Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method



Similar documents
Comprehensive LC/MS Analysis of Illicit and Pain Management Drugs, Including Their Metabolites, in Urine

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System

AppNote 6/2011. Determination of Pain Management Drugs using Automated Disposable Pipette Extraction and LC-MS/MS KEYWORDS ABSTRACT

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications

Simultaneous Quantitation of 78 Drugs and Metabolites in Urine with a Dilute-And-Shoot LC MS-MS Assay

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Improved Method for the Analysis of 31 Drugs of Abuse in Oral Fluid samples using the Thomson extreme FV by LC-MS/MS

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504

LC-MS for Pain Management Support. Gwen McMillin, PhD, DABCC(CC,TC) University of Utah ARUP Laboratories

LC-MS for Pain Management Support

Indiana State Department of Toxicology 2012 Annual Report

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Toxicology CPT Code Changes for 2016

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

PtProtect Pain Medication Management Program Monitors Patient Compliance

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

An Overview of Mayo Clinic Tests Designed for Detecting Drug Abuse

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples

AppNote 7/2009. Analysis of Drugs of Abuse using Automated Disposable Pipette Extraction and LC/MS/MS KEYWORDS ABSTRACT

DUI in Southern Ohio MATT 2006

Interagency Guideline on Opioid Dosing for Chronic Non-cancer Pain (CNCP)

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

What Does the Drug Test Tell Us

Development and Validation of Methods for Chemotherapy Drugs on the New Prelude SPLC LC-MS/MS System

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA

Thermo Scientific ConFlo IV Universal Interface

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Urine Drug Testing. Why drug test? Set the Standard!

AppNote 4/2014. Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS.

What Americans Believe. Grant Beardsley, MS, MT(ASCP) 10/12/2015. Interpreting: Urine Drug Test Results in Chronic Opioid Therapy and Drugs of Abuse

Prevalence and risk of driving under influence of psychoactive substances: Results from epidemiological studies

Thermo Scientific SIEVE Software for Differential Expression Analysis

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare

Thermo Scientific ExactFinder Software

How To Test For Contamination In Large Volume Water

Appropriate utilization of drug tests for pain management patients

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

Opiates in Urine by SAMHSA GC/MS

Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION

Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow

SIMULTANEOUS QUANTITATIVE DETERMINATION OF OPIOID DEPENDENCY TREATMENT DRUGS IN HUMAN URINE USING UPLC/MS/MS

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

A Complete Solution for Method Linearity in HPLC and UHPLC

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Practical Aspects of Drug Testing in Human Hair

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Urine Drug Testing Methadone 101 Methadone for hospitalists

Collection Instructions for ARUP Drug Screen Kit #49204

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Screening and quantitative determination of drugs in diluted urine by UPLC-MS-MS

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Technical Procedure for the Solid Phase Extraction of Acidic, Neutral and Basic Drugs for GC-MS Analysis

Oral Fluid as an alternative matrix in workplace drug testing: which drugs at which cutoff concentration?

FLORIDA DEPARTMENT OF LAW ENFORCEMENT SEPTEMBER Report

WATERS QUANTITATIVE ANALYSIS solutions

Determination of Tapentadol and its Metabolite N

Successful Sample Preparation Strategies for LC/MS/MS Analysis of Drugs in Complex Biological Matrices for Forensic Toxicology Applications

Thermo Scientific Vanquish UHPLC System. Revolutionize Your. UHPLC Experience

Clinic Reference Guide

The CPT code for Human papillomavirus (HPV) has changed from to This code (87624) covers testing for HPV High Risk genotypes.

How To Control A Record System

Urine Drug Toxicology and Pain Management Testing. Kara Lynch, PhD, DABCC University of California San Francisco San Francisco, CA

How To Understand The Health Effects Of Prescription Drugs

2013 Annual Toxicology Report

Glyphosate and AMPA. Medical Laboratory Bremen, Haferwende 12, Bremen, Germany. BUND, FoE

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

What is Eluent Generation?

