ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS

Size: px
Start display at page:

Download "ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS"

Transcription

1 ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS PRESENTATION TO RACI 20 TH APRIL 2016 PAUL MORGAN DIRECTOR CRITICAL CERTIFICATION PTY LTD

2 CHANGES TO ISO14644 PART 1 CHANGES TO ISO14644 PART 2 EXPECTED CHANGES TO ISO14644 PART 3 HOW WILL THIS LIKELY AFFECT EU / GMP PIC/S ISO29643 PARTICLE COUNTER REQUIREMENTS ISO21501 PART 4

3 CURRENT ISO STANDARDS: Part 1: Classification of air cleanliness Part 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration Part 3: Test methods Part 4: Design, construction and start-up Part 5: Operations Part 6: Vocabulary Part 7: Separative devices (clean air hoods, gloveboxes, isolators and mini-environments) Part 8: Classification of airborne molecular contamination Part 9: Classification of surface cleanliness by particles (Under preparation) Part 10: Classification of air cleanliness by nanoparticles (Under preparation)

4 ISO14644 PART 1: 2015 MAJOR CHANGES: EQUIPMENT DESCRIPTION MOVED FROM PART 3 NUMBER OF LOCATIONS INCREASED SO THAT 95% UPPER CONFIDENCE LIMIT IS NO LONGER REQUIRED SOME REQUIREMENTS TO SAMPLE LARGER MICRON SIZE PARTICLES REMOVED

5 ISO Class number (N) ᶢ Maximum allowable concentrations (particles/m3) for particles equal to and greater than the considered sizes, shown belowᵃ 0.1µm 0.2µm 0.3µm 0.5µm 1.0µm 5.0µm 10ᵇ ᵈ ᵈ ᵈ ᵈ ᵉ ᵇ 10ᵇ ᵈ ᵈ ᵉ ᵇ ᵈ ᵉ ᵇ ᵉ ᵈᵉᶠ ᶜ ᶜ ᶜ ᶜ ᶜ ᶜ ᶜ ᶜ ᶜ ᵃ All concentrations in the table are cumulative, e.g. for ISO Class 5, the particles shown at 0,3 μm include all particles equal to and greater than this size. ᵇ These concentrations will lead to large air sample volumes for classification. Sequential sampling procedure may be applied; see Annex D. ᶜ Concentration limits are not applicable in this region of the table due to very high particle concentration. ᵈ Sampling and statistical limitations for particles in low concentrations make classification inappropriate. ᵉ Sample collection limitations for both particles in low concentrations and sizes greater than 1 μm make classification at this particle size inappropriate, due to potential particle losses in the sampling system. ᶠ In order to specify this particle size in association with ISO Class 5, the macroparticle descriptor M may be adapted and used in conjunction with at least one other particle size. (See C.7.) ᶢ This class is only applicable for the in-operation state.

6 THE EFFECT OF REMOVING THE REQUIREMENT FOR HIGHER PARTICLE SIZE: SECTION A.4.4: At each sampling location, sample a volume of air sufficient to detect a minimum of 20 particles if the particle concentration for the largest selected particle size were at the class limit for the designated ISO Class In effect, under the previous standard, when the largest particle size of 5.0µm was used for ISO Class 5, then determined that a cubic metre was to be sampled. However, this requirement for Class 5 has now been removed, however, the standard further states: If measurements are to be made at more than one considered particle size, each larger particle diameter (e.g. D2) shall be at least 1,5 times the next smaller particle diameter.

7 SAMPLING LOCATIONS: In the previous standard, the number of locations required to be sampled was determined as the square root of the floor area. i.e. for 100m² floor area, 10 locations were required. However, under this standard, that is no longer the case, for instance, in a 100m² room it is now required that 16 locations are sampled, see table on following slide. This is required to ensure that it provides at least 95 % confidence that at least 90 % of the cleanroom or clean zone area does not exceed the class limits. In order to position the sampling locations a) use the minimum number of sampling locations NL derived from Table A.1, b) then divide the whole cleanroom or clean zone into NL sections of equal area, c) select within each section a sampling location considered to be representative of the characteristics of the section, and d) at each location, position the particle counter probe in the plane of the work activity or anotherspecified point. Additional sampling locations may be selected for locations considered critical. Their number and positions shall also be agreed and specified.

8 AREA OF ROOM (m²) LESS THAN, OR EQUAL TO MINIMUM NUMBER OF SAMPLE LOCATIONS TO BE TESTED (Nᶫ)

9 Where the size of the room exceeds 100m², the following formula is to be applied: NL = 27x ( A ) (1000) where NL is the minimum number of sampling locations to be evaluated, rounded up to the next whole number; A is the area of the cleanroom in m². If the considered area falls between two values in the table, the greater of the two should be selected. I.e. If the floor area = 42m², then 10 positions are required

10 ISO PART 2: Monitoring to provide evidence of cleanroom performance related to air cleanliness by particle concentration The previous version of this standard was titled Specifications for testing and monitoring to prove continued compliance with ISO This standard is now directed more towards Continuous Monitoring. However, this leaves a problem as the schedule for retesting has now been removed from this standard

11 ISO14644 Parts 1 & 2 have now been published. Part 3 was to have been published at the same time, however this did not occur. Major changes: Table for scheduled testing has been moved form Part 2 and will be included in Part 3. Determination of Room Recovery Rates has been amended.

