Guideline for the validation of packaging processes according to ISO
|
|
|
- Claude Lawrence
- 10 years ago
- Views:
Transcription
1 D 2596 F Supplement 2012 Guideline for the validation of packaging processes according to ISO Official publication of the German Society for Sterile Supply (DGSV e.v.)
2 HAWO. PERFECTLY VALIDATED SEALING PROCESSES Being a part of the sterile goods packaging process, the sealing process also has to be validated in accordance with ISO the new packaging guideline sets out what has to be done. hawo offers compatible heat sealers and testing systems. HEAT SEALERS FOR HOSPITALS AND INDUSTRY Rotary sealers are used to package instruments in hospitals and medical devices in the industrial sector. From the extremely compact and award-winning hd 680/hm 780 series to the particularly powerful pro-class hm 850/880 DC-V (see illustration) and hm 3010/20 DC-V models, virtually all hawo devices in this class feature intelligent monitoring, documentation technologies and intuitive device operation (IntelligentScan). All of the sealing devices marked with V (e.g. hm 850 DC-V) satisfy the requirements for process validation in accordance with ISO and have interfaces for connection to tracking systems. HEAT SEALERS FOR DOCTOR S SURGERIES For the secure sterile packaging of instuments in doctors and dentists practices, hawo offers the particular compact bar and rotary sealers. The top-of-the-range ValiPak hd 380 WSI-V model (see illustration) sets the standard in the clinical practice sector with a fully ISO validatable process and interface for connection to practice software. Tyvek is a registered trademark of E. I. du Pont de Nemours. TESTING SYSTEMS hawo offers two testing systems for the routine monitoring of sealing seams. > Seal Check: The Seal Check med indicator strips for fi lm pouches and reels made from paper / fi lm and Seal Check HDPE (Tyvek / fi lm) make faulty areas visible. > hawo InkTest: The new dye penetration test for testing the seal integrity in accordance with ISO is distinguished by its simple handling and delivers objective results. Scan the QR code and download the guideline free of charge.
3 Central Service Suppl foreword 1 Foreword The main purpose of any packaging system used for terminally sterilized medical devices is to preserve sterility until use as well as to allow aseptic presentation at the point of use on the patient. Validation of packaging processes is crucial to guarantee that the integrity of the packaging system is always assured and maintained during transport and storage until the time of use. The international packaging standard ISO calls for suitable validated packaging processes for medical devices. This standard is applicable to the medical industry, to health care facilities (hospitals, doctors and dentists), and wherever medical devices are packaged and sterilized. The packaging process is one of the links in the process chain of medical device reprocessing and, as such, must be validated. The establishment of a quality management system is an indispensable prerequisite for validation and for assuring reproducibility and ongoing effectiveness of medical device reprocessing. Without a quality management system validation is not possible since all steps must be defined and documented. All products and materials used must in principle meet the normative requirements. The quality management system must specify how bought-in products and services are audited and evaluated. However, the focus of this Guideline is not on audit and evaluation. The international standard ISO describes essential requirements for sterile barrier systems, while the ISO standard describes validation of packaging processes. Detailed quality requirements for sterile barrier systems are outlined in the European CEN standards EN to 10. They serve as a basis for this Guideline and as an orientation guide for conducting validation in practice. Experiences gained from the implementation of the requirements for validation of cleaning, disinfection and sterilization processes have highlighted the need for a practice-oriented and feasible guide for the implementation of the normative requirements so that, as far as possible, they will be similarly interpreted by operators and validators. The focus on uniform and proper conduct of validation of packaging processes is of paramount importance for everyone involved in this process as well as for the supervisory authorities and certification bodies, not least to avoid «confusion». The authors point out that this Guideline is meant as a practical orientation guide. No guarantee of completeness is given. Authors of the Guideline: German Society of Sterile Supply (Deutsche Gesellschaft für Sterilgutversorgung, DGSV e. V.): M. Peißker, A. Schlepp, M. Lüttenberg, A. Carter, A. Jones 1, B. Amann, K. Wiese 1 Section «pouch, bag or reel sealing».
4 Table of Contents Guideline for Validation of Packaging Processes as per ISO Foreword 5 1 Scope 5 2 Normative bases 5 3 Prerequisites 5 4 Validation of packaging processes Drafting of a validation plan Conduct of validation Installation Qualification (IQ) Operational Qualification (OQ) Performance Qualification (PQ) Drafting of a validation report Formal approval of the validation process Process control and monitoring Process changes and revalidation Annex A 11 Annex A.1: Validation plan checklist «pouch, reel or bag sealing» 14 Annex A.2: Installation qualification (IQ) checklist «pouch, reel or bag sealing» 18 Annex A.3: Operational qualification (OQ) checklist «pouch, reel or bag sealing» 19 Annex A.4: Performance qualification (PQ) checklist «pouch, reel or bag sealing» 20 Annex A.5: Example for determining the scope of process validation per heat sealer 21 Annex A.6: Sample standard operating procedure «pouch, reel or bag sealing» 23 Annex A.7: Sample standard operating procedure for verification of seal seams
5 Annex B 25 Annex B.1: Validation plan checklist «sterilization sheets' folding and wrapping» 30 Annex B.2: Installation qualification (IQ) checklist «sterilization sheets' folding and wrapping» When you are surfing the web, you can now visit Central Service. 31 Annex B.3: Operational qualification (OQ) checklist «sterilization sheets' folding and wrapping» 32 Annex B.4: Performance qualification (PQ) checklist «sterilization sheets' folding and wrapping» 34 Annex B.5: Example for determining the scope of process validation per packaging material in combination with the sterilization processes 35 Annex B.6: Sample standard operating procedures «sterilization sheets' folding and wrapping» 39 Annexe B.7 : Sample specification and sample data sheet, e. g. for «sterilization sheets» Annex C 42 Annex C.1: Validation plan checklist for «filling and closing of reusable sterilization containers» Central Service is now accessible online as E-Paper, featuring the most convenient online platform currently available on the market. Order your access code now via from [email protected]. 48 Annex C.2: Installation qualification (IQ) checklist (IQ) «filling and closing of reusable sterilization containers» 50 Annex C.3: Operational qualification (IQ) checklist (OQ) «filling and closing of reusable sterilization containers» 51 Annex C.4: Performance qualification (PQ) checklist (PQ) «filling and closing of reusable sterilization containers» 53 Annex C.5: Example for determining the scope of process validation per sterilization container in combination with the sterilization processes 54 Annex C.6: Sample standard operating procedure: «reusable sterilization containers» Annex D 55 Annex D: Sample data sheet «sterilization markers»
6 Jan Huys HEART Consultancy STERILIZATION of Medical Supplies by Steam Volume I General theory Third Revised Edition mhp Verlag GmbH Wiesbaden Jan Huys HEART Consultancy STERILIZATION of Medical Supplies by Steam Volume I General theory Third Revised Edition mhp Verlag GmbH Wiesbaden
7 Central Service Suppl recommendations 5 Guideline for Validation of Packaging Processes according to ISO Scope The standard series ISO stipulates validation of the packaging processes used for industry, health care facilities and wherever medical devices are packaged and sterilized (examples of health care facilities include hospitals, doctors' and dentists' surgeries). The ISO 11607, Part 2 standard (Article 5.1.1) explicitly calls for validation of all packaging processes. The present Guideline deals with the following packaging processes: pouch, reel or bag sealing 2 sterilization sheets folding and wrapping filling and closing of reusable sterilization containers Likewise, packaging processes not dealt with here must also be validated as per ISO Non-validable packaging processes are not acceptable in practice anymore (Self Seal pouches or taped paper bags). 2 Normative bases The bases for drafting this Guideline include, inter alia, the following standards 3 : 2 If the sealing processes were already validated in accordance with the «Guideline for validation of the sealing process as per ISO (Revision 1, status: July 2008)», there is no need to repeat initial validation. 3 The publication years of the pertinent standards are only given here. 4 EN 868, Part 1 has been replaced by the ISO standard. 5 German Standard DIN 58953, Parts 2 5 have been replaced by EN 868, Parts 2 5. ISO :2009 ISO :2006 EN 868:2009, Part ISO :2009 ISO 9001:2008 ISO 13485:2010 DIN 58953:2010, Part 1, 6, 7, 8, 9 5 (German Standard) The standards stated in table 1 are of relevance for validation and should be made accessible to the user. 3 Prerequisites The packaging materials used must be suited to and defined for the intended packaging and sterilization processes. Suitability shall be determined on the basis of the information provided by the manufacturer. This includes confirmation of conformity with the ISO standard and pertinent sections of the EN 868, Parts 2 10 standard series, in respect of: microbial impermeability compatibility with the sterilization process. The number of process validations to be conducted can be elucidated and defined on the basis of Table 2 (see example Annex A.5, B.5 and C.5). The number of combinations outlined in the table can be reduced by taking account of only the maximum material stress Table 1: Standards of relevance for the validation ISO ISO (worst-case scenario, while providing documentary proof to justify this). Worst-case examples: Gusseted pouches and reels are more critical than flat pouches and reels. Steam sterilization at 134 C/18 min is more critical than at 134 C/5 min and 121 C/20 min. A further reduction can be achieved by a deliberate choice of packaging materials (e. g. see through pouch instead of paper bag). Annex A.5, B.5 and C.5 show practical examples. 4 Validation of packaging processes In principle, a documented process must be available for validation. This process comprises: 4.1 Drafting of a validation plan 4.2 Validation of packaging processes Installation qualification (IQ) Operational qualification (OQ) Performance qualification (PQ) 4.3 Drafting of a validation report 4.4 Formal approval of validation 4.5 Process control and monitoring 4.6 Process changes and revalidation 4.1 Drafting of a validation plan The validation plan should contain, at least, the following details: Requirements for materials, sterile barrier systems and packaging systems Validation requirements
8 6 recommendations Central Service Suppl Table 2: Number of process validations to be conducted The terms used for sterilization processes are based on the standard ISO Sterile barrier system (SBS) Material A Material B Material C Material D 134 C/ 5 min STEAM 134 C/ 18 min 121 C/ 20 min Competences Description of the packaging process Description of the materials /equipment Description of the sterilization processes Qualification steps (IQ, OQ and PQ) The «Validation plan» checklists in Annex A.1, B.1 and C.1 can be used. 4.2 Conduct of validation Installation qualification (IQ) Definition: «Process of obtaining and documenting evidence that equipment has been provided and installed in accordance with the specification.» That means that technical equipment (e. g. heat sealers) must have been properly installed and users trained. In general, the packing processes involving sterilization sheets as well as reusable sterilization containers are purely manual processes, which is why proof of IQ is based on documentation of training of staff. It is recommended that the corresponding checklists be used to conduct installation qualification (IQ). The «Installation qualification (IQ)» checklists in Annex A.2, B.2 and C.2 can be used for documentary purposes Operational qualification (OQ) Definition: «Process of obtaining and documenting evidence that installed equipment operates within predetermined limits when used in accordance with its operational procedures.» FORM (Formaldehyde) EO (Ethylene oxide) VH2O2 (vaporized hydrogenperoxyde; «Plasma») The «Operational qualification (OQ)» checklists in Annex A.3, B.3 and C.3 can be used for documentary purposes. In principle, a distinction must be made here between automated and manual processes. Automated processes Here: pouch, reel or bag sealing. The heat sealing process is defined on the basis of the following parameters: Sealing temperature, Contact pressure and Sealing time/speed (dwell). The contact pressure and sealing speed or time (dwell) are generally set by the manufacturer of the heat sealer. The optimum sealing temperature for the respective packaging material must be determined by the user. To that effect, the technical data sheet supplied by the manufacturer of the packaging material is needed. This must specify the sealing temperature (e. g C). Sealing samples must be produced for the respective lower and upper limits. The quality properties listed in ISO , b must be assured: intact seal for a specified seal width no channels or open seals no punctures or tears no material delamination or separation These quality properties must be verified and documented by means of suitable processes. The test methods in Table 3, for example, can be used as a guide. Then the sealing temperature must be specified for routine operations. In general this is calculated from the mean value of the limit values (e. g. mean value from 170 C and 200 C is 185 C). Manual processes Here: sterilization sheets' folding and wrapping; filling and closing of reusable sterilization containers 9. First, the most critical packaging configuration must be determined (worst case). Examples include: the heaviest and largest tray (container) large, unwieldy single instruments Then these configurations must be packed according to the standard operating procedures. When checking the sterile barrier systems produced all defined quality properties as well as the correct packing method set out in the standard operating procedure (see Annex B.6 and C.6) must be assured. Pursuant to the ISO , c standard the quality properties required for sterilization sheets and reusable sterilization containers are as follows: continuous closeness/integrity no punctures or tears (not applicable to reusable sterilization containers) no other visible damage or material irregularities 10. The quality properties must be verified and documented by means of suitable processes or tests. For the combinations specified in the validation plan, 10 sterile barrier systems of the same material must be packed and their quality properties checked. To document the quality properties it is recommended that at least one photo is taken of each sample Performance qualification (PQ) Definition: «Process of obtaining and documenting evidence that the equipment, as installed and operated in accordance with operational procedures, consistently performs in accordance with predetermined criteria and thereby yields product meeting its specification.» During performance qualification proof must be provided after sterilization that the process is under control and produces optimally sealed or closed sterile barrier systems.
