Particle Counting. - Theory - Guidelines - Monitoring
|
|
|
- George Chapman
- 9 years ago
- Views:
Transcription
1 - Theory - Guidelines - Monitoring 1
2 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 2
3 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 3
4 Why do we care about Particle Counts? Particles as contamination negatively affect a process or product Change chemical composition Affect stability Affect purity Affect safety Affect reliability Contamination affects the yield of a process Increases cost to produce Increases cost to maintain or service Increases cost of ownership [COO] Decreases ROI of capital equipment purchase 4
5 Why do we care about Particle Counts? Pharmaceutical, Biotechnology Particles in injection could cause occlusion of blood vessels Red Blood cells are about 5 µm Capillary (5 to 10 µm) Large veins (10 to 50 µm) Viables in injection can trigger infection Possibility of reaction to foreign substances (RES/allergic reaction) Medical Devices Poor adhesion of medicated coating in stent creates embolisms Medical and photographic films Contamination prevents complete image recovery 5
6 Why monitor for particles? Greatest concern is for viable microorganisms Technology is not available today to measure viable counts in real time Requires incubation time Total non-viable particle counts used as a surrogate Non-viable counts Includes all types of airborne material Solid particles Fibers Microorganisms Skin flakes Non-viable particle count monitoring is a critical component of the total Environmental Monitoring program Non-viable particle counting offers potential for real-time response 6
7 Relative Sizes of Particulate Matter Human hair: µm Visible: µm Skin Flake: about 20 µm Bacteria: 1 to 8 µm 7
8 Relative Sizes of Particles Table salt Bacteria Pencil cells 8
9 Particles Found in a Cleanroom Aluminosilicate 9
10 Guidance Suggested Monitoring Event driven Bacterium 2-8 µm 5 µm Baseline µm Virus: µm 10
11 How an airborne particle counter works Detector Particle Light Trap Laser Diode Mirror Principle: Light Scattering 11
12 Three Common Methods of Sampling Manually with a handheld or portable particle counter Sequentially with a manifold/scanner and particle counter Continuously with portable or remote particle counters 3400 Series Cleanroom classification Environmental monitoring Short-term online sampling HHPC Handheld particle counter Used for troubleshooting 6000 & 7000 Series for production monitoring Grade A & B areas 12
13 Manual Monitoring Advantages: - Traditional method of sampling -Simple - Flexible - Low Initial Cost Disadvantages: - Inconsistent in Time - Inconsistent in Position of Probe - Intrusive to Work Process - Usually is not continuous - Production worker often not informed of high counts - High Labor cost (labor intensive) 13
14 Sequential Monitoring Advantages: - Consistently samples fixed positions - Samples every position one or more times per hour - Can operate 24 hours per day - Frequent sampling of each point - Consistent sampling position - Statistical data for historical purposes - Consistency of data allows SPC analysis - Can be part of an FMS system - Single counter to calibrate A2400 Disadvantages: - Samples at fixed locations - Can miss events - Particle loss in tubing - Requires auxiliary pump(s) to maintain constant flow through all sample tubes R
15 Tubing Transport Loss 15
16 Advantages: - Able to detect fast, random events - Uses dedicated counter at each sample point - Avoids effects of long tubing runs - Consistent sample points - All points monitored simultaneously - Quick operator feedback - Build statistical database - Relate particle events to process activity opportunity to improve process Disadvantages: Continuous Monitoring - Requires good plan of monitoring strategy Sample points must be chosen at the beginning Decisions must be made about alarms and reports - More expensive to implement More instruments Utility support: data cables, vacuum lines - Calibration needed for each counter - Data must be processed or summarized
17 Flexibility Cost Labor Calibration Quality Flexibility Initial Cost Daily Labor Calibration Quality of data Manual Sequential Continuous 17
18 Cleanroom and Clean Zone Classification Classification is the formal process of qualifying the environment by the number of particles using a standard method (ISO 14644) Performed on a regular basis but not frequently Grade A areas: Six months, ISO 4.8 Grade B areas: Six months, ISO 5 Grade C, D areas: Annually, ISO 7/8 Standards define minimum number of points Based on area of cleanroom or clean zone Standards define minimum amount of air to be sampled Minimum volumes for statistically valid samples (typically 1 minute at 1 cfm) Grade A requires minimum 1 m 3 (Annex 1) Classification is a rigid protocol 18
19 Environmental Monitoring Determine readiness of room to carry out designated task Performed whenever relevant activity will occur daily work Typically performed with portable counters Number of sample points defined by risk assessment Area of cleanroom or clean zone Activity to be performed Risk to product Frequency and volume of sample points defined by risk assessment Frequent enough to show control Frequent enough to manage risk of product contamination Inferred guidance, not rigid by regulation 19
20 Aseptic Process Monitoring Governed by EMEA and FDA Guidance Augments ISO requirements for critical (Grade A) areas Recommends continuous monitoring Number of sample points based on risk assessment Not rigidly defined by guidance Location of points based on risk assessment Limited guidance on sampling location Sample volumes defined by type of equipment Published guidance, not rigid company specific SOPs 20
21 Regulations and Guidance for the Manufacture of Sterile Medicinal Products ISO and EU GMP Annex 1 21
22 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 22
23 Classification Standards for Airborne Particles General Air Monitoring Standards 1999 ISO Classification of air cleanliness ISO Specifications for testing and monitoring to prove continued compliance with ISO ISO Guidance on instrumentation to be used for testing for compliance with ISO ISO
24 ISO General Standard Covers All Industries Electronics Semiconductor Flat Panel Circuit Board Optical MEMS/Nanomachines Life Sciences Pharmaceutical Biotechnology Medical Devices Hospitals/Pharmacies Aerospace Launch Vehicles Satellites Commercial/Military Aircraft Laboratories Analytical Laboratories Universities Other Nuclear Photographic, X-ray films Automobile Painting Laboratory Aerospace Other Life Sciences Electronics 24
25 Classification Standard ISO Defines cleanroom classes in a zone Establishes minimum sampling volumes Purpose: Gather a sample volume with theoretically at least 20 particles for a statistically valid sample Establishes minimum number of points to classify area, based on statistical criteria Gather from a valid number of locations for a representative sample of the total air volume 25
26 Classification: ISO A sample size with at least 20 theoretical particles, or Minimum sample volume 2.0 liter Minimum sample time 1 minute Minimum number of locations whichever is greatest 1 with at least 3 samples total 26
27 Classification Limits ISO Class Number of Particles per Cubic Meter by Micrometer Size 0.1 μm 0.2 μm 0.3 μm 0.5 μm 1 μm 5 μm ISO ISO ISO 3 1, ISO 4 10,000 2,370 1, ISO 5 100,000 23,700 10,200 3, ISO 6 1,000, , ,000 35,200 8, ISO 7 352,000 83,200 2,930 ISO 8 3,520, ,000 29,300 ISO 9 35,200,000 8,320, ,000 27
28 What volume do I need to collect? ISO requires sample with 20 theoretical particles ISO allows 3520 particles (0.5um) in 1m 3 (1000 liters) Volume = 20 particles 3520 particles X 1000 liters Volume = 5.7 liters Greater than 2.0 liters so minimum sample volume is satisfied 28
29 ISO Minimum Sample Time at 1 CFM Time required (in minutes) at 1 cfm (28.3 lpm) flow rate with 1-minute limit imposed 0.1 um 0.2 um 0.3 um 0.5 um 1 um 5 um ISO Class ISO Class ISO Class ISO Class ISO Class ISO Class ISO Class ISO Class ISO Class
30 Continued Compliance: ISO Schedule of Tests to Demonstrate Continuing Compliance Test Parameter Class Maximum Time Interval Test Procedure Particle Count Test All classes ISO Months months > ISO 5 12 Months ISO Annex A Important Air Pressure note: If Particle counts and Differential ISO All Classes 12 Months Pressure Difference are monitored continuously (for Annex B5 example, by a Facility Monitoring System) then ISO the Airflow Maximum All Time Classes Interval 12 for Months re-certification Annex B4 can be extended 30
31 Example: ISO Classification Vial Washing Station 5 m Lyo 1 8 m Lyo 2 Calculations for Number of Points: Area of clean zone = 80 m² Take the SQRT (80) = 8.94 Rounding up to next integer = 9 sample positions 5 m Lyo 3 4 m 31
32 Example: ISO Classification Vial Washing Station Lyo 1 7 Lyo 2 8 Calculations for Number of Points: Area of clean zone = 80 m² Take the SQRT (80) = 8.94 Rounding up to next integer = 9 sample positions 9 Lyo 3 32
33 Example ISO Calculations Vial Washing System Freeze Dryer 1 Need to adjust for equipment in room. Under ISO , if you sample at 10 or more positions, you can avoid the added calculation of the UCL (Upper Confidence Limit). Calculation of the UCL is only mandated when the number of positions used is between 2 and 9. Best to sample near potential problem spots which are near entrances and exits and near operator positions. Freeze Dryer 2 Freeze Dryer 3 33
34 Example ISO Calculations Vial Washing System Freeze Dryer Need to adjust for equipment in room. 12 Freeze Dryer 2 Under ISO , if you sample at 10 or more positions, you can avoid the added calculation of the UCL (Upper Confidence Limit). Calculation of the UCL is only mandated when the number of positions used is between 2 and 9. Best to sample near potential problem spots which are near entrances and exits and near operator positions Freeze Dryer 3 34
35 Example ISO Calculations 1. Average sample data values at each position 2. Normalize the average to number of particles per cubic meter 3. Compare normalized value to the target class limit; normalized value at each and every sample point must be less than the limit for the given size and target room classification If the number of points sampled is more than 1 but less than 10, then the UCL factor must be applied: a) Calculate the standard deviation b) Use Student s T-factor from tables c) Calculate UCL d) Compare to classification limit; UCL must not exceed the applicable limit 35
