1 What Cytometry Can Do for You: The Pros and Cons of Image and Flow Cytometry Webinar 31 October 2012 [0:00:10] Slide 1 Hello everyone and welcome to this Science/AAAS webinar. I m Sean Sanders, editor for custom publishing at Science. Today s webinar will be about all things cytometric. Flow cytometry has been used for decades as the method of choice to gather valuable data on an individual cell level. More recently, image cytometry technologies including high content analysis have emerged to enable more detailed cellular and subcellular analysis, particularly of adherent cell types. Image cytometry has become less expensive and more accessible, giving experts and non experts alike fast and easy access to detailed data about their cells. In this webinar, our esteemed speakers will share their views on the pros and cons of image and flow cytometry and how these technologies might be applied to their own research. It s my pleasure to introduce my studio guests for today. They are Dr. Bill Telford from the NIH in Bethesda, Maryland, Dr. Paul Gokhale from the University of Sheffield in the United Kingdom, and Dr. Liz Roquemore from GE Healthcare based in Cardiff in the UK. A very warm welcome to you all. Thanks for being here. Dr. Bill Telford: Dr. Liz Roquemore: Dr. Paul Gokhale: Thank you, Sean. Thank you. Pleasure to be here. You will notice that Dr. Gokhale is not actually with us in the studio today. Due to the big storm on the east coast of the US, his flight over was unfortunately cancelled, but we are very pleased that he could join us through an audio linkup. So welcome to you, Dr. Gokhale. It s good to be here, Sean. Thank you. 1

2 Before we get started, I have some important information for our audience. Note that you can resize or hide any of the windows in your viewing console. The widgets at the bottom of the console control what you see. Click on these to see the speaker bios additional information about technologies related to today's discussion, or to download a PDF of the slides. Each of our speakers will talk briefly about their work. After which we will have a Q&A session during which our guests will address the questions submitted by our live online viewers. So if you're joining us live, start thinking about some questions right now and submit them at any time by typing them into the box on the bottom left of your viewing console and clicking the submit button. If you can't see this box, just click the red Q&A widget at the bottom of the screen. Please remember to keep your questions short and concise. This will give them the best chance of being put to our panel. You can also log in to your Facebook, Twitter, or LinkedIn accounts during the webinar to post updates or send tweets about the event, just click the relevant widgets at the bottom of the screen. For tweets, you can add the hash tag, #sciencewebinar. Finally, thank you to GE Healthcare for sponsoring today's webinar. Slide 2 Dr. Bill Telford: Slide 3 Now, I'd like to introduce our first speaker, Dr. Bill Telford. Dr. Telford is currently the director of the flow cytometry core laboratory in the NCI Experimental Transplantation and Immunology Branch at the NIH. His main research interests include instrument development, particularly in the area of novel solid state laser integration into flow cytometers, flow cytometric stem cell detection and characterization, and functional characterization of early apoptosis by flow and image cytometry. Welcome, Dr. Telford. Thank you, Sean, for having me here today. Okay. So what I d like to do in the next few minutes is give a brief overview of just a review of flow cytometry for those who need it and also to talk a little bit about image cytometry, which is proven to be an important adjunct technique to flow cytometry. Slide 4 2

3 I think many people in our audience are aware of what goes into a traditional flow cytometer. You need first a biological sample, usually labeled with some sort of fluorescent marker. You need to move these cells in a linear stream through a focused light source, usually a laser, it doesn t have to be but it usually is. You then need to detect the fluorescence in a PMT with a filter and then you need to process that data out the other side. Slide 5 [00:05:01] Slide 6 Slide 7 Slide 8 So this is what the data ends up looking like and when we do flow cytometry, we re generally analyzing three major applications and I ll go through them one at a time. The first is fluorescent probes for cell surface or intracellular labeling usually using an antibody that s tagged with a fluorescent marker like fluorescein or PE. Here, we ve labeled mouse T cells for fast expression and CD8 expression using two fluorescent probes, FITC and PE. What I want to point out and I think this is obvious for many people who are familiar with flow but remember that all of this data is inter relational so that we can look at the cells that are CD8 positive in this star plot over here on the right and look at the Fas expression within those cells and also the Fas expression in the CD8 negative cells as well. Now this technology has become quite powerful. We can do 14, 26 parameter flow cytometry and the inter relational nature of it gives us tremendous power. Here, we re looking at lymphoctyes and monocytes and macrophages based on their forward and side scatter and we re able to look at the expression of numerous markers based on the ability to gate on one parameter and view another. All parameters are inter relational relative to the other parameters. Fluorescent proteins is probably the second most popular application for flow cytometry now. This started with GFP back in the mid 90s. There are now literally hundreds of fluorescent proteins that span the visible spectrum and this is another major application for flow cytometric analysis. Finally, we have a large variety of physiological fluorescent probes that can measure the physiological state of a cell. We can look at the mitochondrial membrane potential, the plasma membrane potential. We can visualize mitochondria, look at DNA content. The 3

4 number of probes available of this type literally are in the hundreds right now and another common application for flow cytometry. Slide 9 However, traditional flow has its limits. We re basically analyzing cells that are flying through a laser beam at about 10 meters/second and all we re generally able to do in that format is measure the total fluorescence of the cell as it passes through. We re not getting any subcellular data. We re able to look at the area, the width of the pulse, the height of the pulse here as the cell goes through. But generally the information we re getting per cell is very, very limited. It s just total fluorescence. This is where image cytometry comes in. Image cytometry has allowed us to actually visualize the cell and collect cytometric data at the same time. So we re collecting the same data that we would with a flow cytometry, but we re also able to get a picture, an image of the cell and various aspects of that image can become parameters just like fluorescence or forward scatter is a parameter. Another key characteristic of image cytometry is that the image and the cytometric data are correlated with each other. So you can look at a dot plot in a histogram and go back and see what that cell actually is. Slide 10 Slide 11 So this is what a typical image cytometer looks like. This is a Compycyte icys instrument. You can see a lot of the components are pretty much exactly the same as you d find on a flow cytometer. You ve got lasers, you ve got mirrors, dichroics, and filters. You ve got photomultiplier tubes for detecting that data. But rather than the data coming from cells in a stream, it s now coming from cells sitting on a slide or on a horizontal surface. The cells are static, they re not moving. Now we often refer to this type of analysis as high content analysis because we re getting more than just raw fluorescent data here. We re actually getting data pertaining to the interior structure of the cell. Another key aspect of image cytometry is that like flow cytometry it is event thresholded or event based. On a flow cytometer, a cell will go flying for the beam. You ll use a characteristic like forward scatter to determine whether an object is there or not and then you ll take fluorescence and other measurements off that cell. We do exactly 4

5 the same thing in image cytometry. We set a parameter, it can be scatter, it can be fluorescence to identify whether cells are present. The instrument will then segment or select that object as it has in these scans here and then the cytometric data is taken. So we have both image data, cytometric data and the two are correlated with one another. Slide 12 It can be really quite powerful and the imagery can be quite dramatic. These are mouse lymphoma cells that have been induced to undergo apoptosis using a drug, using camptothecin. We ve labeled them with three apoptotic markers, a caspase substrate labeled with fluorescein, Annexin V labeled with a red dye, Alexa Fluor 647 plus 7 AAD. We re able to get the image data and we re able to see the presence of Annexin V labeling or caspase activation in the cells. These images here are not photographs. They are actual PMT reconstructed data and this PMT reconstructed data, this image data is the data by which the cytometric data is then derived. You can see the cytometric data on the right hand side of the slide looking very much like you would see off of a traditional flow cytometry but you ve got the image data as well. Slid 13 [00:10:24] Slide 14 Now you can do a lot of very powerful things with this. We can look at the subcellular structure of a cell and express that structure as a fluorescent parameter. Here, we ve taken U2OS cells. This is a human osteosarcoma line that s been transfected with a GFP LC3 fusion protein. The LC3 protein during autophagy segregates into these very tight, punctate regions. It s normally distributed evenly throughout the cell. We can measure those punctate regions using image cytometry and express them as a parameter. So we re not just taking total fluorescence here, we re looking at the fluorescence structure of something inside the cell. Now there are a number of good image cytometer systems out there. This is a sampling of them. I ve shown you some data on the Compycyte icys. The BD Biosciences Pathway systems, the 450 and the 850 work very well in this regard. The GE IN Cell platforms, the 2000 and the 6000, the Cellomics and PerkinElmer systems as well. These are generally very sophisticated systems. They can collect many fluorescent parameters. They often have multiple excitation 5

