GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards
|
|
|
- Ambrose Walsh
- 9 years ago
- Views:
Transcription
1 March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often deteriorate with time or use. The design of a calibration system would be incomplete without some established means of determining how often to calibrate instruments and standards. A calibration performed only once establishes a one-time reference of uncertainty. Recalibration detects uncertainty growth and serves to reset values while keeping a bound on the limits of errors. A properly selected interval assures that an item will receive recalibration at the proper time. Proper calibration intervals allow specified confidence intervals to be selected and they support measurement traceability. The following practice establishes calibration intervals for standards and instrumentation used in measurement processes Prerequisites 1.3. Safety 2. Methodology Calibration history for laboratory standards Expected tolerance limits if applicable No outstanding safety concerns 2.1. Summary Recommended calibration intervals are based on various examples of traceability as described in GMP 13. As data is collected and evaluated, the laboratory technical manager may adjust the calibration intervals to ensure that measurement results are not invalidated by the intervals selected Apparatus None. GMP 11 Page 1 of 6
2 March Procedure Identification of Parameters Critical Parameters Components that contribute more than 25 % of a measurement s uncertainty are identified as critical parameters. To ensure an accurate evaluation of performance, calibration intervals are determined to meet a 99 % reliability target. Critical parameters are checked and defined below: Mass Critical Parameters (The example provided below corresponds to Option A in Mass Traceability in GMP 13.) Balance Performance Balance performance in Echelon 1 weighing processes is evaluated in every measurement series. An F-test ratio evaluates the observed standard deviation of the process against the accepted standard deviation of the process. The balance performance component is also tested with a check standard in each weighing series. The check standard value is evaluated with a t-test by compared the observed value to the accepted reference value. All other weighing processes have incorporated measurement control procedures and control charts that are evaluated as data is collected. Mass Standards Mass standards are dynamic with use. Wear, contamination and other factors can cause drift from accepted values. Thus, the following intervals have been set: Table 1. Calibration intervals for mass standards P1. kg + P1.. kg 48 NIST C1. kg + C1.. kg (alternating 2 years) 48 NIST P30 kg - P2 kg 12 Lab P500 g - P1 mg 6 Lab C500 g - C1 mg 6 Lab W25 kg - W1 mg 12 Lab P Pound Standards 24 Lab W Pound Standards 12 Lab P = primary; C = check/control; W = working standards GMP 11 Page 2 of 6
3 Length Critical Parameters March 2003 Length Comparison Performance The measurement performance of each length calibration is evaluated with a check standard (when performing tape-to-tape comparison). Initial Length Intervals Length standards are dynamic with use. Wear, contamination and other factors can cause drift from accepted values. The following intervals have been set due to these factors: Table 2. Calibration intervals for length standards 100 ft Tape #1 60 NIST 100 ft Tape #2 60 NIST 25 ft or 7 m Tape 60 NIST 18 in Steel Rule 120 NIST Length Bench 24 (if used or moved) Lab Volume Critical Parameters (Example shown corresponds to Option A for volume calibration in GMP 13.) Volume Comparison Performance The measurement performance of a volume transfer calibration is evaluated in each use with a repeatability check. Use of check standards is the preferred method for evaluating the measurement process over time. Traceability for volume standards may be established through gravimetric calibrations using traceable mass standards. Initial Volume Intervals Volume standards are dynamic with use. Wear, contamination and other factors can cause drift from accepted values. Calibration intervals are as follows: GMP 11 Page 3 of 6
4 Table 3. Calibration intervals for volume standards March 2003 P100 gal standard 60 NIST P25 gal standard * 60 Lab P5 gal standard * 12 Lab Glassware- Autopipetes 5 L to 100 ml 120 Lab *Gravimetric calibration for volumes 5 gallon or smaller, and all slicker plate standards. Laboratory must be qualified for performing gravimetric calibrations. Volume transfer is acceptable above 5 gallon. *May be a slicker plate type. None are hand-held, dump style, test measures. Temperature Critical Parameters Temperature Comparison Performance The measurement performance of each temperature comparison calibration is evaluated with a check standard and can be verified periodically using triple point cells, melting point cells, and ice baths (using documented procedures). Initial Intervals Temperature standards are dynamic with use. Shock, contamination and other factors can cause drift from accepted values. Recalibration intervals are as follows: Table 4. Calibration intervals for temperature standards 25.5 ohm SPRT 36 NIST 100 ohm PRT s 12 Lab Standard Thermistor 12 Lab Check Standards 6 Lab Liquid-in-glass standards 6 Lab Secondary Parameters Components that contribute less than 25 % but more than 1 % of a measurement s uncertainty are identified as secondary parameters. Secondary parameters are assigned calibration intervals designed to meet a 95 % reliability target. Secondary parameters are defined below: Mass and Gravimetric Volume Secondary Parameters GMP 11 Page 4 of 6
