PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS
|
|
|
- Arlene Hoover
- 10 years ago
- Views:
Transcription
1 PT/EQA STANDARDS AND GUIDELINES: QUALITY AND RELIABILITY OF TEST ITEMS Maria Belli Istituto Superiore per la Protezione e la Ricerca Ambientale (former APAT)
2 ISPRA activities production and characterization of matrix reference materials methods development and harmonization organization of Proficiency testing organization of collaborative studies for method validation
3 Test material/test item IUPAC The International Harmonized Protocol for the proficiency testing of (chemical) Analytical Laboratories does not define specifically the Test materials, but reports RM and CRM definitions as given by ISO REMCO IUPAC The International Harmonized Protocol for the proficiency testing of (chemical) Analytical Laboratories 2006 cross-refers to ISO standard definitions. It is given a definition for Distribution unit a packaged portion of the test material that is sent to participant laboratories
4 Test Item material or artefact presented to the participating laboratory for the purpose of proficiency testing ISO Guide 43-1:1997 a sample, product, artefact, piece of equipment or measurement standard sent to one or more participants in a proficiency testing scheme ILAC G13:2007
5 Test Item sample, product, artefact, reference material, piece of equipment, measurement standard or data set provided to one or more participants, or submitted by participants, in a proficiency testing round ISO-IEC CD 17043: March 2008
6 Test item in analytical chemistry RM family (Measurement Standards - ISO/IEC guide 99) in-house RMs, RMs QCMs, PCMs CRMs, CALs H. Emons et. al., 2006, AQUAL 10, 576 Certified reference materials, calibrators
7 Quality of test items Matrix matching Homogeneity Stability Assigned values
8 Test item - Matrix matching 1993 Pure & Appl. Chem. Vol.65 n 9 (IUPAC Protocol) Qualitative definition 2008 ISO/IEC CD 17043
9 Matrix Matching Matrix proficiency test items should, were practicable, have the same or nearly the same matrix as routine test materials in order to simulate the measurement process as close as possible ILAC-G13:2007
10 Commutability ISO Guide 99 gives the definition of commutability of reference material ISO Dguide 34 reports a method to assess commutability of reference material: If the ratio between the results on reference material with 2 measurement procedures is the same as the ratio for routine samples, the reference material is commutable
11 Commutability ARPA FVG (ng/m3) PAH in PM10 Benzo[a]pyrene (ng/m3) Procedures CRM ratio GC-MS/HPLC BCR 535 sediment GC-MS/HPLC BCR088 sludge GC-MS/HPLC NIST1650b y=0.7409x ISPRA (ng/m3) Procedures ratio GC-MS/HPLC 1.7±0.2
12 Homogeneity APAT RM014 - Ni between bottles homogeneity Homogeneity study results are affected by: measurement repeatability between bottles variability
13 Homogeneity Study Standards and Guidelines requirements Standards/Guidelines IUPAC Harm. Protocol (1993) ISO Guide 43-1:1997 ISO 13528:2005 Statistical methods in PT IUPAC Harm. Protocol (2006) ILAC G13:08/2007 How* 10 test items - 2 replicates ANOVA s bb /σ PT <0.3 Cross reference to IUPAC (1993) 10 test items - 2 replicates s bb /σ PT < test items - 2 replicates σ 2 samp <F 1 0.3σ2 PT +F 2 s2 an ANOVA to obtain σ 2 samp e s2 an Recommendation σ an /σ PT <0.5 Where appropriate * Measurements in repeatability conditions
14 Homogeneity Study Standards and Guidelines requirements Standards/Guidelines ISO/IEC CD 17043:2008 Conformity assessment General requirement for PT ISO Guide 35:2006 RM- General and Statistical principles for certification ISO DGuide 34:2008 General requirements for the competence of RM producers ISO CD Guide 80:2008 Guidance for production of RM for precision control (PCMs) How* As above, but cross reference to ISO IUPAC (2006), ISO Guide 34 and ISO Guide or more test items 3 replicates - ANOVA to assess the uncertainty contribution for residual heterogeneity Always RM producers shall carry out homogeneity study in compliance with ISO test items 2 replicates ANOVA to assess homogeneity between bottles *measurements in repeatability conditions are always required
15 APAT RM014 Ni homogeneity Before the PT ISO Guide 35: After the PT APAT - RM014 - Nichel 71 Valore di riferimento : 256 ± 8 mg/kg Incertezza estesa con k=2 HF+HCl+HNO3 HNO3+HF HCl+HNO3+H2O2/HClO4 HCl+HNO3 HNO3 HNO3+H2O2/HClO4 11 u bb =1.2 µg/g mg/kg Laboratorio σ PT =31.8 µg/g U bb <9.5 µg/g ISO 13528:2005
16 Stability Stability study results are affected by: measurement repeatability between bottles variability
17 Stability Study Standards and Guidelines requirements Standards/Guidelines IUPAC Harm. Protocol (1993) ISO Guide 43-1:1997 ISO 13528:2005 Statistical methods in PT IUPAC Harm. Protocol (2006) ILAC G13:08/2007 How* Stability of the matrix and analytes must be determined Where possible the coordinator should provide evidence of stability of the test items during the PT Test items of the hom. test 3 2 replicates measured at the end of PT x mh -y ms 0.3σ PT Isochronous stability study for the PT period Where appropriate * Measurements in repeatability conditions
18 Stability Study Standards and Standards/Guidelines ISO/IEC CD 17043:2008 Conformity assessment General requirement for PT ISO Guide 35:2006 RM- General and Statistical principles for certification Guidelines requirements How* As above, but cross reference to ISO IUPAC (2006), ISO Guide 34 and ISO Guide 35 Short term stability by isochronous; long term stability by classical method ISO DGuide 34:2008 General requirements for the competence of RM producers ISO CD Guide 80:2008 Guidance for production of RM for precision control (PCMs) Always RM producers shall carry out stability study in compliance with ISO isochronous method * Measurements in repeatability conditions
