Algorithmes : comment calculer et rapporter la dose
|
|
|
- Beatrix Arleen Allen
- 10 years ago
- Views:
Transcription
1 Algorithmes : comment calculer et rapporter la dose N. Reynaert Medical Physics Department, Centre Oscar Lambret Lille
2 Introduction Until 1970: No CT data available: Patient consists of water 2D planning approximate dose calculation
3
4 Introduction Now: Very precise 4D CT information, MR, PET 3D dose calculation Dose escalation IMRT: Large dose gradients Extra cranial SBRT: hypofractionation Small irradiation fields lack of lateral CPE AAPM Report No. 85 (Papanikolaou et al, 2004), Ahnesjö and Aspradakis (1999), AAPM TG101 (2010).
5 Required Dose accuracy Different factors: The slope of dose-effect curves The level of dose differences that can be detected clinically Clinical studies What is practically achievable (evolving )
6 The slope of dose effect curves (TCP, NTCP) Early stage tumours: 1 % increase accuracy 2 % increase in cure rate (Boyer and Schultheiss, 1988) Mid range dose effect curve: 5 % change in dose % change in TCP, % change in NTCP
7 Clinically detectable effects Clinical observation: tumor regression (Dutreix, 1984): Study the difference between electron and photon treatment (squamous cell carcinoma tonsil (amigdale)) smaller efficiency electrons: but 7 % dose error in electron dose calculation Gynecological patients: 25 MV: radiotherapist reported skin reactions and diarrhea Dosimetric study: calibration problem linac: 7 %
8 H&N Patotids
9 Clinical trials Dose escalation studies: e.g. lung: impact of accurate dose calculations > 20 % (what if 3x20 Gy is actually 3x20 for some patients, but only 3x15 for other patients?)
10 Achievable accuracy AAPM Report 85, Mijnheer et al. 1987, ICRU 1976, Brahme 1988: 5 % global uncertainty (1 stdev) TPS: 3 % accuracy decreasing in the future 2 % providing an overall uncertainty of 3 to 4 % Dose calibration: Now 2.5 % going to 1 % Positioning: Now: 2.5 % going to 2 % Patient Data: 1.5 % going to 1 %. Accurate algorithms needed, especially in lung!
11 2 main categories: Overview Dose calculation algorithms Correction based algorithms: Based on measured data (PDD, TMR, OAR, OFs, ) Model-based algorithms: Modeling dose distribution based on first principles: source model, patient information
12 Correction based algorithms Start from phantom measurements: PDD, TAR, TPR, OAR, OFs Contour correction/ beam obliquity: Effective SSD, Isodose shift, TAR/TMR method Heterogeneity corrections: Multiple solutions ranging from 1D density corrections to 3D correction methods
13
14
15 1D/3D corrections for heterogeneities: Density scaling Radiological pathlength: ρ.t (g/cm 2 ): Compton electron density Dose = Dose_primary + Dose_scatter If beam traverses layer density < 1.0 Attenuation Dose_primary Scatter Dose_scatter 1D correction ignores reduction of scatter contribution overestimates effect heterogeneity Also field size has to be scaled: D(t,r) = D(ρt,ρr)
16 Heterogeneity corrections: Categories According to AAPM Report No. 85 Electron transport: using kernels 1D, 3D correction: scaling of primary dose, scatter dose, sampling anatomy along one dimensional rays, or the full three dimensions Category 1: 1D correction, no electron transport (local energy deposition) Category 2: 3D correction, no electron transport Category 3: 1D correction, electron transport Category 4: 3D correction, electron transport
17 Model-based algorithms Primary dose plus first order Compton scatter: monoenergetic parallel beam ignoring heterogeneities Convolution-superposition Monte Carlo
18 Categories Category 1: Linear attenuation, RTAR (effective SSD, isodose shift), Power law (Batho) Category 2: ETAR, DSAR, DVOL, dtar, 3D beam subtraction Category 3: PB convolution, FFT techniques Category 4: Superposition/convolution, Monte Carlo, Acuros XB (grid-based Boltzmann equation solver*) * Vassiliev et al, 2010; Fogliata, 2011
19
20 Practical Superposition/convolution To increase calculation speed Pre-convolve in depth: Pencil Beam Correction factor interpolation (Aspradakis and Redpath (1997) Collapsed cone convolution method: kernel represented by set of cones Ahnesjö, 1992: CCC algorithm of Helax, Masterplan Mackie et al 1985, Reckwerdt and Mackie, 1992: Pinnacle Fast Fourier transform: spatially invariant kernel (Wong et al 1997) AAA (Ulmer et al 2005): PB algorithm: additional degrees of freedom for lateral scatter kernels: inhomogeneities: lateral density scaling
21 Kernels : simplification 2D = "pencil beam" Pre convolution => pencil beams Faster calculation Lateral scattering
22 Superposition/convolution (Battista and Sharpe, 1992)
23 Dose calculations algorithms used in commercial TPSs Iplan (Brainlab): PB (category 3), Monte Carlo dose engine (based on XVMC, M. Fippel). Multiplan (Accuray): Ray-tracing of 1 st category, Monte Carlo dose engine (based on MCSIM, C. Ma). GammaPlan: Ray-tracing of 1 st category (only intracranial), Collapsed cone (category 4) Tomotherapy: superposition/convolution: category 4 (Lu et al 2005) Masterplan (Nucletron): PB (category 3), CCC (category 4) and MC for electrons (based on VMC++, Kawrakow) Xio (CMS, Elekta): superposition/convolution, Monaco: Monte Carlo (XVMC) Eclipse (Varian): Batho, Modified Batho (1 st category), AAA (category 3-4), Acuros XB (category 4) Dosisoft: Monte Carlo (based on Penelope, Salvat).