9 Central Service Suppl recommendations 7 Table 3: Test methods for verification of quality properties Test method Seal integrity test (e. g. «dye penetration test/inktest» according to ISO , Annex B 6 ) Suitable for verification of the following quality properties channels or open seals punctures or tears no channels or open seals no punctures or tears no material delamination or separation These quality properties must be verified and documented by means of suitable processes. The test methods in Table 3, for example, can be used as a guide. Seal integrity indicator 7 (e. g. Seal Check) Peel test according to EN 868, Annex E Visual inspection 8 intact seal for a specified seal width channels or open seals punctures or tears material delamination or separation intact seal for a specified seal width punctures or tears Manual processes For the test, sterilized packaging systems must be taken from the running processes. From three different cycles (batches) one sample must be taken in each case. The batch documentation (protocols) of the respective sterilization processes is part of validation. The «Performance qualification (PQ)» checklists in Annex A.4, B.4 and C.4 can be used for documentation purposes. Here, too, a distinction must be made between automated and manual processes. Automated processes Verification is done by means of the seal strength test as per EN 868-5, Annex D 11. The packaging must be sterilized before verification. The protocols/logs (batch documentation) related to the sterilization processes are part of validation. For the defined combinations (see also Annex A.5) three empty pouches or reels of the same material must be sealed, clearly labelled (sealing device, serial number, sealing parameters) and then sterilized with the specified sterilization program (reels must be sealed at both ends). Each pouch must be added to a different sterilization load to take account of all the factors exerting an influence on the sterilization loads. The test (as per EN 868-5, Annex D) is carried out as follows: Cuts measuring 15 mm in width are taken of the dried samples and at an angle of 90 to the seal seam. At least one sample of a produced seal seam must be taken from each packaging 12. If only one sample of a seal seam is taken, the sample must be taken from around the centre. Simulation of the peeling process at a speed of 200 mm/min Recording of the seal seam strength 13 Evaluation and documentation of the results The results of the seal strength test are confirmed in a report, containing at least the following information: Manufacturer and type of heat sealer Serial number of heat sealer Specification of the sealing parameters Identification of the verified product Maximum strength of seal of each sample measured in N/15 mm width Whether verification was done with the free end supported or not The frequency used (data per second of measurement) Test device (manufacturer, designation)/ last calibration Graphic display of resistance Date of test Testing of the sealed and sterilized pouches can, for example, be carried out by an accredited test laboratory or by the device/ material manufacturer. The maximum strength must be entered into the table in Annex A.4. The maximum strength is the relevant value for assessment and, as per EN 868-5, must be greater than or equal to 1.5 N/15 mm width 14. If the maximum tensile strength of one of the three tests is less than 1.5 N/15 mm width, PQ is deemed to have failed. In addition the quality properties listed in ISO , b must be assured: intact seal for a specified seal width 6 The basis for this test method is ASTM F «Standard test method for detecting seal leaks in porous medical packaging by dye penetration» 7 The seal integrity indicator must not under any circumstances be cut since it must always be guaranteed that the entire pinch roller of the sealing device is printed off. Furthermore, the seal indicator shall always be made of the same type of material as the porous part of the packaging (medical grade paper as per EN or HDPE as per EN 868-9/10) 8 For visual inspection standardized test methods can be used (e. g. ISO , Annex B [ASTM F1886]) 9 The partial step «Filling of pouches and reels» is also a manual process and must be set out in a standard operating procedure. The heat sealing process itself is normally fully automated. 10 The ISO standard uses «No material delamination or separation» here. 11 Alternatively, the test method as per ASTM F88 can be used (validated and round robin approved test method). 12 EN 868-5:1999 specified five samples per seal seam. EN 868-5:2009 stipulates only one sample per seal seam. Additional samples may be needed if the length of a seal is more than 500 mm. 13 For further evaluation and documentation it is advisable to specify as a value the maximum (required as per EN Annex D.3) and additionally the average tensile strength. 14 EN 868-5, «The minimum seal strength value for steam sterilization processes must be 1.5 N per 15 mm in health care facilities and 1.2 N per 15 mm in other sterilization processes in health care facilities». However, stipulation of a minimum value of 1.5 N/15 mm is recommended for all sterilization processes.
10 8 recommendations Central Service Suppl Assurance of the quality properties must be verified for each packaging system (sample). Pursuant to standard ISO , c the quality properties for sterilization sheets and reusable sterilization containers are as follows: continuous closeness/integrity no punctures or tears (not applicable to reusable sterilization containers) no other visible damage or material irregularities 15. These quality properties must be verified and documented by means of suitable processes or tests. The sterile barrier systems or packaging systems are opened one step after the other, verified and documented (for photographic documentation see Annex B.8/C.7). 4.3 Drafting of a validation report The validation procedures and results must be documented in a summary report. The checklists, protocols and any photographic documentation used serve as evidence and must be enclosed in an annex to the report. The report must contain, at least, the following information: Validation plan Evidence of implementation of the validation plan (IQ, OQ and PQ checklists completed as per Annex) Evaluation of the results Photographic documentation for manual packing processes Details and explanation of any deviations from validation plan Formal approval of validation Process control and monitoring Process changes and revalidation 4.4 Formal approval of validation Validation, as documented and evaluated in the report, must be formally approved, and duly documented, by the competent person appointed by the operator. This can be recorded, for example, in a field provided to that effect in the validation plan. If all validation results are not accepted, this must be clearly documented, including assessment of any remaining risks. 4.5 Process control and monitoring The routine tests that are established during the validation as being necessary must be documented (e. g.in the standard operating procedure). This is intended as a means of ensuring that changes in the packaging process are detected on time before they compromise the sterile barrier systems and the requirements are no longer met. These include, e. g.: Visual inspection 16 Peelability (e. g. peel test as per EN 868-5, Annex E «Method for determination of the peel characteristics of paper/plastic laminate products») Seal integrity test (e. g. dye penetration test/ink test as per ISO , Annex B 17 ) Seal integrity indicator 18 (e. g. Seal Check) Tensile strength of seal seam (e. g. determination of seal seam strength as per EN 868-5, Annex D «Method for determination of the seal seam strength of pouches and reels» Stepwise opening of packaging (in the case of sterilization sheets or reusable sterilization containers). Intervals (e. g. daily, weekly, monthly, yearly) and acceptance values must be defined for the routine tests needed, including the action to be taken if a test result is not satisfactory. The routine test results must be documented. This procedure must be set out in the quality management system. 4.6 Process changes and revalidation Processes must be revalidated: Unscheduled revalidation, for example in the event of changes to materials, processes, including changes to equipment or occurring during sterilization (revalidation) Scheduled revalidation, at regular intervals, i. e. in general after one year if no changes were made to materials, sealing process or sterilization (performance requalification). provides evidence that the packaging process continues to be within the limits defined at the time of ini- tial validation (IQ, OQ and PQ). That no changes were made to materials, processes or sterilization compared to the previous validation must be confirmed in the revalidation report. If changes are made to materials, processes or sterilization how such changes will affect the packaging process results must be elucidated. The results must be documented. Based on these, an individual revalidation plan must be drafted. Accordingly, in the event of material changes, for example, operational qualification (OQ) and performance qualification (PQ) must be partially or fully repeated, and if changes are made to the packaging process or to the equipment used installation qualification (IQ) must also be repeated. For revalidation it must be ensured that the documents used meet the current requirements. The checklists must be updated if necessary. An individual validation plan is required for each revalidation or performance requalification. The «Validation plan» checklists in Annex A.1, B.1 and C.1 can be used. 15 The ISO standard uses «No material delamination or separation» here. 16 For visual inspection standardized test methods can be used (e. g. ISO , Annex B [ASTM F1886] for seal seams or EN for reusable sterilization containers. 17 The basis for this test method is ASTM F1929 «Standard test method for detecting seal leaks in porous medical packaging by dye penetration». 18 The seal integrity indicator must not under any circumstances be cut since it must always be guaranteed that the entire pinch roller of the sealing device is printed off. Furthermore, the seal integrity indicator shall always be made of the same type of material as the porous part of the packaging (medical grade paper as per EN or HDPE as per EN 868-9/10)
11
12 Responsibility is our commitment Medical packaging made in Germany. Stericlin quality manufacturing, setting and upholding standards for more than 40years worldwide. Laminates paper/film Laminates Tyvek /film Laminates non woven /film Non woven wrapping SMS wrapping materials Crepe paper wrapping Bowie&Dick test Indicators Batch Monitoring Systems Labels Packaging materials for medical device manufacturers Our quality assures customers around the world that stericlin is a superior choice. It is our daily mission to manufacture the highest quality product in the world market. Tyvek is a registered trademark of DuPont Vereinigte Papierwarenfabriken GmbH
13 Central Service Suppl ANNEX A.1 11 Annex A.1: Validation plan checklist «pouch, reel or bag sealing» 19 Initial validation Revalidation (at regular intervals, only performance requalification) Revalidation for special reasons (e. g. new materials) a) Competences Name of institution (operator) Location Validator (Name of persons, or companies, conducting validation) Responsible for overall validation b) Description of sealing device Manufacturer of sealing device Type of sealer (e. g. rotary sealer) Serial number Supplier Last calibration Contact person c) Description of material Manufacturer Type of material Manufacturer s QM certificate available?* Yes No Evidence Supplier Contact person CE conformity?* Yes No Evidence Specification of material to be sealed */** Paper/foil Tyvek 20 /foil Nonwovens/foil Paper/paper Nonwovens/nonwovens Other: ISO Part 1 conformity?* 21 Yes No Evidence Sealing temperature range (in C)* from to Specification of: Evidence available Compatible with sterilization process* Yes No Evidence * Information featuring an * must, in accordance with EN and ISO , be made available by the manufacturer of the packaging material. ** For each material combination or each category of heat sealable sterile barrier systems a complete checklist must be filled out and the validation process conducted. 19 If other sealing methods are used, a customized checklist must be compiled if necessary. 20 Tyvek is a registered trademark of E.I. du Pont de Nemours. 21 Conformity with ISO is an absolute prerequisite and in general includes conformity with EN Often, CE conformity and conformity with ISO Part 1 are declared jointly in one document.
