MEDICAL APPLICATIONS OF FLUKA
|
|
|
- Arron Robbins
- 10 years ago
- Views:
Transcription
1 MEDICAL APPLICATIONS OF FLUKA Fisica Medica, 2012 G. Battistoni 1
2 FLUKA applicazioni: IORT Fisica Medica, 2012 G. Battistoni 2
3 FLUKA Applications: Linac Head Fisica Medica, 2012 G. Battistoni 3
4 Radiotherapy 4
5 FLUKA radiotherapy simulation 5
6 6 MeV Accelerator electron fluence distribution 6
7 6 MeV Accelerator photon fluence distribution 7
8 FLUKA MONTE CARLO SIMULATION FOR THE LEKSELL GAMMA KNIFE PERFEXION TM Fisica Medica, 2012 G. Battistoni 8
9 6 MeV Accelerator 9
10 6 MeV Accelerator 10
11 The Leksell Gamma Knife Perfexion: The Leksell Gamma Knife Perfexion (LGK-PFX) is a 60 Co based medical device, manufactured by Elekta AB Instruments Stockholm, Sweden. The It is emplyed in the cure of different brain pathologies: small brain and spinal cord tumors (benign and malignant), blood vessel abnormalities, as well as neurologic problems can be fully treated. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
12 The Leksell Gamma Knife Perfexion: The ionizing gamma radiation is emitted from Co sources (average activity ~1TBq each). The sources are arranged on 8 identical sectors of 24 elements. The sectors can be placed in correspondence of three different collimation set able to focus the gamma rays on a common spot, called the isocenter of the field, having a radial dimension of about 4, 8 and 16 mm respectively. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
13 Implementation of the Simulation: Geometry Modeling; Source Modeling; Definition of the bodies; Definition of the regions (Boolean algebra); Definition of the materials; Assignment of the materials to each region. Simulation Optimization. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
14 FLUKA Geometry Modelization: Thanks to the collaboration with ELEKTA, which provided, under a confidential agreement the detail of the geometry and all the involved material, has been possible to implement an accurate model for the radiation unit. ~ 1350 bodies Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
15 Geometry Modelization: Materials Protective shield; LEAD. Collimator channels; TUNGSTEN. Gammex 457 Solid Water Isocenter of the field. Radiative sources encapsulated in stainless steel bushings. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
16 Implementation of the Simulation: Geometry Modeling; Source Modeling; Definition of the geometries and materials. Definition of the particle type and energies. Simulation Optimization. Definition of the starting point of the beam. Definition of the trajectory of the beam. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
17 Source Modeling: Geometry and materials Metallic bushing. 60 Co cylindrical pellets of 1 mm in diameter and 1 mm in length. Fabrizio Cappucci INFN, Milan. The β- electron (average energy of about 315 kev) is supposed to be absorbed from the source or the bushing itself, therefore, each MC primary history is composed only by the two photons. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
18 Source Modeling: Implementation of FLUKA External Routine. 8mm 4mm 16mm Z Axis Randomly selects one of 192 available source position; Samples the starting point of the beam uniformly inside the source volume; Z Axis Samples the beam trajectories. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
19 Implementation of the Simulation: Geometry Modeling; Source Modeling; Optimizing the energy thresholds for production and transport of the secondary particles. Simulation Optimization. CPU TIME AVOID the production of secondary charged particles in the radiation unit body. Fabrizio Cappucci INFN, Milan. FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
20 Relative dose distribution: FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion System, Homogeneous media.
21 Results: We have investigated the relative linear dose distribution along the three coordinated axes for each kind of collimator size and the Relative Output Factors (ROF). Monte Carlo results have been compared with LGP data provided by Elekta in the same homogeneous conditions of the target primary histories (total calculation time of about 20h on 26 nodes) have been performed for each simulation. The simulations have been performed splitting the runs in a Macintosh cluster, made available by the Medical Physics department of Niguarda Ca Granda Milan Hospital.
22 Relative Output Factors (ROF): ROFs are the ratio between the dose given by a set of collimators and the dose given by the largest collimators, i.e. the 16 mm. The Elekta ROFs values are: (16 mm); (8 mm); (4 mm). The FLUKA scoring volume has been chosen as a cylinder of 1 mm in radius an 1 mm length placed around the isocenter point, with longitudinal axis lying along the Z axis.
23 Relative Output Factors (ROF): Collimator size Elekta ROF FLUKA ROF MC Statistical Error 8 mm % 0.43% 4 mm % 0.63% Δ is the percentage difference between the results from Monte Carlo calculation and the Elekta values: ELEKTA OF FLUKA OF
24 Relative dose profiles: quality index definition The output data distributions of the FLUKA simulation have been smoothed by means of a linear interpolation (spline) with 2% of tolerance. The quality index we used to perform the comparison is base on the Gamma Index (GI) method. This index takes into account the relative shift both in terms of intensity and in terms of position, combining together the dose difference (DD) and the distance to agreement (DTA) methods. DD is the difference between the dose value in a measured data point and the dose value in a point of the calculated distribution which has the same coordinates. By definition, if GI>1 the test is not passed and the two distribution are not in agreement. DTA is the distance between a measured data point and the nearest point in the calculated dose distribution that has the same dose value.