19 APAT RM014 stability Isochronous method 3 months Effects at +20 C Reference group at 18 C Measurements under repeatability condition By Linear Regression b n t s b 1 0,95; 2 b 1 = slope; s(b 1 ) = slope uncertainty; t 0,95 = t-student. ( 1 )
20 Assigned values All standards and guidelines agree on methods to determine the assigned values and their associated uncertainties: Reference values Consensus from participants Consensus from expert laboratories By formulation
21 Conclusions Production of test items is not an easy activity Laboratories in charge for their preparation shall demonstrate competence in the measurement of properties being determined Is the ISO accreditation enough for test items production?
General and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
Certifying Reference Materials to ISO Guide 34
Certifying Reference Materials to ISO Guide 34 Gill Holcombe Head of Reference Materials Production Environmental Laboratory Quality Day 17 November 2011 4005 Outline LGC background Quality systems - ISO
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION
International Document OIML 1 st CD GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Organization of Legal Metrology Contents 01 Scope 2 12
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company
CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Performance and advantages of qnmr measurements
Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended
Interlaboratory studies
Interlaboratory studies Department of Food Chemistry and Analysis, ICT Prague Vladimír Kocourek Prague, 2012 Interlaboratory studies Various titles: Interlaboratory proficiency test or studies Interlaboratory
There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.
0(752/2*
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective
IAEA-TECDOC-1350. Development and use of reference materials and quality control materials
IAEA-TECDOC-1350 Development and use of reference materials and quality control materials April 2003 The originating Section of this publication in the IAEA was: Industrial Applications and Chemistry Section
Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team
Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for
1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.
Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements
Validation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
Training Programme. Laboratory Quality Manager and Laboratory Assessor. 16 27 November 2015. Vienna, Austria
Training Programme Laboratory Quality Manager and Laboratory Assessor 16 27 November 2015 Vienna, Austria IMPORTANT INFORMATION OVERVIEW The European Organization for Quality training course is designed
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
NVLAP Assessor Training. Assessor Qualification & Training Requirements
NVLAP Assessor Training Assessor Qualification & Training Requirements References ISO/IEC 17011: Conformity assessment General requirements for accreditation bodies accrediting conformity assessment bodies
OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL DOCUMENT OIML D 18 Edition 2008 (E) The use of certified reference materials in fields covered by metrological control exercised by national services of legal metrology. Basic principles
NABL NATIONAL ACCREDITATION
NABL 162 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDELINES FOR PROFICIENCY TESTING PROGRAM for TESTING AND CALIBRATION LABORATORIES ISSUE NO : 04 AMENDMENT NO : 00 ISSUE
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 LABORATORIES
Page 1 of 7 R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC 17025 November 2009 2009 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
The age of competence: an update on the international laboratory accreditation scene for veterinary testing laboratories
J Vet Diagn Invest 14:89 96 (2002) SPECIAL ARTICLE The age of competence: an update on the international laboratory accreditation scene for veterinary testing laboratories Ann L. Wiegers Abstract. Many
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
The Fitness for Purpose of Analytical Methods. A Laboratory Guide to Method Validation and Related Topics
The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method Validation and Related Topics Second Edition 2014 The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method
List of EA Publications. And International. Documents
EA/INF-01 List of EA Publications and International Documents Publication Reference EA-INF/01: 2016 List of EA Publications And International Documents PURPOSE This publication gives the list of EA documents
TRACEABILITY. Commercially available single-element solutions are used. of Single-Element Calibration Solutions
Elemental solution standards can be produced that are almost clones of NIST SRMs. TRACEABILITY of Single-Element Calibration Solutions Commercially available single-element solutions are used to calibrate
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES Reviewed by: Natalija Jovičić Zarić MSc, Quality Manager Approved by: Milica Lukešević, Acting Director Any
Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014 1 TABLE OF CONTENTS FOREWORD... 1 1.0 SCOPE... 1 1.5 Safety Requirements... 1 4.0 MANAGEMENT REQUIREMENTS... 1 4.1
QA/QC. Standard methods in field. QA/QC in the field. QA/QC in analysis
QA/QC To identify, measure and keep systematic and random errors under control in planning, sampling, analysis and reporting of data. Quality assurance = how samples and data are collected and managed.