5 March 2003 Environmental Measurement Equipment Table 5. Calibration intervals for environmental equipment Item Initial Calibration Interval (months) Source Barometer 12 NIST Hygrometer 24 NIST Thermometer 12 Lab Length Secondary Parameters - No secondary parameters. Volume Secondary Parameters Water Temperature Measuring Device Table 6. Calibration intervals for volume secondary parameters Thermometer 12 Lab Temperature Secondary Parameters - No secondary parameters. 3. Calculations 3.1. Initial Intervals Assignment of Initial Intervals Assignment of initial intervals is based on these recommendations. Subsequent intervals may be adjusted based on analysis of Check standard data that is recorded on control charts. The initial interval is equivalent to the amount of time that passes before a metrologist makes the first observation of a result lying outside the warning limits of the chart when the investigation yields no apparent correctable cause. Warning limits are established at ± 2 standard deviations of the measurement process around the accepted value of the check standard Absence of Control Charts If no control charts are available, the laboratory s Technical Manager will assign the initial interval based on engineering evidence, manufacturer s specifications, NIST recommendations, and experience. GMP 11 Page 5 of 6
6 3.2. Adjustment of Intervals March Intervals will be adjusted when determined to be necessary by the laboratory s Technical Manager The intervals will be adjusted by taking the following factors into consideration as appropriate: 4. Assignment of Uncertainty calibration history; measurement assurance data; data for the population of similar equipment; NIST recommendations; statistical analysis methods; and manufacturer s recommendations. The uncertainty associated with the establishment of calibration intervals is not included as a part of the uncertainties associated with the respective measurement SOPs. See SOP 29 for calculating uncertainties for standards and the instruments. GMP 11 Page 6 of 6
Calibration Service Selection Guide
Calibration Service Selection Guide Certification of Accreditation available for download at www.ricelake.com/accreditation. Answer Questions to determine the calibration service you need Need accredited*
Calibration Certificate
P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: [email protected] SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING
Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
Gravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
User checks and maintenance of laboratory balances
Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National
Estimates of Uncertainty of the Calibration of Balances
Estimates of Uncertainty of the Calibration of Balances Chart shows the uncertainty (estimates) of the calibration of representative electronic balances. Chart. Uncertainty of calibration (estimates) Minimum
Case study Realistic and best uncertainties with humidity instruments
Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against
Process Measurement Assurance Program For U.S. State Metrology Laboratories 1
Process Measurement Assurance Program For U.S. State Metrology Laboratories 1 Jerry L. Everhart JTI Systems, Inc. And Georgia L. Harris NIST Office of Weights and Measures Abstract This paper describes
CALIBRATION PRINCIPLES
1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they
Assessment of Accuracy and Precision
2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter
State Laboratory Program Workload Survey
214 State Laboratory Program Workload Survey Published by the NCSL International Legal Metrology Committee 156 SLP Survey 214 - Page 1 of 132 Contents Acknowledgements... 8 Objectives and History... 9
Infrared Thermometer Calibration A Complete Guide
Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate
State Water Resources Control Board Environmental Laboratory Accreditation Program
State Water Resources Control Board Environmental Laboratory Accreditation Program P.O. Box 100, Floor 24 850 Marina Bay Parkway, Build. P 500 N Central Ave, Suite 500 Sacramento, CA 95812-0100 Richmond,
calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper
calibrate with confidence White Paper Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Authored by Keith Bennett Director of Metrology Howard Zion Director of Technical Operations TRANSCAT
Determination of Moisture Content
Determination of Moisture Content 1. Introduction: Drying, storage, marketing, and roasting are four important aspects of coffee handling in which moisture plays an important role: Drying: moisture measurement
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
Temperature Calibration; Depths of Immersion
Temperature Calibration; epths of Immersion Author: John P. Tavener Company Isothermal Technology Limited, Pine Grove, Southport, England Abstract Of all the sources of errors and uncertainties in thermal
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL
1 of 8 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.
ALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Chemicals Needed:
Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Your Name: Date: Partner(s) Names: Objectives: React magnesium metal with hydrochloric acid, collecting the hydrogen over water. Calculate the grams
COPY. Revision History: Changes were made to reflect the current practice of only including the most recent changes in the revision history.
Revision History COPY The top row of this table shows the most recent changes to this controlled document. For previous revision history information, archived versions of this document are maintained by
How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of
Experiment 1: Measurement and Density
Experiment 1: Measurement and Density Chemistry 140 Learning Objectives Become familiar with laboratory equipment and glassware Begin to see the link between measurement and chemical knowledge Begin to
Calibration of Volumetric Glassware
CHEM 311L Quantitative Analysis Laboratory Revision 2.3 Calibration of Volumetric Glassware In this laboratory exercise, we will calibrate the three types of glassware typically used by an analytical chemist;
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
Beamex. Calibration White Paper. www.beamex.com [email protected]. How often should instruments be calibrated
Beamex Calibration White Paper [email protected] How often should instruments be calibrated How often should instruments be calibrated Plants can improve their efficiency and reduce costs by performing calibration
Physical Properties of a Pure Substance, Water
Physical Properties of a Pure Substance, Water The chemical and physical properties of a substance characterize it as a unique substance, and the determination of these properties can often allow one to
A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS
Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,
Ti-Pure TITANIUM DIOXIDE. DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T4400.535.01.WP
Ti-Pure TITANIUM DIOXIDE DETERMINATION OF BROOKFIELD VISCOSITY, TiO 2 SLURRY METHOD: T4400.535.01.WP 1 TITANIUM DIOXIDE (Ti-Pure ) Determination of Viscosity by Brookfield Viscometer I. Principle Viscosity
Figure 1. A typical Laboratory Thermometer graduated in C.
SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES
ACL 800 MEGOHMMETER OPERATION MANUAL
ACL 800 MEGOHMMETER OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor. Calibration is recommended every twelve months. NOTE: THIS MANUAL IS NOT TO BE SEPARATED
New York State Department of Health (NYSDOH) Bureau of Immunization Vaccine Storage and Handling Plan
Vaccines need to be properly stored and administered to ensure maximum efficacy and safety. Check yes or no for each item. Implement changes for items marked no. Personnel: 1. We have designated a primary
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
Beamex. Calibration White Paper. www.beamex.com [email protected]. How often should instruments be calibrated?
Beamex Calibration White Paper [email protected] How often should instruments be calibrated? How often should instruments be calibrated? An analysis will tell. Plants can improve their efficiency and reduce
Life & Times of a PM2.5 Filter: Quality Control in the Laboratory and Beyond
Life & Times of a PM2.5 Filter: Quality Control in the Laboratory and Beyond Brenda Saldana, Manager Inorganic Laboratory Section Monitoring and Laboratory Division Air Resources Board 1 Field Sampling
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
Calibration of temperature control and monitoring devices
WHO Vaccine Calibration of temperature control and monitoring devices Technical supplement to WHO Technical Report Series, No. 961, 2011 Annex 9: Model guidance for the storage and transport of time and
3M CORPORATE METROLOGY SERVICES St. Paul, MN
A2LA has accredited THE AMERICAN ASSOCIATION FOR LABORATORY ACCREDITATION ACCREDITED LABORATORY 3M CORPORATE METROLOGY SERVICES St. Paul, MN for technical competence in the field of Calibration This laboratory
Measurement and Calibration
Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding
Relative Humidity Calibration Kit
Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration
Measurement of RF Emissions from a Final Coat Electronics Corrosion Module
Engineering Test Report No. 37802-02 Rev. A Measurement of RF Emissions from a Final Coat Electronics Corrosion Module For : Canadian Auto Preservation, Inc. 390 Bradwick Drive Concord, Ontario CANADA
The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications
The Uncertainty of Reference Standards A Guide to Understanding Factors Impacting Uncertainty, Uncertainty Calculations, and Vendor Certifications Kevin Gates, Ning Chang, Isil Dilek, Huahua Jian, Sherri
Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration?
