1 Calibration of Electronic Test Equipment: An Introduction to ISO & Accreditation
2 Presentation s Objectives What does Calibration mean? What is ISO 17025, what is Accreditation, and what is the difference? What s good and what s not so good? What does Agilent Italy offer? Page 2
3 Defining Calibration ISO-VIM-1993 International Vocabulary of Basic and General Terms in Metrology: Calibration: a set of operations that establish, under specified conditions, the relationship between values of quantities indicated by a measuring instrument or measuring system, or values represented by a material measure or reference material, and the corresponding values realized by standards. Agilent Technologies / other vendors / Industry required: Calibration: comparing an instrument s performance with respect to the original equipment manufacturer s specifications and full traceability to the appropriate regional Authority. If it is needed, proper adjustments is done. This is applicable and required especially for complex Measurement & Test Equipment (M&TE) Page 3
4 What is ISO 17025? ISO/IEC 17025:2005 is an international standard for calibration and testing labs Requires labs to demonstrate that they: Operate a quality system Processes, documentation and quality management system Generate technically valid results Traceability, equipment, procedures, personnel etc... It takes relevant parts of ISO 9001 and adds the former ISO Guide 25 to form a standard that is focused on the measurement industry Page 4
5 Who does use ISO 17025? All national accreditation bodies have adopted ISO Several industries or countries have incorporated it into their industry-specific or application-specific regulations Some examples ISO9000: par (1994) par.7.6 (2000) via ISO10012 QS-9000: quality standard for the automotive industry AQAP Military Standards OILM regulations Page 5
6 Why this Standard? A global standard to provide consistency Customers need to determine the relative quality and capability of different cal labs Customers can compare the measurement results from different vendors For multi-national companies, they can compare vendors in different countries Customers gain a higher confidence level in the calibrations they purchase International trade is increased by reducing potential quality barriers A third-party process to assess compliance Vendors need to confirm their quality and competence Can conduct an internal audit through their own quality department Can undergo an external audit from each of their major customers Vendors can now schedule a single third-party audit and become accredited to a common set of requirements Page 6
7 What is Accreditation? A formal recognition process where an authoritative body reviews and approves a calibration lab s quality system and technical competence Each country has its own accreditation body A2LA in U.S., UKAS in U.K., JCSS in Japan, SIT in Italy, etc. Mutual recognition via ILAC/EA agreements All accreditation bodies use ISO ISO17025 does not replace ISO9000 Page 7
8 International Standards Quality Management System Requirements: ISO ISO 9000:1994 ISO/IEC 9000:2000 QS ISO/TS 16949:2002 Technical Management System Requirements: ISO Guide 25 ISO Guide 25 ISO/IEC 17025:1999 ISO/IEC 17025:1999 ISO/IEC 17025:2005 ISO/IEC 17025:2005 ISO/IEC & 2:1992 ISO/IEC & 2:1992 ISO/IEC 10012:2003 ISO/IEC 10012:2003 Page 8
9 What is the Scope of Accreditation? Specific measurement parameters are accredited Some small vendors are only accredited for few parameters such as DC & AC voltage, DC & AC current, resistance Some larger vendors are accredited for 15 or more parameters Accredited parameters are documented in a public document, the scope of accreditation Includes approved parameters, ranges, and best measurement capabilities Page 9
10 ISO versus Accreditation Comparing Requirements Accreditation ISO Page 10
11 ISO versus Accreditation Ideal Situation Accreditation ISO ISO and Accreditation requirements are identical Page 11
12 ISO versus Accreditation Typical Single Country Situation Accreditation ISO Accreditation requirements are a super-set of ISO requirements Page 12
13 ISO versus Accreditation Typical Single Country Situation Accreditation ISO Some differences between them Page 13
14 ISO versus Accreditation Current Global Situation Accreditation Accreditation Local differences ISO Accreditation requirements (and sometimes how ISO is interpreted) are slightly different from country-to-country Page 14