25 Relative dose profile 16mm X profile 4mm Z profile 8mm Y profile The blue line represents the gamma index trend. It results < 1 for each simulated point, attesting the goodness of the FLUKA model. The acceptance threshold values for DD and DTA are derived from the Report of the American Association of Physicists in Medicine for stereotactic radiosurgery, respectively as DD=3% and DTA=2.4 mm
26 Radioactive source decay FLUKA contains data about decaying schemes of radioactive isotopes, allowing to select an isotope as radiation source. Complete databases are generated from the data collected from National Nuclear Data Center (NNDC) at Brookhaven National Laboratory. Routines are available to discriminate only a component of the emission spectrum, simulating for example only the beta emission and disregarding the gamma one. Application in nuclear medicine Fisica Medica, 2012 G. Battistoni 26
27 Application in nuclear medicine Calculation of absorbed dose at voxel level starting from 3D images of activity distribution (SPECT, PET images) Simulations in homogeneous water Simulated 99 Tc-SPECT of water phantoms (SIMIND code): # 1 # 2 Dose calculation: Cylinder + spheres filled with 90 Y Fisica Medica, 2012 G. Battistoni 27
28 SPECT/PET - CT images handling DOSE Maps Collaboration INFN and IEO VOXEL Dosimetry MONTE CARLO 10 9 particles With 10 9 particles simulated, FLUKA and VOXEL DOSIMETRY results in water agree within 5% Fisica Medica, 2012 G. Battistoni 28
29 Rationale for MC in hadron-therapy Biological calculations in tumour therapy with ions depend on a precise description of the radiation field. In 12 C ion irradiation, nuclear reactions cause a significant alteration of the radiation field. contribution of secondary fragments needs to be taken into account for accurate planning of the physical and biological dose delivery in the scheduled treatment. Treatment Planning Systems (TPS) for ion beam therapy essentially use analytical algorithms with input databases for the description of the ion interaction with matter. Monte Carlo codes with sophisticated nuclear models are more efficient (though slower) computational tools to handle the mixed radiation field. Fisica Medica, 2012 G. Battistoni 29
30 Rationale for MC in hadron-therapy In practice MC codes can be used for: startup and commissioning of new facilities beamline modeling and generation of TPS input data validate analytical TPSs in water/ct systems both for physical and biological aspects Prediction/Analysis of in-beam PET application Biological calculations for cell survival experiments Additional advantage to describe complex geometries (and interfaces between rather different materials!): Accurate 3D transport Fully detailed description of the patient anatomy CT image converted into a MC geometry Model challenge: interface to radiobiological model to predict biological dose ( actual effect) and not only physical dose Fisica Medica, 2012 G. Battistoni 30
31 Nuclear reactions: what really matters? Proton therapy (p, E: MeV): Reaction cross sections (beam attenuation) +++ Elastic cross sections + Particle (p,n,α..) emission + (mostly for background, ++ radiobiology) Positron emitter production + (data available) Therapy with ions (ions, E: MeV/n): Reaction cross sections (beam attenuation) +++ Fragment (α included) production +++ Particle emission, p +++, others + Positron emitter production +++ Conventional therapy (γ, E: 3-30 MeV) (γ,x) (particularly (γ,n)) + (mostly for background) Fisica Medica, 2012 G. Battistoni 31
32 Main FLUKA developments in view of medical applications (and hadron therapy in particular) Models for nucleus-nucleus interactions : Improvement of models for evaporation/fission/fragmentation used in fragment final de-excitation. Prediction of radionuclide production Improvement of de/dx models (Z 2 +Z 3 corrections, molecular effects, nuclear stopping power) Run time application of linear-quadratic models describing radiobiological effects Extensions and improvement of neutron library (thermal + ephithermal region) Voxel geometry Time-varying geometry Routines to import CT scans, material/density/composition assignment to CT Fisica Medica, 2012 G. Battistoni 32
33 Main references to FLUKA in ion-therapy related matters: 1) F.Sommerer, K.Parodi, A.Ferrari, K.Poljanc,W.Enghardt and H.Aiginger, Investigating the accuracy of the FLUKA code for transport of therapeutic ion beams in matter, Phys. Med. Biol. 51 (2006) ) K.Parodi, A.Ferrari, F.Sommerer and H.Paganetti, Clinical CT-based calculations of dose and positron emitter distributions in proton therapy using the FLUKA Monte Carlo code, Phys. Med. Biol. 52 (2007) ) A. Mairani, Nucleus-Nucleus Interaction Modelling and Applications in Ion Therapy Treatment Planning, PhD Thesis, Univ. Pavia, ) G.B. et al. (FLUKA collaboration), The FLUKA code and its use in hadron therapy, Il Nuovo Cimento 31C, no. 1 (2008) 69. 5) F.Sommerer, F.Cerutti, K.Parodi, A.Ferrari, W.Enghardt and H.Aiginger, In-beam PET monitoring of mono-energetic 16O and 12C beams: experiments and FLUKA simulations for homogeneous targets, Phys. Med. Biol. 54 (2009) ) A.Mairani, S.Brons, A.Fassò, A.Ferrari, M.Krämer, K.Parodi, M.Scholz and F. Sommerer, Monte Carlo based biological calculations in carbon ion therapy: the FLUKA code coupled with the Local Effect Model, submitted to PMB 2010 Fisica Medica, 2012 G. Battistoni 33
