CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS
|
|
|
- Willis Maxwell
- 9 years ago
- Views:
Transcription
1 CALIBRATION DATA MANAGEMENT: MEETING THE REPORTING REQUIREMENTS OF ISO/IEC FDIS Nicholas Mason Metrology Software Project Manager Calibration Group, Fluke Corporation 6920 Seaway Blvd. Everett, WA Abstract - ISO Guide 25, General Requirements for the Competence of Testing and Calibration Laboratories requires the collection and reporting of various information about instruments calibrated by a metrology laboratory. Many Metrology Laboratories are using personal computers (PCs) connected to increase productivity and collect and report information required by the standards. This paper discuses PC based tools to assist in the collection and reporting calibration information required by ISO17025 such as estimates of measurement uncertainty, calibration certificates, and calibration reports. Shown are examples using commercial Calibration and Asset Management software. INTRODUCTION This paper is about collecting and reporting quality information using computer software in the calibration lab. This task could be accomplished with paper and pencil, but the advantages of automating data collection and reporting have been previously documented [1]. It is based on examining the data collection and report requirements of ISO/IEC FDIS 17025: General Requirements for the Competence of Testing and Calibration Laboratories. Calibration laboratories that comply with ISO/IEC17025 also comply with ISO 9001:1994 or ISO 9002:1994[2]. SECTION GENERAL DOCUMENT CONTROL The first references to software we find are in section Document Control General The laboratory shall establish and maintain procedures to control all documents that form part of its quality system (internally generated or from external sources) such as regulations, software, specifications, instructions, and manuals[2]. From this section we see that the standard considers software a document to be controlled. Note 2 refers us to section and 4.12 for additional information. Continuing on, section states:
2 All documents issued to personal in the laboratory as part of the quality system shall be reviewed and approved for use by authorized personnel prior to issue [2]. Since software is considered to be a document, this means that all software must go through a review and approval process before it can be used in a laboratory. This review process is straightforward for a primary software package such as Fluke Metrology Software or other packages. The laboratory compares features between different packages and examines sample software available from the manufacturer. The laboratory can also consult other users of the software packages or attend user group meetings to see what kinds of issues arise for users of each package. Laboratory personnel can then make a decision on the suitability of a particular software package. However, "calibration" software typically allows the user to create calibration procedures or scripts, or uses external files, such as accuracy files, to provide information to the software. These procedures and external files are often written by the laboratory, or are provided "as is" by the manufacturer. For a lab to comply with ISO 17025, these procedures and files must be reviewed and approved before they can be used to for calibration. This topic is discussed in more detail in section 5.4.7, Control of Data. The remainder of section 4.3 states that documents (software) should be periodically reviewed to make sure they are still suitable, that the lab has procedures in place to ensure that personnel are using the correct version, and that change control procedures are in place [2]. SECTION 4.9 CONTROL OF NONCONFORMING CALIBRATION WORK Section discusses the need to control nonconforming calibrations and states where necessary, the client is notified and work is recalled [2]." Suppose a standard is found out-of-calibration? It would be nice to automatically print notices to all users whose calibration may have been affected. This requires a reverse traceability report to identify the affected equipment. Reports can be performed on the PC using off-the-shelf software. If the database has an open architecture then a variety of tools can be used to query the database and print these reports. Fluke Metrology Software is Microsoft Open Database Connectivity (ODBC) compliant [3]. This means reports can be made using tools like Microsoft Word, Excel, Borland ReportSmith, Crystal Reports, or many other tools. Fluke Metrology Software uses Crystal Reports as its reporting mechanism [3]. The report examples shown in this document are Crystal Reports and were printed by the Fluke Metrology Software Manual Entry application. An example of a reverse traceability report is shown in figure 1. Figure 1 is the top of a simple reverse traceability report for instrument R4. Each level of instrumentation is indented to show the next level of calibration. For example, instrument R4 directly calibrated instrument 752, while 752 calibrated The calibration date is shown for each instrument. This allows for the retrieval of detailed information about an instrument.
3 Figure 1. Reverse Traceability Report for Asset R4. SECTION 4.12: CONTROL OF RECORDS The next section of interest states: The laboratory shall have procedures to protect and back-up records stored electronically and to prevent unauthorized access to or amendment of these records [2]. In no uncertain terms, this means that the laboratory MUST have a documented procedure to back-up the data on their computer system. A very good backup system is described in chapter 2 of the Fluke Metrology Software Users Manual [3]. In addition, the software system must prevent access or modification of the data by unauthorized users. This is easily accomplished by using Fluke Metrology Software, because it uses an SQL Client/Server Database. In the client server environment, only the data engine can read and write to the data file. The data file is typically placed on a computer that users do not have access to. The data engine controls all access to the data by requiring users to login using a username and password. Security enforcement is compromised in a flat-file database by auxiliary programs that can be used to directly edit the data file and change calibration information. SECTION ESTIMATION OF UNCERTAINTY OF MEASUREMENT We now move on to section 5, technical requirements. Section 5.4.6, Estimation of uncertainty of measurement requires that all laboratories analyze measurement uncertainty A calibration laboratory, or a testing laboratory performing its own calibrations, shall have and apply a procedure to estimate the uncertainty of measurement for all calibrations and types of calibrations When estimating the uncertainty of measurement, all uncertainty components which are of importance in the given situation, shall be taken into account using the appropriate methods of analysis [2]. Fluke Metrology Software version 6.0 product can provide this capability required by section An overview of how Fluke Metrology Software handles measurement uncertainty is discussed here.