24 Monte Carlo: the truth???
25 Monte Carlo
26 MCTP: fundamentals Variance reduction and approximations CT conversion Dose to medium/dose to water Linac head modelling De-noising 4D treatment planning
27 Approximations - Transport parameters: - PCUT - ECUT - Region rejection: When electron is in less interesting region: stop transport - Kerma approximation (ECUT ) Bremsstrahlung and secondary electrons!!!
28 Variance reduction techniques Spend more calculation time on «interesting particles» «Does not introduce a bias» Particle splitting Russian roulette Interaction forcing Woodcock tracing Importance sampling Stratified sampling (or quasi RNs) Correlated sampling
29 Example: Particle splitting Split particle in e.g. 10 sub-particles with a weight of 1/10 e.g. Splitting of bremsstrahlung photons in target Risk: under-sampling
30 Example: under-sampling
31 Denoising Smoothing of DVHs Denoising 3D distributions Factor 2 to 3. Otherwise smoothing (gaussian filter)
32
33 Linac head modelling for MCTP Fixed components Multiple Source model Beam modifiers: ray tracing or other approximations
34 Virtual source model Origin plane relative = rel = s - o r o o (x o,y o ) PS-scoring plane (x,y,u,v,e) r s Sampling plane s (x s,y s,e) = (r s, s,e)
35 Beam modifiers
36 4D TP e.g. lung treatment MCTP ideal for 4D TP CT data at different time intervals breathing cycle Calculation time 4D = 3D (Keall et al)
37 4D MC
38 MCTP dose engines For QA: - DOSXYZnrc/BEAMnrc engines - Engines based on MCNP, GEANT, Penelope - DPM (Sempau et al) - MCV (Siebers et al) Commercially used: - MCDOSE, MCSIM (Ma et al) - VMC, VMC++ (Kawrakow) - XVMC (Fippel) - Peregrine (Hartmann Siantar et al)
39 VMC, XVMC, VMC++ HU Cross sections Efficient Boundary crossing VRTs: particle splitting, russian roulette, history repitition Higher ECUT, transporting low energy e- with CSDA range Kerma approximation to higher order photons Directional brems splitting
40 MCDOSE/MCSIM EGS4/BEAM Simplified virtual source model Ray tracing through beam modifiers VRTs
41 MCDOSE (Li et al, 2000) XVMC (Fippel et al, 1999)
42 Multiplan (Accuray) (Wilcox and Daskalov 2008)
43 I.5 Why MCTP? Overall uncertainty clinical dose delivery: 5 %; 3% on dose calculation State of the art non-mctp: superposition/convolution: deviations above 5% due to hererogeneities; composition tissue not taken into account MC: Generally stated: within 2 %
44 QA: Phantom studies
45 Martens et al, 2002, Med. Phys. 29, Why MC? (Martens et al, 2002)
46 A. Fogliota et al. 2007
47 Leal et al, 2003: PB/vs Monte Carlo, prostate plan Clinical cases
48 MCDE/Peregrine (Reynaert et al, 2005)
49 DVHs (MCDE/Peregrine)
50 Dose (Gy) 2x2 offset field: AB Measurement MCDE Peregrine X (cm)
51 VMC++ Masterplan (Nucletron), Cygler et al, 2004
52
53 Extreme case: Cyberknife Small target (but tracking was possible) Extremely low lung density
54 Ray- Tracing : 3 x 20 Gy Monte-Carlo :
55 Dose (Gy) Ray-Tracing MC Ray-Tracing MC GTV PTV D2% (max) D50% D98% (min)
56 Same Patient: Masterplan (Nucletron) SBRT Plan for 6 MV Clinac
57 Pencil beam vs Collapsed cone
58 Very important difference between algorithms of category 3 and 4 Especially for the PTV (low density part) -- Collapsed cone Pencil Beam
59 Summary For SBRT lung treatment, a dose calculation of category 4 is required. Monte Carlo or superposition/convolution algorithms Prescription should be based on these algorithms to have less inter-patient differences (clinical trials e.g.) Even these 4 th category dose calculation algorithms should be carefully benchmarked. For Monte Carlo algorithms: impact of noise, denoising, resolution, commissioning (output factors measurements, ) 4D optimization: Added value of MC algorithm Dose to water/dose to medium?
Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations
Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations Daryoush Sheikh-Bagheri, Ph.D. 1, Iwan Kawrakow, Ph.D. 2, Blake Walters, M.Sc. 2, and David W.O. Rogers, Ph.D.
Variance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS
AAPM REPORT NO. 85 TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS Report of Task Group No. 65 of the Radiation Therapy Committee of the American Association of Physicists in Medicine Members
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
Advanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
Independent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN 92-0-107304-6):
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
Our Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers
Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement
Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.
Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
IAEA-TECDOC-1540. Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems April 2007 IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY W. PARKER, H. PATROCINIO Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 7.1. INTRODUCTION
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical
A review of the Monte Carlo method in external photon beam treatment planning. Indrin J. Chetty Henry Ford Health System
A review of the Monte Carlo method in external photon beam treatment planning Indrin J. Chetty Henry Ford Health System Outline. Definition of Efficiency - Factors that render current MC codes fast enough
Bremsstrahlung Splitting Overview
March 2007 Bremsstrahlung Splitting Overview Jane Tinslay, SLAC Overview & Applications Biases by enhancing secondary production Aim to increase statistics in region of interest while reducing time spent
Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY D.I. THWAITES Department of Oncology Physics, Edinburgh Cancer Centre, University of Edinburgh, Western General Hospital, Edinburgh, United Kingdom
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
The Monte Carlo Simulation of Radiation Transport
The Monte Carlo Simulation of Radiation Transport Iwan Kawrakow Ionizing Radiation Standards, NRC, Ottawa, Canada The Monte Carlo Simulation of Radiation Transport p.1/35 Contents History & application
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
Innovative RT SBRT. The variables with REQ in superscript are required.
The variables with REQ in superscript are required. The variables with a are single-select variables; only one answer can be selected. The variables with a are multi-select variables; multiple answers
The feasibility of a QA program for ISIORT Trials
Baveno, Italy June 22-24, 2012 The feasibility of a QA program for ISIORT Trials Frank W. Hensley 1, Don A. Goer 2, Sebastian Adamczyk 3, Falk Roeder 1, Felix Sedlmayer 4, Peter Kopp 4 1 University Clinics
Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD
Good Morning Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD Barbara Agrimson, BS RT(T)(R), CMD Steve Rhodes, BS RT(T), CMD Disclaimer This presentation
Challenges in small field MV photon dosimetry
Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland [email protected] Why are
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
Key words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect [email protected]. Protect, Enhance and Save Lives
Vincent Bossier System Architect [email protected] IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
CHAPTER 11. COMPUTERIZED TREATMENT PLANNING SYSTEMS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 11. COMPUTERIZED TREATMENT PLANNING SYSTEMS FOR EXTERNAL PHOTON BEAM RADIOTHERAPY MICHAEL D.C. EVANS Department of Medical
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
ACCELERATORS AND MEDICAL PHYSICS 2
ACCELERATORS AND MEDICAL PHYSICS 2 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 2-28.10.10 - U. Amaldi 1 The icone of radiation therapy Radiation beam in matter EPFL 2-28.10.10 - U.
Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.