6 lines either lasers or lamps. They archive the data and the data can be brought up later and correlated with the cytometric data. Slide 15 Slide 16 There is one stream based image cytometer system available. It s the Amnis ImageStream. Rather than capturing images with cells that are sitting on a slide or on a flat surface, this instrument actually captures the images in the stream. Here we have I believe these Raji cells that have been again labeled with multiple apoptotic markers, caspase, Annexin V and DNA and you can see brightfield and also the individual fluorescent data and this data can also be expressed cytometrically as dot plots and histograms. Finally, there are a number of special purpose image cytometers out there currently. The GE Life Sciences Cytell is one of these, the TTL Labtech Accumen, the Cytellect Celigo, the Beckman Coulter Vi cell. These instruments generally do not archive the data, but they use the data, the image data as the basis for the cytometric data they display. These systems are often designed for specialty or specific applications like viability apoptosis or basic immunolabeling. The imagery is generally not archived but sometimes intracellular features can still be quantified and these systems also distinguish themselves in generally being of lower cost than the more sophisticated systems. So they re very good for specialty applications. They can also have very rapid throughput for applications where you need to run large numbers of samples very quickly and Dr. Gokhale will talk about that in a few minutes. So I will conclude my talk at that point. Slide 17 Dr. Bill Telford: Great. And give it back to Sean. Thank you so much, Dr. Telford. Our second speaker for today is Dr. Paul Gokhale. Dr. Gokhale is currently a research associate at the University of Sheffield. Over the past seven years, he has been involved in the international stem cell initiative first to develop standards to characterize human embryonic stem cells and second to identify common genetic changes that occur when human ES cells are cultured in vitro. More recently, Dr. Gokhale has been involved in projects to use real time imaging and image analysis to understand 6

7 how human embryonic stem cells self renew and how early differentiation decisions are controlled. Welcome, Dr. Gokhale. Slide 18 Dr. Paul Gokhale: Slide 19 Slide 20 [00:15:04] Okay. Thank you very much, Sean. So in my short presentation, I m going to take you through a brief overview of high content analysis, briefly talk about why it fits in with other cytological techniques that people are familiar with, some of the types of assays we commonly use with these systems, some factors to think about on the pros and cons compared with flow cytometry. Then I ll go through examples of how we do the segmentation analysis so you can see how all this fits together. Okay. So where does high content analysis fit in the scheme of the sort of cell biological techniques? Well it fits really in between flow cytometry on the one hand which has very, very good throughputs and stats for generating signal cell data and the data analysis is very well refined over decades now. At the other end of the spectrum is the conventional microscopy, which is quite labor intensive, you get fantastic data, but it s not very good for looking at large numbers of cells and large numbers of conditions. So what assay do we really use high content for? Well there s quite a range now and it gets bigger all the time. So we ve got things like morphology so we can stain the samples for morphology for example with phalloidin to look at the actin cytoskeleton. We can look for things like blebbing. People look at neurite outgrowth and how neurons are branching, differentiation because of the cells change shape when they differentiate typically. The fluorescent intensity type measures, which is probably the one you would guess we would use it for which things like cell cycle, for marker expression, which is very similar to the flow cytometric type assays. Then we ve got things like these fluorescence distribution sorts of assays and this is where we look for translocation of proteins from the membrane to the cytosol or the cytosol into the nucleus. We can look at DNA damage by looking for foci in a nuclei and we can also look for co localization. So if we can label two sets of structures within the cell fluorescently, we can see how they co localize. Then we ve got the next sort of set of two which perhaps ones which are becoming more common now. The first one is cell behavior over 7

8 time. So this is now becoming more and more important particularly in stem cell biology is to actually look at cells using time lapse type systems and also high content imaging systems can do this now. It s used to actually get real time measurements of things like the cell division times, the lineage tracing, whether the cells are migrating, and again shape changes based on the imaging. Then finally another aspect which is becoming more and more common now is to look at what we term higher order structures and this is where we combine various features that we ve imaged together to create larger objects that you might be imaging. So the simplest one is that we create virtual cells within the software. So we tell the software that we need a nucleus and we tell we re going to use a cytosol and we tell the software that those tissue be linked together and then we can ask questions about the relationship between fluorescence. Within those two compartments if we combine lots of cells together particularly with this is pretty relevant for pluripotent stem cells, you get colonies and you can ask about what the marker expression is across the whole colony. This has now been extended to embryos such as zebrafish embryos and things like that and tissue distribution by looking at slides as tissue stained slides. Slide 21 So what factors do you need to think about when you re doing these assays? Remember, these are imaging based assays so they go into a microscope. So one of the problems that we encounter of course is that typically people will grow cells in tissue culture plates for example and these can vary greatly in their ability to be imaged accurately and easily. So 6 well plates can really very, very thick plastic actually compared with coverslips for example and so you may find you have problems with sensitivity in using those sorts of plates so you have to change to either much thinner base plates or change completely to using coverslips. The resolution required, you need to think about. As you use higher and higher NA lenses, they become more and more difficult to use with standard tissue culture plastic just because of the thickness and because of the way the plastic is made. It s not very even and things like that. So again, you may need to think about slides. The well size, this is one that most people don t really think about but can you really see enough of what you need to see in one well? You know, if the well size is small and you need to look at lots and 8

9 lots of cells, you may not have enough cells in one particular well. In the case of human ES cells, colonies, it becomes a problem that we need to get enough colonies in our well to be able to see something meaningful. Then the caveat of that, of course the counter thing to that is of course remember that, you know, different size magnifications it will take you longer to cover a particular area of a well or particularly a number of wells so that can slow the assay down and that maybe a factor for you. Then the final thing you have to consider is of course the number of colors. Remember that typically we can only deal with four colors, which is in contrast with flow cytometry, which can deal with more and also for certain fluorescent reporters and things they may be very, very weak and you may have to use specialized filter sets, which you do need to think about before you start these experiments. Slide 22 So as I mentioned, there are some pros and cons regarding when you compare with flow cytometry. So in the acquisition side, the first really big, first major difference you ll notice is the throughput. The flow cytometers are very, very good at getting single cell data from large numbers of cells. The high content analysis type systems are much better for large numbers of conditions with a sort of moderate number of cells. So that s the sort of thing you need to think about there. As I ve mentioned, the parameters, the flow cytometers have a huge color part they can see. It s typically up to 12, it can go even higher depending on the machine configuration. For microscope systems it s typical just four. The sensitivity again as mentioned before, the flow cytometers are very, very sensitive instruments because of the way the optics are arranged and the type of detection they use. The high content analysis systems tend to be less sensitive in part because it s going through a lot of glass in the objectives, in part because of the plates people use and the fact they re using CCD detection. Having said that, the machines are getting more and more sensitive due to better camera technology and better light technologies. So in theory, they should become more sensitive but you do have to watch that. 9

10 [00:20:17] Obviously, subcellular imaging then apart from the imaging flow cytometers, flow cytometers don t let you do subcellular analysis whereas high content analysis that s really routine for this type of analysis. Okay. On the data analysis side, the flow cytometers again with single data, it collects single cell data very, very easily and this is for all systems that you will use it will be very, very easy to get this sort of single cell data. With the high content analysis because the cells are stuck down and next to each other, you have to do image analysis to separate them into single cells first of all. The major problem there is of course some cells will pile up on each other where you may seeded them too densely and then that causes quite a lot of problems when we start to do the actual analysis. Another factor which again as I mentioned is becoming more fashionable to do is time series or time lapse type things and of course on flow cytometry, you can only really do that by sampling but on high content analysis systems now that s becoming more routine. The systems can be supplied with CO2, humidified CO2 systems and heated chambers and the software will enable you to screen the time lapse images together to get that sort of data really quite easily now. Of course, the final thing, which is again the major thing about high content analysis, is the cell cell relations are lost with flow cytometry. So in high content analysis, we preserve all that. We know exactly where we got the information from so those relationships can be reconstructed. Slide 23 Okay. So what s the actual process that we go through? So the process is summarized here. We acquire the image. We do what s called segmentation. Segmentation is where we pull out the features we re interested in from the image and this is by sophisticated image analysis algorithms. Then what we do is we then have to tell the software what the features are and how we wanted to handle it, which is called feature extraction. We then put some parameters into the software to tell it how we want to classify those features. The example that we got here, just the most simple example is you ve got some Hoechst stained nuclei from human ES cells so you can see these colonies in this picture. They ve been segmented. If we then do some methodical treatment to that segmentation, we can 10