4 At the top level, the measurement uncertainty calculation is simply: Expanded Uncertainty = EU = K * SU where K is the coverage factor, and SU is the Standard Uncertainty. The Standard Uncertainty is: SU = U U 2 + U U 10 U1 is the Normalized System Accuracy, based on the accuracy of the calibration standard and is determined automatically by the software. U2 is uncertainty of the UUT based on a sequence of actual measurements and the resolution of the UUT and is determined automatically by the software. U3, U4, U5, U6, U7, U8, U9, and U10 are optional uncertainty components that may be specified by the laboratory. If not specified, they default to zero and do not affect the uncertainty calculation. For a complete discussion of how Fluke Metrology Software provides measurement uncertainty capability, see "Implementing ISO Measurement Uncertainty Requirements in Software". 4 Most parameters used in the measurement uncertainty calculation can be overridden at the procedure level. SECTION CONTROL OF DATA Paragraph Control of Data states: Where computers or automated equipment are used for the acquisition, processing, recording, reporting, storage or retrieval of test or calibration data, the laboratory shall ensure that: 2 a) computer software developed by the user is documented in sufficient detail and suitably validated as being adequate for use; b) procedures are established and implemented for protecting the data; such procedures shall include, but not be limited to, integrity and confidentiality of data entry or collection, data storage, data transmission, and data processing; c) computer and automated equipment are maintained to ensure proper functioning and are provided with the environmental and operating conditions necessary to maintain the integrity of test and calibration data [2]. Subparagraph c is straightforward. Computer equipment must be in good working order and used within the environmental conditions to which it was designed. Subparagraph b discusses how data is handled. Part of this paragraph is very similar to section discussed earlier, and essentially means that you should have a good backup system. You should also have a computer system that allows you to control user access, who can change the data, and keep out unauthorized users. According to subparagraph a, any software written by the lab needs to be validated that it is suited to the task. At a minimum, this requires user documentation, a test plan, and proof that the test plan was executed. The note for section is useful, stating that Commercial off-the-shelf software in general use within its designed application range may be considered sufficiently validated [2]. This means that programs like Fluke Metrology Software may be used with their default
5 configuration without any additional validation work. It is important to read the rest of the note However, laboratory software configuration/modifications should be validated as in a [2]. Many of the software packages today, including Fluke Metrology Software, are highly customizable. This means that if the laboratory changes the product from its default configuration, it needs to validate that the changes from the default configuration work as expected, and that the changes do not adversely affect the adequacy of the software. One final comment on user developed software is the point previously discussed about calibration procedures and external files. Commercial calibration software allows the laboratory to write calibration procedures or scripts or to modify many of the input data sets for the software [3]. These procedures and modifications are considered to be software. Thus, for a lab to comply with ISO 17025, these procedures must also be documented and validated before they can be used for calibration. SECTION 5.5 EQUIPMENT Paragraphs and address data collection requirements. These paragraphs are: Each item of equipment and its software used for testing and calibration and significant to the result shall, when practicable, be uniquely identified Records shall be maintained of each item of equipment and its software significant to the tests and/or calibrations performed. The records shall include at least the following: a) identity of the item of equipment and its software; b) manufacturer s name, type identification, and serial number or other unique identification; c) checks that equipment complies with the specification (see 5.5.2); d) current location, where appropriate; e) the manufacturer s instructions, if available, or reference to their location; f) dates, results, and copies of reports and certificates of all calibrations, adjustments, acceptance criteria; and due date of next calibration; g) maintenance plan, where appropriate, and maintenance carried out to date; h) damage, malfunction, modification or repair to the equipment [2]. These paragraphs list specific data elements the database must maintain. These elements are divided into 4 types of information: Inventory, calibration, location, and maintenance information. Inventory information is information about an instrument or asset that does not change over time. These are listed in paragraph 5.5.4, a, b, e, and part of g. Information about where the asset is, or location information is called out in d c, and f requires calibration information. Repair and maintenance information is called out in g and h. Therefore, we need to gather inventory data, and calibration, location, and repair data over time to satisfy ISO
6 The Fluke Metrology Software product has been designed to gather these required data elements. It can also gather other information, since we may want information in addition to that required by the standard. Figure 2 shows an example Inventory data entry form. The title bar on the form shows this is the inventory form, that the user is changing asset number SAMPLE-11. Note that the date is shown using decimals for separators, allowing customization for local configurations in different countries. Linked validation is superior to having a validation list for each field. First, it reduces the amount of entry the user has to make, but more importantly it guarantees that the data makes sense. For example, 5700 may be a valid model, and GAGE BLOCK may be a valid description, but they do not make sense if these values both appear for the same instrument. The linking and validation values of fields are under user control. Figure 2. Inventory Manual Data Entry Form. The data elements required by ISO/IEC are present. Of special interest is the "validation" box that is open on the model field where only models listed in the box are allowed as valid entries in the field. In addition, there can be linked fields. For example, we could link the model to the manufacturer and description fields. This would mean that if the operator enters 3458, the manufacturer field will be automatically filled in with AGILENT and the description field will be automatically filled in with DIGITAL MULTIMETER. Figure 3 shows an example of Calibration data entry form. The form contains the date and time of calibration, its pass/fail status, when it is due, the standards used, and many other pieces of information.