34 Nuclear interactions: a nuisance x therapy easy to switch off in MonteCarlo, a bit less in real life nice at least I have a good job Nuclear (and particle) physics cannot be simulated at % level, sometimes you are happy with results within a factor of 2 That s why we always plot results in log-log scales Fisica Medica, 2012 G. Battistoni 34
35 From Ref. 1) Projectile Fragmentation: 12 C, 14 N, 16 O at 640 MeV/n Sommerer et al PMB Fisica Medica, 2012 G. Battistoni 35
36 The FLUKA voxel geometry The CT scan contains integer values Hounsfield Unit reflecting the X-ray attenuation coefficient m x HU x = 1000 (m x -m H20 ) / m H20, typically HU 3500 We will use loosely the word organ to indicate a group of voxels (or even more than one group) made of the same tissue material (same HU value or in a given HU interval) The code handles each organ as a CG region, possibly in addition to other conventional non-voxel regions defined by the user The voxel structure can be complemented by parts written in the standard Combinatorial geometry The code assumes that the voxel structure is contained in a parallelepiped. This RPP is automatically generated from the voxel information. To describe a voxel geometry, the user provides the CT scan or equivalent data in a format understood by FLUKA by means of an external conversion program Fisica Medica, 2012 G. Battistoni 36
37 Voxel geometries: examples The anthropomorphic GOLEM phantom Implementation in FLUKA (radioprotection applications) Now available the official ICRP Human Phantom ICRP Publication 110: Adult Reference Computational Phantoms - Annals of the ICPR Volume 39 Issue 2 Petoussi-Henss et al, 2002 Fisica Medica, 2012 G. Battistoni 37
38 The FLUKA voxel geometry This stage should : Assign an organ index to each voxel. In many practical cases, the user will have a continuum of CT values (HU), and may have to group these values in intervals Each organ is identified by a unique integer The organ numbering does not need to be contiguous (i.e. holes in the numbering sequence are allowed.) One of the organs must have number 0 and plays the role of the medium surrounding the voxels (usually vacuum or air). The user assigns to each NONZERO organ a voxel-region number. The voxel-region numbering has to be contiguous and starts from 1. Fisica Medica, 2012 G. Battistoni 38
39 Practical issues for Medical Applications General problems for MC calculations on CT scans How to assign realistic human tissue parameters (= materials) for MC Calculation? How to find a good compromise between the number of different HU values (~ ) and the materials to be considered in the MC? (issues on memory and computation speed when attempting to treat each HU number as a different material!!!) How to preserve continuous, HU-dependent information when segmenting the HU numbers into intervals sharing the same tissue material? (critical for ion range calculation in charged hadron therapy!!!) Fisica Medica, 2012 G. Battistoni 39
40 CT stoichiometric calibration (I) CT segmentation into 27 materials of defined elemental composition (from analysis of 71 human CT scans) Air, Lung, Adipose tissue Soft tissue Skeletal tissue Fisica Medica, 2012 G. Battistoni Schneider et al PMB 45,
41 CT stoichiometric calibration (II) Assign to each material a nominal mean density, e.g. using the density at the center of each HU interval (Jiang et al, MP 2004) Schneider et al PMB 45, 2000 But real density (and related physical quantities) Fisica varies Medica, 2012 continuously with G. HU Battistoni value!!! 41
42 Rasterscan GSI / HIT scanning of focussed ion beams in fast dipole magnets active variation of the energy, focus and intensity in the accelerator and beam lines Haberer et al., NIM A, 1993
43 MC for the Heidelberg Ion Therapy Center (HIT) Ion species p, 12 C (later also 3 He, 16 O) Heidelberg Ion Beam therapy Center Beam delivery Scanning with active energy variation GSI) Required parameters: 255 Energy steps 4 Foci 10 Intensities Start of clinical operation: November 2009
44 Mixed Radiation Field in Carbon Ion Therapy 12 C (400 MeV/u) on water Attenuation of primary beam Build-up secondary fragments Exp. Data (STARS) FLUKA (POINTS - LINE) Exp. Data (points) from Haettner et al, Rad. Prot. Dos Simulation: A. Mairani PhD Thesis, 2007, PMB to be published
45 Mixed Radiation Field in Carbon Ion Therapy 12 C (400 MeV/u) on water Angular distribution Energy distribution H He Li Be B cm 2 Exp. Data (points) from Haettner et al, Rad. Prot. Dos Simulation: A. Mairani PhD Thesis, 2007, PMB to be published TOF spectra in progress
46 The role of MC in ion therapy Treatment planning systems (TPS) use analytical models MC are increasingly used computational tools to support: Startup and Commissioning of new facilities: e.g., beamline modeling and generation of TPS input data ( meas. time) Validation of TPS absorbed and biological dose calculations: in water-equivalent system and in patient anatomy (CT) Dedicated applications: imaging of secondary emerging radiation for treatment verification, like PET monitoring of ion therapy
