Calibration of low-level laser therapy equipment*
|
|
|
- Godfrey Perkins
- 10 years ago
- Views:
Transcription
1 Original Article ISSN Rev Bras Fisioter, São Carlos, v. 14, n. 4, p , July/Aug Revista Brasileira de Fisioterapia Calibration of low-level laser therapy equipment* Aferição dos equipamentos de laser de baixa intensidade Thiago Y. Fukuda 1,2, Julio F. Jesus 1, Marcio G. Santos 1, Claudio Cazarini Junior 2, Maury M. Tanji 3,4, Helio Plapler 2 Abstract Background: Despite the increase in the use of low-level laser therapy (LLLT), there is still a lack of consensus in the literature regarding how often the equipment must be calibrated. Objective: To evaluate the real average power of LLLT devices in the Greater São Paulo area. Methods: For the evaluation, a LaserCheck power meter designed to calibrate continuous equipment was used. The power meter was programmed with data related to the laser s wavelength to gauge the real average power being emitted. The LLLT devices were evaluated in two ways: first with the device cooled down and then with the device warmed up for 10 minutes. For each condition, three tests were performed. The laser probe was aligned with the power meter, which provided the real average power being emitted by the LLLT device. All of the data and information related to the laser application were collected with the use of a questionnaire filled in by the supervising therapists. Results: The 60 devices evaluated showed deficit in real average power in the cooled-down and warmedup condition. The statistical analysis (ANOVA) showed a significant decrease (p<0.05) in the real average power measured in relation to the manufacturer s average power. On average, the most common dose in the clinics was 4 J/cm², and the most desired effects were healing and anti-inflammatory effects. According to the World Association for Laser Therapy (WALT), 1 to 4 J of final energy are necessary to achieve these effects, however only one device was able to reach the recommended therapeutic window. Conclusion: The LLLT devices showed a deficit in real average power that emphasized a lack of order in the application of this tool. The present study also showed the need for periodical calibration of LLLT equipment and a better technical knowledge of the therapists involved. Key words: low-level laser therapy; gauging; calibration. Resumo Contextualização: A laserterapia de baixa intensidade (LBI) vem sendo cada vez mais utilizada, porém ainda não há consenso na literatura quanto ao tempo em que os equipamentos devem ser submetidos à aferição ou calibragem. Objetivo: Analisar a potência média real (PmR) dos equipamentos de LBI na região da Grande São Paulo. Métodos: Para análise dos equipamentos, utilizou-se um potenciômetro (Lasercheck), próprio para aferição de equipamentos contínuos, o qual foi programado com dados referentes ao comprimento de onda do laser a ser avaliado, obtendo-se assim a PmR emitida. Os equipamentos foram analisados de duas formas: uma, com o LBI desaquecido, e outra, após 10 minutos de uso (aquecido), sendo que três análises foram feitas para cada condição. A caneta emissora foi acoplada ao potenciômetro, o qual fornecia a PmR emitida pelo LBI. Todos os dados e informações referentes à aplicação do laser foram coletados por um questionário respondido pelos responsáveis. Resultados: Os 60 equipamentos avaliados mostraram déficit na PmR com os equipamentos desaquecidos e aquecidos. A análise estatística (ANOVA) mostrou diminuição significativa (P<0,05) da PmR aferida em relação à potência média do fabricante (PmF). Em média, a dose mais empregada nas clínicas foi de 4 J/cm², tendo os efeitos de cicatrização e anti-inflamatório como os mais desejados. Segundo a World Association for Laser Therapy (WALT), para atingir esse efeito, necessita-se de 1 a 4 J de energia final, sendo que apenas um dos 60 aparelhos conseguiria atingir a janela terapêutica preconizada. Conclusão: Os equipamentos de LBI apresentam um déficit acentuado na PmR, o que mostra uma desordem na utilização desse recurso. Neste estudo, observou-se a necessidade de aferição periódica dos aparelhos de LBI bem como melhor conhecimento técnico dos profissionais envolvidos. Palavras-chave: terapia laser de baixa intensidade; aferição; calibragem. Received: 16/02/2009 Revised: 29/06/2009X Accepted: 21/10/ Physical Therapy Sector, Irmandade Santa Casa de Misericórdia de São Paulo (ISCMSP), São Paulo (SP), Brazil 2 Experimental Surgery, Universidade Federal de São Paulo (UNIFESP), São Paulo (SP), Brazil 3 Laboratory of Investigation in Dermatology and Immunodeficiency (LIM56), Faculdade de Medicina, Universidade de São Paulo (FMUSP), São Paulo (SP), Brazil 4 Universidade do Grande ABC (UNIABC), Santo André (SP), Brazil Correspondence to: Thiago Yukio Fukuda, Setor de Fisioterapia, Santa Casa de São Paulo, Rua Dr. Cesário Motta JR., 112, Santa Cecília, CEP , São Paulo (SP), Brasil, [email protected]; [email protected] *Preliminary analysis of this research was presented and awarded as one of the top ten studies at the Poster Session of the 6th Conference of the Brazilian Society for Laser and Photodynamic Therapy in
2 Thiago Y. Fukuda, Julio F. Jesus, Marcio G. Santos, Claudio Cazarini Junior, Maury M. Tanji, Helio Plapler Introduction Low-level laser therapy (LLLT) has been investigated and used clinically for over 30 years, which justifies the increasing interest in the effects of laser and the significant amount of scientific publications in the literature 1-3. Therefore, LLLT equipment users should have knowledge of physical and biochemical principles to make better use of its resources. LLLT devices can have pulsed or continuous emission and the wavelengths most commonly used range from 630 nm to 1300 nm, thus including visible and invisible (infrared) light spectra 1. According to in vitro studies, the effects of light activate mechanisms of cellular metabolic control. Such mechanisms involve acceleration of the electron transport chain, increase in the synthesis of adenosine triphosphate (ATP) and decrease in intracellular ph. These reactions form the basis of the effects of LLLT 4-6. Laboratory studies, conducted mostly in animals, show strong evidence that LLLT has the ability to modulate inflammatory processes and