11 then dilate where we think the nuclei are and of course that will show us where the colonies are because it all merge together. Then what we do is we just tell the software well the colonies are the big blobs, everything else as you go smaller is not the colonies. That s what you see in the final picture there where the colonies have been labeled in green and the rest of the cells have been labeled in red and that s how we then would extract where the colonies are from these images. Slide 24 So how do we do that extraction, that segmentation extraction or at least what does it look like? So we ve got some examples in the next few slides, which will just illustrate to you the types of segmentation you can do. So on the left here, we ve got nuclei segmentation and what will typically happen here is we will apply the segmentation algorithm and then we have to try and separate the nuclei using other algorithms to separate them out from each of the it s called clump breaking or anything like this. So that s because of they re very close to each other in some situations. On the right hand side, you ve got mitochondrial segmentation, which is of course they re really quite tiny features. For this, you would have to use high NA lenses. You need a nice dye for this and then there are algorithms, which can handle these types of small vesicular features. They are different from the ones which do the nuclei segmentation, but most software will have various algorithms which can be applied to different features. Slide 25 Slide 26 We ve got for the whole cell sort of the cytosolic level. This is just very, very simply you can label the cytosol with various dyes as Dr. Telford alluded to there or with fluorescent proteins and it becomes relatively simple then to segment the cytosol, tell the software that s the cytosol and away you go. Then finally sort of larger scale features again as I mentioned is the colonies and things like this, which is caused when we merge the nuclei that we segmented from the original Hoechst stain picture and this gives us where the colonies are. The red dots in this or the red features in this picture here in fact are fibroblasts which we use as feeder the cells, which are quite easily separable from the colonies. Okay. So what s the sorts of measures we can do? Well it s really anything you can think of that you could measure from an image 11

12 once you ve extracted the feature from it. So we get all sorts of things from images once we started this process going. Again, in this example of human ES cell colonies, which have been stained with Hoechst, have been stained with pluripotency marker with SSEA 3, we can get the number of cells in the colony, the number of cells positive for the SSEA3, how far apart the nuclei are., and then going up to the colony level how many colonies are in the well, what the shape is, what the area is, and also the spatial relationship between the colonies and things like the number of positive cells in the colonies. So you can really get a lot, vast amounts of information about what this culture is really like and remember all of this is the spatial information in particular was lost with flow cytometry so we tend to use these things in conjunction with each other. [00:25:34] Slide 27 This is an example of what the data would come out like eventually. So we ve done exactly the same thing here but we ve got an additional marker which is OCT4 another pluripotency marker. As you can see by going through all this analysis process, you can get flow cytometric like data, i.e., population histograms after these images either as univariate plots or bivariate plots. You can simulate setting gates just like you would in a flow cytometer and you can get numbers that are proportionate to each fraction that you ve identified. As Dr. Telford alluded to the advantage is we can go back to look at the individual cells in these images because the software records where it s made all these images, where it s made all this analyses from. Slide 28 Slide 29 This is another example just to show you how you can get this flow cytometry like data. This is stained for BrDU and the DNA stain and you get this sort of classical horseshoe plot to show the cell cycle phases, which is similar to what we would see with a flow cytometer. So it s really quite remarkable that you can actually get this out of really quite messy and complex images. So I ll finish there and it just remains for me thank various people in various labs, which help us do all this sort of stuff because we actually make reports and all sorts of things to actually test these 12

13 systems out so it s actually quite a complex exercise and I ll stop there. Slide 30 Wonderful. Thank you so much, Dr. Gokhale, and thank you very much for joining us even though you couldn t make it into the studio. It s great to have you on the line. Our final speaker for this webinar is Dr. Liz Roquemore. Dr. Roquemore s training is in diverse aspects of intracellular protein regulation, membrane protein trafficking, and cell phenotypes. She is now the technology manager for cell applications within the Cell Technologies Division of GE Healthcare Life Sciences and has played a leading role in the development and validation of cell lines, reagents, assays, and imaging tools for research and drug discovery. Her major focus area for her work has been the application of high throughput microimaging and analysis technologies for cell biology research and drug development, most recently investigating the combined use of stem cell derived model systems and high content analysis techniques to enable earlier predictive toxicity testing. Welcome, Dr. Roquemore. Slide 31 Slide 32 Slide 33 Thank you very, a pleasure to be here. So after that great overview of HCA, I d like to take a step back and just give you sort of an overview of image cytometry and the range of instrumentation that s available for doing image cytometry and also the types of applications that those platforms are best suited to. So image cytometry in its broadest sense can be defined as the process of extracting and making sense of cytometric data obtained from images of cells. So as alluded to in the previous talks, those images can be obtained in a variety of ways with a variety of systems, but the emphasis is not so much on the images themselves in image cytometry but on the data that we can extract from those cells and the knowledge that we can gain from that process. Image cytometry has four basic components. Those include the cellular samples of course that we re working with, the probes and sensors that we add to the cells to understand their function phenotype, the imaging platforms that we re working with, and not least of all the image analysis software which allows us to turn those 13

14 images into data. All four of those components are continually evolving together so that advances in any one of those areas propels advances in the others. So we re seeing a lot of very exciting developments now in each of these four areas as image cytometry progresses. Slide 34 So why imaging as an approach to cytometry? Well I think the two previous speakers have summarized very nicely the fact that with image cytometry, we can gain quite a lot of information about morphology, structure, kinetics and position and location of cells and also the orientation of cells and groups of cells that we may be studying. Also as a lot of the higher content imaging systems can process many samples in parallel, many samples very quickly, we can do higher throughput studies, larger scale studies all the while collecting very comprehensive data from the cell population and building up unique fingerprints as to the cell phenotype. [00:30:13] Slide 35 Slide 36 Slide 37 Now we re also seeing the emergence of more convenient image cytometry systems, which are more compact, more affordable, more accessible to the members of the laboratory and they can be located throughout the lab, and you don t necessarily need an expert to run those systems or any extensive training. So image cytometers really fall on a spectrum from those which focus on depth and breadth of information that we can gain and the others that are specialized more for ease of use and accessibility. So let s take an example form the depth and breadth end of that spectrum and we ll look at cell cycle analysis as an example. So perhaps one of the simplest ways of looking at cell cycle phase is to do DNA content analysis and when we re working with adherent cells, it really makes a lot of sense to go for an image based cytometry to do that because we don t need to disturb the cells to be able to quantify the DNA content. Doing that, we can get results, which are strikingly similar to the results that we would get with flow cytometry. So here for example, 14

15 you can see that we re getting very similar cell cycle phase profiles but there are two main differences. The first is that because we re working with an imaging technology, we can process the samples very quickly and also because we re not having to remove the cells from the dish or undergo any centrifugation wash steps, the whole process is taking us a lot less time on a high content imager for these adherent cell types. The second big difference is that not only do we get these cell cycle phase distributions from the intensity information, but we have a lot of morphological information that we get virtually for free with no extra time or effort. Slide 38 Slide 39 Slide 40 So here you can see for example that when we treat with mitomycin C, we re getting a dramatic increase in the nuclear size. We can follow that for every cell in every image and that can give us more confidence in the conclusions that we draw from our cell cycle phase analysis, but it also can lead to new insights into the function and the mechanism of action of our drugs. We can add a second sensor as shown in the previous talk to further delineate the S phase cells. In this case, we re looking again at incorporation of bromodeoxyuridine and we re getting that classic horseshoe shaped distribution of the cells. But things get really interesting if we get a third sensor to the mix. Here, this is something that we could only do with an image based platform because this is a GFP sensor, which marks the cell cycle phase position of the cell by not only its intensity but the position of this probe within the cell. So now, we see something very interesting here. So we ve again got that classic horseshoe shaped distribution of our control cells in blue and our cells treated with a test compound in red. You can see that the test compound is causing that horseshoe shaped plot to shift to the right, but now we ve got a third dimension added to this plot and that s represented by the diameter of the spheres that are plotted here. Those diameter of the spheres are telling us about the cell cycle phase position as reported by this GFP reporter. What we find is that even though the cells on the right which are 8n are in the position that you d expect G2 cells to be, they actually 15