15 ISO measurement requirements Validated Procedure As suggested in: ISO17025:2005 par ISO10012:2003 par it can be based on manufacturer performance test procedure. Page 15
16 ISO measurement requirements Measurement Uncertainties Is: Point-by-point measurement uncertainty Per ISO Guide to the Expression of Uncertainty in Measurement (GUM) or EA4/02 guide or UKAS M3003 etc. Measurement uncertainty must be provided for all parameters Is not: general table of best measurement uncertainties Page 16
17 ISO measurement requirements Statement of Compliance If a vendor states that the unit complies with a specification, then the measurement uncertainty must be taken into account Specification can be manufacturer specification as suggested in ISO10012:2003 par and To take into account uncertainty you can follow ILAC-G8 guide or M3003 guide If adjustments are done, PRE & POST results MUST be reported Page 17
18 Accreditation requirements/differences Country-specific requirements Validated Procedure Must be approved by local accreditation body Could be different of manufacturer recommended Measurement Uncertainties Should be the same as the ISO calibration In Italy measurement uncertainty can be provided only for parameters/ranges in the Scope of Accreditation State of Compliance Should be the same as the ISO calibration In Italy not applicable, only to specific standard. Under Accreditation, the Calibration definition is only the VIM Page 18
19 Accreditation requirements/differences Country-specific requirements Often this means limited calibration compared with manufacturer requirements for Complex M&TE Page 19
20 Checklist for Calibration Customers Confirm what level of compliance is required for your industry or country For example, QS-9000 or Military standards requirements are more rigorous Ask yourself the following questions: Which kind of vendor do you need? Complies with ISO 17025, but may not meet all the requirements of the local accreditation body Accredited to ISO17025 Which kind of calibration do you need? Complies with ISO Accredited to ISO o but, is the vendor accredited for all the parameters that your equipment requires? o Are the uncertainties corrects for your equipment requirements? Page 20
21 Four aspects that need to be considered... A lab lab claim to to be be process-compliant. internal internal quality/technical audit audit Self Self assessed A lab lab can can sell sell a conformant (non-accredited) calibration service need need external audit audit A lab lab is is accredited Assessed by by an an Accreditation body body A lab lab can can sell sell a Accredited calibration service Check Check its its scope scope of of accreditation Could Could need need external audit audit Page 21
22 ISO & Accreditation Summary Good Points One standard covers all Internationally recognised Intends to introduce standardisation into the industry Provides users with better quality assurance Ideally suited to a laboratory specialist who knows exactly what he wants Points to ponder Partial testing allowed Still creates confusion Can give false sense of security Not ideal for the person who manages a large volume/range of equipment Not so focused on measurement industries requirements Page 22
23 What does Agilent Italy offer? Since 1964 a Calibration and Repair Service Center in Italy Centro SIT n.05 - In 1979 Agilent s metrological standard laboratory was the first one to be accredited by SIT for electronic quantities Quality System registered to ISO9001:2000 Technical System accredited to ISO17025:2005 It can do Accredited and ISO17025 conformant calibration Page 23
24 Accredited Laboratory Centro SIT 05 Uncertainty Uncertainty 3.5% 3.0% 2.5% 2.0% 1.5% 1.0% 0.5% 0.0% Power Sensor CF Calibration 1 mw-1 GHz 0.7% 1.2% 1.2% 1.5% 1.6% 1.3% 2.4% 3.2% AgilentLab A Lab B Lab C Lab D Lab E Lab F Worst Case Reference Oscillator Calibration Fequency 10MHz 3E-11 3E-11 3E-11 2E-11 2E-11 1E-11 7E-12 7E-12 1E-11 3E-12 7E-12 5E-12 2E-12 3E-13 0E+00 AgilentLab ALab B Lab C Lab DLab E Lab FWorst Case Accredited quantities: DC Voltage, DC Current, DC Resistance, AC Voltage, AC Current, AC Resistance, Capacitance, Inductance, Attenuation, Frequency, Reflection Coefficient, RF Power (& Cal Factor) Coming soon: Optical Power, Optical Attenuation, Optical Wavelenght Best uncertainties on the market Since 1992 has successfully partecipated to more than 20 ILC national (SIT/INRIM) and international (EA/ILAC) Calibration on Agilent & Multivendor Page 24