relieve acute pain conditions triggered by lesions in soft tissues. This activity may occur through the decrease in nerve conduction, release of endogenous opioids, increase in angiogenesis and, consequently, increase in local microcirculation 7,8. It may also have inhibitory effects on the release of prostaglandins, cytokine levels, and cyclooxygenase (Cox2), and it may accelerate cell proliferation, collagen synthesis and tissue repair 9,10. However, several topics still need to be clarified and standardized for a safe and effective use. Some of these topics concern the type of LLLT, wavelength and dose, which may change the desired effect during the use of this equipment 1. According to Fukuda and Malfatti 11, many therapists and researchers have based their choice of laser dose on energy density or fluence (ΔE), but the wide variety of LLLT equipment may lead to differences in therapeutic results because the parameters vary according to manufacturer. This poses another problem in relation to the clinical reproducibility of the research, because when the same values are used in different equipments, there are differences in the total energy emitted to the tissue. Other important aspects in the variability of clinical outcomes and in the quality of the proposed treatments are the electrotechnical failures of LLLT equipment, as well as the amount of energy being delivered to the tissue. There is still little concern among therapists and manufacturers about the frequency with which the equipment needs to be calibrated Thus the aim of the present study was to analyze the real average power (RAP) of LLLT devices and therapeutic doses applied during the use of this tool in clinics, physical therapy practices, universities, outpatient units, and hospitals in the Greater São Paulo area. Methods Survey of LLLT equipment This was a calibration study of LLLT equipment used in clinics, physical therapy practices, universities, outpatient units, and hospitals with physical therapy service in the Greater São Paulo area. Initially, 261 locations were found through healthcare directories, Internet search engines and personal knowledge. All locations were contacted by phone or personal visit. Only 140 had LLLT equipment, and 52 of them could not be evaluated because the supervising therapist did not allow access or because the equipment was being repaired. At the 88 establishments where the tests were performed, there were 127 devices, 60 of which were selected for evaluation as they were continuous-wave devices. They were divided into seven brands (six national and one imported) with 11 different models. Thirteen devices had red laser (eight with 660nm wavelength and five with 670nm) and 47 infrared (ten with 808nm wavelength, 26 with 830nm, and 11 with 850nm; Figure 1). The study included LLLT devices in perfect condition, with a minimum of three months of use within the scope of physical therapy. Questionnaire For the data collection, we designed a questionnaire to be filled in before the evaluation with information about the equipment, such as brand, model, laser color, manufacturer average power (MAP), and wavelength. We also included questions about the device s main operator, such as main desired therapeutic effect, dose used to achieve this effect 15, information about their knowledge of the need of calibration, and whether the device had undergone maintenance and calibration. Information was collected from the user manuals, and when these were not available, we contacted the respective companies by telephone and/or . Procedures Before the data collection, a consent form was given to the therapists responsible for the equipment to inform them 304
3 Low-level laser equipment calibration of the absence of physical hazards to the equipment, physical and emotional stress, or expenses to the therapists. The RAP was gauged with a power meter (Lasercheck, Coherent, USA) calibrated by the manufacturer prior to the study and with ±5% accuracy. The power meter has a full wavelength range between 400 and 1064nm, and it is intended for the evaluation of continuous emission equipment. This tool has a protective lens used for equipment with a MAP above 10 mw. Below this level, the protective lens remained open according to manufacturer instructions. As a standard, all devices were evaluated in two ways: first, immediately after being switched on (cooled-down condition), then after 10 minutes of use (warmed-up condition). There were three tests with the warmed-up device and three with the cooled-down device, after which the mean for each condition was calculated. The lens of the laser probe was cleaned with disposable gauze and a swab moistened in hydrated ethyl alcohol, and the evaluation was performed in a room with the lowest possible lighting to avoid interference. To begin the evaluation, the dose was set in two ways as each device has a different form of emission (energy density and final energy). It must be noted, however, that this difference did not interfere with data collection because the evaluated parameter was the device s RAP compared to the MAP. Thus, after performing the pilot project, the standard dose of 4 J or 4 J/cm 2 was set depending on each device s form of emission. Next, the power meter was set to the device s wavelength, and the laser probe was aligned at a 90 angle. The laser beam was fired and the power meter was turned on simultaneously, which provided the RAP emitted by the LLLT device (Figure 2). These RAP values were also within the of the Brazilian National Standards Association (Associação Brasileira de Normas Técnicas [ABNT]), which allows an output variation of up to 20% Pilot project A prior study was conducted to standardize the dose to be gauged on all devices. Different doses of energy were tested (2, 4 and 6 J) on the same LLLT device with MAP equal to 100mW and 808nm wavelength. After data analysis, no significant difference was found in the MAP gauged in the described doses (p=0.2). Therefore, to gauge the RAP, the standard energy dose was set to 4 J or 4 J/cm 2 because it allows a radiant exposure time compatible with the performance of the entire gauging procedure. With this selected dose, the analysis would not exceed 15 minutes, thus avoiding 67 pulsed inconvenience to the