36 Probable Revisions to ISO , Frequency of re-certification Now: If zone is ISO Class 5, every 6 months If zone is ISO Class 6, every 12 months Proposed: For all zones, every 12 months 36
37 Probable Revisions to ISO , Eliminate Student s T test for sample plans with 1 to 9 sample positions 3. Method of determining number of sample positions Replace with stated number of minimum sample positions as a look-up chart Based on 95% confidence levels not on SQRT of area May mean a small increase in the number of sample points 37
38 B Derive the minimum number of sampling locations NL from table 3. Table 3 shows the number of sample locations related to the area of each cleanroom or clean zone to be classified and provides at least 95% confidence that at least 90 % of all locations do not exceed the class limits. Different levels of confidence and verification can be specified and agreed upon by the customer and supplier. 38
39 Room Size (ft 2 ) Room Size (m 2 ) ISO : 1999 ISO : 2007 (95,90) FS209E Unidirectional (all classes) FS209E Turbulent Class 1000 FS209E Turbulent Class 10K FS209E Turbulent Class 100K
40 Probable Revisions to ISO , Remove possibility to classify at 5 micron for ISO Class 5 Limit number of 29 removed Note (b) Due to sampling difficulties affecting collection of large particles in low concentrations, specification of sampling by DPC is inappropriate for classification at the indicated class. 5. Indicate that single digit limits for ISO Class 1 and 2 create challenges to timely execution 40
41 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 41
42 Guidance for Airborne Particles Pharmaceutical Industry Compliance Guidance FDA cgmp Guidance for Industry Sterile Drug Products Produced by Aseptic Processing EU GMP Annex I (EMEA) Manufacture of Sterile Medicinal Products EU GMP Annex 1 is more rigid than FDA cgmp Guidance 42
43 EU GMP Annex From EU Annex 1 The guidance has been reviewed in the light of the international standard EN/ISO and amended in the interests of harmonisation but taking into account specific concerns unique to the production of sterile medicinal products. Emphasizes continuous monitoring for Grade A and recommends for Grade B areas A continuous measurement system should be used for monitoring the concentration of particles in the Grade A zone, and is recommended for the surrounding Grade B areas. Introduces concept of measuring one cubic meter for routine testing (Meaning: classification, not in-process monitoring) 43
44 Revisions EU GMP Annex μ particle counts particularly significant when several are seen consecutively and indicates up to 20/m 3 for Grade A and 29/m 3 for Grade B (now equivalent to ISO 5) Accepts that particle counting may not be possible at the point of fill due to the generation of particles or droplets from the product itself Clearly differentiates classification and monitoring The sample sizes taken for monitoring purposes using automated systems will usually be a function of the sampling rate of the system used. It is not necessary for the sample volume to be the same as that used for formal classification of clean rooms and clean air devices. - 1minute sampling time for continuous monitoring - 1m3 for classification 44
45 EU GMP Annex 1 Revised: 1 March 2009 Limits at 5 microns for Grade A 1 per cubic meter 20 per cubic meter At Rest In Operation Grade Maximum permitted number of particles/m 3 equal to or above 0.5 µm 5 µm 0.5 µm 5 µm A B C D not defined not defined 45
46 Intuitive User Interface! Wizard to conduct pass/fail test for ISO, EU-GMP, FS and BS standards Improved area and sample location management Easy sampling recipe management through Group menu Multiple user level log in for data security 46
47 Intuitive User Interface! Test wizard for ISO, EU-GMP, FS, and BS standard compliance Wizard to conduct pass/fail test for ISO, EU- GMP Annex I, FS 209E and BS5295 standards No expertise in standards required. Few clicks to compliance The wizard guides an operator step by step to sample data, process data and product reports 47
48 Intuitive User Interface! Vial Filling Room Freeze Dryer 1 12 Freeze Dryer 2 Better area/location management Freeze Dryer 3 Multiple locations with unique location ID and repeatable location name can be created A group of locations can be assigned to a larger cleanroom space called an area Location setting can be copied 48
49 Intuitive User Interface! Better area/location management 3400 # # # 3 49
50 Intuitive User Interface! Location #1 Location #2 Group = Sampling Recipe Group is a collection of pre-set sampling parameters and alarm limits Better group management Area #1 Location #4 Location #3 Location #N Individual Location setting Group is a collection of desired sampling parameters New group can be created or existing group can be loaded A location can be added to or removed from a group Settings button allows pre-selecting sampling parameters such as sampling time, delay, hold time, count and environmental alarm, etc. 50
51 Intuitive User Interface! 21 CFR Part 11 compliance: unique user log in Multiple level of user accounts Basic (no log in) Operator Administrator No limits to number of log in accounts An administrator can create operator level user account and assign selective access rights and group access 51
52 Met One : 100 LPM Flow Rate Higher flow rate Same packaging Same accessories and support documents Same battery functionality Wireless capability maintained Meets ISO Resolution Sensitivity 52
53 Met One 3445 Rapid classification and routine sampling of Pharmaceutical areas Battery-powered yet still light-weight Smooth Stainless Steel enclosure is designed with the needs to maintain ultra-clean or aseptic areas Color touchscreen provides a clear window into cleanroom contamination Supported by software for stand-alone use and for continuous monitoring requirements PortAll 2.4 UVO EnVigil Lite OPC Server 53
54 Met One 3445 revised 3400 Datasheet 54
55 Met One 3445 revised 3400 Datasheet 55
56 Met One Specifications 56
57 PortAll Software Organize, archive and report particle count data Download Utility for Airborne Particle Counters Two Versions Generic Industries Life Science Demo version provided with all portable and handheld counters 30-day operation Can shift or upgrade License type Generate Spreadsheets and Graphs of Data Generate reports based on Standards/Guidances ISO FS209E EU GMP Annex 1 57
58 PortAll 2.4 Life Sciences 21 CFR Part 11 Support Lockout for multiple attempts Auto Logoff for Inactivity Password Expiration Unique User Accounts Satisfy Corporate standards Length of User Name Length of Password Audit Trail of User actions 58
59 What s new in PortAll 2.4 Paperless Secure Data Transfer in a 21CFR Part 11 Environment USB Memory Stick data are now transferred in a binary format Supports 21CFR Part11 for data security and integrity Report Generation Just Got Easier Report wizard now fully supports EU GMP Annex 1 reporting Secure, Paperless Compliance Reports Export and archive electronic reports in a secure PDF format No Need to Bring Laptops into the Cleanroom Leverage your existing company network to transfer data Ethernet, wireless and serial communications networks fully supported 59
60 PortAll 2.4 What s new Support for USB Memory Stick transfer Binary format cannot be edited Only recent Raptor versions of
61 PortAll 2.4 What s new Crystal Reports engine Standard reports Custom by user Custom by Systems Group New! EU GMP Annex 1 report 61
62 PortAll 2.4 What s new Compliance reports stored in a locked PDF format Suitable for archive in LIMS or other data management system Fully supports 21 CFR 11 62
63 PortAll 2.4 New Data Sheet 63
64 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 64
65 EU Annex 1 Summary: Monitoring Section 12: The sample sizes taken for monitoring purposes using automated systems will usually be a function of the sampling rate of the system used. It is not necessary for the sample volume to be the same as that used for formal classification of clean rooms and clean air devices. It is not necessary to sample 1m 3 during verification or monitoring Particle counters used for monitoring may have the same or different flow rate from those used for classification. Note: Revision of EU Annex completed, effective date of 01 March
66 EU Annex 1 Summary: Monitoring Section 8 Clean rooms and clean air devices should be routinely monitored in operation and the monitoring locations based on a formal risk analysis study and the results obtained during the classification of rooms and/or clean devices Note: Revision of EU Annex completed, effective date of 01 March
67 EU Annex 1 Summary: Monitoring Section 9 The Grade A zone should be monitored at such a frequency and with suitable sample size that all interventions, transient events and any system deterioration would be captured and alarms triggered if alert limits are exceeded. = continuous!!! Note: Revision of EU Annex completed, effective date of 01 March
68 Monitoring Positions: Risk-based Approach Vial Sterilizing Tunnel Lyo Monitoring must follow the workflow, covering areas where product is exposed Annex 1 Where open vials exit de-pyrogenation human interaction (1) Where vials are filled (2,3) Surrounding Grade B background (4) Where the vials are partially stoppered (5) Loading area in front of lyophilizers must be Grade A if product is not fully stoppered (6,7) 6 Lyo 2 Lyo 3 68
69 Use Monitoring System for Classification Reduce the number of manually monitored classification points SAVE TIME! Add a wireless portable for room classification 69
70 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 70
71 Placement of Sample Probes Annex 1: No guidance FDA: Sample near to exposed product Generally near work height and exposed product If liquid sterile fill, guidance is to sample air approaching the product within 12 (30 cm) of exposed Sample near to points of intervention by operators Examples: Descrambler table Filling needles Stoppering process Less than 1 foot (12 inches or 30 cm) 71
72 Sampling Probes and Mounting Unidirectional air bathing the exposed product during manufacturing Isokinetic sample probe for particle monitoring (not positioned directly over exposed product) Exposed product or vials/ampoules 72
73 Where to Monitor? 73
74 Positioning the Probes Turntable Probe shown with Cap in place 74
75 Positioning the Probes Filling Line <305mm 75
76 Positioning the Probes Stoppering Station <305mm 76