Calibration Tvinnargatan 25 SE-507 30 Brämhult SWEDEN Phone: +46 33 22 56 30 Fax: +46 33 13 88 71 [email protected] www.kalibrera.se In May 2011 Elastocon purchased the company Mätcentrum in Eskilstuna,
The Basics of Fertilizer Calculations for Greenhouse Crops Joyce G. Latimer, Extension Specialist, Greenhouse Crops; Virginia Tech
The Basics of Fertilizer Calculations for Greenhouse Crops Joyce G. Latimer, Extension Specialist, Greenhouse Crops; Virginia Tech Publication 430-100 Meeting the Plant s Needs Fertilizers are designed
NIST Calibration Services for Liquid Volume NIST Special Publication 250-72
NIST Calibration Services for Liquid olume NIST Special Publication 250-72 ern E. Bean, Pedro I. Espina, John. D. Wright, John F. Houser, Sherry D. Sheckels, and Aaron N. Johnson March 24, 2006 Fluid Metrology
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company
CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM
Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13. Pi Tape Corporation. 344 North Vinewood St. Escondido CA 92029-1337 USA
Quality Assurance Manual January 2008 Revision I Revision date of 01/17/13 Pi Tape Corporation 344 North Vinewood St. Escondido CA 92029-1337 USA Phone # 760 746-9830 Fax # 760 746-9196 In conformance
COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS
COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS Primer CONTENTS INTRODUCTION...3 QUALITY AND COMPLIANCE IN QUALITY CONTROL LABORATORIES...5 Compliance
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion
A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion Objective In the experiment you will determine the cart acceleration, a, and the friction force, f, experimentally for
ISO/IEC 17025 QUALITY MANUAL
1800 NW 169 th Pl, Beaverton, OR 97006 Revision F Date: 9/18/06 PAGE 1 OF 18 TABLE OF CONTENTS Quality Manual Section Applicable ISO/IEC 17025:2005 clause(s) Page Quality Policy 4.2.2 3 Introduction 4
Keysight Technologies Specifications Guidelines. Technical Overview
Keysight Technologies Specifications Guidelines Technical Overview Introduction Keysight Technologies, Inc. has definitions for its Test & Measurement product specifications and how they are presented.
ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards.
MICRO MOTION WHITE PAPER BY M. LEE, D. STANDIFORD, MICRO MOTION, INC. A. PRUYSEN, EMERSON PROCESS MANAGEMENT ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters
1. The Determination of Boiling Point
1. The Determination of Boiling Point Objective In this experiment, you will first check your thermometer for errors by determining the temperature of two stable equilibrium systems. You will then use
How To Test A Titrette And Digital Burette
Titrette and Digital Burette (all models) Testing Instructions (SOP) May 2013 1. Introduction The standard ISO DIS 8655 describes both the design and the testing of bottle-top burette. The following Testing
The Molar Mass of a Gas
The Molar Mass of a Gas Goals The purpose of this experiment is to determine the number of grams per mole of a gas by measuring the pressure, volume, temperature, and mass of a sample. Terms to Know Molar
TEST REPORT for the ARID TECHNOLOGIES VAPOR RECOVERY UNIT installed in LANTANA, FL
Project No. 148 TEST REPORT for the ARID TECHNOLOGIES VAPOR RECOVERY UNIT installed in LANTANA, FL Prepared for: Mr. Ted Tiberi Arid Technologies, Inc. 130 North LaGrange Rd. LaGrange, IL 60525 Prepared
Custody Transfer Measurement. with the V-Cone Flowmeter
Custody Transfer Measurement with the V-Cone Flowmeter Stephen A. Ifft McCrometer Inc. Hemet, California, USA Abstract This paper will discuss the approval of the McCrometer V-Cone flowmeter for custody
EXERCISE # 1.Metric Measurement & Scientific Notation
EXERCISE # 1.Metric Measurement & Scientific Notation Student Learning Outcomes At the completion of this exercise, students will be able to learn: 1. How to use scientific notation 2. Discuss the importance
Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document 74-461
Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/0/202 Adapted from National Bureau of Standards Document 74-46 The purpose of calibrating glassware is to determine the volume of
Hand-held thermometer Model CTH7000
Calibration technology Hand-held thermometer Model CTH7000 WIKA data sheet CT 55.50 Applications Precision thermometer for very accurate temperature measurements in a range of -200... +962 C Reference
Environment-Laboratory Ambient Conditions
ISSUE 3.0: MADE MINOR CLARIFICATION CHANGES. FOR CLIENT LABS Purpose To assure laboratory conditions are stable during the performance of testing and data collection. Why this requirement is important?