12 ISO14644 PART Occupancy states as-built condition where the installation is complete with all services connected and functioning but with no production equipment, materials, or personnel present at-rest condition where the installation is complete with equipment installed and operating in a manner agreed upon by the customer and supplier, but with no personnel present operational condition where the installation is functioning in the specified manner, with the specified number of personnel present and working in the manner agreed upon

13 A3 Pre-test conditions: ISO14644 PART 3 Special care should be taken when determining the sequence for performing tests for cleanroom performance. The results of each cleanroom test may be dependent upon other pre-conditioning requirements being met. For example, some pre-conditions for the Pressure Difference Test are stated in B1.2.

14 ISO14644 PART 3 Table A2 Typical Test Frequency Type of selected test Pressure differentials Air supply volume in non-unidirectional airflow Air velocity in unidirectional airflow Installed filter system leakage - non-unidirectional airflow cleanroom Installed filter system leakage - unidirectional airflow cleanroom Air direction & visualisation Recovery rate Containment test Particle desposition rates Segregation tests Temperature & humidity Electrostatic & ion generator Typical period to retest 12 months 12 months 12 months 24 months 12 months 12 months

15 Revision to Room Recovery Rates Section B.4: For non-unidirectional cleanroom only This test now incorporates 100:1 and 10:1 tests. Test should be performed when the clean room is in the As Built or At Rest states, or after major modification to the cleanroom, or its operation. No requirement to retest every 2 years. It is not recommended that the 100:1 test be used for ISO 8 & 9

16 B4.3.4 Evaluation of Recovery rate: Recovery performance can be determined from the slope of particle concentration decreasing curve, as follows: a) Commence measurements and record time and concentration continuously. Sampling time should be as short as possible but sampling should be such that the count has statistical relevance. Time intervals between the samplings should be as short as possible. b) Plot the data of decreasing particle concentration on a semi-log chart (concentrations on the ordinate by the logarithmic scale, and the time values on the abscissa by the linear scale) c) Decide upper and lower concentration limits as to the decreasing curve measured is accepted as almost straight line. d) Cleanliness recovery rate is obtained from the slope of the line between the upper and lower concentrations. The cleanliness recovery rate between two measurements is calculated from the following equation: r = -2.3 x 1 log ( C1) t ( C0)

17 Where r is the cleanliness recovery rate; t1 is the time elapsed between measured concentration crosses C0 and C1 C0 is the upper concentration; C1 Is the lower concentration at t1 If it is possible to measure concentration at plural timing until the cleanliness is recovered, recovery rate can be evaluated using the least squares method. Average five to ten of recovery rate values obtained in a measurement. The recovery rate and 100:1 recovery time can be related as follows: r = -2.3 x 1 log¹º ( 1 ) = -2.3 x 1 (-2) = 4.6 x 1 t0.01 (100) t0.01 t0.01 where r is the cleanliness recovery rate; t0.01 is the time elapsed between measured concentration crosses initial and 0.01 of initial concentration.;

18 Annex C lists the Test apparatus required for each test. This annex describes the measuring apparatus that should be used for the recommended tests given in this part of ISO In this annex, data given in Tables C.1 to C.22 indicate the minimum necessary requirements for each item of apparatus. Items are listed and numbered to correspond with Annex B, e.g. the apparatus numbered C.1 is used in the test procedure given in B.1. Those responsible for planning tests may refer to Annex C for the selection of test apparatus and to Annex A for a checklist of recommended tests of an installation and the sequence in which to carry them out. Measuring apparatus should be chosen subject to agreement between the customer and supplier. This annex is informative, and should not prevent the use of improved apparatus as it becomes available. Alternative test apparatus may be appropriate and may be used subject to agreement between customer and supplier. Test apparatus should be selected with measurement limits and range that are appropriate for its application. The apparatus should also be calibrated with calibration points covering the range of its intended use. All test apparatus sensitivity should be 1.

19 Major differences when comparing ISO14644 Part 3 to testing performed by most Australian Testing Companies: Many different piece of test equipment can be utilised. Other test methods included in Part 3, not normally performed in Australia i.e. Sample testing of installed duct HEPA Filter installations. Testing of HEPA filters using DPC (Discrete Particle Counter). Air volume testing using Airflow Hoods / pitot tubes

20 ISO29643 ISO29643 High efficiency filters and filter media for removing particle in air is the new ISO Standard for HEPA Filters. It was published in October, It is broken into 5 parts, as follows: Part 1: Classification, performance testing & marking Part 2: Aerosol production, measuring equipment & particle counting statistics Part 3: Testing flat sheet filter media Part 4: Test method for determining leakage of filter elements Scan method Part 5: Test method for filter element