7 Figure 3. Calibration Data Entry Forms. Similar forms are available for calibration results, location, repair and customer information. This covers data collecting but what does the standard say about data reporting? There are several sections that discuss reporting. SECTION 5.6 MEASUREMENT TRACEABILITY Section requires the lab to have an established programme and procedure for the calibration of its equipment" [2]. The note for this section indicates that the program should include a system for selecting and maintaining standards and equipment used to perform calibrations. One aspect of such a program is the periodic recall of instruments when they are due for calibration. It would be useful if a software system could print some kind of recall for calibration report. Figure 4 is a partial example Due for Calibration Report. This report identifies everything due and overdue for calibration. Overdue items are shown with a * character in the left column.
8 Figure 4. Calibration Due and Overdue Report. Section states that calibration laboratories shall ensure that measurements are traceable by an unbroken chain of calibrations or comparisons linking them to relevant primary standards [2]. One way to ensure this traceability is a forward traceability report. Figure 5 is an example Forward Traceability Report. It shows an unbroken chain from asset 8842A to the local standards that were sent to the vendor for calibration. As with the reverse traceability report, each level of instrumentation is indented. Note that this report has identified two circular calibrations for this asset, a serious problem. One of the circular calibrations is an endless calibration, meaning one instrument in the chain points to a calibration event already in the chain. This is shown by asset calibrating itself using the calibration event of 5/1/95 both times. The other circular calibration is not quite as serious, as two different calibration events are in this chain. Asset 5700 (caldate 4/30/95) is calibrating itself using a different calibration event on 4/10/95.
9 Figure 5 Forward Traceability Report SECTION 5.10 REPORTING THE RESULTS Paragraph Test reports and calibration certificates, lists 11 (a through k) items that should be found on a calibration certificate. Each test report or calibration certificate shall include at least the following information, unless the laboratory has valid reasons for not doing so: a) a title, (e.g. Test Report or Calibration Certificate); b) name and address of the laboratory, and location where the tests and/or calibrations were carried out if different from the address of the laboratory; c) unique identification of the test report or calibration certificate (such as serial number) and on each page an identification in order to ensure that the page is recognized as part of the test report or calibration certificate, and a clear identification of the end of the test report or calibration certificate; d) name and address of the client; e) identification of the method used; f) description, condition, and unambiguous identification of the item(s) tested or calibrated; g) date of receipt of test or calibration item(s) where critical to the validity and application of the results and date(s) of performance of the test or calibration; h) reference to sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results; i )test or calibration results with, where appropriate, the units of measurement;
10 j) the name(s), function(s), and signature(s) or equivalent identification of person(s) authorizing the test report or calibration certificate; k) where relevant, a statement to the effect that the results relate only to the items tested or calibrated In addition to the requirements listed in , calibration certificates shall include the following, where necessary for the interpretation of calibration results: a) the conditions (e.g. environmental) under which the calibrations where made that have an influence on the measurement results; b) the uncertainty of measurement and/or a statement of compliance with an identified metrological specification or clauses thereof; c) evidence that the measurements are traceable (see note 1) [2]. Figure 6 shows a calibration certificate generated by Fluke Metrology Software for asset sample-11. The items called out in paragraph 5.10 are visible, such as the title, identification of the instrument, name and address of the client and laboratory, calibration results with measurement uncertainty, and evidence that the measurements are traceable.