therapists. To compare both evaluators, we included statistical analyses performed by means of the t-test for independent samples (p=0.80) and the intraclass correlation coefficient (ICC=0.81). According to the results, both evaluators were considered to be appropriately trained to perform the measurement. Data analysis 88 establishments 127 devices 60 continuous 1 imported 11 models 13 LLLT red 47 LLLT infrared 7 brands 6 national Figure 1. Diagram with the model of the study in relation to the survey of the equipments. Figure 2. Simulation of the analysis of a LLLT equipment and the consequent gauging of the real average power by the power meter. To better understand and employ the analysis, the devices were divided into four groups according to the time of use: group 1, 3 months to 2.5 years (n=18); group 2, 2.5 to 5 years (n=12); group 3, 5 to 7.5 years (n=10); and group 4, 7.5 to 10 years (n=20). After data collection, the statistical software GraphPad InStat was used for processing. First, the Kolmogorov-Smirnov (K-S) test was performed to verify 305
4 Thiago Y. Fukuda, Julio F. Jesus, Marcio G. Santos, Claudio Cazarini Junior, Maury M. Tanji, Helio Plapler data normality, with a significance level of 5%. We chose a non-parametric test for analysis of variance (ANOVA) with Friedman s post-test to compare MAP, RAP (warmed-up) and RAP (cooled-down). Results The average MAP was 30.7 mw, the RAP of the cooleddown devices was 18.1 mw, and the RAP of the warmed-up devices was 18.3 mw. The correlation between RAP and MAP, including all cooled-down devices, was extremely significant (p<0.001). The same happened when comparing MAP and RAP with the warmed-up devices (p<0.001; Figure 3). The percentage of RAP deficit in relation to MAP with the cooleddown devices was 64.3%, and 63.7% with the warmed-up devices (Table 1). The analysis results showed that among the 60 evaluated devices, only eight were within the set by the ABNT 12,14, therefore, 52 devices had RAP outside the standard range. The analyses according to time of use showed the following average deficits: group 1, 34.73%; group 2, 65%; group 3, 68.40%; and group 4, 90.70%. The ΔE most commonly used by the therapists was 4 J/cm² per point, and the most desired therapeutic effects were healing and antiinflammatory effects. It must be noted that, among the 18 devices in group 1, only two had undergone maintenance, and six were within ABNT. The questionnaire filled in by the therapists showed that 16 of them used the dose in ΔE and two used total energy, but none of the devices reached the desired therapeutic window. Of the 12 devices in group 2, only six had undergone maintenance, and only one was within ABNT. All therapists used the dose in ΔE, and none of the devices reached the therapeutic window. Among the ten devices in group 3, seven had undergone maintenance, but none were within ABNT. As noted in the previous group, all therapists used the dose in ΔE, and none of the devices reached the therapeutic window. For group 4, only ten of the 20 devices had undergone maintenance, and one was within ABNT ; the dose was in ΔE, and only one device reached the desired therapeutic window (Table 2). It is worth noting that the distribution into four groups was only used to analyze the data from the questionnaire; it was not maintained, therefore, for gauging the RAP. Discussion The present study aimed to show the actual condition of LLLT equipment used in clinics and physical therapy practices in the Greater São Paulo area, finding a lack of order in the use and maintenance of this therapeutic resource. The evidence shows that the devices are not within the of inspection agencies. The therapists also lack technical knowledge in choosing the ideal dose, type of laser and methods to achieve the real desired therapeutic effect. After prolonged use, LLLT devices are prone to degradation of the laser radiation structure, which decreases the power of radiation emitted by the devices 16. Therefore, annual calibration should be conducted in accordance with the specifications found in the manuals provided by manufacturers Table 1. Results of the evaluations performed in the three conditions. Type MAP N a RAP b RAP b Deficit Deficit R * I.R. ** (mw) (mw) C+ (mw) W++ C+ (%) W++ (%) (±4) 18.1 (±3.6) 18.3 (±3.6) 64.3 (±4.6) 63.7 (±4.6) * Red; ** infrared; a manufacturer average power; b real average power; + cooled-down; ++ warmed-up. 306 * Significant difference in relation to the MAP; ** Manufacturer average power; *** Real average power with cooled-down equipment; **** Real average power with warmed-up equipment. Figure 3. Average power (average±sem) of devices for the three evaluated conditions. Table 2. Information of the data contained in the questionnaire, distributing the equipments in four groups. Groups Time of Use n ABNT Group 1 3 mo.-2.5 yrs 18 Group yrs 12 Group yrs 10 Group yrs 20 6 within 1 within 0 within 1 within Therapeutic Window (WALT) 18 did not reach 12 did not reach 10 did not reach 1 reached / 19 did not reach
5 Low-level laser equipment calibration in order to achieve a truly functional application of this therapeutic method. Considering that electrical devices used by therapists may be mishandled over the years, it is very difficult to predict when and how the faults will occur, or even how to prevent them The analysis conducted in the present study shows the reality of LLLT equipment and highlights the fact that the minority of the analyzed devices had undergone calibration, further compromising the quality and effectiveness of treatment. ABNT regulations NBR IEC and IEC and ABNT Technical Report IEC recommend a variation no greater than ±20% in relation to the output power of the manufacturer 12-14, however the present study found that only eight of the 60 evaluated LLLT devices were within this standard. The evaluation of the MAP in relation to the RAP found a difference of 64%, reflecting the lack of order in the use of this tool. These findings explain why only one device reached the therapeutic window recommended by the World Association for Laser Therapy (WALT) 15. Regarding the application method, the vast majority of therapists used the dose based on ΔE. Of the 60 devices, only two provided the direct calculation of the final energy as a