SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma



Similar documents
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

How To Test For Contamination In Large Volume Water

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Determination of Tetracycline Antibiotics in Milk Using a Simple Strong Cation-Exchange SPE Cleanup Procedure and LC-MS/MS Analysis

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

A Novel Approach to Low Volume Sample Preparation

Oasis HLB Cartridges and 96-Well Plates

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Overview. Purpose. Methods. Results

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science

CORESTA RECOMMENDED METHOD N 72

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

A Complete Solution for Method Linearity in HPLC and UHPLC

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Thermo Scientific Mass Spectrometric Immunoassay (MSIA)

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

A Determination of Veterinary Drug Residues in Milk Using Polymeric SPE and UHPLC-MS/MS Analysis

Opiates in Urine by SAMHSA GC/MS

Development and Validation of Methods for Chemotherapy Drugs on the New Prelude SPLC LC-MS/MS System

LC-MS/MS for Chromatographers

Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Matrix Effect in Bio-Analysis of Illicit Drugs with LC-MS/MS: Influence of Ionization Type, Sample Preparation, and Biofluid

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Gas Chromatography Liner Selection Guide

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis

Structural Analysis of Labeled N-Glycans from Proteins by LC-MS/MS Separated Using a Novel Mixed-Mode Stationary Phase

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

ORGANIC SAMPLE PREPARATION

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

ApplicationNOTE Introduction

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

AMD Analysis & Technology AG

AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Guide to achieving reliable quantitative LC-MS measurements

Rapid LC/MS/MS Method Development for Drug Discovery

Transcription:

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application Note 20939 Key Words SPE, SOLA, Syncronis C18, ethinyl estradiol Abstract A simple, rapid, and sensitive method for determination of ethinyl estradiol in human plasma by liquid chromatography-tandem mass spectrometry was developed and evaluated. The drug was extracted from a plasma matrix using a Thermo Scientific SOLA SCX solid phase extraction device with a reversed-phase protocol. After extraction it was derivatized and again cleaned using the SOLA SCX device with a reversed-phase protocol. The resultant extracts were separated on a Thermo Scientific Syncronis C18 PLC column under reversed-phase, gradient conditions. Detection was performed on a Thermo Scientific TSQ Vantage triple quadrupole mass spectrometer using positive polarity, heated electrospray ionization (-ESI) conditions operating in selected reaction monitoring (SRM) mode. igh extraction efficiency, excellent peak shape, and linearity over the range 5 to 200 were achieved with high precision. Introduction Ethinyl estradiol is commercially available as a single therapy or in combination with other progestogens for use as oral contraceptives. The purpose of this particular study was to demonstrate the quantitative determination of ethinyl estradiol in human plasma over the concentration range of 5 200. This was achieved by a three-step sample preparation process, followed by separation on a Syncronis C18 PLC column and subsequent MS/MS detection using the TSQ Vantage mass spectrometer. The three-step sample preparation consisted of the following: Extraction of ethinyl estradiol from human plasma using a SOLA SCX solid phase extraction device Derivatization of extracted sample with dansyl chloride in order to enhance sensitivity Post-derivatization clean up by a SOLA SCX solid phase extraction device for removal of excess derivatization reaction mixture constituents and other interferences A mixed-mode cation exchange chemistry was found to be more suitable than a reversed-phase chemistry in terms of sample clean up. The structures of derivatized/underivatized ethinyl estradiol are shown in Figure 1. 3 C N O C 3 O S O C 3 Derivatized ethinylestradiol MW 529.3 O O C Underivatized ethinylestradiol MW 296.2 Figure 1: Structures of derivatized/underivatized ethinyl estradiol C 3 O C