47 Relative Dose in Water Passively formed proton MGH FLUKA coupled with phase space information from a separate Geant4 MCcalculation (Paganetti et al, MP 31, 2004) modeling nozzle and modifiers Parodi et al, Med Phys SOBP MC vs. IC meas.: 80% and 90% fall-off positions agree within 1 mm Depth (mm)
48 Focus (FWHM) in Isocentre (mm) MC for the Heidelberg Ion Therapy Center (HIT) Parodi, Brons, Naumann, Haberer et al, to be published Ion species p, 12 C (later also 3 He, 16 O) 12 HIT (Nominal focus from FLUKA b) Heidelberg Ion Beam therapy Center Beam delivery Scanning with active energy variation GSI) Required parameters: 255 Energy steps 4 Foci 10 Intensities Energy (MeV/u) Start of clinical operation: November 2009 Taking into account desired 4-10 mm ISO for higher E and physical limitation from scattering in monitor system and air
49 Dose (MC: MeVcm 2 /g; Meas. Norm. to area) Dose (MC: MeVcm 2 /g; Meas. Norm. to area) MC for the Heidelberg Ion Therapy Center (HIT) Generation of basic input data for TPS depth dose profiles FLUKA calculation of depth-dose profiles in water w/wo ripple filter (RiFi) for all the 255 energies delivered by accelerator for p and 12 C HIT 12 C + 3mm HIT Depth (cm) Parodi, Mairani et al to be published Depth (cm)
50 MC for the Heidelberg Ion Therapy Center (HIT) Generation of basic input data for TPS Fragment Spectra FLUKA calculation of fragment spectra in water w/wo ripple filter (RiFi) for 12 C ( MeV/u) 100 MeV/u Beam attenuation profile FLUKA calculated database 200 MeV/u 300 MeV/u 400 MeV/u Mairani, Parodi et al to be published
51 MC for the Heidelberg Ion Therapy Center (HIT) Generation of basic input data for TPS Fragment Spectra FLUKA calculation of fragment spectra in water w/wo ripple filter (RiFi) for 12 C ( MeV/u) 12 C ions (400 MeV/u) - Energy Spectra at 25 cm FLUKA calculated database Mairani, Parodi et al to be published
52 MC for the Heidelberg Ion Therapy Center (HIT) Forward recalculation of TPS treatment plans in water (i.e. medium where plans are experimentally verified) FLUKA coupled with control file of raster scanning system + modeling of the beam-line (F. Sommerer et al, EWG-MCTP Workshop, Ghent 2006, A. Mairani et al to be submitted) Carbon ion Gy 90 % FLUKA Exp Data 80 % Primary Beam Secondary Fragments Mairani, Parodi et al to be published
53 MC for the Heidelberg Ion Therapy Center (HIT) Forward recalculation of TPS treatment plans in water Gy FLUKA Exp Data 30 mm 70 mm 110 mm Primary Beam H He Li Be B C 30 mm 70 mm 110 mm
54 MC for the Heidelberg Ion Therapy Center (HIT) Forward recalculation of TPS treatment plans in water (i.e. medium where plans are experimentally verified) FLUKA coupled with control file of raster scanning system + modeling of the beam-line Protons FLUKA Exp Data Primary Beam Secondaries Mairani, Parodi et al to be published
55 MC versus TPS for proton MGH mgy Clivus Chordoma Patient mgy XiO/CMS FLUKA XiO/CMS mgy FLUKA mgy Parodi et al, JPCS 74, 2007
56 MC versus TPS for carbon ion GSI Clivus Chordoma Patient Absorbed Dose TRiP FLUKA CT Absorbed Dose Spread-Out Bragg Peak in the patient A. Mairani PhD thesis 2007, Pavia A. Mairani et al IEEE 2008 TRiP mgy FLUKA mgy
57 Biological MC calculations: FLUKA coupled with external bio database Biaggi et al NIM B 159, 1999
58 CHO Cells Photon Dose-Effect LOCAL EFFECT MODEL Radial Dose Distribution M. Scholz et al (GSI) Microscopic dose distribution
59 Biological MC calculations: FLUKA coupled with the Local Effect Model Starting from the intrinsic LEM coefficients, the α D and β D parameters are obtained in terms of macroscopic dose applying the rapid approach described in : PMB 51 (2006) 1959 M. Krämer and M. Scholz The coupling of the LEM with FLUKA has been carried out using the Theory of Dual Radiation Action (TDRA) (A. M. Kellerer and H. H. Rossi, Radiat. Res. 75 (1978) 471): a biological system exposed to more than one radiation type will show synergism, implying that the total number of lesions is larger than the sum of the lesions produced by each single beam component, due to interactions between sub-lesions produced by different components. A. Mairani et al, IEEE 2008 and PMB 2010 submitted
60 Biological MC calculations: comparison with experimental data and analytical calculations Benchmark in water ( A. Mairani PhD Thesis 2007): 270 MeV/n carbon beam 187 MeV/n carbon beam FLUKA LEM Exp. data (points) courtesy of M. Scholz (GSI) Radiat. Environ. Biophys. 36 (1997) 59 Exp. data (points) courtesy of M. Scholz (GSI Radiat. Environ. Biophys. 36 (1997) 59
61 Biological MC calculations: comparison with experimental data and analytical calculations Clinical calculation: Two opposing dose ramps with tissue sparing (as brain-stem) TRiP98 Analytical calculations: biological planning and optimization for CHO cells FLUKA-LEM: forward calculation of the optimized plan Exp. Data and analytical calculations: Krämer et al, PMB 48 (2003) 2063 MC Calculations: A. Mairani et al, PMB 2010 submitted