parameter, but due to their deficit in RAP, they failed to reach the desired therapeutic window. One device reached the treatment threshold as it was the only one that applied a high ΔE and the only one within ABNT, leading to the real desired therapeutic effect. In a study that conducted comparative simulations between national LLLT devices, the authors concluded that ΔE does not seem to be the parameter that best describes the dose to be used, as it can vary from device to device given that its parameters will be different when compared to other brands and LLLT models 11. This fact corroborates the results of the present analysis. Furthermore, it was observed that, even in devices with high RAP and within ABNT, it would not be possible to reach the recommended therapeutic window due to the therapist s lack of knowledge regarding the desired dose. In the present study, the therapists based their dose on ΔE only, and the main value was 4 J/cm², aiming to achieve anti-inflammatory and healing effects. According to the WALT, 1 to 4 J of final energy are needed to achieve these effects, thus the therapists would need to use a higher ΔE 15. This can be seen in the evaluation of two randomized controlled trials. The first trial applied LLLT in patients with osteoarthritis of the hand, with standard dose in ΔE equal to 3 J/cm². The results obtained in the study showed no significant differences between groups, indicating that the ΔE may have been too low, not reaching the therapeutic window 15,17. The second trial was conducted in patients with low back pain, divided into three groups: the first received LLLT combined with exercises; the second received LLLT only; and the third performed exercises only. According to the dose parameters provided in this trial, the parameters established by the WALT were met as were the therapeutic effects desired by the researchers. Thus, the two groups that received LLLT showed a significant difference in pain level when compared to the group that performed exercises only 15,18. The current lack of standardization in the calibration of devices used in physical therapy also includes therapeutic ultrasound (TUS). The results showed an excessively long period between calibrations, thus interfering in the therapeutic effect of the device. Some studies point to the need for periodic calibration of TUS devices 19,20. This emphasizes the importance of conducting periodical checks on the devices, giving the therapist a resource with greater reliability and reproducibility. It has also been stated that scientific research should follow the same path, i.e., calibrate equipment prior to a study The analysis of these studies and their results leads us to believe that, to make better and more efficient use of the beneficial effects of LLLT, it is extremely important that the average power of the device be within ABNT and that the physical therapist be able to dose the applied energy correctly Therefore, there is a need for annual or even biannual calibration of LLLT devices, investments in quality improvement by manufacturers, and more technical knowledge for therapists who use these devices. A limitation of the present study was that only laser devices with continuous emission were evaluated, as the power meter used was specific to continuous emission. For future studies, we propose that the devices be calibrated again to analyze a possible change in the current scenario and the inclusion of devices with pulsed emission. Conclusion LLLT devices used in clinics, physical therapy practices and hospitals located in the greater São Paulo area showed a marked deficit in average power, which shows a lack of order in the clinical use of this tool. Moreover, many of the devices were not within ABNT, and the applications may not be reaching the recommended therapeutic window, showing the need for periodic calibration. 307
6 Thiago Y. Fukuda, Julio F. Jesus, Marcio G. Santos, Claudio Cazarini Junior, Maury M. Tanji, Helio Plapler References 1. Kitchen SS, Partridge CJ. A review of low level laser therapy. Physiotherapy. 1991;77(3): Moshkovska T, Mayberry J. It is time to test low level laser therapy in Great Britain. Postgrad Med J. 2005;81(957): Gam AN, Thorsen H, Lonnberg F. The effect of low level laser therapy on musculoskeletal pain: a meta-analysis. Pain. 1993;52(1): Karu TI. Photobiology of low power laser effects. Health Phys. 1989;56(5): Karu TI, Pyatibrat LV, Afanasyeva NI. Cellular effects of low power laser therapy can be mediated by nitric oxide. Lasers Surg Med. 2005;36(4): Karu TI. Molecular mechanism of low-power lasertherapy. Lasers Life Sci. 1998;2: Bjordal JM, Johnson MI, Iversen V, Aimbire F, Lopes-Martins RAB. Photoradiation in acute pain: A systematic review of possible mechanisms of action and clinical effects in randomized placebo-controlled trials. Photomed Laser Surg. 2006;24(2): Bjordal JM, Couppé C, Chow RT, Tunér J, Ljunggren EA. A systematic review of low level laser therapy with location-specific doses for pain from chronic disorders. Aust J Physiother. 2003;49(2): Enwemeka CS, Parker JC, Dowdy DS, Harkness EE, Sanford LE, Woodruff LD. The efficacy of low-power lasers in tissue repair and pain control: a meta-analysis study. Photomed Laser Surg. 2004;22: Amaral AC, Parizotto NA, Salvini TF. Dose-dependency of low-energy HeNe laser effect in regeneration of skeletal muscle in mice. Lasers Med Sci. 2001;16(1): Fukuda TY, Malfatti CA. Analysis of low-level laser therapy doses in Brazilian equipment. Rev Bras Fisioter. 2008;12(1): Associação Brasileira de Normas e Técnicas. NBR IEC : Equipamento eletromédico: Prescrições particulares para a segurança de equipamento terapêutico e de diagnóstico a laser. Rio de Janeiro; 1997 (Pt. 2). 13. IEC. IEC : Safety of laser products: Equipment classification, requirements and user s guide. 1ª e 2ª ed. Switzerland; 2001 (Pt.1). 14. IEC. IEC TR : Safety of laser products: Guidelines for the safe use of medical laser equipment. Switzerland; 1999 (Pt. 8). 15. WALT World Association for Laser Therapy [homepage na internet]. acesso em 26 Jan 2009]. Disponível em: Bettiati M, Starck C, Pommies M, Broqua N, Gelly G, Avella M, et al. Gradual degradation in 980nm InGaAs/AlGaAs pump lasers. Material Science & Engineering B. 2002;91(92): Brosseau L, Wells G, Marchand S, Gaboury I, Stokes B, Morin M, et al. Randomized controlled trial on low level laser therapy (LLLT) in the treatment of osteoarthritis of the hand. Lasers Surg Med. 2005;36(3): Gur A, Karakoc M, Cevik R, Nas K, Sarac AJ. Efficacy of low power laser therapy and exercise on pain and functions in chronic low back pain. Lasers Surg Med. 2003;32(3): Artho PA, Thyne JG, Warring BP, Willis CD, Brismée JM, Latman NS. A calibration study of therapeutic ultrasound units. Phys Ther. 2002;82(3): Guirro R, Serrão F, Elias D, Bucalon AJ. Calibration of acoustic intensity of therapeutic ultrasound equipment in use in the city of Piracicaba. Rev Bras Fisioter. 1997;2(1): Guirro R, Santos SCB. A realidade da potência acústica emitida pelos equipamentos de ultrasom terapêutico: Uma revisão. Rev Fisioter Univ São Paulo. 1997;4(2):