Transcription:

Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access MAX, forensic toxicology, drugs of abuse, pain management drugs, urine, quantitation Goal The goal of this work was to develop a simple dilute-and-shoot liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the simultaneous quantitation of 4 drugs of abuse, including pain management drugs, in human urine for forensic toxicology purposes. The drugs to be analyzed included opioids, amphetamines, benzodiazepines, cocaine, buprenorphine, methadone, and some of their metabolites. An additional objective was to use ultra-high-pressure liquid chromatography (UHPLC) to improve throughput and sensitivity of the method. Introduction LC-MS/MS has become more accepted as the tool for quantitative analysis of drugs in forensic toxicology laboratories. This technique enables simultaneous detection of multiple analytes of interests and is compatible with a simple dilute-and-shoot sample preparation method for urine samples. Methods Sample Preparation Nine individual human urine and pure water samples were spiked with and of the 4 drugs of abuse, pain management drugs, and with internal standards (IS). The samples were then mixed with β-glucuronidase and incubated at 6 C for hydrolysis. Methanol was added to the mixture and the supernatant was diluted with water. The final dilution factor was. The mixture was centrifuged at 17, g for minutes. Fifty microliter injections of the supernatant were analyzed by LC-MS/MS. LC-MS/MS Conditions LC-MS/MS analysis was performed on a Thermo Scientific Accela 12 pump and Accela Open autosampler coupled to a Thermo Scientific TSQ Quantum Access MAX triple stage quadrupole mass spectrometer. The analytical column was a Thermo Scientific Accucore PFP column ( 2.1 mm, 2.6 µm particle size) maintained at room temperature. Details of the LC gradient and mobile phases (MP) are as follows: Time Flow rate Gradient MPA MPB MPC (min) (ml/min) (%) (%) (%)..7 Step 9..7 Ramp 6 4 2.6.7 Ramp 9 4. 1. Step 1. 1. Step 1.7 1. Step 9 MPA: 1 mm NH 4 Ac and.1% formic acid in water MPB: 1 mm NH 4 Ac and.1% formic acid in methanol MPC: acetonitrile/isopropanol/acetone 9:9:2 (v/v/v) The mass spectrometer was operated with a heated electrospray ionization (HESI-II) source in positive ionization mode. The MS conditions were as follows: Spray voltage (V) 4 Vaporizer temperature ( C) Sheath gas pressure (arbitrary units) Auxiliary gas pressure (arbitrary units) 1 Capillary temperature ( C) Blank human urine was used as the matrix for calibration samples. The concentrations of the calibrators were 1, 2,, 1,,, 1,,, and 1. Concentration of the internal standards in all samples was 2.