62 Biological MC calculations: comparison with experimental data and analytical calculations Clinical calculation: Two opposing dose ramps with tissue sparing (as brain-stem) TRiP98 Analytical calculations: biological planning and optimization for CHO cells FLUKA-LEM: forward calculation of the optimized plan Solid line : FLUKA Dashed line: TRiP98 Squares: Exp. Data Exp. Data and analytical calculations: Krämer et al, PMB 48 (2003) 2063 MC Calculations: A. Mairani et al, PMB 2010 submitted
63 A. Mairani et al, to be published MC versus TPS for carbon ion GSI Clivus Chordoma Patient Absorbed/Biological Dose mgy mgye FLUKA FLUKA - LEM TRiP mgy TRiP mgye
64 The principle of PET verification Before collision Projectile 12 C ion Atomic nucleus of tissue After collision C C 16 O 15 O Projectile fragment Neutrons Target fragment Before collision Proton Atomic nucleus of tissue After collision 16 O 15 O Target fragment Proton Neutron 12 C ions, E=212 A MeV 11 C, 10 C 15 O, 11 C,... Protons, E=110 MeV 15 O, 11 C,... K. Parodi et al, IEEE TNS 2005 By-product of irradiation ( 15 O, 11 C, 13 N...with T 1/2 ~ 2, 20 and 10 min)
65 Before collision Projectile 12 C ion Atomic nucleus of tissue The principle of PET verification After collision C C 16 O 15 O Projectile fragment Neutrons Target fragment A(r) D(r) Before collision Proton Atomic nucleus of tissue After collision 16 O 15 O Measured activity compared with calculation Target fragment Proton Neutron 12 C ions, E=212 A MeV 11 C, 10 C 15 O, 11 C,... Protons, E=110 MeV 15 O, 11 C,... K. Parodi et al, IEEE TNS 2005 By-product of irradiation ( 15 O, 11 C, 13 N...with T 1/2 ~ 2, 20 and 10 min)
66 Calculation model of + -activation FLUKA Monte Carlo code using Field-specific beam source information from Geant4 modeling of the nozzle and beam modifiers (Paganetti et al, MP 31, 2004) Planning CT (segmented into 27 material) and same CT-range calibration curve as TPS (Parodi et al MP 34, 2007, PMB 52, 2007) Experimental cross-sections for + -emitter production Semi-empirical biological modeling (Parodi et al IJROBP 2007) Convolution with 3D Gaussian kernel (7-7.5 mm FWHM) and other reaction channels on N, O, Ca yielding, e.g., 13 N, 38 K, Data from IAEA Nuclear Data Section (Parodi et al, PMB 45, 2002, Parodi et al, PMB 52, 2007)
67 In-beam PET: ion beam fragmentation Final goal: simulation of β + emitters generated during the irradiation In-beam treatment plan verification with PET Work in progress: FLUKA validation Comparison with experimental data on fragment production (Schall et al.) 12C, 14N, 16O beams, 675 MeV/A Adjustable water column cm Z spectra of escaping fragments for Z 4 Cumulative yield of light fragments Simulation: corrections applied for angular acceptance and for material in the beam upstream the water target Comparison with experimental data on + -emitter production (Fiedler Fisica Medica, et al.) G. Battistoni 67
68 + -Activity after Irradiation graphite PMMA Measured min after irradiation, therefore dominated by 11 C water Further work: processing with same software than experiment more primaries for better statistic year Fisica Medica, 2012 Final results are expected for the end of this G. Battistoni 68
69 F. Sommerer PhD Thesis, 2007, F. Sommerer et al PMB Backprojections: FLUKA vs Exp data 12 C 260 MeV/A on PMMA 12 C 260 MeV/A on PMMA, simulated relative production rate of different isotopes Both the data and the FLUKA calculations are normalized to the same area
70 Applications of FLUKA to p MGH Input phase-space provided by H. Paganetti, MGH Boston TPS mgy FLUKA mgy Parodi et al, JPCS 74, 2007 Prescribed dose: 1 GyE MC : ~ protons in 10 independent runs (11h each on Linux Cluster mostly using 2.2GHz Athlon processors) Fisica Medica, 2012 G. Battistoni 70
71 Applications of FLUKA to p MGH mgy mgy Parodi et PMB 52, 2007 TPS FLUKA TPS mgy FLUKA mgy Fisica Medica, 2012 G. Battistoni 71
72 TPS mgy Applications of FLUKA to p MGH FLUKA mgy metallic implants Prescribed dose: 2 GyE MC : ~ p in 12 independent runs (~ 130h each on 2.2 GHz Linux cluster) K. Parodi et al, IJROBP 2007 Fisica Medica, 2012 G. Battistoni 72
73 Applications of FLUKA to p MGH Clival Chordoma, 0.96 GyE / field mgy Planned dose mgy Fisica Medica, 2012 G. Battistoni 73
74 Post-radiation MGH Average Activity Clival Chordoma, 0.96 GyE / field, T1 ~ 26 min, T2 ~ 16 min Bq / ml Bq / ml 2 Field 1 Field K. Parodi et al, IJROBP 2007 and FLUKA-voxel functionalities being also G. Battistoni used at HIT Fisica Medica,
75 β + emitters for ion beams: phantom experiments Application of FLUKA to PET monitoring of ion species (e.g. 12 C, 16 O) based on internal nuclear models Simulation of imaging process (β + -decay, propagation of e + and annihilation photons, detection) same as for measured data Exact replica of the experimental setup, PET heads included FLUKA irradiation+decay features exploited MC γ s reaching PET heads converted to list-mode data by modified PETSIM 1 Backprojection with same routines as in experiment In-beam GSI 1 Pönisch et al. PMB MeV/u 12 C ion on Graphite, backprojections Measurement FLUKA Fisica Medica, 2012 G. Battistoni 75 F. Sommerer PhD Thesis, 2007, F. Sommerer et al PMB
76 END Fisica Medica, 2012 G. Battistoni 76
Development of on line monitor detectors used for clinical routine in proton and ion therapy
Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary
Status and perspective of emission imaging techniques for ion beam therapy in Lyon