11 Figure 6. Calibration Results Report.
12 CONCLUSIONS This paper has examined the requirements for data collection and reporting as described by ISO Inventory, Customer, calibration history, location history, and maintenance history data need to be collected about instruments calibrated and used for calibration. In addition, the calibration results must report measurement uncertainty. [4] Nicholas, Matt, Implementing ISO Measurement Uncertainty Requirements in Software. NCSL Conference, Charlotte, SC, Four different types of reports are necessary to satisfy the standard. We may need more reports for other purposes. These reports are: Report Section notes Calibration 5.10 Need all data. certificate Due for calibration Determine instruments that have exceeded calibration interval. reverse traceability report Forward traceability report Identify all the instruments that are affected Show unbroken chain of traceability from an instrument to national standards. These data collection and reporting requirements can be performed by the Fluke Metrology Software package. This package will assist your lab in meeting the requirements of ISO/IEC FDIS References. [1]Mason, Nicholas, Software Strategies for On-Site Calibration. Measurement Science Conference: 1999 Workshop and Symposium, MSC, Anaheim, CA, [2] ISO/IEC FDIS General Requirements for the Competence of Testing and Calibration Laboratories, ISO/IEC, April 10, [3] Metrology Software Users Manual, Fluke Corporation, Everett, WA, July 1999.
Quality Management System Policy Manual
Quality Management System Burns Engineering, Inc. 10201 Bren Road East Minnetonka, MN 55343 4. 4.1 General Requirements The Quality Manual consists of the quality system policy documents and the quality
MET/TEAM Test Equipment Asset Management Software
MET/TEAM Test Equipment Asset Management Software NEW, easy-to-use, browser-based calibration asset management software Modular, browser-based calibration asset management solution MET/TEAM software is
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
Tools to Aid in 21 CFR Part 11 Compliance with EZChrom Elite Chromatography Data System. White Paper. By Frank Tontala
Tools to Aid in 21 CFR Part 11 Compliance with EZChrom Elite Chromatography Data System White Paper By Frank Tontala Agilent Technologies Software & Informatics Life Sciences & Chemical Analysis Group
Metrology, Quality and ISO 9000
Metrology, Quality and ISO 9000 Application Note What is ISO 9000 is becoming the internationally recognized model for the development of a documented quality system. A quality system is the organizational
Software Verification and Validation
Software Verification and Validation Georgia L. Harris Carol Hockert NIST Office of Weights and Measures 1 Learning Objectives After this session, using resources and references provided, you will be able
ISO/IEC 17025 QUALITY MANUAL
1800 NW 169 th Pl, Beaverton, OR 97006 Revision F Date: 9/18/06 PAGE 1 OF 18 TABLE OF CONTENTS Quality Manual Section Applicable ISO/IEC 17025:2005 clause(s) Page Quality Policy 4.2.2 3 Introduction 4
OPERATIONAL STANDARD
1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.
July 12, 2013 Page 1 of 5 BellHawk Systems Corporation
BellHawk Compliance with CFR 21 Part 11 Introduction This document details the compliance of the BellHawk software with CFR 21 Part 11 (Part 11) dated March 20, 1997 and the document General Principles
4Sight Calibration Management Software
GE Measurement & Control Solutions 4Sight Calibration Management Software 4Sight calibration and maintenance management software provides visibility to the assets, data, and resources that affect maintenance,
unless the manufacturer upgrades the firmware, whereas the effort is repeated.
Software Validation in Accredited Laboratories A Practical Guide Gregory D. Gogates Fasor Inc., 3101 Skippack Pike, Lansdale, Pennsylvania 19446-5864 USA [email protected] www.fasor.com Abstract Software
EffiValidation 3.0 software basic training module
EffiValidation 3.0 software basic training module Oct 3, 2012 www.effichem.com Agenda EffiChem company EffiValidation 3.0 software : overview International standards and guidelines Software features Structure
The Impact of 21 CFR Part 11 on Product Development
The Impact of 21 CFR Part 11 on Product Development Product development has become an increasingly critical factor in highly-regulated life sciences industries. Biotechnology, medical device, and pharmaceutical
Automated Calibration Software Guardbanding. Author: Matt Nicholas Fluke Corporation. Speaker: Lars Andersson Fluke Europe B.V
Automated Calibration Software Guardbanding Author: Matt Nicholas Fluke Corporation Speaker: Lars Andersson Fluke Europe B.V 1 Matt Nicholas Automated Calibration Software Guardbanding July 06, 2004 ABSTRACT
Company Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature
Far West Technology, Inc. ISO 9001 Quality Manual Document No.: QM Revision: 0 Issue Date: 27 August 1997 Approval Signatures President/CEO Executive Vice President Vice President/CFO Change Record Rev
Translation Service Provider according to ISO 17100
www.lics-certification.org Certification Scheme S06 Translation Service Provider according to ISO 17100 Date of issue: V2.0, 2015-11-15 Austrian Standards plus GmbH Dr. Peter Jonas Heinestraße 38 1020
A ChemoMetec A/S White Paper September 2013
NucleoCounter NC-3000, NucleoView NC-3000 Software and Code of Federal Regulation 21 Part 11; Electronic Records, Electronic Signatures (21 CFR Part 11) A ChemoMetec A/S White Paper September 2013 ChemoMetec
Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA 92029-1337 USA
Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13 Pi Tape Corporation 344 North Vinewood St. Escondido CA 92029-1337 USA Phone # 760 746-9830 Fax # 760 746-9196 In conformance
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements
PART 10 COMPUTER SYSTEMS
PART 10 COMPUTER SYSTEMS 10-1 PART 10 COMPUTER SYSTEMS The following is a general outline of steps to follow when contemplating the purchase of data processing hardware and/or software. The State Board
Warehouse R x Inventory Management Software. Technical Overview
Warehouse R x Inventory Management Software Technical Overview January 19, 2009 System Overview Warehouse R X is the latest version of Daifuku America s Warehouse Control System (WCS) software. It provides
State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist
This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to
Global Value 7. Productivity Suite for GammaVision. Optimizing Gamma Spectrometry Processes through Secure Data Management and Measurement Automation.