2 Data were acquired in selected-reaction monitoring (SRM) mode. SRM transitions for the 4 drugs and their internal standards are shown in Table 1. For each analyte and internal standard, two SRM transitions were monitored. One of transition was used as the quantifier and the other as the qualifier. The signal ratio between the qualifier and the quantifier was used to evaluate the validity of the results. Validation The validation procedure included tests for the following: 1) matrix effects; 2) lower limit of quantitation (LLOQ), linear range, accuracy, and precision; and ) carryover. Table 1. Drug analytes, their corresponding internal standards, and the SRM transitions for both analytes and internal standards Analyte Precursor Quantifier Qualifier Ratio (%) Corresponding Internal Standard Precursor Quantifier Qualifier 6-MAM 28.1 16.1 211. 86. 6-MAM-d 1. 16.1 211.1 7-Amino-clonazepam 286. 222.1 2.1 9. 7-Amino-clonazepam-d4 29.11 226.1 24.1 7-Amino-flunitrazepam 284.1 1.1 226.1 2. 7-Amino-flunitrazepam-d7 291.11 18. 2.1 7-Aminonitrazepam 22.1 121.1 224.1 16. 7-Amino-clonazepam-d4 29.11 226.1 24.1 a-hydroxy-alprazolam 2.1 216.1 297.1 2. α-hydroxy-alprazolam-d.1 221.1 2.1 Alprazolam 9.4. 281. 76. Temazepam-d 6.1 26.1 288.1 Amphetamine 16.1 6. 91. 1. Amphetamine-d 141.1 92. 9. Benzoylecgonine 29.1 1.1 168.1. Benzoylecgonine-d 29.1 1.1 171.1 Benzylpiperazine 177.1 6. 91. 16. Benzylpiperazine-d7 184.1 7. 98. Buprenorphine 468. 96. 414. 1. Diazepam-d 29.12 198.1 227.1 Carisprodol 261. 62.1 97.1 8. α-hydroxy-alprazolam-d.1 221.1 2.1 Clonazepam 1.9 214. 27. 26.6 Temazepam-d 6.1 26.1 288.1 Cocaine 4.1 82. 182.1 17. Amphetamine-d 141.1 92. 9. Codeine.1 16. 21. 91. Codeine-d. 16.1 21.1 Diazepam 28.1 19.1 222.1 72. Diazepam-d 29.12 198.1 227.1 EDDP 279. 2. 2. 4. Temazepam-d 6.1 26.1 288.1 Fentanyl 7. 1. 188. 67. Temazepam-d 6.1 26.1 288.1 Flunitrazepam 14.4 29.1 268.1 2. Temazepam-d 6.1 26.1 288.1 Flurazepam 88.1 288.1 1.1 11. Temazepam-d 6.1 26.1 288.1 Hydrocodone. 171. 199. 4. MDA-d 18.1 11. 17.1 Hydromorphone 286.11 17.1 18.1 64. Benzylpiperazine-d7 184.1 7. 98. Lorazepam 21. 27.. 64. α-hydroxy-alprazolam-d.1 221.1 2.1 MDA 18.1 1. 1.1 79. MDA-d 18.1 11. 17.1 MDEA 8.1 1.1 16. 24. Nordiazepam-d 276.1 16. 21.1 MDMA 194.1 1.1 16.1 4. MDMA-d 199.1 1.1 16.1 Meperidine 248. 174. 2.1 28. Diazepam-d 29.12 198.1 227.1 Methadone 1. 1.1 26.1 29. Diazepam-d 29.12 198.1 227.1 Methamphetamine 1.1 6. 91. 9. Methamphetamine-d 1.1 91. 92. Midazolam 26.1 249. 291. 28. Diazepam-d 29.12 198.1 227.1 Morphine 286.1 12.1 16. 78. Morphine-d 289.1 12.1 16.1 Naloxone 28.21 212. 1.1 2. 7-Amino-clonazepam-d4 29.11 226.1 24.1 Naltrexone 42. 27.1 24. 16. MDA-d 18.1 11. 17.1 Norbuprenorphine 414. 187.1 4. 99. Temazepam-d 6.1 26.1 288.1 Nordiazepam 271. 14.1 8.1 1. Nordiazepam-d 276.1 16. 21.1 Norfentanyl 2.. 84. 16. MDMA-d 199.1 1.1 16.1 Normeperidine 24. 111.1 16.1. Temazepam-d 6.1 26.1 288.1 Oxazepam 287. 241. 269. 82. Oxazepam-d 292.1 246.1 274.1 Oxycodone 16. 241. 298. 22. Benzoylecgonine-d 29.1 1.1 171.1 Oxymorphone 2.1 227.1 284.. 7-Amino-clonazepam-d4 29.11 226.1 24.1 PCP 244. 86. 19.1 84. Diazepam-d 29.12 198.1 227.1 Propoxyphene 4. 8. 91.1 1. Diazepam-d 29.12 198.1 227.1 Temazepam 1. 2. 28. 6. Temazepam-d 6.1 26.1 288.1 Tramadol 264. 8. 246.1. Temazepam-d 6.1 26.1 288.1