14th International Conference on Nuclear Reaction Mechanisms, Varenna 18th June 2015 Status and perspective of emission imaging techniques for ion beam therapy in Lyon on behalf of the CAS-PHABIO group
Medical Applications of radiation physics. Riccardo Faccini Universita di Roma La Sapienza
Medical Applications of radiation physics Riccardo Faccini Universita di Roma La Sapienza Outlook Introduction to radiation which one? how does it interact with matter? how is it generated? Diagnostics
ACCELERATORS AND MEDICAL PHYSICS 2
ACCELERATORS AND MEDICAL PHYSICS 2 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 2-28.10.10 - U. Amaldi 1 The icone of radiation therapy Radiation beam in matter EPFL 2-28.10.10 - U.
Secondary Neutrons in Proton and Ion Therapy
Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles
Quality Assurance for Treatment Planning Systems
Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline
Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters
C-A/AP/#122 November 23 Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters J. Beebe-Wang, P. Vaska, F.A. Dilmanian, S.G. Peggs and D.J. Schlyer "This paper
Variance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
Therapy with protons and ion beams
Therapy with protons and ion beams Biomedical Physics Lecture WiSe 2012/13 many slides from J. Willkens (TUM) Key quesions: How does ion cancer therapy work? What is the physics behind it? How can physics
HADRON THERAPY FOR CANCER TREATMENT
HADRON THERAPY FOR CANCER TREATMENT Seminar presented by Arlene Lennox at Fermilab on Nov 21, 2003 CANCER STAGES LOCAL TUMOR REGIONAL METASTASIS SYSTEMIC DISEASE CANCER TREATMENT SURGERY RADIATION THERAPY
Monte Carlo Simulations in Proton Dosimetry with Geant4
Monte Carlo Simulations in Proton Dosimetry with Geant4 Zdenek Moravek, Ludwig Bogner Klinik und Poliklinik für Strahlentherapie Universität Regensburg Objectives of the Study what particles and how much
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
How To Use A Proton For Radiation Therapy
Proton Cancer Therapy: Using Laser Accelerate Protons for Radiation Therapy A Thesis Submitted in Partial Fulfillment of the Requirements for Graduation with Research Distinction in Engineering Physics
Overview of Proton Beam Cancer Therapy with Basic Economic Considerations
Overview of Proton Beam Cancer Therapy with Basic Economic Considerations Wayne Newhauser, Ph.D. Proton Therapy Project, Houston Cyclotron 235 MeV 300 na Extraction Channel Radial Probe Energy Degrader
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
Radiotherapy in Hungary: present status and future needs. Tibor Major, PhD National Institute of Oncology Radiotherapy Department Budapest, Hungary
Radiotherapy in Hungary: present status and future needs Tibor Major, PhD National Institute of Oncology Radiotherapy Department Budapest, Hungary Academia Europaea Section Workshops, Bergen, 10 September,
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect [email protected]. Protect, Enhance and Save Lives
Vincent Bossier System Architect [email protected] IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
World-first Proton Pencil Beam Scanning System with FDA Clearance
Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi
Current and Future Trends in Proton Treatment of Prostate Cancer
Current and Future Trends in Proton Treatment of Prostate Cancer Reinhard W. Schulte Assistant Professor Department of Radiation Medicine Loma Linda University Medical Center Loma Linda, CA, USA Outline
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Basic Radiation Therapy Terms
Basic Radiation Therapy Terms accelerated radiation: radiation schedule in which the total dose is given over a shorter period of time. (Compare to hyperfractionated radiation.) adjuvant therapy (add-joo-vunt):
90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons
SLAC-PUB-7722 January 9 degrees Bremsstrahlung Source Term Produced in Thick Targets by 5 MeV to GeV Electrons X. S. Mao et al. Presented at the Ninth International Conference on Radiation Shielding, Tsukuba,
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy
Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.176-180 (2011) ARTICLE Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy Fada
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
Use of 3D Printers in Proton Therapy
Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology [email protected] 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing
The Science behind Proton Beam Therapy
The Science behind Proton Beam Therapy Anthony Zietman MD Shipley Professor of Radiation Oncology Massachusetts General Hospital Harvard Medical School Principles underlying Radiotherapy Radiation related
ADROTERAPIA A MEDAUSTRON
ADROTERAPIA A MEDAUSTRON Antonio Carlino Qualified Medical Physicist EBG MedAustron GmbH, Wiener Neustadt, Austria Outline Introduction and milestones Building and construction Treatment beam lines and
Seminar 7. Medical application of radioisotopes - radiotherapy
Seminar 7 Medical application of radioisotopes - radiotherapy Radioisotopes in medical diagnosis. Gamma camera. Radionuclide imaging (PET, SPECT). Radiotherapy. Sources of radiation. Treatment planning.