14 12 ANNEX A.1 Central Service Suppl d) Description of sterilization process Sterilization process STEAM Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process EO (ethylene oxide) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process VH2O2 (plasma) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process FORM (formaldehyde) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation:
15 Central Service Suppl ANNEX A.1 13 Sterilization process Other : Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: e) Qualification steps If this is an initial validation, all three qualification steps (IQ, OQ and PQ) must be carried out as per the checklists in Annex A.2, A.3 and A.4. For revalidation/performance requalification it may be possible to omit some steps. Installation qualification (IQ) executed already executed during validation on: passed failed Date/signature : Operational qualification (OQ) executed already executed during validation on: passed failed Date/signature : Performance qualification (PQ) executed passed failed Date/signature : f) Formal approval of validation/revalidation by the operator All parts of validation/revalidation passed Parts of validation/revalidation failed Measures have been defined and documented Place, date Name Signature
16 14 ANNEX A.2 Central Service Suppl Annex A.2: Installation qualification (IQ) checklist «pouch, reel or bag sealing» 22 Are standard operating procedures (SOPs) available? (example, see Annex A.6) Yes No Where? a) General data Device (designation/number) Manufacturer Manufacturer s address Quality management system Evidence available (certificate): Type of sealer (e. g. rotary sealer) Serial number Year of manufacture Location Responsible for validation Other IQ inspectors Date of test Type of device Bar sealer Rotary sealer Serial device Special device from manufacturer Modified device modified by: CE conformity? 23 Yes No Evidence ISO conformity? 24 Yes No Evidence Service team Address Telephone number Contact person Authorized by the manufacturer Yes, evidence 25 : No 22 If other sealing methods are used, a customized checklist must be compiled if necessary. 23 A heat sealer is neither a medical device nor an accesory to a medical device according the European Medical Device Directive. 24 Conformity with ISO is an absolute prerequisite. 25 Authorization by the manufacturer must be available in the written form.
17 Central Service Suppl ANNEX A.2 15 b) Installation conditions Parameters Required Available (measured) Tension in volts Volt Yes Frequency in Hz 50/60 Hz Yes Fuse protection in ampere 26 Air flow rate (only for vacuum devices) 27 Yes Yes Compliance Yes No Date/signature : c) Documentation Document Available Where (archival site) Operating instructions Yes No Spare parts/order list Yes No Compliance Yes No Date/signature : d) Safety features Parameters Required Available Seal seam width 6 mm 28 Distance to medical device 30 mm 29 Compliance Yes No Date/signature : In general, the operating instructions suffice as evidence of these aspects. In addition, the following aspects must be verified by an authorized person: Description Compliance Remarks Has the sealing device been properly connected? Is the sealing device free of visual safety defects (defective casing, power cables, connector, etc.)? Is the sealing device free of functional defects (unknown running noise, clattering, grating, etc.)? Yes Yes Yes No No No Compliance Yes No Date/signature : 26 Please consult the manufacturer s instructions for the fuse protection required. 27 Please consult the manufacturer s instructions for the air flow rate required. 28 EN «The overall width of the seal(s) shall be not less than 6 mm. For ribbed seals, the sum of the widths of the ribs shall be not less than 6 mm». 29 German standard DIN «Beneath the seal seam at least 30 mm must be left between the sterile item and the seal seam».
18 16 ANNEX A.2 Central Service Suppl e) Critical parameters The following other aspects must be defined or verified by the user (evidence required in some cases): critical during process development? 30 Sealing time Sealing speed Which parameters have been defined as Sealing temperature Contact pressure Issues to be clarified Compliance Evidence based on Are the critical parameters monitored? Yes No Are there systems available which, in the event of deviation from pre-determined limit values for critical process parameters, trigger an alarm or warning or bring the device to a standstill? 31 Yes No Are these critical process parameters routinely controlled and monitored? 32 Yes No Compliance Yes No Date/signature : The following other aspects must be confirmed by providing appropriate evidence: Issues to be clarified Compliance Evidence based on Has the sealing device been serviced and are written servicing plans available? Have the essential sensors (e. g. temperature sensor and DMS module) to the process been calibrated and are written calibration plans available? Yes Yes No No Compliance Yes No Date/signature : In addition, the following must be simulated and documented: Are the parameter settings preserved in the event of power failure? Yes No Compliance Yes No Date/signature : 30 ISO «Critical process parameters shall be defined». Note: For rotary sealers the critical parameters include at least the sealing temperature and contact pressure (monitoring of the sealing speed is recommended additionally). For bar sealers the critical parameters are sealing temperature, contact pressure and sealing time». 31 ISO «Alarms, warning systems or machine stops shall be challenged in the event that critical process parameters exceed predetermined limits». 32 ISO «The critical process parameters shall be controlled and monitored».
19 Central Service Suppl ANNEX A.2 17 f) Induction/Training Name of trained staff member Training Signature By Qualification Date Trainer Trainee Only if all questions have been answered with «Yes», the required sources of evidence provided and users inducted/trained will installation qualification be deemed to have been passed.
20 18 ANNEX A.3 Central Service Suppl Annex A.3: Operational qualification (OQ) checklist «pouch, reel or bag sealing» 33 Criterion Lower limit (LL) Upper limit (UL) 1. Target temperature (as per packaging manufacturer = M 34 ) LLM = ULM = 2. Actual temperature during test (measured/read) LL = UL = 3. Requirement LL LLM UL ULM 4. Compliance with requirement from line 3 Yes No Quality properties Compliance Compliance Intact seal for a specified seal width Yes No Yes No Evidence based on Test method: * No channels or open seals Yes No Yes No Evidence based on Test method: * No punctures or tears Yes No Yes No Evidence based on Test method: * No material delamination or separation Yes No Yes No Evidence based on Test method: * Temperature (T) defined for PQ (mean value from upper and lower limit values of actual temperature at the time of testing) * Test methods are given in Table 3. T = 33 If other sealing methods are used, a customized checklist must be compiled if necessary. 34 If special materials are used (e. g. HDPE), limit values must also be calculated in sample seals if necessary.
21 Central Service Suppl ANNEX A.4 19 Annex A.4: Performance qualification (PQ) checklist «pouch, reel or bag sealing» 35 Temperature defined for the sealing process in the decontamination circuit (carried forward from OQ checklist) Target temperature for operational qualification (carried forward from OQ checklist) T = LL = UL = Switch-off tolerance in degree Celsius as per DIN :2010 (max. ± 5 C) 36 SO = Resultant upper and lower value T SO = T + SO = Requirements T SO LL T + SO UL Compliance with requirements Yes No Yes No Criteria Date/time of sterilization Sterilization protocol (log) available and correct process sequence confirmed Sealing parameters: Sealing temperature Contact pressure Sealing speed/sealing time (dwell) Seal strength test Sterilization cycle (batch) A Sterilization cycle (batch) B Sterilization cycle (batch) C Yes No Yes No Yes No Free end supported Yes No Yes No Yes No Maximum strength Sample A : B : C : Strength value (Smax) Test passed (if all values Smax 1.5 N) Evidence based on (name of laboratory or company) Verification of quality properties: Yes No Yes No Yes No Sample A : B : C : Intact seal for a specified seal width Test method: * No channels or open seals Test method: * No punctures or tears Test method: * No material delamination or separation Test method: * * Test methods are given in Table 3. Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No Yes No 35 If other sealing methods are used, a customized checklist must be compiled if necessary. 36 If special materials are used (e. g. HDPE), narrower switch-off tolerances must be defined if necessary (e. g. ± 3 C instead of ± 5 C).
22 20 ANNEX A.5 Central Service Suppl Annex A.5: Example for determining the scope of process validation per heat sealer Example from everyday practice A Central Sterile Supply Department (CSSD) has two heat sealers, three different steam sterilization programs as well as one formaldehyde sterilizer and one «plasma sterilizer», each with one program. Materials are assigned as follows: Sealer C 5 min STEAM 134 C 18 min 121 C 20 min FORM (formaldehyde) EO (ethylene oxide) VH2O2 (plasma) Material A (see through flat pouch) Material B (see through gusseted pouch) * Material C (Tyvek ) Material D (paper bag) * Sealer C 5 min STEAM 134 C 18 min 121 C 20 min FORM (formaldehyde) EO (ethylene oxide) VH2O2 (plasma) Material A (see through flat pouch) Material B (see through gusseted pouch) Material C (Tyvek ) * Material D (paper bag) The ten combinations outlined in the table can be reduced by taking account of only the maximum material stress (worst-case scenario, while providing documentary proof to justify this; in this example for material A and B: 134 C/18 min as well as see through gusseted pouch). This combination is marked with an * in the table. The seal seam is subjected to the greatest stress during steam sterilization, hence this must be viewed as a «worst case». Here in turn the program with the higher temperature must be first considered and then the longer exposure time with the same temperature. This example shows that in total validation must be carried out three times. A further reduction can be achieved by a deliberate choice of sterile barrier system (e. g. see through flat pouch instead of paper bag). Accordingly, for this example the number of validations needed would be reduced from three to two.
23 Central Service Suppl ANNEX A.6 21 Logo of Institution Quality Management Manual Page of page Department Scope Revision Annex A.6: Sample standard operating procedure «heat sealing» Note: the German standard DIN , 6.3 gives a guide to packing in pouches and reels. That guide has been used as a basis for compiling this sample standard operating procedure (SOP). 1. Selecting pouches or reels Select preformed pouches in accordance with the size of the medical device (MD). If no preformed pouches are available in the correct size, cut reels to an appropriate size and seal at the lower edges such that the reel section can be filled like a pouch. Alternatively, a preformed pouch can also be shortened. Neither the sterile barrier system nor the protective packaging should be kinked or folded. The MD may occupy at most 75 % of the pouch (DIN ). The width chosen must allow for unimpeded introduction of the MD, but it is not advisable to use a bigger size. The space between the upper end of the MD and the seal seam on the peeling side must be at least 3 cm (DIN ). After sealing, an excess of at least 1 cm must be left above the seal seam (recommended in practice: 2 3 cm) to allow for unimpeded peeling as well as aseptic withdrawal (DIN ). When using gusseted pouches or reels the distance to the seal seam should be markedly more than 3 cm to permit orderly sealing of original folds (the folded foil lies evenly on the paper side to prevent formation of any additional folds). 2. Packing the medical device Insert the MD into the see through pouch such that the user can hold the gripping end (on the peel side). For reels, pay attention to the opening direction/peeling direction. A protective must be fitted to any pointed or sharp instruments before they are placed in pouches or reels. MDs with a cavity (e.g. kidney dish) must be arranged such that their opening will face the paper side. 3. Sealing pouches and reels Pull tightly on the open end of pouches or reels so that the foil and paper lie evenly and free of folds in the guide mechanism on the feed-in side of the heat sealer until the device has transported the pouches or reels and a seam has been sealed. If necessary, manually support transport while the seal seam is being produced. Special care has to be taken when sealing gusseted pouches and reels: formation of any additional compression or shrinkage folds, giving rise to channels in the seal seam, must be avoided. Recommendation: if gusseted pouches or reels can be replaced with larger sizes without a gusset this should be done in the interest of risk minimization. 4. Visual inspection of the seal seam Each seal must extend along the total width and length of the seal lines. There must not be any channels, kinks, folds, air pockets or notches. There must not be any signs of burning or melting The test method ASTM F1886 listed in ISO Annex B «Standard test method for determining integrity of seals for medical packaging by visual inspection» can be used for routine visual inspection. Compiled: Date: Reviewed: Date: Approved: Date:
24 22 ANNEX A.6 Central Service Suppl Logo of Institution Quality Management Manual Page of page Department Scope Revision 5. Protective packaging in the form of an outer see through wrap If a second wrap is specified in the packing instructions for the respective instrument, repeat steps 1 to 5, while paying attention additionally to the following: The pouch or reel size must permit unimpeded introduction of the inner wrap. The inner see through foil must not be kinked or folded. Attention must be paid to ensuring that the inner wrap is not sealed into the seal seam of the outer wrap. Make absolutely sure that the paper side of inner pouches and reels face the paper side of the outer pouches and reels. 6. Labelling Labels should as a rule be affixed to the foil side. If the label is to be affixed to the paper side, the size of the label must not exceed 20 % of the paper surface. Do not affix labels to the seal seam. Label only outside the seal seam and outside the area surrounding the sterile MD. To that effect, use ink cassettes that meet the requirements of DIN In exceptional cases a suitable pen may be used to label outside the seal seam and the area enclosing the sterile MD. Here use only pens that meet the requirements of DIN (see Annex D for Sample Data Sheet for Sterilization Markers). 7. Using a further protective packaging after sterilization This can be done, e. g. for transport and storage, protection as well as extension of the storage time, and is documented in the packing lists. Compiled: Date: Reviewed: Date: Approved: Date:
25 Central Service Suppl ANNEX A.7 23 Logo of Institution Quality Management Manual Page of page Department Scope Revision Annex A.7 : Sample standard operating procedure for verification of seal seams (daily when using) Scope This operating procedure is intended for all CSSD personnel who have successfully completed at least Specialist Training Course 1. Aim Daily routine visual inspection of the integrity and peelability of self-produced seal seams. Standard reference: Dye penetration test (ink test): ISO designates the dye penetration test as a test method for verification of the integrity of seal seams (e. g. ASTM F1929: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration). Peel test: EN 868-5, Annex E: «Method for determination of the peel characteristics of paper/plastic laminate products». Materials and prerequisites: 1) Sealing device must be switched on and ready for operation (target temperature reached). 2) Dye penetration test pack (InkTest) 38 : Suitable test ink with defined, very low viscosity Pipette Liquid-impermeable underlay If necessary, small disposable cloth, handkerchief, or similar 3) Reel sections or pouches (approx. 20 cm width) of all see through packaging needed for the dye penetration test. 4) Reel sections of all see through packaging needed for the peel test Complete test packs are commercially available. 39 If only pouch packaging is used, the peel test can be omitted after sterilization. Compiled: Date: Reviewed: Date: Approved: Date:
26 24 ANNEX A.7 Central Service Suppl Logo of Institution Quality Management Manual Page of page Department Scope Revision Procedure 1) Switch on the sealing device and wait until it has reached operating temperate. Dye penetration test (InkTest) 40 : 2) Switch sealing device to test mode (if applicable) 41. 3) Seal an empty pouch or reel section; width at least 20 cm/length approx. 10 cm. 4) Cut the pouch approximately 5 cm above the sealing seam (the reel section is already open at the top). 5) Using a pipette, inject around 2 ml of dye penetrant into the opened pouch or reel section just above the sealing seam. Using a finger or cloth, rub the testing ink along the sealing seam from the outside. 6) After around 20 seconds, check whether the sealing seam is intact. 7) Seal leaks in the sealing seam will be visible from the penetration of test ink. Note: If left for a long time the extremely thin-liquid test ink can penetrate the porous material (paper or Tyvek 42 ) of the pouch or reel. This is not a leak. Peel test: 8) Introduce reel section into sealing device and seal on peel side. 9) Expose sealed reel section to a sterilization cycle. 10) Slowly and carefully peel the seal joints apart by hand. Visually check that the seal extends along the total width and length of the seal lines. There must be no splitting of the paper more than 10 mm from the seal 43. The results must be documented. 40 Seal integrity indicators (e. g. Seal Check) can also be used for routine checks of seal seams. 41 In the test mode (Seal Check mode) the critical sealing parameters as well as the name of test person, test date/time and serial number can be printed on the test packaging. 42 Tyvek is a registered trademark of E.I. du Pont de Nemours. 43 Requirement as per EN 868-5, Annex E Compiled: Date: Reviewed: Date: Approved: Date:
27 Central Service Suppl ANNEX B.1 25 Annex B.1 : Validation plan checklist «sterilization sheets folding and wrapping» Initial validation Revalidation (at regular intervals, only performance requalification) Revalidation for special reasons (e. g. new materials) a) Competences Name of institution (operator) Location Validator (name of persons, or companies, conducting validation) Responsible for overall validation (name/position) b) Description of reusable container Manufacturer Designation Supplier Contact person Manufacturer s CE conformity declaration available?* 44 Yes No Evidence ISO Part 1 conformity?* 45 Yes No Evidence Manufacturer s QM certificate available?* Yes No Evidence Description of packaging material (porous material)** Crepe paper Textile materials Nonwovens SMS nonwovens Other: Manufacturer s specifications and/or data sheet available 46 Yes No Evidence with information on: Surface weight* (rated weight) g/m 2 Compatibility with respective sterilization process* STEAM EO (ethylene oxide) FORM (formaldehyde) VH2O2 (plasma) Other: Label on protective and inner packaging (EN 868-2:2009)* Yes No Evidence 44 The CE mark must be affixed to the outer packaging. The CE mark must not be affixed to the sheets supplied by the manufacturer (preformed sterile barrier system). 45 Conformity with ISO is an absolute prerequisite and in general includes conformity with EN Often, CE conformity and conformity with ISO Part 1 are declared jointly in one document. 46 See Annex F: sample data sheet on crepe sheet materials.
28 26 ANNEX B.1 Central Service Suppl with information on: Reference, raw material or catalogue number* Quantity* Name of manufacturer or supplier or trademark and address* Batch number* Rated dimensions of sheets or rated width of rolls in millimetres as well as length in metres* Date of manufacture as per ISO 28601* 47 Recommended storage conditions* * Information featuring an * must, in accordance with EN 868-2, be made available by the manufacturer of the packaging material. ** For each material a complete checklist must be filled out and the validation process conducted. c) Description of closing system with or without indicator Manufacturer/supplier Contact person Type/designation of closing system Adhesive tape without indicator (process d additionally) Adhesive tape with indicator Other: Manufacturer s/supplier s QM certificate available? Yes No Evidence Have the recommended storage conditions been met? Yes No Evidence Compatibility with packaging material Crepe paper Yes No Nonwovens Yes No Textile material 49 Yes No Compatibility with respective sterilization process Product characteristics of closing system No toxicity STEAM VH2O2 (plasma) EO (ethylene oxide) Other: Information from manufacturer s data sheet Evidence 48 : Evidence Evidence FORM (formaldehyde) Evidence Adhesive tape with indicator* Type/designation of indicator Other with indicator* Evidence * Conformity of indicator used with ISO ? Yes No 47 EN does not call for specification of expiry date. 48 Evidence can be provided on basis of data sheet or documented experience. 49 Here a sterile barrier system is understood to mean only qualified materials as per EN
29 Central Service Suppl ANNEX B.1 27 d) Description of another indicator used (EN ISO ) Manufacturer/supplier Contact person Designation Manufacturer s/supplier s QM certificate available? Yes No DIN EN ISO Part 1 conformity? (e. g. non-toxic) Have the storage conditions as recommended in the data sheet been met? Yes Yes No No Where? Evidence: Evidence: Compatibility with packaging material Crepe paper Yes No Evidence Nonwovens Yes No Evidence Textile material 50 Yes No Evidence Compatibility with respective sterilization process STEAM VH2O2 (plasma) EO (ethylene oxide) Other: FORM (formaldehyde) e) Description of sterilizations process Sterilization process STEAM Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process EO (ethylene oxide) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: 50 Here a sterile barrier system is understood to mean only qualified materials as per EN
30 28 ANNEX B.1 Central Service Suppl Sterilization process VH2O2 (plasma) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process FORM (formaldehyde) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process Other: Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation:
31 Central Service Suppl ANNEX B.1 29 f) Qualification steps If this is an initial validation, all three qualification steps (IQ, OQ and PQ) must be carried out as per the checklists in Annex B.2, B.3 and B.4. For revalidation/performance requalification it may be possible to omit some steps. Installation qualification (IQ) executed already executed during validation on passed failed Date/signature : Operational qualification (OQ) executed already executed during validation on passed failed Date/signature : Performance qualification (PQ) executed passed failed Date/signature : g) Official approval of validation/revalidation by the operator All parts of validation/revalidation passed. Parts of validation/revalidation failed. Measures have been defined and documented. Place, date Name Signature
32 30 ANNEX B.2 Central Service Suppl Annex B.2 : Installation qualification (IQ) checklist «sterilization sheets folding and wrapping» Are standard operating procedures available (SOPs)? (e. g. as in Annex B.6) Yes No Where? a) Training Name of trained staff member Training Signature By Qualification Date Trainer Trainee Only if all users are inducted/trained will installation qualification be deemed to have been passed.
33 Central Service Suppl ANNEX B.3 31 Annexe B.3 : Operational qualification (OQ) checklist «sterilization sheets folding and wrapping» If the packaging system is composed of a sterile barrier system and protective packaging, the quality properties of both the sterile barrier system and protective packaging have to be verified for OQ. Requirement for sample size (S) 51 S 10 Sample size (S) S = Compliance with requirement Yes No Quality properties Compliance Intact closeness/integrity Yes No Evidence based on Test method: No punctures (perforation) or tears Evidence based on Test method: No other visible damage or material irregularities Evidence based on Test method: Protective packaging Sterile barrier system Yes No Yes No Nom/signature Protective packaging Sterile barrier system Yes No Yes No To document the quality properties, it is recommended that at least one photo be taken in addition of each sample. 51 ISO ( 4.2) «The sampling plans used for selection and testing of packaging systems shall be applicable to the process being evaluated. Sampling plans shall be based upon a statistically valid rationale». The value of 10 is based on the experience made in practice. It can be seen as a statistical valid rational in real life.
34 32 ANNEX B.4 Central Service Suppl Annex B.4 : Performance qualification (PQ) checklist «sterilization sheets folding and wrapping» Criteria Date/time of sterilization Sterilization protocol available and correct process sequence confirmed Sterilization cycle (batch) A Sterilization cycle (batch) B Sterilization cycle (batch) C Yes No Yes No Yes No Cycle (batch) A quality properties Compliance Intact closeness/integrity Yes No Evidence based on Test method: No punctures (perforation) or tears Evidence based on Test method: No other visible damage, contamination, material irregularities or residual moisture Evidence based on visual inspection Test method: Compliance with defined packing method (DIN Annex A) Protective packaging Sterile barrier system Yes No Yes No Protective packaging Sterile barrier system Yes No Yes No Protective packaging Sterile barrier system Yes No Yes No Evidence based on photographic documentation
35 Central Service Suppl ANNEX B.4 33 Cycle (batch) B quality properties Compliance Intact closeness/integrity Yes No Evidence based on Test method: No punctures (perforation) or tears Evidence based on Test method: No other visible damage, contamination, material irregularities or residual moisture Evidence based on Test method: Compliance with defined packing method (DIN Annex A) Protective packaging Sterile barrier system Yes No Yes No Protective packaging Sterile barrier system Yes No Yes No Protective packaging Sterile barrier system Yes No Yes No Evidence based on photographic documentation Cycle (batch) C quality properties Compliance Intact closeness/integrity Yes No Evidence based on Test method: No punctures (perforation) or tears Evidence based on Test method: No other visible damage, contamination, material irregularities or residual moisture Evidence based on Test method: Compliance with defined packing method (DIN Annex A) Protective packaging Sterile barrier system Yes No Yes No Protective packaging Sterile barrier system Yes No Yes No Protective packaging Sterile barrier system Yes No Yes No Evidence based on photographic documentation
36 34 ANNEX B.5 Central Service Suppl Annex B.5 : Example for determining the scope of process validation per packaging material in combination with the sterilization process Example from everyday practice A Central Sterile Supply Department (CSSD) has three different steam sterilization programs as well as one formaldehyde sterilizer and one «plasma sterilizer», each with one program. Materials are assigned as follows: Packaging STEAM 134 C/5 min 134 C/18 min 121 C/20 min FORM (formaldehyde) EO (ethylene oxide) VH2O2 (plasma) Material A (crepe paper) * Material B (nonwovens) Material C (SMS nonwovens) Material D (textile materials) * * * * * * The 13 combinations outlined in the table can be reduced by taking account of only the maximum material stress (worstcase scenario, while providing documentary proof to justify this; in this example for material A, B and C: 134 C/18 min). These combinations are marked with an * in the table. This shows that in this example validation needs to be carried out in total seven times. A further reduction can be achieved by a deliberate sterile barrier system (e. g. by using only two different materials). Accordingly, for this example the number of validations would be reduced from seven to five or even four. Note: When using packaging sheets for FORM or EO sterilization one must ensure that the maximum residual content of sterilant permitted is not exceeded. Packaging sheets containing paper (cellulose) absorb a certain amount of moisture with dissolved sterilization gases. When sheet packaging is used for packing purposes, a larger packaging surface is used and this increases the absolute residual content of sterilization gases compared with see through packaging. The most important thing is to measure the residual content of the entire packaging at the time of sterilization process validation.