21 COMPARISON OF ISO29643 & EN1822 Efficiency (%) Penetration (%) E 10 > ISO 15 E E 11 > 95 5 ISO 20 E > 99 1 ISO 25 E E 12 > 99,5 0,5 ISO 30 E > 99,9 0,1 ISO 35 H H 13 > 99,95 0,05 ISO 40 H > 99,99 0,01 ISO 45 H H 14 > 99,995 0,005 ISO 50 U > 99,999 0,001 ISO 55 U U 15 > 99, ,000 5 ISO 60 U > 99, ,000 1 ISO 65 U U 16 > 99, , ISO 70 U > 99, , ISO 75 U U 17 > 99, ,

22 PARTICLE COUNTERS ISO21501 PART 4 The new released version ISO14644 Part 1 states: Not all instruments will be able to meet the requirements of this standard, in this case, the use of non compliant instruments will require an additional explanations

23 PARTICLE COUNTERS ISO21501 PART 4 ISO has the title Determination of particle size distribution Single particle light interaction methods. Included with ISO are four parts: Part 1: Light scattering aerosol spectrometer Part 2: Light scattering liquid-borne particle counter Part 3: Light extinction liquid-borne particle counter Part 4: Light scattering airborne particle counter for clean spaces The documentation and approval of ISO , -3, and -4 subsequently replaces and cancels ISO :2000. The ISO standard widens the scope of analysis to include methodology for airborne particle counting and liquid particle counting (both light scattering and extinction methods). Specifically, ISO provides a calibration procedure and verification method for airborne particle counters to minimize inaccurate measurements and reduce variations between different instruments. These new guidelines require pulse height analysis (PHA) for particle counter calibrations, which reduces inconsistencies.

24 Pulse Height Analyzer (PHA) Particle counters employ solid-state photodetectors that convert detected light energy into electrical signals. The particle signals, known as pulses, leave the detector with proportional amplitudes of energy that represent the particles or photons witnessed by the detectors. If these pulses are sorted according to their height or magnitude, we could equate the pulses to energy received from the particle or photon. Electronic systems that analyze particle pulses and categorize their respective intensities are called Pulse Height Analyzers (PHAs).

25 Single channel PHAs count only pulses of certain amplitudes, which are equivalent to a particle s or photon s perceived energy. Each pulse is placed into a counting bin, called a channel, and totaled with other pulses that are of the same amplitude. In effect, many smaller 0.1cfm DPC s, and those manufactured prior to say 2009 are no longer compliant with this standard.

26 THE END!!! THANK YOU

27 QUESTIONS? Connecting Pharmaceutical Knowledge ispe.org 27

EU GMP Annex 1 Update 2008: Airborne Particle Counting

EU GMP Annex 1 Update 2008: Airborne Particle Counting EU GMP Annex 1 Update 2008: Airborne Particle Counting Airborne Particle Counting for Pharmaceutical Facilities: Update 2008, EU GMP Annex 1 Morgan Polen - VP of Applications Technology, Lighthouse Worldwide

More information

International Standard ISO 14644

International Standard ISO 14644 International Standard ISO 14644 Cleanrooms and associated controlled environments Praphon Angtrakool Food and Drug Administration 1 International Standard : ISO 14644 ISO 14644 consists of the following

More information

Regulations Concerning

Regulations Concerning Regulations Concerning Airborne Particle Counting Two types of activities performed with an airborne particle counter: Classification focuses primarily on the environment Particle Counting Applications

More information

Particle Monitoring Requirements in Pharmaceutical Cleanrooms

Particle Monitoring Requirements in Pharmaceutical Cleanrooms Particle Monitoring Requirements in Pharmaceutical Cleanrooms All drugs must be manufactured in accordance with the current Good Manufacturing Practice (cgmp) regulations. Pharmaceutical manufacturers

More information

ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners

ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners Tony Harrison and Bob Latimer ISO 21501- A Standard Methodology to Optical Particle Counter

More information

Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested

Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested DOP Solutions Technical Document: Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested Issue 1: Prepared by Neil Stephenson and Tim Triggs and edited by

More information

New changes to cleanroom & clean air device classifications: ISO 14644 1 & 2

New changes to cleanroom & clean air device classifications: ISO 14644 1 & 2 New changes to cleanroom & clean air device classifications: ISO 14644 1 & 2 This white paper describes the changes outlined in the Draft International Standard (DIS) editions of ISO 14644-1 and ISO 14644-2.

More information

Cleanrooms and associated controlled environments. Part 1: Classification of air cleanliness by particle concentration

Cleanrooms and associated controlled environments. Part 1: Classification of air cleanliness by particle concentration INTERNATIONAL STANDARD ISO 14644-1 Second edition 2015-12-15 Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness by particle concentration Salles propres et environnements

More information

Air and Water Balancing - Sound and Vibration Testing - Cleanroom Services - Building Commissioning Building Envelope Testing Thermal Imaging

Air and Water Balancing - Sound and Vibration Testing - Cleanroom Services - Building Commissioning Building Envelope Testing Thermal Imaging NEUDORFER ENGINEERS, INC. provides our customer with unbiased, professional, performance testing of Cleanrooms. We are committed to providing you, our customer, with value-added expertise in an effort