An extraordinary claim requires extraordinary proof. Marcello Truzzi
How Not To Promote Laser Therapy By Jan Tunér, DDS Lars Hode, DrSci (Swedish Laser Medical Society) Peter A Jenkins, MBA (Australian Medical Laser Association) We can talk to the cells, but we must learn
Lasers and Specturbility of Light Scattering Liquids - A Review
Artigo Original Revista Brasileira de Física Médica.2011;5(1):57-62. Reproducibility of radiant energy measurements inside light scattering liquids Reprodutibilidade de medidas de energia radiante dentro
The use of laser therapy in small animal patients has
PEER REVIEWED RecoVeRy & Rehab What It Is, How It Works, & When It Benefits Patients David Dycus, DVM, MS, Diplomate ACVS (Small Animal) Regional Institute for Veterinary Emergencies and Referrals (RIVER),
WHAT YOU SHOULD KNOW BEFORE YOU BUY A THERAPEUTIC LASER
WHAT YOU SHOULD KNOW BEFORE YOU BUY A THERAPEUTIC LASER Laurie Edge- Hughes, BScPT, MAnimSt(Animal Physio), CAFCI, CCRT Faculty, Canine Rehab Institute, Wellington, FL Co- Owner, The Canine Fitness Centre
BIOPHYSICAL AND PHYSIOLOGICAL MECHANISMS OF LOW ENERGY LASERS INTERACTIONS WITH LIVING CELLS AND THEIR IMPLICATIONS IN PAIN TREATMENT
ANALELE ŞTIINŢIFICE ALE UNIVERSITĂŢII AL. I. CUZA IAŞI Tomul I, s. Biofizică, Fizică medicală şi Fizica mediului 2005 BIOPHYSICAL AND PHYSIOLOGICAL MECHANISMS OF LOW ENERGY LASERS INTERACTIONS WITH LIVING
Endnote Web tutorial for BJCVS/RBCCV
Oliveira MAB, SPECIAL et al. - Endnote ARTICLE Web tutorial for BJCVS/RBCCV Endnote Web tutorial for BJCVS/RBCCV Tutorial do Endnote Web para o BJCVS/RBCCV Marcos Aurélio Barboza de Oliveira 1, MD, PhD;
Mark Silverman DVM, MS. Sporthorse Veterinary Services
Mark Silverman DVM, MS Sporthorse Veterinary Services Light has long been recognized as being crucial to normal human function and health. Our ability to remain alert and active, along with some of the
DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 DEVELOPMENT AND IMPLEMENTATION
Healing your pain changing your life.
Healing your pain changing your life. About K-Laser What is Laser Therapy? Laser Therapy, or photobiomodulation, is the use of specific wavelengths of light (red and nearinfrared) to create therapeutic
THE LABEL EFFECTS ON THE CONSUMER INSTANT COFFEE PERCEPTION
THE LABEL EFFECTS ON THE CONSUMER INSTANT COFFEE PERCEPTION DELIZA, R. 1 ; MacFIE, H.J.H. 2 e HEDDERLEY, D. 3 1 EMBRAPA - Food Technology, Av. das Américas, 29501 CEP 23020-470 Rio de Janeiro-Brazil; ;
Example of SOP for IR Laser
Example of SOP for IR Laser Standard Operating Procedures (SOP) for Class 4 HPD semiconductor Laser Manufacturer HPD Building & Room Number EH&S 101 Model Number HPD-1/100/975 Laser Class 4 Serial Number
Class IV Laser Therapy Interventional and case reports confirm positive therapeutic outcomes in multiple clinical indications
Published by: LiteCure, LLC. 2009 Class IV Laser Therapy Interventional and case reports confirm positive therapeutic outcomes in multiple clinical indications Brian A. Pryor, PhD ABSTRACT Tissue that
CLASS IV THERAPY LASERS
CLASS IV THERAPY LASERS 80% More than 80% of your patients can benefit from K-Laser therapy. 970 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - PEAK OF WATER 905 - - - - - - - - - - - - - -
QUALITY KNOWLEDGE INTEGRATION: A BRAZILIAN COMPARISON ANALYSIS
QUALITY KNOWLEDGE INTEGRATION: A BRAZILIAN COMPARISON ANALYSIS Úrsula Maruyama [email protected] CEFET/RJ, Departamento de Ensino e Administração (DEPEA) Av. Maracanã 229, Maracanã CEP 20.271-110
Evaluation of the image quality in computed tomography: different phantoms
Artigo Original Revista Brasileira de Física Médica.2011;5(1):67-72. Evaluation of the image quality in computed tomography: different phantoms Avaliação da qualidade de imagem na tomografia computadorizada:
QUESTIONÁRIOS DE AVALIAÇÃO: QUE INFORMAÇÕES ELES REALMENTE NOS FORNECEM?
QUESTIONÁRIOS DE AVALIAÇÃO: QUE INFORMAÇÕES ELES REALMENTE NOS FORNECEM? Grizendi, J. C. M [email protected] Universidade Estácio de Sá Av. Presidente João Goulart, 600 - Cruzeiro do Sul Juiz de Fora
Can Noise in Dental Clinic Produce Hearing Loss?
Original Article Can Noise in Dental Clinic Produce Hearing Loss? Ruído em Consultórios Odontológicos pode Produzir Perda Auditiva? Edmir Américo Lourenço*, Janaína Medina da Rocha Berto**, Sávio Butignolli
Rice University Laser Safety Manual
Rice University Laser Safety Manual Environmental Health and Safety MS 123 P.O. Box 1892 Houston, TX 77251-1892 December 2012 RICE UNIVERSITY 1 Introduction This manual is intended to provide the basic
SUITABILITY OF RELATIVE HUMIDITY AS AN ESTIMATOR OF LEAF WETNESS DURATION
SUITABILITY OF RELATIVE HUMIDITY AS AN ESTIMATOR OF LEAF WETNESS DURATION PAULO C. SENTELHAS 1, ANNA DALLA MARTA 2, SIMONE ORLANDINI 2, EDUARDO A. SANTOS 3, TERRY J. GILLESPIE 3, MARK L. GLEASON 4 1 Departamento
Continuing Medical Education Program: a credit system evaluation
Original Article Continuing Medical Education Program: a credit system evaluation Programa de Educação Médica Continuada: avaliação de um sistema de créditos Claudio Schvartsman 1, Milton Glezer 2, Renato
PowerLight LED Light Therapy. The FUTURE of corrective skin
PowerLight LED Light Therapy The FUTURE of corrective skin care TODAY LED facial treatments Effective when used with correct protocols Non thermal stimulation of collagen Increases circulation and lymphatic
Making our pets comfortable. A modern approach to pain and analgesia.