37 Central Service Suppl ANNEX B.6 35 Logo Quality Assurance Manual Page X / X Department Scope Revision Annex B.6 : Sample standard operating procedure «sterilization sheets folding and wrapping» Note: German Standard DIN , 6.2 and Annex A give a guide to packing with sterilization sheets. That guide has been used as a basis for compiling this sample standard operating procedure. a) Version A, diagonal packaging The item to be sterilized is placed in the centre of the sheet of paper such that its edges are at a right angel with the diagonals of the sheet of paper. The sheet of paper is pulled upwards across the breadth of the item to be sterilized and folded back parallel to the longitudinal edge such that the item to be sterilized is fully covered. A triangle is now formed (point), providing for opening under aseptic (handling that ensures sterility) conditions. Proceed as in Fig. 2, but now working from the right and from the left. Compiled: Date: Approved: Date: Released: Date:
38 36 ANNEX B.6 Central Service Suppl Logo Quality Assurance Manual Page X / X Department Scope Revision Repeat same procedure on opposite side, as in Fig. 3. An open pocket is now formed at the top of the package on a longitudinal side. The last part of the sheet of paper is now pulled over the object to be packed and the point of the paper is inserted into the pocket until it just about sticks out. The paper is then closed with a suitable closing system (e. g. adhesive tape and/or Class A indicator tape). Compiled: Date: Approved: Date: Released: Date:
39 Central Service Suppl ANNEX B.6 37 Logo Quality Assurance Manual Page X / X Department Scope Revision b) Version B, parallel packaging Place sterilization supplies (e. g. instrument tray) on centre of paper. Place front of paper over the instrument tray Fold edge of paper outwards, around as high as the sterilization supplies. Fold back of paper forwards. Fold edge of paper outwards; the paper closes with the front upper edge. Compiled: Date: Approved: Date: Released: Date:
40 38 ANNEX B.6 Central Service Suppl Logo Quality Assurance Manual Page X / X Department Scope Revision Fold paper at the side and place over the sterilization supplies, see Figs. 4 and 5. The paper is then closed with a suitable closing system (e. g. adhesive tape and/or Class A indicator tape). Compiled: Date: Approved: Date: Released: Date:
41 Central Service Suppl ANNEX B.6 39 Annex B.7 : Sample specification and sample data sheet, e. g. for «sterilization sheets» Specifications and/or data sheets are descriptions of product characteristics compiled by the manufacturer or suppliers and they contain additional or more detailed information, in general on the minimum characteristics set out in the standard. Whereas CE conformity is legally declared by the manufacturer or supplier through the use of the CE mark, legally binding compliance with product characteristics must be expressed separately in specifications and/or data sheets. This can be done e.g. by drawing attention to the specification on the invoice or delivery note. Product Specification Crepe paper sterilization sheet «sample brand» Article group 0310_01, 0310_02 Manufacturer/supplier «sample enterprise» Revision: 1 Date: 03/01/2011 The technical values are guide values subjected to typical process fluctuations. They do not constitute grounds for dispensing with validation and operational qualification in any individual case. Any measurement tolerances and packaging/labelling specifications (on agreement) deviating from the above shall be confirmed in the article text or in the print area indication/print drawing. Product description Sizes Packaging material for medical devices Intended purpose Depending on client s needs Sterilization suitability Steam, EO/FO gas and GAMMA sterilization The packaging material complies with ISO Part Standards 1 and EN 868 Part 2 Sections and When stored as directed, products can be used for five Wear resistance years from date of manufacture (see recommended storage conditions) mm to mm (tolerance ± 5 mm) Labelling cardboard boxes Affix a label to the upperside of cardboard box. The label must contain the following information: Supplier batch code Material designation Article No. Size Package contents (number of items) Date (date of manufacture) Expiry date Packaging Sheet materials are sealed in foil and packed in cardboard boxes Technical data Quality features Value Unit Surface weight: 60 ± 5 % g/m 2 Colours: 0310_01 white 0310_02 green Recommended storage conditions Temperature: + 15 C to + 25 C, Relative ambient humidity: 35 % 50 % RH, store in dry place Protect against light or direct sun radiation. Open outer packaging only when product is to be used. Do not store close to: Chemicals Detergents
42 40 ANNEX B.7 Central Service Suppl Technical Data Sheet Crepe paper sterilization sheet «sample brand» Manufacturer/supplier «sample enterprise» This sterile barrier system complies with the following standards and directives: ISO :2009 EN 868-2:2009 European Medical Devices Directive 93/42/EEC Technical Data Sheets in compliance with EN 868-2: Chapter Aspect Test method Unit Requirement General Typical values Raw materials Primary raw material Compliance Colour fastness ISO No leaching of colour from hot-water extract Mass/surface weight ISO 536 g/m 2 Mass must be within ± 5 % of rated value Compliance ph value ISO ph Chloride content ISO 9197 % Sulphate content ISO 9198 % Fluorescence NF Q % Crepe paper Mass portion of chlorides NaCl 0.05 % Mass portion of sodium sulphate Na 2 SO % Brightness 1 %, 5 spots of 1 mm 2 per 0.01 m 2 60 g/m 2 ± 2 g 0.03 % % Compliance Creping Creping for increased flexibility Compliance Fracture elongation ISO % Water resistance EN Annex A 10 % in machined direction (MD) 2 % in traverse direction (TD) 13 % 5 % s Penetration time 20 s Pore diameter Annex B µm Maximum pore diameter 50 µm 20 µm in Ø Stretching Annex C mm Tensile strength ISO kn/m Wet strength ISO 3781 kn/m Microbial impermeability as per ISO 11607:2009 Part 1: Max. stretching in MD 125 m In TD 160 mm MD 1.33 kn/m TD 0.67 kn/m MD 0.33 kn/m TD 0.27 kn/m 85 mm 148 mm Sample enterprise Place, date* Name, Position* Microbial impermeability in dry state Microbial impermeability when moist DIN : 2010, 2.14 DIN : 2010, 2.15 No colonies on agar plates No Max 20 % cycles No * Specifications and/or data sheets can, but need not, be different.
43 Products and systems from a single source......the specialist in sterile goods logistics! Handling sterile goods effi ciently and successfully in hospitals, clinics and sterilisation operations is not only a challenge, it is also a complex management function. HUPFER - the specialist in sterile goods logistics - has the products and expertise not only to rise to the challenges of daily operational processes, but also to optimise established procedures. The individual logistics functions of these operational processes such as sorting, packing, arranging, transportation, storage and distribution are the basis and starting point for a smooth sterile goods cycle. HUPFER develops and manufactures products and systems tailored to each individual logistics function. This product range enables the sterile goods cycle to be confi gured so as to create a complete process chain. The comprehensive range of items in each HUPFER product line ensures specifi c support in making daily tasks easier, more effi cient and more economical. HUPFER Metallwerke GmbH & Co. KG Dieselstrasse 20 D Coesfeld Postfach 1463 D Coesfeld Tel.: +49 (0) 25 41/805-0 Fax: +49 (0) 25 41/ [email protected]
44 42 ANNEX C.1 Central Service Suppl Annex C.1: Validation plan checklist for packaging process with «filling and closing of reusable sterilization containers» Initial validation Revalidation (at regular intervals, only performance qualification) Revalidation for special reasons (e. g. new materials) a) Competences Name of institution (operator) Location Validator (name of persons, or companies, conducting validation) Responsible for overall validation (name/position) b) Description of reusable sterilization container Manufacturer Designation Is the manufacturer s name visible on the product? (ISO ) Supplier Has the supplier been authorized by the manufacturer? Yes No Evidence Supplier s contact person Name: Telephone number: Manufacturer s CE conformity declaration available? 52 Yes No Evidence ISO Part 1 conformity?* Yes No Evidence EN 868-8** conformity? Yes No Evidence Manufacturer s QM certificate available?*** Yes No Evidence Is the operating manual available? Yes No Evidence Is information available on cleaning and disinfection processes as per ISO 17664? 53 Manual Yes No Automated Yes No 52 Based on the Medical Devices Directive a sterilization container is a Class 1 medical device (medical device accessory). 53 Preference must be given to automated cleaning and disinfection processes.
45 Central Service Suppl ANNEX C.1 43 Compatibility with existing sterilization process (according to operating manual) STEAM EN 285 STEAM EN EO (ethylene oxide) Other: FORM (formaldehyde) Sterile supplies inner wrap as per DIN ? Yes No Are other consumables needed? If yes, from the same manufacturer as the sterilization container? Yes No (c and d omitted) Filter Seals Other (compile another table based on c and d) Yes (c omitted) Yes (d omitted) Yes No No No * Information marked with * must, in accordance with ISO , be supplied by the manufacturer of the packaging material. ** Information marked with ** is normally available when there is compliance with the provisions of the CE conformity declaration and with the provisions of ISO *** Information marked with *** is normally available when there is compliance with the provisions of the CE conformity declaration.
46 44 ANNEX C.1 Central Service Suppl c) Description of microbial barrier Type of microbial barrier Single-use filters Reusable filters Number of decontamination cycles 54 :... Closed valve Pasteur loop Manufacturer Designation Is the manufacturer s name visible on the product/outer packaging? Supplier Contact person Manufacturer s CE mark and conformity declaration available? 55 Yes No Evidence ISO Part 1 conformity?* Yes No Evidence EN 868-2** conformity? Yes No Evidence Manufacturer s QM certificate available?*** Yes No Evidence Compatibility with respective sterilization process STEAM VH2O2 (plasma) EO (ethylene oxide) Other: FORM (formaldehyde) Compatibility with reusable sterilization container named in b) Yes No Evidence Reprocessable? Yes No d) Description of seals 56 Manufacturer Designation Is the manufacturer s name visible on the product/outer packaging? (ISO :2009,...)? Supplier Contact person Manufacturer s QM certificate available? Yes No Evidence Compatibility with respective sterilization process STEAM EO (ethylene oxide) FORM (formaldehyde) VH2O2 (plasma) Other: Compatibility with reusable sterilization container named in b) Yes No Evidence * Information marked with * must, in accordance with ISO , be supplied by the manufacturer of the packaging material. ** Information marked with ** is normally available when there is compliance with the provisions of the CE conformity declaration and with the provisions of ISO *** Information marked with *** is normally available when there is compliance with the provisions of the CE conformity declaration. 54 The number of cycles must be documented. 55 The CE mark must be affixed to the sterilization container. 56 CE mark not required
47 Central Service Suppl ANNEX C.1 45 e) Description of the sterilization process Describe only the sterilization processes with which the sterile barrier system described under b) is sterilized. Sterilization process STEAM Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process EO (ethylene oxide) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process VH2O2 (plasma) Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation: Sterilization process Other: Sterilization process validated? Yes No Validated by: Last validation: Validation report number(s): (if there is more than one sterilizer) Next validation:
48 46 ANNEX C.1 Central Service Suppl f) Qualification steps If this is an initial validation, all three qualification steps (IQ, OQ and PQ) must be carried out as per the checklists in Annex C.2, C.3 and C.4. For revalidation / performance requalification it may be possible to omit some steps. Installation qualification (IQ) executed already executed during validation on passed failed Date/signature : Operational qualification (OQ) executed already executed during validation on passed failed Date/signature : Performance qualification (PQ) executed passed failed Date/signature : g) Official approval of validation/revalidation by the operator All parts of validation/revalidation passed Parts of validation/revalidation failed Measures have been defined and documented Place, date Name Signature
49 klsmartin.com 100 % GreenNo filter. paper No waste. No costs. The MicroStop sterilization container. Sustainability that really pays off.
50 48 ANNEX C.2 Central Service Suppl Annex C.2: Installation qualification (IQ) checklist «filling and closing of reusable sterilization containers» Are standard operating procedures available (SOPs)? (example as in Annex C.6) Yes No Where? Document Available Where (archival site) Operating manual Yes No CE conformity declaration 57 Yes No Consumables order list Yes No Compliance Yes No Date/signature : Individual labelling of sterilization container (including lid) Year of manufacture Is the sterilization container defective? (if No, approve and sign) Take remedial action if defective Is the sterilization container defective after taking remedial action? (if No, approve and sign) Approval/ signature Yes No Yes Which: Yes No Signature: Yes No Yes Which: Signature: Yes No 57 The CE conformity declaration is normally part of the operating manual.