77 Monitoring Powder Fill Section 9: For Grade A zones, particle monitoring should be undertaken for the full duration of critical processing, including equipment assembly, except where justified by contaminants in the process that would damage the particle counter or present a hazard, e. g. live organisms and radiological hazards. In such cases monitoring during routine equipment set up operations should be undertaken prior to exposure to the risk. Monitoring during simulated operations should also be performed. Note: Revision of EU Annex completed, effective date of 01 March
78 Interference from Product (Powder Filling) Unidirectional air bathing the exposed product during manufacturing Higher sample probe for monitoring during production Sample probe to demonstrate air quality before filling process Exposed product or vials/ampoules 78
79 Recommendations Powder Fill Line Inlet HEPA Filter Outfeed from tunnel = Position of monitoring during filling = Positions of monitoring at rest 79
80 Continuous Monitoring An FMS System can monitor: Airborne particulates, Room pressures and air flow Room temperature and humidity Other parameters such as door interlocks, etc System configurations available: Multiple Pumps: Multiple sampling positions, greatest flexibility/redundancy, but higher install cost Central Vacuum: Multiple sampling positions, good flexibility/redundancy and medium install cost Integrated Pumps: Multiple sampling positions, good flexibility/redundancy and lowest install cost 80
81 System Overview Multiple Pumps 230VAC Host RS485, ethernet Pump Controller With Flow sensor Pump Controller With Flow sensor Vacuum Pump Vacuum Pump 24Vdc PSU Ceiling void Computer with FMS Software Wall Plate Clean room Portable APC Remote APC in Enclosure WiFi capable 81
82 System Overview Central Vacuum 24 VDC Host RS485, ethernet Vacuum Ring Main Pump Control Host PC Smart Socket Controller, Flow sense and vacuum control valve Smart Socket Controller, Flow sense and vacuum control valve Central Vacuum Pumps Main /Standby Plant room Computer with FMS Software Clean room Wall Plate Portable APC Remote APC in Enclosure WiFi capable 82
83 230VAC System Overview Integrated Pumps Host RS485, ethernet Host PC 24VDC 6A PSU 24VDC 6A PSU 24VDC 1 A PSU Ceiling void Computer with FMS Software Clean room R2315 Portable APC Remote APC with integrated pump WiFi capable 83
84 Networking Ethernet FMS System Server Node QA Network Client View Node Production Area Prep area Filling Vial capping Address 0 Address 0 Address 0 Engineering Network Client View Node Address 1 Address 2 Address 3 Smart Sockets Connected via Local Ethernet or company wide Intranet services 84
85 Maintaining Compliance Early detection of out of compliance Walking window technique Previous results per cubic foot Newest results per cubic foot x x x x Sum over last 36 samples is 2 ALARM raised for 5.0 µm particle counts 36 rolling samples 85
86 Software Software is designed to be compliant with GAMP, Title 21CFR Part 11 and EU-GMP Annex 1 86
87 Human Machine Interface Plan views of Facilities animated to show alarm conditons from Autocad files. Password Access to all panels and system functions ensure security Menu screens can be configured in local language 87
88 Real Time Trending Trends provide multiple traces, multiple Y axes, real time and historic switching with alarm limits Cursor provides analysis features to read actual point values, min,max etc 88
89 Historic Trending Calendar date entry enables user to select start and stop date required Cursor allows trace values to be read, traces and alarm levels may be switched on/off. Up to six signature boxes for sign off. 89
90 Alarm Handling Alarms can be manually or automatically acknowledged and are time and date stamped on generation, acknowledgement and clearing down. 90
91 Alarm Reconciliation Alarm reconciliation provides users with detailed information concerning alarm transitions and in compliance and out of compliance operation 91
92 6000/7000 Series Designed to meet ISO requirements. Built for sterile processing environments with VHP-resistant optics, automatic flow control valve and NEMA-rate enclosure (7000) Built-in user notification of measurement alarms, flow problems, and sensor condition Multiple communication and mechanical installation options External alarm and environmental sensor capabilities reduce monitoring system complexity and costs Built-in memory and Hach OPC Server provide complete data security during network/computer problems Featuring Long Life Laser technology 92
93 Introduction to 6000 Series Multi-color status indicator LED Detail diagnostics by status indicator LED 0.2 µm minimum sensitivity Long Life Laser Models 6002, 6003, 6015 Standard 2 channel, optional 4 channel Built-in flow sensor option. Multiple I/O option Remote status indicator light option RH-Temp probe option Relative Humidity-Temperature probe connection Service port for instrument set-up Also used for remote indicator light stack 93
94 Introduction to 6000 Series Improved Sensitivity Flexible Communications Multiple Installation Options Enhanced Instrument Diagnostics Long Life Laser Diode Worldwide service and support 94
95 Improved Sensitivity and Long Life Laser Identify smaller particles reliably and accurately that could potentially improve product quality and yield with a sensitivity range of 0.2 to 5.0 μm 0.5 to 10.0 μm The Long Life Laser with improved MTTF reduces overall cost of ownership Design for Electronics application OR Life Science application (if sensor installed outside cleanroom) 95
96 Flexible Communications Save FMS integration costs, repair and inventory related costs with multiple communications Serial I/O (Pulse, RS-232, RS-485 with Modbus RTU or FX protocol) Ethernet Analog Wireless 5-Pin Phoenix connector and RJ-45 Ethernet I/O 10-Pin Phoenix connector Pulse, Serial RS232 Serial RS485 Modbus (RTU or FX) Analog 96
97 Installation Options - Connections Reduces re-installation time during routine calibration and maintenance with DIP switch option Easy Installation with multiple connection options Side vacuum connection Bottom vacuum connection DIP switch for RS485 instrument address setting Makes installation and removal easier Optional side vacuum connection Bottom vacuum connection 97
98 Installation Options - Mounting Various mounting options offer shorter installation and removal time DIN Rail (Standard) Mounting Option Utility Terminal Box Mounting Option Wall Plate Mounting Option 98
99 7000 Series 99
100 Introduction to 7000 Series ISO compliant design Multiple I/O option Built-in enhanced instrument status and user notification indicator Long Life Laser VHP tolerant flow path and optics Wall plate, umbilical cords and tubes with quick connect fittings for ease of installation and removal Up to 1000 sampling data storage 0.3 µm sensitivity, 0.1 or 1.0 cfm flow rate Built-in flow sensor option Remote status indicator light option Standard 2 channel, optional 4 channel Sealed enclosure with sealed connectors RH-Temp probe option Free set-up utility program Optional Relative Humidity- Temperature probe connection Multi-color status indicator LED Optional remote indicator light port Service port for instrument set-up Also can be used for optional remote display unit (future enhancement) 100
101 Flexible Communication Options Save FMS integration costs, repair and inventory related costs with multiple communications Serial I/O (RS-485 with Modbus RTU or FX protocol) Ethernet Analog Wireless RJ 45 sealed connector for Ethernet, RS485 serial or Analog I/O Makes integration to any FMS easier 101
102 Ease of Installation Features Easy installation with AC or DC input power options 100~230 VAC 50/60 Hz house power 24 VDC local power network Reduces re-installation time during routine calibration and maintenance with DIP switch feature for serial units Quick-connect type sealed vacuum connection Quick connect type vacuum connector For ease of installation DIP switch for setting address of RS485 serial I/O instrument Makes installation and removal easier AC or DC input power option Makes installation easier 102
103 Ease of Installation Features Wall plate for terminating end-user cables and tubing, reduces disruption to connectivity during routine calibration Umbilical with quick connect sealed connectors protect counter from wash down and make reinstallation easy Quick-connect type sealed vacuum connection makes reinstallation during routine calibration easy Wall Plate - Front Makes installation and removal of counter easier User terminates wiring, tubing at the rear end of the wall plate Makes installation easier Umbilical cords and tubing supplied to connect the counter with wall plate For ease of installation and subsequent removal for routine calibration and maintenance. Ensure sealed connections 103
104 Installation Features - Standard Accessories Reduces initial installation time Minimizes removal and re-installation time during routine calibration and maintenance Provides wash-down tolerance with sealed umbilical and tubing Item Number Description Counter 2 Mounting Plate 3, 4 Isokinetic Probe (0.1 or 1.0 cfm) 5 Wall Plate 6 Communication Cable with Sealed RJ- 45 Connectors 7 Power Cable AC or DC 8 Vacuum Tubing with Sealed Quick Connectors 104
105 Enhanced Instrument Diagnostics Enhanced diagnostics reduce trouble shooting time and downtime Built in or optional remote multi-color status indicator LED Instrument can be diagnosed for flow, sensor, communication failures, and count alarm or count alert through built-in or external light stack Steady Green = Normal (set by software or internal) Flashing Red = Count Alarm (set by software or internal) Flashing Yellow = Count Alert (set by software only) One short flash, one long flash Blue = Flow Fail (set by software or internal) Solid Blue = Sensor Fail (set by software or internal) Flashing Blue (2 Hz) = Communication Failure (internally set based on communication timeout) 105
106 Enhanced Instrument Diagnostics Enhanced instrument status indication reduce troubleshooting time and downtime Built in or optional remote multi-color status indicator LED Instrument can be diagnosed for flow, sensor, communication failures, and count alarm or count alert through built-in or external light indicator Steady Green = Normal (set by software or internal) Flashing Red = Count Alarm (set by software or internal) Flashing Yellow = Count Alert (set by software only) One short flash, one long flash Blue = Flow Fail (set by software or internal) Solid Blue = Sensor Fail (set by software or internal) Flashing Blue (2 Hz) = Communication Failure (internally set based on communication timeout) 106
