, S. Rana, M. Vincent, and J. Soares. Analytical procedure for the determination of the marijuana metabolite, 11-nor-Ag-tetra-hydrocannabinol-9-carboxylic acid (THCA), in oral fluid specimens. J. Anal. Toxicol. 30(7): 409412 (2006). 12. C. Moore, S. Rana, and C. Coulter. Determination of meperidine, tramadol and oxycodone in human oral fluid using solid phase extraction and gas chromatography-mass spectrometry. J. Chromatogr. B 850:370-375 (2007). 13. S. Rana, C. Moore, A. Agrawal, C. Coulter, M. Vincent, and J. Soares. Determination of propoxyphene in oral fluid. J. Anal. Toxicol. 30(8): 516-518 (2006). 14. O. Quintela, D. J. Crouch, and D.M. Andrenyak. Recovery of drugs of abuse from the Immunalysis Quantisal oral fluid collection device. J. Anal. Toxicol. 30(8): 614-616 (2006). 15. N. Samyn, G. de Boeck, V. Cirimele, A. Verstraete, and P. Kintz. Detection of flunitrazepam and 7-aminoflunitrazepam in oral fluid after controlled administration of Rohypnol. J. Anal. Toxicol. 26(4): 211-215 (2002). Manuscript received April 30, 2007; Revision received June 7, 2007. 600