16 have characteristics of G1 phase cells. So by using this GFP reporter, we ve learned a little bit more about the mechanism of action of our compound, which happens to be where our kinase inhibitor and it s causing the cells to undergo a process called endoreduplication where the nuclei divide but the cells cannot undergo cytokinesis. Slide 41 So I hope that gives you a flavor of some of the many types of things you can do with an image cytometer that you maybe you couldn t do with other technologies. There are some emerging trends in the instruments that are available to do this and some of those include a more complete solution. So increasingly, we re seeing systems which take us all the way from acquisition through to analysis and even interpretation of our results seamlessly. We re also seeing more fluorescence based kits and assays available that can really cut down on our assay development times. Another thing that we re seeing is that the systems are all getting quite fast and they re able to collect much higher quality images than ever before. That s very important because the quality of the data that we get from HCA is only as good as the images we start with. A third trend that we re seeing is towards more flexibility with confocality. So instead of being locked in for example to a particular pinhole sized confocal aperture, we re now seeing systems which give us variable aperture and that can really give us more flexibility in terms of the applications that we can cover. All of these improvements are being made possible by new technologies being incorporated into the systems including scientific grade CMOS cameras, diode lasers and LED technology. [00:35:20] Slide 42 Slide 43 But what about the more routine assays that we do every day in the laboratory like counting cells and doing cell health and phenotype analysis? For those systems, maybe high content analysis is a little bit of an overkill but we still want to get some of the benefits of imaging. There, we might want to turn to systems, which are more focused on ease of use and accessibility. A good example there is cell count and viability. So this is an assay that we do every day in the lab. It s very simple and we really want 16

17 to be able to do it anywhere, anytime and not have to turn to an expert to do it. So typically we ll run a two color assay like this one where we re marking the total cell population with a red dye and the dead cell population with a green dye and in this case we were running it on a Cytell Image Cytometer and you can see that we re able to get very familiar results. We can set our thresholds to separate the live from the dead cells and we ve got a selection of gating tools so that we can do further subpopulation analysis. Slide 44 Then we can generate dose response curves from our data and here s an example of a camptothecin dose response curve and its effect on cell viability. In case you re wondering what that little disc is in the upper right hand corner, that is the sample holder for Cytell. It can hold eight different sample cartridges. So we d simply put a small for this dose response curve, we just put a small volume of sample into each of those eight cartridges and we get the results for the entire dose response curve in about two and a half minutes. Now these are the results with Jurkat cells, which are a suspension cell type. But what we find is that whether we re working with suspension cells or adherence cells, we re getting results which are strikingly comparable to other image cytometers. Slide 45 Then my last example is one where we were working with peripheral blood monocytes and we were developing an enrichment procedure for T cells and this is another one where an image cytometer came in handy. So here for example, you can see our results are showing us that our cells are about 96% CD3 positives. So this was a good verification for us that our T cell enrichment process was working well. The other thing that we wanted to know about was the ratio of CD4 helper cells to CD8 cytotoxic T cells. That s typically about 1.5 to 1 in the average individual and here you can see our results that we got with Cytell showed we had 1.5 ratio, which corresponded very well to published results. Another thing is that we wanted to migrate this assay from flow cytometry, which we do quite a lot of in the laboratory, to an image based cytometer so that we could handle this routine assay a little bit more readily within the lab. So we did a little bit of benchmarking and here you can see that we ve got strikingly comparable results between the Cytell Image Cytometer and the flow cytometer that we had in our laboratory. 17

18 Slide 46 So what are the things that we had to consider when we were thinking about doing these more routine assays on an image cytometer? Well of course, we thought about the level of expertise available in the lab and where we would like this instrument to be located and we re finding that some of these more value image cytometers can be located anywhere and don t require a lot of expertise. Another thing to consider is whether we go with predeveloped acquisition and analysis protocol or user defined ones. We like a mixture of both because the predeveloped ones can save us time and expertise but the freedom to create your own can give you some flexibility. Speaking of flexibility, it s also good to have flexibility for the various dyes so that we re not locked into any particular assays. If we re working with an open system like this, we need to think about the number of excitation and emission channels that are going to be available. Last but not least, speed is something that needs to be considered. So when we want to think about the speed all the way from the time we load the sample to the time we get the results, and if we can run multiple samples in parallel that can really save us time. Another thing to remember is about the hidden cost, time penalties that might be associated. So for example if a system needs to have a lot of setup time or if there s calibration involved or wash steps, those can be time penalties that maybe you hadn t counted upon. Slide 47 [00:40:03] So just to summarize, I think in our experience we found that image cytometry and flow cytometry are very complementary. One is not necessarily going to replace the other but with image cytometry, we can get quite a lot of additional morphological information. We can use adherent cells or suspension cells and increasingly the technology to do that is becoming easier and more accessible. Now we re in a state where we ve got quite a lot of instruments to choose from and so we can actually afford to match the image cytometry solution to our needs rather than being driven by the instrumentation in our assay choices, and that s really a smart way to go. 18

19 So for detailed phenotyping and functional studies and screens, we re finding HCA has quite a lot to offer in terms of the power and flexibility that we have. For the more routine assays, I think it s really worth looking at some of these value image cytometers that really can bring that capability right to your bench top without a lot of added expertise required. Slide 48 Slide 49 Slide 50 So finally, just to show you some of the image cytometry solutions available from GE healthcare that we ve used to generate the data in this presentation. At the bottom I ve got the IN Cell Analyzer 6000 which is our confocal version of the high content imager, it s got variable aperture technology, the IN Cell Analyzer 2000 which is our very flexible wide field system. Our most recent introduction is the Cytell Image Cytometer, which is quite compact. It s got two lasers and four emission channels and really does allow us to do quite a lot of the routine assays everywhere in the lab. So that brings me to the end of the talk and I just like to thank people for their attention and hand back over to Sean. Great. Thank you so much, Dr. Roquemore, and thank you to all of our speakers for the wonderful presentation. We re going to move right on to the questions submitted by our online viewers. Just a quick reminder to those watching us live that you can still submit your questions by typing them into the text box and clicking the submit button. If you don t see the box on your screen, click the red Q&A widget and it should appear. So we ve got a lot of questions so I m going to try knock through as many of them as possible. But the first question I m going to ask is about the percentage of assays that are flow versus HCA in your experience. So I know all of you have done both of these so maybe you can talk to that in your particular laboratory, Dr. Telford? Dr. Bill Telford: Sure. Well I run a core facility at the National Cancer Institute so we probably have about 80% of our work being flow cytometry and the other 20% being image cytometry. But the image cytometry tends to be done for very specific purposes where the imagery is really critical. We do a lot of apoptosis analysis on the image cytometer 19

20 because you want the morphological data along with the fluorescence data and it s very powerful for that purpose. We have some other interesting applications too. We do a lot of cellcell based assays where we want to look at cell cell interactions and that s something again where you re not going to get that information out of traditional flow in the stream. You ve got to have the cell sitting on a dish to collect that data. We have people who do very small numbers of cells where they ll just dab them on a plate and allow them to settle. We have applications where people want to analyze their GFP expression in a 96 well plate, but they don t want to discard the cells. They want to put them back in an incubator and keep running them. With these image cytometers, with many of them, you can put the plate with the lid on, on the instrument, scan it, and then put it back in the incubator. Again, you haven t destroyed your cells as part of the analysis. So a lot of specialty applications like that where flow cytometry really doesn t provide the advantages that image cytometry does. Dr. Liz Roquemore: Dr. Paul Gokhale: Great. Dr. Roquemore? Well I think our experience is very similar to yours in that we probably have about 30% of the people in the lab doing flow cytometry and then maybe 70% of the assays would be high content analysis. We do a lot of cell production and I think flow cytometry is very useful there. But increasingly we re finding these bench top image cytometers can really get the job done quickly. We don t need to have a lot of trained experts to do it and so we re getting those being used more and more often. Then on the high content analysis side, we re finding that we can get really detailed cellular phenotype data from the HCA, which we just couldn t get by any other means and that s really breaking new ground in terms of understanding how compounds act on a cell and predicting their toxicology, toxicity for example. Great. Dr. Gokhale? Yeah it s very similar actually to Dr. Telford. It s about 80%, 20% and it s for very similar reasons. It s where we need morphological information, we will do the in situ high content assay or for time lapse imaging. Uh hum. Excellent. 20