Making our pets comfortable. A modern approach to pain and analgesia. What is pain? Pain is an unpleasant sensory and emotional experience with awareness by an animal to damage or potential damage to its
HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL 1
Harvesting and Wood Transport Planning with SNAP III... 831 HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL
New Methods for Guidance of Light-Based Treatments Using Objective Melanin Measurements
New Methods for Guidance of Light-Based Treatments Using Objective Melanin Measurements Sean Doherty, MD, 1 Richard Cohen, PhD, 2 Felicia Whitney, RN, 2 Oksana Bradley, MD, 2 (Ukraine) Brooke Seckel, MD
Independent Clinics Lasers and Intense Pulse Light Source Treatments
Independent Clinics Lasers and Intense Pulse Light Source Treatments Registration & Inspection Unit May 2008 (amended October 2013) Department of Health and Social Care Rhyenn Slaynt as Kiarail y Theay
Corporate Medical Policy
Corporate Medical Policy Electrical Stimulation for the Treatment of Arthritis File Name: Origination: Last CAP Review: Next CAP Review: Last Review: electrical_stimulation_for_the_treatment_of_arthritis
Benefits of Cold Laser Therapy in Pain Management CURE SOME, COMFORT ALL. Disclaimer, Conflict of Interest. Objectives
Benefits of Cold Laser Therapy in Pain Management Disclaimer, Conflict of Interest A Guide for Health Care Professionals Jason Reed, MD, Medical Director 2010 Partly Adapted/Revised from Meditech-Bioflex
IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 IMPROVEMENT OF GAMMA
Original Article INTRODUCTION. Abstract
Original Article DOI: 10.17354/ijss/2015/399 Comparative Evaluation between the Effects of Low-Intensity Laser Therapy and Transcutaneous Electric Nerve Stimulation on Temporomandibular Joint Disorders
TEMPOROMANDIBULAR DISORDER (TMD) is the communal
Photomedicine and Laser Surgery Volume 24, Number 1, 06 Mary Ann Liebert, Inc. Pp. 45 49 Management of Mouth Opening in Patients with Temporomandibular Disorders through Low-Level Laser Therapy and Transcutaneous
Laser Classification. Laser Classes
LASER SAFETY Laser Classification It is the responsibility of the laser manufacturer to provide the correct classification of a laser product. This classification is made on the basis of a combination
How To Test Water Penetration On Concrete Block And Mortar
WATER PENETRATION TEST ON CONCRETE BLOCK MASONRY Vilató, Rolando Ramirez 1 1 PhD, Prof., Mackenzie Presbyterian University, School of Engineering, São Paulo, [email protected] This paper considers
Laser therapy for the treatment of arthritic knees: a clinical study F. Kahn a, R. Liboro a and F. Saraga* a
Laser therapy for the treatment of arthritic knees: a clinical study F. Kahn a, R. Liboro a and F. Saraga* a a Meditech Laser Rehabilitation Centre, 415 Horner Ave, Toronto, ON, Canada, M8W4W3 ABSTRACT
ENERGETIC EFFICIENCY OF AN AGRICULTURAL TRACTOR IN FUNCTION OF TIRE INFLATION PRESSURE
ENERGETIC EFFICIENCY OF AN AGRICULTURAL TRACTOR IN FUNCTION OF TIRE INFLATION PRESSURE LEONARDO DE A. MONTEIRO 1, DANIEL ALBIERO 2, KLEBER P. LANÇAS 3, ANDRÉ V. BUENO 4, FABRICIO C. MASIERO 5 ABSTRACT:
Temperature Increase in the. Human Eye When Subjected to a. Laser Source
Validation of a Model to Predict Temperature Increase in the Human Eye When Subjected to a Laser Source Corinna Sue Thompson, Antonio Campo, PhD., University of Vermont, Undergraduate Research Endeavors
Lasers and Pain Treatment
Oficiální orgán Společnosti pro využití laseru v medicíně ČLS JEP of the Czech Society for Official paper the Use of Laser in Medicine Vydáváno s oficiální odbornou podporou EMLA Edited under official
MLS Laser Therapy: demonstrated scientifically.
MLS Laser Therapy: demonstrated scientifically. ASAlaser: valid innovation. We have increased our experience day after day since 1983, successfully dedicating ourselves to therapy solutions in Pain Management
QUALITY OF LIFE IN PATIENTS WITH NARCOLEPSY
Arq Neuropsiquiatr 2008;66(2-A):163-167 QUALITY OF LIFE IN PATIENTS WITH NARCOLEPSY A WHOQOL-Bref study Heloísa Rovere 1, Sueli Rossini 2, Rubens Reimão 3 Abstract Objective: To evaluate the perception
Development of a parallel plate ion chamber for radiation protection level
Artigo Original Revista Brasileira de Física Médica.2011;5(1):9-14. Development of a parallel plate ion chamber for radiation protection level Desenvolvimento de uma câmara de ionização de placas paralelas
Laser Safety Self-Audit Checklist
Laser Safety Self-Audit Checklist www.uic.edu/depts/envh [email protected] Ph: 312-996-7411 24-hour Emergency: 312-996-SAFE (7233) Building Room Principal Investigator Date Audit Performed by A. Administrative
ArcHC_3D research case studies (FCT:PTDC/AUR/66476/2006) Casos de estudo do projecto ArcHC_3D (FCT:PTDC/AUR/66476/2006)
ArcHC_3D research case studies (FCT:PTDC/AUR/66476/2006) Casos de estudo do projecto ArcHC_3D (FCT:PTDC/AUR/66476/2006) 1 Casa de Valflores - Loures 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Capela de S. Frutuoso
BRADEN SCALE: THE RESPONSIBILITY OF NURSES IN THE PROMOTION OF PATIENT SAFETY 1
BRADEN SCALE: THE RESPONSIBILITY OF NURSES IN THE PROMOTION OF PATIENT SAFETY 1 FERREIRA, Emanuelli Mancio 2 ; MAGNAGO, Tânia Solange Bosi de Souza 3 ; CERON, Marinez Diniz 4 ; PASA, Thiana Sebben 5 ;
StreamServe Project Guide and Framework Versão 1.4 / Maio-2013
StreamServe Project Guide and Framework Versão 1.4 / Maio-2013 PAF011 Version control V 1.1 Revisão do documento 09-04-2012 Rui Miguel (DSI/DIS) V 1.2 Alteração da forma de obtenção do PIARQT012 29-08-2012
Low-level laser therapy (LLLT) stimulates cell
The American Journal of Cosmetic Surgery Vol. 26, No. 3, 2009 1 6 ORIGINAL SCIENTIFIC RESEARCH Low-Level Laser Therapy Effectiveness for Reducing Pain After Breast Augmentation Robert F. Jackson, MD; Gregory
Marcio J. Santos, P.T., Ph.D.