51 Central Service Suppl ANNEX C.2 49 Induction/Training Name of trained staff member Training Signature By Qualification Date Trainer Trainee Only if all users have been inducted/trained will installation qualification be deemed to have been passed.
52 50 ANNEX C.3 Central Service Suppl Annex C.3: Operational qualification (OQ) checklist «filling and closing of reusable sterilization containers» If the sterilization container has an inner wrap, the quality properties of both the sterilization container and inner wrap have to be verified for OQ. Requirement for sample size (S) 58 S 10 Sample size (S) S = Compliance with requirement Yes No Quality properties Intact closeness/integrity Compliance Yes No Evidence based on Test method: No visible damage or material irregularities Sterilization container Inner wrap Yes No Yes No Evidence based on Test method: To document the quality properties, it is recommended that at least one photo be taken in addition of each sample. 58 ISO ( 4.2) «The sampling plans used for selection and testing of packaging systems shall be applicable to the process being evaluated. Sampling plans shall be based upon a statistically valid rationale». The value of 10 is based on the experience made in practice. It can be seen as a statistical valid rational in real life.
53 Central Service Suppl ANNEX C.4 51 Annex C.4: Performance qualification (PQ) checklist «filling and closing of reusable sterilization containers» Criteria Date/time of sterilisation Sterilization protocol available and correct process sequence confirmed Sterilization cycle (batch) A Sterilization cycle (batch) B Sterilization cycle (batch) C Yes No Yes No Yes No Cycle (batch) A quality properties Intact closeness/integrity Compliance Yes No Evidence based on Test method: No visible damage, contamination, material irregularities or residual moisture Evidence based on Test method: Compliance with defined packing method Evidence based on photographic documentation Sterilization container Inner wrap Yes No Yes No Sterilization container Inner wrap Yes No Yes No Cycle (batch) B quality properties Intact closeness/integrity Compliance Yes No Evidence based on Test method: No visible damage, contamination, material irregularities or residual moisture Evidence based on Test method: Compliance with defined packing method Evidence based on photographic documentation Sterilization container Inner wrap Yes No Yes No Sterilization container Inner wrap Yes No Yes No
54 52 ANNEX C.4 Central Service Suppl Cycle (batch) C quality properties Intact closeness/integrity Compliance Yes No Evidence based on Test method: No visible damage, contamination, material irregularities or residual moisture Evidence based on Test method: Compliance with defined packing method Evidence based on photographic documentation Sterilization container Inner wrap Yes No Yes No Sterilization container Inner wrap Yes No Yes No
55 Central Service Suppl ANNEX C.5 53 Annex C.5: Example for determining scope of process validation per sterilization container in combination with the sterilization processes Packaging STEAM 134 C/5 min 134 C/18 min 121 C/20 min EO (ethylene oxide) VH2O2 (plasma) 1) Sterilization container from manufacturer A (with permanent filter, without inner wrap * 2) Sterilization container from manufacturer B/filters from manufacturer B with inner wrap * 3) Sterilization container from manufacturer B/filters from manufacturer C with inner wrap 4) Sterilization container from manufacturer B/filters from manufacturer C without inner wrap * * * The 14 combinations outlined in the table can be reduced by taking account of only the maximum material stress (worstcase scenario, while providing documentary proof to justify this). These combinations are marked with an * in the table. This example shows that in total validation must be carried out five times. A further reduction can be achieved by opting for standardization (e. g. using only one filter or sterilization container or sterilization container/filter system).
56 54 ANNEX C.6 Central Service Suppl Logo Quality Assurance Manual Page X / X Department Scope Revision Annex C.6: Sample standard operating procedure «filling and closing of reusable sterilization containers» Note: German Standard DIN , 6 gives a guide to packing reusable sterilization containers. That guide has been used as a basis for compiling this sample standard operating procedure. 1. Aim: After this procedural step, the packaging system must be available and ready for the sterilization process step. 2. Scope of application: Clean side of CSSD. 3. Preparation: 3.1 Trays must be first packed as a precondition for packing in sterilization containers. 3.2 Compliance with the maximum loading heights specified in the manufacturer s instructions must be assured. 3.3 For ergonomic reasons and to avoid excessive condensation, the weight of the load should not exceed 10 kg (as per EN 868, Part 8) Workflow pattern: 4.1 Perform a functional test in accordance with the instructions of the manufacturer of the respective sterilization container. 4.2 If necessary, fit a microbial barrier in the packaging system at the sites specified in the manufacturer s instructions. 4.3 Insert the prepared trays with or without an inner wrap. 4.4 The sterilization container lid must be fitted to the container tank without exerting any pressure and closed in accordance with the instructions of the manufacturer of the respective closing system. 4.5 If necessary, fit a sealing system to the prescribed sites to protect against unauthorized opening, e.g. in the form of a seal. 4.6 The sterilization container must feature at least the following information: Name of packer, Proprietors and content, Documentation of sterilization date. 4.7 Load trolley for the sterilizer as per the manufacturer s instructions. 4.8 Last visual inspection before closing door. 4.9 Before activating start button, verify whether prescribed program has been selected. 5. Accompanying documents: Operating instructions Sterilization container Sterilizer Preliminary and subsequent CSSD operating procedures Validation documentation Compiled: Date: Approved: Date: Released: Date:
57 Central Service Suppl ANNEX D 55 Annex D: Sample data sheet «sterilization markers» Technical Data Sheet Sterilization Markers Manufacturer/supplier «sample enterprise» Description: Sterilization markers Features: n-propanol/ethanol, does not contain xylol or toluol. Waterproof on most surfaces. Odourless. Colours: Organic colours. Ingredients based on latest technical information sources. Sheath: Polypropylene PP Test standard: ISO 554 Inspection: Based on Batch No. Test : As per prescribed procedure Shelf life: 2 years after date of manufacture
58 56 Notes Central Service Suppl. 2012
59 Offizielles Mitteilungsorgan Deutsche Gesellschaft für Sterilgutversorgung e.v. Central Service Suppl MASTHEAD Editor in Chief P. Heeg, Tübingen, Germany Editors P. Heeg, Tübingen (D), W. Renders, Brugge (B), W. Spencer, Selby (GB), D. Talon, Paris (F) Advisory Editorial Board W. Accoe, Gent (B), M. Alfa, Winnipeg (CDN), M. Borneff-Lipp, Halle (D), A. Carter, Rödinghausen (D), F. Cavin, Lausanne (CH), E. Dennhöfer, Köln (D), C. Faber, Brüssel (B), T. Fengler, Berlin (D), D. Goullet, Lyon (F), P. Holland, Kingston-upon-Thames (GB), U. Junghannß, Köthen (D), W. Kohnen, Mainz (D), S. Krüger, Grünendeich (D), H. Martiny, Berlin (D), A.P. Mercieca, Sydney (AUS), T. Miorini, Graz (A), K. Roth, Tübingen (D), A. Schneider, Pforzheim (D), G.A. Sills, Nottingham (GB), Y. Uetera, Tokyo (Japan), P.A. de Vries, Utrecht (NL), T. Zanette, Tübingen (D) Central Service is the official publication of the German Society for Sterile Supply e.v. (DGSV). Publisher mhp-verlag GmbH, CEO Andreas Klein, Marktplatz 13, D Wiesbaden. Tel.: +49 (0) 611/ , Fax: -11, Editorial Staff Dr. Gudrun Westermann, Tel.: +49 (0) 611/ , Fax: -11, Advertising Sales Walter Bockemühl Tel.: +49 (0) 611/ Rate Card No. 20/2012 Printing Druckerei Chmielorz GmbH, Ostring 13, D Wiesbaden-Nordenstadt Paper Terraprint mat «chlorine-free», 80 g/sq.m Publication Schedule Bimonthly: February, April, June, August, October, December Subscription Rates Single issue: (+ postage). Annual subscription (incl. postage): Print: Germany: 66.60; Europe: 77.00; World (air mail, + tax): E-Paper: Germany: 65.00; Europe: 65.00; World (w/o tax): Kombi Print/E-Paper: Germany: 82.85; Europe: 93,25; World (air mail included, + tax): 103,16. Unless cancelled 6 weeks before the end of the current year, the subscription is automatically continued for another year. Subscription orders should be sent to the publisher. For more information see Copyright All rights reserved. No part of this publication may be reproduced, transmitted in any form or by any means without the prior written permission of mhp-verlag GmbH. One photocopy of individual articles or parts thereof may be made for personal use only. mhp-verlag GmbH 2012 Trademarks The use of general descriptive names, trade names, trademarks, etc., even if these are not especially identified, is not to be taken as sign that such names may be used freely by anyone. Editor in Chief, Editorial Board and publisher do not guarantee for articles published outside the editorial part of the journal (esp. advertising, industrial information). Authors who are mentioned with their full name and who are not member of the editorial staff publish articles in their own responsibility. ISSN D 2596 F 20. Jahrgang l YES, I would like to subscribe to Central Service. l I will pay after receipt of your invoice. Standardization: Simplified Validation Normen: Vereinfachte Validierung Cleaning, Disinfection, Sterilization: Report on the DGKH Congress in Berlin Reinigung, Desinfektion, Sterilisation: Bericht vom DGKH- Kongress in Berlin 4 th Heidelberg Sterilisation Symposium 4. Heidelberger Sterisymposium Sterilization requirements for materials used for patients undergoing transplantation Sterilisationsanforderungen an Materialien zur Verwendung bei Transplantationspatienten l I wish to pay by Credit Card: l Visa l Master Card l AmEx No. l l l l l l l l l l l l l l l l Safety no. l l l l Surface changes on surgical instruments Oberflächenveränderungen auf chirurgischen Instrumenten Quality Task Group: Reprocessing Transmission Instruments in Dentistry AK Qualität: Aufbereitung von Übertragungsinstrumenten im zahnmedizinischen Bereich Photo: Thinkstock FAX +49 (0) 611/ You can cancel your order by writing to the publisher within 14 days from the date of submission. Sender Name Establishment Address Date, Signature Suppl. LLV_en/12
60 A-ST12007»AIOS The Smile System. Aesculap Instrument Organisation System» Organise and protect your surgical instruments Our new AIOS organisation system for the supply of sterile products makes clever organisation and efficient processes easy. The multi-facetted system is specially designed for Aesculap instrument trays and is easy to assemble. The racks are fitted in seconds with pins or clamps. Done! Why not try the AIOS starter kit for an easy, practical introduction. For more information, please visit: Aesculap AG Am Aesculap-Platz Tuttlingen Germany Phone +49 (0) Fax +49 (0) Aesculap a B. Braun company
Packaging Validation according to ISO 11607
Synergy makes sence Packaging Validation according to ISO 11607 MG-FSI72-104 Last revision: March 2011 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel. 33 (0)4 72 81 22 62 - Fax : 33 (0)4
STERILIZATION WRAP. STERISHEET user guide for the validation of the packaging THE BACTERIAL BARRIER FOR HUMAN PROTECTION
STERILIZATION WRAP STERISHEET user guide for the validation of the packaging ARJOWIGGINS in few words A large range of sterilization wrap from standard creped paper to non-woven A large range of standard
Process Validation for Medical Devices
Process Validation for Medical Devices Dan O Leary CBA, CQE, CRE, CSSBB, CIRM, LLC 603-209-0600 [email protected] Copyright 2010 by, LLC Process Validation for Medical Devices 1 Instructor Introduction
Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables
(1987) Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables 1. General 1.1 Introduction 1.1.1 The present guidelines are to serve as a supplement to the IACS
THE BASICS OF PACKAGING
THE BASICS OF PACKAGING Objectives: Review the regulatory requirements for packaging Present the purpose, function and essential characteristics of packaging Describe packaging options, their use, and
New Applications for Dye Penetration Testing of SBS Packages and Films
New Applications for Dye Penetration Testing of SBS Packages and Films Geoff Pavey, Principal Engineer Oliver Medical March 3-5, 2009 www.healthpack.net What will be covered Intro to Integrity failures/need
Naue GmbH&Co.KG. Quality Control and. Quality Assurance. Manual. For Geomembranes
Naue GmbH&Co.KG Quality Control and Quality Assurance Manual For Geomembranes July 2004 V.O TABLE OF CONTENTS 1. Introduction 2. Quality Assurance and Control 2.1 General 2.2 Quality management acc. to
VIEW-PACK STERILISATION POUCHES
VIEW-PACK STERILISATION POUCHES E DESCRIPTION ITEM NUMBER SALES UNIT MINIMUM SALES UNIT View-Pack sterilisation pouches, 42 x 60 cm 500 pcs per box 1 box NL OMSCHRIJVING ARTIKELNUMMER VERKOOPSEENHEID MINIMUM
T 300 Tray sealer. Automatic tray packaging in the smallest of spaces
T 300 Tray sealer Automatic tray packaging in the smallest of spaces MULTIVAC T 300 T 300 Tray sealer: Automatic tray packaging in the smallest of spaces The especially compact T 300 tray sealer packs