25 Agilent ISO Conformant Calibration Applicable to Agilent Complex M&TE, i.e. Spectrum, Network, RF & MW Generator, Mobile Test-set etc... Supported list on the Web. Validated Procedure Full parameters following Manufacturer Procedures Measurement Uncertainty All measurements following ISO GUM Traceability Guaranteed by Centro SIT 05, others Accredited Agilent Labs or NMIs. Statement of Compliance Following ILAC-G8 & M3003 Page 25
26 Agilent s treatment of the indeterminate situation A = PASS. B, C = PASS #. Performance falls within the measurement uncertainty on the in-spec side. The customer is notified on the measurement report that: All measured values indicated by this data report lie within specification. Due to measurement uncertainty, the true values of one or more of these points may lie outside of the specification, but there is a high probability that these values lie within specification. A statement of specification compliance is issued. B and C are not adjusted to be closer to centre. Measured values D, E = FAIL. D s Performance falls within measurement uncertainty on the out-of-spec side and is considered a FAIL E is totally out-of-spec with and without uncertainty This pragmatic approach has met with general approval from users Page 26
27 For more information Agilent Metrology Forum General information on calibration and standards List of Agilent service centers/standards labs and their accredited parameters Calibration myths, metrology humor, and other articles of interest Information on the Centro SIT 05 and on the Calibration and Repair Service Center in Italy Page 27
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z540-1-1994
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z540-1-1994 TEKTRONIX, INC. 13725 SW Karl Braun Drive Beaverton, OR 97077 Andrew Mumford Phone: 503 627 2476 CALIBRATION Valid to: June 30, 2018
Guidelines on the Reporting of Compliance with Specification
Guidelines on the Reporting of Compliance with Specification ILAC-G8:03/2009 Copyright ILAC 2009 ILAC encourages the authorized reproduction of this publication, or parts thereof, by organisations wishing
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements
RECOMMENDATION. on the Evaluation of Quality Management System of National Metrology Institutes
EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS COOMET ЕВРО-АЗИАТСКОЕ СОТРУДНИЧЕСТВО ГОСУДАРСТВЕННЫХ МЕТРОЛОГИЧЕСКИХ УЧРЕЖДЕНИЙ КООМЕТ Approved on the COOMET Quality Forum 25 th February 2002
High Level Importance: ISO 17025 Microbiology compliance
High Level Importance: ISO 17025 Microbiology compliance Thamolwan Laoviitayanurak, 3M FSD Professional Service APEC Regional Workshop, BKK Thailand on Aug 25, 2011 3M 2009. All Rights Reserved. Topic
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994. ALLTITE, INC. 1600 Murdock, Wichita, KS 67214 Bruce Demaree Phone: 800 498 6721
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 ALLTITE, INC. 1600 Murdock, Wichita, KS 67214 Bruce Demaree Phone: 800 498 6721 CALIBRATION Valid To: September 30, 2016 Certificate
OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL DOCUMENT OIML D 18 Edition 2008 (E) The use of certified reference materials in fields covered by metrological control exercised by national services of legal metrology. Basic principles
What Are the Differences?
Comparison between the MSA manual and VDA Volume 5 What Are the Differences? MSA is short for Measurement Systems Analysis. This document was first published by the Automotive Industry Action Group (AIAG)
Don t risk it... ...use a UKAS Accredited Laboratory www.ukas.org
Don t risk it... A short introductory guide that explains how yo u can minimise risk by using a UKAS accredited l a b o ra t o r yfor all your testing and calibration needs...use a UKAS Accredited Laboratory
Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK
code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475
ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES
ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.
Standards and accreditation. Tools for delivering better regulation
Standards and accreditation Tools for delivering better regulation Introduction Standards and accreditation are marketbased tools that can be used by Government policy makers to deliver better regulation.
Accreditation and Competence?