21 Dr. Paul Gokhale: Dr. Liz Roquemore: [00:45:02] Dr. Liz Roquemore: Dr. Bill Telford: Dr. Bill Telford: For the rest, we do flow cytometry. Great. So the next question I m going to ask, there s been a couple of questions about resolution and size limit for flow and image cytometry and I know that one of you did show a slide looking at the mitochondria. So really, what is the resolution of image cytometry current? What can you get down to? Maybe Dr. Roquemore, we ll start with you. I think with the high content systems particularly if you re going for the confocal ones, we can go quite small down to the micrometer size. All right. So we can see small things like mitochondria. We get great detail from the cells. On some of the more bench top image cytometers, for Cytell for example the size range would be from 5 microns up to about 25 microns. Okay. Dr. Telford? Yeah. Again, it really varies depending on the system, although some systems have some flexibility. A lot of the data I showed in my talk was from the Compycyte icys system and it can go down to I believe 0.5 micron resolution. So you re not visualizing individual organelles at that level but you can look at translocation from the cytoplasm to the nucleus. You can look at fluorescence polarity if the fluorescence is greater on one side of the cell or another and you can also vary that resolution. So the higher the resolution, the slower the scan. If you need to scan 96 well plates for viability, you can scan them but a very low resolution like a 5 or 10 micron resolution and do a 96 well plate in 15 minutes. Uh hum. And the you can do the converse situation where you need to look at individual cells and look at subcellular detail and the scans can be much longer than that. Great. Dr. Gokhale, how about in your experience? 21

22 Dr. Paul Gokhale: Dr. Bill Telford: Dr. Bill Telford: Dr. Paul Gokhale: It s typically down to mitochondrial size to be honest with the much smaller you re starting to use very, very specialized optics done. I mean you can get 60 times and a hundred times air lenses for example and water immersion on these systems. But as I alluded to in my talk, it becomes progressively more difficult to maintain focus on things in an automated system because you got other factors affecting you like autofluorescence caused by the plastic ware or the fact the plastic ware is not made flat. So you then start to really need to be thinking about systems, which look at coverslips and then maybe more of a hybrid system. You can now get software for the conventional microscopes which behaves much more like a high content system where you can take multiple fields in a well or on a slide. So you start to think about systems like that where you re using proper high confocal optics and things so. Great. Excellent. So I m going to come to you, Dr. Telford, for a more specific question about stem cells. We have a couple of questions on that here. This person asks what markers and staining protocols you use to identify your stem cells and how you manage quality control? Well we certainly you know, most of our stem cell analysis is in the flow cytometry platform, although we had done some image cytometry as well. So we use a wide variety of cell surface markers. Most of our work is murine so we ll look at [c Kit+, Sca 1. We ll look at the SLAM markers, CD150, 244, 48. We still use Hoechst side population quite a bit. This is an older technique. It was developed by Peggy Gooddell back in the mid 90s, but it s proven to be relatively resilient in the sense that it does while certainly not universal it does appear in many mammalian stem cell hematopoietic and non hematopoietic populations. Uh hum. So we still include that in our analysis and it s actually quite compatible with immunophenotyping so we don t lose anything if we add in the Hoechst red and Hoechst blue or dicycle red and blue. I think probably Dr. Gokhale can speak more to the imaging aspects of stem cells than we can. Sure. Sure. Dr. Gokhale? Yeah. I mean we actually use the same antibodies we would use on flow cytometers, although of course from these antigens and things we discover, the image cytometer didn t really exist so it s exactly 22

23 the same antibodies. So it s very the stains and everything are really very similar apart from the trypsinization of the cells to what we would use for flow cytometers. So and you have the additional thing that we would coat Hoechst in with thick cells just to locate the cell so very, very similar. With regards to quality control, all the antibodies are titered against cells that we re using to check that they are in fact at the correct place in the titration curve and the same for the secondaries. Also, we do have cell lines that we know express the markers that we use that can be used as positive control as well. So we tend to find that gives us a bit more consistency if we include those controls. Dr. Paul Gokhale: [00:50:26] Dr. Bill Telford: Uh hum. Now stem cells tend to be rather delicate cells and flow cytometry can be a little bit harsh as far as conditions are concerned so I m guessing that maybe the HCA systems might be a little bit better for looking at stem cells? So maybe Dr. Gokhale, can you comment on that? Well yes. Well particularly human pluripotent cells, like ES cells and human embryonic stem cells, do not like being on their own. So when you trypsinize them, you ll start to induce cell death in a large proportion of them in theory. So whilst you can analyze them on a traditional flow cytometer, you know, if you just put in and so you don t care the cells die afterwards, if you want to recover them with the flow sorter, I m afraid you re going to get quite high cell losses. We typically factor it about 50% cell loss. The other factor when going through a flow sorters that we found is that human embryonic stem cells have very little cytosol so they don t like going through these machines at very high speeds because they can t endure shearing forces which can rip them apart. So we have to turn all the pressures down and actually sort at a slower rate than you might be able to in some of the cell types. Yeah, it s just one of those things. The mouse cells clone a lot better. They re actually better at handling this but for some reason human embryonic stem cells just don t like being on their own. Uh hum. So Dr. Telford, I know you have some experience with the use of flow. Yeah. We ve observed exactly the same thing. Really the only strategy you can do is turn down the pressure on your instrument. Use a larger nozzle. Basically go back to the pressures that we used 23

24 15, 20 years ago for sorting all cells to try to, you know, minimize the damage. Stem cells, particularly human stem cells they re like dendritic cells and apoptotic cells, they tend to be very, very fragile and they don t tend to deal with the high differential pressures in stream and air sorters very well. So really just turn down the pressure and maximize the nozzle size. Dr. Bill Telford: Dr. Liz Roquemore: Dr. Liz Roquemore: Dr. Liz Roquemore: Dr. Liz Roquemore: Dr. Liz Roquemore: Right. If you can. Dr. Roquemore, any comments on stem cells? Yes. We work with stem cells quite a lot in the lab and I think our experience is similar to the other two speakers really that the cells like to be together. They don t like a lot of shear forces and so a lot of time we are using the high content system to characterize those cells. It gives them time to recover on the dish, makes sure that they re expressing everything they should be expressing. Uh hum. And they re just generally happier cells then when we go to image them. Right. So the next question I have is the application of immunocytometry in microbiology and I m not sure which of the speakers have much experience but maybe Dr. Roquemore, are you aware of anyone that s using it for bacteria We are aware of some applications where people are using high content for very small organisms like yeast and bacteria but you don t hear about that a lot. Uh hum. So I know that the resolution is there to do it but I m not you know, you don t hear about a lot of applications in that area. Right. Particularly because I think people are looking for things they can do at high speed screening. 24

25 Dr. Bill Telford: Okay. So the next question, the viewer seems very impressed by the data. He says it seems that the assays are great, there s no need and it doesn t seem there s a need for an expert to be involved. But is this the truth, is this, the high content systems, are they something that you can set up and somebody maybe with less experience can use? Dr. Telford, you want to try this? You know, what we tell our users with the image cytometry is that the sample is 2/3 of the battle because unlike flow cytometry where you re triggering on individual cells and most of your trash and debris are flying away and you re not seeing them, with imaging, you re seeing everything. You re seeing all the garbage and everything else. So you really need to do a very high quality sample preparation and what we tell our investigators is the same techniques and care that you apply to your microscopy should be applied to image cytometry as well because the same problems that you ll have with microscopy will also occur with image cytometry too. You choose fluorochromes that don t bleach easily. A lot of these systems have lamp or laser dwell times in the seconds where a typical flow cytometer illuminates that the cell for about a microsecond. So we like to use some of the newer fluorinated, sulfonated low molecular weight probes like the Alexa Fluor dyes work rather well, Quantum dots. Quantum nanoparticles work very well. Take the same precautions we would from microscopy, use reagents that will prevent photobleaching. Anything you can do to maximize the intensity of your fluorescence and to keep your samples as clean as possible should be applied here. Dr. Paul Gokhale: Dr. Paul Gokhale: Uh hum. Great. Dr. Gokhale, I ve got a question for you that came in during your presentation about the software packages that you use when you re doing your analysis. Are these available commercially or are these something that s being developed in your lab? Well the images, the thing that I showed are actually from commercial packages, which is the GE developer toolbox. Uh hum. There are other commercial packages out there, which will do similar things and you really do need to check them out if they will do what you want them to do with the I would strongly recommend that you test it on real images of your biology that you re trying to look at 25