Productivity Suite for GammaVision Optimizing Gamma Spectrometry Processes through Secure Data Management and Measurement Automation. Global Value was designed to transform the standard GammaVision spectroscopy
TIBCO Spotfire and S+ Product Family
TIBCO Spotfire and S+ Product Family Compliance with 21 CFR Part 11, GxP and Related Software Validation Issues The Code of Federal Regulations Title 21 Part 11 is a significant regulatory requirement
The Fulfillment of AS 9100 Rev C Requirements by EnterpriseIQ
The Fulfillment of AS 9100 Rev C Requirements by EnterpriseIQ Manufacturers have worked to meet the stringent requirements of AS9100 for decades; the recent introduction of Revision C has further expanded
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company
CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM
NIST HANDBOOK 150 CHECKLIST
NIST HANDBOOK 150 CHECKLIST Instructions to the Assessor: This checklist addresses the general accreditation criteria prescribed in NIST Handbook 150, NVLAP Procedures and General Requirements (2006 edition).
Full Compliance Contents
Full Compliance for and EU Annex 11 With the regulation support of Contents 1. Introduction 2 2. The regulations 2 3. FDA 3 Subpart B Electronic records 3 Subpart C Electronic Signatures 9 4. EU GMP Annex
FDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry
FDA 21 CFR Part 11 Electronic records and signatures solutions for the Life Sciences Industry The Rule 21 CFR Part 11 Handwritten signature means the scripted name or legal mark of an individual handwritten
August 2012 Web Application for Enprise Job Costing. 1.0.2 1.2.3863 SAP 2007A, SAP 2007A SP01, SAP 8.8 All regions
August 2012 Document Version: Product Version: SAP Version: Localization: 1.0.2 1.2.3863 SAP 2007A, SAP 2007A SP01, SAP 8.8 All regions 2012 Enprise Solutions All rights reserved. No parts of this work
Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
21 CFR Part 11 Implementation Spectrum ES
21 CFR Part 11 Implementation Spectrum ES INFRARED SPECTROSCOPY T E C H N I C A L N O T E Introduction Compliance with 21 CFR Part 11 is mandatory for pharmaceutical companies and their suppliers to sell
Bogue Machine Company, Inc. Business Instruction
Bogue Machine Company, Inc. Business Instruction Document Title: Supplier Quality Control Requirements Document Number: Revision: Document Owner: Mark Bogue Backup Owner: Craig Bogue Parent Document: PURP-004,
City of Madison. Information Services. Crystal Enterprise Polices, Standards, and Guidelines
City of Madison Information Services Crystal Enterprise Polices, Standards, and Guidelines March 2006 City of Madison Crystal Enterprise Policies, Standards, and Guidelines Table of Contents Crystal Enterprise
Eagle Machining, Inc. Quality Management System
Eagle Machining, Inc. Quality Management System 1 of 10310 Antoine Drive Bldg D, Houston, Texas 77086 BUSINESS OPERATING MANUAL (QUALITY MANUAL) Revision Date: 08/01/2014 Approved By: Joseph Vu Date: 08/01/2014
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES October 2008 Revision 08 Great Lakes Institute for Environmental Research University of Windsor Windsor, Ontario
Measuring equipment in a laboratory.
Measuring equipment in a laboratory. Laboratory and metrology are two words that practically do not function separately. They are integrally connected to measuring equipment for tests and measurements.