1.1.2 Polypropylene The polypropylene must be a white opaque film, 1 2 mil thick. Biaxial orientation is preferred.
Library of Congress Preservation Directorate Specification Number 700 704 09 Specifications for Pressure Sensitive Adhesive Labels For Application to Single Paper Sheets and Text Pages of Bound Books For
Sterile Medical Device Impulse Sealer Validation: Trials and Tribulations
Sterile Medical Device Impulse Sealer Validation: Trials and Tribulations Jan Gates PackWise Consulting March 14, 2013 General Outline Validation review Sealing review Sealing elements/information required
Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS
Art.S001-S002 SOLAR MODULE WITH DELIVERY AND RETURN CONNECTIONS Technical Information sheet 0001/07/Rev01 ENG FUNCTION Series S001 and S002 circulation units are applied to the primary circuit of solar
Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice
EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 Working Party on Control
MB-Accumulator Operator's Manual
Mathias Bäuerle GmbH Gewerbehallestraße 7-11 D-78112 St.Georgen Telefon (07724) 882-0 Telefax (07724) 882-111 Internet: http://www.mbfold.de E-mail: [email protected] MB-Accumulator Operator's Manual CONTENTS
Guidance on safe use of Autoclaves
Safety Office Reviewed: Dec 2013 Reviewed: DD Month Year Guidance on safe use of Autoclaves 1. Hazards & Uses There are several different types of hazard associated with the use of autoclaves. The main
VdS 2100-09en. VdS Guidelines for water extinguishing systems. Non-return valves. Requirements and test methods. VdS 2100-09en : 2011-05 (01)
VdS Guidelines for water extinguishing systems VdS 2100-09en Requirements and test methods VdS 2100-09en : 2011-05 (01) Publishing house: VdS Schadenverhütung GmbH Amsterdamer Str. 172-174 50735 Köln,
T A B L E T 1 T E S T S A N D I N S P E C T I O N C A B L E P C U T A N D P C U T - A
T A B L E T 1 1 of 7 Tests & Inspection Cable PCUT & PCUT-A (Table T1) T E S T S A N D I N S P E C T I O N C A B L E P C U T A N D P C U T - A No. Test Scale MOC Requirements G 20:10:001:01 Defined Test
Case Study in Test Method Validation - ASTM F1886 (2004), Visual Examination
Case Study in Test Method Validation - ASTM F1886 (2004), Visual Examination Todd Dehmer CPP Senior Packaging Engineer St. Jude Medical, AFD Division March 3-5, 2009 www.healthpack.net 1 Presentation Overview
Standard laryngoscope blade assemblies Directions for use
1 2014 Welch Allyn, Inc. MM 721358 Ver. C Standard laryngoscope blade assemblies Directions for use Intended use About this document The rigid laryngoscope is a device used to examine and visualize a patient
Installation and Operational Qualification Protocol (Reference: SOP )
Project Name Equipment Process Line/Location Project Number Serial Number Model Number Protocol number WRITTEN BY: REVIEWED BY: Position APPROVAL TO EXECUTE: Position: PROTOCOL COMPLETION APPROVAL: Position:
GSI. Geosynthetic Institute GRI. 475 Kedron Avenue Folsom, PA 19033-1208 USA TEL (610) 522-8440 FAX (610) 522-8441 GII GAI GCI. GRI Test Method GM21
Geosynthetic Institute 4 Kedron Avenue Folsom, PA 133-1208 USA TEL (610) 522-8440 FAX (610) 522-8441 GEI GRI GSI GAI GCI GII Revision 4: December 17, 2012 Revision schedule on pg. 11 GRI Test Method GM21
Equipment Information
3002A Equipment Information INDEX: DESCRIPTION AND KEY FEATURES LAYOUT NOMINAL TECHNICAL DATA CONFIGURATION Sealed Air GmbH Customer Systems Center D4 Park 4 CH-6039 Root-Längenbold Telephone: +41 (0)
MELA quick 12+ Compact. Fast. Strong.
MELA quick 12+ Compact. Fast. Strong. The requirement. Requirements for the preparation of dental instruments are defined by official hygiene experts. Instruments for semicritical use (restorative treatment)
Sterilization of narrow channels: how to ensure effectiveness in daily routine? Dr.ir. Josephus Paulus Clemens Maria Van Doornmalen Gomez Hoyos
April 2015 Sterilization of narrow channels: how to ensure effectiveness in daily routine? Dr.ir. Josephus Paulus Clemens Maria Van Doornmalen Gomez Hoyos 3M Deutschand GmbH, IPD, Decontamination University
Cliniklav 25. For highest quality in surgical applications MELAG. Evidence Based Sterilization
Cliniklav 25 For highest quality in surgical applications MELAG Evidence Based Sterilization Cliniklav 25 The big one from MELAG a powerful piece of technology Built-in printer for documenting program
Safety. matters. CryoMACS Products
Safety matters CryoMACS Products CryoMACS Freezing Bags The right step towards safe more cryopreservation safety in cryopreserv Overwrap bag protects both sample and freezing bag during cryopreservation.
NORSOK STANDARD M-650 Edition 4. Qualification of manufacturers of special materials
Date: 2014-05-30 NORSOK STANDARD M-650 Edition 4 Qualification of manufacturers of special materials A: Frequently Asked Questions FAQ B: Clarification and Comment List NORSOK standard M650 Ed. 4 INTRODUCTION
Material test chamber with mechanical convection
FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and
TECHNICAL DATA SHEET GRILON BG-15 S
TECHNICAL DATA SHEET GRILON BG-1 S Grilon BG-1 S is a heat stabilised PA6 injection moulding grade with 1% glass fibres. Grilon BG-1 S has the following important properties: Excellent surface finish Easy
peel adhesion 2 breaking load 4 rolling ball 6 shear adhesion 7 carton test 9 summary test methods 11
CONTENT peel adhesion 2 breaking load 4 rolling ball 6 shear adhesion 7 carton test 9 summary test methods 11 PEEL ADHESION Testing the peel adhesion on stainless steel 1. Definition Peel adhesion - The
Peel & Seal Self-Stick Roofing Solutions
Peel & Seal Self-Stick Roofing Solutions Ideal for Low Slope Applications Waterproof and Weatherproof Energy Efficient Tough and Durable Versatile, Effective Sealing Around Protrusions Perfect for Vents,
Butt Fusion Welding of HDPE Pipes Work Procedure
Butt Fusion Welding of HDPE Pipes Work Procedure A pipeline is as good as its weakest point. Accordingly it is not only important that quality pipes and components are used in a piping system, but also
11.I In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07
In-process control In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07 Here you will find answers to the following questions: What are the in-process control tasks? Where
OVERVIEW OF STANDARD 7E & STANDARD 20
Standard for Thermal Transport Packaging Used in Parcel Delivery System Shipment 7E 2 010 ISTA 7E Usage Modes VERSION DATE OCTOBER 2010 Initial Release ISTA, Distributing Confidence, Worldwide ISTA 7-Series
THERMAL TRANSFER PRINTING PROBLEM SOLVER
THERMAL TRANSFER PRINTING PROBLEM SOLVER Top quality thermal transfer printing is always a result of compatibility between material and hardware. Print quality problems can be caused by a multitude of
Testing Automated Manufacturing Processes
Testing Automated Manufacturing Processes (PLC based architecture) 1 ❶ Introduction. ❷ Regulations. ❸ CSV Automated Manufacturing Systems. ❹ PLCs Validation Methodology / Approach. ❺ Testing. ❻ Controls
Body Art Facility Infection Prevention And Control Plan Guideline
Body Art Facility Infection Prevention And Control Plan Guideline In accordance with the California Health and Safety Code, Section 119313, a body art facility shall maintain and follow a written Infection
UGVCL/SP/591/11KV HT AB CABLE
TECHNICAL SPECIFICATION FOR 11KV AERIAL BUNCHED CABLES FOR OVERHEAD LINES (CROSSED LINKED POLYTHENE DRY GAS CURED) 1. SCOPE This specification covers requirements of XLPE insulated, 11 KV Aerial Bunched
PQS Independent type-testing protocol
PQS Independent type-testing protocol WHO/PQS/E10/SB01-VP.1 Original: English Distribution: General TITLE: Safety box for the disposal of used sharps Product verification protocol: E10/SB01-VP.1 Applies
3.1.1 Full Type Tests & Routine Tests according to Clause 8 2 & 8 3. 4.0 Instructions For Installation, Operation & Maintenance
SPECIFICATION FOR LOW VOLTAGE SWITCHBOARD SEN I N D E X Description 10 STANDARD TECHNICAL REQUIREMENTS 11 Standards 12 General Operating Conditions 13 General Description Of Switchboard 131 Structure 132
Cavity Drain R20 is manufactured from1.0mm thick black high density polyethylene with studs approximately 20mm high.
CI/SfB Tn6 L6813 September 2013 RIW CAVITY DRAIN R20 R20 is manufactured from1.0mm thick black high density polyethylene with studs approximately 20mm high. Other systems are available for specific uses,
Brake module AX5021. Documentation. Please read this document carefully before installing and commissioning the brake module!
Documentation Brake module AX5021 Please read this document carefully before installing and commissioning the brake module! Version : 1.2 : 2012.03.05 Date Article-no. : TDmlAX-5021-0000-0200 Page 2/8
Monitoring the autoclaving process in the pharmaceutical industry
Application Description AD/RandC/006-EN Monitoring the autoclaving process in the pharmaceutical industry - Provides independent verification and validation monitoring of the autoclaving process - Enables
Sterile Cleanroom Management
Sterile Cleanroom Management Lynn Stanard, Sr. Quality Manager, Berkshire Corporation When manufacturing in an aseptic environment, it is critical to ensure that the various cleanroom consumables, such
4 Installation Requirements
4 Installation Requirements 9 4.1 Code Reference The authority having jurisdiction should be referenced to determine what law, ordinance or code shall apply in the use of flexible duct. Ducts conforming
German engineering under high pressure
German engineering under high pressure Over 25 years of experience as a manufacturer of sterilizers, autoclaves and lyophilization systems for laboratories, production and research Production-scale autoclaves
FMC Technologies Measurement Solutions Inc.
Table of Contents 1 Revision History... 4 2 Company... 5 2.1 Introduction... 5 2.2 Quality Policy... 6 2.3 Quality Organization... 6 2.4 Company Processes... 7 3 Scope and Exclusions... 8 3.1 Scope...
Understanding High Perm vs. Low Perm
Understanding High Perm vs. Low Perm Building Science Bulletin Figure A High perm(eability) vs. low perm(eability) within the housewrap category what does it mean and how does it affect your building practices?
Patient Information Guide Morpheus CT Peripherally Inserted Central Catheter
Patient Information Guide Morpheus CT Peripherally Inserted Central Catheter IC 192 Rev C A measure of flexibility and strength. Table of Contents 1. Introduction 2. What is the Morpheus CT PICC? 3. What
DuPont Kapton. polyimide film. General Specifications
DuPont Kapton polyimide film General Specifications Introduction DuPont manufactures and sells a variety of highquality polyimide film products in conformance with ISO 9002 certification. These specifications
Incubator with mechanical convection
BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,
Aseptic preparations, including TPN, for a limited number of patients
Aseptic preparations, including TPN, for a limited number of patients Group F (TPN) 1 Objective Presentation of our business case: setting up an aseptic TPN production in the Hilton Pharmacy instead of
OPERATIONAL STANDARD
1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.