2 SOLA is a revolutionary solid phase extraction (SPE) device. This first-in-class SPE product range introduces next generation, innovative technological advancements, giving unparalleled performance characteristics compared to conventional SPE, phospholipid, and protein precipitation products. These include: igher levels of reproducibility igher levels of extract cleanliness Reduced sample and solvent requirements Increased sensitivity SOLA has significant advantages for the analyst when processing compounds in complex matrices, particularly in high-throughput bioanalytical and clinical laboratories where reduced failure rate, higher analysis speed, and lower sample/solvent requirements are critical. The increased performance of SOLA products gives higher confidence in analytical results and lowers cost without compromising ease of use or requiring complex method development. The chromatography of the derivatized ethinyl estradiol was achieved using a Syncronis C18 column, which gave the most suitable retention when compared to other C18 chemistry columns. One of the key goals for the chromatographer is to achieve a consistent, reproducible separation. The selection of a highly reproducible PLC column is essential if this goal is to be attained. The Syncronis column range has been engineered to provide exceptional reproducibility due to its highly pure, high surface area silica, dense bonding, and double end capping, all controlled and characterized through the use of rigorous testing. Experimental Details Consumables Part Number Fisher Scientific Optima LC/MS grade acetonitrile A955-1 Fisher Scientific LC/MS grade water W/011217 Fisher Scientific Optima LC/MS grade formic acid (90%) A117-50 Ethinyl estradiol and d4-ethinyl estradiol, kindly supplied by a customer SOLA SCX 10 mg/1 ml 60109-002 Thermo Scientific Syncronis C18 50 mm 2.1 mm, 1.7 µm 97102-052130 Thermo Scientific Micro+ Vial 300 μl, Fused Insert 60180-507 Thermo Scientific 9 mm Screw Top Cap W/ PTFE/Silicone septa 60180-516 Sample andling Equipment Part Number Thermo Scientific FinnPipette (100 1000 µl) 4642090 Thermo Scientific FinnPipette (20 200 µl) 4642080 Thermo Scientific FinnPipette (2 20 µl) 4642060 Thermo Scientific Finntip Flex 1000 94060720 Thermo Scientific Finntip Flex 200 94060320 Sample Pretreatment A standard spiking solution of ethinyl estradiol was prepared in methanol at a concentration of 0.2 mg/ml. An internal standard solution of ethinyl estradiol d4 was prepared in methanol at a concentration of 0.1 mg/ml. For the preparation of standards and quality control (QC) samples, 475 µl of drug-free human plasma was taken and 25 µl of standard spiking solution and 50 µl of internal standard solution were added. For blank standards, 75 µl of water was added. Samples were subsequently diluted with the addition of 500 µl of 5 mm ammonium formate at p 4.5.

Sample Preparation The sample preparation protocol consisted of initial extraction of the drug from human plasma using SOLA SCX SPE. The drug in the extract was then derivatized. This was followed by additional clean up (to remove excess reagent) using SOLA SCX SPE. Extraction: Cartridge: Condition: Application: Wash one: Wash two: Elution: SOLA SCX 10 mg/1 ml 1000 µl methanol then 1000 µl water Load pre-treated sample (1050 µl) 1000 µl water / methanol (95:5 v/v) (twice) 1000 µl water / methanol (80:20 v/v) 1000 µl methanol Dry down: Under a stream of nitrogen at 50 C Derivatization: The dried sample was reconstituted in 200 µl of 100 mm sodium bicarbonate, p 10.5 (adjusted with 1 M sodium hydroxide) and vortexed for few seconds. An additional 200 µl of dansyl chloride in acetone (1 mg/ml) was added and the sample was vortexed for few seconds. The sample was incubated at 60 C for 30 minutes and then cooled. Then, 200 µl water was added and the sample was vortexed for few seconds. Cleanup: SPE: Condition: Application: Wash one: Wash two: Elute: SOLA SCX 10 mg/1 ml 1000 µl methanol then 1000 µl water Load derivatized sample 1000 µl water / methanol (95:5 v/v) (twice) 1000 µl water / methanol (80:20 v/v) 1000 µl methanol Dry down: Under a stream of nitrogen at 50 C Reconstitute: In 250 µl 5 mm ammonium formate, p 4.5 / acetonitrile (20:80 v/v) and vortex for few seconds 3 Separation Conditions Recommended instrumentation: Thermo Scientific Dionex UltiMate 3000 RSLC Column: Syncronis C18 50 mm 2.1 mm, 1.7 µm Mobile phase A: Water + 0.1% formic acid Mobile phase B: Acetonitrile Mode: Gradient (refer to Table 1) Flow rate: 0.5 ml/min Column temperature: 30 C Injection details: 10 µl Time (min) % B 0 60 5.0 90 5.2 60 6.0 60 Table 1: Mobile phase gradient