Marcio J. Santos, P.T., Ph.D. Postdoctoral research associcate University of Illinois at Chicago Department of Physical Therapy 1919 W. Taylor m/c 898 Chicago, Il 60612 Phone: (312) 355 0902 Fax: (312)
Low Level Laser Therapy (LLLT) for Orofacial Pain
Review Article Low Level Laser Therapy (LLLT) for Orofacial Pain Seyyed Amir Seyyedi 1, Pooya Olyaee 2, Zohreh Dalirsani 3, Farnaz Falaki 3 1Department of Oral Medicine, Faculty of Dentistry, Urmia University
Hydrocarbon reservoir modeling: comparison between theoretical and real petrophysical properties from the Namorado Field (Brazil) case study.
ANNUAL MEETING MASTER OF PETROLEUM ENGINEERING Hydrocarbon reservoir modeling: comparison between theoretical and real petrophysical properties from the Namorado Field (Brazil) case study. Marcos Deguti
Inovando sistemas com arquiteturas elásticas
Inovando sistemas com arquiteturas elásticas Renato Bognar Principal System Engineer 1 Agenda Quais são os desafios de construir ua aplicação? Quais os pontos de atenção? Vai construir Apps móveis? Desfazendo
LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes
LBS-300 Beam Sampler for C-mount Cameras P/N SP90183, SP90184, SP90185 and SP90186 YAG Focal Spot Analysis Adapter P/N SP90187, SP90188, SP90189, SP90190, SP90191 User Notes Ophir-Spiricon Inc. 60 West
HYBRID INTELLIGENT SUITE FOR DECISION SUPPORT IN SUGARCANE HARVEST
HYBRID INTELLIGENT SUITE FOR DECISION SUPPORT IN SUGARCANE HARVEST FLÁVIO ROSENDO DA SILVA OLIVEIRA 1 DIOGO FERREIRA PACHECO 2 FERNANDO BUARQUE DE LIMA NETO 3 ABSTRACT: This paper presents a hybrid approach
Scientific fraud in laser therapy
Scientific fraud in laser therapy By Lars Hode & Jan Tunér Swedish Laser Medical Society A number of reports on scientific fraud have recently been highlighted in the international press. The main stories
Do home-use hair removal lasers & intense light devices deliver what they promise and are they safe?
Do home-use hair removal lasers & intense light devices deliver what they promise and are they safe? Godfrey Town, Innovation Scholar, Global Academy, University of Wales; Swansea Metropolitan University,
hire the lasers Light therapy proves valuable in many veterinary and rehabilitation clinics 01 Barbara Dobbins, August 2013
hire the lasers Light therapy proves valuable in many veterinary and rehabilitation clinics by barbara dobbins reprinted with permission. originally appeared in whole dog journal, august 2013 One look
CRM: customer relationship management: o revolucionário marketing de relacionamento com o cliente P
CRM: customer relationship management: o revolucionário marketing de relacionamento com o cliente Download: CRM: customer relationship management: o revolucionário marketing de relacionamento com o cliente
User Manual Table of Contents. Instructions for Using the LaserTouchOne Charging the Unit... 5 Holding the Device... 5 Treatment Steps...