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
SUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536
SUS Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0208/658535 Fax 0049(0208/658536 Company Profile SUS was founded in 1986 in Oberhausen in the Ruhr area (close to Düsseldorf, in order to meet
Stimuli to the Revision Process
2 STIMULI TO THE REVISION PROCESS of the USPC or the USP Council of Experts Vol. 33(6) [Nov. Dec. 2007] The Application of Uncertainty to USP s Compendial Reference Standards Program: Certified Reference
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19
Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19 By Patricia DALMAS, Product Specialist, Radiometer Analytical Pharmaceutical laboratories working
Selection, Use and Interpretation of Proficiency Testing (PT) Schemes
Selection, Use and Interpretation of Proficiency Testing (PT)Schemes es Second Ed dition 2011 Selection, Use and Interpretation of Proficiency Testing (PT) Schemes Second Edition 2011 Editors Ian Mann
Preparation of Reference Materials for Analytical Quality Control in the Water Laboratories
UNDP/GEF Danube Regional Project Preparation of Reference Materials for Analytical Quality Control in the Water Laboratories Report on Homogeneity Tests of the Second Sediment Reference Materials October,
Guidelines on the Reporting of Compliance with Specification
Guidelines on the Reporting of Compliance with Specification ILAC-G8:03/2009 Copyright ILAC 2009 ILAC encourages the authorized reproduction of this publication, or parts thereof, by organisations wishing
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
DMA-1 Direct Mercury Analyzer
DMA-1 Direct Mercury Analyzer www.milestonesrl.com MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the milestone DmA-1 1. no sample preparation 2. ease of use 3. lowest cost of analysis
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
THE INTERNATIONAL HARMONIZED PROTOCOL FOR THE PROFICIENCY TESTING OF ANALYTICAL CHEMISTRY LABORATORIES
Pure Appl. Chem., Vol. 78, No. 1, pp. 145 196, 2006. doi:10.1351/pac200678010145 2006 IUPAC INTERNATIONAL UNION OF PURE AND APPLIED CHEMISTRY ANALYTICAL CHEMISTRY DIVISION* INTERDIVISIONAL WORKING PARTY
ALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
R221 - Specific Requirements: Forensic
R221 - Specific Requirements: Forensic Page 1 of 6 R221 - Specific Requirements: Forensic (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by
International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection
International Laboratory Accreditation Cooperation Laboratory Accreditation or ISO 9001 Certification? Laboratory Accreditation or ISO 9001 Certification? When selecting a supplier to fulfill your testing,
Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007
Proficiency Testing In A Global Analytical Environment Benefits & Challenges AOAC September 2007 1 The Coca-Cola Company The Coca-Cola Company is the largest manufacturer, distributor and marketer of nonalcoholic
Overview of ISO 17511 for 'Biologicals'
JCTLM Symposium on Reference Measurement Systems for Biologicals International Bureau of Weights and Measures Pavillon de Breteuil, Sèvres, FR 2004-12-15 Overview of ISO 17511 for 'Biologicals' R. Dybkaer
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials
COOMET Recommendation Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials CООМЕТ R/RM/22:2013 Agreed at the 17 th meeting of COOMET
Accreditation according to ISO/IEC 17025 in Europe. Historic EA Europeing Accreditation Assessment Experience from Technical assessment
Accreditation according to ISO/IEC 17025 in Europe Historic EA Europeing Accreditation Assessment Experience from Technical assessment 1 In year 1803 the foots in Europe had a big differences, depending
EffiValidation 3.0 software basic training module
EffiValidation 3.0 software basic training module Oct 3, 2012 www.effichem.com Agenda EffiChem company EffiValidation 3.0 software : overview International standards and guidelines Software features Structure
EUROLAB 25 years. The role of metrology and testing to characterize materials and products. EUROLAB 25th Anniversary Seminar
EUROLAB 25 years The role of metrology and testing to characterize materials and products April 9, 2015, Odense Danmark 1 Origin of Conformity Assessment The New Approach for the European Economic Community
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES October 2008 Revision 08 Great Lakes Institute for Environmental Research University of Windsor Windsor, Ontario
MANAGEMENT REVIEW FOR LABORATORIES AND INSPECTION BODIES
APLAC TC 003 MANAGEMENT REVIEW FOR LABORATORIES AND INSPECTION BODIES Issue No. 4 Issue Date: 09/10 Page 1 of 7 PURPOSE This document gives laboratories and inspection bodies guidance on how to establish
To avoid potential inspection
Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become
Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites
Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites 1. OVERVIEW In Ireland at present there are no mandatory minimum criteria specified in relation to the completion of stationary source
18.6.1 Terms concerned with internal quality control procedures
18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
High Level Importance: ISO 17025 Microbiology compliance
High Level Importance: ISO 17025 Microbiology compliance Thamolwan Laoviitayanurak, 3M FSD Professional Service APEC Regional Workshop, BKK Thailand on Aug 25, 2011 3M 2009. All Rights Reserved. Topic
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
Observations on the determination of gold in gold alloys cupellation method (Fire Assay)
Observations on the determination of gold in gold alloys cupellation method (Fire Assay) Dippal Manchanda MSc CSci CChem FRSC Technical Director & Chief Assayer WHAT WE WANT TO ACHIEVE TODAY 1. Understand
How do I gain confidence in an Inspection Body? Do they need ISO 9001 certification or ISO/IEC 17020 accreditation?
What should I look for when I have an Inspection need? 3 How do I gain confidence in an 4 How can accreditation of the inspection body by an ILAC accreditation body member provide confidence? 6 How can
ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES
ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.
A commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
Federal Institute for Materials Research and Testing
Federal Institute for Materials Research and Testing CERTIFIED REFERENCE MATERIAL FOR THE GAS ADSORPTION BAM-PM-101 Material: SiO 2 with specific surface area (BET) of 0.177 0.004 m² g -1 Mean of means
Basic Steps to Establish ISO/IEC 17025 Compliant Testing Laboratories
Workshop for Caribbean countries to promote the development and implementation of Conformity Assessment programmes St. Augustine (Trinidad and Tobago) 2-4 December Basic Steps to Establish ISO/IEC 17025
Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
DHI DANETV Test Centre Quality Manual
Purpose This quality manual has been prepared to comply with the requirements from the EU ETV pilot programme and the international standard for inspection, ISO/IEC 17020. The quality manual is integrated
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x
CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this
EA-4/18 - guidance on the level and frequency of proficiency testing participation. Guidance on the level
EA-4/18 INF: 2010 Guidance on the level and frequency of proficiency expression testing participation PURPOSE The aim of this paper is to promote harmonization between Accreditation Bodies on how the level
MILESTONE H E L P I N G C H E M I S T S
MILESTONE H E L P I N G C H E M I S T S 8 good reasons to choose the Milestone DMA-80 1. No sample preparation The DMA-80 does not require any sample preparation or other wet chemistry prior the analysis.
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
DATA FOR PUBLIC HEALTH DECISION MAKING
Asia Pacific Economic Cooperation (APEC) Laboratory Capacity Building Workshop Thailand August 2011 DATA FOR PUBLIC HEALTH DECISION MAKING María Estela Ayala Galdós INSTITUTO TECNOLÓGICO PESQUERO DEL PERÚ
Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters. Environment Agency July 2015 Version 2
Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters Environment Agency July 2015 Version 2 Foreword We set up our Monitoring Certification Scheme
American Association for Laboratory Accreditation
Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number
TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS
TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified
LAB 37 Edition 3 June 2013
LAB 37 Edition 3 June 2013 Accreditation Requirements for Sampling and Testing in Accordance with the Drinking Water Testing Specification (DWTS) CONTENTS SECTION PAGE 1 Introduction 2 2 Regulatory requirements
PTA proficiency testing for metal testing laboratories
PTA proficiency testing for metal testing laboratories BRIGGS Philip (Proficiency Testing Australia PO Box 7507, Silverwater, Rhodes NSW 2128 Australia ) Abstract: Proficiency testing is used internationally
Quality Assurance/Quality Control in Acid Deposition Monitoring
Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial
CERTIFICATION REPORT
CERTIFICATION REPORT The certification of the mass fractions of As, Br, Cd, Cl, Cr, Hg, Pb, S and Sb and the assignment of indicative values for Sn and Zn in two polyethylene reference materials Certified
ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:
How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure
Appendix H: Probability of Detection (POD) as a Statistical Model for the Validation of Qualitative Methods
Appendix H: Probability of Detection (POD) as a Statistical Model for the Validation of Qualitative Methods European validation organizations utilize ISO standard 16140:003(E), Protocol for the Validation