Cancer Therapy with Particle Beams
Cancer Therapy with Particle Beams The potential to use high energy particle beams to treat many types of cancer has been known even before their creation. The availability of these treatments has been
MEDICAL DOSIMETRY. COLLEGE OF APPLIED SCIENCES AND ARTS Graduate Faculty: therapy.
Graduate Catalog 2012-2013 Medical Dosimetry / 339 MEDICAL DOSIMETRY COLLEGE OF APPLIED SCIENCES AND ARTS Graduate Faculty: Collins, Kevin Scott, Associate Professor, Ph.D., Jensen, Steve, Emeritus Professor,
Proton beam. Medical Technical Complex Joint Institute for Nuclear Research, Dubna, Russia
Proton beam Medical Technical Complex Joint Institute for Nuclear Research, Dubna, Russia Egyptian Candidates Proton therapy It can do more!!! Effectiveness of the proton therapy in targeting and curing
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
Proton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
Lecture 2 Macroscopic Interactions. 22.106 Neutron Interactions and Applications Spring 2010
Lecture 2 Macroscopic Interactions 22.106 Neutron Interactions and Applications Spring 2010 Objectives Macroscopic Interactions Atom Density Mean Free Path Moderation in Bulk Matter Neutron Shielding Effective
Treating Thyroid Cancer using I-131 Maximum Tolerable Dose Method
Treating Thyroid Cancer using I-131 Maximum Tolerable Dose Method Christopher Martel, M.Sc., CHP Lisa Thornhill,, NRRPT, RT(NM) Boston University Medical Center Thyroid Carcinoma New cases and deaths in
Protons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.
Protons vs. CyberKnife UC SF Protons vs. CyberKnife UC SF Alexander R. Gottschalk, M.D., Ph.D. Associate Professor and Director of the CyberKnife Radiosurgery Program Department of Radiation Oncology University
Laue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator
Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator R. Gupta and M.Harrison, Brookhaven National Laboratory A. Zeller, Michigan State University
Current Status and Future Direction of Proton Beam Therapy
Current Status and Future Direction of Proton Beam Therapy National Cancer Center Hospital East Division of Radiation Oncology and Particle Therapy Tetsuo Akimoto Comparison of status of particle therapy
Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations
Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations Daryoush Sheikh-Bagheri, Ph.D. 1, Iwan Kawrakow, Ph.D. 2, Blake Walters, M.Sc. 2, and David W.O. Rogers, Ph.D.
Medical Physics and Radioactivity
Medical Physics and Radioactivity Radioactivity Unstable nucleus Electromagnetic wave particle Atoms which emit electromagnetic radiation or a particle by the spontaneous transformation of their nucleus
HITACHI Proton Beam Therapy System
HITACHI Proton Beam Therapy System Masumi Umezawa, M.S. Hitachi, Ltd., Hitachi Research Laboratory HITACHI Proton Beam Therapy System Contents 1. Overview of HITACHI 2. HITACHI Proton Beam Therapy and
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
Calorimetry in particle physics experiments
Calorimetry in particle physics experiments Unit n. 8 Calibration techniques Roberta Arcidiacono Lecture overview Introduction Hardware Calibration Test Beam Calibration In-situ Calibration (EM calorimeters)
Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy
Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino
Status of Radiation Safety System at
Status of Radiation Safety System at Taiwan Photon Source Joseph C. Liu Radiation and Operation Safety Division National Synchrotron Radiation Research Center, Taiwan NSRRC layout 1.5 GeV, 120m, 400 ma
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
Proton Radiotherapy. Cynthia Keppel Scientific and Technical Director Hampton University Proton Therapy Institute. Lead Virginia.
Proton Radiotherapy Cynthia Keppel Scientific and Technical Director Hampton University Proton Therapy Institute Lead Virginia g October 16, 2009 About Proton Therapy Cancer is the (2nd) largest cause
Energy Deposition in MICE Absorbers and Coils
in MICE Absorbers and Coils November 2, 2003 Video Conference Page 1 An Application for G4Mice We would like to estimate how much energy is deposited in magnet coils and the hydrogen absorber. Most of
Neutron Detection Setups proposed for
Neutron Detection Setups proposed for DESPEC D. Cano-Ott on behalf of the WG members CIEMAT, IFIC, LNL, FYL, UPC, UU, UW Motivation GOAL: to measure neutron emission probabilities and energies for neutron
Radioactivity III: Measurement of Half Life.