26 [00:55:09] Dr. Liz Roquemore: Dr. Liz Roquemore: Dr. Bill Telford: because you can have all sorts of problems trying to segment images with certain packages depending on how they do it. For the time lapse imaging, we do develop our own software for that but that s because it s using phase contrast imaging so it s using very sophisticated image analysis techniques to do that so that s not commercially available. But I fully expect that sort of systems to be available in the future actually. It s becoming more and more common to incorporate that sort of analysis into these software packages. Uh hum. Excellent. So a question now from somebody who s obviously looking to purchase a high content system and they re interested in what you would say are the most important features to consider when you re looking at an instrument. Dr. Roquemore, I know you might be a little bit biased towards the GE machines but maybe you can all talk about the things that you look for. So we ll start with you, Dr. Roquemore. Yeah. I ll try to be as unbiased as possible. I have been doing microscopy for a number of years and Right. used a lot of different systems. I think key when thinking about a system is going to be the flexibility because you know you re investing quite a lot in a system and you want to make sure that it s going to be able to do the applications that you re doing now and the things that maybe are unforeseeable that you want to do in the future. Another thing has got to be image quality. I think, you know, you want to make sure you got the best optics that are available to generate those images because the data that you get really is only going to be as good as the images. The third thing I would say is speed. If you re turning to a high content system, it s not like a research grade microscope. You re investing in one so that you can get a lot of data from many different samples in order to be able to get that high content information so speed is really important. So those are three key things that I would look for. Right. Dr. Telford? Yeah. I would also think about some of the ways, some of the internal aspects of the instrument whether it s for example laser based or lamp based. Most image cytometer systems either use 26

27 lasers where they raster an image or use some of their line scanning or they use a lamp like a xenon lamp to excite the sample. On the downstream detection, they can use sometimes photomultiplier tubes. CCD cameras and chips now are becoming more common in some systems. As far as the laser based systems, I think you re going to have better sensitivity. You may have better edge to edge uniformity of the sample than you will with the lamp based systems. But laser systems tend to be slower and lamp based systems can often be very, very rapid, have very high throughput. With the xenon lamp, you can also filter off any bandwidth you want really from the violet all the way to up to the red. So there s a certain amount of flexibility in excitation out there. But certainly some of the systems that we ve talked about today, we ve talked about the Compycyte icys, the GE IN Cell Systems, the PerkinElmer systems. These are all being used very actively in laboratories now and we ve tested most of them and they re all very, very good at what they do. So they re all worth considering. Just think about what your application is going to be and whether sensitivity is your primary concern, is high throughput your primary concern, high versus low resolution and so forth. Dr. Paul Gokhale: Uh hum. Great. Dr. Gokhale, do you have anything to add? Well there s a couple of things, some very, very sort of practical aspects is also focus systems. So with HCA systems one of the things that you want it to is go from field to field and be able to focus as it goes from well to well. You really got to have autofocus systems that can work very, very well. You know, so you must test that out with whatever plates or whatever you re planning to use. The other thing is the software, will it do the things that you want it to do and is it easy to learn and stuff like that. You know, you want to be concentrating on looking at the biology and understanding the biology not having to, you know, relearn programming or something, you know. So I think those are the additional things. The points that the others speakers made though are excellent sort of guides to what you should look for. Fantastic. Well we re at the top of the hour but I m going to shoot one more question to you that maybe you can answer very briefly in 30 seconds. That is where do you see the field of high content analysis moving in the next five to ten years or what would you like to see that will really drive your research? So we ll start with you, Dr. Telford. 27

28 Dr. Bill Telford: [01:00:14] Dr. Paul Gokhale: Dr. Liz Roquemore: Slide 51 What I d like to see is all flow cytometers have imaging where imaging will become a parameter just like forward scatter or fluorescence is a parameter. I think you re seeing it in the Amnis Image Stream. It s sort of the forerunner of this, which is currently the one system that will do in stream imaging. I think all the major manufacturers are looking at that very closely and I would very much like to see a picture of the cell be a parameter and all the subsequent intracellular data that you can pull out of that be a parameter just like fluorescence or forward scatter is. I think the slide based systems are quite well developed now. I think we re going to see increases in speed and resolution but that s another direction that has proven to be very productive in biomedical sciences, and I think we ll continue. Uh hum. Dr. Gokhale, now about you? More colors will be useful. And when we compare it with flow cytometers, at the moment we re quite limited which is partly reagents so it s not really the manufacturer s fault. But it would be good to have systems which could handle more colors if we could actually have the if we have the reagents to do it. Right. Great. And Dr. Roquemore, last word for you. I think we will see quite a lot of good developments in this area but one of the main things I d like to see is the systems continually become easier to use. So the emphasis is no longer on being an expert to use the instrument but putting the scientist is then freed up to put the effort in to developing the assay and handling the cells properly and the instrument is just there and does its job. And also I think we will see great developments on the image analysis side of things so that not only will this software be analyzing the data for you but it will be interpreting it for you and helping you process all that great amount of high content data. Fantastic. I just wanted to mention very briefly before we close out the webinar, I know some of you did have some trouble downloading the slide deck. We will make sure that that is available on the on demand webinar so come back any time in about 48 hours to look for the on demand. You will get an alerting you to that, the availability of that archive. 28

29 [01:02:40] End of Audio But we are unfortunately out of time now for this webinar so on behalf of myself and our viewing audience, I wanted to thank our speakers very much for being with us today: Dr. Bill Telford from the NIH, Dr. Liz Roquemore from GE Healthcare, and Dr. Paul Gokhale on the phone from the University of Sheffield. Please go to the URL now at the bottom of your slide viewer to learn more about resources related to today s discussion, and look out for more webinars from Science available at webinar.sciencemag.org. This webinar will be made available again as an on demand presentation within approximately 24 to 48 hours from now. We'd love to know what you thought of the webinar, send us an at the address now up in your slide viewer, [email protected]. Again, thank you to our wonderful panel and to GE Healthcare for their kind sponsorship of today's educational seminar. Goodbye. 29

Cytell Cell Imaging System

Cytell Cell Imaging System Cytell Cell Imaging System Simplify your cell analysis and start changing tomorrow today www.gelifesciences.com/cytell The Cytell Cell Imaging System is an intuitive cell analysis instrument that streamlines

More information

Introduction to flow cytometry

Introduction to flow cytometry Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer

More information

Uses of Flow Cytometry

Uses of Flow Cytometry Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7

More information

IncuCyte ZOOM Whole Well Imaging Overview

IncuCyte ZOOM Whole Well Imaging Overview IncuCyte ZOOM Whole Well Imaging Overview With the release of the 2013B software, IncuCyte ZOOM users will have the added ability to image the complete surface of select multi-well plates and 35 mm dishes

More information

use. 3,5 Apoptosis Viability Titration Assays

use. 3,5 Apoptosis Viability Titration Assays Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

a measurable difference

a measurable difference Thermo Scientific Cell Analysis Software and Informatics Products a measurable difference for everyone Thermo Scientific HCS Studio Cell Analysis Software Thermo Scientific Store Image and Database Management

More information

Immunophenotyping peripheral blood cells

Immunophenotyping peripheral blood cells IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune

More information

Thermo Scientific CellInsight CX5 HCS Platform. discover more uncover the possibilities

Thermo Scientific CellInsight CX5 HCS Platform. discover more uncover the possibilities Thermo Scientific CellInsight CX5 HCS Platform discover more uncover the possibilities high content screening Driving your Productivity with Innovated Technology What is High Content Screening (HCS)? High

More information

EDF Extended Depth of Field

EDF Extended Depth of Field EDF Extended Depth of Field An upgrade for the ImageStream system Think outside the dot. Break the classic depth of field barrier Improve precision Enhance discrimination Simplify analysis Increase resolution

More information

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes

More information

FlowSight. Flow cytometry with vision

FlowSight. Flow cytometry with vision FlowSight Flow cytometry with vision Flow cytometry with vision Introducing FlowSight Capable: Sensitive and flexible for every need Intuitive: Easy-to-use, with imagery for every cell Affordable: Designed

More information

An Overview of Cells and Cell Research

An Overview of Cells and Cell Research An Overview of Cells and Cell Research 1 An Overview of Cells and Cell Research Chapter Outline Model Species and Cell types Cell components Tools of Cell Biology Model Species E. Coli: simplest organism