Accreditation and Competence? South African National Accreditation System Shadrack Phophi Presentation Layout 1. Brief history of SANAS 2. SANAS scope of activity 3. International recognition 4. Current
American Association for Laboratory Accreditation
Page 1 of 15 R104 GENERAL REQUIREMENTS: ACCREDITATION OF FIELD TESTING AND FIELD CALIBRATION LABORATORIES July 2015 2015 by A2LA All rights reserved. No part of this document may be reproduced in any form
ISO/IEC 17021:2011 Conformity assessment Requirements for bodies providing audit and certification of management systems
ISO/IEC 17021:2011 Conformity assessment Requirements for bodies providing audit and certification of management systems The publication of ISO/IEC 17021:2011 introduces some important new requirements
International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection
International Laboratory Accreditation Cooperation Laboratory Accreditation or ISO 9001 Certification? Laboratory Accreditation or ISO 9001 Certification? When selecting a supplier to fulfill your testing,
The IAF Multilateral Recognition Arrangement (MLA) Certified Once Accepted Everywhere
The IAF Multilateral Recognition Arrangement (MLA) Certified Once Accepted Everywhere Supporting the acceptance of goods and services across national borders Accreditation provides an assurance to government,
1 For more information T: +1.847.664.1499 / E: [email protected] / W: ul.com/dap
DATA ACCEPTANCE PROGRAM UL Client Test Data Program Investigations of products by UL typically involve the testing of product samples. While UL maintains extensive facilities for these purposes, customers
R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES
Page 1 of 7 R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 November 2009 2009 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
Info 15:2 TRAINING 2015/2016. Info 13:19
Info 15:2 TRAINING 2015/2016 Info 13:19 1 2 Info 13:19 S WELCOME TO SWEDAC ACADEMY... 4 TRAINING PROGRAM AND COURSES 2015/2016... 7 1. HOW TO RUN AN ACCREDITATION BODY ISO/IEC 17011... 7 2. LEAD ASSESSOR
Memorandum of Understanding
Memorandum of Understanding between Department for Business, Innovation and Skills and United Kingdom Accreditation Service Page 1 of 13 Contents 1 Purpose... 3 2 Background... 3 3 Scope of activity...
The BIPM key comparison database
The BIPM key comparison database Dr Claudine Thomas BIPM KCDB Coordinator The BIPM key comparison database The KCDB is a public-access website http://www.bipm.org/kcdb built up upon a number of databases
American Association for Laboratory Accreditation
Page 1 of 5 INTRODUCTION ISO/IEC 17011 Conformity Assessment General requirements for accreditation bodies accrediting conformity assessment bodies, the standard by which A2LA operates in order to be recognized
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING. The Expression of Uncertainty in Quantitative Testing
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING Accreditation of Laboratories Publication Reference EAL-G23 The Expression of Uncertainty in Quantitative Testing
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
ipimms is an industry primary impedance measurement service supplied and maintained by UK s National Physical Laboratory. [4]
Page: 1 of 25 LABORATORY LOCATION: (PERMANENT LABORATORY) AGILENT TECHNOLOGIES MICROWAVE PRODUCTS (M) SDN. BHD. BAYAN LEPAS FREE INDUSTRIAL ZONE 11900 PENANG, MALAYSIA This laboratory accredited under
ISO17025 INTERPRETATION DOCUMENT FOR CEC TEST METHODS
COORDINATING EUROPEAN COUNCIL ISO17025 INTERPRETATION DOCUMENT FOR CEC TEST METHODS Prepared by CEC European Accreditation Uniformity Project COORDINATING EUROPEAN COUNCIL FOR THE DEVELOPMENT OF PERFORMANCE
Basic Steps to Establish ISO/IEC 17025 Compliant Testing Laboratories
Workshop for Caribbean countries to promote the development and implementation of Conformity Assessment programmes St. Augustine (Trinidad and Tobago) 2-4 December Basic Steps to Establish ISO/IEC 17025
Accreditation according to ISO/IEC 17025 in Europe. Historic EA Europeing Accreditation Assessment Experience from Technical assessment