Francis Mukalazi Stephen Fisheries Resources Department Uganda P.O. Box 4 Entebbe Tel +25677405204 [email protected] or mukalazif@yahoo.
P.O. Box 1390, Skulagata 4 120 Reykjavik, Iceland Final Project 2003 CALIBRATION ACCORDING TO ISO/IEC 17025 IN THE OPERATION OF MICROBIOLOGICAL AND CHEMICAL LABORATORIES: AN EXERCISE IN CREATING CONTROL
Estimating Weighing Uncertainty From Balance Data Sheet Specifications
Estimating Weighing Uncertainty From Balance Data Sheet Specifications Sources Of Measurement Deviations And Uncertainties Determination Of The Combined Measurement Bias Estimation Of The Combined Measurement
11.I In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07
In-process control In-process control Authors: Dr. Christian Gausepohl / Paolomi Mukherji / Update 07 Here you will find answers to the following questions: What are the in-process control tasks? Where
Calibration of Glassware
Calibration of Glassware Introduction Glassware is commonly calibrated using a liquid of known, specific density, and an analytical balance. The procedure is to determine the mass of liquid the glassware
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
SUPPLIER Form #: 007 AUDIT QUESTIONNAIRE
1.0 BACKGROUND INFORMATION Company Name: Company Address: Phone Number: Fax Number: Number of years in business: Number of employees: Annual Sales in Dollars: Size of Facility in Square Feet: Name of Senior
ME 315 - Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS
ME 315 - Heat Transfer Laboratory Nomenclature Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS A heat exchange area, m 2 C max maximum specific heat rate, J/(s
Inventory Control. Inventory Control Tank & Silo Weighing
Inventory Control Tank & Silo Weighing Gravimetric level control is one of the most accurate inventory control methods available. Its high degree of accuracy makes it particularly helpful when measuring
Understanding electrical conductivity
World Bank & Government of The Netherlands funded Training module # WQ - 08 Understanding electrical conductivity New Delhi, May 1999 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi 11
OVERVIEW OF STANDARD 7E & STANDARD 20
Standard for Thermal Transport Packaging Used in Parcel Delivery System Shipment 7E 2 010 ISTA 7E Usage Modes VERSION DATE OCTOBER 2010 Initial Release ISTA, Distributing Confidence, Worldwide ISTA 7-Series
Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley. Norton 0
Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley Norton 0 Norton 1 Abstract The charge of an electron can be experimentally measured by observing an oil
Phase Diagram of tert-butyl Alcohol
Phase Diagram of tert-butyl Alcohol Bill Ponder Department of Chemistry Collin College Phase diagrams are plots illustrating the relationship of temperature and pressure relative to the phase (or state
EQUIPMENT CALIBRATION/MAINTENANCE
EQUIPMENT CALIBRATION/MAINTENANCE The following chart provides a guide for calibration of standard equipment used in the Fish Health Laboratory. All Intervals suggested are subject to individual manufacturer
18.6.1 Terms concerned with internal quality control procedures
18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable
Basic requirements for quality management in plant pest diagnosis laboratories
Blackwell Publishing Ltd European and Mediterranean Plant Protection Organization PM 7/84 Organisation Européenne et Méditerranéenne pour la Protection des Plantes Diagnostics Diagnostic Basic requirements
State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist
This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to
Testing Your Laboratory Balance
Testing Your Laboratory Balance If you use a laboratory balance in your day-to-day work, you may have wondered whether it is accurate. Here are four tests that you can use to confirm your instrument's