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples

More information

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection CyTOF2 Mass cytometry system Unveil new cell types and function with high-parameter protein detection DISCOVER MORE. IMAGINE MORE. MASS CYTOMETRY. THE FUTURE OF CYTOMETRY TODAY. Mass cytometry resolves

More information

Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model

Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model A p p l i c a t i o n N o t e Z-Stacking and Z-Projection using a Scaffold-based 3D Cell Culture Model Brad Larson and Peter Banks, BioTek Instruments, Inc., Winooski, VT Grant Cameron, TAP Biosystems

More information

APPLICATION INFORMATION

APPLICATION INFORMATION DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL

More information

Standardization, Calibration and Quality Control

Standardization, Calibration and Quality Control Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic

More information

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required. PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo ([email protected]) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,

More information

NyONE - Cell imaging in a bird s eye view 4. NyONE...resolution matters! 8. Features & benefits 10. Fluorescence excitation channels 12

NyONE - Cell imaging in a bird s eye view 4. NyONE...resolution matters! 8. Features & benefits 10. Fluorescence excitation channels 12 Envisions confirmed Content NyONE - Cell imaging in a bird s eye view 4 From cells to numbers 6 NyONE...resolution matters! 8 Features & benefits 10 Fluorescence excitation channels 12 Technical specifications

More information

Automation of Cell Staining Technical Information Bulletin

Automation of Cell Staining Technical Information Bulletin Automation of Cell Staining Technical Information Bulletin Automation of Cell Staining Using the Biomek 4000 Laboratory Automation Workstation Amy N. Yoder, Sr. Development Scientist and Li Liu, Sr. Development

More information

How to operate the BD FACSCanto flow cytometer

How to operate the BD FACSCanto flow cytometer How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.

More information

Cell Cycle Tutorial. Contents

Cell Cycle Tutorial. Contents Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6

More information

BD FACSComp Software Tutorial

BD FACSComp Software Tutorial BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS

More information

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor

More information

MetaMorph Software Basic Analysis Guide The use of measurements and journals

MetaMorph Software Basic Analysis Guide The use of measurements and journals MetaMorph Software Basic Analysis Guide The use of measurements and journals Version 1.0.2 1 Section I: How Measure Functions Operate... 3 1. Selected images... 3 2. Thresholding... 3 3. Regions of interest...

More information

Annexin V-EGFP Apoptosis Detection Kit

Annexin V-EGFP Apoptosis Detection Kit ab14153 Annexin V-EGFP Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in various samples This product is for research use only and is not intended

More information

Below is a list of things you should be aware of before you schedule your sort.

Below is a list of things you should be aware of before you schedule your sort. Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as: INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals

More information

Thermo Scientific ArrayScan XTI High Content Analysis Reader. revolutionizing cell biology with the power of high content

Thermo Scientific ArrayScan XTI High Content Analysis Reader. revolutionizing cell biology with the power of high content Thermo Scientific ArrayScan XTI High Content Analysis Reader revolutionizing cell biology with the power of high content learn more about your cells using high content technology Thermo Scientific High

More information

Practical Cell Analysis

Practical Cell Analysis Practical Cell Analysis Dimitri Pappas Dept of Chemistry & Biochemistry, Texas Tech University, USA WILEY A John Wiley and Sons, Ltd, Publication Contents Preface Acknowledgments xiii xix 1 Getting Started

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Item No. 10009853 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions

More information

Mike: Alright welcome to episode three of Server Talk, I m here with Alexey. I m Mike. Alexey, how are things been going, man?

Mike: Alright welcome to episode three of Server Talk, I m here with Alexey. I m Mike. Alexey, how are things been going, man? Mike: Alright welcome to episode three of Server Talk, I m here with Alexey. I m Mike. Alexey, how are things been going, man? Alexey: They re doing pretty good. Yeah, I don t know, we ve launched two

More information

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.

More information

The Olympus stereology system. The Computer Assisted Stereological Toolbox

The Olympus stereology system. The Computer Assisted Stereological Toolbox The Olympus stereology system The Computer Assisted Stereological Toolbox CAST is a Computer Assisted Stereological Toolbox for PCs running Microsoft Windows TM. CAST is an interactive, user-friendly,

More information

Neuro imaging: looking with lasers in the brain

Neuro imaging: looking with lasers in the brain Neuro imaging: looking with lasers in the brain Aim: To image life cells, label free, with cellular resolution in deep tissue Marloes Groot Vrije Universiteit Amsterdam Faculteit Exacte Wetenschappen Natuurkunde

More information

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University

More information

Flow Cytometry A Basic Overview

Flow Cytometry A Basic Overview Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,

More information

Automotive Applications of 3D Laser Scanning Introduction

Automotive Applications of 3D Laser Scanning Introduction Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc

More information

COMPENSATION MIT Flow Cytometry Core Facility

COMPENSATION MIT Flow Cytometry Core Facility COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,

More information

BD Pathway Bioimaging Systems

BD Pathway Bioimaging Systems BD Pathway Bioimaging Systems Cellular Imaging for High-Content Analysis Delivering industry-leading bioimaging, BD Pathway high-content cell analyzers combine superior image quality, flexible image capture,

More information

PROTOCOL. Immunocytochemistry (ICC) MATERIALS AND EQUIPMENT REQUIRED

PROTOCOL. Immunocytochemistry (ICC) MATERIALS AND EQUIPMENT REQUIRED PROTOCOL Immunocytochemistry (ICC) 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 11-07 MATERIALS AND EQUIPMENT REQUIRED Materials: MitoSciences primary monoclonal antibody/antibodies Fluorophore-conjugated

More information

Introduction to Open Atrium s workflow

Introduction to Open Atrium s workflow Okay welcome everybody! Thanks for attending the webinar today, my name is Mike Potter and we're going to be doing a demonstration today of some really exciting new features in open atrium 2 for handling

More information

INSIDE THE BLACK BOX

INSIDE THE BLACK BOX FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

LOAD & GO and make flow cytometry simplicity itself. the full spectrum of Flow Cytometry

LOAD & GO and make flow cytometry simplicity itself. the full spectrum of Flow Cytometry LOAD & GO and make flow cytometry simplicity itself the full spectrum of Flow Cytometry Introducing Aquios CL...the first true LOAD & GO flow cytometer What would it mean for your laboratory s productivity

More information

Application Note No. 2 / July 2012. Quantitative Assessment of Cell Quality, Viability and Proliferation. System

Application Note No. 2 / July 2012. Quantitative Assessment of Cell Quality, Viability and Proliferation. System Application Note No. 2 / July 2012 Quantitative Assessment of Cell Quality, Viability and Proliferation System Quantitative Assessment of Cell Quality, Viability and Proliferation Introduction In vitro

More information

Fluorescent Array Imaging Reader

Fluorescent Array Imaging Reader Fluorescent Array Imaging Reader Multiplex Enabled For 2-Color Fluorescent Microarrays Extended Dynamic Range Automated Spot Analysis Compact and Affordable Microarray Analysis With FLAIR FLAIR Sensovation

More information

Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis

Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis Cam CANIMPEX CPX-SOLUTIONS 3D Digital Microscope Camera FAST PRECISE AFFORDABLE 3D CAMERA FOR MICROSCOPY Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis

More information

The Basics of Scanning Electron Microscopy

The Basics of Scanning Electron Microscopy The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single

More information

EXPERT: Rich Sloan, co-founder and CEO, StartupNation.com; Darrin. VIDEO DESCRIPTION: See how Wells Fargo s OptRight Payroll Solutions can

EXPERT: Rich Sloan, co-founder and CEO, StartupNation.com; Darrin. VIDEO DESCRIPTION: See how Wells Fargo s OptRight Payroll Solutions can WELLS FARGO BUSINESS INSIGHT SERIES VIDEO TITLE: Welcome to OptRight Payroll Solutions MODERATOR: N/A EXPERT: Rich Sloan, co-founder and CEO, StartupNation.com; Darrin Northagen, Sales Manager, Wells Fargo

More information

CyAn : 11 Parameter Desktop Flow Cytometer

CyAn : 11 Parameter Desktop Flow Cytometer CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on

More information

full library phenotypic screening without compromise www.ttplabtech.com

full library phenotypic screening without compromise www.ttplabtech.com R full library phenotypic screening without compromise www.ttplabtech.com full library phenotypic screening is within your reach acumen is unique, has sophisticated software and offers the best characteristics