More information

CLASSIFICATION OF CLEANROOMS

CLASSIFICATION OF CLEANROOMS CLASSIFICATION OF CLEANROOMS Cleanrooms are classified by the cleanliness of their air. The method most easily understood and most universally applied is the one suggested in the earlier versions (A to

More information

CLEAN ROOM MONITORING REGULATORY STANDARDS

CLEAN ROOM MONITORING REGULATORY STANDARDS CLEAN ROOM MONITORING REGULATORY STANDARDS Air Classification as per Schedule M Maximum permitted number of particles / m3 equal or above at rest in operation 0.5µm 5.0µm 0.5µm 5.0µm A 3,520 29 3,500 29

More information

FS209E and ISO Cleanroom Standards

FS209E and ISO Cleanroom Standards FS209E and ISO Cleanroom Standards Terra Universal is the leading expert in the design and fabrication of critical-environment applications. We offer a complete range of equipment, furnishing and supplies

More information

Environmental Monitoring

Environmental Monitoring Environmental Monitoring Purpose of Environmental Monitoring Critical process within the pharmaceutical and biotechnology industries. Determines the microbial and particulate content of cleanroom air and

More information

GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS

GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 032-2 8 January 2010 RECOMMENDATION GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE

More information

TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS

TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS APPLICATION NOTE CC-105 (US) Introduction This application note will demonstrate how real-time viable particle count data will

More information

Particle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences

Particle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences Particle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences LIGHTHOUSE WORLDWIDE SOLUTIONS WWW.GOLIGHTHOUSE.COM Introduction The pharmaceutical and medical device industries are developing

More information

VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING. F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group

VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING. F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group tel + 3226431600 E-Mail: [email protected] VALIDATION OF CLEAN

More information

Risk-Based Environmental Monitoring. Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI

Risk-Based Environmental Monitoring. Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI Risk-Based Environmental Monitoring Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI Presenter Marsha Stabler Hardiman Over 20 years experience in the

More information

Revision of ISO 14644-1:1999 A progress report and explanation of some of the key issues and principles June 2012

Revision of ISO 14644-1:1999 A progress report and explanation of some of the key issues and principles June 2012 Revision of ISO 14644-1:1999 A progress report and explanation of some of the key issues and principles June 2012 Gordon J Farquharson Convenor ISO TC209 WG1, Critical Systems Ltd, 4 Greencroft, Guildford,

More information

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS UNDERSTANDING, ACCELERATED TSI SETS THE STANDARD FOR PARTICLE COUNTING Performance Semiconductor, hard disk drive, and flat panel manufacturers

More information

Qualification of Cleanrooms and Environmental Monitoring - Practical Approaches in Europe

Qualification of Cleanrooms and Environmental Monitoring - Practical Approaches in Europe Qualification of Cleanrooms and Environmental Monitoring - Practical Approaches in Europe 1 Dr. Doris Herrmann Zum Eselsbachtal 20 67657 Kaiserslautern [email protected] www.gmp-unlimited.com

More information

PROCEDURAL STANDARDS FOR CERTIFIED TESTING OF CLEANROOMS

PROCEDURAL STANDARDS FOR CERTIFIED TESTING OF CLEANROOMS PROCEDURAL STANDARDS FOR CERTIFIED TESTING OF CLEANROOMS 2009 THIRD EDITION National Environmental Balancing Bureau 8575 Grovemont Circle Gaithersburg, Maryland 20877 301-977-3698 301-977-9589 FAX www.nebb.org

More information

Cleanroom. For. Sterile Manufacturing Facilities

Cleanroom. For. Sterile Manufacturing Facilities Cleanroom For Sterile Manufacturing Facilities Praphon Angtrakool Food and Drug Administration 1 WHO TRS No. 823 Annex 1, 1992 (1) General 17.1 The production of sterile preparations should be carried

More information

Environmental Monitoring of Clean Rooms

Environmental Monitoring of Clean Rooms Environmental Monitoring of Raul Duarte President DDK Scientific, Corp. Copyright DDK Scientific, Corp. 2008, 2009, DDK Scientific, Corp. Proprietary A manufacturing facility for pharmaceutical products

More information

Manufacturing. Environmental Monitoring Particle Counts Are Easy B IOP ROCESSTECHNICAL. Scott E. Mackler

Manufacturing. Environmental Monitoring Particle Counts Are Easy B IOP ROCESSTECHNICAL. Scott E. Mackler B IOP ROCESSTECHNICAL Environmental Monitoring Particle Counts Are Easy Scott E. Mackler Manufacturing pharmaceutical and biotechnology products requires that the appropriate level of quality be designed

More information

cleanroom validation Cleanroom Validation

cleanroom validation Cleanroom Validation Cleanroom Validation cleanroom applications laboratories production environments laminar flow boxes cleanroom concepts microelectronics pharmaceutical industry, food industry nuclear industry, aerospace

More information

Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations

Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations B e c k m a n C o u l t e r L i f e S c i e n c e s 2 5 0 S. K r a e m e r B o u l e

More information

Contamination Control and Environmental Monitoring Systems

Contamination Control and Environmental Monitoring Systems Life Science Contamination Control and Environmental Monitoring Systems Hospital Electronics Brookhaven Instruments deliver, install and validate particle counters, Microbial Air Samplers and Monitoring