User Manual User Manual Table of Contents Introduction The LaserTouchOne... 3 Device Features and Controls (Photo of device)... 4 Product Contents... 4 Laser Safety... 4 Electrical Safety... 5 Instructions
Introduction to Low Level Light Therapy (LLLT) for pain relief. James D Carroll. THOR Photomedicine Ltd (UK) www.thorlaser.com
C M Y CM MY CY CMY K THOR logo 450 WIDE.pdf 26/9/07 22:55:23 Introduction to Low Level Light Therapy (LLLT) for pain relief James D Carroll THOR Photomedicine Ltd (UK) www.thorlaser.com Low Level Laser
IN-SERVICE NURSING EDUCATION DELIVERED BY VIDEOCONFERENCE
GODOY, S.; MENDES, I.A.C.; HAYASHIDA, M., NOGUEIRA, M.S., ALVES, L.M.M. Inservice nursing education delivered by videoconference. Journal of Telemedicine and Telecare, v.10, n.5, p.303-305, 2004. IN-SERVICE
PPTA Payer Summit Medical Review Challenges and Red Flags in Documentation. CPT Coding for Physical Therapy Services 97000 Series and Beyond
PPTA Payer Summit Medical Review Challenges and Red Flags in Documentation November 19, 2014 Presented by Sandra McCuen, PT PPTA Reimbursement Specialist [email protected] 717.623.6135 CPT Coding
Lasers and Pain Treatment Dr. Kevin Moore Department of Anaesthesia The Royal Oldham Hospital, UK
Used by permission of the Czech Society for the Use of Laser in Medicine, www.laserpartner.org Lasers and Pain Treatment Dr. Kevin Moore Department of Anaesthesia The Royal Oldham Hospital, UK Summary
HIGH INTENSITY LASER A REVOLUTION IN THERAPEUTIC LASER TECHNOLOGY
HIGH INTENSITY LASER A REVOLUTION IN THERAPEUTIC LASER TECHNOLOGY [email protected] www.btlnet.com All rights reserved. Although every care has been taken to provide accurate and up-to-date information,
Low back pain. Quick reference guide. Issue date: May 2009. Early management of persistent non-specific low back pain
Issue date: May 2009 Low back pain Early management of persistent non-specific low back pain Developed by the National Collaborating Centre for Primary Care About this booklet This is a quick reference
Dose Thresholds and Effect Mechanisms for Pain Management with LASER Phototherapy
Dose Thresholds and Effect Mechanisms for Pain Management with LASER Phototherapy E.-L. Laakso School of Physiotherapy and Exercise Science, Griffith University, Gold Coast, 4222, Australia Summary Arguably,
ORIGINAL ARTICLE. Ernesto Cesar Pinto Leal-Junior & Douglas Scott Johnson & Anita Saltmarche & Timothy Demchak
DOI 10.1007/s10103-014-1592-6 ORIGINAL ARTICLE Adjunctive use of combination of super-pulsed laser and light-emitting diodes phototherapy on nonspecific knee pain: double-blinded randomized placebo-controlled
Home advantage in football in Brazil: differences between teams and the effects of distance traveled
: differences between teams and the effects of distance traveled Vantagem em casa no futebol no Brasil: diferenças entre clubes e efeitos de distância de viagem Pollard R 1, Silva CD 2, Medeiros NC 2 1-
The effective treatment of acne vulgaris by a high-intensity, narrow band 405 420 nm light source
J Cosmetic & Laser Ther 2003; 5: 111 116 # J Cosmetic & Laser Ther. All rights reserved ISSN 1476-4172 DOI: 10.1080/14764170310001276 111 The effective treatment of acne vulgaris by a high-intensity, narrow
Linear programming applied to healthcare problems
105 Linear programming applied to healthcare problems Programação linear aplicada a problemas da área de saúde* Frederico Rafael Moreira 1 ORIGINAL ARTICLE ABSTRACT Objective and Method: To present a mathematical
The current situation of sanitary landfills in Brazil and the application of economic models
The current situation of sanitary landfills in Brazil and the application of economic models Raul Oliveira Neto, Ph.D. Mining technology course Federal University of Pampa - UNIPAMPA Avenue Anunciação
Prof.:Renata P. de Brito. July 2013
Sustainability in the Brazilian Business Context Prof.:Renata P. de Brito July 2013 Development Model Our Common Future -1987 Gro Brundtland Report Sustainable Development: Sustainability "Development
Exacta ISSN: 1678-5428 [email protected] Universidade Nove de Julho Brasil
Exacta ISSN: 1678-5428 [email protected] Universidade Nove de Julho Brasil Paschoalin Filho, João Alexandre; Guerner Dias, Antonio José; Brum de Lima Duarte, Eric The use of polymer slurry as an economic
Experimental model for low level laser therapy on ischemic random skin flap in rats 1
Experimental model for low level laser therapy on ischemic random skin flap in rats 1 Modelo experimental para laserterapia de baixa intensidade em retalho cutâneo randômico isquêmico em ratos 13 - TECHNICAL
Infrared Viewers. Manual
Infrared Viewers Manual Contents Introduction 3 How it works 3 IR viewer in comparison with a CCD camera 4 Visualization of infrared laser beam in mid-air 4 Power Density 5 Spectral sensitivity 6 Operation
Articles. Funding None.
Efficacy of low-level laser therapy in the management of neck pain: a systematic review and meta-analysis of randomised placebo or active-treatment controlled trials Roberta T Chow, Mark I Johnson, Rodrigo
Mellen Center for Multiple Sclerosis
Mellen Center Cleveland Clinic Marie Namey, RN, MSN, MSCN Mellen Center Cleveland Clinic Cleveland, OH Home of. Mellen Center for Multiple Sclerosis Mellen Center Mission The Mellen Center remains committed
Technological platforms
Advitech Advitech is a life sciences & technology company Its mission is to discover and commercialize proprietary and evidence-based natural health products Focus on milk, whey and bovine colostrum R&D,
adj., departing from the norm, not concentric, utilizing negative resistance for better client outcomes
Why Eccentrics? What is it? Eccentric adj., departing from the norm, not concentric, utilizing negative resistance for better client outcomes Eccentrics is a type of muscle contraction that occurs as the
Chapter 4 Physiological Therapeutics. 3 Therapeutic Ultrasound
Chapter 4 Physiological Therapeutics 3 Therapeutic Ultrasound Therapeutic Ultrasound PHYSIOLOGIC EFFECTS OF ULTRASOUND 1. Increased collagen extensibility 5. Increased nerve conduction velocities 2. Increased
COLD LASERS IN PAIN MANAGEMENT
COLD LASERS IN PAIN MANAGEMENT Low energy laser therapy has been shown at appropriate dosimetry, wavelength, duration, and site-specific application to reduce tissue pain/ tenderness, normalize circulation
Abstract Number: 002-0427. Critical Issues about the Theory Of Constraints Thinking Process A. Theoretical and Practical Approach
Abstract Number: 002-0427 Critical Issues about the Theory Of Constraints Thinking Process A Theoretical and Practical Approach Second World Conference on POM and 15th Annual POM Conference, Cancun, Mexico,
Sharon Gorman, PT, DPTSc, GCS October 22, 2011 Acute Care: The Front Line UCSF / SFSU & Kaiser Permanente Northern California Community Benefits
Sharon Gorman, PT, DPTSc, GCS October 22, 2011 Acute Care: The Front Line UCSF / SFSU & Kaiser Permanente Northern California Community Benefits Division Where have we been? Changes over time Practice