107 107 Met One 6000/7000 Setup Utility Program Reduce set-up time with the Met One 6000/7000 setup utility program Local Setup LAN Setup Data Display Reduce trouble shooting time Run Met One 6000/7000 series remote counters without FMS software Obtain instrument information such as model, serial number and built in flow sensor
108 Local Setup Screen Capture 108
109 LAN Setup Screen Capture 109
110 Data Display Screen Capture 110
111 Environmental Sensors Measurement of Temp/RH, Differential Pressure, Air Flow and Temperature of Fridges, Freezers and Incubators Up to 1 km and 1000 channels Differential Pressure Sensor Temp/RH Sensor 111
112 Environmental Sensors DP Panels provide central location for signal connection for Temp/RH, PT100s, 4-20mA, DP tubes and digital I/O ( alarms etc.) 112
113 Alarm Indication Must provide alarm feedback e.g. via traffic light stack via alarm message display via volt free contact to BMS via network via pager/sms text 113
114 114 Alarm Indication Delivered where it is needed Red/Green lamps at each point of fill Alarm Message displays in area Computer screen in clean area
115 Alarm Message Display Alarm Indicators and Message Displays can be distributed around the facility 115
116 Alarm Message Display Alerts staff within facility that monitored parameters are out of limits Four line LCD display shows local parameters currently in alarm Built - in Sounder announces arrival of new alarm conditions Local Mute Button, silences sounder without leaving the room 116
117 Scalable Software Solutions For small FMS systems using only Met One 6000/7000 (best at < 20 counters), Pharmagraph envigil-lite provides an easy-to-implement solution Large FMS systems requiring: more than sensors, alarm messaging, remote client PCs, custom screens & maps, consolidated reports, batch features, etc. are best supported with custom integrated systems such as Hach UltraVision Online, Pharmagraph envigil FMS, Wonderware, Intellution, etc Hach s OPC Server enables flexibility for ALL software solutions 117
118 Low-cost, Simple envigil Lite 2 for Aseptic Production Monitoring Ethernet RS
119 Large, Customized, Batch-Driven UVO Systems for Aseptic Production Monitoring Client # 1 Fill Line A Client # 2 Fill Line B Ethernet RS-485 Server/client with database Fill Line A Ethernet or RS-485 Grade C & D monitoring Fill Line B 119
120 OPC Server for Aseptic Production Met One OPC Server Customer Software Real-time data & control Sensor Control & OPC Server Sensor network Sensors controlled via MODBUS / TCP (SCADA, UVO, etc.) Historical electronic data Database Data from sensors configured for online operation Paper data & PDF files Crystal Reports Data from sensors configured for offline portable operation Translator that converts data from all Met One sensors into the language of the customers preferred system. 120
121 Pharmaceutical FMS Installations Many hundreds since 1987 Every major pharmaceutical company on every continent 121
122 Summary Review your whole particle monitoring program relative to evolving regulations and standards Review the amount of manual, paper-based monitoring Use automation as an opportunity to improve compliance while simultaneously gaining efficiency Involve all internal stake holders Work with flexible suppliers using open architectures to enable maximum integration and preserve options for future expansion 122
123 Table: Sampling Strategies based on ISO and EU GMP Annex 1 Grade Classification Regular EM Program Process Monitoring A 1. Frequency = 6 months 2. Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: Annex 1 = 1 cubic meter 1. Frequency = a/r (daily????) 2. a/r; sometimes based on Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: a/r 4. best practice: (3) 1-minute counts 1. Continuous during process 2. Key high risk positions 3. No minimum volume defined B 1. Frequency = 6 months 2. Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: Annex 1 = based on ISO minimum 4. Min volume (5µm) = cubic feet (0.69 m3) 1. Frequency = a/r (daily????) 2. a/r; sometimes based on Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: a/r 4. best practice: (3) 1-minute counts 1. Frequent during process; continuous may be preferred 2. Key elevated risk positions 3. No minimum volume defined a/r = as required 123
124 Table: Sampling Strategies based on ISO and EU GMP Annex 1 Grade Classification Regular EM Program Process Monitoring C 1. Frequency = 12 months 2. Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: (5 µm) Annex 1 > ISO = 0.24 cubic foot 1. Frequency = a/r (weekly????) 2. a/r; sometimes based on Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: a/r 4.best practice: (3) 1-minute counts 1. no requirement 2. Low risk 3. No minimum volume defined D 1. Frequency = 12 months 2. Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: (5 µm) Annex 1 > ISO = cubic foot 1. Frequency = a/r (monthly????) 2. a/r; sometimes based on Minimum number of sample positions: ISO = SQRT area 3. Minimum sample volume: a/r 4.best practice: (3) 1-minute counts 1. no requirement 2. Low risk 3. No minimum volume defined a/r = as required 124
125 Regulations and Guidance for the Manufacture of Sterile Medicinal Products ISO and EU GMP Annex 1 125
126 Particle Count Theory and Calibration (ISO 21501) 126
127 Agenda Introduction to Cleanroom Classification of Air Cleanliness: ISO Guidance for the Manufacture of Sterile Medicinal Products EU GMP, Annex 1 Classification In-Process Monitoring Best Practices for FMS Solutions Particle Counter Theory and Calibration: IS On-line TOC Analyzer EP&USP Harmonization for Liquid Particle 127
128 Optical Particle Sensor Configuration 128
129 Sensor View Volume Inlet Tube The View Volume is The area where the air stream and the laser beam intersect Laser beam viewed in cross section Requires: Even illumination Precise alignment 129
130 Particle Sizing The larger the particle, the larger the corresponding output pulse from the sensor. 130
131 Counting Electronics +V Threshold Circuit This circuitry is duplicated from one channel up to six channels depending on the model counter Analog Signal in +V CH3 CH2 Digital signals to counting circuitry +V CH. 3 CH. 2 CH. 1 CH1 131
132 Air Particle Counter Calibration ISO and the link to GMP Regulators inspect to EU GMP, which calls up ISO14644 Next revision ISO14644 will refer to ISO ISO states Instruments that conform to this part of ISO are used for the classification of air cleanliness in cleanrooms and associated controlled environments in accordance with ISO
133 What is ISO 21501? ISO is a new family of standards describing the instruments and calibration requirements for determining particle size distribution using light interaction methods for both liquid and airborne particle counters. ISO represents the culmination of work by instrumentation manufacturers and industry users and comes at a critical time for the life science industry with the increasing trend for real-time air particle monitoring in cleanrooms using light scattering air particle counters. 133
134 What is ISO replacing? Previous calibration methods guidelines: ASTM F (2003) Standard Practice for Calibration of an Airborne Particle Counter Using Monodisperse Spherical Particles (withdrawn May 2007). IEST-RP-CC014.1 Calibration and Characterization of Optical Airborne Particle Counters (providing actual methods to perform the calibration). JIS B 9921: Japanese standard which comprehensively deals with OPC design performance, most notably in the area of counting efficiency. 134
135 ISO Additional Tests Before ISO Size calibration Size resolution False count rate Sampling time ISO Size calibration Verification of size setting Counting efficiency Size resolution False count rate Concentration limit Sampling flow rate Sampling time Sampling volume 135
136 Parameter ISO Requirements LS Liquid LE Liquid Airborne Size calibration +/- 2.5% +/- 2.5% +/- 2.5% Verification of size setting +/- 15% Cal curve (minimum = 3 points) +/- 10% Reference material +/- 10% Cal curve (minimum = 3 points) 50% Counting efficiency 50 +/- 30% N/A 50 +/- 20% 100% Counting efficiency 100 +/- 30% 1.5 to 3X minimum size 100 +/- 20% Reference material < 2X minimum size 100 +/- 10% 1.5 to 2X minimum size Size resolution 10% 10% 15% False count rate Particles per liter (minimum size) ISO Requirements N/A Comments Standard uncertainty of the mean size Median voltage Use 10% (As-received) Close to the minimum detectable size. Concentration < 25% of concentration limit. Use 1.5 to 2X minimum size Manufacturer's recommended particle size. Use area under the curve. >1.5X CC sets markers at 5% medium value (counts) Particles per cubic Poisson distribution with a 95% confidence limit meter (minimum size) On screen calculator Maximum particle concentration Particles per cubic centimeter Particles per cubic centimeter Particles per cubic 10% Coincidence loss (calculated) meter (minimum size) On screen calculator Sample flow rate MFR standard uncertainty MFR standard uncertainty +/- 5% Volumetric flow rate If unit has flow rate control system Include flow rate uncertainty (all) Sample time +/- 1% +/- 1% +/- 1% On screen calculator Sample volume +/- 5% +/- 5% N/A On screen calculator Response rate N/A N/A 0.5% 10 min at concentration limit (minimum size), Sample (<60 sec) then filter (10 sec), then sample (< 60 sec) 136
137 Calibrating Air Counters with PSL Spheres f Mean size Std Deviation size NIST-traceable standard particles have normal distribution 137
138 Particle Count Distribution Number of Particles millivolts (Particle Size ) 138
139 Resolution/Counting Efficiency (Accuracy) N Particle size 0.3 µm 0.5 µm Good resolution showing separation of channels 139
140 Resolution/Counting Efficiency (Accuracy) N Particle size 0.3 µm 0.5 µm Poor resolution leads to poor counting accuracy: undercounting in smaller sized channels 140
141 Counting Efficiency ISO At the counter minimum specified size Using PSL that is times the minimum specified size Counting efficiency must be 50% ±20% Counting efficiency must be 100% ±10% AND Channel 1 Channel 1 The counting efficiency for PSL particles of the minimum measurable particle size value marked in the specifications shall be within the range from 30% to 70% AND PSL particles of times minimum measurable particle size shall be 90% to 110% 141
142 Illumination Uniformity Affects Resolution Laser beam in cross section Particle = Ideal: Even illumination in the view volume provides the same pulse height regardless of where the particle passes through the beam = = Not ideal: Uneven illumination causes pulse height variance. High illumination intensity leads to high intensity peak, and low intensity illumination leads to low peak height 142