Protons vs. CyberKnife UC SF Protons vs. CyberKnife UC SF Alexander R. Gottschalk, M.D., Ph.D. Associate Professor and Director of the CyberKnife Radiosurgery Program Department of Radiation Oncology University
External dosimetry Dosimetry in new radiotherapeutic techniques
External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France [email protected] Objective : To describe
Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation
Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation Deborah Schofield, Laurence Court, Aaron Allen, Fred Hacker, Maria Czerminska Department of Radiation Oncology Dana
Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
Conversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry
Conversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry ORAMED project Rapport CEA J. Daures (1), J. Gouriou (1), JM Bordy (1) (1) CEA, LIST, Laboratoire National
Status and perspective of emission imaging techniques for ion beam therapy in Lyon
14th International Conference on Nuclear Reaction Mechanisms, Varenna 18th June 2015 Status and perspective of emission imaging techniques for ion beam therapy in Lyon on behalf of the CAS-PHABIO group
Accreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
02 CyberKnife: Treatment Delivery
TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options
Oncentra Brachy Comprehensive treatment planning for brachytherapy. Time to focus on what counts
Oncentra Brachy Comprehensive treatment planning for brachytherapy Time to focus on what counts Time to focus on what counts In contemporary brachytherapy, the medical physicist needs to process an increasing
REPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE
E S T I M A T I O N O F T H E I M A G I N G D O S E F O R T H E C Y B E R K N I F E R O B O T I C R A D I O S U R G E R Y S Y S T E M REPORT ORGANIZATION ABSTRACT I. INTRODUCTION II. III. IV. CALCULATION
Chapter XIII. PDD TAR TMR Dose Free Space
Ver 2.11 Chapter XIII PDD TAR TMR Dose Free Space Joseph F. Buono RTT Allied Health Science Nassau Community College 1 Education Drive Garden City, NY 11530-6793 phone: 516-572 - 7536 oice - 9460 Secretary
QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS
QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS Ben Mijnheer Agnieszka
CSE168 Computer Graphics II, Rendering. Spring 2006 Matthias Zwicker
CSE168 Computer Graphics II, Rendering Spring 2006 Matthias Zwicker Last time Global illumination Light transport notation Path tracing Sampling patterns Reflection vs. rendering equation Reflection equation
CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY WILLIAM PARKER Department of Medical Physics McGill
Quality Assurance in Stereotactic. Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
Quality Assurance for Treatment Planning Systems
Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline
MONTE CARLO MODELING OF A VARIAN 2100C 18 MV MEGAVOLTAGE PHOTON BEAM AND SUBSEQUENT DOSE DELIVERY USING MCNP5
MONTE CARLO MODELING OF A VARIAN 2100C 18 MV MEGAVOLTAGE PHOTON BEAM AND SUBSEQUENT DOSE DELIVERY USING MCNP5 A Thesis Presented to The Academic Faculty by Jared Hoover In Partial Fulfillment of the Requirements
3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer
Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used
IMRT: dal presente al futuro L esperienza italiana
IMRT: dal presente al futuro L esperienza italiana Giuseppe Scielzo U.O.A. Fisica Sanitaria Ordine Mauriziano Reggio Emilia, 14-15 Ottobre 2003 Contents IMRT in Italy - AIFM BEV -AIRO BEV - OEV Breaking
MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
VARIANCE REDUCTION TECHNIQUES FOR IMPLICIT MONTE CARLO SIMULATIONS
VARIANCE REDUCTION TECHNIQUES FOR IMPLICIT MONTE CARLO SIMULATIONS An Undergraduate Research Scholars Thesis by JACOB TAYLOR LANDMAN Submitted to Honors and Undergraduate Research Texas A&M University
Übersicht von IMRT Konzepten bei CMS. Norbert Steinhöfel DGMP AK IMRT 19.-20.2.2002
Übersicht von IMRT Konzepten bei CMS Norbert Steinhöfel DGMP AK IMRT 19.-20.2.2002 1 CMS: Radiation Oncology Software An Integrated Department From Start to Finish CMS provides integration from CT simulation
Radiation Protection in Radiotherapy
Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France [email protected] Radiation therapy The lecture is oriented to
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
How To Improve Efficiency In Ray Tracing
CS 563 Advanced Topics in Computer Graphics Russian Roulette - Sampling Reflectance Functions by Alex White Monte Carlo Ray Tracing Monte Carlo In ray tracing, use randomness to evaluate higher dimensional
Veterinary Oncology: The Lumps We Hate To Treat
Veterinary Oncology: The Lumps We Hate To Treat Michelle Turek, DVM, DACVIM (Oncology), DACVR (Radiation Oncology) College of Veterinary Medicine University of Georgia Athens, GA Veterinary Oncology Veterinary
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