PHY 192 Half Life 1 Radioactivity III: Measurement of Half Life. Introduction This experiment will once again use the apparatus of the first experiment, this time to measure radiation intensity as a function
Deformable Registration for Image-Guided Radiation Therapy
Symposium Multimodality Imaging for Radiotherapy : State of the art, needs and perspectives 1/33 Deformable Registration for Image-Guided Radiation Therapy David Sarrut «Radiation, images, oncology» team
EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction
ACTA PHYSICA DEBRECINA XLVI, 25 (2012) EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES L. Csedreki 1 1 Institute of Nuclear Research of the Hungarian Academy of Sciences,
Dynamic Load Balancing of Parallel Monte Carlo Transport Calculations
Dynamic Load Balancing of Parallel Monte Carlo Transport Calculations Richard Procassini, Matthew O'Brien and Janine Taylor Lawrence Livermore National Laboratory Joint Russian-American Five-Laboratory
Nuclear data for science & technology: Centres for development
Nuclear data for science & technology: Centres for development An update on the IAEA's nuclear data centre, its services, and the contributions to the global network from developing by Hans Lemmel f\ typical
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Advanced Radiation Therapy of Cancer by Proton Beam
March 20th, 2014 Advanced Radiation Therapy of Cancer by Proton Beam Fukui Prefectural Hospital Proton Therapy Center Yamamoto, Kazutaka Wave (electromagnetic wave) IR (Ionizing) Radiation 700 400 350~100
Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission
Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission MAVEN Science Community Workshop December 2, 2012 Particles and Fields Package Solar Energetic Particle Instrument (SEP) Davin Larson and the SEP
VARIANCE REDUCTION TECHNIQUES FOR IMPLICIT MONTE CARLO SIMULATIONS
VARIANCE REDUCTION TECHNIQUES FOR IMPLICIT MONTE CARLO SIMULATIONS An Undergraduate Research Scholars Thesis by JACOB TAYLOR LANDMAN Submitted to Honors and Undergraduate Research Texas A&M University
Brain Tumor Treatment
Scan for mobile link. Brain Tumor Treatment Brain Tumors Overview A brain tumor is a group of abnormal cells that grows in or around the brain. Tumors can directly destroy healthy brain cells. They can
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
Experimental study of beam hardening artefacts in photon counting breast computed tomography
Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,
The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.
The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. KAROLINA KOKUREWICZ S U P E R V I S O R S : D I N O J A R O S Z Y N S K I, G I U S E P P E S C
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range The radiation from orthovoltage units is referred to as x-rays, generated by bombarding
Proton Therapy Center at the Institute of Nuclear Physics in Kraków - from the Eye Treatment to the Scanning Gantry
Proton Therapy Center at the Institute of Nuclear Physics in Kraków - from the Eye Treatment to the Scanning Gantry Teresa Cywicka - Jakiel Institute of Nuclear Physics (IFJ PAN), Kraków, POLAND National
Scripps Proton Therapy Center: Configuration and Implementation
Scripps Proton Therapy Center: Configuration and Implementation Anthony Mascia Proton Symposium 2013 Annual AAPM Meeting Indianapolis, Indiana USA Facility Configuration Scripps Proton Therapy Center ***All
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
Cross section, Flux, Luminosity, Scattering Rates
Cross section, Flux, Luminosity, Scattering Rates Table of Contents Paul Avery (Andrey Korytov) Sep. 9, 013 1 Introduction... 1 Cross section, flux and scattering... 1 3 Scattering length λ and λ ρ...
Advanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Our Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
Geometrical importance sampling in Geant4: from design to verification
Geometrical importance sampling in Geant4: from design to verification Michael Dressel CERN, CH-1211 Geneva 23, Switzerland Abstract CERN-OPEN-2003-048 Sep 18 2003 The addition of flexible, general implementations
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
Objectives 404 CHAPTER 9 RADIATION
Objectives Explain the difference between isotopes of the same element. Describe the force that holds nucleons together. Explain the relationship between mass and energy according to Einstein s theory
Mission: Cure, Research and Teaching. Marco Pullia Cnao design and commissioning
CNAO design and commissioning Marco Pullia, CNAO Foundation What is the CNAO Foundation No profit organisation (Foundation) created with the financial law 2001 to build the national center for hadrontherapy
Solar Energy. Outline. Solar radiation. What is light?-- Electromagnetic Radiation. Light - Electromagnetic wave spectrum. Electromagnetic Radiation
Outline MAE 493R/593V- Renewable Energy Devices Solar Energy Electromagnetic wave Solar spectrum Solar global radiation Solar thermal energy Solar thermal collectors Solar thermal power plants Photovoltaics
The Master course Medical Radiation Sciences and the new profile Medical Radiobiology
OncoRay National Center for Radiation Research in Oncology, Dresden The Master course Medical Radiation Sciences and the new profile Medical Radiobiology Mario Helm Outline Teaching at OncoRay Postgraduate
Ionizing Radiation, Czech Republic, CMI (Czech Metrology Institute)
Ionizing Radiation, Czech Republic, (Czech Metrology Institute) Calibration or Measurement RADIOACTIVITY 1.0E+00 1.0E+02 Bq cm -2 C-14 1.0E+01 1.0E+02 Bq cm -2 Co-60 1.0E+01 1.0E+02 Bq cm -2 Sr-90 1.0E+01
Antoine Henri Becquerel was born in Paris on December 15, 1852
Discovery Antoine Henri Becquerel was born in Paris on December 15, 1852 Summit Environmental Technologies, Inc. Analytical Laboratories 3310 Win Street Cuyahoga Falls, Ohio 44223 Fax: 1-330-253-4489 Call
First Three Years After Project Proton Therapy Facility:
39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per