143 Effects of Flow Rate on Pulse Height The amplitude of the sensor output is a function of the particle s residence time in the view volume lpm flow rate At 28.3 LPM, residence time yields a strong response from 0.5µ particle 50 lpm flow rate As flow rate is increased, residence time decreases and response decreases 100 lpm flow rate At very high flow rates, performance can be challenging as the particles pass through the view volume very quickly and calibration is based on a weaker signal 143
144 Sensor View Volume Optics and Flow Small inlet/outlet Nozzle Good optical uniformity Low flow turbulence Large inlet/outlet Nozzle Poor optical uniformity Higher flow turbulence 144
145 145 Design Considerations Pump Type Image Advantages Disadvantages Roots (Hach Met One Patent) Blower High flow Low power Lightweight Best vacuum High flow Low power Lightweight Poor vacuum (requires larger inlet nozzle) Carbon Vane High flow High power Heavyweight Maintenance
146 Hach Ultra and ISO Designed Compliant Instruments: 3400 Series 6000 Series 7000 Series 146
147 HACH Calibration Software Core Cal 3 Ability to provide a premium ISO calibration to the customer resulting in increased service opportunity. Same procedures & software for all HACH distributors Equivalent to a Factory calibration 147
148 Core Cal 3 Features Continued Automatic counter detection Real-time graphing capability 148
149 149 Professional Documentation Professional Abobe PDF certificates Custom logo capable Incorporated into Hach calibration procedures
150 Competitor s 100 lpm PHA Curve (0.5 um) Poor optical performance leads to very poor resolution Poor signal-to-noise leads to poor differentiation between noise and real particle counts This particle counter does not meet the JIS 9921 or ISO standard for counting efficiency 150
151 Ultra High Flow Considerations (100 lpm) Impact on immediate environment and airflow patterns in critical areas Isokinetic probe must be matched to particle counter flow rate High exhaust rate of sampled air may disturb room air flow Disturbance of airflow patterns in restricted areas due to high rate of sampled air Modest air supply of most LAF cabinets may not be sufficient to support high flow rates without substantial disturbance Recommend smoke studies to evaluate potential disturbance of controlled air flow Accuracy of size resolution and counting efficiency Pulsations in high flow pumps cause inconsistent air flow within the counter resulting in poor size resolution performance Long term performance higher maintenance requirements 151
152 Sampling Probes and Mounting High-Flow Unidirectional air bathing the exposed product during manufacturing Air from outside Grade A zone REMEMBER: There is no requirement to sample 1m 3 of air when monitoring 152
153 Effect of Exhaust Air: Issues in using High-Flow No exhaust air flow REMEMBER: There is no requirement to sample 1m 3 of air when monitoring Exhaust air flow creates turbulence and worse 153
154 Met One 3400 Roots Pump Operating at 100 lpm (0.5 μm) Clearly identified peak, excellent resolution Exceeds ISO21501 calibration standard 154
155 155 ISO Calibration Standard Improved compliance Removes ambiguity by providing a single internationally recognized standard method for calibration Harmonization between ISO and GMP guidance Improved unit-to-unit reproducibility All particle counters are referenced to a gold standard Improved counting accuracy Reduces false alarms in Grade A areas All current Met One particle counters from Hach may be calibrated using ISO at your facility
156 Portfolio 156
157 Total Organic Carbon 157
158 Why? Regulation TOC tousp<643>, EP Conductivity to USP<645>, EP Suitable for WFI and PW Water contamination Biofilm 158
159 TOC increase and Biofilm 159
160 Correlation between endotoxines and TOC 160
161 On-line vs. conventional lab testing 161
162 Anatel Theory 162
163 USP <643> & EP Requirements Attribute USP <643> EP Technology to be used share the objective of completely oxidizing the organic molecules in an aliquot of sample water to carbon dioxide have in common the objective of completely oxidising the organic molecules in the sample water to produce carbon dioxide Apparatus on-line or off-line using a calibrated instrument Use a calibrated instrument on-line or offline has demonstrated acceptable system suitability shown to have acceptable system suitability. 163
164 164 USP <643> & EP Requirements Attribute USP <643> EP Limit of detection 0.05 mg of carbon per liter 0.05 mg of carbon per litre Blank and water (r w ) TOC Water high purity water as defined under <661> containers (conductivity measured just prior to dispensing, of not more than 0.15 µs/cm) TOC level of not more than 0.25 mg per liter Reagent Water not greater than 1.0 µs.cm -1. not greater than 0.1 mg/l.
165 165 USP <643> & EP Requirements Attribute USP <643> EP System Suitability frequency Standard Solution (r s ) System Suitability Solution (r ss ) Response Efficiency periodically demonstrated USP Sucrose USP 1,4-benzoquinone RE = 100[(r ss r w )/(r s r w ] not less than 85% and not more than 115% System suitability SOPs should include error trapping for systematic errors. at suitable intervals Reagent grade sucrose Reagent grade 1,4-benzoquinone RE = 100[(r ss r w )/(r s r w ] not less than 85% and not more than 115% System suitability SOPs should include error trapping for systematic errors. Limit response r s r w limit response (not necessarily 500 ppb) r s r w limit response (not necessarily 500 ppb)
166 USP <643> & EP Requirements Attribute USP <643> EP On-line locations reflect the quality of the water used. (changed from earlier draft specifying quality of the water at the use points ) representative of the water used. 166
167 Anatel Theory 167
168 Anatel Theory of Operation The organic compounds in each sample are completely oxidized to CO 2 in the presence of UV light and a TiO 2 catalyst Conductivity of the water sample is measured before and after oxidation The difference in conductivity is due to the conductive species produced from the dissolved CO 2 in water, and is the basis for calculating the amount of organic carbon in the sample The end point of the oxidation process is determined precisely and dynamically by a sophisticated set of algorithms Inorganic species that contribute conductivity are constant, and are compensated for in the background measurement 168
169 Principle of operation Organics oxidized to CO 2 Conductivity δ conductivity UV on Time 169
170 Product Portfolio Positioning Anatel TOC Pharmaceuticals Electronics PAT700 A1000XP A643a A
171 171 Anatel Performance Criteria Low detection limit of 1.0 ppb TOC and conductivity in a single sensor Wide sample temperature range Complete, rapid oxidation Multiple sensor network for data acquisition and control Analog, digital and serial data outputs Convenient system suitability Multi-point user calibration Easy validation check Confirm conductivity meter accuracy and cell constant Measure grab samples (for model A643 & PAT700)
172 A643 overview On Line TOC analyzer Available in 2 versions (Portable or Stationary) Calibration, Suitability & Validation tests pre-programmed on the transmetter Completed oxydation of the sample Grab sample feature 172
173 PAT
174 Using On-line TOC analyzers to Meet USP <643> TOC and <645> Conductivity Process piping Isolation valve Cell Constant can be verified without removing from the line Water system does not have to be re-sanitized Drain 174
175 PAT 700 Key features - Video On-line TOC analysis with complete sample oxidation Science-based, risk managing instrument designed to meet all requirements of USP, EP and JP OASIS TM Onboard, Automated Standards Introduction System Integral color touch screen display Multiple inputs/outputs including separate analogs for TOC, temperature and conductivity Dual UV lamps with UV Detect technology for improved reliability and diagnostics IP 56 stainless-steel enclosure improves protection from water and particulates 175
176 PAT700 TOC Analyzer with OASIS TM Onboard, Automated Standards Introduction System 176
177 PAT700 TOC Analyzer with OASIS TM Reduces operator intervention inserting/changing out bottles and entering data All bottles for a test installed at once and used in sequence -Reduces risk Information about each standard stored in RFID (Radio Frequency Identification) tag on the bottle (concentration, C of A value, lot number, expiration date ) -Information can be written to the bottle by the analyzer Analyzer tracks amount of standard used and number of reps Voids bottle following test - avoids reuse of bottles Sensor serial number on excursion bottles 177
178 OASIS TM Technology - RFID RFID - Radio Frequency Identification Store all information concerning standards 178
179 Bottle Mode Excursion Mode Empty sample bottle with RFID tag can be loaded for excursion sampling Analyzer automatically fills bottle with sample from water system under User configured TOC value USP conductivity 35 or 36 error conditions Manually Information on sample programmed to RFID tag on the bottle Date, time, serial number, previous TOC, previous conductivity and previous temperature Validation bottle can be run automatically to verify analyzer Selection of excursion with or without validation sample Reinforces PAT capability through process analysis and validation 179
180 UV Detect TM Diagnostics Direct measurement of UV source performance Diagnosis of UV when TOC readings are skeptical Fast, real-time lamp feedback On-line: Verified after every TOC analysis Off-line: Can be conducted in diagnostic mode after lamp replacement Provides confidence for on-line water release Can be used for rouge monitoring Supports PAT initiative through reduced risk More reliable than firmware or simple hours-of-operation counters Reduces cost-of-ownership 180
181 Integral Color touch screen Integral color touch screen display for ease of use and access to information, configuration and analyzer operation 181
182 182 PAT700 Portable or Stationary Optional handle sold separately Customer installed Recommended to be used with quick connect version of analyzer
183 reflect the quality of the water used Water treatment plant Purified Water Tank Points Of Use WFI treatment plant WFI Tank Points Of Use 183
184 184 Cleaning in Place Validation Typical Process: Rinse vessel Grab samples Perform laboratory with TOC HPLC Drawback Labor intensive Material intensive Slow process Vessel being cleaned unavailable.