More information

Environmental Monitoring

Environmental Monitoring Chapter 10 Environmental Monitoring By the end of this chapter, you will be able to: l List the types of environmental testing l Name three types of checks that are carried out to ensure the isolator is

More information

ENVIRONMENTAL MONITORING

ENVIRONMENTAL MONITORING ENVIRONMENTAL MONITORING Assessment and verification of the adequacy of the aseptic compounding environment is essential. Environmental monitoring programs are designed to promptly identify potential sources

More information

Validating and Monitoring the Cleanroom

Validating and Monitoring the Cleanroom 6 Validating and Monitoring the Cleanroom Kevina O Donoghue Overview... 6-1 Cleanroom Design... 6-1 Principles of Cleanroom Validation and Testing... 6-2 Cleanroom Validation... 6-2 So How Does One Determine

More information

Qualification of an Environmental Monitoring Program

Qualification of an Environmental Monitoring Program [ Scott Sutton Qualification of an Environmental Monitoring Program Scott Sutton Microbiology Topics discusses various topics in microbiology of practical use in validation and compliance. We intend this

More information

Indoor air quality monitoring made easy

Indoor air quality monitoring made easy Indoor air quality monitoring made easy with the Fluke 985 Particle Counter Application Note Introduction Energy savings is top of facilities owners agenda. HVAC system tends to be the largest contributor

More information

Surface Scanning Probes for Lighthouse Particle Counters Solair Gen D/E

Surface Scanning Probes for Lighthouse Particle Counters Solair Gen D/E 1. Why surface particulate detection? In many branches of industry the manufacture of an ultraclean surface is crucial to the functionality and quality of the product. Particles have an impact on the overall

More information

GUIDELINES ON TEST METHODS FOR ENVIRONMENTAL MONITORING FOR ASEPTIC DISPENSING FACILITIES

GUIDELINES ON TEST METHODS FOR ENVIRONMENTAL MONITORING FOR ASEPTIC DISPENSING FACILITIES GUIDELINES ON TEST METHODS FOR ENVIRONMENTAL MONITORING FOR ASEPTIC DISPENSING FACILITIES PRODUCED BY A WORKING GROUP OF THE SCOTTISH QUALITY ASSURANCE SPECIALIST INTEREST GROUP SECOND EDITION FEBRUARY

More information

Non-destructive testing of steel forgings - Part 2: Penetrant testing (BS EN 10228-2:1998)

Non-destructive testing of steel forgings - Part 2: Penetrant testing (BS EN 10228-2:1998) Non-destructive testing of steel forgings - Part 2: Penetrant testing (BS EN 10228-2:1998) 1 Scope This part of EN 10228 describes the method and acceptance criteria to be used for the penetrant testing

More information

Experiment (13): Flow channel

Experiment (13): Flow channel Introduction: An open channel is a duct in which the liquid flows with a free surface exposed to atmospheric pressure. Along the length of the duct, the pressure at the surface is therefore constant and

More information

Laminar Flow Cabinets

Laminar Flow Cabinets Laminar Flow Cabinets The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. 24 36 48 Simple, Effective Protection for Samples and Work Processes, Uncompromised Performance, Economical

More information

Testing HEPA/ULPA filters at the manufacturer s facility and in situ on the user s premises

Testing HEPA/ULPA filters at the manufacturer s facility and in situ on the user s premises Publication Series Viledon Testing HEPA/ULPA filters at the manufacturer s facility and in situ on the user s premises Dipl.-Ing. Thomas Schroth Dr.-Ing. Thomas Caesar Lecture at 4th CLEANROOMS Europe

More information

ASTACEA4 Inspect complex/central air conditioning systems

ASTACEA4 Inspect complex/central air conditioning systems Overview This Unit covers the competences required to inspect complex/central air conditioning systems as defined by the CIBSE TM 44 Figure 1.1: Summary of system types and their component parts. The air

More information

SPECIAL APPLICATIONS

SPECIAL APPLICATIONS Page 1 SPECIAL APPLICATIONS The protection of microelectronics clean rooms using aspirating smoke detection systems requires special consideration. The physical configuration of the room(s) and the direction

More information

CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010

CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010 CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010 1. Background Class II Type B Biosafety Cabinets (BSCs) are unique laboratory

More information

1.0 What Are the Purpose and Applicability of Performance Specification 11?

1.0 What Are the Purpose and Applicability of Performance Specification 11? While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data.

In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data. MATHEMATICS: THE LEVEL DESCRIPTIONS In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data. Attainment target

More information

APPENDIX A-1 SAMPLE CLEANROOM SPECIFICATIONS MICROELECTRONICS & SEMICONDUCTORS

APPENDIX A-1 SAMPLE CLEANROOM SPECIFICATIONS MICROELECTRONICS & SEMICONDUCTORS APPENDIX A-1 SAMPLE CLEANROOM SPECIFICATIONS SECTION xxxxx CLEANROOM PERFORMANCE TESTING PART 1 -- GENERAL 1.1 WORK INCLUDED A. This Section specifies the requirements of the NEBB Certified Cleanroom Performance

More information

Stainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6.

Stainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6. Product Sampler #1 Laminar Flow Hoods / Laminar Flow Cabinets: 1- Benchtop Vertical Laminar Flow Hoods- Air Recirculation Available s: Clean Bench Class 100 Bench-top Vertical Laminar Flow Hoods with Air-recirculation

More information

Approval Block. Prepared by: Signature Date. Evan Parnell 08 NOV 2013. Reviewed by: Signature Date. Approved by: Signature Date

Approval Block. Prepared by: Signature Date. Evan Parnell 08 NOV 2013. Reviewed by: Signature Date. Approved by: Signature Date ATS-ATP-3460 Page: 1 of 5 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 14644-4 First edition 2001-04-01 Cleanrooms and associated controlled environments Part 4: Design, construction and start-up Salles propres et environnements maîtrisés apparentés

More information

The use of risk assessment tools for microbiological assessment of cleanroom environments. by Tim Sandle

The use of risk assessment tools for microbiological assessment of cleanroom environments. by Tim Sandle The use of risk assessment tools for microbiological assessment of cleanroom environments by Tim Sandle Email: [email protected] / [email protected] Web: www.pharmig.blogspot.com Environmental

More information

Annex 4 Supplementary guidelines on good manufacturing practices: validation

Annex 4 Supplementary guidelines on good manufacturing practices: validation World Health Organization WHO Technical Report Series, No. 937, 2006 Annex 4 Supplementary guidelines on good manufacturing practices: validation 1. Introduction 2. Scope 3. Glossary 4. Relationship between

More information

How To Control A Medical Device With A Pc System

How To Control A Medical Device With A Pc System Clean room monitoring and disinfection FMS 5.0, the Leading Facility Monitoring Software TSI FMS system is an advanced, reliable, and user-friendly monitoring software suite that features a truly open

More information

SECTION 23 05 93 TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION

SECTION 23 05 93 TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION SECTION 23 05 93 TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION The purpose of this guide specification is to assist the specifier in correctly specifying Mechanical System Testing

More information

Duct Leakage Testing. Ken Helfers, TSI Inc. 2011, TSI Incorporated

Duct Leakage Testing. Ken Helfers, TSI Inc. 2011, TSI Incorporated Duct Leakage Testing Ken Helfers, TSI Inc. Agenda Need for duct leakage testing Duct leakage testing standards and regulations Performing duct leakage testing Why Test Ducts for Leakage? Conserve energy

More information

Certification system of air filter quality (CAF mark)

Certification system of air filter quality (CAF mark) Certification system of air filter quality (CAF mark) KOREA AIR CLEANING ASSOCIATION 1. Overview What is the certification system of air filter quality (CAF mark)? It is a mark granted by the KACL depending

More information

Hazardous Drug. Cleanroom. Standards

Hazardous Drug. Cleanroom. Standards Hazardous Drug Cleanroom Standards Module 3 Table of Contents Section A... 1 A.1 International Standards Organization (ISO) Classifications... 1 A.2 Hazardous Drug Cleanroom and Anteroom... 1 A.2.1 Location

More information

Duplexvent DV72 EC So light, so slim it fits inside a 600mm cupboard

Duplexvent DV72 EC So light, so slim it fits inside a 600mm cupboard Duplexvent DV72 EC So light, so slim it fits inside a 600mm cupboard Duplexvent DV72 EC MVHR is designed to provide constant volume ventilation levels to exceed the latest Building Regulations 2010. The

More information

HEPA Filters for Pharma Applications Cleanrooms & Safety Applications. Gordon Farquharson June 2012

HEPA Filters for Pharma Applications Cleanrooms & Safety Applications. Gordon Farquharson June 2012 HEPA Filters for Pharma Applications Cleanrooms & Safety Applications Gordon Farquharson June 2012 1 Air Filters at work Images courtesy of CamfilFarr 2 How filters work Sieving Effect Straining 3 Filter

More information

Pharmaceutical Wholesaler Site Inspection Checklist

Pharmaceutical Wholesaler Site Inspection Checklist Pharmaceutical Wholesaler Site Inspection Checklist Date and time of inspection: Legal business name and plant name: Location (address): Phone number: Fax number: Email address: Web site URL: Contact Person:

More information

Check-list for auditors. Audit visit at candidate registered laboratories

Check-list for auditors. Audit visit at candidate registered laboratories Check-list for auditors Audit visit at candidate registered laboratories 1 Contents Introduction Laboratory documentation required by the standard(s), available for inspection by the auditor Items to be

More information

Summarizing and Displaying Categorical Data

Summarizing and Displaying Categorical Data Summarizing and Displaying Categorical Data Categorical data can be summarized in a frequency distribution which counts the number of cases, or frequency, that fall into each category, or a relative frequency

More information

Risk Based Environmental Monitoring (EM) and EM Data Management and Trending

Risk Based Environmental Monitoring (EM) and EM Data Management and Trending Risk Based Environmental Monitoring (EM) and EM Data Management and Trending An Industry Roundtable Discussion Marsha Stabler Hardiman and Parsa Famili March 11, 2015 PDA NE Chapter Meeting Presenters