Calibration Control. Calibration Management Software. Tools for Management Systems
Calibration Control Calibration Management Software Calibration Control CREATED FROM THE MANUFACTURING USER S PERSPECTIVE MAJOR FEATURES WINDOWS 7 & 8 21 CFR PART 11 COMPLIANT MS ACCESS OR SQL SERVER DANISH,
Request for Proposal. Contract Management Software
Request for Proposal Contract Management Software Ogden City Information Technology Division RETURN TO: Ogden City Purchasing Agent 2549 Washington Blvd., Suite 510 Ogden, Utah 84401 Attn: Sandy Poll 1
B.Com(Computers) II Year DATABASE MANAGEMENT SYSTEM UNIT- V
B.Com(Computers) II Year DATABASE MANAGEMENT SYSTEM UNIT- V 1 1) What is Distributed Database? A) A database that is distributed among a network of geographically separated locations. A distributed database
Office of the Auditor General Performance Audit Report. Statewide Oracle Database Controls Department of Technology, Management, and Budget
Office of the Auditor General Performance Audit Report Statewide Oracle Database Controls Department of Technology, Management, and Budget March 2015 071-0565-14 State of Michigan Auditor General Doug
Planning the Installation and Installing SQL Server
Chapter 2 Planning the Installation and Installing SQL Server In This Chapter c SQL Server Editions c Planning Phase c Installing SQL Server 22 Microsoft SQL Server 2012: A Beginner s Guide This chapter
Systems Software. Introduction to Information System Components. Chapter 1 Part 2 of 4 CA M S Mehta, FCA
Systems Software Introduction to Information System Components Chapter 1 Part 2 of 4 CA M S Mehta, FCA 1 Systems Software Learning Objects Task Statements 1.1 Identify deployment of different components
Agilent MicroLab Software with Spectroscopy Configuration Manager and Spectroscopy Database Administrator (SCM/SDA)
Agilent MicroLab Software with Spectroscopy Configuration Manager and Spectroscopy Database Administrator (SCM/SDA) Compliance with 21 CFR Part 11 Introduction Part 11 in Title 21 of the Code of Federal
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
Testing Facility Laboratory Manual
Quality Synthetic Rubber's Testing Facility Laboratory Manual Quality Synthetic Rubber, Inc. 1700 Highland Road P.O. Box 1030 Twinsburg, Ohio 44087-9030 Phone: 330-425-8472 Duns Code: 01-753-2458 Z-Code:
NABL NATIONAL ACCREDITATION
NABL 160 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDE for PREPARING A QUALITY MANUAL ISSUE NO. : 05 AMENDMENT NO : 00 ISSUE DATE: 27.06.2012 AMENDMENT DATE: -- Amendment
InfoCenter Suite and the FDA s 21 CFR part 11 Electronic Records; Electronic Signatures
InfoCenter Suite and the FDA s 21 CFR part 11 Electronic Records; Electronic Signatures Overview One of the most popular applications of InfoCenter Suite is to help FDA regulated companies comply with
Access Control and Audit Trail Software
Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA Access Control and Audit Trail Software Operation Manual Varian, Inc. 2002 03-914941-00:3 Table of Contents Introduction... 1 Access Control
INFORMATION SYSTEMS SPECIALIST 8 1488
INFORMATION SYSTEMS SPECIALIST 8 1488 SERIES DESCRIPTION The INFORMATION SYSTEMS SPECIALIST (ISS) classification series has eight levels that describe technical and professional non-supervisory positions
Database Assistant. Once Database Assistant is installed you must login to gain access to the database. Copyright 2009
TOSHIBA Strata CIX Technical Bulletin March 2008 is a companion package to Voice Assistant. provides an interface between any Open Database Connectivity (ODBC) compliant database and Voice Assistant. This
ICT USER ACCOUNT MANAGEMENT POLICY
ICT USER ACCOUNT MANAGEMENT POLICY Version Control Version Date Author(s) Details 1.1 23/03/2015 Yaw New Policy ICT User Account Management Policy 2 Contents 1. Preamble... 4 2. Terms and definitions...
Introduction. Editions
Introduction TRAIN TRACK Employee Training Management Software gives you the tools you need to make sure employee training requirements are met. Assign required training for groups or individuals. Easily
EUROLAB Cook Book Doc No. 13 ELECTRONIC RECORDS
ELECTRONIC RECORDS Introduction The increased use of computers and computer systems in laboratories have lead to an increased number of electronic records. There are a lot of advantages with electronic
FMC Technologies Measurement Solutions Inc.
Table of Contents 1 Revision History... 4 2 Company... 5 2.1 Introduction... 5 2.2 Quality Policy... 6 2.3 Quality Organization... 6 2.4 Company Processes... 7 3 Scope and Exclusions... 8 3.1 Scope...
Guidelines on the Reporting of Compliance with Specification
Guidelines on the Reporting of Compliance with Specification ILAC-G8:03/2009 Copyright ILAC 2009 ILAC encourages the authorized reproduction of this publication, or parts thereof, by organisations wishing
Quality Management System Manual Revision L
This Page 1 of 35 of the Quality Management System Manual If issued as a controlled copy, the serial number of this copy is Quality Management System Manual Certified to AS9100 Revision C Printed copies
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
Continuous Monitoring Data Acquisition System
Continuous Monitoring Data Acquisition System Finally, a company that offers a cost effective solution for my data acquisition system requirements Data management is vital to your company s bottom line
ATTACHMENT III Tender No. 10000265-HD-48009. Laboratory Information Management System (LIMS)
Objective: ATTACHMENT III Tender No. 10000265-HD-48009 Laboratory Information Management System (LIMS) HPCL Mumbai refinery is looking for LIMS system to automate the Refinery Laboratory operation and
GE Power Electronics Business Total Quality Management for Suppliers
GE Power Electronics Business Total Quality Author: Reviewers: Contributors: Alessandro A. Cervone Technical Manager Component & Reliability Engineering Alicia Scott Title Erika Nannis Title Robert Roessler
COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.