Accreditation according to ISO/IEC 17025 in Europe Historic EA Europeing Accreditation Assessment Experience from Technical assessment 1 In year 1803 the foots in Europe had a big differences, depending
Calibration Certificate
P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: [email protected] SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING
National Institute of Justice
U.S. Department of Justice Office of Justice Programs National Institute of Justice National Institute of Justice Law Enforcement and Corrections Standards and Testing Program NIJ Duty Holster Certification
European Type Approval. for. Automotive Systems and Components
European Type Approval for Automotive Systems and Components VCA004 Page 1 of 5 Revision No 11 About VCA The Vehicle Certification Agency (VCA) is an Executive Agency of the United Kingdom Department for
Procedures and General Requirements
NIST HB 150-2016 National Voluntary Laboratory Accreditation Program Procedures and General Requirements Warren R. Merkel Vanda R. White This publication is available free of charge from: http://dx.doi.org/10.6028/nist.hb.150-2016
EMR COMPLIANCE HOW TO ENSURE YOUR NATA ACCREDITED CONTRACTOR IS ABLE TO PROVIDE A NATA ENDORSED REPORT
EMR COMPLIANCE HOW TO ENSURE YOUR NATA ACCREDITED CONTRACTOR IS ABLE TO PROVIDE A NATA ENDORSED REPORT Chris Zombolas Technical Director, EMC Technologies Pty Ltd 57 Assembly Drive Tullamarine Vic. 3043,
How do I gain confidence in an Inspection Body? Do they need ISO 9001 certification or ISO/IEC 17020 accreditation?
What should I look for when I have an Inspection need? 3 How do I gain confidence in an 4 How can accreditation of the inspection body by an ILAC accreditation body member provide confidence? 6 How can
MET/TEAM Test Equipment Asset Management Software
MET/TEAM Test Equipment Asset Management Software NEW, easy-to-use, browser-based calibration asset management software Modular, browser-based calibration asset management solution MET/TEAM software is
Experis Gas Mixtures. When accuracy, stability and reliability are a must
Experis Gas Mixtures When accuracy, stability and reliability are a must Making the Difference For over 70 years, Air Products has enabled customers to become more productive, energy efficient and sustainable.
American Association for Laboratory Accreditation
Page 1 of 18 Traceability for Life Sciences Testing and (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written
Enhanced calibration High quality services from your global instrumentation partner
Products Solutions Services Enhanced calibration High quality services from your global instrumentation partner Services 2 Calibration One trusted advisor Get the best to meet all your critical application
(Uncertainty) 2. How uncertain is your uncertainty budget?
(Uncertainty) 2 How uncertain is your uncertainty budget? Paper Author and Presenter: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc 10219 Coral Reef Way, Indianapolis, IN 46256 Phone: (317) 849 9577,
IMQ RULES. Contents. IMQ Rules - Product certification
IMQ RULES PRODUCT CERTIFICATION "ELECTRIC ENERGY METERS AND METER UNITS" PRODUCT OF NEW MANUFACTURING Contents IMQ Rules - Product certification Particular Rules for the Electric energy meters and meter
ROLE OF THE APEC SPECIALIST REGIONAL BODIES
ROLE OF THE APEC SPECIALIST REGIONAL BODIES ND A S ARD URE D N STA TRUCT E H F T FRAS O S NT NCE IN E M ELE FORMA CON Why is the Standards and Conformance Infrastructure important? ~ Improves everyday
Criminal Justice Offender Tracking System Certification Program Requirements
U.S. Department of Justice Office of Justice Programs National Institute of Justice Criminal Justice Offender Tracking System Certification Program Requirements NIJ CR-1004.00.00 June 2012 NCJ xxxxxx National
Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements
Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements Nick M. Ridler 1 and Nils Nazoa 2 1 National Physical Laboratory, UK (www.npl.co.uk) 2 LA Techniques Ltd,
R221 - Specific Requirements: Forensic
R221 - Specific Requirements: Forensic Page 1 of 6 R221 - Specific Requirements: Forensic (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company
CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM
Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Old Bracknell Lane West Bracknell Berkshire RG12 7AH Contact: Mr M Trotter Tel: +44 (0)1344 459314 Fax: +44 (0)1344 465556 E-Mail: [email protected]
Quality Manual. ISO/TS 16949:2009 Quality Management System. Sensing and Control
Sensing and Control ISO/TS 16949:2009 Quality Management System CORPORATE HEADQUARTERS: OPTEK Technology 1645 Wallace Drive Carrollton, TX 75006 MANUFACTURING: Optron de Mexico Avenue Rio Bravo #1551 Parque
USE OF FIBRE OPTICS INTERNATIONAL STANDARDS FOR CALIBRATION LABORATORY ACCREDITATION INTERNATIONAL ELECTROTECHNICAL COMMISSION
USE OF FIBRE OPTICS INTERNATIONAL STANDARDS FOR CALIBRATION LABORATORY ACCREDITATION INTERNATIONAL ELECTROTECHNICAL COMMISSION USE OF FIBRE OPTICS INTERNATIONAL STANDARDS FOR CALIBRATION LABORATORY ACCREDITATION
Steps to ISO/IEC 17025 Accreditation
Steps to ISO/IEC 17025 Accreditation Perry Johnson Laboratory Accreditation, Inc. 755 W. Big Beaver Rd., Suite 1325 Troy, MI 48084 U.S.A. Phone: 1-877-369-LABS (248) 519-2603 Fax: 248-213-0737 www.pjlabs.com
Customer application package. Included in this package are the following documents. Customer application form
Customer application package Included in this package are the following documents. Customer application form European Type Approval for Automotive Systems and Components Setting out some of the basics
Introduction to EMC Antenna Calibration Methods:
Introduction to EMC Antenna Calibration Methods: An overview of new antenna developments, related standards, calibration and what you need to know for efficient and compliant EMC testing Doug Kramer ETS-Lindgren
Non-Controlled Copy. Quality Manual Revision Log. Rev # Date Description
Quality Manual Revision Log Rev # Date Description 1 Oct 1984 Original Document. 2 Aug 1986 Document Hillsboro Move. Complete Revision. 3 May 1988 Incorporate/Beaverton Move. Reformat to Word Perfect 4.0.
GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E)
GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E) Guidelines for the determination of calibration intervals of measuring instruments Guide pour la détermination
DOCUMENT CS/1: SCHEME DESCRIPTION AND BENEFITS
WELDING FABRICATOR CERTIFICATION SCHEME DOCUMENT CS/1: SCHEME DESCRIPTION AND BENEFITS 6 th Edition July 2012 Issued under the authority of the Governing Board for Certification All correspondence should
2. The Vector Network Analyzer
ECE 584 Laboratory Experiments 2. The Vector Network Analyzer Introduction: In this experiment we will learn to use a Vector Network Analyzer to measure the magnitude and phase of reflection and transmission
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
Radio Spectrum and Technical Standards Advisory Committee
Radio Spectrum and Technical Standards Advisory Committee SSAC Paper 7/2015 for information Update on Testing and Certification of Telecommunications Equipment Office of the Communications Authority 15
Accreditation in Europe
Accreditation in Europe Facilitating regulatory compliance and international trade ACCREDITATION INSPECTION TESTING CALIBRATION EXAMINATION VERIFICATION CERTIFICATION About the EA The EA is appointed by
NVLAP Assessor Training. Assessor Qualification & Training Requirements
NVLAP Assessor Training Assessor Qualification & Training Requirements References ISO/IEC 17011: Conformity assessment General requirements for accreditation bodies accrediting conformity assessment bodies
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
Measuring equipment in a laboratory.
Measuring equipment in a laboratory. Laboratory and metrology are two words that practically do not function separately. They are integrally connected to measuring equipment for tests and measurements.
SADCAS Experiences- Medical Laboratories Accreditation Programme. Presented by. Maureen P Mutasa (Mrs) SADCAS CEO
SADCAS Experiences- Medical Laboratories Accreditation Programme Presented by Maureen P Mutasa (Mrs) SADCAS CEO 1 Presentation Overview Overview of SADCAS The SADCAS MLAP Accreditation process Trends from
CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS 17025.
CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS 17025. Nicholas Mason Metrology Software Project Manager Calibration Group, Fluke Corporation 6920 Seaway Blvd. Everett,
IAF Mandatory Document for the use of Computer Assisted Auditing Techniques ( CAAT ) for Accredited Certification of Management Systems
IAF MD 4:2008 International Accreditation Forum, Inc. IAF Mandatory Document IAF Mandatory Document for the use of Computer Assisted Auditing Techniques ( CAAT ) for Accredited Certification of (IAF MD
Testing Facility Laboratory Manual
Quality Synthetic Rubber's Testing Facility Laboratory Manual Quality Synthetic Rubber, Inc. 1700 Highland Road P.O. Box 1030 Twinsburg, Ohio 44087-9030 Phone: 330-425-8472 Duns Code: 01-753-2458 Z-Code:
ISO/IEC 17025 QUALITY MANUAL
1800 NW 169 th Pl, Beaverton, OR 97006 Revision F Date: 9/18/06 PAGE 1 OF 18 TABLE OF CONTENTS Quality Manual Section Applicable ISO/IEC 17025:2005 clause(s) Page Quality Policy 4.2.2 3 Introduction 4
Recognition of my Product (Testing Laboratory) Why I need recognition, what is the effect?
Page 1 of 23 Accreditation of Testing Laboratories Accreditation of (Sports Surface) Testing Laboratories Konrad Binder (OIST / Austria) ISSS Year 2000 Forum (Schaffhausen) Note: This is a summary of the
IAF Informative Document for the Transition of Food Safety Management System Accreditation to ISO/TS 22003:2013 from ISO/TS 22003:2007
IAF Informative Document IAF Informative Document for the Transition of Food Safety Management System Accreditation to ISO/TS 22003:2013 from ISO/TS 22003:2007 (IAF ID 8:2014) Page 2 of 6 The (IAF) details
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP operates under one quality program, whether you purchase products that are commercial grade, nuclear safety-related or industrial safety compliant (IEC 61508).
3M CORPORATE METROLOGY SERVICES St. Paul, MN
A2LA has accredited THE AMERICAN ASSOCIATION FOR LABORATORY ACCREDITATION ACCREDITED LABORATORY 3M CORPORATE METROLOGY SERVICES St. Paul, MN for technical competence in the field of Calibration This laboratory
Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials
COOMET Recommendation Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials CООМЕТ R/RM/22:2013 Agreed at the 17 th meeting of COOMET
International Accreditation Forum, Inc.
IAF ML 3:2012 IAF MLA Document GUIDANCE FOR RESPONDING TO INQUIRIES ON IAF MLA SIGNATORY EQUIVALENCE AND ON THE ACCEPTANCE OF CERTIFICATION DOCUMENTS (IAF ML 3:2012) SIGNATORY EQUIVALENCE AND ON ACCEPTANCE
Metrology, Quality and ISO 9000
Metrology, Quality and ISO 9000 Application Note What is ISO 9000 is becoming the internationally recognized model for the development of a documented quality system. A quality system is the organizational
BOBS COURSES OVERVIEW
BOBS COURSES OVERVIEW The Botswana Bureau of Standards was established by the Standards Act, 1995. The major function of the Bureau is to promote standardization and quality assurance in industry and commerce
Field Calibration Software
SIGNAL HOUND Field Calibration Software User s Manual Version 1.1.0 7/8/2016 This information is being released into the public domain in accordance with the Export Administration Regulations 15 CFR 734
KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results.
utfärdat av riksmätplats 01 issued by a Swedish National Laboratory Contact person Lars-Erik Josefson 2014-03-27 1 (3) MTv4F006969-K01 Measurement Technology +46 10 516 54 56 [email protected] EcoGuard
Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative
Check-list for auditors. Audit visit at candidate registered laboratories
Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be
Report Of. Shielding Effectiveness Test For. SafeSleeve Radiation Shielding Technology. Test Date(s): June 07 June 09, 2015
Report Of Shielding Effectiveness Test For SafeSleeve Radiation Shielding Technology Test Date(s): June 07 June 09, 2015 Issue Date: June 10, 2015 UST Project No: 15-0137 Total Number of Pages Contained
Lab Service, Asset Management and Extending the Life of Your Lab Equipment. Scott Greenwood Director, Services
Lab Service, Asset Management and Extending the Life of Your Lab Equipment Scott Greenwood Director, Services PLEASE STOP ME AT ANY TIME TO ASK QUESTIONS OR MAKE A COMMENT Is this a business manager...