Results and Discussion Matrix Effects Matrix effects were assessed with the nine individual human urine samples. Absolute recovery was determined by comparing the signals of unlabeled drugs in urine and water samples. Relative recovery was determined by comparing the analyte/is ratio in urine and water samples. The recovery/matrix effects results are summarized in Table 2. Summary of matrix effects Drug Average Absolute Recovery (%, n=9) Table 2. All 4 drugs had almost full absolute recovery (between 8% and 1%), except morphine for which the matrix effect was compensated by the use of its internal standard, morphine-d. The observed precision from the nine individual human urine samples was below 1% for most of the 4 drugs. CV (%, n=9) Average Relative Recovery (%, n=9) CV (%, n=9) 6-MAM 86.7 92. 16.2 12.2 9.1 1.2.6.7 7-Amino-clonazepam 96.4 18.4 11. 11.6 9. 1.7 6.1.8 7-Amino-flunitrazepam 86.8 9. 11.4 8.4 97.1 12.1 6..1 7-Aminonitrazepam 86. 8. 12. 9.6 8.6 81.9 9.9 8.2 a-hydroxy-alprazolam 87.4 87.4 12.9 6.7 99.4 96.2 1.4 4.1 Alprazolam 94. 89.1 26. 16.6 9.1 84.8 22.9 1.9 Amphetamine 19. 112. 11.8 8. 112.4 11.8 16.7.8 Benzoylecgonine 82.7 8.7 1. 12.7 98.7 1.9 4.6.6 Benzylpiperazine 87. 8.4 1. 7.2 1.4 1.6 8.7 7. Buprenorphine 18.4 96.9 1. 6.2 118.1 97.2 14.6.4 Carisprodol 88. 96. 1.1 11. 1. 1.8 1.6 8.4 Clonazepam 1.7 98.4 9. 6.9 1.4 94.4 1.6 9. Cocaine 9.6 9. 7.4 8.2 9. 92.2.. Codeine 9.9 98.9 8.6 8.2 99. 98.. 7.2 Diazepam 98. 96.6 14. 9.1 16. 96.7 11.7 6. EDDP 1.8 99.2 6.8 2.9 16.8 9. 1.7 6.2 Fentanyl 98.6 1.9 4.1 2.8 11.4 96.7 1.7.8 Flunitrazepam 8.7 86.9 18.8 14.7 87.1 82.9 14.6 12.1 Flurazepam 97. 1.1 4.2.9 1.2 98.8 11.7.8 Hydrocodone 91. 96.4 1.1 1. 9.2 97.9 7.8 9.6 Hydromorphone 91.2 94. 11. 1.4 14.6 11.8 7.2.4 Lorazepam 1.7 9. 16. 6. 1.7 99.7 17.2.7 MDA 96.6 1.8 16.1 9.6 1.6 17.9 8.7 6.4 MDEA 9.6 94. 11.8 1.1 99. 82.8 9.7 1.6 MDMA 92. 94. 9. 7.8 16.1 12.4 2.4 4.1 Meperidine 88.4 88.4 9.8 9.9 96.2 88. 7.4 7.8 Methadone 11.6 1.2.2.4 111.1 1.6 8.9.4 Methamphetamine 94.6 86.2 12. 11.1 1. 94.1 8.4.6 Midazolam 98.4 97.4 9.. 17.1 97.6 6.. Morphine 48.1.8 6. 8.2 9. 98.4 6.9.4 Naloxone 124.2 129.4 17.9 16.1 116.1 12. 9.9 7.7 Naltrexone 96.1 1.2 12.6 1.9 1. 11.9.2 6.1 Norbuprenorphine 76.9 14.6 19.4 14.2 78.9 99.9.4 11.6 Nordiazepam 12.8 17.1 21. 7. 16. 94.2 19.8 7. Norfentanyl 89. 92.2 11.4 8.2 1.2 1.1 12.4 4. Normeperidine 81.7 92. 11.9 11.6 8.1 87.7 7.8 8.4 Oxazepam 9.8 91. 1.8.1 11.4 12.1 6.8 4.7 Oxycodone 8.4 84.7 8.8 1.9 97. 1.1 9.7 6.1 Oxymorphone 17. 11.4 1.2 12.9 1.1 97. 8.2 9. PCP 1.8 1. 4.2 4. 11. 1.9 8.8.1 Propoxyphene 11. 1.8 6.8.8 111.2 14.1 1. 4.4 Temazepam 9. 12.2 14.2 7.1 97. 97.7 12. 4.6 Tramadol 77.8 84.8 14.1 12.9 78.9 8.8 1.1 1.4