Revision History COPY The top row of this table shows the most recent changes to this controlled document. For previous revision history information, archived versions of this document are maintained by
How To Ensure Quality Assurance
1001 Robb Hill Road Oakdale, PA 15071-3200 412/923-2670 PROCUREMENT SPECIFICATION: SUPPLIER QUALITY MANAGEMENT SYSTEM REQUIREMENTS Specification Number: ATI-QA-703, Rev 4 Date: 11/12/13 Page: 1 of 7 Written
SWIFT SILLIKER - LABWARE LIMS AND ISO 17025 ACCREDITATION Paul Roberts
SWIFT SILLIKER - LABWARE LIMS AND ISO 17025 ACCREDITATION Paul Roberts Contents About Swift Silliker a Mérieux NutriSciences Company. LabWare LIMS compliance with SANAS & ISO 17025 requirements. Location
Complete fleet management package for Qualitrol Information Products
+ iq Master Station Configuration, collection and analysis platform Complete fleet management package for Qualitrol Information Products Scalable client-server solution Advanced Record Analysis diagnostics
05.0 Application Development
Number 5.0 Policy Owner Information Security and Technology Policy Application Development Effective 01/01/2014 Last Revision 12/30/2013 Department of Innovation and Technology 5. Application Development
Enhanced calibration High quality services from your global instrumentation partner
Products Solutions Services Enhanced calibration High quality services from your global instrumentation partner Services 2 Calibration One trusted advisor Get the best to meet all your critical application
R214 SPECIFIC REQUIREMENTS: INFORMATION TECHNOLOGY TESTING LABORATORY ACCREDITATION PROGRAM
The American Association for Laboratory Accreditation Document Revised: R214: Specific Requirements: Information Technology Testing Laboratory Accreditation July 13, 2010 Program Page 1 of 26 R214 SPECIFIC
What s New in Microsoft Dynamics GP 2015 R2. Microsoft Dynamics GP June 2015
Microsoft Dynamics GP June 2015 What s New in Microsoft Dynamics GP 2015 R2 This document describes enhancments add for the Microsoft Dynamics 2015 R2 release, including where needed, step-by-step instructions
Sage MAS 90 and 200 ERP
Sage MAS 90 and 200 ERP What s New in Sage MAS 90 and 200 4.4? Something for Everyone in 4.4 Based on the solid foundation that Sage MAS 90 and Sage MAS 200 ERP solutions have continued to build on for
INTRODUCTION: SQL SERVER ACCESS / LOGIN ACCOUNT INFO:
INTRODUCTION: You can extract data (i.e. the total cost report) directly from the Truck Tracker SQL Server database by using a 3 rd party data tools such as Excel or Crystal Reports. Basically any software
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (08) 69 3R Full document title and reference Document type VALIDATION OF COMPUTERISED SYSTEMS Legislative basis - CORE DOCUMENT
www.novell.com/documentation Database Maintenance ZENworks Mobile Management 2.7.x August 2013
www.novell.com/documentation Database Maintenance ZENworks Mobile Management 2.7.x August 2013 Legal Notices Novell, Inc., makes no representations or warranties with respect to the contents or use of
Musina Local Municipality. Information and Communication Technology User Account Management Policy -Draft-
Musina Local Municipality Information and Communication Technology User Account Management Policy -Draft- Version Control Version Date Author(s) Details V1.0 June2013 Perry Eccleston Draft Policy Page
21 CFR Part 11 Electronic Records & Signatures
Gap Analysis - Checklist 21 CFR Part 11 Electronic Records & Signatures his document is a proposal and starting point only. he type and extent of documentation depends on the process environment. he proposed
SAM Benefits Overview SAM SOFTWARE ASSET MANAGEMENT
SAM Benefits Overview SAM SAM is critical to managing an IT environment because effectiveness is seriously compromised when an organization doesn t know what software assets it has, where they are located,
ISMS Implementation Guide
atsec information security corporation 9130 Jollyville Road, Suite 260 Austin, TX 78759 Tel: 512-615-7300 Fax: 512-615-7301 www.atsec.com ISMS Implementation Guide atsec information security ISMS Implementation
Non-Controlled Copy. Quality Manual Revision Log. Rev # Date Description
Quality Manual Revision Log Rev # Date Description 1 Oct 1984 Original Document. 2 Aug 1986 Document Hillsboro Move. Complete Revision. 3 May 1988 Incorporate/Beaverton Move. Reformat to Word Perfect 4.0.