185 185 Cleaning in Place Validation TOC & Conductivity Conductivity/TOC Process: Rinse vessel Monitor conductivity When conductivity reach a low level, start TOC Valid CIP after TOC reach low value Advantage PAT Quick answer Lab independent
186 186
187 Cleaning in Place Validation TOC & Conductivity Pre-rinse 1 & 2 : Low or ambient temperature. Aim is to remove majority of product residues. Wash 1: Use detergent (sodium hydroxide based). Aim is to clean the vessel. Increase of temperature in order to boos this process Post Rinse 1: Aim is to remove caustic residues Wash 2: Use detergent (citric acid). Aim is to remove caustic residues Post rinse 2: Aim is to remove acidic residues. High temperature Final Rinse: Ambient temperature, for accurate measurement. 187
188 Cleaning in Place Validation TOC & Conductivity Conductivity TOC Time Final Rinse Start TOC measure Stop CIP 188
189 Lab move to online TOC 189
190 Total Organic Carbon 190
191 Liquid 191
192 Effects of Particulate Contamination Focus has been on injectable liquids Possibility to occlude (block) capillaries and arteries Red Blood cells are about 5 µm Capillary (5 to 10 µm) Large veins (10 to 50 µm) Threat of microbial infection Possibility of reaction to foreign substances (RES/allergic reaction) 192
193 193 Global Regulations:Particles in Liquids Primary method Optical Particle Counter [OPC] Light Obscuration Counter Secondary method Optical microscope Subjective Labor intensive Requires more time to process samples
194 194 Volume Definitions: SVI, LVI SVI ex., Antibiotics, Insulin, Cytotoxic drug Small Volume Injectable Less than or egal to100 ml Also known as SVP [Small Volume Parenteral] LVI ex., NaCl solution ( IV bag ) Large Volume Injectable More than 100 ml Also known as LVP [Large Volume Parenteral]
195 Global Regulations: Particles in Liquids USP 32-NF 26 <1> Injections USP 32-NF 26 <788> Particulate Matter in Injections USP 32-NF 26 <789> Particulate Matter in Ophthalmic Solutions 195
196 Global Regulations: Particles in Liquids EP 6 (6.3) Parenteral Preparations EP 6 (6.3) - (2.9.19) Particulate Contamination: Sub-visible Particles 196
197 197 Harmonization of USP 788 with EP Current version USP 32 May 2009 Current version EP6 <788> Particulate Matter in Injection
198 Harmonization of USP 788 Almost a verbatim copy of EP Change in calibration management At least (4) sample aliquots to be taken USP 788 has previously required only (3) At least 25 ml pooled sample for SVI<25mL USP 788 has previously required only 20 ml Environment test changed Test at 10 microns only; no 25 micron limit now Exactly (5) sample draws of (5) ml At 10 microns, must be no more than 25 particles in 25 ml 198
199 Comparison of Compendial Requirements for Liquid Particles in Injections Feature USP 31-NF 26 (USP30-33(2) IRA) EP JP/KP Number of samples Size of aliquot 5 ml 5 ml 5 ml or portion Discard #1 Yes Yes Yes SVI: 10 um 6000 per container 6000 per container 6000 per container counts SVI: 25 um 600 per container 600 per container 600 per container counts LVI: 10 um 25 per ml 25 per ml 25 per ml counts LVI: 25 um 3 per ml 3 per ml 3 per ml counts Environment 25 per 25 ml 25 per 25 ml 5 per um 25 um No test No test 2 per 10 ml 199
200 Comparison of Compendial Requirements for Liquid Particles in Injections Container Type USP 29 <788> Volume <25 ml SVI/SVP Table 1 SVI 25 ml <= Volume <100 ml SVI/SVP Table 1 SVI Volume = 100 ml SVI/SVP Table 1 SVI 100 ml > Volume LVI/LVP Table 1 LVI EP USP 31- NF 26 Test 1.B Test 1.B Counts per container Test 1.B Test 1.B Counts per container Test 1.B Test 1.B Counts per container Test 1.A Test 1.A Counts per ml 200
201 Ophtalmic products USP <789> Title: Particulate Matter in Ophthalmic Solutions Based on methods and standardization tests of USP <788> Increases particulate matter limits for ophthalmic solutions 50 or fewer particles of 10 micron or larger 5 or fewer particles of 25 micron or larger Excludes these formulations from particulate testing Gels Suspensions Emulsions Medical devices EP By microscope only, and only for eye-drops: No more than 20 particles 25 microns and larger No more than 2 particles 50 microns and larger. No particles greater than 90 microns. [Refer to General section, 1163, Eye Preparations]. 201
202 Calibration & System Suitability USP <788> 31 and earlier Instrument Standardization IST - Sample Volume accuracy - Sample Flow Rate - Calibration (Moving Windows) - Sensor Resolution - Accuracy with USP Particle Count RS Do be performed every 6 months 202
203 Calibration & System Suitability What is the USP Particle Count RS? 2 bottles of blank water 2 bottles containing 10 and 15µm particles, with a well know 10µm particles concentration and a well know ratio of counts at 10µm to the counts at 15µm 203
204 Calibration & System Suitability 204
205 Calibration & System Suitability USP <788> 32 Harmonized with EP & JP - Calibration with Particle Sphere Latex (sizes between 10 and 25µm) - System Suitability with USP Particle Count RS No timeframe commonly every year 205
206 Calibration & System Suitability Hach recommendation Calibration + IST every 6 months (Sample Volume accuracy; Sample Flow Rate; Calibration; Sensor Resolution; Particle Counting Accuracy USP Particle Count RS) On a regular basis (1 to 3 months), system suitability with USP Particle Count RS. 206
207 PharmSpec 2.2 Software Added USB to connect 9705 to computer Supports 21 CFR Part 11 compliance Familiar environment minimizes learning curve and training time Network backup and archiving Designed for Windows 2000 and XP Create and control user-defined tests 207
208 PharmSpec 2.2 Software Supports Compendial standards from USP, EP, JP and KP for Injectables and Ophthalmics 208
209 PharmSpec 2.2 Software Support Standardization tests: -Moving Windows -Sensor resolution -Volume Accuracy -USP Counting accuracy -Flow rate -Electronic resolution 209
210 PharmSpec 2.2 Software Permits users to comply with FDA 21 CFR Part 11 - Electronic Records, Electronic Signatures 210
211 Update for PharmSpec Customer can download compendial update at no charge from hachultra.com website 211
212 Product Portfolio Positioning
213 9705 Features: Sides Handles Easily moved Sensor Cleanout Quick access for removing blockages Thermal printer Delivers data immediately Counts Averages Standard Deviation Software for computer PHARMSPEC Include all EP & USP procedure test User procedure builder IQ/OQ doc 213
214 Liquid 214
Regulations Concerning
Regulations Concerning Airborne Particle Counting Two types of activities performed with an airborne particle counter: Classification focuses primarily on the environment Particle Counting Applications
Workflow System for Paperless Air Monitoring Powered by MODA. Presented by
Introducing Hach s MET ONE Compliance Workflow System for Paperless Air Monitoring Powered by MODA Presented by Scott Hjelmervik Bob Toal Systems Director Vice President Hach Company MODA Technology Partners
Particle Monitoring Requirements in Pharmaceutical Cleanrooms
Particle Monitoring Requirements in Pharmaceutical Cleanrooms All drugs must be manufactured in accordance with the current Good Manufacturing Practice (cgmp) regulations. Pharmaceutical manufacturers
ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners
ISO 21501 A Standard Methodology to Optical Particle Counter Calibration and What It Means to Cleanroom Owners Tony Harrison and Bob Latimer ISO 21501- A Standard Methodology to Optical Particle Counter
EU GMP Annex 1 Update 2008: Airborne Particle Counting
EU GMP Annex 1 Update 2008: Airborne Particle Counting Airborne Particle Counting for Pharmaceutical Facilities: Update 2008, EU GMP Annex 1 Morgan Polen - VP of Applications Technology, Lighthouse Worldwide
Pharmagraph. GMP/GLP Monitoring System
Pharmagraph GMP/GLP Monitoring System Pharmagraph GMP/GLP Monitoring System Demonstrates GMP/GLP compliance for cleanrooms and laboratories Monitors, Alarms, Logs and Reports Airborne particle counts,
How To Control A Medical Device With A Pc System
Clean room monitoring and disinfection FMS 5.0, the Leading Facility Monitoring Software TSI FMS system is an advanced, reliable, and user-friendly monitoring software suite that features a truly open
GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS
PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 032-2 8 January 2010 RECOMMENDATION GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS UNDERSTANDING, ACCELERATED TSI SETS THE STANDARD FOR PARTICLE COUNTING Performance Semiconductor, hard disk drive, and flat panel manufacturers
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS APPLICATION NOTE CC-105 (US) Introduction This application note will demonstrate how real-time viable particle count data will
ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS
ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS PRESENTATION TO RACI 20 TH APRIL 2016 PAUL MORGAN DIRECTOR CRITICAL CERTIFICATION PTY LTD CHANGES TO ISO14644 PART 1 CHANGES
Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations
Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations B e c k m a n C o u l t e r L i f e S c i e n c e s 2 5 0 S. K r a e m e r B o u l e
Particle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences
Particle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences LIGHTHOUSE WORLDWIDE SOLUTIONS WWW.GOLIGHTHOUSE.COM Introduction The pharmaceutical and medical device industries are developing
Monitoring manufacturing, production and storage environments in the pharmaceutical industry
Application Description AD/RandC/005-EN Monitoring manufacturing, production and storage environments in the pharmaceutical industry - Provides independent verification and validation of the manufacture,
Environmental Monitoring
Environmental Monitoring Purpose of Environmental Monitoring Critical process within the pharmaceutical and biotechnology industries. Determines the microbial and particulate content of cleanroom air and
USB-500/600 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
Contamination Control and Environmental Monitoring Systems
Life Science Contamination Control and Environmental Monitoring Systems Hospital Electronics Brookhaven Instruments deliver, install and validate particle counters, Microbial Air Samplers and Monitoring
Wireless Particle Monitoring of Pharmaceutical Cleanrooms
Wireless Particle Monitoring of Pharmaceutical Cleanrooms Introduction Wireless technology has brought many innovative changes, enabling us to communicate in new ways. This article will explain how wireless
JNIOR. Overview. Get Connected. Get Results. JNIOR Model 310. JNIOR Model 312. JNIOR Model 314. JNIOR Model 410
The INTEG is an Ethernet I/O (digital, analog) device that monitors and controls a small set of process signals. functions as both basic I/O for integration with another application or system AND as a
USB-500 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
Indoor air quality monitoring made easy
Indoor air quality monitoring made easy with the Fluke 985 Particle Counter Application Note Introduction Energy savings is top of facilities owners agenda. HVAC system tends to be the largest contributor
site monitoring Kit Site Monitoring Kit User Manual we prove it.