4 MS Conditions Instrumentation: TSQ Vantage system Ion Source Type: ESI-2 Polarity: Positive Spray voltage: 3500 V Vaporizer temperature: 500 C Sheath gas pressure: 40 Arb Ion sweep gas pressure: 0 Arb Auxiliary gas pressure: 20 Arb Capillary temperature: 375 C Declustering voltage: 0 V Collision pressure: 1.5 mtorr Scan width: 0.2 m/z Scan time: 0.1 s Q1 (FWM): 0.2 Q3 (FWM): 0.7 The compound transition details are given in Table 2. Compound Dansyl-ethinyl estradiol Dansyl-ethinyl estradiol-d4 (IS) Parent m/z 530.2 m/z 534.2 Products m/z 171.0 m/z 171.0 Collision energy 34 V 36 V S-lens 162 V 162 V Table 2: Compound transition details Data Processing Software: Thermo Scientific LCQUAN Results Derivatized ethinyl estradiol standards extracted from human plasma gave a linear response over the range of 5 to 200 with an r 2 coefficient of 0.997 (Figure 2 and Table 3). Chromatography at the limit of quantitation (LOQ) of 5 is shown in Figure 3. 1.8 1.6 1.4 1.2 Area Ratio 1.0 0.8 0.6 0.4 0.2 0.0 0 20 40 60 80 100 120 140 160 180 200 220 [Ethinyl Estradiol, ] Figure 2: Derivatized ethinyl estradiol linearity over the dynamic range 5 200

100 90 80 0.63 5 70 Relative Abundance 60 50 40 0.79 3.04 Derivatized ethinyl estradiol RT: 4.44 SN: 250 4.22 30 20 10 0.47 1.01 1.20 3.30 3.58 2.15 2.17 2.78 1.47 2.08 4.09 4.69 5.02 5.28 5.52 0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 Time (min) Derivatized ethinyl estradiol-d4 100 RT: 4.42 SN: 1912 90 80 70 Relative Abundance 60 50 40 30 20 0.56 0.62 1.19 10 0 0.0 0.40 0.5 3.12 3.20 2.27 2.49 1.57 2.13 3.56 3.99 4.55 4.75 5.20 5.55 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 Time (min) Figure 3: Representative chromatograms of derivatized ethinyl estradiol SRM, extracted from human plasma (at 5 top) and derivatized ethinyl estradiol-d4 (IS) (bottom)

6 QC samples were analyzed in replicates of six at concentrations of 5, 15, 100, and 170 (Table 4). Overspikes (of ethinyl estradiol) were analyzed at concentrations of 5, 15, 100, and 170 and used to calculate recovery and matrix interference (Table 5). Accuracy and Precision Standard Specified [Ethinyl estradiol] Calculated [Ethinyl estradiol] %Diff S1 5 5.4 8.1 S2 10 9.1-8.9 S3 20 20.5 2.5 S4 40 40.7 1.8 S5 80 73.8-7.7 S6 120 123 2.5 S7 160 159-0.6 S8 180 191 6.0 S9 200 193-3.7 Table 3: Accuracy data for extracted standards over the linear range 0.2 40 ng/ml Standard Specified [Ethinyl Estradiol] Number of Samples (n) Calculated [Ethinyl Estradiol] Precision (%CV) LLOQ 5 6 4.9 11.6 QCL 15 6 15.0 2.5 QCM 100 6 94.7 2.2 QC 170 6 160.3 7.2 Table 4: Average precision data for six replicate QCs for derivatized ethinyl estradiol Recovery Standard % Recovery at Each Level % Matrix Interference at Each Level QC Lower Limit of Quantitation 108.0-6.3 QC Low 102.1-3.5 QC Medium 100.1-1.8 QC igh 109.8-12.4 Table 5: Recovery and matrix interference data for derivatized ethinyl estradiol

Conclusion SOLA SCX SPE cartridges and Syncronis C18 PLC columns coupled with the TSQ Vantage mass spectrometer allow for simple and effective extraction, separation, and quantification of derivatized ethinyl estradiol from human plasma. The method exhibited good linearity (r 2 = 0.997) for concentrations of derivatized ethinyl estradiol in the range 5 200. A limit of quantitation of 5 for derivatized ethinyl estradiol in plasma was achieved. Extraction recovery and matrix interference were found within the limits of acceptance generally applied to bioanalytical methods. Application Note 20939 Reference http://www.fda.gov/downloads/drugs/guidancecomplianceregulatoryinformation/guidances/ ucm070107.pdf thermoscientific.com/columns 2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 800 332 3331 Australia 1300 735 292 (free call domestic) China 800 810 5118 (free call domestic) 400 650 5118 France +33 (0)1 60 92 48 34 Germany +49 (0) 2423 9431 20 or 21 India +91 22 6742 9494 +91 27 1766 2352 Japan 0120 753 670 (free call domestic) 0120 753 671 fax United Kingdom +44 (0) 1928 534 110 New Zealand 0800 933 966 (free call domestic) Singapore +65 6289 1190 All Other Enquiries +44 (0) 1928 534 050 Technical Support For advice and support, please visit our website: www.thermoscientific.com/chromexpert AN20939_E 04/14S