Section i POLICY AND AUTHORITY STATEMENT
Section i POLICY AND AUTHORITY STATEMENT It is the policy of Laboratory Testing, Incorporated (LTI)*, as a Material Organization, to furnish and perform testing services in nondestructive, mechanical,
Vector HelpDesk - Administrator s Guide
Vector HelpDesk - Administrator s Guide Vector HelpDesk - Administrator s Guide Configuring and Maintaining Vector HelpDesk version 5.6 Vector HelpDesk - Administrator s Guide Copyright Vector Networks
Infrastructure Technical Support Services. Request for Proposal
Infrastructure Technical Support Services Request for Proposal 15 May 2015 ISAAC reserves the right to reject any and all proposals, with or without cause, and accept proposals that it considers most favourable
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP operates under one quality program, whether you purchase products that are commercial grade, nuclear safety-related or industrial safety compliant (IEC 61508).
Matisse Installation Guide for MS Windows. 10th Edition
Matisse Installation Guide for MS Windows 10th Edition April 2004 Matisse Installation Guide for MS Windows Copyright 1992 2004 Matisse Software Inc. All Rights Reserved. Matisse Software Inc. 433 Airport
MICHIGAN AUDIT REPORT OFFICE OF THE AUDITOR GENERAL THOMAS H. MCTAVISH, C.P.A. AUDITOR GENERAL
MICHIGAN OFFICE OF THE AUDITOR GENERAL AUDIT REPORT THOMAS H. MCTAVISH, C.P.A. AUDITOR GENERAL ...The auditor general shall conduct post audits of financial transactions and accounts of the state and of
ISO 9001:2008 QUALITY MANUAL. Revision B
ISO 9001:2008 QUALITY MANUAL Revision B Because we want you to achieve the highest levels of performance, we connect care Because with leading we want science you to achieve continuously the highest improve,
Request for Information PM07212011 CONTRACT MANAGEMENT SOLUTION
Request for Information PM07212011 CONTRACT MANAGEMENT SOLUTION July 21, 2011 1 Table of Contents I. Introduction and Purpose of Request For Information... 1 II. Background... 2 III. Description of Information
MapReduce. MapReduce and SQL Injections. CS 3200 Final Lecture. Introduction. MapReduce. Programming Model. Example
MapReduce MapReduce and SQL Injections CS 3200 Final Lecture Jeffrey Dean and Sanjay Ghemawat. MapReduce: Simplified Data Processing on Large Clusters. OSDI'04: Sixth Symposium on Operating System Design
Data Management Implementation Plan
Appendix 8.H Data Management Implementation Plan Prepared by Vikram Vyas CRESP-Amchitka Data Management Component 1. INTRODUCTION... 2 1.1. OBJECTIVES AND SCOPE... 2 2. DATA REPORTING CONVENTIONS... 2
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective
SCHEDULE 25. Business Continuity
SCHEDULE 25 Business Continuity 1. Scope 1.1 This schedule covers TfL s requirements in respect of: any circumstance or event which renders, or which TfL considers likely to render, it necessary or desirable
CALIBRATION PRINCIPLES
1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they
Client SSL Integration Guide
Client SSL Integration Guide Version 8.2 December 15, 2015 For the most recent version of this document, visit our documentation website. Table of Contents 1 Client SSL integration overview 3 2 System
Turbo Resources International Quality Procedure
TURBO RESOURCES INTERNATIONAL QUALITY MANAGEMENT SYSTEM MANUAL APPROVALS: Revision : 46 Page : 1 of 53 TABLE OF CONTENTS INTRODUCTION 3 1. SCOPE 4 1.1 GENERAL 4 1.2 APPLICATION 5 2. NORMATIVE REFERENCE
JOHN HART GENERATING STATION REPLACEMENT PROJECT. Schedule 9. Quality Management
JOHN HART GENERATING STATION REPLACEMENT PROJECT Schedule 9 Quality Management SCHEDULE 9 QUALITY MANAGEMENT TABLE OF CONTENTS 1. QUALITY MANAGEMENT SYSTEM... 1 1.1 Quality Management System...1 1.2 Project
Electronic records and electronic signatures in the regulated environment of the pharmaceutical and medical device industries
White Paper No 01 I December 2010 Implementation of 21 CFR Part 11 in the epmotion Software Electronic records and electronic signatures in the regulated environment of the pharmaceutical and medical device
White Paper BMC Remedy Action Request System Security
White Paper BMC Remedy Action Request System Security June 2008 www.bmc.com Contacting BMC Software You can access the BMC Software website at http://www.bmc.com. From this website, you can obtain information
The following items are trademarks or registered trademarks of Kaba Mas in the United States and/or other countries. GITCON
The Gitcon Access Management Software Installation Guide is a publication of Kaba Mas LLC (hereinafter Kaba Mas). No part of this book may be reproduced or transmitted in any form or by any means, electronic