4 Lower Limit of Quantitation, Linear Range, Accuracy, and Precision The LLOQ of these 4 drugs and other aspects of analytical performances of this method are summarized in Table. Linear fit with 1/X weighting was used for calibration curves of all the 4 drugs. The LLOQ for these 4 drugs was determined to be between 2 and except for tramadol, which was. At the LLOQ, the accuracy ranged between 89.9% and 118.4%, and precision ranged between.6% and 19.%. The method was linear to 1 for all the drugs. Figure 1 shows the calibration curves of six typical pain management drugs in human urine. Table. Lower limit of quantitation, linear range, accuracy, and precision Drug Retention Time (min) LLOQ () Accuracy at LLOQ (%, n=4) CV at LLOQ (%, n=4) Linear Range () R 2 Precision (%, n=6) Precision (%, n=6) 6-MAM 2.97 2 9. 14.6 2 1.99.8 2.8 7-Amino-clonazepam 2.76 9.1 1.1 1.9988.4 4. 7-Amino-flunitrazepam.1 11. 1.7 1.998. 4. 7-Aminonitrazepam 2.1 2 12. 9.6 2 1.9972.7.2 a-hydroxy-alprazolam.87 94. 1. 1.9972 6.9.9 Alprazolam 4.11 94.1 1.1 1.99 2.7 1.2 Amphetamine 2.97 94.9 7.7 1.9944.4. Benzoylecgonine 2.99 92..6 1.999 2.7 2. Benzylpiperazine 2.7 1 96. 1.4 1 1.9979 9.. Buprenorphine 4. 94.7 17. 1.9976 6. 6.1 Carisprodol.8 1 14. 11. 1 1.99 9. 6. Clonazepam 4. 92.7 7.7 1.994 9.1 6.1 Cocaine 4.2 11.2 7.4 1.9969 4..7 Codeine 2.82 1 11.4 18. 1 1.9978 6.6.8 Diazepam 4.24 9. 11.9 1.9979 7..4 EDDP 4.9 1 16..9 1 1.9944 4. 2.2 Fentanyl 4.62 2 18.9.7 2 1.997 4.8 1.8 Flunitrazepam 4.12 9.7 17.2 1.994 6.4 4.4 Flurazepam 4.7 2 118.4.6 2 1.9961 4.9 2.4 Hydrocodone.16 2 16.6 9.6 2 1.9988 7.8 2.8 Hydromorphone 2.2 2 89.9 1.2 2 1.9979 8.2.1 Lorazepam.86 92. 17. 1.994 2.2 9.6 MDA.16 1 9.1 6.8 1 1.9974 1.1.1 MDEA.97 2 14. 4. 2 1.997 7. 4.4 MDMA.61 97. 4. 1.997 7.6 2.2 Meperidine 4. 11.2 9.6 1.9986. 4.6 Methadone 4.9 1..8 1.9982 4.2. Methamphetamine.1 16..1 1.9979. 4. Midazolam 4.48 2 117.1 12.7 2 1.998 7. 4. Morphine 1.71 9. 1.6 1.999.. Naloxone 2.86 1 12. 1.9 1 1.9944. 2.9 Naltrexone.11 11.9 7. 1.998.1 1.6 Norbuprenorphine 4.1 11.4 14.4 1.99.9 8.4 Nordiazepam 4.6 1 97.1 19. 1 1.9948 8.4.8 Norfentanyl.68 1 12. 6. 1 1.998 7.1 2. Normeperidine 4. 2 116.2 11.2 2 1.9982 7. 4.2 Oxazepam.88 18. 1. 1.997 1.9 6.4 Oxycodone. 91.9 11.7 1.9982 2.6 2. Oxymorphone 2.1 2 9. 9. 2 1.9946 1. 2.6 PCP 4.8 2 1.9 4. 2 1.9981 7.8. Propoxyphene 4.7 1 11.6 4.1 1 1.9978 7..2 Temazepam 4. 14.6 16.9 1.9981.6 2.2 Tramadol 4.4 98.8 2. 1.997 NA 2.