Active Monitor Box McCrypt S.T.E.V.E. 15. Order No. 30 17 06
Active Monitor Box McCrypt S.T.E.V.E. 15 Order No. 30 17 06 1 McCrypt Active Monitor 15 Introduction Dear Customer, Thank you for purchasing this McCrypt Active Monitor. You have chosen a quality product
Quick Start Guide See Inside for Use and Safety Information
3 rd Generation Personal 3D Printer Quick Start Guide See Inside for Use and Safety Information The USB Mass Storage Device Contains the User Guide and Quick Start Guide in other Languages Congratulations
FAT Factory Acceptance Testing SAT Site Acceptance Testing
FAT Factory Acceptance Testing SAT Site Acceptance Testing Dr. Ingrid Walther May 2015, Page 1 Factory Acceptance Testing - FAT Site Acceptance Testing - SAT FAT and SAT - Important Milestones in the Project
GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS
PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 032-2 8 January 2010 RECOMMENDATION GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE
738-B-297 POLYMERIC CONCRETE BRIDGE DECK OVERLAY. (Adopted 02-20-14)
POLYMERIC CONCRETE BRIDGE DECK OVERLAY (Adopted 02-20-14) Description The polymeric concrete bridge deck overlay shall consist of an epoxy polymer that acts together with special aggregate to form an overlay
MHRA GMP Data Integrity Definitions and Guidance for Industry March 2015
MHRA GMP Data Integrity Definitions and Guidance for Industry Introduction: Data integrity is fundamental in a pharmaceutical quality system which ensures that medicines are of the required quality. This
Test Report No. C829QPEN
Institut für Solartechnik Hochschule für Technik Rapperswil Oberseestrasse 10, CH 8640 Rapperswil Tel +41 55 222 48 21, Fax +41 55 222 48 44 www.solarenergy.ch Ordered by: Westech Components Wuxi Co.,
By Vince Paulauskas, Technical Service Manager, Winpak December 2002
QUALITY ASSURANCE OF CUP PACKAGING FOR FOOD AND BEVERAGE APPLICATIONS By Vince Paulauskas, Technical Service Manager, Winpak December 2002 There are numerous tests and checks that a processor/packager
500/600/1000. User Manual June 2011 Rev. 061311. 1 v061311
500/600/1000 User Manual June 2011 Rev. 061311 1 v061311 http://shopping.netsuite.com/digitalcheck TABLE OF CONTENTS Contents Radio Frequency Interference... 3 About the SB Scanner... 3 SB Scanner Equipment...
QFP910. Water detector. Building Technologies. CE1N2732en. Synco living
2 732 Synco living Water detector QFP910 Wireless radio sensor to record water puddles RF communication based on KNX standard (868 MHz unidirectional) Battery operation with standard 1.5 V batteries Use
QFP910. Water detector. Building Technologies. CE1N2732en 2016-03-01. Synco living
2 732 Synco living Water detector QFP910 Wireless radio sensor to record water puddles RF communication based on KNX standard (868 MHz unidirectional) Battery operation with standard 1.5 V batteries Use
Subminiature Load Cell Model 8417
w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...
1.4 Metallic Impurities Iron must not exceed 150 ppm and copper shall not exceed 6 ppm when tested according to TAPPI T 266.
Library of Congress Preservation Directorate Specification Number 100 132 09 Specifications for 3 x 10ʺ Paper Enclosures For the Storage of Strip Photographic Negatives This specification is provided as
DuPont Tyvek Commercial Air Barrier Assemblies Exceed Air Barrier Association of America, ASHRAE 90.1
Building Science Bulletin DuPont Tyvek Commercial Air Barrier Assemblies Exceed Air Barrier Association of America, ASHRAE 90.1 and IECC Air Leakage Requirements When Tested in Accordance with ASTM E2357
Length measuring instruments
INTERNATIONAL OIML R 66 RECOMMENDATION Edition 1985 (E) Length measuring instruments Instruments mesureurs de longueurs OIML R 66 Edition 1985 (E) ORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE INTERNATIONAL
SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE
Company Name Street Address City, State, Zip code Phone Number Fax Company Website Email Address ORGANIZATION NAME PHONE NUMBER EMAIL ADDRESS President/CEO General Manager Engineering Manager Production
site monitoring Kit Site Monitoring Kit User Manual we prove it.
site monitoring Kit 1 User Manual Site Monitoring Kit User Manual we prove it. 2 site monitoring Kit Content 1. Introduction 3 2. Content of the Site Monitoring Kit 4 3. Preparation 6 1. Positioning of
File:Quality-Management Manual-V2
Quality- Management Manual ChenYang Technologies Gmbh & Co.KG Markt Schwabener Str. 8 85464 Finsing Telephone:: 08121/ 25 74 100 Fax:: 08121 / 25 74 101 E-Mail: [email protected] Internet: www.chenyang.de
Peel Adhesion of Pressure Sensitive Tape
Harmonized International Standard Call Letters PSTC 101 Date of Issuance 10/00 Revised 10/03 Revised 05/07 Peel Adhesion of Pressure Sensitive Tape 1. SCOPE 1.1 These methods cover the measurement of the
Gamma Sterilisation Validation according to ISO 11137 - Sterilising dose -
Synergy makes sence Gamma Sterilisation Validation according to ISO 11137 - Sterilising dose - MG-FSI72-105 Last revision: March 2011 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel. 33 (0)4
Steam Sterilization and the 2007 Revision of PDA Technical Report 1
Steam Sterilization and the 2007 Revision of PDA Technical Report 1 14 November 2007 Presented By: Mike Finger (Tunnell Consulting) Don Drew (Abbott Bioresearch Center) Agenda: Introduction Sterilization
High Capacity Hot Air Dryer
High Capacity Hot Air Dryer MADE IN AUSTRIA autumn 2015 www.zimmer-austria.com page 1 Hot air dryer type: Compact HC The modular construction of the high efficiency and powerful hot air nozzle dryer is
Biological Indicators and
& and BIs play an important role in the monitoring of sterilizers, along with physical and chemical indicators. However, only BIs contain live spores that directly measure the lethality of a sterilization
Tabletop Line. Automatic and Semi-Automatic Autoclave Series
Tabletop Line Automatic and Semi-Automatic Autoclave Series Tuttnauer Your Sterilization & Infection Control Partner Company Profile For over 80 years, Tuttnauer s sterilization and infection control products
Particle Monitoring Requirements in Pharmaceutical Cleanrooms
Particle Monitoring Requirements in Pharmaceutical Cleanrooms All drugs must be manufactured in accordance with the current Good Manufacturing Practice (cgmp) regulations. Pharmaceutical manufacturers
Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm
Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm 1. How to prepare straws for pronuclear stages and embryos (use straws with plug, 66.5 mm) Straws are
Installation Instructions for Fletco Carpets
Table of Contents The backing is a very important part of a carpet, as well as the installation and the daily use are concerned, so therefore these installations instructions are based on the different
alpha DSL WTM HS Instructions for use ADSL modem alpha DSL WTM HS
alpha DSL WTM HS Instructions for use ADSL modem alpha DSL WTM HS Safety precautions Safety precautions The DSL modem alpha DSL WTM HS complies with the European norm EN60950 safety of information technology
Pharmaceutical Rotary Presses. PharmaPress 800. High Speed Double Rotary Press
Pharmaceutical Rotary Presses PharmaPress 800 High Speed Double Rotary Press Highest Output with... The KORSCH PharmaPress 800 is a high capacity, double sided tablet press for fully automated and unattended
Aesculap Service Systems. Consumables instacount, sterile containers and instrument care
Aesculap Service Systems Consumables instacount, sterile containers and instrument care Rev. May 2011 Consumables Steam sterilization IC-V001 Endless labels Double-sided self adhesive, outside wound, with
MHRA GMP Data Integrity Definitions and Guidance for Industry January 2015
MHRA GMP Data Integrity Definitions and Guidance for Industry Introduction: Data integrity is fundamental in a pharmaceutical quality system which ensures that medicines are of the required quality. This
L-LAS-TB-CL serie. laser light curtains for inline measuring tasks
L-LAS-TB-CL serie laser light curtains for inline measuring tasks Maximum distance 2.000 mm Measurement range 6 up to 98 mm Resolution up to 4 µm up to 1,5 khz scan rate 0-10 V or 4-20 ma analogue output
Quality Management System Manual
Quality Management System Manual Assurance ISO / AS Manual Quality Management System EXCEEDING ALL EXPECTATIONS Since our inception in 1965, Swiss-Tech has supplied the medical, aerospace, hydraulic, electronic
Regulations Concerning
Regulations Concerning Airborne Particle Counting Two types of activities performed with an airborne particle counter: Classification focuses primarily on the environment Particle Counting Applications
3039878000or8009926372
3039878000or8009926372 bi s a l e s @me s a l a bs. c om Regulatory officials and sterilization experts have voiced concerns regarding the appropriateness of using a Biological Indicator (BI) Ampoule interchangeably
HIGH PERFORMANCE PRE-APPLIED SYSTEM FOR BLIND SIDE & BELOW GRADE WATERPROOFING APPLICATIONS
BSW HIGH PERFORMANCE PRE-APPLIED SYSTEM FOR BLIND SIDE & BELOW GRADE WATERPROOFING APPLICATIONS BSW is a fully reinforced Pre-Applied system membrane designed for horizontal and vertical external blind-side
Precision Leaded Resistors
FEATURES Approved according to CECC 40101-806 Advanced thin film technology Low TC: ± 15 to Precision tolerance of value: ± 0.1 % and ± 0.25 % Superior overall : class 0.05 Wide precision range: 10 Ω to
How to Build a Printed Circuit Board. Advanced Circuits Inc 2004
How to Build a Printed Circuit Board 1 This presentation is a work in progress. As methods and processes change it will be updated accordingly. It is intended only as an introduction to the production
ECMO AND LIFE SUPPORT SYSTEMS QUADROX PLS AND ROTAFLOW HARDWARE AND ACCESSORIES
ECMO AND LIFE SUPPORT SYSTEMS QUADROX PLS AND ROTAFLOW HARDWARE AND ACCESSORIES The Gold Standard Cardiovascular QUADROX PLS and ROTAFLOW 3 QUADROX PLS AND ROTAFLOW INNOVATIVE PERFORMANCE AND RELIABILITY
KGI MEDICAL WASTE MANAGEMENT PLAN
1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:
Testing guideline. "special protected vehicles"
"" ( Sondergeschützte Fahrzeuge ) bullet resistance Englische Übersetzung; es gilt immer die deutsche Originalfassung! English translation; however the original German version always prevails! Publisher:
CERTIFICATE NAME OF PRODUCT MANUFACTURER PRODUCT DESCRIPTION CERTIFICATION PROCEDURE. No VTT C-6044-10 Date of issue 11.10.2010, Updated July 1, 2011
CERTIFICATE NAME OF PRODUCT No VTT C-6044-10 Date of issue 11.10.2010, Updated July 1, 2011 H-CONTROL REFLEX+ reflective insulating vapour control layer for roof and wall applications MANUFACTURER ACTIS
EUROPEAN STANDARD EN 343:2003 PROTECTION AGAINST RAIN
EUROPEAN STANDARD EN 343:2003 PROTECTION AGAINST RAIN Published by W. L. Gore & Associates Editorial support Petra Jackisch, Zentrum für Sicherheitstechnik, Erkrath, Germany Suitable protection against
Heat Transfer. Condensation. Revaporization
Objectives Wrapping a Towel Around the Problem of Wet Packs Identify causes of wet packs in steam sterilization Review suggested documentation that can assist in solving wet pack problems Why are Wet Packs
ALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
Produce Traceability Initiative Best Practices for Direct Print
Produce Traceability Initiative Best Practices for Direct Print About This Best Practice Guideline (Revision 1.0) Best practices are generally accepted, informally-standardized techniques, methods or processes
Check-list for auditors. Audit visit at candidate registered laboratories
Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be