More information

Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation

Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation Comparison of Trypan Blue and Fluorescence-Based Viability Detection Methods Via Morphological Observation Nexcelom Bioscience LLC. 360 Merrimack Street, Building 9 Lawrence, MA 01843 T: 978.327.5340 F:

More information

International Beryllium Conference, Montreal, Canada March 10, 2005

International Beryllium Conference, Montreal, Canada March 10, 2005 Alternative Lymphocyte Proliferation Tests: BrdU and Flow Cytometry Based Tests International Beryllium Conference, Montreal, Canada March 10, 2005 Tim K. Takaro Department of Environmental and Occupational

More information

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie [email protected] MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment

More information

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows

More information

Islet Viability Assessment by Single Cell Flow Cytometry

Islet Viability Assessment by Single Cell Flow Cytometry Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing

More information

Flow Cytometry. flow cytometer DNA apoptosis ph

Flow Cytometry. flow cytometer DNA apoptosis ph Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal

More information

Automated Library Preparation for Next-Generation Sequencing

Automated Library Preparation for Next-Generation Sequencing Buyer s Guide: Automated Library Preparation for Next-Generation Sequencing What to consider as you evaluate options for automating library preparation. Yes, success can be automated. Next-generation sequencing

More information

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating

More information

Animal & Plant Cell Slides

Animal & Plant Cell Slides Animal & Plant Cell Slides Category: Biology Type: Class Experiment, 60 min class Materials: 2 Glass Slides 2 Cover Slips 1 Bottle of methylene blue (optional) 1 Plastic tray 1 Bottle of iodine 1 Plastic

More information

MIRAX SCAN The new way of looking at pathology

MIRAX SCAN The new way of looking at pathology Microscopy from Carl Zeiss MIRAX SCAN The new way of looking at pathology Greater reliability. Greater efficiency. Virtual microscopy for research & analysis. Better. More efficient. Quality as the basis

More information

Fighting the Battles: Conducting a Clinical Assay

Fighting the Battles: Conducting a Clinical Assay Fighting the Battles: Conducting a Clinical Assay 6 Vocabulary: In Vitro: studies in biology that are conducted using components of an organism that have been isolated from their usual biological surroundings

More information

UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a

More information

Proposal Writing in a Nutshell

Proposal Writing in a Nutshell Proposal Writing in a Nutshell Context Resources http://www.sci.utah.edu/~macleod/grants/ Find Your Purpose Who is the Audience? If you don t know, find out.?? Don t assume too much!! What are Review Criteria?

More information

Chronoforums. Written by ClubAero.nl, 8 th December 2013

Chronoforums. Written by ClubAero.nl, 8 th December 2013 Written by ClubAero.nl, 8 th December 2013 Chronoforums ClubAero.nl is an association set up in the Netherlands to lease or charter a regional airliner for short single day or weekend ski-trips. After

More information

SETTING UP THE NEW FACEBOOK BUSINESS PAGE DESIGN

SETTING UP THE NEW FACEBOOK BUSINESS PAGE DESIGN SETTING UP THE NEW FACEBOOK BUSINESS PAGE DESIGN CREATIVE After logging in and accessing your Facebook business page as an administrator, you ll be presented with the following message: Start Tour: Click

More information

Flow cytometry basics fluidics, optics, electronics...

Flow cytometry basics fluidics, optics, electronics... Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence

More information

Cell Viability Assays: Microtitration (MTT) Viability Test Live/Dead Fluorescence Assay. Proliferation Assay: Anti-PCNA Staining

Cell Viability Assays: Microtitration (MTT) Viability Test Live/Dead Fluorescence Assay. Proliferation Assay: Anti-PCNA Staining Cell Viability Assays: Microtitration (MTT) Viability Test Live/Dead Fluorescence Assay Proliferation Assay: Anti-PCNA Staining Spring 2008 1 Objectives To determine the viability of cells under different

More information

Outline. 1. Experiment. 2. Sample analysis and storage. 3. Image analysis and presenting data. 4. Probemaker

Outline. 1. Experiment. 2. Sample analysis and storage. 3. Image analysis and presenting data. 4. Probemaker Tips and tricks Note: this is just an informative document with general recommendations. Please contact [email protected] should you have any queries. Document last reviewed 2011-11-17 Outline 1. Experiment

More information

Principles of Flowcytometry

Principles of Flowcytometry Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline

More information

Cell Discovery 360: Explore more possibilities.

Cell Discovery 360: Explore more possibilities. Cell Discovery 360: Explore more possibilities. Flexible solutions for the BioPharma laboratory Blood Banking Capillary Electrophoresis Centrifugation Flow Cytometry Genomics Lab Automation Lab Tools Particle

More information

Annexin V-FITC Apoptosis Detection Kit

Annexin V-FITC Apoptosis Detection Kit ab14085 Annexin V-FITC Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Apoptosis in living cells (adherent and suspension). This product is for research

More information

History of Innovation

History of Innovation 1 History of Innovation Only Immucor consistently delivers the next-generation of blood bank automation to labs worldwide ABS HV Galileo NEO 1998 1999 2004 2007 2010 ABS2000 ABS Precis Echo 2 Innovation

More information

The Secret Formula for Webinar Presentations that Work Every Time

The Secret Formula for Webinar Presentations that Work Every Time The Secret Formula for Webinar Presentations that Work Every Time by Gihan Perera www.webinarsmarts.com Sponsored by In an online presentation, your slides aren t visual aids; they ARE the visuals. One

More information

Software Outsourcing - Software Development. [email protected]

Software Outsourcing - Software Development. info@westtownwebservices.com Hi I m Tony Radford from West Town Web Services. We area UK based great value, high quality software development and outsourcing solutions business. If you need software built or looked after please get

More information

DMBI: Data Management for Bio-Imaging.

DMBI: Data Management for Bio-Imaging. DMBI: Data Management for Bio-Imaging. Questionnaire Data Report 1.0 Overview As part of the DMBI project an international meeting was organized around the project to bring together the bio-imaging community

More information

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:

More information

These particles have something in common

These particles have something in common These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative

More information

WEB ANALYTICS Where to Begin

WEB ANALYTICS Where to Begin WEB ANALYTICS Where to Begin e-book SPOILER ALERT! People love the Internet. A lot. In fact, 92% of online adults rely on search engines when looking for information 1. That means your audience s first

More information

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that

More information

BD CellQuest Pro Software Analysis Tutorial

BD CellQuest Pro Software Analysis Tutorial BD CellQuest Pro Analysis Tutorial This tutorial guides you through an analysis example using BD CellQuest Pro software. If you are already familiar with BD CellQuest Pro software on Mac OS 9, refer to

More information

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is

More information

Meso Scale Discovery. WINDOWS is a registered trademark of Microsoft Corporation

Meso Scale Discovery. WINDOWS is a registered trademark of Microsoft Corporation Meso Scale Discovery M S D M S D Meso Scale Discovery, a division of Meso Scale Diagnostics, LLC. (MSD) 9238 Gaither Rd, Gaithersburg, MD 20877 Phone: 240.631.2522 Fax: 240.632.2219 email: [email protected]

More information

Boundary-breaking acoustic focusing cytometry

Boundary-breaking acoustic focusing cytometry Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory

More information

Getting Started in Tinkercad

Getting Started in Tinkercad Getting Started in Tinkercad By Bonnie Roskes, 3DVinci Tinkercad is a fun, easy to use, web-based 3D design application. You don t need any design experience - Tinkercad can be used by anyone. In fact,

More information

Animal Cell Culture. Third Edition. A Practical Approach OXJORD VNIVVRSITY 1'RVSS

Animal Cell Culture. Third Edition. A Practical Approach OXJORD VNIVVRSITY 1'RVSS Animal Cell Culture Third Edition A Practical Approach Edited by John R. W. Masters 3rd Floor Research Laboratories, University College London OXJORD VNIVVRSITY 1'RVSS Contents List of protocols page xiii

More information

Image Analysis Using the Aperio ScanScope

Image Analysis Using the Aperio ScanScope Image Analysis Using the Aperio ScanScope Allen H. Olson, PhD Algorithm Development Engineer Aperio Technologies INTRODUCTION Why should I choose the Aperio ScanScope over competing systems for image analysis?

More information

CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles

CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results

More information