EA-7/01. EA Guidelines. on the application. Of EN 45012. Publication Reference PURPOSE
Publication Reference EA-7/01 EA Guidelines on the application Of EN 45012 PURPOSE The purpose of the document is to provide explanations with a view to harmonise the application of ISO/IEC Guide 62/EN
Quality Management Systems for Seed Testing Laboratories: Presented to the 2010 CSAAC Meeting. Valerie Martz Senior Laboratory Accreditation Officer
Quality Management Systems for Seed Testing Laboratories: Presented to the 2010 CSAAC Meeting Valerie Martz Senior Laboratory Accreditation Officer Outline 1. Quality Management Systems (QMS): Generic
Derbyshire Trading Standards Service Quality Manual
Derbyshire Trading Standards Service Quality Manual This Quality Manual has been developed to give a broad outline of how the Trading Standards Division s range of services comply with the requirements
Fire Protection Industry Scheme Reference SP205 Part 1
BAFE Scheme: SP205 Version 4(a): July 2015 Fire Protection Industry Scheme Reference SP205 Part 1 Life Safety Fire Risk Assessment BAFE (British Approvals for Fire Equipment) The Fire Service College,
Chain of Custody Standard
Responsible Supply of Fishmeal and Fish Oil Chain of Custody Standard A Tool for Voluntary Use in Markets for Products of Fishmeal and Fish oil Contents Page A Foreword... 3 B Principles of the Process.
RFID Design Principles
RFID Design Principles Harvey Lehpamer ARTECH HOUSE BOSTON LONDON artechhouse.com Contents Introduction 2 2.1 2.1.1 2.1.2 2.1. 2.1.4 2.2 2.2.1 2.2.2 2. 2..1 2..2 2.4 2.4.1 2.4.2 2.5 2.5.1 2.5.2 Comparison
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
The Calibration Chain: Role of BIPM,
The Calibration Chain: Role of BIPM, PSDLs and ADCLs J. Seuntjens and M. McEwen McGill University & National Research Council CANADA Clinical Dosimetry for Radiotherapy, 2009 AAPM Summer School PSDL ADCL
SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION
1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.
RG 7 Accreditation for Inspection Bodies Performing Non-Destructive Testing
This publication contains policy, recommendations and guidance applicable to UKAS accredited inspection bodies RG 7 Accreditation for Inspection Bodies Performing Non-Destructive Testing Contents Section
Selection and use of ISO 9000
Selection and use of ISO 9000 ISO in brief ISO is the International Organization for Standardization. It is made up of national standards institutes from countries large and small, industrialized and developing,
CHAPTER 1 Introduction 1
Contents CHAPTER 1 Introduction 1 CHAPTER 2 Short-Range Communications Systems 3 2.1 Radio-Frequency Spectrum and Propagation 3 2.1.1 Theory of Electromagnetism and Maxwell s Equations 3 2.1.2 RF Propagation
How DCMA Helps To Ensure Good Measurements
How DCMA Helps To Ensure Good Measurements Speaker/Author: Robert Field Defense Contract Management Agency 605 Stewart Avenue Garden City, New York 11530 Email: [email protected] Phone: (516) 228-5886;
Guidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC. GNB-CPD position paper from SG17 - EN 1090-1:2009
GNB-CPD SG17 Guidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC NB-CPD/SG17/09/069 Issued: 7 September 2009 APPROVED GUIDANCE GNB-CPD position paper from SG17
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES Reviewed by: Natalija Jovičić Zarić MSc, Quality Manager Approved by: Milica Lukešević, Acting Director Any