site monitoring Kit 1 User Manual Site Monitoring Kit User Manual we prove it. 2 site monitoring Kit Content 1. Introduction 3 2. Content of the Site Monitoring Kit 4 3. Preparation 6 1. Positioning of
Environmental Monitoring of Clean Rooms
Environmental Monitoring of Raul Duarte President DDK Scientific, Corp. Copyright DDK Scientific, Corp. 2008, 2009, DDK Scientific, Corp. Proprietary A manufacturing facility for pharmaceutical products
Environmental Monitoring
Chapter 10 Environmental Monitoring By the end of this chapter, you will be able to: l List the types of environmental testing l Name three types of checks that are carried out to ensure the isolator is
Wireless Data Logger Monitoring Systems
Wireless Data Logger Monitoring Systems Table of Contents About MadgeTech, Inc. Wireless Systems & Components Wireless Data Loggers 2000A Wireless Data Logger Models MadgeTech Cloud Services MadgeTech
How To Use A Ds340 (Dsp)
Ideal for monitoring dust levels in the exhaust gas of industrial combustion or air filtration processes. Innovative Dynamic Detection Principle (DDP) measurement technique Immune to gradual reductions
Temperature & Humidity SMS Alert Controller
Temperature & Humidity Alert Controller METERS 3 simple steps starting the unit: Insert the SIM card Plug in the sensors connectors Connect the AC power cord. Specifications: AC 90~260V Auto Select Internal
4Sight Calibration Management Software
GE Measurement & Control Solutions 4Sight Calibration Management Software 4Sight calibration and maintenance management software provides visibility to the assets, data, and resources that affect maintenance,
Mitos P-Pump. product datasheet
product datasheet page Product description 2 Main benefits 3 Why choose the Mitos P-Pump? 3 Closed-loop flow control 4 Product specifications 5 Accessories - overview 6 Mitos P-Pump Basic 7 Mitos P-Pump
Smart. Scalable. Simple.
Thermo Scientific Wireless Monitoring Solution Smart. Scalable. Simple. providing the protection you need with the simplicity you deserve Thermo Scientific Wireless Monitoring Solution Optimized Sample
USB-500 Series Stand-Alone, Low-Cost Data Loggers
Stand-Alone, Low-Cost Data Loggers Features Stand-alone, remote data loggers Measure temperature, humidity, voltage, or current 1 or 2 channels Low cost, small size Battery powered Software Included, easy-to-use
OPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
International Standard ISO 14644
International Standard ISO 14644 Cleanrooms and associated controlled environments Praphon Angtrakool Food and Drug Administration 1 International Standard : ISO 14644 ISO 14644 consists of the following
Cleanroom. For. Sterile Manufacturing Facilities
Cleanroom For Sterile Manufacturing Facilities Praphon Angtrakool Food and Drug Administration 1 WHO TRS No. 823 Annex 1, 1992 (1) General 17.1 The production of sterile preparations should be carried
Compact Monitoring Technology your environmental monitoring solution
Compact Monitoring Technology your environmental monitoring solution Monitor your GxP environmental parameters and cut your validation costs. Solutions for Life Science and Hospitals www.siemens.com/cmt
450TOC Analyzer. Portable, Real-time, Continuous Total Organic Carbon Measurement. 450TOC Analyzer
450 Analyzer 450 Analyzer Fast, Continuous Measurement Compliant with USP, Ch P, EP and JP USB Data Logging and Printer Output Robust Portable Design Portable, Real-time, Continuous Total Organic Carbon
NEW. EVEN MORE data acquisition and test stand automation
NEW EVEN MORE data acquisition and test stand automation the new class of data The plug&play complete package User benefits Expert Series is the latest generation of data acquisition Complete hardware
NETWORK ENABLED EQUIPMENT MONITOR
NETWORK ENABLED EQUIPMENT MONITOR Remotely Monitor Sensors over the Internet Connect Sensors to the Web to Remotely Monitor Equipment, Processes or Other Applications A Complete, Easy to Deploy, Stand-Alone
PE6108/PE6208/PE8108/PE8208
Energy Intelligence Solutions PE6108/PE6208/PE8108/PE8208 eco PDU Distribution Unit ATEN has developed a new generation of green energy power distribution units (PDUs) to effectively increase the efficiency
Inova SV125 Machine Number 4016589
Inova SV125 Machine Number 4016589 Date of Manufacture: 2011 o Never installed or used; in original crate Configured for 10x10 nested 1.0mL Long syringe Fully automatic vacuum filling and vacuum stoppering
Temperature Alert. LAN-based Temperature and Humidity Data Logger
Temperature Alert LAN-based Temperature and Humidity Data Logger LAN-based Temperature and Humidity Monitoring - Supports Simultaneous WiFi and Hardwired Interfaces -20 C to +60 C (-4 F to +140 F) Standard
Short Form Catalogue. Alarm Systems. Reliable Supervision and Control
Short Form Catalogue Alarm Systems Reliable Supervision and Control Alarm Monitors and Indicators The SELCO product range includes a number of alarm monitors and indicator panels for use in numerous applications.
USB-500/600 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
SECTION 13850 DETECTION AND ALARM
SECTION 13850 DETECTION AND ALARM PART 1 GENERAL 1.01 SUMMARY A. Section Includes 1. Control Panel 2 Associated Equipment B. Products Installed But Not Supplied Under This Section 1. Section 16140 - Wiring
Monitoring the autoclaving process in the pharmaceutical industry
Application Description AD/RandC/006-EN Monitoring the autoclaving process in the pharmaceutical industry - Provides independent verification and validation monitoring of the autoclaving process - Enables
Streaming Potential System (SPT1000, SPD1000)
Streaming Potential System (SPT1000, SPD1000) User's Manual Sentrol Systems, Inc. 3949 Cotswold Dr. SW, Lilburn, GA 30047-2371 Tel: 770-564-1541, Fax: 770-564-8605 www.sentrolsystems.com Since the streaming
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
Power & Environmental Monitoring
Data Centre Monitoring Made Easy Power & Environmental Monitoring Features & Benefits Packet Power provides the easiest, most cost effective way to capture detailed power and temperature information for
CLASSIFICATION OF CLEANROOMS
CLASSIFICATION OF CLEANROOMS Cleanrooms are classified by the cleanliness of their air. The method most easily understood and most universally applied is the one suggested in the earlier versions (A to
Pharmaceutical Industry Whitepaper
Pharmaceutical Industry Whitepaper Cleanroom Data Management Challenges Labor and time intensive particle counting documentation can be streamlined Cleanroom Data Management Challenges Introduction Pharmaceutical
Emerson Smart Firewall
DeltaV TM Distributed Control System Product Data Sheet Emerson Smart Firewall The Emerson Smart Firewall protects the DeltaV system with an easy to use perimeter defense solution. Purpose built for easy
DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
TrueAlarm Fire Alarm Systems
TrueAlarm Systems UL, ULC, CSFM Listed; FM Approved* Network Annunciators; es Features Provides a dedicated local area network (LAN) for connection of a TrueSite workstation server to remote clients: es
Energy Monitoring Solutions
Energy Monitoring Solutions Energy Monitoring Solutions from The Industry Leader Every energy professional is faced with a difficult challenge; ensure maximum building performance while reducing energy
Application/Connection Examples
This Quick Start Guide is designed to familiarize the user with the connection and configuration of the DTS-305 DIN rail mounted single / 3 phase power & energy meter with RS-485 or TCP communications.
The Breather Box. High Performance Grade-D Breathing Air Filtration system. BB50-cO
The Breather Box High Performance Grade-D Breathing Air Filtration system Specifications Filter change Indicators Incoming Air Plug 1/2 Industrial Interchange 2 Automatic Drains with Discharge tubes 3
Mobile Data Power Model: MDP-25
Mobile Data Power Model: MDP-25 Topic Section Features... 2 Operational Features Summary... 2 Back-up Battery Power Internal Charger Voltage Spike Protection RF Noise Filtering Warning of Imminent Loss
LyoBeta. R+D Freeze-Dryers
LyoBeta R+D Freeze-Dryers LyoBeta series A Responsible Choice New demands in R+D centers are related to scale-up feasibility and quality of procedures and research methods. FDA investigators consider that
WEATHER STATION FOR SOLAR FARM MONITORING
WEATHER STATION FOR SOLAR FARM MONITORING SOLAR FARM MONITORING SYSTEM: Measures global, horizontal, & background irradiance. Measures wind speed, wind direction, ambient temperature, and relative humidity.