More information

Hydraulics Laboratory Experiment Report

Hydraulics Laboratory Experiment Report Hydraulics Laboratory Experiment Report Name: Ahmed Essam Mansour Section: "1", Monday 2-5 pm Title: Flow in open channel Date: 13 November-2006 Objectives: Calculate the Chezy and Manning coefficients

More information

THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS

THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS INDOOR AIR QUALITY INSTRUMENTS UNDERSTANDING, ACCELERATED BREATHE A LITTLE EASIER WITH TSI Indoor air quality is a growing concern. With the increasing

More information

How to Estimate the Cost of a. Laboratory Renovation. Candidate Number: 0714144

How to Estimate the Cost of a. Laboratory Renovation. Candidate Number: 0714144 How to Estimate the Cost of a Laboratory Renovation Candidate Number: 0714144 December 15, 2014 1 Table of Contents Cover Page Page 1 Table of Contents Page 2 Introduction Page 3 Types & Methods of Measurements

More information

Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice

Working Party on Control of Medicines and Inspections. Final Version of Annex 15 to the EU Guide to Good Manufacturing Practice EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 Working Party on Control

More information

G U I D E L CERTIFICATION EDUCATION TRAINING. Procurement Guideline. National Air Filtration Association Recommended Practice

G U I D E L CERTIFICATION EDUCATION TRAINING. Procurement Guideline. National Air Filtration Association Recommended Practice G U National Air Filtration Association Recommended Practice I D E L I N E Procurement Guideline TRAINING EDUCATION CERTIFICATION HVAC Air Filter Purchasing Specification Guideline 1.0 Purpose To provide

More information

Acceleration of Gravity Lab Basic Version

Acceleration of Gravity Lab Basic Version Acceleration of Gravity Lab Basic Version In this lab you will explore the motion of falling objects. As an object begins to fall, it moves faster and faster (its velocity increases) due to the acceleration

More information

MATHS LEVEL DESCRIPTORS

MATHS LEVEL DESCRIPTORS MATHS LEVEL DESCRIPTORS Number Level 3 Understand the place value of numbers up to thousands. Order numbers up to 9999. Round numbers to the nearest 10 or 100. Understand the number line below zero, and

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

Surface Cleanliness Martin Derks

Surface Cleanliness Martin Derks Surface Cleanliness Martin Derks www.golighthouse.com 1 Contamination Monitoring Solutions: Contamination: "Any foreign material or energy that has a detrimental effect on a product or process Monitoring

More information

HVAC Air Systems. Recirculation Air System Types

HVAC Air Systems. Recirculation Air System Types HVAC Air Systems Recirculation Air System Types Summary Recirculation air handler fan energy can account for 10 to 30% of total cleanroom energy use. There are three fundamental recirculation system configurations:

More information

White Paper. Life Sciences Industry Solution Blueprint for Environmental Monitoring Systems (EMS) ISPE Guidance and Best Practices.

White Paper. Life Sciences Industry Solution Blueprint for Environmental Monitoring Systems (EMS) ISPE Guidance and Best Practices. White Paper Life Sciences Industry Solution Blueprint for Environmental Monitoring Systems (EMS) Author: Roberto Zerbi, Global Industries Solutions Life Sciences, Invensys What s Inside: 1. Introduction

More information

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE Company Name Street Address City, State, Zip code Phone Number Fax Company Website Email Address ORGANIZATION NAME PHONE NUMBER EMAIL ADDRESS President/CEO General Manager Engineering Manager Production

More information

Workflow System for Paperless Air Monitoring Powered by MODA. Presented by

Workflow System for Paperless Air Monitoring Powered by MODA. Presented by Introducing Hach s MET ONE Compliance Workflow System for Paperless Air Monitoring Powered by MODA Presented by Scott Hjelmervik Bob Toal Systems Director Vice President Hach Company MODA Technology Partners

More information

Perimeter, Area and Volume What Do Units Tell You About What Is Being Measured? Overview

Perimeter, Area and Volume What Do Units Tell You About What Is Being Measured? Overview Perimeter, Area and Volume What Do Units Tell You About What Is Being Measured? Overview Summary of Lessons: This set of lessons was designed to develop conceptual understanding of the unique attributes

More information

HEPA FILTER DEFECTS ARE COMMON

HEPA FILTER DEFECTS ARE COMMON HEPA FILTER DEFECTS ARE COMMON Kenneth F. Cherry 1 and Kenneth A. Smigielski 2 1 NCEC Inc. 2 First Solar LLC. ABSTRACT HEPA filters (High Efficiency Particulate Air) filters are claimed by vendors to have

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

SAM HOUSTON STATE UNIVERSITY

SAM HOUSTON STATE UNIVERSITY PART 1: GENERAL 1.01 General Requirements A. This standard is intended to provide useful information to the Professional Service Provider (PSP) to establish a basis of design. The responsibility of the

More information

Information. Cleanroom Product Range-subjected to change

Information. Cleanroom Product Range-subjected to change Cleanroom Product Range-subjected to change 123 General information Introduction to Cleanroom The term Cleanroom is something you associate with in modern industries. However, the root of cleanroom design