Calibration of Volumetric Glassware
Chemistry 119: Experiment 2 Calibration of Volumetric Glassware For making accurate measurements in analytical procedures, next in importance to the balance is volumetric equipment. In this section volumetric
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
VERIFYING CALIBRATION OF WEIGHING DEVICES USED FOR LABORATORY TESTING
Test Procedure for VERIFYING CALIBRATION OF WEIGHING DEVICES USED FOR Texas Department of Transportation TxDOT Designation: Tex-901-K Effective Date: August 1999 1. SCOPE 1.1 All weighing devices shall
Experimental Uncertainties (Errors)
Experimental Uncertainties (Errors) Sources of Experimental Uncertainties (Experimental Errors): All measurements are subject to some uncertainty as a wide range of errors and inaccuracies can and do happen.
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results
Weight Calibration ComparatorPac TM Simple Calibration Efficient Workflow Reliable Results Safe and Simple Weight Calibration with ComparatorPac TM ComparatorPac Weight Calibration for Traceable Results
Experiment 1: Colligative Properties
Experiment 1: Colligative Properties Determination of the Molar Mass of a Compound by Freezing Point Depression. Objective: The objective of this experiment is to determine the molar mass of an unknown
Apr 17, 2000 LAB MANUAL 1302.0. 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified)
Apr 17, 2000 LAB MANUAL 1302.0 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified) 1302.1 SCOPE This method describes a procedure for the quantitative determination of the distribution
FederalDemocratic RepublicofEthiopia
FederalDemocratic RepublicofEthiopia EDICTOFGOVERNMENT± Inordertopromotepubliceducationandpublicsafety, equaljusticeforal,abeterinformedcitizenry,therule oflaw,worldtradeandworldpeace,thislegaldocumentis
Grammage of paper and paperboard (weight per unit area) (Revision of T 410 om-08) (underscores and strikeouts indicate changes from Draft 1)
NOTICE: This is a DRAFT of a TAPPI Standard in ballot. Although available for public viewing, it is still under TAPPI s copyright and may not be reproduced or distributed without permission of TAPPI. This
American Association for Laboratory Accreditation
Page 1 of 18 Traceability for Life Sciences Testing and (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written
EVALUATION AND ACCREDITATION OF SUPPLIER TEST FACILITIES GP 10
EVALUATION AND ACCREDITATION OF SUPPLIER TEST FACILITIES GP 10 EVALUATION AND ACCREDITATION OF TESTING FACILITIES USED BY SUPPLIER OF MATERIALS TO GENERAL MOTORS PREFACE TO GP-10 General Motors expects
US Environmental Protection Agency Office of Pesticide Programs
US Environmental Protection Agency Office of Pesticide Programs Office of Pesticide Programs Microbiology Laboratory Environmental Science Center, Ft. Meade, MD Standard Operating Procedure for Monitoring
calibrate confidence with Risk Management: The Critical Role of Measurement and Calibration WHITE PAPER SERVICES
calibrate with confidence The Critical Role of Measurement and Calibration The ultimate goal of your company is to be profitable in providing a superior, effective, and safe product or service. Your use
Guide to adhesively mounting accelerometers
Guide to adhesively mounting accelerometers cmyk Guide to adhesively mounting accelerometers Purpose This technical paper identifies and clarifies issues regarding adhesive mounting of accelerometers.
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS
DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS Bernie B. Bernard, Heather Bernard, and James M. Brooks TDI-Brooks International/B&B Laboratories Inc. College Station,