Deformable Registration for Image-Guided Radiation Therapy
Symposium Multimodality Imaging for Radiotherapy : State of the art, needs and perspectives 1/33 Deformable Registration for Image-Guided Radiation Therapy David Sarrut «Radiation, images, oncology» team
Proton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota
Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler [email protected] Overview Background of 3D Printing Practical Information Current
Self-adjusting Importances for the Acceleration of MCBEND
Self-adjusting Importances for the Acceleration of MCBEND Edmund Shuttleworth Abstract The principal method of variance reduction in MCBEND is the use of splitting and Russian roulette under the control
IMRT for Prostate Cancer. Robert A. Price Jr., Ph.D. Philadelphia, PA
IMRT for Prostate Cancer Robert A. Price Jr., Ph.D. Philadelphia, PA Number of Patients 16 14 12 1 8 6 4 1481 IMRT Patients at FCCC 293 97 Prostate Breast H&N Other 64 Approximately 13-15 patients per
Use of 3D Printers in Proton Therapy
Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology [email protected] 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing
Medical Applications of radiation physics. Riccardo Faccini Universita di Roma La Sapienza
Medical Applications of radiation physics Riccardo Faccini Universita di Roma La Sapienza Outlook Introduction to radiation which one? how does it interact with matter? how is it generated? Diagnostics
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
Follow this and additional works at: http://digitalcommons.wustl.edu/open_access_pubs
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2012 Evaluation of the efficiency and effectiveness of independent dose calculation followed by machine log file
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Current Status of Proton Clinical Activities at PTC H
Current Status of Proton Clinical Activities at PTC H Michael Gillin, PhD, Professor, Chief of Clinical Services Department of Radiation Physics, UT MDACC The University of Texas M.D. Anderson Cancer Center
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
Measuring Line Edge Roughness: Fluctuations in Uncertainty
Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as
High Quality Image Magnification using Cross-Scale Self-Similarity
High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg
Machine Learning for Medical Image Analysis. A. Criminisi & the InnerEye team @ MSRC
Machine Learning for Medical Image Analysis A. Criminisi & the InnerEye team @ MSRC Medical image analysis the goal Automatic, semantic analysis and quantification of what observed in medical scans Brain
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.
The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. KAROLINA KOKUREWICZ S U P E R V I S O R S : D I N O J A R O S Z Y N S K I, G I U S E P P E S C
Project Management Triangle
Session: Optimizing the Treatment Planning Process Maintain the Quality of Treatment Planning for Time-Constraint Cases Jenghwa Chang, Ph.D. Radiation Oncology, NewYork-Presbyterian Hospital/Weill Cornell
SBRT IMRT SBRT IMRT. SBRT Process in a Nut Shell. SBRT: Simulation, Localization, and Delivery. Moyed Miften, PhD University of Colorado TG 101
SBRT: Simulation, Localization, and Delivery Moyed Miften, PhD University of Colorado SBRT Process in a Nut Shell Immobilization and Simulation Motion Management Treatment Planning (TP) Image Fusion Planning
Appendix A. An Overview of Monte Carlo N-Particle Software
Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation
Particle Therapy for Lung Cancer. Bradford Hoppe MD, MPH Assistant Professor University of Florida [email protected]
Particle Therapy for Lung Cancer Bradford Hoppe MD, MPH Assistant Professor University of Florida [email protected] Content Rationale for Particle Therapy in Lung Cancer Proof of Principle Treatment
RADIOTHERAPY Giovanna Mantello
F The impact on RADIOTHERAPY Giovanna Mantello AOU Osp. Riuniti Ancona [email protected] RADIOTHERAPY OF PULMONARY LESIONS INTENSITY MODULATION HIGH GRADIENT DOSE DOSE ESCALATION IPOFRACTIONATION STEREO BODY
Bildverarbeitung und Mustererkennung Image Processing and Pattern Recognition
Bildverarbeitung und Mustererkennung Image Processing and Pattern Recognition 1. Image Pre-Processing - Pixel Brightness Transformation - Geometric Transformation - Image Denoising 1 1. Image Pre-Processing
Monte Carlo Simulations in Proton Dosimetry with Geant4
Monte Carlo Simulations in Proton Dosimetry with Geant4 Zdenek Moravek, Ludwig Bogner Klinik und Poliklinik für Strahlentherapie Universität Regensburg Objectives of the Study what particles and how much
Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,
HADRON THERAPY FOR CANCER TREATMENT
HADRON THERAPY FOR CANCER TREATMENT Seminar presented by Arlene Lennox at Fermilab on Nov 21, 2003 CANCER STAGES LOCAL TUMOR REGIONAL METASTASIS SYSTEMIC DISEASE CANCER TREATMENT SURGERY RADIATION THERAPY