More information

Chapter 12 Industrial Ventilation

Chapter 12 Industrial Ventilation Chapter 1 Industrial Ventilation Introduction Industrial Ventilation is an important method for reducing employee exposures to airborne contaminants. Ventilation is used to reduce or remove contaminates

More information

SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES (GMP): VALIDATION

SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES (GMP): VALIDATION WORLD HEALTH ORGANIZATION ORGANISATION MONDIALE DE LA SANTE Working document QAS/03.055/Rev.2 RESTRICTED SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES (GMP): VALIDATION This document has followed

More information

Pharmagraph. GMP/GLP Monitoring System

Pharmagraph. GMP/GLP Monitoring System Pharmagraph GMP/GLP Monitoring System Pharmagraph GMP/GLP Monitoring System Demonstrates GMP/GLP compliance for cleanrooms and laboratories Monitors, Alarms, Logs and Reports Airborne particle counts,

More information

SUPPLEMENTARY GUIDELINES ON

SUPPLEMENTARY GUIDELINES ON August 2015 Draft document for comment 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 SUPPLEMENTARY GUIDELINES ON GOOD MANUFACTURING PRACTICES

More information

Lecture 30: Cleanroom design and contamination control

Lecture 30: Cleanroom design and contamination control Lecture 30: Cleanroom design and contamination control Contents 1 Introduction 1 2 Contaminant types 2 2.1 Particles.............................. 2 2.2 Metal ions............................. 4 2.3 Chemicals.............................

More information

New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01)

New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01) Safeguarding public health New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01) Tony Orme, Senior GDP Inspector Inspection, Enforcement and Standards Division

More information

Test Code: 8297 / Version 1

Test Code: 8297 / Version 1 Blueprint HVAC Maintenance Technology PA Test Code: 8297 / Version 1 Copyright 2014. All Rights Reserved. General Assessment Information HVAC Maintenance Technology PA Blueprint Contents General Assessment

More information

PANDA SPRAY BOOTH FILTER INFORMATION

PANDA SPRAY BOOTH FILTER INFORMATION PANDA SPRAY BOOTH FILTER INFORMATION Our Spray Booth has been designed to reduce the risk of potential hazardous situations occurring. You must adhere to the specifications chart listed below when replacing

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

Fire Alarm system Installation Guide

Fire Alarm system Installation Guide Fire Alarm system Installation Guide Fire Alarm System Categories Systems designed for Protection of Property only, fall into two classifications P1 or P2. but a manual system is also needed, then this

More information

Applying LDV to Air and Water Flow Measurements: Research Activities & Accreditation. Matthew A. Rasmussen, Danish Technological Institute

Applying LDV to Air and Water Flow Measurements: Research Activities & Accreditation. Matthew A. Rasmussen, Danish Technological Institute Applying LDV to Air and Water Flow Measurements: Research Activities & Accreditation Matthew A. Rasmussen, Danish Technological Institute Euramet Flow Meeting Danish Technological Institute, Taastrup 10-12

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

The Prophotonix (UK) Ltd Quality manual

The Prophotonix (UK) Ltd Quality manual The Prophotonix (UK) Ltd Quality manual Date: March 2014 Revision: D Sparrow lane, Hatfield Broad Oak, Herts, UK, CM22 7BA Tel: +44 (0)1279 717170 Fax: +44 (0)1279 717171 e-mail: [email protected] Page

More information

Wireless Particle Monitoring of Pharmaceutical Cleanrooms

Wireless Particle Monitoring of Pharmaceutical Cleanrooms Wireless Particle Monitoring of Pharmaceutical Cleanrooms Introduction Wireless technology has brought many innovative changes, enabling us to communicate in new ways. This article will explain how wireless

More information

Particle Counting. - Theory - Guidelines - Monitoring

Particle Counting. - Theory - Guidelines - Monitoring - Theory - Guidelines - Monitoring 1 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO 14644 Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification

More information

EA-04/15. Accreditation. For Non-Destructive. Testing. Publication. Reference. EA - 4/15 Accreditation for Non-Destructive Testing PURPOSE

EA-04/15. Accreditation. For Non-Destructive. Testing. Publication. Reference. EA - 4/15 Accreditation for Non-Destructive Testing PURPOSE Publication Reference EA-04/15 Accreditation For Non-Destructive Testing PURPOSE This publication provides detailed guidance for organisations carrying out non-destructive testing as an accredited activity

More information

Flow Measurement Options for Pipeline and Open Channel Flow

Flow Measurement Options for Pipeline and Open Channel Flow Flow Measurement Options for Pipeline and Open Channel Flow October 2013 Presented by Molly Skorpik - 2013 Montana Association of Dam and Canal Systems Conference Irrigation Training and Research Center

More information

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter 2 Table of Contents Abstract... 3 Pressure Fundamentals... 3 Applications...

More information

Amplified High Speed Fiber Photodetectors

Amplified High Speed Fiber Photodetectors Amplified High Speed Fiber Photodetectors User Guide (800)697-6782 [email protected] www.eotech.com Page 1 of 7 EOT AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE Thank you for purchasing your Amplified

More information