4.8 4.6 7-Aminoclonazepam 4.4 4.2 4..8.6.4.2. 2.8 2.6 2.4 2.2 2. 1.8 1.6 1.4 1.2 1..8.6.4.2. 4.8 4.6 Benzoylecogonine 16 Fentanyl 4.4 4.2 4..8.6.4.2. 2.8 2.6 2.4 2.2 2. 1.8 1.6 1.4 1.2 1..8.6.4.2. 1 14 1 12 11 1 9 8 7 6 4 2 1 Concentration () Concentration () Concentration () 14 Hydrocodone 1 12 11 14 1 12 11 Midazolam. Morphine 4. 4. 1 1. 9 8 7 6 9 8 7 6. 2. 2. 4 4 1. 1. 2 2 1 1. Concentration () Concentration (). Concentration () Figure 1. Calibration curves of six selected drugs in spiked human urine Method precision was also assessed with spiked human urine samples at low and high quality control (QC) concentrations of and, respectively (Table 2). Precision values at low ( ) and high ( ) quality control concentrations ranged between 1.1% and 1.9% (Table 2). Figure 2 shows both the quantifier and qualifier SRM chromatograms of selected pain management drugs spiked at in human urine. 1 Methadone 8 EDDP Relative Abundance (%) 6 4 Codeine 6-MAM Benzoylecgonine Oxycodone Hydrocodone MDA PCP Fentanyl Cocaine Meperidine MDEA Carisprodol 7-Aminonitrazepam Hydromorphone Oxymorphone Methamphetamine MDMA Morphine. 1. 2.. 4.. Time (min) Figure 2. SRM chromatograms of selected drugs at in spiked human urine

Carryover The lowest calibrator was analyzed after the highest calibrator. No carryover causing elevated measurements of the drugs in the lowest calibrator was observed. Conclusion The developed method provides a simple, fast, and sensitive way for forensic toxicology labs to simultaneously quantify 4 drugs of abuse, including pain management drugs, in human urine by LC-MS/MS. The method provided LLOQ values of 2 for 42 of the 4 drugs, and was linear to 1. Minimal ion suppression and no carryover were observed in matrix samples. At the LLOQ, the accuracy ranged between 89.9% and 118.4%. Method precision ranged between 1.1% and 1.9% at low and high QC samples. Application Note 76 For forensic toxicology use only. www.thermoscientific.com 12 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. Africa-Other +27 11 7 184 Australia +61 977 4 Austria +4 1 4 Belgium +2 7 42 41 Canada +1 8 8447 China +86 1 8419 88 Denmark +4 7 2 62 6 Europe-Other +4 1 4 Finland/Norway/Sweden +46 8 6 468 France + 1 6 92 48 Germany +49 61 48 114 India +91 22 6742 944 Italy +9 2 9 91 Japan +81 4 4 91 Latin America +1 61 688 87 Middle East +4 1 4 Netherlands +1 76 79 New Zealand +64 9 98 67 Russia/CIS +4 1 4 South Africa +27 11 7 184 Spain +4 914 84 96 Switzerland +41 61 716 77 UK +44 1442 2 USA +1 8 2 472 AN6686_E 11/12S