Laminar Flow Cabinets
Laminar Flow Cabinets The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. 24 36 48 Simple, Effective Protection for Samples and Work Processes, Uncompromised Performance, Economical
SMS Alarm Messenger. Setup Software Guide. SMSPro_Setup. Revision 090210 [Version 2.2]
SMS Alarm Messenger SMSPro_Setup Revision 090210 [Version 2.2] ~ 1 ~ Contents 1. How to setup SMS Alarm Messenger?... 3 2. Install the SMSPro_Setup software... 5 3. Connection Type... 6 4. Connection Port
FIRE ALARM SYSTEM. reliable fire detection and alarm signalling system
FIRE ALARM SYSTEM reliable fire detection and alarm signalling system Satel fire alarm devices allow to implement reliable and modern systems responsible for fire safety. A combination of 20-year experience
DREXELBROOK. USonic-R Series. Ultrasonic Level Measurement System. User Friendly
DREXELBROOK A Leader In Level Measurement Solutions USonic-R Series Ultrasonic Level Measurement System User Friendly Set the measurement range directly in inches, feet, millimeters, centimeters, or meters
LNG Monitoring. Fiber-Optic Leakage Detection System. Pipeline leakage detection. Regasification and liquefaction monitoring
LNG Monitoring Fiber-Optic Leakage Detection System Pipeline leakage detection Regasification and liquefaction monitoring Tank annulus and base slab monitoring Spill containment control Intelligent Solutions
S-series SQ Controller
DeltaV Distributed Control System Product Data Sheet December 2015 S-series SQ Controller Scalable controllers Quick assembly Easy-to-use Field proven architecture Designed for Electonic Marshalling Advanced
1.44 kw Programmable DC Power Supplies XLN Series
Data sheet 1.44 kw Programmable DC Power Supplies XLN Series New Family of High Density System Power Supplies The B&K Precision XLN series are compact, programmable, single-output DC power supplies, suitable
Qualification of an Environmental Monitoring Program
[ Scott Sutton Qualification of an Environmental Monitoring Program Scott Sutton Microbiology Topics discusses various topics in microbiology of practical use in validation and compliance. We intend this
Product Description Full Voltage Starting Electric Fire Pump Controllers FTA1000
Product Description Full Voltage Starting Electric Fire Pump Controllers FTA1000 Description Firetrol FTA1000 Full Voltage Fire Pump Controllers are intended for use with electric motor driven fi re pumps
Data Logging Solutions from Lascar Electronics
Data Logging Solutions from Lascar Electronics Remote Monitoring EL-BT-2 Bluetooth and Humidity Data Logging The EL-BT-2 data logger measures and stores over 1 million temperature and humidity readings.
Risk-Based Environmental Monitoring. Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI
Risk-Based Environmental Monitoring Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI Presenter Marsha Stabler Hardiman Over 20 years experience in the
Process Instrumentation
Process Instrumentation Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Process Gas Chromatography Features of Process Chromatographs Agilent 7890A
Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF
Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator
Current Monitoring Kit
Current Monitoring Kit QUICK START GUIDE DO090-6 CONTENTS Issues: 1) 2.10.02 WP A4 format 2) 2.10.02 Added safety warning 3) 17.3.06 Word A5 format. S1: Removed relay modules. S2: Added MI010. S4.1: added
Testing Automated Manufacturing Processes
Testing Automated Manufacturing Processes (PLC based architecture) 1 ❶ Introduction. ❷ Regulations. ❸ CSV Automated Manufacturing Systems. ❹ PLCs Validation Methodology / Approach. ❺ Testing. ❻ Controls
ViZion Installation Guide
ViZion Installation Guide v2.0 1 ViZion Installation Guide Table of Contents Inventory Hardware Setup Understanding Synchronization Cable Begin by taking an inventory of the required equipment DR Unit
Healthcare Solutions Room Pressure Monitors and Controls
Healthcare Solutions s and Controls TOUCHSCREEN ROOM PRESSURE MONITOR The Touchscreen Room Pressure Monitor (PMT) provides precise room pressure monitoring for critical spaces such as isolation rooms,
SCD Server. SCD Server Pro
SCD Server SCD Server Pro SCD Server & SCD Server Pro 9850-000387-01 - Page 1 of 8 SCD Server and SCD Server Pro The SCD Server is a 2U high DMX generator, running the ZerOS Operating System and emulating
User Manual Network connection and Mobics Dashboard (MIS) software for Dryer Controller M720
User Manual Network connection and Mobics Dashboard (MIS) software for Dryer Controller Manual version : v1.00 Networking and MIS Manual Dryer controller Page 1 of 16 Document history Preliminary version
New changes to cleanroom & clean air device classifications: ISO 14644 1 & 2
New changes to cleanroom & clean air device classifications: ISO 14644 1 & 2 This white paper describes the changes outlined in the Draft International Standard (DIS) editions of ISO 14644-1 and ISO 14644-2.
User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction
User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger
Uninterruptible Power Supply
96-01101 / rev. 2e / 2-2-12 Uninterruptible Power Supply EXCEPTIONAL SUPPORT & PROTECTION Uninterruptible Power Supply with Energy Saver design that is optimized to address the needs of A/V systems Features
Case Study - University at Buffalo State University of New York. Lower Energy Usage Costs with Real Time Monitoring, Campus-wide
Case Study - University at Buffalo State University of New York Lower Energy Usage Costs with Real Time Monitoring, Campus-wide Reduce Energy Consumption through Awareness of Energy Use Analyze Power Quality
High Power Programmable DC Power Supplies PVS Series
Data Sheet High Power Programmable DC Power Supplies The PVS10005, PVS60085, and PVS60085MR programmable DC power supplies offer clean output power up to 5.1 kw, excellent regulation, and fast transient
Introducing AUDIT-BUDDY
Introducing AUDIT-BUDDY Optimize the Data Center with AUDIT-BUDDY Executive Summary Proper temperature and humidity for the inlet air into the servers is essential to efficient data center operation. Several
VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING. F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group
VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group tel + 3226431600 E-Mail: [email protected] VALIDATION OF CLEAN
Infinity 2020 Perimeter Intrusion Detection System
Infinity 2020 Perimeter Intrusion Detection System User Guide: Network Application Your First Line of Defense 1 Infinity 2020 Perimeter Intrusion Detection System Table of Contents Section 1: Getting Started...
German engineering under high pressure
German engineering under high pressure Over 25 years of experience as a manufacturer of sterilizers, autoclaves and lyophilization systems for laboratories, production and research Production-scale autoclaves
Laboratory & Scientific Line. Bulk Steam Sterilizers. Since
Laboratory & Scientific Line Bulk Steam Sterilizers Since 1925 Bulk Steam Sterilizers for Laboratory & Scientific Applications The Tuttnauer line of bulk autoclaves are designed to handle the most critical
NGSS Data Consolidator (DC)
NGSS Data Consolidator (DC) The Data Consolidator (DC) facilitates system configuration, monitors network status, and supervises long-term data storage and remote data transmission. The DC consists of
D7050/D7050TH. Installation Instructions. Multiplex Photoelectric Smoke Detectors
D7050/D7050TH EN Installation Instructions Multiplex Photoelectric Smoke Detectors D7050/D7050TH Installation Instructions.0 Description Notices These instructions cover the installation of the D7050/D7050TH
WISE-4000 Series. WISE IoT Wireless I/O Modules
WISE-4000 Series WISE IoT Wireless I/O Modules Bring Everything into World of the IoT WISE IoT Ethernet I/O Architecture Public Cloud App Big Data New WISE DNA Data Center Smart Configure File-based Cloud
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2 Instruction Manual Firmware version: 2.0.1 Issue: 0.6 Version of the GPRS transmitters configurator: 1.3.6.3 Date of issue: 07.03.2012 TABLE OF CONTENTS
Laminar Flow Cabinets and Workstations
VERTICAL AND HORIZONTAL HEPA FILTERED SYSTEMS FOR PROCESS OR OPERATOR PROTECTION Laminar Flow Cabinets and Workstations For procedures requiring protection from particulate contamination and for aseptic
Process Alarm Solutions
Process Alarm Solutions Reliable Supervision and Control www.selco.com SELCO flexible alarm panels for supervision and control SELCO provides efficient and reliable solutions for alarm monitoring of electrical
REAL-TIME DUST AND AEROSOL MONITORING
REAL-TIME DUST AND AEROSOL MONITORING THE DUSTTRAK II AND DRX AEROSOL MONITORS UNDERSTANDING, ACCELERATED REAL-TIME DUST MONITORING. ANY ENVIRONMENT. ANY APPLICATION. The DustTrak II and DRX Aerosol Monitors
Owner's Manual [ S/N LABEL HERE ] Document # BW-BCU5DOC
Owner's Manual [ S/N LABEL HERE ] Document # BW-BCU5DOC BAYweb Remote Model BW-BCU5 Owner's Manual Copyright 2009-2010 Bay Controls, LLC Part Number: BW-BCU5DOC Revision: 0.2 April 18, 2010 BAYweb is a
Dash 18X / Dash 18 Data Acquisition Recorder
75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH
WIRELESS REMOTE MONITORING OF CATHODIC PROTECTION SYSTEMS. John Hawkyard MICorr Deputy General Manager Rawabi Corrosion Technology Co Ltd Al-Khobar
WIRELESS REMOTE MONITORING OF CATHODIC PROTECTION SYSTEMS John Hawkyard MICorr Deputy General Manager Rawabi Corrosion Technology Co Ltd Al-Khobar INTRODUCTION Cathodic Protection is an electrochemical
Product Guide AC Voltage Control Devices. Guide Guide. Popular Variable Transformers and AC Power Supplies. www.stacoenergy.com
Popular Variable Transformers and AC Power Supplies Product Guide AC Voltage Control Devices Guide Guide Your Tailored Power Solutions Provider TM Staco Energy Products Co. has been a leading manufacturer
